K Number
K241374

Validate with FDA (Live)

Date Cleared
2025-02-28

(289 days)

Product Code
Regulation Number
890.3850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Manual Wheelchair is to provide mobility to persons limited to a sitting position.

Device Description

The S002, S004, S005, S006, S007, S008, S009 series are mechanical wheelchairs which are a manually operated devices with wheels that are intended for medical purposes to provide mobility to persons restricted to a sitting position. It can be folded for transport by bring the two sides together. The manual wheelchair incorporates a main frame, a seat, two adjustable footrests and four wheels. The larger rear wheels have hand rims of slightly smaller diameter projecting just beyond the tire. These allows the user to manoeuvre the chair by pushing them on without requiring them to grasp the tires. The manual wheelchairs have brakes that bear on the tires of the rear wheels and two push handles at the upper rear of the frame to allow for manual propulsion by an assistant.

Main Components: Main frame, Backrest, Seat cushion, handgrip, front wheel, rear wheel, hand rim, crossbar, footrests, brake, Anti-tipper, Seat belt.

The device can be operated indoors, or outdoors on dry, level surfaces composed of concrete, blacktop, or asphalt under normal driving conditions.

AI/ML Overview

This FDA 510(k) summary describes a Manual Wheelchair (Models: S002, S004, S005, S006, S007, S008, S009) and its substantial equivalence to a predicate device (K232230). The information provided is for a physical medical device, not an AI/software device. Therefore, many of the requested categories related to AI/software performance criteria (such as sample sizes for test/training sets, data provenance, expert ground truth, adjudication methods, MRMC studies, or standalone algorithm performance) are not applicable to this document.

However, I can extract the acceptance criteria related to the physical device's performance and the study details that demonstrate it meets those criteria.

1. Table of acceptance criteria and the reported device performance:

Acceptance Criteria (Standard Compliance)Reported Device Performance
Mechanical Performance Standards:Compliance Demonstrated:
ISO 7176-1:2014 (Static stability)Complies
ISO 7176-3:2012 (Effectiveness of brakes)Complies
ISO 7176-5:2008 (Overall dimensions, mass and maneuvering space)Complies
ISO 7176-7:1998 (Measurement of seating and wheel dimensions)Complies
ISO 7176-8:2014 (Static, impact and fatigue strengths)Complies
ISO 7176-11:2012 (Test dummies)Complies
ISO 7176-13:1989 (Coefficient of friction of test surfaces)Complies
ISO 7176-15:1996 (Information disclosure, documentation and labeling)Complies
ISO 16840-10:2021 (Resistance to ignition of postural support devices)Complies
ISO 7176-22:2014 (Set-up procedures)Complies
Biocompatibility Standards:Compliance Demonstrated:
ISO 10993-1 (General biological evaluation)Complies
ISO 10993-5:2009 (Cytotoxicity)Non-cytotoxic
ISO 10993-10:2021 (Sensitization)Non-sensitizing
ISO 10993-23:2021 (Irritation)Non-irritating

2. Sample size used for the test set and the data provenance:

  • Sample size: Not specified in the document. For physical device performance testing, samples are typically a representative number of units, but specific numbers are not usually detailed in a 510(k) summary unless a specific statistical power is claimed for a particular test.
  • Data provenance: The tests were conducted by the applicant, Anhui JBH Medical Apparatus Co., Ltd, in China (No.116 QiCang Road Industrial District MingGuang Chuzhou,Anhui, China). The data is retrospective in the sense that the tests were performed on finished devices to demonstrate compliance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable. For physical device performance, "ground truth" is established through standardized testing procedures and objective measurements as defined by the referenced ISO standards, not through expert consensus on interpretation.

4. Adjudication method for the test set:

Not applicable. Physical device performance is determined by meeting objective criteria defined in the ISO standards, not by an adjudication process as used in clinical or imaging studies.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This is a physical medical device, not an AI/software device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This is a physical medical device, not an AI/software device.

7. The type of ground truth used:

The "ground truth" for the device's performance is established by its adherence to the technical specifications and performance limits defined within the cited international ISO standards (e.g., ISO 7176 series for wheelchairs and ISO 10993 series for biocompatibility). This is essentially objective measurement against predefined engineering and biocompatibility standards.

8. The sample size for the training set:

Not applicable. This is a physical medical device, not an AI/software device that requires a training set.

9. How the ground truth for the training set was established:

Not applicable. This is a physical medical device, not an AI/software device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health and Human Services logo is a stylized representation of a human figure, while the FDA acronym and full name are written in blue, with the acronym in a square.

February 28, 2025

Anhui JBH Medical Apparatus Co., Ltd % Jarvis Wu Senior Consultant Shanghai SUNGO Management Consulting Co., Ltd. 14th Floor, Dongfang Building, 1500# Century Ave. Shanghai, Shanghai China

Re: K241374

Trade/Device Name: Manual Wheelchair (Models: S002, S004, S005, S006, S007, S008, S009) Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I, reserved Product Code: IOR Dated: February 26, 2025 Received: February 26, 2025

Dear Jarvis Wu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Tushar Bansal -S

for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K241374

Device Name

Manual Wheelchair (Models: S002, S004, S005, S006, S007, S008, S009)

Indications for Use (Describe)

The Manual Wheelchair is to provide mobility to persons limited to a sitting position.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) Summary

Document Prepared Date: 2025/2/28

A. Applicant:

Name: Anhui JBH Medical Apparatus Co., Ltd Address: No.116 QiCang Road Industrial District MingGuang Chuzhou,Anhui, China

Contact Person: Wei yuan wei Tel: +86- 0550-8108866 Email: outreach@jbhmedical.com

Submission Correspondent: Primary contact: Mr. Jarvis Wu Title: Senior Consultant Shanghai SUNGO Management Consulting Co., Ltd. 14th Floor, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: jiawei.wu@sungoglobal.com

Secondary contact: Mr. Raymond Luo Title: Technical Director Shanghai SUNGO Management Consulting Co., Ltd. 14th Floor, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: zxfda(@sungoglobal.com

B. Device:

Trade Name: Manual Wheelchair (Models: S002, S004, S005, S006, S007, S008, S009) Common Name: Manual Wheelchair Models: S002, S004, S005, S006, S007, S008, S009

Regulatory Information

Classification Name: Wheelchair, Mechanical Classification: Class I Product code: IOR Regulation Number: 21 CFR 890.3850

Review Panel: Physical Medicine

C. Predicate device: K232230 YJ-K1/K2 Wheelchair Zhenjiang Assure Medical Equipment Co., Ltd

{5}------------------------------------------------

D. Device Description:

The S002, S004, S005, S006, S007, S008, S009 series are mechanical wheelchairs which are a manually operated devices with wheels that are intended for medical purposes to provide mobility to persons restricted to a sitting position. It can be folded for transport by bring the two sides together. The manual wheelchair incorporates a main frame, a seat, two adjustable footrests and four wheels. The larger rear wheels have hand rims of slightly smaller diameter projecting just beyond the tire. These allows the user to manoeuvre the chair by pushing them on without requiring them to grasp the tires. The manual wheelchairs have brakes that bear on the tires of the rear wheels and two push handles at the upper rear of the frame to allow for manual propulsion by an assistant.

Main Components: Main frame, Backrest, Seat cushion, handgrip, front wheel, rear wheel, hand rim, crossbar, footrests, brake, Anti-tipper, Seat belt.

The device can be operated indoors, or outdoors on dry, level surfaces composed of concrete, blacktop, or asphalt under normal driving conditions.

E. Indications for Use:

The Manual Wheelchair is to provide mobility to persons limited to a sitting position.

F. Comparison of Technological Characteristics with the Predicate Device Table 1 General Comparison

DeviceProposed DevicePredicate DeviceResults
510K NumberK241374K232230---
ManufacturerAnhui JBH Medical ApparatusCo., LtdZhenjiang Assure MedicalEquipment Co., Ltd.---
Proprietary NameManual WheelchairYJ-K1/K2 Wheelchair---
ModelS002, S004, S005, S006, S007,S008, S009 (7 models)24 models---
ClassificationIISame
Indications for useThe Manual Wheelchair isintended for medical purposesto provide mobility to personsrestricted to a sitting position.The YJ-K1/K2 ManualWheelchair is to providemobility to persons limited toa sitting position.Same
DesignCharacteristicManual Operation, Four-Wheels, Foldable, Cross-Brace, Quick release hub,Seatcushion ,Backrest, Footrests,Protective cover, Anti-tipper,brake.Manual Operation, Four-Wheels, Foldable, Cross-Brace, Pull-to- Lock, Armrest,Backrest, Foot rest, skirtguard , Anti-tipper (selectedmodels), brake extender(selected models)Similar
Brake controloccupant-operated brake onlyoccupant-operated brake onlySame
OperationFor indoor/outdoor useFor indoor/outdoor useSame
Environment
Control ModeMechanicalMechanicalSame
Size(unfold)S002: 815610765mmS004:810595810mmS005:810595790mmS006:815570765mmS007:815590765mmS008:815630765mmS009:815650765mm1220 (L) *660/640/700 (W) *857/908mm (H)Analysis #1
Stowagelength/width/heightS002: 815350765mmS004: 810595650mmS005: 810595620mmS006: 815350765mmS007: 815350765mmS008: 815350765mmS009: 815350765mm1220(L)*280/285/310(W)*857/90 8mm(H)Analysis #1
Weight(Total)12.5kg/14.0kg/14.5kg20.0kg/22.6kgAnalysis #1
Weight Capacity100kg136Kg(300lbs)Analysis #1
Seat Width380mm405/455/505mmAnalysis #1
Seat height510/520mm490mm/445mmAnalysis #1
Seat depth430mm16",18",20"Analysis #1
Back typeFixedFixedSame
TiresFront: 105mmRear:595mmFront: 193mmRear:610mmAnalysis #1
Foot restswing away footrest/ elevatingleg restswing away footrest/ elevatingleg restSame
Rear Axle PositionSingleSingleSame
FrameConstructionFoldable frame Push inwardfrom left and right sides to foldFoldable frame Push inwardfrom left and right sides to foldSame
Safety FeatureManual Wheel LockManual Wheel LockSame
Maximum SafeOperationalinclineVertical forward tilt ≥8°,vertical backward tilt ≥8°,and lateral tilt ≥8°Vertical forward tilt ≥10°,vertical backward tilt ≥10°,and lateral tilt ≥10°Analysis #2
braking timeThe manual wheelchair brakeis mechanical structure andmanual braking, and thebraking time can be ignoredwhen it is used immediately.The manual wheelchairbrake is mechanical structureand manual braking, and thebraking time can be ignoredwhen it is used immediately.Same
braking distancebrake is mechanical structureand manual braking, and thebraking distance can beignored.brake is mechanical structureand manual braking, and thebraking distance can beignored.Same
PerformanceComply with:ISO7176-1ISO7176-3ISO7176-5ISO7176-7ISO7176-8ISO7176-11ISO7176-13ISO7176-15ISO 16840-10Comply with:ISO7176-1ISO7176-3ISO7176-5ISO7176-7ISO7176-8ISO7176-11ISO7176-13ISO7176-15ISO7176-16Same
BiocompatibilityComply with:ISO10993-1ISO10993-5ISO10993-10ISO10993-23Comply with:ISO10993-1ISO10993-5ISO10993-10Same

{6}------------------------------------------------

{7}------------------------------------------------

Difference Analysis:

    1. Compare the predicate device, the subject device have different value on the unfold size, stowage size, device weight, Capacity, Seat Width, Seat height, Seat depth, Tires size. However the subject has pass the <ISO 7176-7-1998 Part7: Measurement of seating and wheel dimensions > and <ISO 7176-5-2008 Part 5: Determination of dimensions, mass and maneuvering space>, so the above different will not raise any new risk of safety or effectiveness.
    1. The Safe Operational incline is not identical, the subject device has passed the bench performance testing. So it will not raise any new risk of safety or effectiveness.

G. Summary of Non-Clinic Performance Testing

Performance Testing

Non-clinical tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • A ISO 7176-1:2014, Wheelchairs - Part 1: Determination of static stability
  • A ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes
  • ISO 7176-5: 2008 Wheelchairs Part 5: Determination of overall dimensions, mass and maneuvering A space
  • ISO 7176-7: 1998 Wheelchairs Part 7: Measurement of seating and wheel dimensions ISO A 7176-8: 2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strengths

{8}------------------------------------------------

No.116 QiCang Road Industrial District MingGuang Chuzhou,Anhui, China

  • ISO 7176-11: 2012 Wheelchairs Part 11: Test dummies V
  • ISO 7176-13: 1989 Wheelchairs Part 13: Determination of coefficient of friction of test surfaces V
  • ISO 7176-15: 1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling

  • ISO 16840-10:2021 Wheelchair seating- Part 10: Resistance to ignition of postural support devices > Requirements and test method
  • ISO 7176-22:2014 Wheelchairs Part 22: Set-up procedures >

Biocompatibility

Biocompatibility testing was conducted in accordance with the FDA guidance Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process. " Testing included:

  • Cytotoxicity (ISO 10993-5:2009)

  • ﺮ Sensitization (ISO 10993-10:2021)
  • Irritation (ISO 10993-23:2021)

The testing supports the biocompatibility of the patient-contacting device materials that were shown to be non-cytotoxic, non-irritating, and non-sensitizing.

H. Clinical Test Conclusion

No clinical study is included in this submission.

I. Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission, the Manual Wheelchair is as safe, as effective, and performs as well as the legally marketed predicate device cleared under K232230.

N/A