K Number
K240977

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-09-13

(156 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DenQ Sub SLA Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. This system is intended for delayed loading.

Device Description

The DenQ Sub SLA Implant System is a dental implant made of titanium metal intended to be surgically placed in the bone of the upper or lower jaw arches. This implant system has internal hex connection, tapered with straight body and bone level that are similar to other commercial available products based on the intended use, technology used, the material composition employed and performance characteristics. This DenQ Sub SLA fixture of implant system is supplied sterile.

The DenQ Sub Abutments are device made of cp titanium grade 4 and titanium alloy intended for use as a Prosthetic restoration. It consists of Abutments(Healing, Solid, Cement, Angled, Temporary, Multi-unit Straight, Multi-unit Angled, Multi-unit Healing Cap, Multiunit Ti Cylinder, Multi-unit Temporary Cylinder and FreeMilling).

AI/ML Overview

This document is a 510(k) Pre-market Notification for a dental implant system (DenQ Sub SLA Implant System). It aims to demonstrate substantial equivalence to previously cleared predicate devices, rather than proving the device meets acceptance criteria through clinical or comparative effectiveness studies in the typical sense of a novel AI/software device.

Therefore, many of the requested elements (like MRMC studies, number of experts for ground truth, training set information for an AI device) are not applicable to a 510(k) submission for a physical dental implant system. The "acceptance criteria" here refers to demonstrating that the device is as safe and effective as a legally marketed predicate device, primarily through non-clinical performance testing and comparison of characteristics.

Here's a breakdown of the available information relevant to your request, adapting it to the context of this specific document:

1. Table of Acceptance Criteria (as demonstrated by testing) and Reported Device Performance:

For a physical device like a dental implant, "acceptance criteria" are the relevant ISO standards and material specifications that the device must meet to be considered safe and effective and substantially equivalent to a predicate. The "reported device performance" is the demonstration that these tests were passed.

Acceptance Criteria (Relevant Standard/Test)Reported Device Performance (Outcome)
Gamma Sterilization Validation (ISO 11137-1, -2)Test performed and met standards.
Shelf-life Validation (ISO 11607-1, -2, ASTM F1980-16, ASTM F1140-13, ASTM A F2096-11, ASTM F88/F88M-15)Test performed and met standards.
BET Validation (USP <85><161>)Test performed and met standards.
User Sterilization Validation (ISO 17665-1, -2)Test performed and met standards.
Biocompatibility (ISO 10993-1, ISO 14971)Test performed and met standards.
Fatigue Testing (ISO 14801:2016) for DenQ Sub straight type Abutment, DenQ Sub Angled Abutment, and DenQ Sub SLA FixtureTest performed and met standards; worn compliant and similar to previously cleared primary predicate devices.
Shear testing (ASTM F1044)Test performed. (Outcome not explicitly stated as "met standards" but implied by overall conclusion of substantial equivalence).
Tension testing (ASTM F1147)Test performed. (Outcome not explicitly stated as "met standards" but implied).
Abrasion characteristics (Scratch test)Test performed; information on protocol, load range, indenter, depth, failure event, and images provided. (Outcome implied as acceptable).
Surface analysis (SEM/EDS analysis per FDA guidance)Test performed. (Outcome implied as acceptable).
MR Environment Safety (Magnetically induced displacement force and torque per FDA Guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment")Non-clinical worst-case MRI review performed using scientific rationale and published literature. (Conclusion: MR Conditional).

2. Sample Size Used for the Test Set and Data Provenance:

This document is for a physical medical device, not an AI/software device that would typically have a "test set" of patient data. The "tests" here refer to laboratory and bench testing on the physical device components.

  • Sample Size: Not specified for individual non-clinical tests (e.g., how many implants were subjected to fatigue testing). This information is typically detailed in the full test reports, which are summarized in the 510(k) submission.
  • Data Provenance: Not applicable in the sense of patient data. The results are from in-vitro (laboratory) and bench testing of the devices themselves. The tests were performed "in accordance with ISO standards and US regulations." The manufacturer is located in the Republic of Korea.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

  • Not applicable. This submission relies on engineering standards and laboratory measurements for its performance criteria, not expert interpretations of medical images or clinical data for ground truth.

4. Adjudication Method for the Test Set:

  • Not applicable. There is no "test set" requiring adjudication in the context of this type of submission.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done (and effect size):

  • No. MRMC studies are typically for diagnostic AI devices assessing human reader performance. This is a physical dental implant.
  • Summary of clinical testing: "No clinical testing was performed for this submission."

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • Not applicable. This is a physical device, not an algorithm.

7. The Type of Ground Truth Used:

  • The "ground truth" for the device's performance is established by engineering and material standards (e.g., ISO, ASTM, USP), and the physical measurements and observations derived from in-vitro and bench testing of the device's components (e.g., fatigue resistance, material properties, sterility).

8. The Sample Size for the Training Set:

  • Not applicable. This is a physical device, not an AI algorithm that requires a "training set" of data.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable. As above, no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

September 13, 2024

DenQ Taehoon Lee Representative 302, PNU Hyowon Industry-Academic Cooperation Hall, 2, Busandaehak-ro, Geumjeong-gu Busan, 46241 REPUBLIC OF KOREA

Re: K240977

Trade/Device Name: DenQ Sub SLA Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: August 14, 2024 Received: August 14, 2024

Dear Taehoon Lee:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Andrew I. Steen -S

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K240977

Device Name

DenQ Sub SLA Implant System

Indications for Use (Describe)

The DenQ Sub SLA Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. This system is intended for delayed loading.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for DenQ Implant. The logo is pink and features the word "DenQ" in a stylized font. Below the word "DenQ" is the word "IMPLANT" in a smaller, sans-serif font. The logo is simple and modern, and the pink color gives it a friendly and approachable feel.

1. 510(k) Summary

Submitter DENQ TAEHOON LEE 302, PNU Hyowon Industry-Academic Cooperation Halll, 2, Busandaehak-ro 63Beon-gil, Gumjeong-gu, Busan, Republic of Korea Tel: +82-51-514-4888 Email: ceo@deng.kr

Device Information

Trade Name: DenQ Sub SLA Implant System Common Name: Endosseous Dental Implant Classification Name: Implant, Endosseous, Root-Form Product Code: DZE Secondary Product Code: NHA Regulation Number: 872.3640 Device Class: Class II Date Prepared: 09/13/2024

Predicate Device(s) Primary Predicate Device

  • K222707-IH Implant System, Sewon Medix Inc. ●

Reference Devices

  • K121995-TS Fixture System, OSSTEM Implant Co., Ltd. ●
  • K213576-Tatum Surgical Dental Implant System, Suncoast Dental, Inc. dba Tatum Surgical
  • K230108-Straumann® BLC and TLC Implants, Straumann USA, LLC
  • K153521-IH Implant System, Sewon medix Inc. ●

Description

The DenQ Sub SLA Implant System is a dental implant made of titanium metal intended to be surgically placed in the bone of the upper or lower jaw arches. This implant system has internal hex connection, tapered with straight body and bone level that are similar to other commercial available products based on the intended use, technology used, the material composition employed and performance characteristics. This DenQ Sub SLA fixture of implant system is supplied sterile.

The DenQ Sub SLA Implant System diameter and lengths are below:

  • The DenQ Sub SLA Fixtures

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for DenQ Implant. The logo is pink and features the word "DenQ" in a stylized font. Below the word "DenQ" is the word "IMPLANT" in a smaller, sans-serif font. The logo is simple and modern, and the pink color gives it a friendly and inviting feel.

ProductnameDivisionPlatform Diameters(Fixture Diameter)Body DiametersLengths
Fixture IRegularØ3.75, Ø3.77-7, 7.5, 8, 8.5, 9,9.5, 10, 10.5,
RegularØ4.25-11, 11.5, 12,
RegularØ4.6, Ø4.63, Ø4.65-12.5, 13, 13.5,
RegularØ5.05, Ø5.08. Ø5.1-14, 14.5, 15 mm
Fixture IWideØ5.95-
Fixture IWideØ6.8-
Fixture IIRegularØ4.21Ø4.07, 7.5, 8, 8.5, 9,9.5, 10, 10.5,
RegularØ4.67Ø4.511, 11.5, 12,
RegularØ5.14Ø5.012.5, 13, 13.5,
Fixture IIWideØ6.2Ø5.014 mm
Fixture IIWideØ6.8Ø5.8
Fixture IIIRegularØ4.25-7, 7.5, 8, 8.5, 9,9.5, 10, 10.5,
RegularØ4.6, Ø4.63, Ø4.65-11, 11.5, 12,
RegularØ5.05, Ø5.08. Ø5.1-12.5, 13, 13.5,
Fixture IIIWideØ5.95-14 mm
Fixture IIIWideØ6.8-

Internal Hex-connected, Bone level fixture, Tapered and Tapered-Straight body Implant Fixture Dimension:

The fixture is made of Pure Titanium Grade 4. The surface is treated by Sand -blasting (Large grit) and acid etching method (SLA). This system only contains the implant bodies with cover screw and are provided as a set-packing.

The DenQ Sub Abutments are device made of cp titanium grade 4 and titanium alloy intended for use as a Prosthetic restoration. It consists of Abutments(Healing, Solid, Cement, Angled, Temporary, Multi-unit Straight, Multi-unit Angled, Multi-unit Healing Cap, Multiunit Ti Cylinder, Multi-unit Temporary Cylinder and FreeMilling). The TiN coating is applied to some abutment, which are solid Abutment, Cement Abutment and Angled Abutment. They have various diameter, height(Length), gingiva height, post height and angled. The Healing Abutment has various diameter 4, 4.5, 5.0, 5.5, 6.0, 6.5 and 7.0, Lengths 8.4, 9.4, 10.4, 11.4 and 12.4mm. The Solid Abutment packaging is only abutment that has various diameters 4.0, 4.5, 5.0, 5.5, 6.0, 6.5 and 7.0, gingiva height 0.5, 1, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5 and 5mm, post height 4, 5.5 and 7mm. The Cement Abutment has various diameters 4.0, 4.5, 5.0, 5.5, 6.0, 6.5 and 7.0, gingiva height 0.5, 1, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5 and 5mm, post height 4. 5.5 and 7mm. The Angled Abutment has post angled 15°, 17°, 20° and 25°. Its packaging is Abutment with Abutment screw and it has hex and non-hex type. The angle abutment has diameter 4.0, 4.5, 5.0, 5.5 and 6.0mm, gingiva height 0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5 and 5.0mm, post height 8mm. It has hex and non-hex, hex type of a and b that a type is same path for post height with hex's edge and b type is same path for hex's plat with post height. Its packaging is abutment with abutment screw as a set packaging. The temporary abutment has diameter 4.0, 4.5, 5.0, 5.5 and 6.0mm, gingiva height 0.5, 1.0, 1.5, 2.0, 2.5 and 3.0mm, height 12.6, 13.1, 13.6, 14.1, 14.6, 15.1 and 12.85, 13.35, 13.85, 14.35, 14.85 and 15.35. Its packaging unit is abutment with abutment screw as a set. The multi-unit Abutment consist straight and angled type and straight type has diameter 4.8, 5.0 and 5.5mm, gingiva height 0.5, 1.0, 1.5, 2.0, 2.5, 4.0, 4.5 and 5.0mm, the hex of abutment 2.0mm, hex height 2.2mm. And angled type has diameter 4.71, 4.92, 5.04, 5.4 and

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for DenQ Implant. The logo is pink and features the word "DenQ" in a stylized font. Below the word "DenQ" is the word "IMPLANT" in a smaller, sans-serif font. The logo is simple and modern, and the pink color gives it a friendly and approachable feel.

5.5mm, angle 15°, 17°, 20° and 30°, gingiva height 1.0, 2.0, 2.5, 3.0, 3.5, 4.0 and 4.5. Its packaging unit is abutment with abutment screw as a set. It has hex and non-hex. The multiunit healing cap has various diameter 4.5, 4.8, 5.0 and 5.5mm, height 4.8 and 5.0mm. The multi-unit Ti cylinder has diameter 4.5, 4.8, 5.0 and 5.5mm, post height 4.1, 5.6 and 7.1mm. Its packaging unit is Ti cylinder with cylinder screw as a set. The multi-unit temporary cylinder has various diameter 4.5, 4.8, 5.0 and 5.5mm. length 12mm. Its packaging unit is temporary abutment with cylinder as a set. The FreeMilling abutment has various diameter 4.5, 5.0, 5.5 and 6.0mm, gingiva height 1, 2, 3, 4 and 5mm, length 15.5 and 15.75mm. It has hex and non-hex and packaging unit is abutment with abutment screw. It is used by hand milling. Its packaging unit is abutment with abutment screw as a set.

Healing Abutment, Temporary Abutment, Multi-unit Abutment and Temporary Abutment are uncoating applied. All Abutment is supplied non-sterilized and should be sterilized before use as given conditions of manufacturer.

Indication for Use

The DenQ Sub SLA Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. This system is intended for delayed loading.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for DenQ Implant. The logo is pink and features the word "DenQ" in a stylized font. Below the word "DenQ" is the word "IMPLANT" in smaller, sans-serif letters. The logo is simple and modern.

Predicate Device and reference devices & Comparison of Fixture

The subject device is substantially equivalent to the following predicate device and reference devices:

  • · K222707, IH Implant System by Sewon Medix Inc. Primary predicate device
  • K121995, TS Fixture System by Osstem Implant Co., Ltd. Reference device .
  • K213576, Tatum Surgical Dental Implant System by Suncoast Dental, Inc. dba Tatum Surgical. Reference device
  • · K230108, Straumann® BLC and TLC Implants by Straumann USA, LLC Reference device
DivisionSubject DevicePrimary PredicatedeviceReference device
Trade NameDenQ Sub SLAImplant SystemIH Implant SystemTS Fixture SystemTatum SurgicalDental ImplantSystemStraumann® BLC andTLC Implants
510(k)NumberK240977K222707K121995K213576K230108
ManufacturerDenQ Co., Ltd.Sewon Medix Inc.Osstem Implant Co.,Ltd.Suncoast Dental, Inc.dba Tatum SurgicalStraumann USA, LLC
Reason forprimary andReferencedeviceN/AMajor technologicalcharacteristicsImplant body shape,wide diameter(Ø6.2,6.8) andlength(7~12.5mm)Implant widediameter(Ø6.0, 7.0)and length(13~15mm)Basic surfacetreatment, implantbody diameter andlength
Indication foruseThe DenQ Sub SLAImplant System isindicated for use inpartially or fullyedentulous mandiblesand maxillae, insupport of single ormultiple unitrestorations including;IH Implant System isdevice made oftitanium and titaniumalloy indicated for inpartially or fullyedentulous mandiblesand maxillae, insupport of single ormultiple-unitThe TS FixtureSystem is indicated foruse in partially or fulledentulous mandiblesand maxillae, insupport of single ormultiple-unitrestorations including;cemented retained.The Tatum SurgicalIntegrity Tapered, "T"and “S” DentalImplant Systems areintended to besurgically placed inthe bone of the upperor lower jaw toprovide support forStraumann® dentalimplants are indicatedfor the functional andesthetic oralrehabilitation of theupper or lower jaw ofedentulous or partiallyedentulous patients.They can be used for
cemented retained, orscrew retained, oroverdenturerestorations, andterminal orintermediate abutmentsupport for fixedbridgework. Thissystem is dedicated forone and two stagesurgical procedures.This system isintended for delayedloading.restorations including;cemented retained orscrew retainedrestorations andterminal or interminalabutment support forfixed bridgework. IHImplant System is forsingle and two stagesurgical procedures. Itis intended for delayedloading.screw retained, oroverdenturerestorations; and finalor temporary abutmentsupport for fixedbridgework. It isintended for delayedloading.removable or fixedprosthesis to restorechewing function. Itmay be used withsingle-stage or two-stage procedures, forsingle or multiple unitrestorations, and maybe loaded immediatelywhen good primarystability is achievedand with appropriateocclusal loading.immediate, early orlate implantationfollowing theextraction or loss ofnatural teeth. Theimplants can be placedwith immediatefunction for single-tooth and/or multiple-tooth restorationswhen good primarystability is achievedand with appropriateocclusal loading torestore chewingfunction.
MaterialCP Titanium Gr.4(ASTM F67-13)CP Titanium Gr.4(ASTM F67-13)CP Titanium Gr.4(ASTM F67-13)Ti 6Al 4V(ASTM F136-13)Titanium-13Zirconium alloy(Roxolid®)
Design(FixtureType)Image: Fixture IImage: MiniImage: TSII SA FixtureImage: Ti 6Al 4VImage: Titanium-13 Zirconium alloy
FixtureDiameterFixture IReg.: Ø3.75, Ø3.77,Ø4.25, Ø4.6, Ø4.63Mini: Ø3.7Reg.: Ø3.8, Ø4.2,Ø4.5, Ø5.0, Ø5.95TSII SA FixtureØ3.5, Ø4.2, Ø4.4,Ø4.9TSIII SA FixtureØ3.7, 4.0, Ø4.5, Ø5.0,Ø6.0, Ø7.0, Ø8.0Ø3.3, Ø3.75, Ø4.5,Ø5.5, Ø6.5
Ø4.65, Ø5.05, Ø5.08,Ø5.1Wide: Ø5.95, Ø 6.8Fixture IIReg.: Ø4.21, Ø4.67,Ø5.14Wide: Ø6.2, Ø6.8Fixture IIIReg.: Ø4.25, Ø4.6,Ø4.63, Ø4.65, Ø5.05,Ø5.08, Ø5.1Wide: Ø5.95, Ø 6.8Ø3.75, Ø3.77, Ø4.2,Ø4.25, Ø4.6, Ø4.63,Ø4.65, Ø5.05, Ø5.08,Ø5.1TSIII SA Ultra-WideFixtureØ5.92, Ø5.95, Ø6,Ø6.8
FixtureLengthFixture I715mmFixture II, III714 mm7.0, 8.5, 10, 11.5, 13,15 mmTS Fixture715 mmTS III SA Ultra-WideFixture712.5 mm9, 11, 14, 17, 20 mm(no 20 mm length forØ7 and 8 mm)BLC: Ø3.3: 818mmØ3.75 and Ø4.5:618mmØ5.5: 616mmØ6.5: 614mmTLC: Ø3.3, Ø3.75and Ø4.5 of BLC aresameØ5.5: 612mmØ6.5: 610mm
Structure- Internal Hex-connected-Bone level- Tapered & straightbody shape- Cutting edge withself-tapping- 0.8mm thread pitch- Internal Hex-connected-Bone level- Tapered body shape- Cutting edge withself-tapping- 0.8mm thread pitch- Internal Hex-connected-Bone level- Tapered body shapeand straight bodyshape- 4 sided cutting edgewith self-tapping- Tissue or BoneLevel- Threaded- Tapered body- Root-form implantApically taperedBone Levelimplant

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for DenQ Implant. The logo is pink and features the word "DenQ" in a stylized font. Below the word "DenQ" is the word "IMPLANT" in smaller, gray letters. The logo is simple and modern.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for DenQ Implant. The logo is pink and features the word "DenQ" in a stylized font. Below the word "DenQ" is the word "IMPLANT" in a smaller, sans-serif font. The "Q" in "DenQ" is stylized to look like a spiral.

{10}------------------------------------------------

Image /page/10/Picture/4 description: The image shows the logo for DenQ Implant. The logo is pink and features the word "DenQ" in a stylized font. Below the word "DenQ" is the word "IMPLANT" in smaller, sans-serif letters. The logo is simple and modern.

ConnectiontypeInternal hexconnectionInternal hexconnectionInternal hexconnectionInternal PentagonconnectionTorcFit(with conicalfitting)
Surgery typeOne or two stageOne or two stageOne or two stageOne or two stageOne or two stage
SurfaceTreatmentSLASLASA(SLA)Aluminum oxideblasted and passivatedHydrophilicSLActive® andSLA®
GammaSterilizedYesYesYesYesYes
Product CodeDZE, NHADZE, NHADZEDZE, NHADZE
Shelf Life5 years5 years8 years5 years5 years
S.EThe Indication for use of DenQ Sub SLA Implant System is similar to the primary predicate device as K222707. Also,the indication for use of reference devices K121995, K213576 and K230108 are the similar to the Subject device. TheDenQ Sub SLA Implant System has taper body of fixture I, III and taper-straight-taper body shape of fixture II. TheDenQ Sub SLA Implant System has the similar raw material, surface treatment, surgical type, sterilization method, shelf-life, connection structure, implant basic components and product code as primary predicate device K222707. Thedifference between the Subject device and primary predicate device are their similar implant body shape, implantdiameter, and length. The implant shape and configuration were similar to the primary predicate device and referencedevice, with bone level, tapered and straight body, cutting edge, and self-tapping structures. The Subject device hastapered body and tapered-straight-tapered body shapes, but the primary device K222707 has only tapered body shape.However, the reference device K121995 has an implant body shape with tapered and straight body shapes, and the bodyshape of subject device and reference device K121995 are very similar. The implant diameter of subject device rangesfrom Ø3.75 to Ø6.8(Ø3.88, Ø3.9 of Fixture II and Ø3.77 of Fixture III are not available), and these diameters have asimilar range to the diameter of primary predicate device. However, implant diameters of Ø6.2 and Ø6.8 are not includedin the primary predicate device. The implant diameters Ø6.2 and Ø6.8 are similar to the reference device K122995 wideimplant Ø6.0 and Ø6.2 of implant diameter of subject device, and the subject device Ø6.8 diameter is similar to Ø6.8diameter of reference device K122995. The implant length of the Subject device is 715mm, and the implant length ofthe primary predicate device is 715mm, but Ø5.95 is 713mm and there is no 15mm. The implant lengths of Subjectdevice Ø6.2 and Ø6.8x715mm are similar to Ø5.92, Ø5.95, Ø6.8x7~12.5mm in reference device K121995. The implant

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for DenQ Implant. The logo is pink and features the word "DenQ" in a stylized font. Below the word "DenQ" is the word "IMPLANT" in smaller, sans-serif letters. The "Q" in "DenQ" is designed to look like a spiral.

lengths of Subject device Ø5.95x15mm, Ø6.2 and Ø6.8x1315mm are similar to the Ø6.0, Ø7.0x1117mm of the reference device K213576. The subject device consists of an implant body with a diameter of 3.7 to 6.8, and the K230108 consists of an implant body with a diameter of 3.3 to 6.5, which constitute implant body diameters similar to those of the subject device and K230108. The Subject device has sand blasting surface it is similar primary device K222707 and the basic surface treatment method is the similar as the reference device K230108.

Reference Devices & Comparison of Abutment

The subject device is substantially equivalent to the following reference device:

  • · K153521, IH Implant System by Sewon Medix Inc. Reference device
DivisionSubject DeviceReference Device
Trade NameDenQ Sub SLA Implant SystemIH Prosthetic System
510(k) NumberK240977K153521
ManufacturerDenQ Co., Ltd.Sewon Medix Inc.
Indication for useThe DenQ Sub SLA Implant System is indicatedfor use in partially or fully edentulous mandiblesand maxillae, in support of single or multiple unitrestorations including; cemented retained, screwretained, or overdenture restorations, and terminalor intermediate abutment support for fixedbridgework. This system is dedicated for one andtwo stage surgical procedures. This system isintended for delayed loading.IH Implant System is device made of titanium andtitanium alloy indicated for in partially or fullyedentulous mandibles and maxillae, in support of singleor multiple-unit restorations including ; cementedretained or screw retained restorations and terminal orinterminal abutment support for fixed bridgework. IHImplant System is for single and two stage surgicalprocedures. It is intended for delayed loading.
MaterialTi 6Al 4V ELI(ASTM F136-13)Healing, Solid, Cement, Multi-unitCP Titanium Gr4(ASTM F67-13)Angled, Temporary, FreeMillingTi 6Al 4V ELI(ASTM F136)Healing, Solid, Cement, Multi-unitCP Titanium Gr4(ASTM F67-13)Angled, Temporary, FreeMilling

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for DenQ Implant. The logo is pink and features the word "DenQ" in a stylized font. Below the word "DenQ" is the word "IMPLANT" in a smaller, sans-serif font. The "Q" in "DenQ" is stylized to look like a spiral.

Image /page/12/Figure/2 description: The image shows a table with different designs of abutments. The designs are Healing, Solid, Cement, Angled, and Temporary. Each design has two different images of the abutment. The abutments are used in dentistry to connect a dental prosthesis to a dental implant.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for DenQ Implant. The logo is pink and features the word "DenQ" in a stylized font. Below the word "DenQ" is the word "IMPLANT" in a smaller, sans-serif font. The logo is simple and modern, and the pink color gives it a friendly and approachable feel.

510(k) Summary K240977

Multi-unit(St. &Angled)Image: Multi-unit (St. & Angled)Image: Multi-unit (St. & Angled)Product CodeNHANHA
FreeMillingImage: FreeMillingImage: FreeMillingThe DenQ Sub Abutment is similar indication for use, Materials, Structure, Connection type, Surface,Sterilization, and product code as the reference device(K153521). But dimension is included referencedevice(K153521). And the design(shape) and description are similar with reference device(K153521). But it isdifferent characteristics for each abutment as blow;
DescriptionThis product is a bone level dental implantAbutment(Endosseous Dental Abutment). It is adental implant to support and maintain dentures byimplanting a fixture into jawbones and prostheticrestoration in case of partial or total loss of naturalteeth. This product is excellent biocompatibility, itis aesthetically treated by mechanical and physicalcoating treatment on titanium alloy. This dentalimplant superstructure can use cement-fixed andscrew fixed prostheses. It is mechanicallyconnected to the bone level dental implant fixture.Healing Abutment is used to make a soft tissue shapebefore setting up prosthetics and removing cover screwafter osteointegration. Angled Abutment is an abutmentwhich has certain angle to easy for adjustment ofinstallation angle of prosthesis. Temporary Abutment isused temporary until final prosthesis is made to maintainesthetic appreciation and chew ability. Multi-UnitAbutment is used for screw retained multiple case. Solid,Cement and FreeMilling Abutment to fabricate aprosthesis of internal single & bridge cement retainedtype.DenQ Sub AbutmentReference device
DimensionHealingDia.: Ø4.0, Ø4.5, Ø5.0, Ø5.5, Ø6.0, Ø6.5, Ø7.0G/H Length: -Post length: -Dia.: Ø4.0, Ø4.5, Ø5.0, Ø5.5, Ø6.0, Ø6.5G/H Length: -Post length: -1. Shape(TiN coating applied)Multi-unit(Angled)Non-appliedPartial applied
SolidDia.: Ø4.0, Ø4.5, Ø5.0, Ø5.5, Ø6.0, Ø6.5, Ø7.0G/H: 0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0mmP/L: 4, 5.5, 7mmDia.: Ø4.0, Ø4.5, Ø5.0, Ø5.5, Ø6.0, Ø6.5, Ø7.0G/H: 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0mmP/L: 4, 5.5, 7mmTechnologicalCharacteristics2. Dimension(Diameter)HealingØ4.0, Ø4.5, Ø5.0, Ø5.5, Ø6.0,Ø6.5, Ø7.0Ø4.0, Ø4.5, Ø5.0, Ø5.5, Ø6.0, Ø6.5
CementØ4.0, Ø4.5, Ø5.0, Ø5.5, Ø6.0,Ø6.5, Ø7.0Ø4.5, Ø5.0, Ø5.5, Ø6.0, Ø6.5
CementDia.: Ø4.0, Ø4.5, Ø5.0, Ø5.5, Ø6.0, Ø6.5, Ø7.0G/H: 0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0mmP/L: 4, 5.5, 7mmDia.: Ø4.5, Ø5.0, Ø5.5, Ø6.0, Ø6.5G/H: 0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0mmP/L: 4, 5.5, 7mmTemporaryØ4.0, Ø4.5, Ø5.0, Ø5.5, Ø6.0Ø4.0, Ø4.5
AngledDia.: Ø4.0, Ø4.5, Ø5.0, Ø5.5, Ø6.0G/H: 0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0mmP/L: 8mmAngle: 15°, 17°, 20°, 25°Dia.: Ø4.5, Ø5.0, Ø5.5, Ø6.0G/H: 2.0, 3.0, 4.0mmP/L: 7.5, 8mmAngle: 15°, 25°Multi-unit StraightØ4.8, Ø5.0, Ø5.5Ø4.8
TemporaryDia.: Ø4.0, Ø4.5, Ø5.0, Ø5.5, Ø6.0G/H: 0.5, 1.0, 1.5, 2.0, 2.5, 3.0mmP/L: -Dia.: Ø4.0, Ø4.5G/H: 1.0, 3.0mmP/L: -Multi-unit AngledØ4.71, Ø4.92, Ø5.04, Ø5.4, Ø5.5Ø4.7, Ø5.13
Multi-unit(St.)Dia.: Ø4.8, Ø5.0, Ø5.5G/H: 0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0mmP/L: 2.2mmDia.: Ø4.8G/H: 1.5, 2.5, 3.5, 4.5mmP/L: 2.2mmFreeMillingØ4.5, Ø5.0, Ø5.5, Ø6.0Ø4.5, Ø5.0, Ø5.5, Ø6.0, Ø6.5
Multi-unit(Angled)Dia.: Ø4.71, Ø4.92, Ø5.04, Ø5.4, Ø5.5G/H: 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5mmP/L: -Angled: 15°, 17°, 20°, 25°, 30°Dia.: Ø4.7, Ø5.13G/H: 2.5, 3.5, 4.5mmP/L: -Angled: 17°, 30°3. Dimension(G/H Length)
FreeMillingDia.: Ø4.5, Ø5.0, Ø5.5, Ø6.0G/L: 1.0, 2.0, 3.0, 4.0, 5.0mmH: 8, 9, 10, 11, 12mmDia.: Ø4.5, Ø5.0, Ø5.5, Ø6.0, Ø6.5G/H: 1.0, 2.0, 3.0mmP/L: 10, 11, 12mmMulti-unit Straight0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0,4.5, 5.0mm1.5, 2.5, 3.5, 4.5mm
StructureInternal HexMulti-unit Angled1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0,4.5mm2.5, 3.5, 4.5mm
Connection typeScrew & Internal Hex-ConnectedTemporary0.5, 1.0, 1.5, 2.0, 2.5, 3.0mm1.0, 3.0mm
SurfaceMachined or TiN4. Dimension(Post Length)
SterilizationNon-SterileAngle8mm7.5, 8mm

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image contains a logo for "DenQ Implant". The logo is pink and features the word "DenQ" in a stylized font. Below the word "DenQ" is the word "IMPLANT" in smaller, sans-serif letters. To the right of the word "DenQ" is a circular design with a dot in the upper right corner.

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image contains the logo for "DenQ Implant". The logo is pink and features the brand name in a stylized font. Below the brand name, the word "IMPLANT" is written in smaller, uppercase letters.

{16}------------------------------------------------

Temporary8, 9, 10, 11, 12mm10, 11, 12mm
5. Dimension(Angle)
DenQ Sub AbutmentReference device
Angled15°, 17°, 20°, 25°15°, 25°
Multi-unit(Angled)15°, 17°, 20°, 25°, 30°17°, 30°

Substantial Equivalence Discussion

The DenQ Sub SLA Fixture has similar indication for use, material, surgery type, sterilization method, connection type, shelf life, product code and similar design, structure, diameter and technological characteristic as the primary predicate device(K222707). The wide implant diameter and length of subject device differ from the primary predicate device K 122995 and K 213576 support substance of subject device's wide implant diameters and lengths. The basic surface treatment of reference device K230108 supports substantial equivalence to subject device's surface treatment.

The DenQ Sub Abutment is similar indication for use, materials, structure, surface, sterilization, and product code as the device but, the technological characteristic is coating area and dimensions between subject device and the angle is included in the range of the reference device. The coating area, dimension, and angle of abutment of reference device K153521 support substantial equivalence of coating area, dimension, and angle of abutment of subject device.

The DenQ Sub SLA System has been subjected to performance and product validations prior to release. Testing including biocompatibility has been performed to ensure the devices comply with the applicable International and US regulations.

The differences between the subject devices are detailed shape and detailed dimension of diameter, length, angle, gingiva height, post height and coating area.

Summary of non-clinical test data

The following tests, verification and validations were performed in accordance with ISO standards and US regulations.

  • Gamma Sterilization Validation test according to ISO 11137-1, -2

  • Shelf-life Validation test according to ISO 11607-1, ISO 11607-2, ASTM F1980-16, ASTM F1140-13, ASTM A F2096-11 and ASTM F88/F88M-15

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the logo for DenQ Implant. The logo is pink and features the word "DenQ" in a stylized font. Below the word "DenQ" is the word "IMPLANT" in smaller, uppercase letters. The logo is simple and modern, and the pink color gives it a friendly and approachable feel.

  • BET Validation test according to USP <85><161> A
  • User Sterilization Validation test according to ISO 17665-1, -2 A
  • Biocompatibility test according to ISO 10993-1 and ISO 14971 A
  • A Fatigue test according to ISO 14801
  • A Shear testing was performed per ASTM F1044
  • A Tension testing was performed per ASTM F1147
  • Abrasion characteristics was determined by a scratch test drawn across the coated surface under an increasing load (either stepwise or continuous) until at some load (termed as critical load) a well-defined failure event occurs. Information clearly indicating the test protocol, selection of the load range used for the scratch test, type of indenter used to perform the scratch of the surface, depth of the scratch with respect to the applied load, failure event selection criteria, and high mages clearly demonstrating the coating detachment was provided.

  • A Surface analysis according to the FDA guidance document including SEM/EDS analysis.

The fatigue test for DenQ Sub straight type Abutment and DenQ Sub Angled Abutment and DenQ Sub SLA Fixture were conducted according to the "Guidance for industry and FDA staff Class II Special Controls Got-form Endosseous Dental Implants and Endosseous Dental Abutment" and ISO 14801:2016 Dentistry-Implant-Dynamic loading test for endosseous dental implant. This was tested based on the wore compliant and similar to previously cleared primary predicate devices. Those tests have been performed to evaluate the surface characteristics compared to the primary predicate device. The results of the above tests have met the standards and demonstrated the substantial equivalence with the primary predicate device. The non-clinical testing demonstrate that the subject device is substantially equivalent to the primary predicate device.

MR Environment: MR Conditional

Non-clinical worst-case MRI review was performed to evaluate the metallic DenQ Sub SLA Implant System devices in the MRI environment using scientific rationale and published literature (i.e., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Nagnetically Induced Displacement Force and Torque on Metal Devices." Journal of Testing and Evaluation 49, no.2 (March/April 2021): 783-795), based on the entire system to include all variations (all compatible implant bodies, dental and fixation screws)

{18}------------------------------------------------

Image /page/18/Picture/0 description: The image shows the logo for DenQ Implant. The logo is pink and features the word "DenQ" in a stylized font. Below the word "DenQ" is the word "IMPLANT" in smaller, gray letters. The logo is simple and modern.

510(k) Summary K240977

and material compositions. The rationale addressed parameters per the FDA Guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment", including magnetically induced displacement force and torque."

Summary of clinical testing

No clinical testing was performed for this submission.

Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification DenQ concludes that the Den 2 Substantially equivalent to the primary predicate devices as described herein.

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.