K Number
K241155

Validate with FDA (Live)

Date Cleared
2024-09-05

(133 days)

Product Code
Regulation Number
878.4860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cold Sore Device is indicated for shortening the time to healing of herpes simplex labialis lesions on or around the lips with time to healing defined as the time to patient described re-epithelialization.

Device Description

The Cold Sore Device is a solid state opto-electronic device that emits a controlled quantity of 1072nm +/- 12nm peak wavelength near infrared light for a period of approximately 3 minutes. The maximum peak light intensity across the treatment surface is 20mW/cm². The light output and duration are monitored by a microprocessor. The tip of the device that contacts the patient is made out of Acrylonitrile Butadiene Styrene (ABS) + PC. Treatment with the device is commenced at the first symptoms of a cold sore 3 times a day with 4 hours in between each treatment for 2 consecutive days. The treatment area is approximately 3 cm². There are 2 light emitting diodes (LEDs) in the treatment area. The light within the device is activated by opening the cover of the device and automatically shut down after the pre-programmed treatment time (3 minutes). The device is designed for external, limited duration skin contact in an environment free from fluids and is provided non-sterile. The LEDs do not come in direct contact with the patient based upon the design of the device. The device is for OTC use and single patient use as described in the patient and box labeling.

AI/ML Overview

The provided text is a 510(k) summary for the Cold Sore Device (QPZ-03) and does not contain the specific information requested about acceptance criteria and a study proving device performance in the context of clinical efficacy (shortening time to healing).

Instead, the summary focuses on demonstrating substantial equivalence to a predicate device (QPZ-01) based on non-clinical tests, physical characteristics, and intended use. The core argument for equivalence is that the subject device's technical parameters (wavelength, treatment time, etc.) are "highly consistent" with the predicate device, which presumably had clinical data supporting its efficacy.

Therefore, many of the requested fields cannot be filled directly from this document. I will highlight what information is available and what is explicitly stated as not performed or applicable.

Here's a breakdown based on the provided text:


Acceptance Criteria and Device Performance Study (as described in the document):

The document does not detail specific acceptance criteria for clinical effectiveness (e.g., a target reduction in healing time) that the new device directly met through a clinical study. Instead, it relies on demonstrating substantial equivalence to a predicate device which implies the predicate device had met such criteria. The reported "performance" for the subject device is largely in terms of its technical specifications aligning with the predicate.

1. Table of Acceptance Criteria and Reported Device Performance

Criterion TypeSpecific Criterion (if stated)Reported Device Performance / Status
Clinical EfficacyShortening time to healing of herpes simplex labialis lesions (implied from predicate's performance)"highly consistent with the predicate device (K22205)" in intended use and technical parameters. No direct clinical study on the subject device's efficacy is presented.
Non-Clinical SafetyConformance to various IEC/ANSI/ISO standards (e.g., electrical safety, EMC, photobiological safety)Non-clinical tests were performed to assure conformance. (Specific results not provided, but presumed to meet standards for clearance).
BiocompatibilityConformance to ISO 10993 series (cytotoxicity, irritation, skin sensitization)Biocompatibility tests performed and conformance stated. (Specific results not provided, but presumed to meet standards for clearance).
Software ValidationConformance to FDA's 2023 Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions."Software verification and validation testing conducted. Documentation provided. Software documentation level considered "basic level" as failure could result in minor injury. (Specific results not provided).
Human Factors/UsabilityAbility of users to make correct self-selection, operate the device, and comprehend labeling.All 45 subjects made correct self-selection (3 found unsuitable). All 42 remaining subjects performed usability operation correctly. All 42 subjects correctly understood label information.
Technical Parameters (vs. Predicate)Wavelength: 1072nm +/- 12nm, Energy Source: LED, Treatment Schedule: 3X/day, 4 hours between treatments, for 2 consecutive days, Auto-off feature: Yes, Treatment time: 3-minute, Treatment area: 3 cm², Energy density: 20 mw/cm²Subject device matches all these parameters with the predicate device.

2. Sample size used for the test set and the data provenance

  • Usability Study:

    • Sample Size: 45 subjects for self-selection, reduced to 42 subjects for usability and label comprehension.
    • Data Provenance: Not explicitly stated, but implies a prospective study conducted by the manufacturer for regulatory submission. Country of origin not specified, but the manufacturer is based in China.
  • For Clinical Efficacy (if a separate study existed): Not applicable, as the document explicitly states "So, there is no clinical test on our device" (referring to the subject device for clinical efficacy). Efficacy is established through substantial equivalence to the predicate.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • For Usability Study: The document mentions "researchers" and "investigator recorded the operation of the subject" but does not detail the number or qualifications of experts involved in objectively establishing "ground truth" (e.g., whether a correct operation was performed). It appears the researchers/investigators themselves made these judgments based on whether the instructions were followed.
  • For Clinical Efficacy: Not applicable, as no clinical study was performed on the subject device.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • For Usability Study: No formal adjudication method like "2+1" or "3+1" is described. The "researchers" and "investigators" are implied to have made direct observations and judgments.
  • For Clinical Efficacy: Not applicable, as no clinical study was performed on the subject device.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is a light-based treatment device, not an AI-assisted diagnostic tool involving "human readers."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is a physical treatment device, not an algorithm. Its operation is standalone in the sense that once activated, it delivers light therapy for a set duration, but this does not equate to "algorithm only performance" in the context of diagnostic AI.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • For Usability Study: The "ground truth" was whether subjects could correctly follow instructions for self-selection, device operation, and label comprehension, as judged by the "researchers/investigators."
  • For Clinical Efficacy: Not applicable for the subject device. For the predicate device, it is implied that the clinical ground truth for "shortening the time to healing of herpes simplex labialis lesions" would have been patient-described re-epithelialization or similar clinical outcomes data.

8. The sample size for the training set

  • Not applicable. This document describes a medical device, not a machine learning algorithm that requires a training set in that sense.

9. How the ground truth for the training set was established

  • Not applicable. See point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 5, 2024

Shenzhen Nuon Medical Equipment Co., Ltd Alain Dijkstra Regulatory Affairs Engineer 1F-3F, No.27-2, Xintang Rd, Xintian Comm, Fuhai Str, Baoan Dist Shenzhen, Guangdong 518000 China

Re: K241155

Trade/Device Name: Cold Sore Device (QPZ-03) Regulation Number: 21 CFR 878.4860 Regulation Name: Light Based Energy Source For Topical Application Regulatory Class: Class II Product Code: OKJ Dated: April 7, 2024 Received: April 25, 2024

Dear Alain Dijkstra:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product: and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by TANISHA TANISHA L. HITHE -S L.HITHE - ○ Date: 2024.09.05
19:19:05 -04'00'

Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K241155

Device Name

Cold Sore Device (QPZ-03)

Indications for Use (Describe)

The Cold Sore Device is indicated for shortening the time to healing of herpes simplex labialis lesions on or around the lips with time to healing defined as the time to patient described reepithelialization.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K241155

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 878.4810.

1. Submitter's Information

Sponsor Name: Shenzhen Nuon Medical Equipment Co., Ltd Address:1st Floor-3rd Floor, No. 27-2, Xintang Road, Xintian Community, Fuhai Street, Baoan District, Shenzhen, Guangdong, China Contact Person (including title): Alain Dijkstra (CEO) Establishment Registration Number: 3030541658 Tel: +86-82129361 Fax: +86-755-25024651 E-mail: alaindijkstra@nuonmedical.com

Application Correspondent:

Contact Person: Alain Dijkstra Sponsor Name: Shenzhen Nuon Medical Equipment Co., Ltd Address: 1st Floor, 3rd Floor, No. 27-2, Xintang Road, Xintian Community, Fuhai Street, Baoan District, Shenzhen, Guangdong, China Tel: +86 755 82129361 Fax: +86 755 25024651 Email: Annahe@nuonmedical.com

2. Subject Device Information

Classification Name: Light based energy source device for topical application Trade Name: Cold Sore Device (Model: QPZ-03) Model Name: QPZ-03 Common Name: Light Based Treatment for Cold Sores Herpes Simplex Virus-1 Review Panel: General & Plastic Surgery Product Code: OKJ Regulation Number: 21 CFR 878.4860 Regulatory Class: II

3. Predicate Device Information

Sponsor: Light Tree Ventures Europe B.V. Trade Name: Cold Sore Device Model Name: QPZ-01 510(K) Number: K222205 Review Panel: General & Plastic Surgery Product Code: OKJ Regulation Number: 21 CFR 878.4860 Classification Name: Light based energy source device for topical application Regulation Class: II

4. Device Description

The Cold Sore Device is a solid state opto-electronic device that emits a controlled quantity of 1072nm +/- 12nm peak wavelength near infrared light for a period of approximately 3 minutes. The maximum peak light intensity across the treatment surface is 20mW/cm². The light output and duration are monitored by a microprocessor. The tip of the device that contacts the patient is made out of Acrylonitrile Butadiene Styrene (ABS) + PC.

{5}------------------------------------------------

K241155

Treatment with the device is commenced at the first symptoms of a cold sore 3 times a day with 4 hours in between each treatment for 2 consecutive days. The treatment area is approximately 3 cm². There are 2 light emitting diodes (LEDs) in the treatment area. The light within the device is activated by opening the cover of the device and automatically shut down after the pre-programmed treatment time (3 minutes). The device is designed for external, limited duration skin contact in an environment free from fluids and is provided non-sterile. The LEDs do not come in direct contact with the patient based upon the design of the device. The device is for OTC use and single patient use as described in the patient and box labeling.

5. Intended Use / Indications for Use

The Cold Sore Device is indicated for shortening the time to healing of herpes simplex labialis lesions on or around the lips with time to healing defined as the time to patient described re-epithelization.

6. Test Summary

6.1 Non-Clinical Tests Performed

Non-clinical tests were performed on the subject device in order to validate the design and to assure conformance with the following voluntary design standards in connection with medical safety, and electromagnetic compatibility:

  • ANSI/AAMI ES60601-1: 2005 & A1:2012 & A2:2021 Medical equipment Part 1: General requirements for � basic safety and essential performance. / IEC 60601-1:2012+AMD2:2020 Edition 3.2
  • � IEC 60601-1-11: 2015+AMD1:2020 Edition 2.1 Medical Electrical Equipment --Part 1: General Requirements for Basic Safety and Essential Performance -- Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment.
  • � IEC 60601-2-57: 2023 Edition 2.0 Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, monitoring and cosmetic/aesthetic use.
  • � IEC 60601-1-2: 2014+AMD1:2020 Edition 4.1 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
  • � IEC 62471: 2006 First edition Photobiological safety of lamps and lamp systems.
  • � IEC 62133-2: 2017+AMD1:2021 Edition1.0 Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems.
  • � IEC 60601-1-6: 2010+AMD1:2020 Edition 3.2 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
  • � IEC 62366-1: 2015+AMD1:2020 Edition 1.1 Medical devices - Part 1: Application of usability engineering to medical devices

6.2 Biocompatibility Test

  • � ISO 10993-1: 2018 Fifth edition Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management.
  • � ISO 10993-5:2009 Third edition Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
  • � ISO 10993-10 Third Edition 2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization

6.3 Discussion of Clinical Tests Performed

The intended use and technical parameters of the subject device are highly consistent with the predicate device (K22205), and important parameters such as output wavelength, treatment time, etc. are all consistent with the predicate device (K22205). At the same time, the safety of the equipment has also been tested by safety regulations, EMC and performance. So, there is no clinical test on our device.

{6}------------------------------------------------

K241155

6.4 Software Validation and Verification Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's 2023 Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions." The software documentation level was considered to be basic level, since a failure or latent design flaw could directly result in minor injury to the patient or operator.

6.5 Human Factors/Usability

According to "Applying Human Factors and Usability Engineering to Medical Devices: Guidance for Industry and Food and Drug Administration Staff", the researchers conducted simulated usability tests, including self-selection, device usability, and label comprehension.

At first, the subjects made self-selection according to the outer packaging to determine whether the subjects can make a correct self-selection, and the researchers made judgments about the subjects. Subjects who were identified as inapropriate equipment in self-selection was withdrawn from the study. The results showed that all 45 subjects could make self-selection correctly, and 3 subjects were not suitable for the device and withdrew from the study.

Next, the 42 subjects underwent a usability study. The researchers simulated the user's home environment and had subjects simulate the operation of the investigator recorded the operation of the subject to determine whether the using instructions provided on the user manual are understood by the results showed that all 42 subjects could perform the device usability operation correctly.

Final, the subjects answered a series of questions related to label comprehension. Investigators recorded and analyzed responses to these questions to determine whether the informations provided on the label were understood by the subjects. The results showed that all 42 subjects could correctly understand the information and instructions provided on the label.

After analyzing, usability test results, the participating subjects could correctly perform self-selection, and understand labeling content.

7. Comparison to predicate device and conclusion

Compare to the predicate device, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between subject device do not raise and new questions of safety or effectiveness.

Elements ofComparisonSubject DevicePredicate DeviceRemark
CompanyShenzhen Nuon Medical EquipmentCo., LtdLight Tree Ventures Europe B.V.
Trade NameCold Sore Device (Model: QPZ-03)Cold Sore Device (Model: QPZ-01)--
Classification NameLight based energy source device fortopical applicationLight based energy source device fortopical application--
510(k) NumberTBDK222205--
Product CodeOKJOKJSame
Intended Use /Indications for UseThe Cold Sore Device is indicated forshortening the time to healing of herpessimplex labialis lesions on or around thelips with time to healing defined as thetime to patient described re-epithelialization.The Cold Sore Device is indicated forshortening the time to healing of herpessimplex labialis lesions on or aroundthe lips with time to healing defined asthe time to patient described re-epithelialization.Same
Wavelengths1072nm +/- 12nm1072nm +/- 12nmSame
Energy SourceLEDLEDSame
Treatment Schedule3X/day, 4 hours between treatments, for2 consecutive day3X/day, 4 hours between treatments,for 2 consecutive daysSame
Auto-off feature (after3 min treatment)YesYesSame
Treatment time3-minute treatment cycle3-minute treatment cycleSame

{7}------------------------------------------------

Elements ofComparisonSubject DevicePredicate DeviceRemark
Treatment area3 cm²3 cm²Same
Energy density(mw/cm²)20 mw/cm²20 mw/cm²Same
Power supplyAdapter:Input: 100-240Va.c., 50/60Hz, 0.35AOutput: 5.0Vd.c., 1.0ALithium battery, 3.7V, 350mAhAdapter:Input: 100-240Va.c., 50/60Hz, 0.35AOutput: 5.0Vd.c., 1.0ALithium battery, 3.7V, 350mAhSame
Location for UseOTCOTCSame
Safety and EMCIEC 60601-1IEC 60601-1Same
IEC 60601-1-2IEC 60601-1-2
IEC 60601-1-11IEC 60601-1-11
IEC 62471IEC 62471
IEC 60601-2-57IEC 60601-2-57
IEC 62133-2IEC 62133-2
BiocompatibilityISO 10993-1ISO 10993-1Same
ISO 10993-5ISO 10993-5
ISO 10993-10ISO 10993-10

Conclusion:

The proposed device uses technology that is similar to the predicate device. The technology and design do not raise new types of questions regarding safety and effectiveness for the proposed indications for use and the performance testing supports that the device can be used safely and effectively for the proposed indications for use. The proposed device is considered to be substantially equivalent to the predicate device K222205.

8. Date of the summary prepared: September 5, 2024

Version: V5.0

§ 878.4860 Light based energy source device for topical application.

(a)
Identification. The device emits light energy at near infrared spectrum and is applied externally to the surface of herpes simplex labialis lesions on or around the lips.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The technical parameters of the device, including wavelength, treatment time, treatment area, energy density, spot size, and power, must be characterized.
(2) The cleaning and disinfection instructions for the device must be validated.
(3) The device must be demonstrated to be biocompatible.
(4) Performance testing must validate electromagnetic compatibility (EMC), ocular safety, and electrical safety of the device.
(5) Labeling must direct end-users to contact the device manufacturer and MedWatch if they experience any adverse events when using this device.
(6) Labeling must include specific information pertinent to use of the device by the intended patient population and the treatment regimen.
(7) Simulated use testing must include information from a usability, label comprehension and self-selection study to demonstrate that the device can be used by the intended patient population without any assistance.
(8) Clinical data must show adequate reduction in time to healing and assess risks of redness, discomfort, burns, and blisters.