K Number
K241106

Validate with FDA (Live)

Date Cleared
2024-05-22

(30 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ice Cooling IPL Home Use Hair Removal Device is an over-the-counter device indicated for the removal of unwanted hair such as but not limited to small as underarm and facial hair below the chin line and large areas such as legs.

The device is also indicated for the permanent reduction in hair regrowth, defined as the long term, stable reduction in the amount of hair regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

Device Description

The Ice Cooling IPL Home Use Hair Removal Device is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user. The device provides hair reduction using Intense Pulsed Light technology and it has the cooling function that effectively cools the skin.

It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the external power adapter and its IPL emission activation is by finger switch. The device contains a Xenon lamp and a skin sensor to detect appropriate skin contact. If the device is not properly and fully applied to the skin of the treatment area, the device will not emit light pulses. Besides, the device has the cooling function that will be activated throughout the whole hair removal process or when the user want to activated to provide users with a more comfortable experience.

AI/ML Overview

The provided document is a 510(k) summary for an Ice Cooling IPL Home Use Hair Removal Device. It focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed study proving the device meets specific performance acceptance criteria through clinical trials.

Therefore, it is not possible to fully answer all aspects of your request based on the provided text alone, as it lacks a standalone clinical study section with detailed acceptance criteria, sample sizes, expert involvement, and ground truth establishment for performance.

However, I can extract information related to the device's alignment with safety and performance standards which serve as a form of acceptance criteria in the context of a 510(k) submission.

Here's the information that can be gleaned from the document:

1. A table of acceptance criteria and the reported device performance

The document lists several international and FDA-recognized standards that the device conforms to. While these are not "performance" metrics in terms of hair reduction efficacy, they represent acceptance criteria for safety, electrical compatibility, usability, and photobiological safety.

Acceptance Criteria (Standards Conformance)Reported Device Performance (Conformance)
IEC 60601-1 (Basic Safety & Essential Performance)Conforms
IEC 60601-1-2 (Electromagnetic Disturbances)Conforms
IEC 60601-1-6 (Usability)Conforms
IEC 60601-1-11 (Home Healthcare Environment)Conforms
IEC 60601-2-57 (Non-laser light source for therapeutic/cosmetic use)Conforms
IEC 60601-2-83 (Home light therapy equipment)Conforms
IEC 62471 (Photobiological Safety)Conforms
IEC 62366-1 (Usability Engineering for Medical Devices)Conforms
ISO 10993-5 (In-Vitro Cytotoxicity)Conforms
ISO 10993-10 (Irritation and Skin Sensitization)Conforms
ISO 10993-23 (Tests for Irritation)Conforms
Software Verification and ValidationPerforms as intended (according to FDA Guidance)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. The submission focuses on non-clinical engineering and biocompatibility testing against standards, and comparison to predicate devices, rather than a clinical performance study with a specific test set of human subjects.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable and therefore not provided, as the document does not describe a clinical study with a ground truth established by experts.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable and therefore not provided, as the document does not describe a clinical study with a test set requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable and therefore not provided. This device is an IPL hair removal device, not an AI-assisted diagnostic tool that would involve human readers interpreting cases.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable and therefore not provided. The device is a hardware product for hair removal, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical tests (electrical safety, photobiological safety, biocompatibility, software V&V), the "ground truth" or reference is adherence to the specified international and FDA-recognized standards. For the claimed "permanent reduction in hair regrowth," the document states it is "defined as the long term, stable reduction in the amount of hair regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime." However, explicit outcomes data from a clinical trial to prove this ground truth for the subject device is not presented in this summary, rather the substantial equivalence to predicate devices that have this indication is leveraged.

8. The sample size for the training set

This information is not applicable and therefore not provided. There is no mention of machine learning or AI models with training sets for this device.

9. How the ground truth for the training set was established

This information is not applicable and therefore not provided, as there is no training set for a machine learning model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

Shenzhen Mlay Intelligent Technology Co., Ltd. Libin Liu General Manager 302,401, Building 1, No. 2, Xiashiwei Road, Huaide Community Fuyong Street, Baoan District Shenzhen. 518000 China

Re: K241106

Trade/Device Name: Ice Cooling IPL Home Use Hair Removal Device (Model(s): T10B, T10C. T15A. T17A, T18A, T14A, T16A, T19A) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: April 22, 2024 Received: April 22, 2024

Dear Libin Liu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Yan Fu -S
Digitally signed by Yan Fu -S
Date: 2024.05.22 13:31:27
-04'00'

for

Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices

{2}------------------------------------------------

Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K241106

Device Name

Ice Cooling IPL Home Use Hair Removal; Device Model(s): T10B, T10C, T15A, T17A, T18A, T19A

Indications for Use (Describe)

The Ice Cooling IPL Home Use Hair Removal Device is an over-the-counter device indicated for the removal of unwanted hair such as but not limited to small as underarm and facial hair below the chin line and large areas such as legs.

The device is also indicated for the permanent reduction in hair regrowth, defined as the long term, stable reduction in the amount of hair regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Research Use (21 CFR 201.3 Subject to)

Non-Therapeutic Use (21 CFR 201.3 Subject to)
__ Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary - K241106 "510(k) Summary" as required by 21 CFR Part 807.92.

1. Submitter Information

Shenzhen Mlay Intelligent Technology Co., Ltd. 302,401, Building 1, No. 2, Xiashiwei Road, Huaide Community, Fuyong Street, Baoan District, Shenzhen, Guangdong, China

Post Code: 518000

LIU LIBIN General Manager Tel.: +86 150 6071 0113 Email: amy@mlays.com

2. Device Information

Trade name:Ice Cooling IPL Home Use Hair Removal Device
Model(s):T10B, T10C, T15A, T17A, T18A, T14A, T16A, T19A
Common Name:Light Based Over-The-Counter Hair Removal
Regulation Class:Class II
Product Code:OHT
Review Panel:General & Plastic Surgery
Regulation Number:21CFR 878.4810
Regulation Description:Laser surgical instrument for use in general and plasticsurgery
Submission Type:Special 510(k)

3. Predicate Device

Primary predicate device:

ManufacturerPredicate Device510k NumberDecision Date
Shenzhen MlayIntelligentTechnology Co.,LtdHome Use Hair RemovalDevice (Model(s):T10A,T10B, T10C, T10D, T11A,T15A, T17A, T18A, T14A,T16A, T19A)K240264March 8, 2024

Predicate device:

ManufacturerPredicate Device510k NumberDecision Date
Shenzhen UlikeSmart ElectronicsCo.,LtdIPL Hair Removal DeviceK223618Feb. 28, 2023
Shenzhen UlikeSmart ElectronicsIPL Hair Removal DeviceK221002June 01, 2022

{5}------------------------------------------------

Co.,Ltd
Shen ZhenCosBeauty Co.,LtdIPL Hair Removal DeviceJoy VersionK173813Sept. 07, 2018

4. Device Description

The Ice Cooling IPL Home Use Hair Removal Device is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user. The device provides hair reduction using Intense Pulsed Light technology and it has the cooling function that effectively cools the skin.

It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the external power adapter and its IPL emission activation is by finger switch. The device contains a Xenon lamp and a skin sensor to detect appropriate skin contact. If the device is not properly and fully applied to the skin of the treatment area, the device will not emit light pulses. Besides, the device has the cooling function that will be activated throughout the whole hair removal process or when the user want to activated to provide users with a more comfortable experience.

5. Indications for Use

The Ice Cooling IPL Home Use Hair Removal Device is an over-the-counter device indicated for the removal of unwanted hair such as but not limited to small areas such as underarm and facial hair below the chin line and large areas such as legs. The device is also indicated for the permanent reduction in hair regrowth, defined as the long term, stable reduction in the amount of hair regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

6. Comparison of Technological Characteristics With the Predicate Device

The Ice Cooling IPL Home Use Hair Removal Device has the same intended use, mode of action and operational characteristics as the models T10B, T15A, T17A, T18A, T14A, T16A, T19 in the predicate devices. The modification between the subject device and the listed predicate devices do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the primary predicate device and predicate device for its intended use. Therefore, the Ice Cooling IPL Home Use Hair Removal Device may be found substantially equivalent to its predicate device.

Ice Cooling IPL Home Use Hair Removal Device is compared with the primary predicate device and predicate device as belows:

{6}------------------------------------------------

ComparisonElementsSubject DevicePrimary Predicate Device K240264Comparison
K NumberPendingK240264/
Trade nameIce Cooling IPL Home Use Hair RemovalDevice, Model(s): T10B, T10C, T15A,T17A, T18A, T14A, T16A, T19AHome use hair removal device, Model(s):T10A, T10B, T10C, T10D, T11A, T15A,T17A, T18A, T14A, T16A, T19ASimilarNote¹
LocationforuseOTCOTCSE
Indications forUseThe Ice Cooling IPL Home Use HairRemoval Device is an over-the-counterdevice indicatedfor the removalofunwanted hair such as but not limited tosmall areas such as underarm and facial hairbelow the chin line and large areas such aslegs.The device is also indicated for thepermanent reduction in hair regrowth,defined as the long term, stable reduction inthe amount of hair regrowing whenmeasured at 6, 9 and 12 months after thecompletion of a treatment regime.The Home use hair removal device is anover-the-counter device intended forremoval of unwanted hair such as but notlimited to small areas such as underarm andfacial hair below the chin line and largeareas such as legs, in patients withFitzpatrick Skin Phototypes I-V.SimilarNote2
Ice coolingfunctionModel(s): T10B, T10C, T15A, T17A,T18A, T14A, T16A, T19A have coolingfunction(1)Model(s): T10A, T10D, T11Ado nothave cooling function.(2)Model(s): T10B, T10C, T15A, T17A,T18A, T14A, T16A, T19A have coolingfunctionSENote3
Coolingpanel'smaterialT10B, T10C: artificial sapphireT15A, T17A, T18A, T14A, T16A, T19A:5052 aluminum alloyT10A,T10D,T 11A: have not coolingfunctionT10B, T10C: artificial sapphireT15A, T17A, T18A, T14A, T16A, T19A:5052 aluminum alloy
EnergymediumXenon Arc FlashlampXenon Arc FlashlampSE
Light SourceIntense Pulsed LightIntense Pulsed LightSE
PulsingcontrolFinger switchFinger switchSE
DeliverydeviceDirect illumination tissueDirect illumination tissueSE
Energy densityT10B, T10C, T15A, T17A, T18A: $1.37 ~ 4.11J/cm²$T14A,T16A,T19A: $1.84~ 5.23J/cm²$T10A, T10B, T10C, T10D, T11A, T15A,T17A, T18A: $1.374.11J/cm²$T14A,T16A,T19A: $1.84 5.23J/cm²$SE
ComparisonElementsSubject DevicePrimary Predicate Device K240264Comparison
K NumberPendingK240264/
Treatmentenergy level(1)Model(s)T17A,T18A:have 3 levels;Level 1 to Level 3(2)Model(s)T10B, T10C, T15A, T14A,T16A, T19A: have 5 levels; Level 1 toLevel 5(1)Model(s)T11A,T17A,T18A:have 3levels; Level 1 to Level 3(2)Model(s)T10A, T10B, T10C, T10D,T15A, T14A, T16A, T19A: have 5 levels;Level 1 to Level 5SENote3
Spot sizeT10B,T10C,T15A,T17A,T18A: 3.5cm²T14A,T16A,T19A: 3.9cm²T10A, T10B, T10C, T10D, T11A, T15A,T17A, T18A: 3.5cm²T14A,T16A,T19A: 3.9cm²SE
Pulse durationT10B, T10C,T15A, T17A, T18A:4.08.0msT14A,T16A,T19A: 0.52.5msT10A, T10B, T10C, T10D, T11A, T15A,T17A, T18A: 4.08.0msT14A,T16A,T19A: 0.52.5msSE
Pulse intervalT10B, T10C: 0.52.8sT15A, T17A, T18A: 0.72.5sT14A,T16A,T19A: 0.5~1.5sT10A, T10B, T10C, T10D: 0.52.8sT11A: 1.02.5sT15A, T17A, T18A: 0.72.5sT14A,T16A,T19A: 0.51.5sSE
Treatmentwindow to bedesigned asremovable ornon-removable(1)Model(s) T10B, T10C,T15A, T17A,T18A:non-removable(2)Model(s)T14A,T16A,T19A:removable(1)Model(s)T10A, T10B, T10C, T10D,T11A,T15A,T17A,T18A:non-removable(2)Model(s)T14A,T16A,T19A:removableSENote3
DesignSoftwareFeaturesMaterialsEnergy sourcePrinciple ofoperationThe models applied for this time are exactly the same as these models T10B, T10C, T15A,T17A, T18A, T14A, T16A, T19A in K240264.SE

{7}------------------------------------------------

{8}------------------------------------------------

Comparison ElementsSubject DevicePredicate Device 1Predicate Device 2Predicate Device 3Comparison
K NumberPendingK223618K221002K173813/
Trade nameIce Cooling IPL Home Use HairRemoval Device, Model(s):T10B, T10C, T15A, T17A,T18A, T14A, T16A, T19AIPL Hair RemovalDeviceIPL Hair RemovalDevice (UI04A, UI04B,UI04C)IPL Hair RemovalDevice Joy Version, CB- 027SimilarNote1
Location for useOTCOTCOTCOTCSE
Indications for UseThe Ice Cooling IPL Home UseHair Removal Device is anover-the-counter deviceindicated for the removal ofunwanted hair such as but notlimited to small areas such asunderarm and facial hair belowthe chin line and large areassuch as legs.The device is also indicated forthe permanent reduction in hairregrowth, defined as the longterm, stable reduction in theamount of hair regrowing whenmeasured at 6, 9 and 12 monthsIPL Hair Removal Deviceis indicated for the removalof unwanted hair. Thedevice is also indicated forthe permanent reduction inhair regrowth, defined asthe long-term, stablereduction in the number ofhairs regrowing whenmeasured at 6, 9 and 12months after the completionof a treatment regime.IPL Hair RemovalDevice is indicated forthe removal ofunwanted hair. Thedevice is also indicatedfor the permanentreduction in hairregrowth, defined as thelong-term, stablereduction in the numberof hairs regrowingwhen measured at 6, 9and 12 months after thecompletion of atreatment regime.The IPL Hair RemovalDevice Joy Version isindicated for the removal ofunwanted hair. The device isalso indicated for thepermanent reduction in hairregrowth, defined as thelong-term, stable reduction inthe number of hairsregrowing when measured at6, 9 and 12 months after thecompletion of a treatmentregime. The device is usedfor adults with Fitzpatrickskin types I - IV.SimilarNote2
Comparison ElementsSubject DevicePredicate Device 1Predicate Device 2Predicate Device 3Comparison
K NumberPendingafter the completion of atreatment regime.K223618K221002K173813/
Applicable skinFitzpatrick Skin Types I-VFitzpatrick Skin Types I-VFitzpatrick Skin TypesI-VFitzpatrick skin types I - IVSimilar
Treatment areaLarge areas(e.g. legs) and smallareas(e.g.underarm, facial hairbelow the chin line)Large areas (e.g. arms, legs,chest) and small areas (e.g.lip)Large areas (e.g.arms,legs, chest) and smallareas (e.g. fingers, lip)Large areas (legs, arms, backand abdomen), face(upper lip, chin andsideburns)Similar
Output specification
Light sourceIntense Pulsed LightIntense Pulsed LightIntense Pulsed LightIntense Pulsed LightSE
Energy mediumXenon Arc FlashlampXenon Arc FlashlampXenon Arc FlashlampXenon Arc FlashlampSE
Wavelength range510~1200nm560~1200nm550~1200nm510~1200nmSimilarNote4
Energy densityT10B, T10C, T15A, T17A,T18A: 1.37~4.11J/cm²T14A,T16A, T19A: 1.84 ~5.23J/cm²3~6J/cm²3.03~5.3J/cm²1.8~5.1J/cm²SimilarNote5
Comparison ElementsSubject DevicePredicate Device 1Predicate Device 2Predicate Device 3Comparison
K NumberPendingK223618K221002K173813/
Output energyT10B, T10C, T15A, 17A,T18A:612JT14A, T16A, T19A: 9 17J9.9~19.8J10.0~17.5JBodylampcartridge11.7722.21J5101200nmSimilarNote5
Faciallampcartridge3.657.04J5121197nm
Bikinilampcartridge3.847.22J5111200nm
Spot sizeT10B, T10C, T15A, 17A,T18A:3.5cm²T14A, T16A, T19A: 3.9cm²3.3cm²3.3cm²Body: 4.2cm²Bikini and face: 2.0cm²SimilarNote6
Pulse durationT10B, T10C, T15A, 17A,T18A:4.08.0msT14A, T16A, T19A: 0.52.5ms1~7ms7~10ms9.2~11.2msSimilarNote7
Pulsing controlFinger switchFinger switchFinger switchFinger switchSE
Delivery deviceDirect illumination tissueDirect illumination tissueDirect illumination tissueDirect illumination tissueSE

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

Comparison ElementsSubject DevicePredicate Device 1Predicate Device 2Predicate Device 3Comparison
K NumberPendingK223618K221002K173813/
Output intensity levelT17A,T18A: 3 levelsT10B, T10C, T15A, T14A,T16A, T19A: 5 levels3 levels5 levels5 levelsSimilar
Software/ Firmware/Microprocess or ControlYesYesYesYesSE
Additional features
Electrical safetyIEC 60601-1IEC 60601-1-2IEC 60601-1-6IEC 60601-1-11IEC 60601-2-57IEC 60601-2-83ANSI AAMI ES60601-1IEC 60601-1-2IEC 60601-1-11IEC 60601-2-57IEC 60601-2-83IEC 60601-1-2IEC 60601-1-11ANSI AAMIES60601-1IEC 60601-2-57IEC 60601-2-83IEC 60601-1IEC 60601-1-2IEC 60601-2-57Similar

{12}------------------------------------------------

Comparison ElementsSubject DevicePredicate Device 1Predicate Device 2Predicate Device 3Comparison
K NumberPendingK223618K221002K173813/
Eye safetyIEC62471IEC 62471IEC 62471IEC 62471SE
BiocompatibilityISO 10993-5ISO 10993-10ISO 10993-12ISO 10993-5ISO 10993-10ISO 10993-5ISO 10993-10ISO 10993-5ISO 10993-10Similar

{13}------------------------------------------------

Notel: These model(s): T10B, T10C, T15A, T17A, T18A, T14A, T16A, T19A all provide hair reduction using Intense Pulsed Light technology and it has the cooling function that effectively cools the skin, so the device name changes from " Home use hair removal device" to "Ice Cooling IPL Home Use Hair Removal Device". The user can get the information about the function of the device by the device name, which is convenient for the user to self-select for the device.

Note2:The Indications for Use descriptions of the subject device and the predicate device are essentially the same, except that the wording is different. 1) The user manual and the outer packaging label specify the hair colors that the device can be used to treat, so the words "in patients with Fitzpatrick Skin Phototypes I-V" have been deleted. 2) In order to allow the user to clearly understand the results they can achieve if they use the device for a long term, the following words have been added: "The device is also indicated for the permanent reduction in hair regrowth, defined as the long term, stable reduction in the amount of hair regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.". The technical parameters of the subject device are similar to those of the predicate device. The results of the comparison show that the subject device can achieve this result.

Note3:The same models are totally same.

Note4:Though the wavelength of subject device is a little different from the predicate device, they all comply with IEC 60601-2-57/IEC 60601-2-83 and IEC 62471 requirements, so this difference will not raise any safety or effectiveness issue.

Note': Though the energy density and the output energy of subject device is a little different from the predicate devices, the energy density and the output energy of subject device is within the range of the maximum value of the predicate devices, and they all comply with IEC 60601-2-57/IEC 60601-2-83 and IEC 62471 requirement, so this difference will not raise any safety or effectiveness issue.

Note6:There is minor difference in spot size between the subject device and the predicate devices. The spot size is related to energy density and since the difference in energy density is not significant as explained in note5, so this difference will not raise any safety or effectiveness issue.

Note : Though the pulse duration of subject device is a little different from the predicate device, it's basically within the range of the predicate devices, and the subject device comply with IEC 60601-2-57/IEC 60601-2-83 and IEC 62471 requirement, so this difference will not raise any safety or effectiveness issue.

{14}------------------------------------------------

7. Non-clinical studies and tests performed

Non-clinical tests have been conducted to verify that the Ice Cooling IPL Home Use Hair Removal Device meets all design specifications which supports the conclusion that it's Substantially Equivalent (SE) to the predicate devices. The testing results demonstrate that the subject device conforms to the following performance standards:

  • A IEC 60601-1, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
  • IEC 60601-1-2, Medical electrical equipment Part 1-2: General requirements for A basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests
  • IEC 60601-1-6, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability

  • A IEC 60601-1-11, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
  • A IEC 60601-2-57, Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic diagnostic monitoring and cosmetic/aesthetic use
  • IEC 60601-2-83, Medical electrical equipment Part 2-83: Particular A requirements for the basic safety and essential performance of home light therapy equipment
  • IEC 62471, Photobiological safety of lamps and lamp systems

  • IEC 62366-1, Medical devices Part 1: Application of usability engineering to > medical devices

The device has been tested for biocompatibility for conformance to the following performance standards:

  • A ISO 10993-5, Biological Evaluation of Medical Devices - Part 5: Tests for InVitro Cytotoxicity
  • ISO 10993-10, Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization

  • ISO 10993-23, Biological evaluation of medical devices - Part 23: Tests for irritation

Software Verification and Validation

Software verification and validation was performed, and it was demonstrated that the software performs as intended according to the FDA Guidance for the Content of Premarket Submissions for Device Software Functions.

{15}------------------------------------------------

8. Conclusion

Based on the above analysis, it can be concluded that the subject device Ice Cooling IPL Home Use Hair Removal Device meets the product requirements, which is based on the primary predicate device , and substantially equivalent to these predicate devices for permanent reduction in hair regrowth. Therefore, the modification resulted in a device that performs within the same specifications of the primary predicate device and the predicate devices, and is therefore substantially equivalent.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.