(191 days)
RemedyLogic AI MRI Lumbar Spine Reader ("RAI") is an image post-processing and measurement software tool that provides quantitative spine measurements from previously-acquired DICOM lumbar spine Magnetic Resonance (MR) images for users' review, and interpretation. It provides the following functionality to assist users in visualizing, measuring and documenting measurements:
· Feature segmentation;
· Feature measurement; and
· Exportation of measurement results in DICOM Structured Report and a downloadable .docx file for users to review and to use full or partial list of software-generated measurements to prepare their own radiology report.
RAI does not produce or recommend any type of medical diagnosis or treatment. Instead, it simply helps users to more easily identify and measure features in lumbar MR images and compile their own reports. The user is responsible for reviewing, verifying, and correcting, if necessary, the software-generated segmentations and measurements, leveraging useful software output and using their medical judgment and discretion to make diagnostic or treatment decisions.
The device is intended to be used only by radiologists, neuro- and spine-surgeons in hospitals and other medical institutions.
Only T2 MRI images in DICOM format, acquired from lumbar spine exams of patients aged 18 and older, are considered to be valid input. RAI does not support DICOM images of patients that are pregnant, undergo MRI scan with contrast media, or have post-operational complications, scoliosis, tumors, infections, and/or fractures.
The RemedyLogic AI MRI Lumbar Spine Reader (RAI) is an MR image post-processing and measurement software tool that provides quantitative spine measurements from previously acquired DICOM lumbar spine Magnetic Resonance (MR) images for qualified users' review, analysis, and interpretation. The qualified users (i.e., radiologists, spine- and neuro-surgeons) are physicians qualified to read and interpret spine MRI exams in a manner consistent with American College of Radiology (ACR) recommendations.
The RAI analyzes the user-uploaded lumbar spine images and provides the following functionalities to assist qualified users in visualizing images, and measuring images, and generating reports:
- Feature segmentation: the software automatically detects the borders of anatomical objects of interest and generates the corresponding contours for these objects.
- 0 Feature measurement: the software automatically generates common measurements for segmented objects.
- 0 Measurement export: a DICOM Structured Report or a .docx file containing the measurement results can be exported for users to review and to use the full or a partial list of the softwaregenerated measurements to prepare their own radiology reports.
The RAI software does not interface directly with any MR scanner or data collection equipment. Rather, a qualified user must upload a previously acquired MR study in DICOM format into the RAI software via their Picture Archiving and Communication System (PACS). The PACS serves as the RAI user interface. After less than two minutes of processing, the RAI software automatically generates and uploads back to PACS the DICOM with segmentations of regions of interest along with corresponding measurements. These measurements are also presented in a DICOM Structured Report and a downloadable .docx file, which is accessible for download from the PACS from PACS system. The user reviews the softwaregenerated measurement list, studies the software-annotated images and/or the original unannotated images when necessary, and reviews other pertinent medical information about the patient. The user can manually segment anatomical objects and mark their own measurements using the DICOM viewer tools. The user can also edit measurements in the downloaded .docx file. The user then writes their own radiology report, incorporating some or all verified or corrected measurements, with diagnosis and/or treatment recommendations.
The purpose of the RAI software is to save time by automating tedious, time-consuming, and potentially error-prone manual tasks. The software does not perform any functions that could not be accomplished by a qualified user. The outputs of the software, i.e. feature segmentations and quantitative measurements, are reviewed, analyzed, confirmed or corrected by the user before any such content is included in the user's final report.
The RAI consists of a cloud-based machine learning (ML) analytical algorithm deployed on a GPU cloud service and an API to integrate directly with the client's PACS system.
Here's a summary of the acceptance criteria and the study proving the device meets them, based on the provided text:
Device Name: RemedyLogic AI MRI Lumbar Spine Reader (RAI)
1. Table of Acceptance Criteria and Reported Device Performance
Primary Endpoints (Measurements)
| Measurement | MAE Limit (Acceptance Criteria) | Reported MAE (95% CI Upper Bound) | Success |
|---|---|---|---|
| Dural Sac Area (Axial) | 20 mm² | 17.9 mm² | Yes |
| Spinal Canal Area (Axial) | 30 mm² | 24.3 mm² | Yes |
| Lordotic Angle (Sagittal) | 6° | 3.9 ° | Yes |
| Vertebral Body Slippage (Sagittal) | 2 mm | 0.8 mm | Yes |
Secondary Endpoints (Segmentations)
| Anatomical Structure Segmentation | MDC Limit (Acceptance Criteria) | Reported MDC (95% CI Lower Bound) | Success |
|---|---|---|---|
| Artery (Axial) | 0.8 | 0.861 | Yes |
| Disc (Axial) | 0.7 | 0.796 | Yes |
| Disc (Sagittal) | 0.7 | 0.910 | Yes |
| Disc Material Outside IV Space (Axial) | 0.7 | 0.793 | Yes |
| Dural Sac (Axial) | 0.8 | 0.924 | Yes |
| Kidney (Axial) | 0.8 | 0.872 | Yes |
| Ligamentum Flavum (Axial) | 0.7 | 0.736 | Yes |
| Muscle (Axial) | 0.8 | 0.945 | Yes |
| Sacrum (Sagittal) | 0.8 | 0.923 | Yes |
| Spinal Canal (Axial) | 0.8 | 0.941 | Yes |
| Spinal Canal (Sagittal) | 0.8 | 0.865 | Yes |
| Vein (Axial) | 0.8 | 0.815 | Yes |
| Vertebral Arch (Axial) | 0.8 | 0.843 | Yes |
| Vertebral Body (Sagittal) | 0.8 | 0.894 | Yes |
Secondary Endpoints (Measurements)
| Measurement | MAE Limit (Acceptance Criteria) | Reported MAE (95% CI Upper Bound) | Success |
|---|---|---|---|
| Anterior Disc Height (Sagittal) | 2 mm | 1.32 mm | Yes |
| Anterior Vertebral Body Height (Sagittal) | 2 mm | 1.81 mm | Yes |
| Dural Sac Anterior-Posterior Diameter (Axial) | 2 mm | 1.52 mm | Yes |
| Dural Sac Transverse Diameter (Axial) | 2 mm | 1.22 mm | Yes |
| Middle Disc Height (Sagittal) | 2 mm | 1.04 mm | Yes |
| Middle Vertebral Body Height (Sagittal) | 2 mm | 1.33 mm | Yes |
| Posterior Disc Height (Sagittal) | 2 mm | 1.07 mm | Yes |
| Posterior Vertebral Body Height (Sagittal) | 2 mm | 1.72 mm | Yes |
| Spinal Canal Anterior-Posterior Diameter (Axial) | 2 mm | 0.83 mm | Yes |
| Spinal Canal Transverse Diameter (Axial) | 2 mm | 1.85 mm | Yes |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: 200 MR image studies for 200 patients.
- Data Provenance: The data was collected from three (3) geographically diverse sites across the U.S. The study was a "standalone software performance study conducted in the U.S." This indicates the data was retrospective as it involved "previously-acquired MR studies" for analysis.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Number of Experts: Five (5) radiologists.
- Qualifications: They were U.S. radiologists (no specific experience years were mentioned, but it's implied they were qualified to interpret spine MRI exams consistent with ACR recommendations, as stated for "qualified users").
4. Adjudication Method for the Test Set
- For Segmentation: For each anatomical structure, the ground truth was established by the per-pixel majority opinion of the five (5) radiologists. Specifically, if at least 3 of the 5 radiologists labeled a pixel as belonging to a particular anatomical structure, the pixel was included; otherwise, it was excluded.
- For Measurement: The ground truth was established by taking the mean of the five (5) radiologists' measurements.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The study was described as a "standalone software performance study."
- Therefore, no effect size for human readers improving with AI vs. without AI assistance was reported.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Yes, a standalone software performance study was conducted. The study compared the RAI software outputs directly to the ground truth established by the five radiologists, without human readers interacting with the AI outputs in a diagnostic setting.
7. The Type of Ground Truth Used
- Expert Consensus: The ground truth was established by the independent assessment and consensus of five (5) U.S. radiologists. For segmentations, it was a majority vote (3 out of 5); for measurements, it was the mean of their individual measurements.
8. The Sample Size for the Training Set
- The document states that "The RAI software machine learning algorithm training and testing data used during the algorithm development... were all independent data sets." However, the specific sample size for the training set is not provided in this document.
9. How the Ground Truth for the Training Set Was Established
- The document mentions that training data was used during algorithm development, but it does not explicitly detail how the ground truth for this training set was established. It only states that the training data and testing data were independent.
{0}------------------------------------------------
October 30, 2024
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
Remedy Logic Inc. % Yu Zhao Principal Advisor Bridging Consulting LLC 2108 N St., Suite N Sacramento, CA 95816
Re: K241108
Trade/Device Name: RemedyLogic AI MRI Lumbar Spine Reader Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: OIH Dated: September 30, 2024 Received: September 30, 2024
Dear Yu Zhao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Re"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Saml for
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K241108
Device Name
RemedyLogic AI MRI Lumbar Spine Reader
Indications for Use (Describe)
RemedyLogic AI MRI Lumbar Spine Reader ("RAI") is an image post-processing and measurement software tool that provides quantitative spine measurements from previously-acquired DICOM lumbar spine Magnetic Resonance (MR) images for users' review, and interpretation. It provides the following functionality to assist users in visualizing, measuring and documenting measurements:
· Feature segmentation;
· Feature measurement; and
· Exportation of measurement results in DICOM Structured Report and a downloadable .docx file for users to review and to use full or partial list of software-generated measurements to prepare their own radiology report.
RAI does not produce or recommend any type of medical diagnosis or treatment. Instead, it simply helps users to more easily identify and measure features in lumbar MR images and compile their own reports. The user is responsible for reviewing, verifying, and correcting, if necessary, the software-generated segmentations and measurements, leveraging useful software output and using their medical judgment and discretion to make diagnostic or treatment decisions.
The device is intended to be used only by radiologists, neuro- and spine-surgeons in hospitals and other medical institutions.
Only T2 MRI images in DICOM format, acquired from lumbar spine exams of patients aged 18 and older, are considered to be valid input. RAI does not support DICOM images of patients that are pregnant, undergo MRI scan with contrast media, or have post-operational complications, scoliosis, tumors, infections, and/or fractures.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
K241108
510(k) Summary
1. Submitter
Remedy Logic Inc. 1177 Avenue of the Americas, 5th Floor New York, NY 10036 USA Contact Person: Andrej Rusakov Date Prepared: March 30, 2024
2. Device
Name of Device: RemedyLogic AI MRI Lumbar Spine Reader Common or Usual Name: RemedyLogic AI MRI Lumbar Spine Reader Classification Name: Medical image management and processing system (21 CFR 892.2050) Product Code: QIH Regulatory Class: II
3. Predicate Devices
Device Name: CoLumbo Manufacturer: Smart Soft Healthcare AD Classification Panel: Radiology Classification Name: Medical image management and processing system (21 CFR 892.2050) Product Code: OIH Device Class: Class II 510(k) Number: K220497 (cleared on June 23, 2022)
4. Device Description
The RemedyLogic AI MRI Lumbar Spine Reader (RAI) is an MR image post-processing and measurement software tool that provides quantitative spine measurements from previously acquired DICOM lumbar spine Magnetic Resonance (MR) images for qualified users' review, analysis, and interpretation. The qualified users (i.e., radiologists, spine- and neuro-surgeons) are physicians qualified to read and interpret spine MRI exams in a manner consistent with American College of Radiology (ACR) recommendations.
The RAI analyzes the user-uploaded lumbar spine images and provides the following functionalities to assist qualified users in visualizing images, and measuring images, and generating reports:
- Feature segmentation: the software automatically detects the borders of anatomical objects of interest and generates the corresponding contours for these objects.
{5}------------------------------------------------
- 0 Feature measurement: the software automatically generates common measurements for segmented objects.
- 0 Measurement export: a DICOM Structured Report or a .docx file containing the measurement results can be exported for users to review and to use the full or a partial list of the softwaregenerated measurements to prepare their own radiology reports.
The RAI software does not interface directly with any MR scanner or data collection equipment. Rather, a qualified user must upload a previously acquired MR study in DICOM format into the RAI software via their Picture Archiving and Communication System (PACS). The PACS serves as the RAI user interface. After less than two minutes of processing, the RAI software automatically generates and uploads back to PACS the DICOM with segmentations of regions of interest along with corresponding measurements. These measurements are also presented in a DICOM Structured Report and a downloadable .docx file, which is accessible for download from the PACS from PACS system. The user reviews the softwaregenerated measurement list, studies the software-annotated images and/or the original unannotated images when necessary, and reviews other pertinent medical information about the patient. The user can manually segment anatomical objects and mark their own measurements using the DICOM viewer tools. The user can also edit measurements in the downloaded .docx file. The user then writes their own radiology report, incorporating some or all verified or corrected measurements, with diagnosis and/or treatment recommendations.
The purpose of the RAI software is to save time by automating tedious, time-consuming, and potentially error-prone manual tasks. The software does not perform any functions that could not be accomplished by a qualified user. The outputs of the software, i.e. feature segmentations and quantitative measurements, are reviewed, analyzed, confirmed or corrected by the user before any such content is included in the user's final report.
The RAI consists of a cloud-based machine learning (ML) analytical algorithm deployed on a GPU cloud service and an API to integrate directly with the client's PACS system.
5. Indications for Use
RemedyLogic AI MRI Lumbar Spine Reader ("RAI") is an image post-processing and measurement software tool that provides quantitative spine measurements from previously-acquired DICOM lumbar spine Magnetic Resonance (MR) images for users' review, analysis, and interpretation. It provides the following functionality to assist users in visualizing, measuring and documenting measurements:
- . Feature segmentation;
- . Feature measurement; and
- Exportation of measurement results in a DICOM Structured Report and a downloadable .docx file for users to review and to use full or partial list of software-generated measurements to prepare their own radiology report.
RAI does not produce or recommend any type of medical diagnosis or treatment. Instead, it simply helps users to more easily identify and measure features in lumbar MR images and compile their own reports. The user is responsible for reviewing, verifying and correcting, if necessary, the software-generated segmentations and measurements, leveraging useful software output and using their medical judgment and discretion to make diagnostic or treatment decisions.
The device is intended to be used only by radiologists, neuro- and spine-surgeons in hospitals and other medical institutions.
Only T2 MRI images in DICOM format, acquired from lumbar spine exams of patients aged 18 and older, are considered to be valid input. RAI does not support DICOM images of patients that are pregnant,
{6}------------------------------------------------
undergo MRI scan with contrast media, or have post-operational complications, scoliosis, turnors, infections, and/or fractures.
6. Comparison of the Technological Characteristics with the Predicate Devices
In comparison to the Predicate Device and the Reference Devices, the Subject Device provides comparable outputs in terms of segmentation, measurement and reporting. A tabular high-level comparison of the Subject Device and the Predicate Device is provided in the table below.
{7}------------------------------------------------
| Predicate Device | Subject Device | |
|---|---|---|
| CoLumbo (K220497) | RAI | |
| Intended User | Radiologist and neuro- and spine-surgeons | Radiologist and neuro- and spine-surgeons |
| Intended PatientPopulation | Patients aged 18 years or older,undergoing clinical lumbar spine MRexams. | Patients aged 18 years or older,undergoing clinical lumbar spine MRexams. |
| CoLumbo does not support DICOMimages of patients that are pregnant,undergo MRI scan with contrast media,or have post-operational complications,scoliosis, tumors, infections, fractures. | RAI does not support DICOM imagesof patients that are pregnant, undergoMRI scan with contrast media, or havepost-operational complications,scoliosis, tumors, infections, fractures. | |
| Supported BodyPart | Lumbar spine | Lumbar spine |
| Segmentation | Software automatically generatessegmentations of anatomical features ofinterest | Software automatically generatessegmentations of anatomical features ofinterest |
| Measurement | Software automatically generatesmeasurements of interest | Software automatically generatesmeasurements of interest |
| Threshold-BasedOut-of-RangeMeasurements | Software automatically highlights andreports out-of-range measurementsbased on predetermined thresholds | Software does not automaticallyhighlight or report any out-of-rangemeasurements |
| Reporting | Exportation of measurement results to awritten report for user's review, reviseand approval | Exportation of measurement results inboth a DICOM Structured Report anda .docx file for users to review and touse full or partial list of software-generated measurements to preparetheir own radiology report. |
| SaMD | Yes | Yes |
| Algorithm | Deep Convolutional Image-to-ImageNeural Network | Convolutional Neural Network |
| SupportedModality | MR | MR |
Comparison of Technological Characteristics with Predicate Device
The Subject Device is substantially equivalent in comparison to the Predicate Device. The information regarding the Subject Device does not raise new questions about safety and effectiveness and demonstrates that RAI is at least as safe and effective as the legally marketed devices.
7. Performance Data
7.1.Biocompatibility Testing
Not applicable.
7.2.Electrical Safety and Electromagnetic Compatibility (EMC)
{8}------------------------------------------------
Not applicable.
7.3.Animal Study
Not applicable.
7.4. Voluntary Conformance Standards
RAI has been tested to meet the requirements of conformity to multiple industry standards. Non-clinical performance testing demonstrated that RAI complies with the following voluntary FDA recognized Consensus Standards listed in the table below.
| Recognition# | Standard |
|---|---|
| 13-79 | IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION, Medical device software— Software life cycle processes |
| 5-125 | ISO 14971 Third Edition 2019-12, Medical devices - Application of risk managementto medical devices |
| 5-129 | IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION, Medical devices -Part 1: Application of usability engineering to medical devices |
| 5-134 | ISO 15223-1 Fourth edition 2021-07, Medical devices - Symbols to be used withinformation to be supplied by the manufacturer - Part 1: General requirements |
| 12-349 | NEMA PS 3.1 - 3.20 2022d, Digital Imaging and Communications in Medicine(DICOM) Set |
| Voluntary Conformance Standards | ||
|---|---|---|
| -- | --------------------------------- | -- |
7.5.Nonclinical Tests
Remedy Logic has performed software design verification testing and a standalone performance assessment study, in accordance with the FDA guidance, General Principles of Software Validation; Final Guidance for Industry and FDA Staff, issued on January 11, 2002. All software requirements and risk analysis have been successfully verified and traced. The performance data demonstrates continued conformance with special controls for medical devices containing software.
Software documentation for Basic Documentation level, per FDA Guidance for Industry and Food and Drug Administration Staff, Content of Premarket Submissions for Device Software Functions, issued on June 14, 2023, were provided.
Remedy Logic conforms to the cybersecurity requirementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed or transferred from a medical device to an external recipient, per FDA Guidance for Industry and Food and Drug Administration Staff, Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions, issued on September 27, 2023, as well as FDA Guidance for Industry and Food and Drug Administration Staff, Postmarket Management of Cybersecurity in Medical Devices, issued on December 28, 2016. The vulnerability assessment and penetration testing demonstrated satisfactory security performance.
The nonclinical test data demonstrated conformance with special controls and substantial equivalence to predicate devices' performance.
Standalone Software Performance Study
To validate the RAI software from a clinical perspective, a clinical data based standalone software
{9}------------------------------------------------
performance study was conducted in the U.S. The standalone software performance study included 200 MR image studies for 200 patients of different ages and racial groups, collected from three (3) geographically diverse sites across the U.S. The standalone software performance study compared the RAI software outputs to the ground truth defined by five (5) radiologists on segmentations and measurements.
| Number of subjects | Percent of total | |
|---|---|---|
| Total number of subjects | 200 | 100% |
| Gender, Male | 92 | 46% |
| Gender, Female | 108 | 54% |
| Age, 18 - 21 | 43 | 21.5% |
| Age, 22 - 50 | 89 | 44.5% |
| Age, 51 - 100 | 68 | 34% |
| Racial, Caucasian | 107 | 53.5% |
| Racial, Black/African American | 16 | 8% |
| Racial, Hispanic | 58 | 29% |
| Racial, Asian or other | 19 | 9.5% |
Study Subjects
Imaging Systems
The 200 MR studies were acquired on MRI imaging systems made by five (5) manufacturers.
| Manufacturer | Model name | Tesla | Number of MRIs | Percent |
|---|---|---|---|---|
| GE MedicalSystems | GENESIS SIGNA | 1.5 | 5 | 2.5% |
| SIGNA Creator | 1.5 | 1 | 0.5% | |
| SIGNA EXCITE | 1.5 | 10 | 5% | |
| Signa HDxt | 1.5 | 21 | 10.5% |
{10}------------------------------------------------
| Philips | Prodiva CS | 1.5 | 5 | 2.5% |
|---|---|---|---|---|
| Philips MedicalSystems | Achieva | 1.5 | 1 | 0.5% |
| Ingenia | 1.5 | 11 | 5.5% | |
| Panorama 0.23 TPower | 0.23 | 21 | 10.5% | |
| SIEMENS | Espree | 1.5 | 25 | 12.5% |
| MAGNETOM Altea | 1.5 | 21 | 10.5% | |
| Skyra | 3 | 21 | 10.5% | |
| Symphony | 1.5 | 7 | 3.5% | |
| TOSHIBA MEC | MRT200PP2 | 1.5 | 51 | 25.5% |
Ground Truth
The ground truths for the segmentation of anatomical structures and the measurements were independently established by five (5) U.S. radiologists without using the RAI software.
For the segmentation, each radiologist used a specialized pixel labeling tool to independently label the pixels of the tissues at the predetermined levels of the preselected axial and sagittal slices. The per-pixel majority opinion of the five (5) radiologists established the ground truth for each anatomical structure. Specially, if at least 3 of the 5 radiologists labeled a pixel as belonging to a particular anatomical structure, the pixel was included. Otherwise, the pixel was excluded.
For the measurement, each radiologist used a commercial software tool to produce a standard set of areal, angular and linear measurements. The ground truth was established by taking the mean of the five (5) radiologists' measurements.
Acceptance Criteria and Study Results
Primary Endpoints:
- -For 4 measurements, the maximum Mean Absolute Error (MAE) as defined as the upper limit of the 95% confidence interval for MAE is below a predetermined allowable error limit (MAE limit) for each measurement listed below.
| Measurement | MAE limit |
|---|---|
| Dural Sac Area (Axial) | 20 mm2 |
| Spinal Canal Area (Axial) | 30 mm2 |
{11}------------------------------------------------
| Lordotic Angle (Sagittal) | 6° |
|---|---|
| Vertebral Body Slippage (Sagittal) | 2 mm |
Secondary Endpoints:
- For segmentations of 14 anatomical structures: the minimum Mean Dice Coefficient, defined as the lower limit of the 95% confidence interval for MDC, is above a predetermined allowable limit (MDC Limit) for each segmentation listed below.
| Anatomical structure | MDC limit |
|---|---|
| Artery (Axial) | 0.8 |
| Disc (Axial) | 0.7 |
| Disc (Sagittal) | 0.7 |
| Disc Material Outside IV Space (Axial) | 0.7 |
| Dural Sac (Axial) | 0.8 |
| Kidney (Axial) | 0.8 |
| Ligamentum Flavum (Axial) | 0.7 |
| Muscle (Axial) | 0.8 |
| Sacrum (Sagittal) | 0.8 |
| Spinal Canal (Axial) | 0.8 |
| Spinal Canal (Sagittal) | 0.8 |
| Vein (Axial) | 0.8 |
| Vertebral Arch (Axial) | 0.8 |
| Vertebral Body (Sagittal) | 0.8 |
- -For 10 measurements: the maximum Mean Absolute Error (MAE) as defined as the upper limit of the 95% confidence interval for MAE is below a predetermined allowable error limit (MAE limit) for each measurement listed below.
| Measurement | MAE limit |
|---|---|
| Anterior Disc Height (Sagittal) | 2 mm |
| Anterior Vertebral Body Height (Sagittal) | 2 mm |
| Dural Sac Anterior-Posterior Diameter(Axial) | 2 mm |
| Dural Sac Transverse Diameter (Axial) | 2 mm |
| Middle Disc Height (Sagittal) | 2 mm |
| Middle Vertebral Body Height (Sagittal) | 2 mm |
| Posterior Disc Height (Sagittal) | 2 mm |
| Posterior Vertebral Body Height (Sagittal) | 2 mm |
| Spinal Canal Anterior-Posterior Diameter(Axial) | 2 mm |
{12}------------------------------------------------
| Spinal Canal Transverse Diameter (Axial) | 2 mm |
|---|---|
| ------------------------------------------ | ------ |
Primary Endpoints Results:
| Measurement | Mean AbsoluteError (MAE) | 95% ConfidenceIntervalLower Bound | 95%ConfidenceIntervalUpper Bound | MAELimit | Success |
|---|---|---|---|---|---|
| Dural Sac Area(Axial) | 17.2 mm² | 16.5 mm² | 17.9 mm² | 20 mm² | Yes |
| Spinal CanalArea (Axial) | 23.3 mm² | 22.3 mm² | 24.3 mm² | 30 mm² | Yes |
| Lordotic Angle(Sagittal) | 3.4 ° | 2.9 ° | 3,9 ° | 6 ° | Yes |
| Vertebral BodySlippage(Sagittal) | 0.6 mm | 0,5 mm | 0,8 mm | 2 mm | Yes |
Secondary Endpoints Results:
| AnatomicalStructureSegmentation | Mean DiceCoefficient (MDC) | 95% ConfidenceIntervalLower Bound | 95% ConfidenceIntervalUpper Bound | MDCLimit | Success |
|---|---|---|---|---|---|
| Artery (Axial) | 0.866 | 0.861 | 0.871 | 0.8 | Yes |
| Disc (Axial) | 0.806 | 0.796 | 0.815 | 0.7 | Yes |
| Disc (Sagittal) | 0.914 | 0.910 | 0.918 | 0.7 | Yes |
| Disc MaterialOutside IVSpace (Axial) | 0.803 | 0.793 | 0.812 | 0.7 | Yes |
| Dural Sac(Axial) | 0.926 | 0.924 | 0.929 | 0.8 | Yes |
| Kidney (Axial) | 0.879 | 0.872 | 0.886 | 0.8 | Yes |
| LigamentumFlavum (Axial) | 0.740 | 0.736 | 0.744 | 0.7 | Yes |
| Muscle (Axial) | 0.946 | 0.945 | 0.947 | 0.8 | Yes |
| Sacrum(Sagittal) | 0.925 | 0.923 | 0.928 | 0.8 | Yes |
| Spinal Canal(Axial) | 0.942 | 0.941 | 0.944 | 0.8 | Yes |
| Spinal Canal(Sagittal) | 0.871 | 0.865 | 0.877 | 0.8 | Yes |
| Vein (Axial) | 0.821 | 0.815 | 0.827 | 0.8 | Yes |
| Vertebral Arch | 0.846 | 0.843 | 0.850 | 0.8 | Yes |
{13}------------------------------------------------
| (Axial) | |||||
|---|---|---|---|---|---|
| Vertebral Body(Sagittal) | 0.900 | 0.894 | 0.905 | 0.8 | Yes |
| Measurement | Mean AbsoluteError (MAE) | 95% confidenceintervalLower bound | 95% confidenceintervalUpper bound | MAElimit | Success |
|---|---|---|---|---|---|
| Anterior DiscHeight (Sagittal) | 1.3 mm | 1.18 mm | 1.32 mm | 2 mm | Yes |
| AnteriorVertebral BodyHeight (Sagittal) | 1.6 mm | 1.48 mm | 1.81 mm | 2 mm | Yes |
| Dural SacAnterior-PosteriorDiameter(Axial) | 1.49 mm | 1.46 mm | 1.52 mm | 2 mm | Yes |
| Dural SacTransverseDiameter(Axial) | 1.15 mm | 1.09 mm | 1.22 mm | 2 mm | Yes |
| Middle DiscHeight (Sagittal) | 1 mm | 0.95 mm | 1.04 mm | 2 mm | Yes |
| MiddleVertebral BodyHeight (Sagittal) | 1.3 mm | 1.21 mm | 1.33 mm | 2 mm | Yes |
| Posterior DiscHeight (Sagittal) | 1 mm | 0.95 mm | 1.07 mm | 2 mm | Yes |
| PosteriorVertebral BodyHeight (Sagittal) | 1.6 mm | 1.44 mm | 1.72 mm | 2 mm | Yes |
| Spinal CanalAnterior-PosteriorDiameter(Axial) | 0.81 mm | 1.78 mm | 0.83 mm | 2 mm | Yes |
| Spinal CanalTransverseDiameter(Axial) | 1.74 mm | 1.64 mm | 1.85 mm | 2 mm | Yes |
The RAI software was shown to produce segmentations and measurements accurate to within a prospectively-defined margin of error around the Ground Truth. This accuracy was preserved for all critical subgroups, including MRI scanner manufacturer, patient age, gender and race, except for the dural sac area measurement from MR studies conducted using Philips MR equipment and for patients in over 51 year-old age group.
{14}------------------------------------------------
Training, Testing and Validation Data Independence:
The RAI software machine learning algorithm training and testing data used during the algorithm development, as well as validation data used in the U.S. standalone software performance assessment study were all independent data sets.
7.6.Clinical Validation Study
No human clinical study was conducted to support the pre-market clearance.
8. Conclusions
The RemedyLogic AI MRI Lumbar Spine Reader ("RAI") software is as safe and effective as the predicate device. The subject device has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor differences between subject and predicate device in indications do not alter the intended use of the device and do not raise new or different questions regarding its safety and effectiveness when used as labeled.
The software verification and validation testing data, including the standalone software performance assessment study data, support the safety of the devices and demonstrate that the RAI software performs as intended in the specified use conditions.
Therefore, the RAI software is substantially equivalent.
N/A