K Number
K240808

Validate with FDA (Live)

Date Cleared
2024-10-03

(192 days)

Product Code
Regulation Number
870.2700
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

WS20A is a pulse oximeter indicated for use in measuring, displaying, storing and transmitting functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for patients(≥10 years old, ≥30kg). It is designed for finger circumference more than 33mm. It is intended for spot-check, continuous data collection, recording and transmitting, not continuous monitoring. It can be used in sleep labs, long-term care, hospitals and home.

Device Description

Pulse Oximeter WS20A is an internally powered pulse oximeter. The main functions of the devices include hemoglobin oxygen saturation (SpO2), pulse rate (PR) measurements and Pulse amplitude index (PAI), data storage and transmission. Place one fingertip into the sensor and the oxygen saturation (SpO2), pulse rate (PR) measurements and pulse amplitude index (PAI) will appear on the display. The device is intended to be applied to adult and pediatric patients in sleep labs, long-term care, hospital and home care environment. The subject device is composed of the following components to achieve the above detection process: power supply module, detector and emitter, signal collection and process module (MCU), TFT display screen and Bluetooth module.

AI/ML Overview

The provided text describes the 510(k) summary for the Hunan Accurate Bio-Medical Technology Co., Ltd.'s Pulse Oximeter (WS20A). While it details the device, its intended use, and comparison to a predicate device, it does not contain a specific table of acceptance criteria with reported device performance in the format of typical medical device performance claims (e.g., sensitivity, specificity, accuracy for an AI/diagnostic device).

Instead, the performance criteria for this pulse oximeter are related to its accuracy in measuring SpO2 and Pulse Rate against a reference standard (CO-Oximetry). The acceptance criteria for pulse oximeters are typically defined by recognized standards such as ISO 80601-2-61, which specifies the accuracy requirements (e.g., ARMS value).

Based on the provided text, here's an attempt to extract the relevant information and present it as requested.


Overview of Device Performance and Study

The Pulse Oximeter (WS20A) underwent clinical studies to verify its accuracy in measuring functional oxygen saturation of arterial hemoglobin (SpO2) against arterial blood CO-Oximetry, in accordance with ISO 80601-2-61:2017 and FDA guidance for Pulse Oximeters - Premarket Notification Submissions.

1. Table of Acceptance Criteria and Reported Device Performance

For pulse oximeters, the key performance metric for SpO2 accuracy is typically the Accuracy Root Mean Square (ARMS) value when compared to a reference method (CO-Oximetry). The acceptance criterion is generally that the ARMS value should be within a specified limit, commonly ± 3% for 70-100% SpO2 range as per ISO 80601-2-61.

Performance MetricAcceptance Criteria (ARMS for SpO2, 70-100% range)Reported Device Performance (ARMS for SpO2, 70-100% range)
SpO2 Accuracy$\leq \pm 3$% (derived from ISO 80601-2-61)Adult Patients: - All subjects: 1.81% - Female subjects: 1.80% - Male subjects: 1.83% - Light-skinned: 1.78% - Dark-skinned: 1.91% Pediatric Patients: - 10 female subjects: 1.80% - Light-skinned: 1.81% - Dark-skinned: 1.74%
PR Accuracy (bpm)$\pm$ 3bpm (as per predicate device spec)Not explicitly reported from clinical study for WS20A. The 510(k) submission document for WS20A lists its PR accuracy as $\pm$ 3 bpm, which aligns with its own stated specifications and is compared to a similar specification for the predicate.

Note: The acceptance criteria for ARMS are not explicitly stated as a number in the provided text but are inferred from common regulatory expectations for pulse oximeters based on the ISO standard cited.

2. Sample Size Used for the Test Set and Data Provenance

  • Adult Patients: 13 healthy adult subjects. 315 data sets (SpO2 vs SaO2) were obtained.
  • Pediatric Patients: 10 healthy female subjects. 243 data sets (SpO2 vs SaO2) were obtained. (Note: The description states "healthy female subjects, aged 22-30" for 'pediatric patients' which seems to be a typo given the 'pediatric' labeling and the age range overlaps with 'adult'. Assuming this refers to subjects representing a pediatric patient population as defined by the study design, or a mislabeling of the subject group based on physical characteristics matching the pediatric profile for device testing.)
  • Data Provenance: The document does not specify the country of origin of the data. The studies were described as "clinical studies" and appear to be prospective in nature, as they involved actively obtaining data sets from healthy subjects under stationary (non-motion) conditions through controlled desaturation.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The ground truth for pulse oximetry accuracy studies is established through arterial blood CO-Oximetry, which is a direct invasive measurement of arterial oxygen saturation (SaO2). This is a gold standard laboratory method, not typically established by physician experts in the same way as image interpretation. The text does not mention the use of experts for ground truth establishment; rather, it relies on the direct physiological measurement.

4. Adjudication Method for the Test Set

Not applicable. Ground truth for oxygen saturation in these studies is derived from direct physiological measurement (CO-Oximetry) rather than expert consensus requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a pulse oximeter for direct physiological measurement, not an AI-assisted diagnostic imaging device requiring human reader interpretation for a comparative effectiveness study.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, the performance presented (ARMS values) represents the standalone accuracy of the pulse oximeter device algorithm in measuring SpO2 against the CO-Oximetry reference. While a human uses the device, the accuracy is derived from the device's measurement capabilities.

7. The Type of Ground Truth Used

The ground truth used was arterial blood CO-Oximetry (SaO2), which is a gold standard for measuring oxygen saturation in the blood. This is a direct physiological measurement, not expert consensus or pathology.

8. The Sample Size for the Training Set

The document describes clinical studies conducted to verify the accuracy of the proposed device. It does not explicitly mention a "training set" for an algorithm in the sense of machine learning. The data described (from 13 adult and 10 "pediatric" subjects) appears to be the test/validation set used to demonstrate the device's accuracy. For traditional medical devices like pulse oximeters, the core measurement principles are physics-based, not reliant on machine learning models that require distinct training sets. Any internal calibration or algorithm development would typically occur during the R&D phase, prior to these described validation studies, and the size of data used for such internal development is not generally disclosed in 510(k) summaries unless it's a novel AI/ML device.

9. How the Ground Truth for the Training Set was Established

Not applicable, as a distinct "training set" in the context of an AI/ML device is not described for this traditional pulse oximeter. If there were internal algorithm development or calibration, the ground truth would likely have been established using similar reference methods (CO-Oximetry) in controlled lab settings.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and name on the right. The FDA acronym and name are in blue, with the acronym in a larger font size than the name. The name is written as "U.S. FOOD & DRUG ADMINISTRATION".

October 3, 2024

Hunan Accurate Bio-Medical Technology Co., Ltd. % Jie Yang Consultant Chonconn Medical Device Consulting Co., Ltd. Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District Shenzhen, Guangdong 518067 China

Re: K240808

Trade/Device Name: Pulse Oximeter (WS20A) Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: March 25, 2024 Received: March 25, 2024

Dear Jie Yang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.

{2}------------------------------------------------

See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bradley Q. Quinn -S

Bradley Quinn Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Submission Number (if known)

K240808

Device Name

Pulse Oximeter (WS20A)

  • Indications for Use (Describe)
    WS20A is a pulse oximeter indicated for use in measuring, displaying, storing and transmitting functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for patients(≥10 years old. ≥30kg). It is designed for finger circumference more than 33mm. It is intended for spot-check, continuous data collection, recording and transmitting, not continuous monitoring. It can be used in sleep labs, long-term care, hospitals and home.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

{4}------------------------------------------------

510(K) Summary

Prepared in accordance with the requirements of 21 CFR Part 807.92

Prepared Date: 2024/10/02

1. Submission sponsor

Name: Hunan Accurate Bio-Medical Technology Co., Ltd. Address: Accurate Industrial Park, No.108, Zhixian Road, Xuelian Community, Xueshi Street of Yuelu District, 410208 Changsha, Hunan Province, PEOPLE'S REPUBLIC OF China Contact person: Zhang Li Title: General Manager E-mail: regulation@accbiomed.com Tel: +86 731-8559853

2. Submission correspondent

Name: Chonconn Medical Device Consulting Co., Ltd.

Address: Room 504, Block C, No. 1029 Nanhai Avenue, Nanshan District, Shenzhen, Guangdong, P.R. China

Contact person: Yang Jie E-mail: yangjie(@chonconn.com

Tel: +86-755 33941160

Trade/Device NamePulse Oximeter
ModelWS20A
Common NameWrist Pulse Oximeter
Regulatory ClassClass II
Classification21CFR 870.2700 / Oximeter / DQA
Submission typeTraditional 510(K)

3. Subject Device Information

4. Predicate Device

Manufacturer: Beijing Choice Electronic Technology Co., Ltd. Device name: Wrist Pulse Oximeter,Model: MD300W314 510(K) Number: K172366

5. Device Description

Pulse Oximeter WS20A is an internally powered pulse oximeter. The main functions of the devices include hemoglobin oxygen saturation (SpO2), pulse rate (PR) measurements and Pulse amplitude index (PAI), data storage and transmission.

{5}------------------------------------------------

Place one fingertip into the sensor and the oxygen saturation (SpO2), pulse rate (PR) measurements and pulse amplitude index (PAI) will appear on the display. The device is intended to be applied to adult and pediatric patients in sleep labs, long-term care, hospital and home care environment.

The subject device is composed of the following components to achieve the above detection process: power supply module, detector and emitter, signal collection and process module (MCU), TFT display screen and Bluetooth module.

6. Intended use & Indication for use

WS20A is a pulse oximeter indicated for use in measuring, displaying, storing and transmitting functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for patients(≥10 years old, ≥30kg). It is designed for finger circumference more than 33mm. It is intended for spot-check, continuous data collection, recording and transmitting, not continuous monitoring. It can be used in sleep labs, long-term care, hospitals and home.

ITEMProposed DevicePredicate DeviceComparison Result
Pulse OximeterWS20AK240808Wrist Pulse Oximeter,Model: MD300W314K172366
ManufactureHunan Accurate Bio-Medical Technology Co.,Ltd.Beijing Choice ElectronicTechnology Co., Ltd.Same
Indications forUseWS20A is a pulseoximeter indicated for usein measuring, displaying,storing and transmittingfunctional oxygensaturation of arterialhemoglobin (SpO2) andpulse rate for patients(≥10years old, ≥30kg). It isdesigned for fingercircumference more than33mm. It is intended forspot-check, continuousdata collection, recordingand transmitting, notcontinuous monitoring. Itcan be used in sleep labsMD300W314 is a wristpulse oximeter indicated foruse in measuring,displaying, storing andtransmitting functionaloxygen saturation of arterialhemoglobin (SpO2) andpulse rate for adult,adolescent, child and infantpatients. It is intended forspot-check and / or datacollection, recording andtransmitting. It can be usedin sleep labs, long-term care,hospitals and home use.Similar
long-term care, hospitalsand home.
Populationadult, adolescent and childpatientsadult, adolescent, child andinfant patientsSimilar
ComponentsPower supply module,detector and emitter LED,signal collection andprocessor module, displaymodule, Bluetoothmodule.Power supply module,detector and emitter LED,signal collection andprocessor module, displaymodule, Bluetooth module.Same
Type of SpO2SensorTransmittance OpticalSensorTransmittance OpticalSensorSame
ApplicationSiteFingerFingerSame
MeasurementWavelengthRed (660 nm)Infrared (905 nm)Red (660 nm)Infrared (905 nm)Same
Display TypeTFTLCDDifference
Power SupplyLithium-ion rechargeablebatteryLithium-ion rechargeablebatterySame
Display DataSPO2 , PR , PAISPO2 , PRSimilar
SPO2SpO2 Display Range:0%100%Measurement range:70100%Accuracy: 70%~100%:±2%;SpO2 Display Range:0%~100%Measurement range:70%~100%Accuracy: 70%~100%:±2%;Same
PRPR Display Range25250bpmMeasurement range:25250bpmResolution: 1bpmAccuracy: $\pm$ 3bpm;PR Display Range30255bpmMeasurement range:30250bpmResolution: 1bpmAccuracy: 3099bpm, $\pm$ 2bpm; 100250bpm, $\pm$ 2%Similar
Pulseamplitudeindex(PAI)Display range: 0.120.0%Measurement range:0.320.0%Resolution: 0.1%Accuracy: 0.31.0% ( $\pm$ 0.2digits); 1.120.0% ( $\pm$ 20%)N/ADifference
NetworkBluetoothBluetoothSame
EnvironmentRequirementsOperating Temperature:5~40 $^{\circ}$ CAmbient Humidity:10%90%, nocondensationStorage/Transportation: -20 $^{\circ}$ C60 $^{\circ}$ C AmbientHumidity: 10%~90% nocondensationOperating Temperature:5~40 $^{\circ}$ CAmbient Humidity:15%93%, no condensationStorage/Transportation: -25 $^{\circ}$ C70 $^{\circ}$ C AmbientHumidity: $\leq$ 93% nocondensationSimilar
ContactingMaterialEnclosure: ABSWatchband: SiliconeFingertip sensor: SiliconeEnclosure: ABSWrist oximeter wrist belt:Nylon brandFingertip sensor: SiliconeSimilar
ElectricalConformed to IEC60601-Conformed to IEC60601-1,Same
  1. Comparison to the Predicate Device

{6}------------------------------------------------

{7}------------------------------------------------

Non-clinical Data 8.

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

The biocompatibility evaluation for the Pulse Oximeter was conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".

{8}------------------------------------------------

Bench testing

The Pulse Oximeter has been tested according to the following standards:

  • IEC 60601-1-2005+CORR.1:2006+CORR.2:2007+A1:2012, Medical Electrical Equipment-. Part 1: General requirements for basic safety and essential performance
  • IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic . safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests
  • IEC 60601-1-11 Edition 2.0 2015-01Medical electrical equipment Part 1-11: General . requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
  • ISO 80601-2-61: 2017 Medical Electrical Equipment Part 2-61: Particular Requirements for . Basic Safety and Essential Performance of Pulse Oximeter Equipment.

Clinical data

Clinical studies were conducted to verify the accuracy of proposed device. The clinical studies were conducted per following standards:

  • ISO 80601-2-61: 2017 Medical Electrical Equipment - Part 2-61: Particular Requirements for Basic Safety and Essential Performance of Pulse Oximeter Equipment.

  • Pulse Oximeters-Premarket Notification Submissions: Guidance for Industry and Food and Drug Administration Staff

One clinical study is conducted for adult and pediatric patients. The purpose of the clinical trial is to validate accuracy of the pulse oximetry (SpO2) of the WS20A pulse oximeter manufactured by Hunan Accurate Bio-Medical Technology Co. Ltd. during stationary (non-motion) conditions over a wide range of arterial blood oxygen saturation levels as compared to arterial blood CO-Oximetry.

  • For adult patients, 315 data sets (SpO2 VS SaO2) were obtained from 13 healthy adult subjects under nonmotion, aged 22-30, with skin tones of Fitzpatrick 1-6 and 3 of subjects with Fitzpatrick 5. According to the clinical trial results by all subject, gender and skin tones subgroup, within the blood oxygen range of 70-100%, the ARMS of the pulse oximeter is: 1.81% for all subjects, 1.80% for female subjects, and 1.83% for male subjects, 1.78% for light-skinned subjects.1.91% for dark-skinned subjects.

  • For pediatric patients, 243 data sets (SpO2 VS SaO2) were obtained from 10 healthy female subjects, aged 22-30, with skin tones of Fitzpatrick 1-6 and 2 of subjects with Fitzpatrick 5. According to the clinical trial results by 10 female subjects and skin tones subgroup, within the blood oxygen range of 70-100%, the ARMS of the pulse oximeter is: 1.80% for 10 female subjects, 1.81% for light-skinned subjects, 1.74% for dark-skinned subjects.

Conclusion 9.

Based on the non-clinical and clinical data as documented in the device development, the subject devices were found to be as safe, as effective, and perform as well as the predicate device.

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).