K Number
K241052
Manufacturer
Date Cleared
2024-10-04

(170 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Bladeless Trocar, Bladed Trocar, and Hasson Trocar have application in a variety of laparoscopic procedures to provide a port of entry for laparoscopic instruments.

Device Description

The Disposable Surgical Trocar is composed of two parts, one obturator and one guide tube (cannula). The obturator of the Bladeless Trocar is plastic allowing the surgeon to dilate between tissues. The obturator of Bladed Trocar is made of PC and a small blade made of stainless steel, which is released when the surgeon presses the trocar to help with cutting. The Disposable Surgical Trocar Kit, bladed or bladeless, has one obturator and two cannulas or two obturators and four cannulas. The device is for single use only. This device is sterilized with ethylene oxide and has a shelf life of 3 years.

AI/ML Overview

This is an FDA 510(k) summary for a medical device called "Bladeless Trocar; Bladed Trocar; Hasson Trocar." The document describes the device, its intended use, and compares it to a legally marketed predicate device (K141594).

However, your request asks for "acceptance criteria and the study that proves the device meets the acceptance criteria" specifically in the context of an AI/ML medical device submission. The provided document is for a mechanical surgical instrument (trocar) and does not involve AI/ML. Therefore, the requested information categories concerning AI/ML performance (e.g., sample size for test set, data provenance, expert ground truth, MRMC studies, standalone performance, training set details) are not applicable to this submission.

The document states:

"The subject device trocar models (Bladeless Trocar, Hasson Trocar) are manufactured by the same company that manufactures those of the predicate device (Unimicro) and are identical to those cleared under K141594. As such, performance testing is not required."

This statement indicates that the regulatory acceptance is based on the substantial equivalence to a predicate device, specifically because the new device is identical and manufactured by the same company as the predicate. Therefore, no new performance studies were conducted or required to demonstrate that the device meets acceptance criteria.

To specifically address the prompts given the available information:

  1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: Substantial Equivalence to predicate device K141594. The key criteria are identical design, materials, manufacturing process, and intended use to a previously cleared device.
    • Reported Device Performance: Not applicable, as performance testing was explicitly stated as "not required" due to the identical nature to the predicate device. The performance is assumed to be equivalent to the predicate device.
  2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable as no performance study was conducted.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable as no performance study was conducted.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable as no performance study was conducted.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is a non-AI device.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a non-AI device.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not applicable as no performance study was conducted. Regulatory acceptance is based on substantial equivalence to the predicate device.
  8. The sample size for the training set

    • Not applicable. This is a non-AI device.
  9. How the ground truth for the training set was established

    • Not applicable. This is a non-AI device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and consists of the letters "FDA" stacked on top of the words "U.S. FOOD & DRUG ADMINISTRATION".

October 4, 2024

Advamedix, GmbH % Marianne Jacklyn President and Principal Consultant TRAXN Consulting LLC 1235 Swift Shore Circle West Linn, Oregon 97068

Re: K241052

Trade/Device Name: Bladeless Trocar; Bladed Trocar; Hasson Trocar Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ Dated: September 5, 2024 Received: September 6, 2024

Dear Marianne Jacklyn:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long H. Chen -> Digitally signed by Long H.Chen -S
Digitally signed Date: 2024.10.04 17:26:30 -04'00'

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

or Use

Submission Number (if known)

K241052

Device Name

Bladeless Trocar;

Bladed Trocar;

Hasson Trocar

Indications for Use (Describe)

The Bladeless Trocar, Bladed Trocar, and Hasson Trocar have application in a variety of laparoscopic procedures to provide a port of entry for laparoscopic instruments.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92

1 Contact Details: 21 CFR 807.92(a)(1)

Applicant NameAdvamedix, GmbH
Applicant AddressGreschbachstr 3/3, OG Karlsruhe 76229Germany
Applicant Contact TelephonePhone: +49 721 94552
Applicant ContactDr. Islam Abed, Director
Applicant Contact Emailabed@advamedix.net
Correspondent NameTRAXN Consulting LLC
Correspondent Address1235 Swift Shore Circle, West Linn, Oregon 97068United States
Correspondent Contact Telephone(503) 729-9633
Correspondent ContactMs. Marianne D. Jacklyn
Correspondent Contact Emailmjacklyn@traxnconsulting.com

2 Device Name: 21 CFR 807.92(a)(2)

Trade NameBladeless Trocar, Bladed Trocar, Hasson Trocar
Common NameTrocar
Classification NameEndoscope and Accessories
Regulation Number876.1500
Product Code(s)GCJ

Legally Marketed Predicate Devices: 21 CFR 807.92(a)(3) ന

K141594, Unimicro Trocar Kit, models: Auto-Locking Trocar, Hasson Trocar, Bladeless Trocar

Device Description Summary: 21 CFR 807.92(a)(4) 4

The Disposable Surgical Trocar is composed of two parts, one obturator and one guide tube (cannula). The obturator of the Bladeless Trocar is plastic allowing the surgeon to dilate between tissues. The obturator of Bladed Trocar is made of PC and a small blade made of stainless steel, which is released when the surgeon presses the trocar to help with cutting.

{5}------------------------------------------------

The Disposable Surgical Trocar Kit, bladed or bladeless, has one obturator and two cannulas or two obturators and four cannulas.

The device is for single use only. This device is sterilized with ethylene oxide and has a shelf life of 3 years.

Intended Use/Indications for Use: 21 CFR 807.92(a)(5) 5

The Bladeless Trocar, Bladed Trocar, and Hasson Trocar have application in a variety of laparoscopic procedures to provide a port of entry for laparoscopic instruments.

6 Indications for Use Comparison: 21 CFR 807.92(a)(5)

The indications for use are the same for the proposed devices and the predicate devices.

Device & PredicateDevice(s)K241052K141594
Principles ofOperationDuring the operation, the trocar sleeveand the obturator are used together.The surgeon uses the obturator toexpand the incision of the abdomenand penetrates the trocar sleevethrough the abdominal surface of thehuman body into the abdominal cavity,thereby delivering gas to the abdominalcavity and establishing a path of entryfor surgical instruments.During the operation, the trocar sleeveand the obturator are used together.The surgeon uses the obturator toexpand the incision of the abdomenand penetrates the trocar sleevethrough the abdominal surface of thehuman body into the abdominal cavity,thereby delivering gas to the abdominalcavity and establishing a path of entryfor surgical instruments.
Main Componentsobturator, cannulaobturator, cannula
ModelsBladelessBladedHassonBladelessBladedHasson
Dimensions(Overall)Diameter: 5-12 mmLength: 100 mmDiameter: 5-12 mmLength: 70 - 120 mm

7 Technological Comparison: 21 CFR 807.92(a)(6)

Non-clinical and/or Clinical Tests Summary & Conclusions: 21 CFR 807.92(b) 8

The subject device trocar models (Bladeless Trocar, Hasson Trocar) are manufactured by the same company that manufactures those of the predicate device (Unimicro) and are identical to those cleared under K141594. As such, performance testing is not required.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.