K Number
K240776

Validate with FDA (Live)

Date Cleared
2024-09-24

(187 days)

Product Code
Regulation Number
878.4350
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Celsio Flexible Cryocatheter System is intended for cryoadhesion applications during interventional procedures such as the removal of tissue, foreign bodies, mucous plugs, blood clots, necrotic tissue, tumors and tissue biopsies, and for palliative devitalization (destruction) of tissue by the application of extreme cold.

Device Description

The Celsio Flexible Cryocatheter System is a cryosurgical device with a disposable carbon dioxide (CO2) cartridge that is used with commercially available endoscopes with a minimum working channel diameter of 2mm and maximum working length of 1052 mm, in interventional endoscopic procedures. The Celsio Flexible Cryocatheter System is comprised of a Flexible Cryocatheter (sterile, single use), a CO2 Refrigerant Cartridge (non-sterile, single use) and a Refrigerant Dispenser (nonsterile, single use). The flexible cryocatheter is typically delivered using a compatible endoscope to the target location. Once the refrigerant dispenser is activated, pressurized refrigerant fluid is delivered to the catheter tip where it undergoes expansion to its lower temperature (the Joule-Thomson effect), causing surrounding tissue in contact with the catheter tip to freeze and adhere.

AI/ML Overview

The provided text is a 510(k) summary for a medical device (Celsio Flexible Cryocatheter System). It details the device's intended use, comparison to a predicate device, and general performance data, but does not contain specific details about acceptance criteria or a study proving that an AI/Software device meets acceptance criteria.

The document discusses "performance data" and "Design Verification and Validation testing" for the cryocatheter system itself, including thermal characteristics (ice ball dimensions, temperature, cooling/thawing rates) and pre-clinical testing (ex-vivo, in-vivo, post-mortem). However, these are standard tests for a physical medical device and do not involve AI or software performance as would be relevant to the questions posed about AI acceptance criteria, sample sizes for test/training sets, expert ground truth, MRMC studies, or standalone algorithm performance.

Therefore, I cannot provide the requested information from the given text as it pertains to a physical medical device clearance, not an AI/software device clearance. The prompt explicitly asks for details related to a "device" meeting acceptance criteria, and then queries a series of points that are specifically relevant to AI/Software performance studies (e.g., sample sizes for test/training sets, expert ground truth, MRMC studies, standalone performance). Since the provided document is for a physical cryosurgical unit, it does not contain the information necessary to answer those specific questions.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 24, 2024

Endocision Technologies Inc. Anthony Sirgi Director, Regulatory Affairs and Quality 338 Rue Saint-Antoine Est, Suite 407 Montreal, QC H2Y1A3 Canada

Re: K240776

Trade/Device Name: Celsio Flexible Cryocatheter System Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit And Accessories Regulatory Class: Class II Product Code: GEH Dated: March 14, 2024 Received: August 30, 2024

Dear Anthony Sirgi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Digitally signed by Long H. Chen Long H. Chen-S Date: 2024.09.24 11:44:11 -04'00'

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K240776

Device Name

Celsio Flexible Cryocatheter System

Indications for Use (Describe)

The Celsio Flexible Cryocatheter System is intended for cryoadhesion applications during interventional procedures such as the removal of tissue, foreign bodies, mucous plugs, blood clots, necrotic tissue, tumors and tissue biopsies, and for palliative devitalization (destruction) of tissue by the application of extreme cold.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K240776

ഫ് EndoCision 510(k) Summary: Celsio Flexible Cryocatheter System

In accordance with the requirements of 21 CFR 807.92 of the Federal Code of Regulations, the following information is a summary of safety and effectiveness for the Celsio Flexible Cryocatheter System.

SUBMITTERS INFORMATION
Submitted by:Endocision Technologies Inc.338 Rue Saint - Antoine Est, Suite 407Montreal, QuebecH2Y1A3, Canada
Contact Person:Anthony SirgiDirector, Regulatory Affairs and QualityTel: 857-529-5014 ext 202
Date Prepared:September 24th, 2024
DEVICE INFORMATION
Registration Number:K240776
Device Trade Name:Celsio Flexible Cryocatheter System
Regulation Name:Cryosurgical Unit and Accessories
Device Class:Class II
Product Code:GEH
Classification Panel:General & Plastic Surgery

21 CFR 878.4350

PREDICATE DEVICE

Regulation Number:

K190651 - ERBECRYO 2 Cryosurgical Unit and Accessories: ERBECRYO 2 cryosurgical unit; Erbe Flexible Cryoprobe

DEVICE DESCRIPTION

The Celsio Flexible Cryocatheter System is a cryosurgical device with a disposable carbon dioxide (CO2) cartridge that is used with commercially available endoscopes with a minimum working channel diameter of 2mm and maximum working length of 1052 mm, in interventional endoscopic procedures.

{5}------------------------------------------------

് Endocision 510(k) Summary: Celsio Flexible Cryocatheter System

The Celsio Flexible Cryocatheter System is comprised of a Flexible Cryocatheter (sterile, single use), a CO2 Refrigerant Cartridge (non-sterile, single use) and a Refrigerant Dispenser (nonsterile, single use).

The flexible cryocatheter is typically delivered using a compatible endoscope to the target location. Once the refrigerant dispenser is activated, pressurized refrigerant fluid is delivered to the catheter tip where it undergoes expansion to its lower temperature (the Joule-Thomson effect), causing surrounding tissue in contact with the catheter tip to freeze and adhere.

INTENDED USE/INDICATIONS FOR USE

The Celsio Flexible Cryocatheter System is intended for cryoadhesion applications during interventional procedures such as the removal of tissue, foreign bodies, mucous plugs, blood clots, necrotic tissue, tumors and tissue biopsies, and for palliative devitalization (destruction) of tissue by the application of extreme cold.

SUBSTANTIAL EQUIVALENCE COMPARISON

Table 1: Comparison of Celsio Flexible Cryocatheter System (Subject Device) andErbecryo 2 (Predicate Device)
CharacteristicsSubject Device (K240776)Predicate Device (K190651)
Trade NameCelsio Flexible CryocathetersystemERBECRYO 2 CryosurgicalUnit and Accessories:ERBECRYO 2 cryosurgicalunit; Erbe Flexible Cryoprobe
ManufacturerEndocision Technologies Inc.Erbe Elektromedizin GmbH
Product CodeGEHGEH
Regulation21 CFR 878.435021 CFR 878.4350
Device ClassClass IIClass II
Indication for use/IntendedUseThe Celsio FlexibleCryocatheter System isintended for cryoadhesionapplications duringinterventional proceduressuch as the removal of tissue,foreign bodies, mucousplugs, blood clots, necrotictissue, tumors and tissueThe Erbe FlexibleCryoprobes are intended forpalliative devitalization(destruction) of tissue duringinterventional procedures bythe application of extremecold and cryoadhesion forapplications such as theremoval of foreign bodies,
biopsies, and for palliativedevitalization (destruction) oftissue by the application ofextreme cold.mucus plugs, blood clots,necrotic tissue, tissue tumors(palliative recanalization) andtissue biopsies.
Target PopulationPatients which requiredremoval of foreign body,tissue and palliativedestruction of tissue throughendoscopic techniquesPatients which requiredremoval of foreign body,tissue and palliativedestruction of tissue throughendoscopic techniques
Intended Anatomical SiteEndoscopic uses throughnatural orifice (respiratorytree, GI tract, urethral tractetc.)Endoscopic uses throughnatural orifice (respiratorytree, GI tract, urethral tractetc.)
Prescription UseYesYes
Environment of Use/UsageHospital Setting / single-patient use deviceHospital Setting / single-patient use catheter
Cryogenic Gas usedCarbon Dioxide (CO2)Carbon Dioxide (CO2)
Cryogen SourceDisposable CO2 CartridgeRefillable CO2 Cylinder
Provided SterileYes, ethylene oxide (FlexibleCryocatheter and tissueremoval tool)Yes, ethylene oxide (FlexibleCryocatheter and tissueremoval tool)
Application of Cryogenic Gas(Cooling activation method)Handheld RefrigerantDispenser with a leverFoot Pedal andElectromechanical console
DimensionsOuter Diameter: 1.7mm (5 Fr)Working Length: 1200mmOuter Diameter: 1.7mm (5 Fr)Working Length: 1150mmUmbilical Cord: 2.5m
Connection of Catheter toUser InterfaceProprietary Quick-ConnectConnectorProprietary Three-ProngedConnector
Thermodynamic EffectJoule-ThomsonJoule-Thomson
Safety and PerformanceStandardsConformity to currentrecognized StandardsConformity to currentrecognized Standards

{6}------------------------------------------------

EndoCision 510(k) Summary: Celsio Flexible Cryocatheter System

The differences are not significant and do not raise additional questions of safety and effectiveness as can be demonstrated by a review of predicate device characteristics, design and development standards, and non-clinical performance testing.

{7}------------------------------------------------

ڪ Endocision 510(k) Summary: Celsio Flexible Cryocatheter System

PERFORMANCE DATA

To support the substantial equivalence, performance testing of the subject device was conducted in accordance with the applicable standards and guidance documents to ensure that the subject device performs as intended. Design Verification and Validation testing was performed to evaluate the physical, performance and the safety specifications of the device. Device sterility, shelf life, and biological safety of device materials to the point of use were validated according to FDA recognized consensus standards and guidance documents. Thermal characteristics were tested and measured to evaluate substantial equivalence by assessing the ice ball dimensions, temperature, cooling and thawing rates. Pre-clinical testing (ex-vivo, in-vivo and post mortem) was conducted to evaluate the performance and safety of the subject device. The tests demonstrated that the technological characteristics of the subject device do not raise any new questions of safety or effectiveness. The acceptance criteria have been satisfied for all the tests.

CONCLUSION

The subject device (Celsio Flexible Cryocatheter System K240776) and the predicate device (ERBEcryo2 Cryosurgical unit and accessories, K190651) are both cryosurgical devices intended for cryoadhesion and cryodevitalization. There are only minor technological differences between the two devices. However, the similarities in function, purpose and the performance testing for the subject device assure that it is at least as safe and effective as the legally marketed predicate device.

§ 878.4350 Cryosurgical unit and accessories.

(a)
Identification —(1)Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold.(2)
Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold.(3)
Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.(b)
Classification. Class II.