K Number
K241031

Validate with FDA (Live)

Date Cleared
2024-07-12

(88 days)

Product Code
Regulation Number
882.4560
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TMINI™ Miniature Robotic System is indicated as a stereotaxic instrumentation system for total knee replacement (TKA) surgery. It is to assist the surgeon by providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of knee implant components. The robotic device placement is performed relative to anatomical landmarks as recorded using the system intraoperatively and based on a surgical plan determined preoperatively using CT based surgical planning tools. It includes a handheld robotic device, an optical sensor navigation system and accessories, software system, surgical instruments and accessories.

The robotic device placement is performed relative to anatomical landmarks as recorded using the system intraoperatively and based on a surgical plan determined preoperatively using CT based surgical planning tools.

It includes a handheld robotic device, an optical sensor navigation system and accessories, surgical instruments and accessories.

The targeted population has the same characteristics as the population that is suitable for the implant(s) compatible with the TMINI™ Miniature Robotic System. The TMINI™ Miniature Robotic System is to be used with the following knee replacement system(s) in accordance with the indications and contraindications:

  • · Enovis™ EMPOWR Knee System®
  • · Ortho Development® BKS® and BKS TriMax® Knee System
  • · Total Joint Orthopedics Klassic® Knee System
  • · United® U2™ Knee Total Knee System
Device Description

The TMINI 1.1 like its predicate, the TMINI™ Miniature Robotic System consists of three primary components: a three-dimensional, graphical, Preoperative Planning Workstation (TPLAN Planning Station), an Optical Tracking Navigation Console (TNav) and a robotically controlled hand-held tool (TMINI Robot) that assists the surgeon in preparing the bone for implantation of TKA components.

The TPLAN Planning Station uses preoperative CT scans of the operative leg to create 3D surface models for case templating and intraoperative registration purposes. The Planning Workstation contains a library of 510(k) cleared knee replacement implant(s) available for use with the system. The surgeon can select an implant model from this library. The planner/surgeon can manipulate the 3D representation of the implant in relation to the bone model to optimally place the implant. The surgeon reviews and approves the case plan once the surgeon is satisfied with the implant selection, location and orientation. The data from the approved plan is written to a file that is used to guide the robotically controlled hand-held tool.

The hand-held robotic tool is optically tracked relative to optical markers placed in both the femur and tibia and articulates in two degrees-of-freedom, allowing the user to place bone pins in a planar manner in both bones. Mechanical quides are clamped to the bone pins, resulting in subsequent placement of cut slots and drill guide holes such that the distal femoral and proximal tibial cuts can be made in the pre-planned positions and orientations, and such that the implant manufacturer's multi-planer cutting block can be placed relative to drilled distal femoral pilot holes. If the surgeon needs to change the plan during surgery, it can be changed intraoperatively.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the TMINI™ Miniature Robotic System (TMINI 1.1) based on the provided text, recognizing that specific numerical values for acceptance criteria and detailed study designs are often proprietary and not fully disclosed in 510(k) summaries.

Important Note: The provided text is a 510(k) summary for a modified device (TMINI 1.1), asserting substantial equivalence to a previously cleared predicate device (TMINI™ Miniature Robotic System, K232802). As such, it primarily refers to the testing methods and acceptance criteria being similar to those of the predicate, rather than detailing new, distinct acceptance criteria and studies for the TMINI 1.1 that would demonstrate its initial safety and effectiveness from scratch. The text emphasizes that the modifications did not alter indications for use or raise new safety/effectiveness questions, and thus, previous testing is largely considered sufficient by extension.


1. Table of Acceptance Criteria and Reported Device Performance

Given the nature of the 510(k) summary for a modified device, specific quantitative acceptance criteria are not explicitly stated for all tests. Instead, the document consistently states that the device "Passed" various tests, implying it met the established (but not detailed) criteria.

Test CategoryAcceptance Criteria (Implied from text)Reported Device Performance
Performance Testing
Full System Run Through TestingMet all test criteria and specifications (similar to predicate)Passed
Cutting Accuracy - Pin & Block PlacementMet all test criteria and specifications (similar to predicate)Passed
Cadaver Lab Validation TestingMet all test criteria (similar to predicate)Passed
System Gap Balance AccuracyMet qualitative surgeon assessment testing and quantitative measurement accuracy testing (for TMINI 1.1) Predicate only listed qualitative surgeon assessment testingPassed
User Needs Validation TestingMet all test criteria (similar to predicate)Passed
Usability TestingMet all test criteria (similar to predicate)Passed
Software TestingMet all test criteria (similar to predicate)Passed
Biocompatibility Testing(Consistent with ISO 10993 requirements for patient-contacting materials)
- CytotoxicityMet relevant ISO 10993 criteriaPassed
- SensitizationMet relevant ISO 10993 criteriaPassed
- Intracutaneous ReactivityMet relevant ISO 10993 criteriaPassed
- Acute Systemic ToxicityMet relevant ISO 10993 criteriaPassed
- PyrogenicityMet relevant ISO 10993 criteriaPassed

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the exact sample sizes for each test set.

  • For "Cadaver Lab Validation Testing," it implies that cadaver models were used, suggesting a prospective design for those specific tests.
  • The data provenance (country of origin) is not mentioned.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: Not explicitly stated.
  • Qualifications of Experts: For "System Gap Balance Accuracy," the text mentions "qualitative surgeon assessment testing." This implies that qualified surgeons were involved in assessing the device's performance, but their specific number and detailed qualifications (e.g., years of experience, specialization) are not provided.

4. Adjudication Method for the Test Set

The document does not describe any specific adjudication method (like 2+1 or 3+1 consensus) for establishing ground truth or evaluating test results. It states that tests "Passed" or "Met" criteria, implying a direct comparison against predefined standards.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No, an MRMC comparative effectiveness study is not mentioned in the provided text. The document focuses on demonstrating substantial equivalence of a robotic surgical system (an instrumentation system) to a predicate, not on a human-reader performance comparison.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • The TMINI™ Miniature Robotic System (TMINI 1.1) is described as an "instrumentation system" that "assists the surgeon." It includes "software-defined spatial boundaries" and a "robotically controlled hand-held tool."
  • While software testing was performed ("Software Testing: Passed"), the overall device function inherently involves human interaction (the surgeon). Therefore, a purely standalone algorithm-only performance as a diagnostic or analytical tool (without any human in the loop for the critical surgical actions) seems unlikely and is not explicitly stated. The described functions (planning, robot guidance, navigation) are all in support of a human-performed surgery.

7. The Type of Ground Truth Used

  • For "Cutting Accuracy - Pin & Block Placement Accuracy," "Cadaver Lab Validation Testing," and "System Gap Balance Accuracy," the ground truth implicitly refers to objective physical measurements and surgical assessments (e.g., accuracy of pin/block placement against a programmed plan, quantitative gap measurements, qualitative surgeon assessment of balance). These are proxies for optimal surgical outcomes and precise execution.
  • The system uses "CT images... to create a 3D model of the bone for surgical planning," and "preoperatively determined landmarks are compared to intraoperatively identified landmarks to complete patient bone registration." This suggests that the ground truth for planning and navigation largely relies on preoperative imaging and anatomical landmarks.
  • No mention of pathology or long-term outcomes data being used as ground truth for this particular submission.

8. The Sample Size for the Training Set

  • The document does not provide information regarding a training set sample size. This submission is for a modification to an existing device, and the focus is on verification and validation testing, not typically on reiterating details of the initial development or training data for the core algorithms (which would have been part of the predicate device's clearance).

9. How the Ground Truth for the Training Set Was Established

  • As the document does not discuss a training set for the current submission, there is no information on how ground truth for any potential training set was established. The existing (predicate) device's software and algorithms would have had their own development and training, but those details are not part of this modification summary.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 12, 2024

Think Surgical, Inc. Meliha Mulalic Vice President. OA/RA 47201 Lakeview Blvd Fremont, California 94538

Re: K241031

Trade/Device Name: TMINI Miniature Robotic System (TMINI 1.1) Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: May 10, 2024 Received: May 13, 2024

Dear Meliha Mulalic:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerelv.

Shumaya Ali-S
Shumaya Ali, M.P.H.

Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K241031

Device Name TMINI™ Miniature Robotic System (TMINI 1.1)

Indications for Use (Describe)

TMINI™ Miniature Robotic System is indicated as a stereotaxic instrumentation system for total knee replacement (TKA) surgery. It is to assist the surgeon by providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of knee implant components. The robotic device placement is performed relative to anatomical landmarks as recorded using the system intraoperatively and based on a surgical plan determined preoperatively using CT based surgical planning tools. It includes a handheld robotic device, an optical sensor navigation system and accessories, software system, surgical instruments and accessories.

The robotic device placement is performed relative to anatomical landmarks as recorded using the system intraoperatively and based on a surgical plan determined preoperatively using CT based surgical planning tools.

It includes a handheld robotic device, an optical sensor navigation system and accessories, surgical instruments and accessories.

The targeted population has the same characteristics as the population that is suitable for the implant(s) compatible with the TMINI™ Miniature Robotic System. The TMINI™ Miniature Robotic System is to be used with the following knee replacement system(s) in accordance with the indications and contraindications:

  • · Enovis™ EMPOWR Knee System®
  • · Ortho Development® BKS® and BKS TriMax® Knee System
  • · Total Joint Orthopedics Klassic® Knee System
  • · United® U2™ Knee Total Knee System
Type of Use (Select one or both, as applicable)
-------------------------------------------------
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Think Surgical. The logo features a blue sphere with a white reflection at the top, resembling a stylized "T". Below the sphere, the word "THINK" is written in bold, black capital letters, with the "I" replaced by a blue, downward-pointing triangle. Underneath "THINK", the word "SURGICAL" is written in smaller, gray capital letters.

510(k) SUMMARY

Applicant Information:

Owner Name:THINK Surgical, Inc.
Address:47201 Lakeview Blvd., Fremont, CA 94538
Phone number:408-921-5648
Fax number:510-249-2396
Establishment Registration Number:3000719653
Contact Person:Meliha Mulalic
Date Prepared:15 April 2024
Device Information:
Device Classification:Class II
Trade Name:TMINITM Miniature Robotic System
Common name:Orthopedic Stereotaxic Instrument
Classification name:Stereotaxic Instrument
Regulation number:882.4560
Product Code:OLO

Predicate Device:

The TMINI™ Miniature Robotic System (TMINI 1.1) is substantially equivalent in intended use, Indications for Use, design, materials, technology, operational principles and performance to the predicate, TMINI™ Miniature Robotic System, cleared via K232802.

Device Modification:

The purpose of this submission is to introduce a series of incremental modifications to the TMINI™ Miniature Robotic System to enhance efficiency, consistency and ease of use.

Device Description:

The TMINI 1.1 like its predicate, the TMINI™ Miniature Robotic System consists of three primary components: a three-dimensional, graphical, Preoperative Planning Workstation (TPLAN Planning Station), an Optical Tracking Navigation Console (TNav) and a robotically controlled hand-held tool (TMINI Robot) that assists the surgeon in preparing the bone for implantation of TKA components.

The TPLAN Planning Station uses preoperative CT scans of the operative leg to create 3D surface models for case templating and intraoperative registration purposes. The Planning Workstation contains a library of 510(k) cleared knee replacement implant(s) available for use with the system. The surgeon can select an implant model from this library. The planner/surgeon can manipulate the 3D representation of the implant in relation to the bone

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Think Surgical. The logo features the word "THINK" in bold, sans-serif font, with the "I" replaced by a blue, three-dimensional teardrop shape. Below the word "THINK" is the word "SURGICAL" in a smaller, sans-serif font. The trademark symbol is located in the upper right corner of the teardrop shape, and the registered trademark symbol is located in the upper right corner of the word "THINK".

model to optimally place the implant. The surgeon reviews and approves the case plan once the surgeon is satisfied with the implant selection, location and orientation. The data from the approved plan is written to a file that is used to guide the robotically controlled hand-held tool.

The hand-held robotic tool is optically tracked relative to optical markers placed in both the femur and tibia and articulates in two degrees-of-freedom, allowing the user to place bone pins in a planar manner in both bones. Mechanical quides are clamped to the bone pins, resulting in subsequent placement of cut slots and drill guide holes such that the distal femoral and proximal tibial cuts can be made in the pre-planned positions and orientations, and such that the implant manufacturer's multi-planer cutting block can be placed relative to drilled distal femoral pilot holes. If the surgeon needs to change the plan during surgery, it can be changed intraoperatively.

Intended Use:

The TMINI 1.1 like the predicate, the TMINI™ Miniature Robotic System is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.

Indications for Use:

The Indications for Use of the TMINI™ Miniature Robotic System (TMINI 1.1) and the predicate are the same (See Table 1). Both describe the use of a system that uses diagnostic imaging of the patient to allow a surgeon to plan and execute orthopedic total knee procedures. The only difference is that the device, that is the subject of this 510(k) submission, implements a series of incremental modifications to the system to enhance system efficiency, consistency and ease of use.

None of the changes made to the predicate, nor the aggregate of all changes, have altered the indications for use or testing, nor have they raised new type of safety or effectiveness questions.

Tables 1 and 2, below, identify the substantial equivalence of the TMINI™ Miniature Robotic System (TMINI 1.1) to the predicate, TMINI™ Miniature Robotic System cleared via K232802.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Think Surgical. The logo features a stylized letter 'T' with a blue sphere at the top, resembling a surgical instrument. Below the 'T', the word 'THINK' is written in bold, sans-serif font, with the 'I' replaced by a blue, needle-like shape. Underneath 'THINK', the word 'SURGICAL' is written in smaller, sans-serif font.

TMINI™ Miniature RoboticTMINI™ Miniature Robotic
ProductSystem (TMINI 1.1)SystemConclusion
510(k) numberSubject DeviceK232802SAME
ManufacturerTHINK Surgical, IncTHINK Surgical, IncSAME
Product CodeOLOOLOSAME
Regulation21 CFR 882.456021 CFR 882.4560SAME
Intended UseIntended to assist the surgeonin providing software definedspatial boundaries fororientation and referenceinformation to anatomicalstructures during orthopedicprocedures.Intended to assist the surgeonin providing software definedspatial boundaries fororientation and referenceinformation to anatomicalstructures during orthopedicprocedures.SAME
Indications for UseThe TMINI™ MiniatureRobotic System is indicatedas a stereotaxicinstrumentation system fortotal knee replacement (TKA)surgery. It is to assist thesurgeon by providingsoftware-defined spatialboundaries for orientation andreference information toidentifiable anatomicalstructures for the accurateplacement of knee implantcomponents.The robotic device placementis performed relative toanatomical landmarks asrecorded using the systemintraoperatively and based ona surgical plan-determinedpreoperatively using CTbased surgical planning tools.It includes a handheld roboticdevice, an optical sensornavigation system andaccessories, software system,surgical instruments andaccessories.The targeted population hasthe same characteristics asthe population that is suitablefor the implant(s) compatiblewith the TMINI™ MiniatureRobotic System.The TMINI™ MiniatureRobotic System is to be usedwith the following kneereplacement system(s) inThe TMINI™ MiniatureRobotic System is indicatedas a stereotaxicinstrumentation system fortotal knee replacement (TKA)surgery. It is to assist thesurgeon by providingsoftware-defined spatialboundaries for orientation andreference information toidentifiable anatomicalstructures for the accurateplacement of knee implantcomponents.The robotic device placementis performed relative toanatomical landmarks asrecorded using the systemintraoperatively and based ona surgical plan-determinedpreoperatively using CTbased surgical planning tools.It includes a handheld roboticdevice, an optical sensornavigation system andaccessories, software system,surgical instruments andaccessories.The targeted population hasthe same characteristics asthe population that is suitablefor the implant(s) compatiblewith the TMINI™ MiniatureRobotic System.The TMINI™ MiniatureRobotic System is to be usedwith the following kneereplacement system(s) inSAME
ProductTMINI™ Miniature Robotic System (TMINI 1.1)TMINI™ Miniature Robotic SystemConclusion
indications and contraindications:- Enovis™ EMPOWR Knee System®- Ortho Development BKS® and BKS TriMax® Knee System- Total Joint Orthopedics Klassic® Knee System- United U2™ Knee Total Knee Systemindications and contraindications:- Enovis™ EMPOWR Knee System®- Ortho Development BKS® and BKS TriMax® Knee System- Total Joint Orthopedics Klassic® Knee System- United U2™ Knee Total Knee System
ProductTMINI™ MiniatureRobotic System (TMIN1.1)TMINI™ MiniatureRobotic SystemSubstantialEquivalenceConclusion
510(k) numberManufacturerSubject DeviceTHINK Surgical Inc.K232802THINK Surgical Inc.
Materials
-Materials UsedUses materials with a longhistory of use in orthopedicprocedures or providedbiocompatibility dataconsistent with ISO 10993requirementsUses materials with a longhistory of use in orthopedicprocedures or providedbiocompatibility dataconsistent with ISO 10993requirementsSAME
TechnologicalCharacteristics
-Major SystemComponentsPlanning and robot controlsoftware, robotic positioningdevice, navigation system,reusable and disposableinstrumentationPlanning and robot controlsoftware, robotic positioningdevice, navigation system,reusable and disposableinstrumentationSAME
-Patient ImagingCT images used to create a3D model of the bone forsurgical planningCT images used to create a3D model of the bone forsurgical planningSAME
-Preoperative planningworkstationTPLAN three-dimensionalpreoperative planningworkstationTPLAN three-dimensionalpreoperative planningworkstationSAME
-Surgical planningsystemTechnician guided surgicalplanning with surgeonreview and approval on adesktop planning stationTechnician guided surgicalplanning with surgeonreview and approval on adesktop planning stationSAME
-Bone Marker Arrays forbone registration andtrackingActive markers on femurand tibia mounted onto thebones via an attachmentassemblyActive markers on femurand tibia mounted onto thebones via an attachmentassemblySAME
-Surgical ExposureSimilar to traditionalsurgical exposureSimilar to traditionalsurgical exposureSAME
-Patient/RobotRegistrationPreoperatively determinedlandmarks are compared tointraoperatively identifiedlandmarks to completepatient bone registrationPreoperatively determinedlandmarks are compared tointraoperatively identifiedlandmarks to completepatient bone registrationSAME
-Camera TrackingTechnologySix camera overheadtracking with a wide-anglefield of viewSix camera overheadtracking with a wide-anglefield of viewSAME
-Cut guide positioningRobotic device places bonepins in the correct plane,then cutguide or drill blockis attached to the pins andboneRobotic device places bonepins in the correct plane,then cutguide or drill blockis attached to the pins andboneSAME
-Intraoperative planningchangesImplant position can be fullyadjusted, allowing deviationfrom the intended implantpositioning philosophy andimplant sizeImplant position can beadjusted along bone axisonly, preserving theintended implant positioningphilosophySubstantiallyEquivalent
ProductTMINI™ MiniatureRobotic System (TMIN1.1)TMINI™ MiniatureRobotic SystemSubstantialEquivalenceConclusion
510(k) numberSubject DeviceK232802
ManufacturerTHINK Surgical Inc.THINK Surgical Inc.
-Bone PreparationTechniqueA surgical saw is used tocut the bone through a cutguideA surgical saw is used tocut the bone through a cutguideSAME
-Intraoperative AnatomicMeasurementsThe tracked bone arraysand bone registration dataare used to determine theknee flexion angle andvarus/valgus laxityThe tracked bone arraysand bone registration dataare used to determine theknee flexion angle andvarus/valgus laxitySAME
-Gap BalancingDisplays the maximumspace in the medial andlateral compartments inmillimeters with the knee inextension and in flexionallowing the surgeon toperform gap balancing, ifdesired.Displays the maximumachievable angle betweenthe medial and lateralcompartments in degreeswith the knee in extensionand in flexion allowing thesurgeon to perform gapbalancing, if desired.SubstantiallyEquivalent
-TKA ComponentImplantation TechniqueImplants are secured to thebone, either with or withoutcement using standardsurgical technique providedby the implant manufacturerImplants are secured to thebone, either with or withoutcement using standardsurgical technique providedby the implant manufacturerSAME
-Compatible KneeImplant Systems- Enovis™ EMPOWR KneeSystem®-Ortho Development®BKS® and BKS TriMax®Knee System-Total Joint OrthopedicsKlassic® Knee System-United® U2™ KneeSystem- Enovis™ EMPOWR KneeSystem®-Ortho Development®BKS® and BKS TriMax®Knee System-Total Joint OrthopedicsKlassic® Knee System-United® U2™ KneeSystemSAME
Performance Testing-Full System RunThrough TestingPassedPassedSAME
Cutting Accuracy-Pin & Block PlacementAccuracy-Cadaver Lab ValidationTestingPassedPassedSubstantiallyEquivalent
-System Gap BalanceAccuracyPassed qualitativesurgeon assessmenttestingPassed quantitativemeasurement accuracytestingPassed qualitativesurgeon assessmenttesting
-User Needs ValidationTestingPassedPassedSAME
-Usability TestingPassedPassedSAME
-Software TestingPassedPassedSAME
ProductTMINI™ MiniatureRobotic System (TMIN1.1)TMINI™ MiniatureRobotic SystemSubstantialEquivalenceConclusion
510(k) numberSubject DeviceK232802
ManufacturerTHINK Surgical Inc.THINK Surgical Inc.
*Biocompatibility
Testing
-Cytotoxicity*PassedPassedSAME
-Sensitization*PassedPassedSAME
-IntracutaneousReactivity*PassedPassedSAME
-Acute Systemic Toxicity*PassedPassedSAME
-Pyrogenicity*PassedPassedSAME

Table 1: Comparison of Intended Use and Indications for Use

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Think Surgical. The logo features a stylized letter "T" in blue, with a sphere behind it. Below the "T" is the word "THINK" in black, with the "I" replaced by a thin blue line extending from the "T" above. Underneath "THINK" is the word "SURGICAL" in a smaller font, also in black.

Substantial Equivalence:

Both the TMINI™ Miniature Robotic System (TMIN 1.1), the subject of this submission, and the predicate device have the same intended use. Both are indicated as a stereotaxic instrumentation system for total knee replacement (TKA) surgery. It is to assist the surgeon by providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of knee implant components during orthopedic procedures.

The robotic device placement is performed relative to anatomical landmarks as recorded using the system intraoperatively and based on a surgical plan determined preoperatively using CT based surgical planning tools. The difference between the new device and the predicate is that the new device incorporates a series of incremental changes that enhance system efficiency, consistency and ease of use. None of these changes either individually or in the aggregate alter the intended use, Indications for Use, design, materials, technology, or operational principles of the TMINI™ Miniature Robotic System and no new questions of safety or effectiveness resulted from the changes.

The Indications for Use of the TMINI™ Miniature Robotic System (TMINI 1.1) and the predicate are the same (See Table 1). Both describe the use of a system that uses diagnostic imaging of the patient to allow a surgeon to plan and execute orthopedic total knee procedures. The only difference is that the device, that is the subject of this 510(k) submission, implements a series of incremental modifications to the system to enhance system efficiency, consistency and ease of use.

Performance testing to verify the cutting accuracy of the subject device was conducted following similar test methods and acceptance criteria to those used for the predicate device. This testing demonstrated that the TMINI™ Miniature Robotic System (TMINI 1.1) met all test criteria and specifications. Validation testing, with methods and acceptance criteria similar to that used for the predicate device, was conducted using simulated surgical testing in a cadaver model and all test criteria were met.

Biocompatibility information for patient contacting materials and testing for the TMINI™ Miniature Robotic System were presented in predicate device submission K232802. There are no material changes to any of the direct patient contact components of the TMINI™ Miniature Robotic System as a result of TMINI 1.1 modifications included in this submission; therefore, no additional biocompatibility testing was required.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Think Surgical. The logo features a stylized letter 'T' with a blue circle behind it, forming the top part of the 'T'. Below the 'T', the word 'THINK' is written in bold, sans-serif font, and below that, the word 'SURGICAL' is written in a smaller, sans-serif font.

TMINI™ Miniature Robotic System (TMINI 1.1) Traditional 510(k) Submission

Substantial equivalence in technological characteristic and performance of the TMINI™ Miniature Robotic System to the predicate device is outlined in Table-2 below:

Table-2: Substantial Equivalence

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Think Surgical. The logo features a stylized letter 'T' with a blue circular shape at the top, resembling a head. Below the 'T', the word 'THINK' is written in bold, sans-serif font. Underneath 'THINK', the word 'SURGICAL' is written in a smaller, sans-serif font.

TMINI™ Miniature Robotic System (TMINI 1.1) Traditional 510(k) Submission

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Think Surgical. The logo features a blue circle with a white "T" inside, and the word "THINK" in black, block letters below it. The "I" in "THINK" is stylized to look like a surgical instrument. Below "THINK" is the word "SURGICAL" in smaller, gray letters.

  • There are no material changes to any of the direct patient contact components of the TMINI™ Miniature Robotic System as a result of TMINI 1.1 modifications included in this submission; therefore, no additional biocompatibility testing was required.

Risk assessment was performed on the device in accordance with ISO 14971:2019 and THINK Surgical Risk Management procedures. Each change was assessed separately and in the aggregate through risk analysis and appropriate performance testing to evaluate the impact of the change. Risk analysis resulted in the identification of new instance of risks; however, no new clinical hazards were identified, and no new questions of safety or effectiveness were identified as a result of these changes. The risks identified have been mitigated to acceptable levels and there is no change in the overall risk profile of the device compared to the predicate cleared in K232802.

Conclusion

The TMINI™ Miniature Robotic System (TMINI 1.1) is substantially equivalent to the predicate, TMINI™ Miniature Robotic System (K232802), in the following ways:

  • it has the same intended use,
  • it has the same Indication for Use, ●
  • it has the same technological characteristics and operating principles and . incorporates the same design and materials.

Performance testing and risk analysis has demonstrated that the performance and risk profile of the TMINI™ Miniature Robotic System (TMINI 1.1) is substantially equivalent to that of the predicate device and does not raise any new question of safety and effectiveness.

THINK Surgical Inc. respectfully submits that this filing contains adequate information and data to demonstrate the substantial equivalence of the TMINI™ Miniature Robotic System (TMINI 1.1) to the legally marketed TMINI™ Miniature Robotic System cleared via K232802.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).