K Number
K241261

Validate with FDA (Live)

Date Cleared
2025-01-17

(266 days)

Product Code
Regulation Number
874.3450
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

INTENDED USE: The passive middle ear implant - tympanoplasty partial prosthesis is intended to be used for replacement of components of the ossicular chain, if at least the stapes footplate is present and functional. The tympanoplasty partial prosthesis is implanted in the middle ear to restore sound transmission from the tympanic membrane to the oval window by replacing the ossicles partially. The tympanoplasty prosthesis is a medical device for single use delivered in sterile condition.

INDICATIONS: The tympanoplasty prosthesis is indicated to treat patients with: - congenital or acquired ossicular chain defects (e.g., chronic otitis media, traumatic injury, malformation, cholesteatoma) - inadequate conductive hearing from previous middle ear surgery. The partial ossicular replacement prosthesis is indicated, if at least the stapes head and its stapes footplate are present and functional.

Tympanoplasty Sizers for partial prosthesis: INTENDED USE: The tympanoplasty Sizers are intended to be used during a partial ossicular replacement surgery to determine the required functional length of the MED-EL tympanoplasty prosthesis. The Tympanoplasty Sizers are optional, single-use surgical tools for intraoperative transient use only and delivered in sterile condition. The Sizers for partial prosthesis are intended to intraoperatively determine the distance between the head of the stapes and the tympanic membrane, thereby determining the required functional length of the MED-EL tympanoplasty prosthesis. INDICATIONS: The indications of the PMEI tympanoplasty partial prostheses apply.

Device Description

The MED-EL "Passive Middle Ear Implants" (PMEI) Tympanoplasty Partial Prostheses are partial ossicular replacement prostheses which restore the mechanical sound transmission to the oval window. They are offered in different designs and fixed or adjustable functional length to allow the surgeon to have more flexibility in terms of selecting the correct prosthesis according to the individual patient's anatomical needs, and the surgeon's surgical preferences and proficiency.

AI/ML Overview

The provided text describes the regulatory clearance for the MED-EL PMEI Tympanoplasty Partial Prostheses and includes details about non-clinical and clinical testing performed to demonstrate substantial equivalence to predicate devices. However, the document does not outline specific, quantitative acceptance criteria with corresponding "reported device performance" in a table format as typically requested for AI/ML device submissions. Instead, it discusses the outcomes of the clinical study in relation to the "state of the art" and provides general performance metrics.

Here's an attempt to extract and frame the information according to your request, with the understanding that specific acceptance criteria in the exact format you provided are not explicitly stated in this document for the device itself, but rather for the performance relative to existing treatments.

1. Table of Acceptance Criteria and Reported Device Performance

As specific quantitative "acceptance criteria" for the device's performance in a tabular format are not clearly defined in the provided document, I will infer them from the clinical outcomes that were deemed acceptable by the manufacturer for demonstrating substantial equivalence. The document primarily focuses on achieving outcomes comparable to the "state of the art."

Acceptance Criteria (Inferred)Reported Device Performance
Effectiveness:
% of patients with postoperative PTA4 Air-Bone Gap (ABG) ≤ 20 dB (comparable to state of the art, which is 62.5-71.3%)72% of patients with postoperative PTA4 ABG ≤ 20 dB
Mean BC-PTA4 (Bone Conduction Pure Tone Average) before surgery19.8 ± 12.2 dB HL
Mean BC-PTA4 (Bone Conduction Pure Tone Average) after surgery17.9 ± 12.2 dB HL
Rate of BC-PTA4 deterioration > 10 dB HL3.2% (6/186 patients)
Safety:
Rate of (recommended) revision surgeries5 (out of 202 patients)
Rate of (potential) device dislocations/extrusions3 (out of 202 patients)
Total Adverse Events (AEs)34 AEs in 21 patients

2. Sample size used for the test set and the data provenance

  • Sample Size for Audiological Outcomes: 186 patients (89 female, 97 male)
  • Sample Size for Adverse Events: 202 patients (98 female, 104 male)
  • Data Provenance: The study was a "large multicenter, retrospective, single-subjects repeated measures (each subject served as his or her own control) Post-Market Clinical Follow-Up (PMCF) investigation with the PMEI Tympanoplasty Partial Prostheses in Europe."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not mention the use of experts to establish "ground truth" in the context of image interpretation or similar diagnostic tasks. The ground truth for the clinical outcomes (audiological data, adverse events) was established by direct clinical measurements and follow-up, not by expert consensus on data interpretation. The "Intended User" is stated as "qualified ENT surgeons only, with adequate skills to perform otological surgeries," who would be responsible for the clinical procedures and likely the reporting of outcomes.

4. Adjudication method for the test set

Not applicable. The study involved direct clinical outcomes and adverse event reporting, not a process requiring adjudication of interpretations (e.g., of medical images) by multiple experts.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This document describes a medical device (a passive middle ear implant) with an associated clinical study, not an AI/ML-powered diagnostic or assistive tool involving human readers. Therefore, an MRMC study and effects on human reader performance are not relevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a physical implant, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth for the clinical study was outcomes data, specifically:

  • Postoperative audiological measurements (PTA4 ABG, BC-PTA4).
  • Reported adverse events, including device dislocations/extrusions and revision surgeries.

These were collected directly from patient follow-ups.

8. The sample size for the training set

Not applicable. This is a physical medical device, not an AI/ML model that requires a "training set."

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the date January 17, 2025. The month is January, the day is the 17th, and the year is 2025. The text is in a simple, sans-serif font. The date is written out in full, with the month spelled out and the day and year written numerically.

Image /page/0/Picture/9 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

MED-EL Elektromedizinische Geräte GmbH Inés Román Santiago Senior Regulatory Affairs Fürstenweg 77a Innsbruck, Tirol 6020 Austria

Re: K241261

Trade/Device Name: mCLIP Partial Prosthesis, mCLIP ARC Partial Prosthesis, mXACT Partial Prosthesis and mXACT PRO Partial Prosthesis Kit Regulation Number: 21 CFR 874.3450 Regulation Name: Partial Ossicular Replacement Prosthesis Regulatory Class: Class II Product Code: ETB Dated: December 18, 2024 Received: December 18, 2024

Dear Inés Román Santiago:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joyce C. Lin -S

for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K241261

Device Name

mCLIP Partial Prosthesis, mCLIP ARC Partial Prosthesis, mXACT Partial Prosthesis and mXACT PRO Partial Prosthesis Kit

Indications for Use (Describe)

INTENDED USE:

The passive middle ear implant - tympanoplasty partial prosthesis is intended to be used for replacement of components of the ossicular chain, if at least the stapes footplate is present and functional. The tympanted in the middle ear to restore sound transmission from the tympanic membrane to the oval window by replacing the ossicles partially. The tympanoplasty prosthesis is a medical device for single use delivered in sterile condition.

INTENDED USER

The tympanoplasty prosthesis is intended to be implanted by qualified ENT surgeons only, with adequate skills to perform otological surgeries. Replacement of the ossicular chain is a standard surgical procedure, no additional specific device training is mandatory or required for the safe and effective use.

TARGET PATIENT POPULATION

The target patient population for the passive middle ear implant are patients of all ages requiring reconstruction of the ossicular chain.

INDICATIONS:

The tympanoplasty prosthesis is indicated to treat patients with:

  • congenital or acquired ossicular chain defects (e.g., chronic otitis media, traumatic injury, malformation, cholesteatoma) - inadequate conductive hearing from previous middle ear surgery

The partial ossicular replacement prosthesis is indicated, if at least the stapes footplate are present and functional.

Tympanosplasty Sizers for partial prosthesis

(Note that the tympanosplasty Sizers are packaged together with the mXACT PRO Partial Prosthesis conforming the mXACT PRO Partial Prosthesis Kit).

INTENDED USE

The tympanoplasty Sizers are intended to be used during a partial ossicular replacement surgery to determine the required functional length of the MED-EL tympanoplasty prosthesis.

The tympanoplasty Sizers are optional, single-use surgical tools for intraoperative transient use only and delivered in sterile condition.

The Sizers for partial prosthesis are intended to intraoperatively determine the head of the stapes and the tympanic membrane, thereby determining the required functional length of the MED-EL tympanoplasty prosthesis.

INDICATIONS

The indications of the PMEI tympanoplasty partial prostheses apply.

{4}------------------------------------------------

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in bold, sans-serif font. To the right of the name is a stylized graphic of a cochlea, which is the part of the inner ear that is responsible for hearing. The logo is simple and modern, and it effectively communicates the company's focus on hearing solutions.

510(k) SUMMARY of Safety and Effectiveness MED-EL Elektromedizinische Geräte GmbH [As Required by 21 CFR 807.92(c)]

Submitter [807.92(a)(1)] 1.0

Manufacturer: MED-EL Elektromedizinische Geräte GmbH (hereafter MED-EL) Fürstenweg 77a, 6020 Innsbruck Austria

Contact Person:

Ines Roman Santiago MED-EL Regulatory Affairs Senior Specialist Phone: +43 577885786 E-Mail: ines.romansantiago@medel.com

FDA Official Correspondent: Elizabeth Gfoeller MED-EL Corporate Director, Regulatory Affairs Phone: +43 577885614 E-Mail: elizabeth.gfoeller@medel.com

Date the Summary was prepared: 17th January 2025.

2.0 Device Names [807.92(a)(2)]

Table 1 Device/Trade Names of MED-EL PMEIs Tympanoplasty Partial Prostheses

Device/TradeNameGeneric category& Medical SpecialtyClassificationProductcodeRegulation(CFR)
mCLIP PartialProsthesisMiddle ear, Prosthesis,Partial OssicularReplacement Prostheses(Ear, Nose & Throat)Class IIETB874.3450
mCLIP ARCPartialProsthesis
mXACT PartialProsthesis
mXACT PROPartialProsthesis KitMiddle ear, Prosthesis,Length -Adjustable andPartial OssicularReplacement Prostheses(Ear, Nose & Throat)

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in bold, sans-serif font. To the right of the "D" in MED is a circular graphic with curved lines, possibly representing sound waves or the inner ear.

Predicate Devices [807.92(a)(3)] 3.0

Table 2 Predicate devices manufactured by Heinz Kurz GmbH
MED-EL devicePrimary Predicate Device APredicate Device B
mXACT Partial ProsthesisN/A
mXACT PRO PartialProsthesis KitTTP-TuebingenType BELLPartialProsthesisTTP- VARIACSystem PartialK990923
mCLIP Partial ProsthesisK972492CLIP PartialProsthesisDresden TypeK013573
mCLIP ARC PartialProsthesisCliP PartialFlexiBALK013573

Description of the Devices [807.92(a)(4)] 4.0

The MED-EL "Passive Middle Ear Implants" (PMEI) Tympanoplasty Partial Prostheses are partial ossicular replacement prostheses which restore the mechanical sound transmission to the oval window.

They are offered in different designs and fixed or adjustable functional length to allow the surgeon to have more flexibility in terms of selecting the correct prosthesis according to the individual patient's anatomical needs, and the surgeon's surgical preferences and proficiency.

Table 3 Description of the MED-EL PMEI Tympanoplasty Partial Prostheses

Device/Trade NamemXACTPartial ProsthesismXACT PROPartial Prosthesis KitmCLIPPartial ProsthesismCLIP ARCPartial Prosthesis
DesignImage: mXACT Partial ProsthesisImage: mXACT PRO Partial Prosthesis KitImage: mCLIP Partial ProsthesisImage: mCLIP ARC Partial Prosthesis

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in bold, sans-serif font. A stylized graphic of an ear is placed between the "D" and "E" in MED-EL.

System/KitComponentsandAccessoriesN/ASystem/Kit content:1 mXACT PRO PartialProsthesis + 5 Sizers forpartial prosthesesN/A
CouplingstructureRound & "bell-shaped" for optimized stabilityon the stapes headUniversal clip for standardizedcoupling on the stapes head
HeadplateOval with optionalbendingpossibilityOval & length adjustable(through snap-mechanism)The surgeon can movethe "shaft" inside theheadplate to the desiredfunctional length,afterwards theconnection betweenheadplate and shaft isfixed through the snap-mechanism inside theheadplate. As a last stepthe overlapping titaniumshaft of the prosthesis iscut-off with a scalpel.Round with aball-jointconnection to theshaft.The additionalball-joint allowsthe prosthesis toadapt itself topostoperativetympanicmembranemovements (forexample: causedby the healingprocess).
HeadplateRound withoptional bendingpossibility
MaterialMedical gradetitaniummXACT PRO PartialProsthesis:Medical grade titaniumSizer for partialprosthesis: PolypropyleneMedical grade titanium
Sterile /Single-UseYes

Statement of the Intended use [807.92(a)(5)] 5.0

The intended use and indications of the MED-EL PMEI Tympanoplasty Partial Prostheses and of the predicate devices manufactured by Heinz Kurz GmbH are substantially equivalent.

Comparing the subject devices and the respective predicate devices, the same medical conditions or underlying pathologies can be treated, and the principal functionality of the devices is the same, hence no different questions of safety and effectiveness arise.

PMEI Tympanoplasty Partial Prostheses

Intended Use

The passive middle ear implant - tympanoplasty partial prosthesis is intended to be used for replacement of components of the ossicular chain, if at least the stapes footplate is present and functional. The tympanoplasty partial prosthesis is implanted in the middle ear to restore sound transmission from the tympanic membrane to the oval window by replacing the ossicles partially. The tympanoplasty prosthesis is a medical device for single use delivered in sterile condition.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for MED-EL. The logo is in red and features the company name in bold, sans-serif font. To the right of the name is a circular graphic with curved lines, also in red.

Intended User

The tympanoplasty prosthesis is intended to be implanted by qualified ENT surgeons only, with adequate skills to perform otological surgeries. Replacement of the ossicular chain is a standard surgical procedure, no additional specific device training is mandatory or required for the safe and effective use.

Target Patient Population

The target patient population for the passive middle ear implant are patients of all ages requiring reconstruction of the ossicular chain.

Indications

The tympanoplasty prosthesis is indicated to treat patients with:

  • congenital or acquired ossicular chain defects (e.g., chronic otitis media, traumatic injury, malformation, cholesteatoma).
  • inadequate conductive hearing from previous middle ear surgery.

The partial ossicular replacement prosthesis is indicated, if at least the stapes head and its stapes footplate are present and functional.

Tympanoplasty Sizers for partial prosthesis4

Intended Use

The tympanoplasty Sizers are intended to be used during a partial ossicular replacement surgery to determine the required functional length of the MED-EL tympanoplasty prosthesis. The Tympanoplasty Sizers are optional, single-use surgical tools for intraoperative transient use only and delivered in sterile condition.

The Sizers for partial prosthesis are intended to intraoperatively determine the distance between the head of the stapes and the tympanic membrane, thereby determining the required functional length of the MED-EL tympanoplasty prosthesis.

Indications

The indications of the PMEI tympanoplasty partial prostheses apply.

1 Note that the "italic text" in this section shall be used if the tympanoplasty Sizers are packaged together with the mXACT PRO Partial Prosthesis conforming the mXACT PRO Partial Prosthesis Kit. Traditional 510(k) PMEI TYM Partial

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in bold letters. To the right of the name is a stylized graphic that resembles an ear.

6.0 Comparison on Technological Characteristics with predicate devices [807.92(a)(6)]

The MED-EL PMEls have the same technological characteristics (i.e. design, material composition, etc.) as the predicate devices identified in paragraph |a||3). Hence, no difference in safety and effectiveness of the MED-EL PMET Tympanoplasty Partial Prostheses as compared to the predicate devices from Heinz CmbH is expected. A summary of the technological characteristics is provided in the following table.

6.1 PMEI Tympanoplasty PARTIAL Prostheses - Comparison with predicate devices

MED-ELDevice /PredicateDevice"mXACT PartialProsthesis"PredicateDevice [A]"TTP-TuebingenType BELLPartialProsthesis"(K972492)"mXACT PROPartialProsthesis Kit"Predicate Device[B]"TTP-VARIACSystem Partial"(K990923)"mCLIP PartialProsthesis"PredicateDevice [B]"CLIP PartialProsthesisDresden Type"(K013573)"mCLIP ARCPartialProsthesis"Predicate Device[B]"Clip PartialFlexiBAL"(K013573)
DesignImage: mXACT Partial ProsthesisImage: TTP-Tuebingen Type BELL Partial ProsthesisImage: mXACT PRO Partial Prosthesis KitImage: TTP-VARIAC System PartialImage: mCLIP Partial ProsthesisImage: CLIP Partial Prosthesis Dresden TypeImage: mCLIP ARC Partial ProsthesisImage: Clip Partial FlexiBAL
mXACT features a slit-headplatedesign for optional bendingpossibilitySystem/Kit content:1x Partial ossicular replacementprosthesis5 Sizers1 Sizer Disc(including 6mCLIP features a slit-headplatedesign for optional bendingpossibility and a clip with 8"legs" instead of 7Both devices have a ball-joint in theheadplate.mCLIP ARC has a round and centricheadplate
Sizers)
Design:
The mXACTheadplateworks with asnap-closingmechanismTTP-VARIACheadplate withplasticdeformation
AttachmentBell-shaped coupling structure for the placement onto the stapes headClip-shaped coupling structure for the placement onto the stapes head
1 length adjustable partialprosthesis
# of sizes1085 Sizers1 Sizer Disccontaining 6Sizers108108
Functionallengths [mm]0.75, 1.00,1.25, 1.50,1.75, 2.00,2.25, 2.50,3.00, 3.500.75, 1.00,1.25, 1.50,1.75, 2.00,2.25, 2.50 (*)length adjustable partialprosthesis can be adjusted to afunctional length from 0.75 to3.50Sizers functional length:1.0, 1.5, 2.0,2.5, 3.51.0, 1.5, 2.0,2.5, 3.0, 3.50.75, 1.00,1.25, 1.50,1.75, 2.00,2.25, 2.50,3.00, 3.500.75, 1.00,1.25, 1.50,1.75, 2.00,2.25, 2.50(*)0.75, 1.00, 1.25,1.50, 1.75, 2.00,2.25, 2.50, 3.00,3.500.75, 1.00, 1.25,1.50, 1.75, 2.00,2.25, 2.50 (*)
Shaft Ø [mm]0.20
Headplatedimensions[mm]Ø2.60 x 3.60 (oval and off-centric)Ø2.60 (round and centric)Ø2.60 (roundand centric)Ø2.60 x 3.60(oval and off-centric)
MaterialsMedical grade titanium ASTM F67Medical gradetitanium ASTMF67 (bell andshaft) &ASTMF136(headplate);Medical grade titanium ASTM F67Medical gradetitanium ASTMF67, except themicro-ball whichcan be out ofASTMF67 or ISO5832-2;Medical gradetitanium ASTMF67
BiocompatibleYes (EN ISO 10993)
Surgical ToolsN/ATitanium MicroN/A
(mandatory)Closing Forceps&Cutting Forceps
Surgical Tools(optional)TympanoplastySizers forpartialprosthesesACSizer DiscPartial.MalleusHandle. CavityBending pliers.TympanoplastySizers forpartialprostheses(included inthe Kit)ACSizer Disc Partial(included in thesales package).KURZ TitaniumTweezers.Micro ScissorsTympanoplastySizers forpartialprosthesesACSizer DiscPartialTympanoplastySizers for partialprosthesesACSizer Disc Partial
PackagingOne length variant per package;Single-Use; SterileOne length variable partialossicular replacement prosthesis +5 Sizers for partial prostheses.Single-Use; SterileOne length variant per package; Single-Use; Sterile
MRIMRI Conditional at 1.5, 3.0 and 7.0 T

Table 4: PMEI Tympanoplasty Partial Prostheses - Comparison of Technical Characteristics

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in a bold, sans-serif font. The "O" in MED-EL is stylized with a circular design, adding a unique touch to the logo.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in bold, sans-serif font. To the right of the "D" in MED, there is a stylized graphic of a cochlea, which is the part of the inner ear responsible for hearing. The logo is simple and modern, and it effectively communicates the company's focus on hearing solutions.

  • The predicate device uses a different type of lent in the finctional length of the device (functional length is the distance between stages head and the tympanic nembrane), whereas for the overal length is specified as part of the mentioned 510k. According to the predicate device (see https://www.kurzmed.com/), the overall length of each variant is the functional length plus 1mm.

Comparing the MED-EL PMEI Tympanoplasty Prostheses and their identified predicate devices manufactured by Heinz Kurz GmbH, the, biocompatibility, packaging configuration and MRI compatibility are IDENTICAL.

The following features are considered EQUIVALENT: device design, number and dimensions of length varials in body contact. As there is no difference in technological characteristics, there is no difference in safety and/or effectiveness.

There is just one difference: the surgical tools.

Sizers that can be used optionally determine the required length of the prosthess to be used are available for both, the subject devices as well as for the identified predicate devices.

The MED-EL tympanoplasty Sizers are either available separately (Class ) or packaged together with the mXACT PRO Partial Proshess conforming the mXACT PRO Partial Prosthess, for the predicate devices, also additional optional tools (Malleus Handle Cavity Bending pliers, KURZ Titanium Tweezers, and Micro Scissors) are prosthesis length, the TTP-VARIAC System Partial (Heinz Kurz GmbH) requires the surgeon to use special tools for closing the headplate of the MED-EL mXACT PRO Partial Prosthesis can be closed with e.g. standard delicate tweezers. As this is considered an improvement, no difference in safety and effectiveness is expected. The safety and effectiveness of the MED-EL mXACT PRO Partial Prosthesis headplate was confirmed in non-clinical testing.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in bold, sans-serif font. To the right of the name is a stylized graphic of an ear, also in red. The logo is simple and modern, and the use of red conveys a sense of energy and innovation.

7.0 Non-clinical Test [807.92(b)(1)]

Non-clinical testing was conducted to assess the performance data for a determination of substantial equivalence.

The following tests were conducted:

  • . MRI environment according to ASTM F2119, F2052, F2182 с MRI Conditional at 1.5T, 3.0T and 7.0T
  • . Biocompatibility according to EN ISO 10993
    • The performed evaluation provides objective evidence to support the o conclusion that the devices can be considered biocompatible for their intended use.
  • Shelf-Life testing according to EN ISO 11607
  • Sterilization Validation according to EN ISO 11137-1, EN ISO 11137-2. ●
    • Gamma Sterilization with a confidence sterility assurance level of <10 o
  • . Packaging Validation according EN ISO 11607
  • . Mechanical & other testing

Safety and effectiveness have been demonstrated with non-clinical testing, concluding that the design and performance requirements are met.

8.0 Clinical Test [807.92(b)(2)]

The benefit for patients of the PMEI Tympanoplasty Partial Prostheses has been successfully demonstrated and the clinical outcomes of the MED-EL PMEL Tympanoplasty Partial Prostheses are comparable to the predicate devices as per the state of the art for these types of devices.

The goal of ossicular replacement prosthesis is to close the postoperative pure tone average (PTA4) air-bone gap (ABG) to within 20 dB. According to the state of the art, as assessed in several systematic literature reviews and meta-analysis, 62.5-71.3 % of patients implanted with a partial prosthesis, achieve a postoperative PTA« ABG ≤20 dB.

MED-EL conducted a large multicenter, retrospective, single-subjects repeated measures (each subject served as his or her own control) Post-Market Clinical Follow-Up (PMCF) investigation with the PMEI Tympanoplasty Partial Prostheses in Europe:

  • Audiological outcomes:
    • O Number of patients: 186 (89 female, 97 male)
    • Age: mean 43.9 ± 18.6 years (range 5-81) o
    • Implanted ear: 95 right, 91 left o
    • Follow-up: median 70.0 days (mean 78.4 ± 55.3; range 3-302 days) O
    • Prosthesis type: 48 mXACT Partial, 15 mXACT PRO Partial, 68 mCLIP o Partial, 55 mCLIP ARC Partial
    • % patients with a postoperative PTA4 ABG ≤20 dB: 72 % O
      • Mean BC-PTA« pre/post-surgery: 19.8 ± 12.2 dB HL / 17.9 ± 12.2 dB HL
    • BC-PTA4 deterioration >10 dB HL: 6/186 patients = 3.2 % (range 11.3- O 23.5 dB HL)
  • . Adverse events:

O

  • Number of analyzed patients: 202 (98 female, 104 male) o

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for MED-EL, a company that specializes in hearing implants. The logo is in red and features the company name in a bold, sans-serif font. To the right of the name is a circular graphic with curved lines, possibly representing sound waves or the inner ear. The logo is simple, modern, and easily recognizable.

  • Age: mean 44.5 ± 18.7 vears (range 5-84) o
  • Implanted ear: 107 right, 95 left o
  • Follow-up: median 275.0 days (mean 284.0 ± 176.9; range 0-666 days) o
  • Prosthesis type: 52 mXACT Partial, 16 mXACT PRO Partial, 72 mCLIP o Partial, 62 mCLIP ARC Partial
  • Number of (recommended) revision surgeries: 5 O
  • Number of (potential) device dislocations/extrusions: 3 o
  • o Total 34 AEs in 21 patients: 7 Device unrelated AEs, 4 Medical related AEs, 15 Clinical/surgical related AEs, 8 AEs "unknown", 2 Follow-up AE

In the above mentioned PMCF investigation the safety and effectiveness of the PMEI Tympanoplasty Partial Prostheses was confirmed as demonstrated by postoperative PTA4 ABG ≤20 dB rates comparable to the state of the art, low rates of device dislocations/extrusions and revision surgeries, and overall stable bone conduction thresholds, an indicator that residual inner ear hearing is not compromised after implantation with an ossicular replacement prosthesis. Safety and effectiveness of the MED-EL PMEI Tympanoplasty Partial Prostheses are comparable to the current state of the art.

9.0 Conclusion [807.92(b)(3)]

The results of the non-clinical testing demonstrate that:

  • the MED-EL PMEI Tympanoplasty Partial Prostheses meet the established . specifications, the design and performance requirements to ensure safety and effectiveness according to the intended use.
  • . there is no difference in safety and/or effectiveness between MED-EL PMEI Tympanoplasty Partial Prostheses and the predicate devices manufactured by Heinz Kurz GmbH.

It can be concluded that the PMEI Tympanoplasty Partial Prostheses are substantially equivalent to the predicate devices manufactured by Heinz Kurz GmbH.

§ 874.3450 Partial ossicular replacement prosthesis.

(a)
Identification. A partial ossicular replacement prosthesis is a device intended to be implanted for the functional reconstruction of segments of the ossicular chain and facilitates the conduction of sound wave from the tympanic membrane to the inner ear. The device is made of materials such as stainless steel, tantalum, polytetrafluoroethylene, polyethylene, polytetrafluoroethylene with carbon fibers composite, absorbable gelatin material, porous polyethylene, or from a combination of these materials.(b)
Classification. Class II.