K Number
K241169

Validate with FDA (Live)

Date Cleared
2024-11-22

(210 days)

Product Code
Regulation Number
876.5980
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Entarik NI Feeding Tubes (FTs) are intended for the administration, fluids and medications by the nasoenteric or orogastric route for neonatal patients who have an intact gastrointestinal tract but are physically unable to manage nutritional intake through normal mastication and deglutition.

The Entarik NI Feeding Tube System (Entarik NI System) is designed to aid, in conjunction with institutional protocols, qualified operators in the placement of the Entarik NI Feeding Tube (FT) into the stomach of neonatal patients requiring enteral feeding. The Entarik FTs are equipped with sensors designed to provide information about the tube tip relative to the stomach, thus assisting in reducing the incidence of misplacement during first positioning. The Entarik NI FTs are provided sterile.

The Entarik NI System also monitors the feeding tube position continuously during the course of feeding and automatically and in real time alerts of tube migration.

The Entarik NI System provides continuous monitoring of gastric and esophageal temperature. The Entarik device can be used solely for the purpose of monitoring gastric and esophageal temperature in situations where invasive monitoring is indicated.

The Entarik NI system is intended to record, store, view and analyze impedance in the pharynx and stomach.

Device Description

The Entarik NI Feeding Tube System (Entarik NI System) consists of the Entarik NI Feeding Tube (disposable) and the Entarik NI Monitor (reusable) with Power Supply. The Entarik NI System is designed for enteral feeding and to aid in the placement of the Entarik NI Enteral Feeding Tube into the stomach of neonatal patients requiring enteral feeding. The Entarik NI System also continuously monitors feeding tube position, gastric and esophageal temperature, and gastrointestinal impedance and reflux. The feeding tubes are single use devices intended for prolonged use (less than 30 days). The Entarik NI Feeding Tubes are provided sterile.

AI/ML Overview

The provided text describes the acceptance criteria and study proving the device meets those criteria for the Entarik NI Feeding Tube System.

1. Table of Acceptance Criteria and Reported Device Performance

The document does not present a formal table of acceptance criteria with corresponding performance metrics in a pass/fail format. Instead, it describes various tests performed and states that the device was found to be "safe and effective". We can infer the "acceptance criteria" from the fact that no issues were raised during these tests, indicating successful outcomes for all mentioned evaluations.

Acceptance Criteria TypeReported Device Performance
SterilizationValidation performed per ISO 11135-1 and ISO 10993-7.
BiocompatibilityEvaluation performed per ISO 10993-1:2018 and FDA guidance. Leveraged existing tests from primary predicate due to identical materials and manufacturing processes (Cytotoxicity, Intracutaneous Irritation, Maximization Sensitization, Acute Systemic Toxicity, Sub-acute Systemic Toxicity, Intramuscular Implantation, Material Mediated Pyrogenicity were successfully demonstrated).
Software QualityVerification and validation testing performed per IEC 62304:2015, demonstrating safety and performance.
Electrical, Mechanical, and Thermal Safety, and EMCDesigned to comply with AAMI/ANSI ES60601-1, IEC 60601-1-6, IEC 60601-1-8, IEC 80601-2-49, ISO 80601-2-56, and IEC 60601-1-2.
Device Performance (Bench)Bench tests performed: Dimensional Verification, Impedance and Temperature Accuracy and Sensing Rate, Insertion and Withdrawal, Flow and Pressure, Buckle and Kink, Tensile Integrity, Viscous Feed Testing, Radiopacity, Product and Label Durability. All tests implicitly passed as no failures are indicated.
Clinical PerformanceIn the validation study, the system was "safe and effective in all eleven neonatal participants to aid qualified operators in the placement of the Entarik FT into the stomach of patients, and monitor the feeding tube position continuously and automatically and in real-time alert of tube migration."

2. Sample Sizes and Data Provenance

  • Training Set Sample Size: 16 subjects.
  • Test Set Sample Size: 11 subjects for the validation study.
  • Data Provenance: The document does not explicitly state the country of origin. The studies appear to be prospective clinical studies, as they involved the placement and monitoring of the device in new participates for the purpose of the study.

3. Number of Experts and Qualifications for Ground Truth

The document does not explicitly state the number of experts used or their specific qualifications (e.g., "radiologist with 10 years of experience") for establishing ground truth. It mentions that the validation study's placement confirmation was "per institutional standard of care," which implies reliance on qualified medical professionals within the clinical setting.

4. Adjudication Method

The document does not describe a specific adjudication method (e.g., 2+1, 3+1, none) for the test set. It mentions "placement was confirmed per institutional standard of care," suggesting that the medical institution's standard procedures were followed for verifying tube placement.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

There is no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study or any effect size of how much human readers improve with AI vs. without AI assistance. The study focuses on the device's ability to aid operators and monitor position, not on comparing human reader performance with and without AI.

6. Standalone (Algorithm Only) Performance

The document describes two clinical studies. The first (observational study) collected data silently for algorithm training and optimization without providing placement guidance. This suggests an initial phase where the algorithm was developed and refined, but standalone performance against a ground truth is not explicitly reported in terms of metrics like sensitivity/specificity for tube placement. The second (validation study) involved the device aiding qualified operators, indicating a human-in-the-loop scenario rather than a purely standalone algorithm evaluation of performance directly supporting a diagnostic claim.

7. Type of Ground Truth Used

  • For the validation study's tube placement: "placement was confirmed per institutional standard of care." This likely refers to methods such as X-ray confirmation, pH testing of aspirate, or visual confirmation, as standard practice in clinical settings for feeding tube placement. While not explicitly stated, this would be a clinical ground truth established by medical professionals following established protocols.
  • For continuous monitoring of tip location: The device itself automatically and in real time alerts of tube migration. The validation study confirmed this functionality was "safe and effective," implying these alerts were indeed accurate relative to actual tube migration conditions (though the specific ground truth for judging "tube migration" during continuous monitoring is not detailed).

8. Sample Size for the Training Set

The training set for the algorithm involved 16 subjects.

9. How the Ground Truth for the Training Set was Established

In the observational study for algorithm training and optimization, the Entarik NI Monitor "did not provide any placement guidance or verification. It collected impedance and temperature data silently that was used to train and optimize the placement and monitoring algorithm." This implies that while the device collected its data, the true tube position (ground truth) during this training phase was independently established by standard clinical methods (e.g., X-ray) by the clinicians or institution, and this independent ground truth was then used to train the algorithm. The document doesn't explicitly state the method for ground truth establishment during the training phase, but it must have been an independent reference standard.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 22, 2024

Gravitas Medical, Inc. % Lucie Dalet Principal Ram+ 2251 San Diego Ave. Suite B-257 San Diego, California 92110

Re: K241169

Trade/Device Name: Entarik NI Feeding Tube System Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal Tube (And Accessories) Regulatory Class: Class II Product Code: KNT, FPD Dated: October 16, 2024 Received: October 16, 2024

Dear Lucie Dalet:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

{2}------------------------------------------------

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Anthony Lee -S

Anthony Lee, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K241169

Device Name Entarik NI Feeding Tube System

Indications for Use (Describe)

Entarik NI Feeding Tubes (FTs) are intended for the administration, fluids and medications by the nasoenteric or orogastric route for neonatal patients who have an intact gastrointestinal tract but are physically unable to manage nutritional intake through normal mastication and deglutition.

The Entarik NI Feeding Tube System (Entarik NI System) is designed to aid, in conjunction with institutional protocols, qualified operators in the placement of the Entarik NI Feeding Tube(FT) into the stomach of neonatal patients requiring enteral feeding. The Entarik FTs are equipped with sensors designed to provide information about the tube tip relative to the stomach, thus assisting in reducing the incidence of misplacement during first positioning. The Entarik NI FTs are provided sterile.

The Entarik NI System also monitors the feeding tube position continuously during the course of feeding and automatically and in real time alerts of tube migration.

The Entarik NI System provides continuous monitoring of gastric and esophageal temperature. The Entarik device can be used solely for the purpose of monitoring gastric and esophageal temperature in situations where invasive monitoring is indicated.

The Entarik NI system is intended to record, store, view and analyze impedance in the pharynx and stomach.

Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Gravitas Medical. The logo features a green and blue graphic on the left, resembling a stylized stomach. To the right of the graphic, the words "GRAVITAS MEDICAL" are written in blue, with the tagline "Enteral Nutrition Intelligence" underneath in a smaller font.

510(k) SUMMARY K241169

DATE PREPARED

November 21, 2024

MANUFACTURER AND 510(k) OWNER

Gravitas Medical, Inc. 2332 Fourth St., Suite E, Berkeley, CA 94710, USA Telephone: Official Contact: Saheel Sutaria, Chief Technical Officer

REPRESENTATIVE/CONSULTANT

Lucie Dalet, PhD. Alexia Haralambous, MS, Allison Komiyama, Ph.D. ROM+ Telephone: +1 877-652-0830 Email: ldalet@rqmplus.com, aharalambous@rqmplus.com, akomiyama@rqmplus.com

DEVICE INFORMATION

Proprietary Name/Trade Name:Entarik NI Feeding Tube System
Common Name:Tubes, Gastrointestinal (And Accessories)
Regulation Number:21 CFR 876.5980
Class:Class II
Product CodesKNT, FPD
Premarket Review:OHT 3 / DHT3A
Review Panel:Gastroenterology/Urology

PREDICATE DEVICE IDENTIFICATION

The Entarik NI Feeding Tube System is substantially equivalent to the following predicates:

510(k) NumberPredicate Device Name / ManufacturerPrimary Predicate
K230206Entarik Feeding Tube System / Gravitas Medical Inc.
K202539Nutriglide™ Nasal Feeding Tube / Applied MedicalTechnology, Inc.
K131590InnerSense Esophageal Temperature Probe, FeedingTube / Philips Medical Systems

The predicate devices have not been subject to a design-related recall.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the text "K241169" and "Page 2 of 5" at the top. Below the text is the Gravitas Medical logo, which includes a stylized image of a stomach in green and blue. Next to the image is the text "GRAVITAS MEDICAL" in blue, with the words "Enteral Nutrition Intelligence" in a smaller font below.

DEVICE DESCRIPTION

The Entarik NI Feeding Tube System (Entarik NI System) consists of the Entarik NI Feeding Tube (disposable) and the Entarik NI Monitor (reusable) with Power Supply. The Entarik NI System is designed for enteral feeding and to aid in the placement of the Entarik NI Enteral Feeding Tube into the stomach of neonatal patients requiring enteral feeding. The Entarik NI System also continuously monitors feeding tube position, gastric and esophageal temperature, and gastrointestinal impedance and reflux. The feeding tubes are single use devices intended for prolonged use (less than 30 days). The Entarik NI Feeding Tubes are provided sterile.

INDICATIONS FOR USE

Entarik NI Feeding Tubes (FTs) are intended for the administration of nutrition, fluids and medications by the nasoenteric or orogastric route for neonatal patients who have an intact gastrointestinal tract but are physically unable to manage nutritional intake through normal mastication and deglutition.

The Entarik NI Feeding Tube System (Entarik NI System) is designed to aid, in conjunction with institutional protocols, qualified operators in the placement of the Entarik NI Feeding Tube (FT) into the stomach of neonatal patients requiring enteral feeding. The Entarik FTs are equipped with sensors designed to provide information about the location of the tube tip relative to the stomach, thus assisting in reducing the incidence of misplacement during first positioning. The Entarik NI FTs are provided sterile.

The Entarik NI System also monitors the feeding tube position continuously during the course of feeding and automatically and in real time alerts of tube migration.

The Entarik NI System provides continuous monitoring of gastric and esophageal temperature. The Entarik device can be used solely for the purpose of monitoring gastric and esophageal temperature in situations where invasive monitoring is indicated.

The Entarik NI System is intended to record, store, view and analyze impedance in the pharynx, esophagus and stomach.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

Gravitas Medical, Inc. believes that the Entarik NI Feeding Tube System is substantially equivalent to the predicate devices based on the information summarized here:

The subject device has the same intended use as the primary predicate cleared in K230206 and similar Indications for Use to the primary and secondary predicates. The subject device has a similar design and uses identical materials as the primary predicate device cleared in K230206. The main differences with the primary predicate device include changes in intended patient population, modifications of the Feeding Tubes, and software modifications. The subject device is intended for use in neonates whereas the primary predicate device is intended for adults. As a result, changes were made to the Feeding Tubes dimensions and electrodes number and spacing to adapt the device to the anatomy of neonates, similar to the predicate device cleared in

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Gravitas Medical Enteral Nutrition Intelligence. Above the logo is the text "K241169" and "Page 3 of 5". The logo is a blue and green graphic with the text "GRAVITAS MEDICAL" in blue and "Enteral Nutrition Intelligence" in a smaller font below in blue.

K202539 for neonatal, pediatric, and adult patients. Software modifications include changes in calculation methods to adjust to the anatomical specificities of neonates, and the addition of temperature and impedance monitoring functionalities, which are similar to the functionalities of the predicate device cleared in K131590, All these technological characteristics have undergone testing to ensure the device is as safe and effective as the predicates.

SUMMARY OF NON-CLINICAL TESTING

The following tests were performed to demonstrate substantial equivalence based on current industry standards.

Sterilization

Sterilization validation was performed in accordance with the following standards:

  • Sterilization validation per ISO 11135-1 Sterilization of health care products Ethylene oxide - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices
  • Ethylene oxide sterilization residuals per ISO 10993-7 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals

Biocompatibility

Biocompatibility evaluation was conducted according to ISO 10993-1:2018 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process and FDA's Biocompatibility Guidance. Since the materials and manufacturing processes of the subject device are identical to the primary predicate device, testing conducted on the predicate was leveraged to demonstrate biocompatibility of the subject device. Leveraged testing included:

  • -Cytotoxicity Study Using the ISO Elution Method
  • ISO Intracutaneous Irritation Test -
  • ISO Maximization Sensitization Test -
  • -ISO Acute Systemic Toxicity Study
  • ISO Sub-acute Systemic Toxicity Study -
  • ISO Intramuscular Implantation Test -
  • ISO Material Mediated Pyrogenicity -

Software

Software verification and validation testing was performed per IEC 62304:2015 Medical device software - Software life cycle processes to demonstrate safety and performance based on current industry standards.

Electrical, Mechanical, and Thermal Safety, and Electromagnetic Compatibility

The Entarik NI Feeding Tube System has been designed to comply with the following standards covering electrical, mechanical, and thermal safety:

  • AAMI/ANSI ES60601-1 Medical electrical equipment Part 1: General requirements for basic safety and essential performance.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Gravitas Medical, which includes a blue and green graphic of a stomach. Next to the graphic is the text "GRAVITAS MEDICAL" in blue, with the words "Enteral Nutrition Intelligence" in a smaller font below. Above the logo is the text "K241169" and "Page 4 of 5".

  • IEC 60601-1-6 Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usabilitv
  • -IEC 60601-1-8 Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical svstems
  • IEC 80601-2-49 Medical electrical equipment Part 2-49: Particular requirements for the basic safety and essential performance of multifunction patient monitoring equipment
  • -ISO 80601-2-56 Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
  • -IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests

Performance Testing

The following bench tests were performed on the Entarik NI Feeding Tube and Monitor (where applicable):

  • -Dimensional Verification
  • Impedance and Temperature Accuracy and Sensing Rate -
  • Insertion and Withdrawal -
  • -Flow and Pressure
  • Buckle and Kink -
  • -Tensile Integrity
  • Viscous Feed Testing -
  • Radiopacity -
  • -Product and Label Durability

SUMMARY OF CLINICAL TESTING

Two clinical studies were conducted with the subject device.

  • Observational study for algorithm training and optimization in 16 subjects ranging from -30 to 39 weeks of gestational age. The Entarik Feeding Tube (FT) was placed in all participants while connected to the Entarik NI Monitor. The Entarik NI Monitor did not provide any placement guidance or verification. It collected impedance and temperature data silently that was used to train and optimize the placement and monitoring algorithm for this population of patients.
  • Validation study conducted on 11 subjects ranging from 32 to 43 weeks of gestational age to assess the efficacy and safety of the system on the neonatal population. The Entarik FT was placed under the guidance of the Entarik NI Monitor, and placement was confirmed per institutional standard of care. The Entarik NI Monitor stayed connected to the Entarik FT for a target of 72 hours in each patient to continuously monitor tip location, as well as provide nutrition and medication to the patients. The study demonstrated that the Entarik NI System was safe and effective in all eleven neonatal participants to aid qualified operators in the placement of the Entarik FT into the stomach

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the text "K241169" at the top, followed by "Page 5 of 5". Below that is the logo for Gravitas Medical, which includes a stylized image of a stomach in green and blue, along with the words "GRAVITAS MEDICAL" and "Enteral Nutrition Intelligence".

of patients, and monitor the feeding tube position continuously and automatically and in real-time alert of tube migration.

CONCLUSION

The Entarik NI Feeding Tube System is substantially equivalent to the primary predicate device cleared in K230206, with the changes being the intended patient population, dimensions changes of the Feeding Tubes, and the addition of temperature and impedance monitoring. The intended patient population is similar to the secondary predicate cleared in K202539 and the added functionalities are substantially equivalent to the secondary predicate cleared in K131590. Based on the testing performed, including software verification and validation testing, electrical, mechanical, and thermal safety testing, electromagnetic compatibility (EMC) testing, bench performance testing, and clinical testing, it can be concluded that the subject device does not raise different questions of safety or effectiveness compared to the predicate devices. The intended use, patient population, technological characteristics, and performance characteristics of the subject device are considered substantially equivalent to the predicate devices.

§ 876.5980 Gastrointestinal tube and accessories.

(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.