K Number
K240877

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-06-04

(67 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Monteris Medical NeuroBlate System is a neurosurgical tool and is indicated for use to ablate, necrotize, or coagulate intracranial soft tissue, including brain structures (e.g., brain tumor, radiation necrosis, and epileptogenic foci as identified by non-invasive and invasive neurodiagnostic testing, including imaging), through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers.

The Monteris Medical NeuroBlate System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI-based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate Laser Delivery Probes. It also provides near real-time thermographic analysis of selected MRI images.

When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate System analysis. The intended patients are adults and pediatric from the age of 2 years and older.

Device Description

The Monteris NeuroBlate System is a collection of MRI-compatible laser devices and accessories that create an MRI guided delivery of precision thermal therapy in the practice of neurosurgery.

The NeuroBlate System components consist of:

  • Families of gas-cooled Laser Delivery Probe (Probe) (SideFire & FullFire) to deliver controlled energy to a target zone.
  • Probe Drivers (Advanced Probe Driver, Robotic Probe Driver) which allow the surgeon to precisely position, stabilize and manipulate a probe, endoscope or other device within the target zone.
  • An Interface Platform, which attaches to the MRI system patient table and provides supporting electronics for the Advanced and Robotic Probe Drivers and interconnections for the Laser Delivery Probes;
  • A System Electronics Rack and Components, which includes necessary umbilicals, cables, penetration panels, and small hardware for system mechanical, electrical, and electronic operation.
  • A Control Workstation including the M-Vision™ and M-Vision Pro™ software, which includes a user interface for procedure planning, interactive monitoring of NeuroBlate procedures, and interfaces to the MRI and hardware subsystems.

The NeuroBlate System is utilized with stereotaxic frames and patient stabilization systems, such as:

The Monteris Cranial Bolt and Mini-Bolt fixation components, and

  • The AtamA Stabilization System and MRI receive-only head coil, as well as, other optional accessories, including: drill bits, bolts, thumbscrews, instrument adaptors, accessory host adaptors, MRI trajectory wands, cranial screws, bone screws, fiducial markers, stereotactic manual driver with mandrel and T-handle, and other manual accessory instruments and tools.
AI/ML Overview

The provided text is an FDA 510(k) clearance letter for the Monteris Medical NeuroBlate System (K240877). This document primarily focuses on demonstrating substantial equivalence to a predicate device (K231061) and modifying the pediatric age range in the Indications for Use. It explicitly states that "no additional testing was required" because the changes are labeling-related and "no physical changes, manufacturing changes, process changes, materials changes, or technology changes" were made to the system.

Therefore, the document does not contain details about specific acceptance criteria, a study proving the device met those criteria, sample sizes for test or training sets, expert qualifications, adjudication methods, MRMC studies, or standalone performance studies for this specific submission (K240877). The information provided heavily references data from the predicate device (K231061) and general clinical literature reviews.

Based on the provided text, here's what can be extracted:


1. Table of Acceptance Criteria and Reported Device Performance

This information is not provided in the document as no new acceptance criteria or device performance data were generated for this specific submission (K240877). The submission focuses on a labeling change to an existing device, relying on previous data.

2. Sample Size Used for the Test Set and Data Provenance

This information is not provided for this specific submission as no new testing was required. The document mentions that "adult and pediatric data from Monteris' prospective LAANTERN Registry was also reviewed," but it does not specify the sample size from this registry or explicitly define it as a "test set" for this submission. The provenance is from Monteris' prospective registry.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided for this specific submission as no new testing was required.

4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

This information is not provided for this specific submission as no new testing was required.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

An MRMC study is not mentioned for this specific submission. The NeuroBlate System is a physical laser thermal therapy system with software for planning and monitoring, not an AI diagnostic tool for interpreting images for human readers.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

A standalone performance study is not mentioned for this specific submission. The device is intended to be used by a "trained physician" for planning and monitoring, with patient management decisions "not made solely on the basis of the NeuroBlate System analysis," indicating a human-in-the-loop system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

This information is not explicitly stated for any data underlying the predicate device, other than general references to "LITT related clinical literature" and "NeuroBlate System commercial clinical experience," and data from the "LAANTERN Registry" (which would likely include outcomes data).

8. The sample size for the training set

This information is not provided for this specific submission as no new technical development or training of algorithms is indicated; the submission relies on previously established performance of the predicate device.

9. How the ground truth for the training set was established

This information is not provided for this specific submission as no new technical development or training of algorithms is indicated.


Summary of Change in K240877 and Rationale:

The core of this 510(k) submission (K240877) is a labeling change to the existing Monteris Medical NeuroBlate System (predicate K231061). The change is to explicitly include "pediatric patients from the age of 2 years and older" in the "Indications for Use" statement, clarifying an existing "adult and pediatric" intended patient population.

The submission asserts that because "no physical changes, manufacturing changes, process changes, materials changes, or technology changes" were made to the device, and the "technical modes of action and technical principles remain the same," no additional testing (in-vitro or in-vivo) was required. The conclusion that the device remains safe and effective for this expanded age range is based on:

  • Review of LITT related clinical literature.
  • Review of Monteris' own NeuroBlate System commercial clinical experience.
  • Review of adult and pediatric data from Monteris' prospective LAANTERN Registry.

The document explicitly states: "Monteris has concluded from these reviews that the NeuroBlate LITT system has been, and remains, an effective and safe tool for use in pediatric patients." This serves as the "study" for this submission, validating the labeling change rather than a new device performance study. However, specific details about the methods or data from these reviews are not included in this FDA letter.

{0}------------------------------------------------

June 4, 2024

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Monteris Medical David H. Mueller Senior Principal Regulatory Affairs Specialist 131 Cheshire Lane, Suite 100 Minnetonka, Minnesota 55305

Re: K240877

Trade/Device Name: Monteris Medical NeuroBlate System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: ONO, GEX, HAW Dated: March 29, 2024 Received: March 29, 2024

Dear David H. Mueller:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Image /page/1/Picture/5 description: The image shows a digital signature and contact information for Adam D. Pierce. The signature indicates it was digitally signed on June 4, 2024, at 12:34:20 -04'00'. Adam D. Pierce is listed as an Assistant Director in the DHT5A division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices, as well as the OHT5 office of Neurological and Physical Medicine Devices, and the Office of Product Evaluation and Quality.

{2}------------------------------------------------

Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K240877

Device Name Monteris Medical NeuroBlate System

Indications for Use (Describe)

The Monteris Medical NeuroBlate System is a neurosurgical tool and is indicated for use to ablate, necrotize, or coagulate intracranial soft tissue, including brain structures (e.g., brain tumor, radiation necrosis, and epileptogenic foci as identified by non-invasive and invasive neurodiagnostic testing, including imaging), through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers.

The Monteris Medical NeuroBlate System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI-based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate Laser Delivery Probes. It also provides near real-time thermographic analysis of selected MRI images.

When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate System analysis. The intended patients are adults and pediatric from the age of 2 years and older.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

CategoryComments
Sponsor / Submitter:Monteris Medical131 Cheshire Lane, suite 100Minnetonka, MN 55305866-799-7655www.monteris.com
Correspondent ContactInformation:David H. MuellerSenior Principal Regulatory Affairs SpecialistMonteris MedicalTEL: 763-333-1614Email: DMueller@Monteris.com
Device Common Name:Magnetic Resonance Image Guided Laser ThermalTherapy System
Device ClassificationRegulation & Name:21 CFR 878.4810Laser surgical instrument for use in general and plastic surgery and in dermatology Neurosurgical Laser With MR Thermography21 CFR 882.4560Stereotaxic instrument
Device Classification &Product Code:Class II, HAWClass II, ONOClass II, GEX
Device Proprietary Name:Monteris Medical NeuroBlate® System

a. Device Information:

Predicate Device Information:

ManufacturerMonteris Medical
Commercial NameNeuroBlate System
Common Name:Laser surgical instrument for use in general and plastic surgery and indermatology;Stereotaxic instrument
PremarketNotification #K231061
Regulation21 CFR 878.481021 CFR 882.4560
Class/ Product CodeClass II; ONO, GEX, HAW

b. Date Summary Prepared

21May2024

c. Description of Device

The Monteris NeuroBlate System is a collection of MRI-compatible laser devices and accessories that create an MRI guided delivery of precision thermal therapy in the practice of neurosurgery.

{5}------------------------------------------------

The NeuroBlate System components consist of:

  • Families of gas-cooled Laser Delivery Probe (Probe) (SideFire & FullFire) to deliver controlled energy to a target zone.
  • Probe Drivers (Advanced Probe Driver, Robotic Probe Driver) which allow the surgeon to precisely position, stabilize and manipulate a probe, endoscope or other device within the target zone.
  • An Interface Platform, which attaches to the MRI system patient table and provides supporting electronics for the Advanced and Robotic Probe Drivers and interconnections for the Laser Delivery Probes;
  • A System Electronics Rack and Components, which includes necessary umbilicals, cables, penetration panels, and small hardware for system mechanical, electrical, and electronic operation.
  • A Control Workstation including the M-Vision™ and M-Vision Pro™ software, which includes a user interface for procedure planning, interactive monitoring of NeuroBlate procedures, and interfaces to the MRI and hardware subsystems.

The NeuroBlate System is utilized with stereotaxic frames and patient stabilization systems, such as:

The Monteris Cranial Bolt and Mini-Bolt fixation components, and

  • The AtamA Stabilization System and MRI receive-only head coil, as well as, other optional accessories, including: drill bits, bolts, thumbscrews, instrument adaptors, accessory host adaptors, MRI trajectory wands, cranial screws, bone screws, fiducial markers, stereotactic manual driver with mandrel and T-handle, and other manual accessory instruments and tools.

d. Indications for Use/ Intended Use

The Monteris Medical NeuroBlate System® is a neurosurgical tool and indicated for use to ablate, necrotize, or coagulate soft intracranial tissue including brain structures (e.g., brain tumor, radiation necrosis, and epileptogenic foci as identified by non-invasive and invasive neurodiagnostic testing, including imaging), through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers.

The Monteris Medical NeuroBlate System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate® Laser Delivery Probes. It also provides real-time thermographic analysis of selected MRI images.

When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate System analysis. The intended patients are adults and pediatric from the age of 2 years and older.

e. Comparison to Predicate Device

The application for the Monteris Medical NeuroBlate System with the modified labeling is substantially equivalent to the predicate Monteris NeuroBlate System (K231061) in intended use, technology, design and physician use. Reference Table 1.

{6}------------------------------------------------

This submission proposes no physical changes to the NeuroBlate System hardware, software, accessories, operating principles or functionality.

The proposed change defines the existing "adult and pediatric" intended patient population to include both adults and pediatric patients from the age of 2 years and older.

The proposed indication presents no (zero) physical or technological changes to the predicate device, and no changes to the Indications for Use (except for adding "The intended patients are adults and pediatric from the age of 2 years and older" to the Indications for Use) in the Instructions for Use/ Directions for Use.

As the modified device presents no additional or different risks or technological characteristics when compared to the predicate devices, no labelling modifications are necessary, e.g., there are no (zero) labelling changes to the Device Description, Contraindications, Warnings, and Precautions, potential Adverse Events sections.

Additionally, there are no manufacturing, process, material, or technology proposed changes to the NeuroBlate System. The technical modes of action and technical principles remain the same as the predicate NeuroBlate device. Considering the proposed changes are labeling related, the previously provided in-vitro (bench) data remains applicable and no additional bench testing was required.

The fundamental functionality and technical characteristics of the proposed modified device is identical to the existing NeuroBlate System (K231061). The conclusion is that the difference in technological characteristics between proposed modified device and the described predicate devices do not raise new or different questions of safety and effectiveness, i.e., the proposed modified device is as safe as the described predicate devices.

f. Summary of Supporting Data

This submission proposes no physical changes, manufacturing changes, process changes, materials changes, or technology changes to the NeuroBlate System. Considering the proposed changes are labeling, the previously provided in-vitro (bench), in-vivo (animal) data remains applicable and no additional testing was required.

Multiple FDA Guidance documents were utilized to ensure substantial equivalency, e.g.:

  • United States Food and Drug Administration. Deciding When to Submit a 510(k) for a ● Change to an Existing Device. Washington, D.C .: FDA, 2017.
  • United States Food and Drug Administration. Guidance for Industry and Food and Drug . Administration Staff, General/Specific Intended Use. Washington, D.C .: FDA, 1998

{7}------------------------------------------------

  • . United States Food and Drug Administration. Premarket Assessment of Pediatric Medical Devices Guidance, Guidance for Industry and Food and Drug Administration Staff. March 24th edition. Washington, D.C. 2014
    The proposed modifications to the NeuroBlate System's Indications For Use (IFU) are to provide clarification for the existing terminology utilized within the existing indications as the device is currently utilized. The proposed changes are not intended to provide specific indications for use which are different than the current language.

In this submission. Monteris monitors LITT related clinical literature, as well as our own NeuroBlate System commercial clinical experience, to ensure our current instructions and device remain appropriate for their use. Additionally, adult and pediatric data from Monteris' prospective LAANTERN Registry was also reviewed. Monteris has concluded from these reviews that the NeuroBlate LITT system has been, and remains, an effective and safe tool for use in pediatric patients.

g. Conclusion

Monteris Medical's Quality Systems utilize various Risk Assessment and Risk Mitigation methodologies. As there are no physical changes, manufacturing changes, process changes, materials changes, or technology changes to the NeuroBlate System, and that the technical modes of action and technical principles remain the same as the predicate devices, and that the proposed changes are labeling related, the previously provided (K231061) Risk Assessment and Risk Mitigation documentation remains applicable. This supporting information demonstrates that the subject NeuroBlate system is as safe and effective as the predicate device.

{8}------------------------------------------------

CharacteristicUNDER REVIEWPRIMARY PREDICATES: K231061Current Monteris NeuroBlate System Laser Probes
IndicationsFor UseNo change fromMonteris predicate.The Monteris Medical NeuroBlate System is a neurosurgical tool and is indicatedfor use to ablate, necrotize, or coagulate intracranial soft tissue, including brainstructures (e.g., brain tumor, radiation necrosis, and epileptogenic foci as identifiedby non-invasive and invasive neurodiagnostic testing, including imaging), throughinterstitial irradiation or thermal therapy in medicine and surgery in the disciplineof neurosurgery with 1064 nm lasers.The Monteris Medical NeuroBlate System is intended for planning and monitoringthermal therapies under MRI visualization. It provides MRI-based trajectoryplanning assistance for the stereotaxic placement of MRI compatible (conditional)NeuroBlate Laser Delivery Probes. It also provides near real-time thermographicanalysis of selected MRI images.When interpreted by a trained physician, this System provides information that maybe useful in the determination or assessment of thermal therapy. Patientmanagement decisions should not be made solely on the basis of the NeuroBlateSystem analysis.
(ADDED)The intended patientsare adults andpediatric from the ageof 2 years and older.Not Applicable
IFU StatedPatientPopulationStated in Indicationsfor UseNot stated
ClinicalEvidenceProvided inSubmissionYesYes

Ca mnori nival .............................................................................................................................................................................. Table 1· Substantial Fa

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.