K Number
K241111

Validate with FDA (Live)

Date Cleared
2024-07-05

(74 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AMT Suture Passer is indicated to pass suture through soft tissues of the body during endoscopic/laparoscopic surgery and interventional radiology procedures in child, adolescent and adult populations.

Device Description

The AMT Suture Passer is a sterile, single-case use, hand-held suture grasping device. The device features a hypodermic needle through which a suture can be passed and retrieved through another. The retrieval component may be offered in two configurations: a non-magnetic version, and a magnetically assisted version, using a grasper with a magnetized arm. Using either configuration, the AMT Suture Passer provides more flexibility in stitch geometry and approach than traditional U-stitches. The preferred method to introduce the suture is through a separate introducer needle; the AMT Suture Passer Kit with Magnet Assist instrument set bundles the magnetically assisted configuration of the AMT Suture Passer with an introducer needle and two magnetic sutures to create a non-procedure-specific kit. For the magnetically assisted configuration, the magnetized arm of the grasper works in conjunction with a magnetic suture to reduce the technical difficulty of intracorporeal suture retrieval under laparoscopic, endoscopic, radiologic, or ultrasound guidance. This magnet-assist technology is derived from the ATLAS Suture Delivery System (K193612), also manufactured by Applied Medical Technology, which performs the same clinical function and is used in similar clinical applications. Both configurations of the AMT Suture Passer feature identical components, except for the end-effectors, and make use of Luer-lock compatible hubs that allow a syringe to be attached for the purpose of administering contrast through the device. The contrast enters through the Luer hub and exits from the distal end of the needle, which allows the needle position to be verified prior to passing the suture during interventional radiology procedures.

AI/ML Overview

This document is an FDA 510(k) clearance letter for a medical device called the "AMT Suture Passer." It outlines the regulatory review process and the basis for the FDA's determination of substantial equivalence to previously cleared devices.

The request asks for details on acceptance criteria and a study proving the device meets these criteria. However, the provided document does not describe a study involving an AI/Machine Learning algorithm, human readers, or a test set with ground truth established by experts.

Instead, the document details the performance data for a physical medical device (AMT Suture Passer) through bench testing, biocompatibility testing, sterilization validation, and shelf-life testing. It explicitly states:

  • "B. Software: There are no software components related in any way to this device. Therefore, Software Validation is not applicable to this device."
  • "E. Animal Study: Animal testing was NOT performed."
  • "F. Clinical Study: Clinical testing was NOT performed."

Therefore, I cannot provide the requested information regarding an AI study, test sets, expert ground truth, adjudication methods, MRMC studies, standalone algorithm performance, or training set details from this document. The information requested aligns with studies typically performed for AI/ML-driven medical devices, which the AMT Suture Passer is not.

However, I can extract the acceptance criteria and performance data for the physical device as presented in the document:


Acceptance Criteria and Device Performance (Based on the Provided Document for a Physical Device)

The "AMT Suture Passer" is a physical medical device, not an AI/ML diagnostic tool. Therefore, the acceptance criteria and performance data relate to its physical properties, sterility, biocompatibility, and functionality as a surgical instrument. There is no mention of AI performance metrics like sensitivity, specificity, AUC, or reader improvement.

Here's a summary of the performance data and acceptance criteria as described:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Test/Standard AppliedAcceptance CriteriaReported Device Performance
BiocompatibilityISO 10993-1, -5, -7, -10, -11, -23; USP-NF <85>, <161>Met acceptance criteria for limited contact (≤24 hours) with tissue/bone/dentin. Suture component: Met acceptance criteria for prolonged contact (>24 hours to 30 days) with tissue/bone/dentin."It was determined that the AMT Suture Passer met the acceptance criteria for limited contact (24 hours or less) with tissue/bone/dentin. The suture component of the subject device met the acceptance criteria for prolonged contact (greater than 24 hours to 30 days) with tissue/bone/dentin."
SoftwareN/A (Device has no software)N/A"There are no software components related in any way to this device. Therefore, Software Validation is not applicable to this device."
Electromagnetic Comp./Electrical SafetyN/A (Device has no electronic components)N/A"There are no electronic components related in any way to this device."
SterilizationISO 10993-7Sterility Assurance Level (SAL) of 10-6; compliance with standards."The subject device is ethylene oxide sterilized, and has been validated to confirm a Sterility Assurance Level (SAL) of 10-6. The sterilization processing complies with the standards."
Shelf LifeISO 11607-1 & 2; ISTA 3A 2018Validated shelf life demonstrated."Testing indicates that the subject device has a validated shelf life of three (3) years."
Bench Testing (Functional & Material)ISO 7864 (hypodermic needles), ISO 9626 (stainless steel needle tubing), USP-NF <881> (Tensile Strength), USP-NF <871> (Sutures Needle Attachment), ISO 80369-1 & -7 (small-bore connectors)Conformance to applicable recognized standards; reliable design and performance under specified parameters; met or exceeded all acceptance criteria."The AMT Suture Passer met or exceeded all the acceptance criteria and does not raise new questions of safety or effectiveness when compared to the predicate."

Since the device is a physical instrument without AI components, the following points as requested in the prompt are not applicable or detailed in the provided document:

  1. Sample size used for the test set and the data provenance: Not applicable for a physical device's bench/biocompatibility testing in this context. These tests are typically performed on a statistically relevant sample size of manufactured devices/materials, but specific numbers are not provided. Data provenance would be from manufacturing and testing labs.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a physical device's performance is established by direct measurement and compliance with engineering/material standards, not by expert interpretation of data.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable (no AI component).
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable (no AI algorithm).
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For a physical device, the "ground truth" for acceptance is determined by adherence to established material specifications, manufacturing tolerances, mechanical performance criteria, and sterility/biocompatibility standards as measured through laboratory testing.
  7. The sample size for the training set: Not applicable (no AI component).
  8. How the ground truth for the training set was established: Not applicable (no AI component).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. To the left of the FDA logo is the Department of Health & Human Services logo.

July 5, 2024

Applied Medical Technology, Inc. Jennifer Hazou Regulatory Affairs Specialist 8006 Katherine Blvd. Brecksville, Ohio 44141

Re: K241111

Trade/Device Name: AMT Suture Passer Regulation Number: 21 CFR 21 CFR 876.1500 Regulation Name: Laparoscope, General & Plastic Surgery Regulatory Class: Class II Product Code: GCJ Dated: April 22, 2024 Received: April 22, 2024

Dear Jennifer Hazou:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Long H. Chen
Long H. Chen -S-S
Date: 2024.07.05 10:56:26 -04'00'

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K241111

Device Name

AMT Suture Passer

Indications for Use (Describe)

The AMT Suture Passer is indicated to pass suture through soft tissues of the body during endoscopic/laparoscopic surgery and interventional radiology procedures in child, adolescent and adult populations.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

AMT Suture Passer

I. SUBMITTER:

Applied Medical Technology, Inc. 8006 Katherine Boulevard Brecksville, OH 44141 Phone: 440-717-4000 Fax: 440-717-4200

Contact Person: Jennifer Hazou - Regulatory Affairs Specialist Email: Jennifer.Hazou@appliedmedical.net Date Prepared: June 25, 2024

II. DEVICE INFORMATION:

Trade/Device Name: AMT Suture Passer Common Name: Laparoscope, General & Plastic Surgery Regulation Number: 21 CFR 876.1500 Regulation Name: Laparoscope, General & Plastic Surgery Regulatory Class: II Product Code: GCJ

III. PREDICATE INFORMATION:

Carter-Thomason Predicate: K980123 (Needle Point Suture Passer Instrument Set - CTSG; Carter-Thomason)

AMT Predicate Device: K193612 (Suture Delivery System; Applied Medical Technology, Inc.)

Teleflex Predicate Device: K192490 (Cottony II Polyester Suture; Teleflex Medical)

Cook Predicate Device: K182832 (Cope Pediatric Gastrointestinal Suture Anchor Set & Enterostomy Suture Anchor Set; Cook Incorporated)

** The predicate devices have not been subject to design-related recalls.

IV. INDICATIONS FOR USE:

The AMT Suture Passer is indicated to pass suture through soft tissues of the body during endoscopic/laparoscopic surgery and interventional radiology procedures in child, adolescent and adult populations.

DEVICE DESCRIPTION: V.

The AMT Suture Passer is a sterile, single-case use, hand-held suture grasping device. The device features a hypodermic needle through which a suture can be passed and retrieved through another. The retrieval component may be offered in two configurations: a non-magnetic version, and a

{4}------------------------------------------------

magnetically assisted version, using a grasper with a magnetized arm. Using either configuration, the AMT Suture Passer provides more flexibility in stitch geometry and approach than traditional Ustitches. The preferred method to introduce the suture is through a separate introducer needle; the AMT Suture Passer Kit with Magnet Assist instrument set bundles the magnetically assisted configuration of the AMT Suture Passer with an introducer needle and two magnetic sutures to create a non-procedure-specific kit.

For the magnetically assisted configuration, the magnetized arm of the grasper works in conjunction with a magnetic suture to reduce the technical difficulty of intracorporeal suture retrieval under laparoscopic, endoscopic, radiologic, or ultrasound guidance. This magnet-assist technology is derived from the ATLAS Suture Delivery System (K193612), also manufactured by Applied Medical Technology, which performs the same clinical function and is used in similar clinical applications.

Both configurations of the AMT Suture Passer feature identical components, except for the endeffectors, and make use of Luer-lock compatible hubs that allow a syringe to be attached for the purpose of administering contrast through the device. The contrast enters through the Luer hub and exits from the distal end of the needle, which allows the needle position to be verified prior to passing the suture during interventional radiology procedures.

{5}------------------------------------------------

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATES:

SubjectCarter-ThomasonPredicateAMT PredicateCottony PredicateCope Predicate
AMT Suture Passer(KXXXXXX)Carter-Thomason SGSuture Passer(K980123)ATLAS Suture DeliverySystem(K193612)Cottony II Suture(K153076, K021019,K192490)Cope EnterostomySuture Anchor Set andCope PediatricGastrointestinal SutureAnchor Set (K182832)Comparison
ClassificationDeviceClassificationClass IIClass IIClass IIClass IIClass IISAME AS PREDICATE
Product CodesGCJGCJKGCGATKGCThe subject device has thesame classification and productcode as the predicate.
Regulation No.876.1500876.1500876.5980878.5000876.5980
PanelGeneral & PlasticSurgeryGeneral & PlasticSurgeryGastroenterology/UrologyGeneral & PlasticSurgeryGastroenterology/Urology
SterilityConditionProvided sterileProvided sterileProvided sterileProvided sterileProvided sterileSAME AS PREDICATE
ReprocessingSingle use only; not re-processableSingle use only; not re-processableSingle use only; not re-processableSingle use only; not re-processableSingle use only; not re-processableThe subject device andpredicate devices are sterile,single-use devices that are notre-processable. Sterilizationmethod (EO) is the same andsimilar packaging is employed,including a tray for organization,and a peel-open Tyvek lid tomaintain the sterile barrier. Thesubject device also uses apouch, which provides asecond sterile barrier.
MethodEthylene OxideEthylene OxideEthylene OxideEthylene OxideEthylene Oxide
PackagingTray w/peel-openTyvek lid inside peel-open Tyvek pouchTray w/ peel-openTyvek lidTray w/peel-openTyvek lid inside peel-open Tyvek pouchPeel-open paperpacket inside peel-openTyvek pouchPeel-open TyvekPouch
AssuranceLevelSAL 10-6SAL 10-6SAL 10-6SAL 10-6SAL 10-6
SubjectCarter-ThomasonPredicateAMT PredicateCottony PredicateCope PredicateComparison
AMT Suture Passer(KXXXXXX)Carter-Thomason SGSuture Passer(K980123)ATLAS Suture DeliverySystem(K193612)Cottony II Suture(K153076, K021019,K192490)Cope EnterostomySuture Anchor Set andCope PediatricGastrointestinal SutureAnchor Set (K182832)
PrescriptionPrescription OnlyPrescription OnlyPrescription OnlyPrescription OnlyPrescription OnlySAME AS PREDICATE
User PopulationIntended UserSurgeons trained inendoscopic, and/orlaparoscopictechniques and/orinterventionalradiologists.Surgeons trained inendoscopic and/orlaparoscopictechniquesSurgeons trained inendoscopic techniquesVarious clinicianstrained in the use ofsuturesSurgeons,gastrointestinal doctors,and interventionalradiologistsThe subject device and thepredicate devices areprescription only devices thatserve the same userpopulations.
Clinical ApplicationIndications forUseIndicated to passsuture through softtissues of the bodyduringendoscopic/laparoscopic surgery andinterventional radiologyprocedures in child,adolescent and adultpopulations.Indicated to passsuture through softtissues of the bodyduring endoscopic/laparoscopic surgeryIntended for anchoringthe wall of a hollowviscus to the abdominalwall prior to theintroduction ofinterventional cathetersand stay in place for upto 14 days, in child,adolescent, and adultpopulationsIndicated for use ingeneral soft tissueapproximation and/orligation, including useincardiovascular,ophthalmic, orthopedicand neurologicalproceduresIntended for anchoringthe wall of a hollowviscus to the abdominalwall prior to theintroduction ofinterventional cathetersand stay in place for upto 14 daysSIMILAR TO PREDICATEThe subject device differs fromthe predicate in the addition ofinterventional radiologyprocedures to the indicationsfor use, and in the inclusion ofthe patient population in theindications for use.
Intended UsePasses a suturethrough soft tissues ofthe bodyPasses a suturethrough soft tissues ofthe bodyPasses a suturethrough soft tissues ofthe body for thepurpose of securing ahollow viscus to theanterior abdominal wallApproximation and/orligation of soft tissuesDelivers a suture with ametal anchor throughsoft tissues of the bodyfor the purpose ofsecuring the stomachto anterior theabdominal wallExpansion of the indication toIR procedures is supportedbecause the subject deviceallows contrast to be injectedthrough the device.
SubjectCarter-ThomasonPredicateAMT PredicateCottony PredicateCope Predicate
AMT Suture Passer(KXXXXXX)Carter-Thomason SGSuture Passer(K980123)ATLAS Suture DeliverySystem(K193612)Cottony II Suture(K153076, K021019,K192490)Cope EnterostomySuture Anchor Set andCope PediatricGastrointestinal SutureAnchor Set (K182832)Comparison
Procedural UseGeneral surgeryGeneral surgeryPexy procedures (e.g.,gastropexy, jejunopexy,cystopexy, etc.)General surgery(includingcardiovascular,ophthalmic, orthopedicand neurologicalprocedures)GastropexyThe addition of the patientpopulation to the indications iskeeping in line with current bestpractices for device labeling.
DeviceNot MR-safeNot MR-safeNot MR-safeN/A - SUTURE ONLYNot MR-safeSAME AS PREDICATE
MR SafetySutureFor the configurationwhich featuresmagnetically assistedend-effectors, suture isMR-safe after magnethead is removed*Suture not included,but required inaccordance withlabeling. Suture is MR-safe after needle isremoved.Suture is MR-safe aftermagnet head isremovedSuture is MR-safe afterneedle is removedSuture is MR-conditional due to metalT-bar anchor thatremains inside thepatient after placementNeither the subject device northe predicate devices are MR-safe. Suture placed using anyof the devices is MR-safe whenplaced in accordance with thelabeling.
ConfigurationsInstrumentKit(s)For the configurationwhich featuresmagnetically assistedend-effectors, YES -AMT Suture Passer KitFor the configurationthat features portclosure elements, YES- Carter-Thomason IIPort Closure SystemNONOYES - Gastropexy KitSIMILAR TO PREDICATEThe subject device uses a smallgauge needle that is shorterthan the predicate and may beoffered in curvedconfigurations.
Needle Sizes17-gauge: May beoffered curved orstraight, depending onconfiguration.14-gauge17-gaugeMultiple sizes:Needle size scales withsuture size18-gaugeThe subject device utilizes anechogenic treatment on theneedle.
TechnologicalCharacteristicsSubjectCarter-ThomasonPredicateAMT PredicateCottony PredicateCope Predicate
AMT Suture Passer(KXXXXXX)Carter-Thomason SGSuture Passer(K980123)ATLAS Suture DeliverySystem(K193612)Cottony II Suture(K153076, K021019,K192490)Cope EnterostomySuture Anchor Set andCope PediatricGastrointestinal SutureAnchor Set (K182832)Comparison
Needle Length2.5 in. [6.35 cm]8.3 in. [120.82 cm]2.5 in. [6.35 cm]Multiple sizes available(needle size scalesrelative to suture size)7 cmSuture is included with oneconfiguration of the subjectdevice, but must be furnishedby the user with the other andthe predicate.
Needle TypeHypodermic (hollow,round cross-section)Hypodermic (hollow,round cross-section)Hypodermic (hollow,round cross-section)Solid, available inmultiple cross-sectionalshapesHypodermic (hollow,round cross-section)the predicate.
Needle FinishAs-drawn; echogenictreatmentAs-drawnAs-drawnMachined/groundAs-drawn
Suture SizeFor the configurationwhich featuresmagnetically assistedend-effectors, USP 3-0*Suture not included.Recommended size isUSP 1USP 3-0USP 7-0 through 9USP 2-0
DesignArchetypeGrasper-type suturepasserGrasper-type suturepasserMagnetic suture passerSurgical sutureT-bar suture anchorSIMILAR TO PREDICATEBoth the subject device and the
OperationalStyleHand-held; manuallyoperated / no externalpower sourceHand-held; manuallyoperated / no externalpower sourceHand-held; manuallyoperated / no externalpower sourceHand-held; manuallyoperated / no externalpower sourceHand-held; manuallyoperated / no externalpower sourceCarter-Thomason predicatedevice follow the same designarchetype and have the sameoperational style. Both devices
SutureSecurementMethodSuture mechanicallytrapped betweengrasper and wall ofneedleSuture mechanicallytrapped betweengrasper and wall ofneedleSuture coupled toretriever by magneticconnectionManual (needle driver,etc.)N/A (suture is notretrieved using device)utilize similar means of suturesecurement, and both work withstandard surgical sutures.
SubjectCarter-ThomasonPredicateAMT PredicateCottony PredicateCope PredicateComparison
AMT Suture Passer(KXXXXXX)Carter-Thomason SGSuture Passer(K980123)ATLAS Suture DeliverySystem(K193612)Cottony II Suture(K153076, K021019,K192490)Cope EnterostomySuture Anchor Set andCope PediatricGastrointestinal SutureAnchor Set (K182832)
CompatibleSutureCompatible withstandard surgicalsutures and magneticsutures, dependingupon configurationCompatible withstandard surgicalsuturesCompatible withincluded magneticsutureN/ACompatible withincluded T-bar sutureThe subject device differs fromthe Carter-Thomason predicatein the use of magnet technologyand provision for fluid exchangethrough the device.
SutureSpecificationFor the configurationwhich featuresmagnetically assistedend-effectors, USP 3-0,polyester, braided,nonabsorbable,magnetic head*Suture not included(USP size 1recommended)USP 3-0, polyester,braided,nonabsorbable,magnetic headUSP 7-0 through 9,polyester, braided,nonabsorbable, needleheadUSP 3-0, Byosyn,monofilament,absorbable, T-bar head
MagnetTechnologyONLY for theconfiguration whichfeatures magneticallyassisted end-effectors -small permanent gold-plated magnets(NdFeB) incorporatedinto grasper arm andsuture headNOSmall permanent gold-plated magnets(NdFeB) incorporatedinto retriever arm andsuture headNONO
Fluid ExchangeLuer-lock compatiblehub for syringeattachment; patentlumen through deviceNONON/ALuer-lock compatiblehub for syringeattachment; patentlumen through device
SubjectCarter-ThomasonPredicateAMT PredicateCottony PredicateCope Predicate
AMT Suture Passer(KXXXXXX)Carter-Thomason SGSuture Passer(K980123)ATLAS Suture DeliverySystem(K193612)Cottony II Suture(K153076, K021019,K192490)Cope EnterostomySuture Anchor Set andCope PediatricGastrointestinal SutureAnchor Set (K182832)Comparison
ConstructionMainComponents• Instrument body w/hypodermicneedle(s)• Grasper assembly• Lock assembly• Suture with magnethead, for theconfiguration whichfeaturesmagneticallyassisted end-effectors• Instrument body w/hypodermic needle• Grasper assembly• Port-closureelements ONLY withthe Carter-Thomason II PortClosure System• Instrument body w/hypodermic needles(2)• Magnetic suturedelivery cannulaassembly• Magnetic sutureretrieval probeassembly• Suture with magnethead• Skin bumper• Suture with needlehead• Instrument body w/hypodermic needle• T-bar suture deliverycannula assembly• Suture with T-barhead• Skin bumper/ sutureclipSIMILAR TO PREDICATEThe main components thatcomprise the subject device aresimilar to the predicate, differingin the inclusion of magneticsutures for the relevantconfiguration and a lockingmechanism.
Materials ofConstruction• 300-series Stainlesssteel• Nitinol• Thermoplastics• Adhesives• Gold-platedneodymium(magnets), ONLY forthe configurationwhich featuresmagneticallyassisted end-effectors• 300-series Stainlesssteel• Thermoplastics• Adhesives• 300-series Stainlesssteel• Nitinol• Gold-platedneodymium(magnets)• Thermoplastics• Adhesives• Silicone• 300- or 400-seriesStainless steel orMaraging Steel• Thermoplastics• 300-series Stainlesssteel• Thermoplastics• Bio-absorbablethermoplastics• Adhesives
SubjectCarter-ThomasonPredicateAMT PredicateCottony PredicateCope Predicate
AMT Suture Passer(KXXXXXX)Carter-Thomason SGSuture Passer(K980123)ATLAS Suture DeliverySystem(K193612)Cottony II Suture(K153076, K021019,K192490)Cope EnterostomySuture Anchor Set andCope PediatricGastrointestinal SutureAnchor Set (K182832)Comparison
Construction (Cont.)ManufacturingProcessesThermoplasticInjection Molding ThermoplasticExtrusion Progressive DieStamping DOM Tube Forming Bending/ Coiling Filament Extrusion +Braiding Cutting/ Grinding/Machining Shape Setting Assembly with UV-Cured Adhesives Assembly withCyanoacrylates Assembly withoutAdhesives Swaging/ Crimping PowderedMetallurgy + GoldPlating, ONLY forthe configurationwhich featuresmagneticallyassisted end-effectorsThermoplasticExtrusion ThermoplasticInjection Molding Metal Stamping DOM Tube Forming Bending/ Coiling Cutting/ Grinding/Machining Assembly withAdhesives Assembly withoutAdhesivesThermoplasticInjection Molding ThermoplasticExtrusion Silicone InjectionMolding Silicone Extrusion Progressive DieStamping DOM Tube Forming Bending/ Coiling Filament Extrusion +Braiding Cutting/ Grinding/Machining PowderedMetallurgy + GoldPlating Autogenous Welding+ Shape Setting Assembly with UV-Cured Adhesives Assembly withCyanoacrylates Assembly withoutAdhesives Swaging/ CrimpingFilament Extrusion +Braiding Cutting/ Grinding/Machining Assembly withoutAdhesives Swaging/ CrimpingThermoplasticInjection Molding Silicone InjectionMolding DOM Tube Forming Filament Extrusion Cutting/ Grinding/Machining Assembly withAdhesives Assembly withoutAdhesives Swaging/ CrimpingSIMILAR TO PREDICATEThe subject device employsmanufacturing processes thatare similar to the predicatedevices.The differences that existbetween the subject device andpredicate device are related tothe differences in componentsdescribed above (e.g.,powdered metallurgy + goldplating relates to theconstruction of the permanentmagnets, etc. ONLY for theconfiguration which featuresmagnetically assisted end-effectors - components whichare not found in the predicatedevice).

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

VII. PERFORMANCE DATA:

  • A. Biocompatibility Testing: Following a Biological Evaluation Plan, the AMT Suture Passer was tested for biocompatibility based on the applicable sections of the following standards:
    • . ISO 10993-1 Fifth edition 2018-08 - Biological evaluation of medical devices --Part 1: Evaluation and testing within a risk management process
    • ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices . --Part 5: Tests for in vitro cytotoxicity
    • ISO 10993-7 Second edition 2008-10-15 Biological evaluation of medical devices . --Part 7: Ethylene oxide sterilization residuals
    • ISO 10993-10 Fourth edition 2021-11 Biological evaluation of medical devices ● -- Part 10: Tests for skin sensitization
    • ISO 10993-11 Third edition 2017-09 Biological evaluation of medical devices ● --Part 11: Tests for systemic toxicity
    • ISO 10993-23 First edition 2021-01 Biological evaluation of medical devices ● --Part 23: Tests for irritation
    • . USP-NF M98830 02 01 <85> Bacterial Endotoxins Test
    • USP-NF M98910 01 01 ● <161> Medical Devices - Bacterial Endotoxin and Pyrogen Tests

It was determined that the AMT Suture Passer met the acceptance criteria for limited contact (24 hours or less) with tissue/bone/dentin. The suture component of the subject device met the acceptance criteria for prolonged contact (greater than 24 hours to 30 days) with tissue/bone/dentin.

  • B. Software: There are no software components related in any way to this device. Therefore, Software Validation is not applicable to this device.
  • C. Electromagnetic Compatibility & Electrical Safety: There are no electronic components related in any way to this device.

D. Performance Testing:

1. Sterilization

Testing has been completed to evaluate the sterilization process for the subject device, and is outlined below:

  • Testing per ISO 10993-7. Second edition 2008-10-15 .
    • Biological evaluation of medical devices Part 7: Ethylene oxide sterilization o residuals

{13}------------------------------------------------

The subject device is ethylene oxide sterilized, and has been validated to confirm a Sterility Assurance Level (SAL) of 106. The sterilization processing complies with the standards.

2. Shelf Life

Applicable shelf life testing was conducted in accordance with the following:

  • . Testing per ISO 11607-1 & 2. Second edition 2019-02:
    • o Packaging for terminally sterilized medical devices
  • Testing per ISTA 3A 2018: Packaged products for parcel delivery system shipment 70kg (150 lb) or less o

Testing indicates that the subject device has a validated shelf life of three (3) years.

3. Bench Testing

Bench tests have been carried out to demonstrate conformance to applicable recognized standards and to assure reliable design and performance under the specified testing parameters according to predetermined criteria. The tests carried out include:

  • Testing per ISO 7864 Fourth edition 2016-08-01 ●
    • Sterile hypodermic needles for single use Requirements and test methods o
  • . Testing per ISO 9626 Second edition 2016-08-01
    • Stainless steel needle tubing for the manufacture of medical devices -O Requirements and test methods
  • Testing per USP-NF M99670 02 01
    • <881> Tensile Strength о
  • Testing per USP-NF M99660 03 01 ●
    • o <871> Sutures Needle Attachment
  • Testing per ISO 80369-1 Second edition 2018-11 ●
    • Small-bore connectors for liquids and gases in healthcare applications Part 1: O General requirements
  • Testing per ISO 80369-7 Second edition 2021-05
    • Small-bore connectors for liquids and gases in healthcare applications Part 7: O Connectors for intravascular or hypodermic applications

The AMT Suture Passer met or exceeded all the acceptance criteria and does not raise new questions of safety or effectiveness when compared to the predicate.

  • E. Animal Study: Animal testing was NOT performed.
  • F. Clinical Study: Clinical testing was NOT performed.

{14}------------------------------------------------

VIII. CONCLUSION:

The subject device is comparable in intended use, operative principles, materials of construction, methods of manufacture, and technology to the predicate devices and therefore do not raise new concerns regarding the safety or effectiveness of the subject device. Minor differences between the subject device and the predicate devices do not result in new concerns regarding safety and effectiveness, either. Therefore, the subject device may be claimed to be substantially equivalent to the predicate devices.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.