K Number
K240916

Validate with FDA (Live)

Date Cleared
2024-10-18

(198 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non sterile.

Device Description

The Surgical face mask is single use, three-layer, flat-pleated style with ear loops/ties and nose piece. The mask is manufactured with three layers, the inner and outer layers, and the straps are made of nonwoven fabrics, the middle layer is made of melt blown fabrics.

The model of proposed device, tie-on, is held in place over the user's mouth and nose by four ties welded to the face mask. The tie is made of spun-bond polypropylene.

The model of proposed device, ear loops, is held in place over the user's mouth and nose by two elastic ear loops welded to the face mask. The elastic ear loops are not made with natural rubber latex.

The ties/ear loops are not made with natural rubber latex. The nose piece on the layers of face mask is to allow the user to fit the face mask around their nose, which is made of Polypropylene with iron core. The Surgical face mask will be provided in blue. The masks are sold non-sterile and are intended to be single use, disposable devices.

AI/ML Overview

This document is a 510(k) summary for a Surgical Face Mask. It details the non-clinical testing performed to demonstrate substantial equivalence to a predicate device.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

ItemAcceptance Criteria (ASTM F2100-23 Level 3)Reported Performance
Performance Testing
Bacterial filtration efficiency (BFE) (%)≥9899.7%-99.9% (Pass)
Differential pressure (mmH2O/cm²)<6.0 mmH2O/cm²3.2-3.7 mmH2O/cm² (Pass)
Sub-micron particulate filtration efficiency at 0.1 micron, % (PFE)≥8597.16-99.52% (Pass)
Resistance to penetration by synthetic blood, Minimum pressure29 of 32 test articles passed at 160mmHg32 of 32 test articles Pass (at 160mmHg, implied)
Flame spreadClass 132 of 32 test articles Pass (Class 1, implied)
Biocompatibility Testing
Cytotoxicity (ISO 10993-5)Viability < 70%, it has a cytotoxic potential. (Acceptance if no cytotoxic potential)Tie on: 91.0%; Ear loops: 70.5% (No potential toxicity)
Irritation (ISO 10993-23)Primary Irritation-Index ≤ 2.0 (Response category Negligible)Primary irritation indexes of both polar and non-polar test groups were 0. (No skin irritation)
Sensitization (ISO 10993-10)Provided grades less than 1, otherwise sensitization. (Acceptance if no sensitization)Skin sensitization rates of both polar and non-polar extract groups were 0%. (No sensitization)

2. Sample size used for the test set and the data provenance

The document does not specify the exact sample sizes used for each individual performance test (BFE, differential pressure, PFE). However, for Resistance to penetration by synthetic blood and Flame spread, it states that 32 test articles were used.

The data provenance is not explicitly stated in terms of country of origin or retrospective/prospective. Given this is a 510(k) submission for a Chinese manufacturer (Xiantao Daoqi Plastic Co., Ltd.) and tests are cited against international (ISO) and US (ASTM) standards, the testing was likely conducted in laboratories, possibly in China or other accredited facilities. The nature of these tests (bench testing) implies they are prospective assessments of product samples.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable to this type of submission. The ground truth for these medical device tests (surgical face mask performance and biocompatibility) is established by adherence to recognized international and national standards (ASTM, ISO). These standards define specific methodologies and acceptance criteria. No human experts are involved in establishing "ground truth" in the way they would for a diagnostic AI study. The "experts" are the technicians performing the standardized tests and interpreting the results against the defined criteria.

4. Adjudication method for the test set

This information is not applicable. The tests are objective measurements against predefined criteria in recognized standards. There is no adjudication process involving multiple human observers or adjudicators for these types of physical and chemical tests.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices or AI-powered clinical decision support systems, not for a physical device like a surgical face mask.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No, a standalone (algorithm only) performance study was not done. This is not an AI/algorithm-based device. The device is a physical surgical face mask.

7. The type of ground truth used

The "ground truth" for this device is established by objective measurements and assessments against recognized performance and biocompatibility standards (ASTM F2100-23, ASTM F2101-23, ASTM F1862/F1862M-17, ISO 10993-5, ISO 10993-23, ISO 10993-10). These standards define the acceptable range of physical properties and biological responses for surgical masks.

8. The sample size for the training set

Not applicable. This is a physical device, not an AI or machine learning model, so there is no "training set."

9. How the ground truth for the training set was established

Not applicable. As there is no training set, there is no ground truth for a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 18, 2024

Xiantao Daoqi Plastic Co., Ltd. Honghong Bie Quality Manager No.83, Wagou New District Pengchang Town Xiantao, Hubei 433018 China

Re: K240916

Trade/Device Name: Surgical Face Mask (Tie on/ Ear loops) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: September 12, 2024 Received: September 12, 2024

Dear Honghong Bie:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Allan Guan -S

For Bifeng Qian, M.D., Ph.D. Assistant Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K240916

Device Name Surgical Face Mask (Tie on/ Ear loops)

Indications for Use (Describe)

Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non sterile.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

| | Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary (K240916)

This summary of 510(k) safety and effectiveness information is being submitted in accordance with 21 CFR 807.92.

1.0 Submitter's information

Name: XIANTAO DAOQI PLASTIC CO., LTD. Address: No.83, Wagou New District, Pengchang Town, Xiantao City, Hubei Province, 433018, China Phone Number: +86-728 2620008 Contact: Ms. Bie Honghong Date of Preparation: 10/18/2024

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device information

Trade name:Surgical Face Mask
Common name:Surgical Mask
Classification name:Mask, Surgical
Model(s):Ear loops (17.5cm×9.5cm)
Tie-on (17.5cm×9.5cm)

3.0 Classification

Production code: FXX Regulation number: 21CFR 878.4040 Classification: Class II Panel: Surgical apparel

4.0 Predicate device information

Manufacturer: Hubei Qianjiang Kingphar Medical Material Co., Ltd Device: Surgical face mask 510(k) number: K220777

{5}------------------------------------------------

5.0 Indication for Use Statement

Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non sterile.

6.0 Device description

The Surgical face mask is single use, three-layer, flat-pleated style with ear loops/ties and nose piece. The mask is manufactured with three layers, the inner and outer layers, and the straps are made of nonwoven fabrics, the middle layer is made of melt blown fabrics.

The model of proposed device, tie-on, is held in place over the user's mouth and nose by four ties welded to the face mask. The tie is made of spun-bond polypropylene.

The model of proposed device, ear loops, is held in place over the user's mouth and nose by two elastic ear loops welded to the face mask. The elastic ear loops are not made with natural rubber latex.

The ties/ear loops are not made with natural rubber latex. The nose piece on the layers of face mask is to allow the user to fit the face mask around their nose, which is made of Polypropylene with iron core. The Surgical face mask will be provided in blue. The masks are sold non-sterile and are intended to be single use, disposable devices.

7.0 Technological Characteristic Comparison Table

Table 1 - General Comparison

ItemProposed devicePredicate deviceComparison
Product CodeFXXFXXSame

{6}------------------------------------------------

Regulation No.21 CFR 878.404021 CFR 878.4040Same
ClassIIIISame
Product nameSurgical Face MaskSurgical Face MaskSame
510(k) No.K240916K220777-
Mask stylesEar loops/Tie onEar loops/Tie onSame
Intended UseThe surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non sterile.The Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non sterile.Same
OTC useYesYesSame
Design FeaturesFlat Pleated, 3 layersFlat Pleated, 3 layersSame
MaterialsInternal layerSpun-bond polypropyleneSpun-bond polypropyleneSame
Middle layerMelt blown polypropyleneMelt blown polypropyleneSame
External layerMelt blown polypropyleneSpun-bond polypropyleneSame
Nose piecePolypropylene+iron corePolypropylene+Galvanize d-iron dual coreSimilar
Ear loopsPolyester+spandexPolyester, spandex* Gap 1
TiesSpun-bond polypropylene
ColorBlueBlueSame
Dimension (Length)17.5cm±0.5cm17.5cm±0.5cmSame
Dimension (Width)9.5cm±0.5cm9.5cm±0.5cmSame
SterilityNon-SterileNon-SterileSame
Single UseYesYesSame
SterileNoNoSame
ASTM F2100 LevelLevel 3Level 3Similar
Shelf Life2 years5 yearsDifferent
  • Gap analysis:

Gap 1: the two devices have some difference in materials, product materials safety is proved by its biocompatibility, and the difference does not raise additional questions for safety and effectiveness of device.

ItemProposed devicePredicate deviceRemark
Bacterial filtrationefficiency (BFE) (%)ASTM F2100-23 Level 3, ≥98ASTM F2100-19 Level 3, ≥98Same
Different pressure(mmH2O/cm²)ASTM F2100-23 Level 3, <6.0ASTM F2100-19 Level 3, <6.0

Table 2 Performance Comparison

{7}------------------------------------------------

Sub-micron particulatefiltration efficiency at0.1 micron, % (PFE)ASTM F2100-23 Level 3, ≥85ASTM F2100-19 Level 3, ≥98Similar
Resistance topenetration by syntheticblood, Minimumpressure in mmHg forpass resultASTM F2100-23 Level 3, 160ASTM F2100-19 Level 3, 160Same
Flame spreadClass 1Class 1Same

Biocompatibility Comparison Table 3

ItemProposed DevicePredicate DeviceRemark
CytotoxicityComply with ISO 10993-5Comply with ISO 10993-5Same
IrritationComply with ISO 10993-23Comply with ISO 10993-10Same
SensitizationComply with ISO 10993-10Comply with ISO 10993-10Same

8.0 Summary of Non-Clinical Testing

The proposed device was tested and conformed to the related recognized standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks - Premarket Notification [510(k)] Submission issued on March 5, 2004 and following recognized standards:

Biocompatibility Testing

2-245 ISO 10993-5 Third edition 2009-06-01 Biological Evaluation of Medical Devices -Part 5: Tests For In Vitro Cytotoxicity

2-291 ISO 10993-23 First edition 2021-01 Biological evaluation of medical devices- Part 23: Tests for irritation

2-296 ISO 10993-10 Fourth edition 2021-11 Biological Evaluation of Medical Devices -Part 10: Tests for Skin Sensitization.

Performance Testing (Bench)

6-492 ASTM F2100-23 Standard Specification for Performance of Materials Used in Medical Face Masks 6-493 ASTM F2101-23 Standard Test Method of Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of

Staphylococcus aureus

{8}------------------------------------------------

6-406 ASTM F1862/F1862M-17 Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)

ItemsPurposeAcceptance CriteriaLevel 3( ASTM F2100-23)Result
Bacterial filtrationefficiency (BFE) (%)The purpose of the test isto evaluate the Bacterialfiltrationefficiency (BFE) (%)≥9899.7%-99.9%Pass
Different pressure(mmH2O/cm2)The purpose of the test isto evaluate the Differentpressure(mmH2O/cm2)<6.0 mmH2O/cm²3.2-3.7mmH2O/cm²Pass
Sub-micron particulatefiltration efficiency at0.1 micron, % (PFE)The purpose of the test isto evaluate theSub-micron particulatefiltration efficiency at 0.1micron, % (PFE)≥8597.16-99.52%Pass
Resistance topenetration bysynthetic blood,Minimum pressure inmmHg for pass resultThe purpose of the test isto evaluate theResistance to penetrationby synthetic blood,Minimum pressure inmmHg for pass result29 of 32 test articles passedat 160mmHg32 of 32 testarticlesPass
Flame spreadThe purpose of the test isto evaluate theFlammabilityClass 132 of 32 testarticlesPass

Table 4 - Performance Testing

Table 5 - Biocompatibility Testing

ItemsPurposesAcceptance CriteriaResultSummary
CytotoxicityTo evaluate therisk of the sampleto induce biologicalresponse of L-929ISO 10993-5:2009Viability < 70 %, it hasa cytotoxic potential.Tie on:the viability ratio of the100% test article extractwas 91.0%, no potentialPass
cellstoxicity of L-929 cells.
Ear loops:the viability ratio of the100% test article extractwas 70.5%, no potentialtoxicity of L-929 cells.
IrritationTo evaluate thepotential of skinirritation ofsamples under ofthe test conditionsISO 10993-23:2021Tie on:The primary irritationindexes of the polar andnon-polar test group wereboth calculated to be 0.The extract of the testarticle did not induce skinirritation.Ear loops:The primary irritationindexes of the polar andnon-polar test group wereboth calculated to be 0.The extract of the testarticle did not induce skinirritation.Pass
If the primary irritationindex is 0-0,4, theresponse category isNegligible.0,5-1,9 means slight2-4,9 means moderate5-8 means severePrimaryIrritation-Index≤ 2.0
SensitizationTo evaluate thepotential of a testarticle to causeskin sensitizationusing Guinea PigMaximization TestISO 10993-10:2021Tie on:The skin sensitization ratesof polar and non-polarextract group were bothdetermined with 0%.Ear loops:The skin sensitization ratesof polar and non-polarextract group were bothdetermined with 0%Pass
provided grades lessthan 1, otherwisesensitization.

{9}------------------------------------------------

9.0 Summary of Clinical Testing

No clinical study conducted for the Surgical Face Mask.

10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.