K Number
K240758

Validate with FDA (Live)

Date Cleared
2024-12-11

(266 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Radiofrequency Generator System is used in conjunction with Cura RF Electrode for coagulation of tissue during percutaneous, laparoscopic, and intraoperative surgical procedures.

The Cura RF Electrode is to be used with Radiofrequency Generator System intended for coagulation of tissue during percutaneous, laparoscopic, and intraoperative surgical procedures.

Device Description

Radiofrequency Generator System consists of a genstaltic pump for electrode cooling, a foot switch and a respiratory locator. This Radiofrequency Generator System is used in conjunction with the Cura RF electrode for coagulation of tissue during percutaneous, laparoscopic, and intraoperative surgical procedures.

The Radiofrequency Generator System is capable of delivering up to 250 W of RF power is limited through software control, and the power, impedance, current are monitored by software. The temperature of the electrode's tip is monitored for charring.

The generator generates high frequency alternating current and transferred to the RF electrode and generator is interrupted, the generator will stop the power output and trigger reminder. When the temperature of RF electrode detected by the generator exceeds 99℃, or the temperature of the neutral electrode reaches the setting temperature, the device will automatically out off the power output and give a corresponding reminder. The power output can also be manually stopped.

AI/ML Overview

This document is a 510(k) Premarket Notification from the FDA, a regulatory submission for a medical device. It does not describe an AI/ML-based device, but rather a Radiofrequency Generator System and Cura RF Electrode used for tissue coagulation. Therefore, many of the typical acceptance criteria and study details related to AI/ML performance (e.g., sample size for test set, number of experts, MRMC studies, ground truth establishment for training set, etc.) are not applicable or present in this type of submission.

Specifically, the document focuses on demonstrating substantial equivalence to existing predicate devices based on:

  • Indications for Use: The devices have similar intended uses.
  • Technological Characteristics: Comparison of physical specifications, operating principles, and safety features.
  • Performance Data (Non-Clinical): Testing against recognized standards and ex-vivo tissue tests, rather than clinical efficacy studies on human subjects that would involve extensive human reader studies or large clinical trial data sets.

Therefore, the requested information cannot be fully provided as it pertains to AI/ML device evaluations, which is not the subject of this FDA submission.

However, I can extract the relevant information regarding the non-clinical performance data and the "acceptance criteria" presented in this type of regulatory context, which are primarily compliance with relevant medical device standards.

Here's a breakdown based on the provided document and the non-applicability of certain AI/ML-specific questions:


Acceptance Criteria and Device Performance for a Radiofrequency Generator System

The acceptance criteria for this device are not defined in terms of AI/ML performance metrics (like accuracy, sensitivity, specificity, AUC) but rather through compliance with established medical device safety and performance standards, and demonstration of heating effects in ex-vivo tissue that align with the intended function.

1. A table of acceptance criteria and the reported device performance

The "acceptance criteria" for this specific device (a radiofrequency generator and electrode) are primarily demonstrated through compliance with various national and international standards for electrical safety, electromagnetic compatibility, usability, biological compatibility, and sterilization. The "reported device performance" refers to the successful completion of these tests.

Acceptance Criteria Category (Demonstrated by Compliance)Specific Standards/Tests MentionedReported Device Performance
Electrical SafetyANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] (Medical electrical equipment - Part 1: General requirements for basic safety and essential performance)"The test results demonstrated that the proposed device complies with the following standards:" (followed by a list of standards, implying successful compliance).
IEC 60601-2-2 Edition 6.0 2017-03 (Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories)"The test results demonstrated that the proposed device complies with the following standards:"
IEC 60601-2-18: Edition 3.0 2009-08 (Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment)"The test results demonstrated that the proposed device complies with the following standards:"
Electromagnetic Compatibility (EMC)A IEC60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION (Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests)"The test results demonstrated that the proposed device complies with the following standards:" Also explicitly stated in the comparison table of Radiofrequency Generator System: "Complied with IEC 60601-1-2".
UsabilityIEC 60601-1-6 Edition 3.1 2013-10 (Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability)"The test results demonstrated that the proposed device complies with the following standards:"
Biological EvaluationISO 10993-5 Third edition 2009-06-01 (Tests for in vitro cytotoxicity)"The test results demonstrated that the proposed device complies with the following standards:"
ISO 10993-10 Third Edition 2010-08-01 (Tests for irritation and skin sensitization)"The test results demonstrated that the proposed device complies with the following standards:"
ISO 10993-11 Third edition 2017-09 (Tests for systemic toxicity and Pyrogen test)"The test results demonstrated that the proposed device complies with the following standards:"
Sterilization & ResidualsISO 10993-7:2008 (Ethylene oxide sterilization residuals)"The test results demonstrated that the proposed device complies with the following standards:"
ISO 11135 Second edition 2014-07-15 [Including: Amendment 1 (2018)] (Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices)"The test results demonstrated that the proposed device complies with the following standards:"
ISO 11137-2 Third edition 2013-06 [Including AMD1:2022] (Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose)"The test results demonstrated that the proposed device complies with the following standards:"
Packaging & Shelf-lifeASTM F1980-16 (Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices)"The test results demonstrated that the proposed device complies with the following standards:"
Software PerformanceIEC 62304:2015 and FDA guidance "Content of Premarket Submissions for Device Software Functions" (Software verification and validation testing for embedded software monitoring power, time, impedance, and temperature)."Software verification and validation testing were conducted... to support enhanced level." (Implies successful completion and meeting requirements, as the device received 510(k) clearance).
Thermal Effects (Ex-vivo)FDA Guidance 'Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery' (Test on thermal effects to evaluate width and depth of thermally damaged zone in relation to active part length, gauge, working mode, and power setting on liver, kidney, and muscle tissue)."The test on thermal effects was performed... The test results demonstrated that the proposed device complies with the following standards:" (implies the thermal effects were within expected/acceptable parameters to show equivalence for tissue coagulation, aligning with the FDA guidance).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Test Set Sample Size: For the non-clinical tests, sample sizes are not explicitly stated in the summary, but they would be typical for engineering verification and validation testing of medical devices according to the specified standards (e.g., a certain number of units for electrical tests, specific tissue samples for ex-vivo tests). These are typically small, controlled engineering samples, not large patient datasets.
  • Data Provenance: The tests are likely performed by the manufacturer (Zhejiang CuraWay Medical Technology Co., Ltd. in China) or contract testing organizations on their behalf. The data is prospective in the sense that the tests were specifically conducted to support this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable as this is not an AI/ML diagnostic device requiring expert consensus for "ground truth" on diagnoses. The "ground truth" for this device's performance is objective measurements (e.g., electrical parameters, temperature readings, physical dimensions of coagulation zones) and compliance with specified engineering and medical device standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This question is not applicable as it pertains to expert consensus/adjudication in diagnostic studies, which is not relevant for this type of device submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable. An MRMC study is relevant for AI/ML diagnostic aids that assist human readers. This device is a therapeutic/surgical tool (radiofrequency generator), not a diagnostic AI.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable for the same reasons as above. The device's performance is its direct physical output (e.g., RF energy, coagulation effect), not an algorithm-only output like an AI diagnostic prediction. Its "standalone" performance is assessed through its compliance with technical standards and ex-vivo tests.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's performance is based on:

  • Pre-defined technical specifications and ranges for electrical output, temperature monitoring, and safety features.
  • The ability to achieve intended thermal effects (coagulation zone) in ex-vivo tissue models, compared against expectations from existing predicate devices and literature.
  • Compliance with recognized national and international consensus standards (e.g., IEC, ISO, ASTM).

8. The sample size for the training set

This question is not applicable. This is not an AI/ML device that requires a "training set" in the machine learning sense. The device is a hardware system with embedded software, developed using traditional engineering principles and verified/validated against specifications and standards.

9. How the ground truth for the training set was established

This question is not applicable for the reasons stated above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.

December 11, 2024

Zhejiang CuraWay Medical Technology Co., Ltd. Ruiran Song International Registration Engineer Room 106, Building 1,Room 293, Building 2, No. 600, 21st Avenue, Baiyang Sub-district, Qiantang New District Hangzhou. Zhejiang 310000 China

Re: K240758

Trade/Device Name: Radiofrequency Generator, Cura RF Electrode Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: November 14, 2024 Received: November 14, 2024

Dear Ruiran Song:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Long H. Long H. Chen -Chen -S Date: 2024.12.11 10:39:49 S -05'00'

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K240758

Device Name Radiofrequency Generator System Cura RF Electrode

Indications for Use (Describe)

Radiofrequency Generator System is used in conjunction with Cura RF Electrode for coagulation of tissue during percutaneous, laparoscopic, and intraoperative surgical procedures.

The Cura RF Electrode is to be used with Radiofrequency Generator System intended for coagulation of tissue during percutaneous, laparoscopic, and intraoperative surgical procedures.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

1 SUBMITTER

Company Name: Zhejiang CuraWay Medical Technology Co., Ltd. Contact details Address: Room 106, Building 1, No. 600, 21st Avenue, Baiyang Subdistrict, Qiantang New District, 310018 Hangzhou City, Zhejiang Province, China Phone: +86-057187016876 Contact person: Ruiran Song (Miss) Date prepared: 12/11/2024

2 SUBJECT DEVICE

Trade name: Radiofrequency Generator System

Cura RF Electrode

Common name: Electrosurgical cutting and coagulation device and accessories

Classification name: Electrosurgical cutting & coagulation device & accessories

Review panel: General & Plastic Surgery (SU)

Regulation number: 878.4400

Product code: GEI

3 PREDICATE DEVICES

Primary predicate device: The VIVA combo RF System (K163450)

Secondary predicate device: Star RF Electrode, VIVA RF Electrode (K172012)

{5}------------------------------------------------

4 DESCRIPTION

Radiofrequency Generator System consists of a genstaltic pump for electrode cooling, a foot switch and a respiratory locator. This Radiofrequency Generator System is used in conjunction with the Cura RF electrode for coagulation of tissue during percutaneous, laparoscopic, and intraoperative surgical procedures.

The Radiofrequency Generator System is capable of delivering up to 250 W of RF power is limited through software control, and the power, impedance, current are monitored by software. The temperature of the electrode's tip is monitored for charring.

The generator generates high frequency alternating current and transferred to the RF electrode and generator is interrupted, the generator will stop the power output and trigger reminder. When the temperature of RF electrode detected by the generator exceeds 99℃, or the temperature of the neutral electrode reaches the setting temperature, the device will automatically out off the power output and give a corresponding reminder. The power output can also be manually stopped.

5 INDICATIONS FOR USE

Radiofrequency Generator System is used in conjunction with Cura RF Electrode for cosgulation of tissue during percutaneous, laparoscopic, and intraoperative surgical procedures.

The Cura RF Electrode is to be used with Radiofrequency Generator System intended for coagulation of tissue during percutaneous, laparoscopic, and intraoperative surgical procedures.

{6}------------------------------------------------

6 SUBSTANTIAL EQUIVALENCE COMPARISON WITH PREDICATE DEVICE

Detailed substantial comparison was made between subject device. Please refer to below tables for details.

Comparison ElementsSubject DevicePredicate Device
Classification Regulation878.4400878.4400
Product CodeGEIGEI
510(k) NumberK240758K172012
Indications for useThe Cura RF Electrode is to be used with RadiofrequencyGenerator System intended for coagulation and ablation oftissue during percutaneous, laparoscopic, and intraoperativesurgical procedures.star/VIVA RF Electrode is intended for usein percutaneous and intraoperativecoagulation and ablation of tissue.
Type of usePrescription usePrescription use
Monopolar or bipolarMonopolar and BipolarMonopolar
ModelCura RF Electrode has three models:- Cura RF Electrode Fixed- Cura RF Electrode Adjustable- Cura RF Electrode Doublestar/VIVA RF Electrode has two models:- star RF Electrode_Fixed- VIVA RF Electrode (which is adjustable)
Electrode length (mm)Monopolar: 70, 100, 150, 200, 250Bipolar: 43, 103, 143Monopolar: 70, 150, 200
Active part length (mm)Cura RF Electrode Fixed:5, 7, 10, 15, 20, 25, 30, 40Cura RF Electrode Adjustable: 5 to 40, and adjustable step is continuously.VIVA RF Electrode (Adjustable): 5 to 30, and adjustable step is 5.
Cura RF Electrode Double: 15, 20star RF Electrode Fixed:5, 7, 10, 20, 30
Diameter (Gauge)Monopolar: 15,17,18Bipolar: 15, 16, 17Monopolar: 15, 17, 18
PatientcontactingmaterialsElectrode MaterialsStainless steel SUS 304
Insulation film MaterialsPolyester, Polyimide

Substantial equivalence Comparison table of Cura RF Electrode

{7}------------------------------------------------

{8}------------------------------------------------

Connector type4 pin4 pin
NeutralelectrodeConductive orcapacitiveConductiveConductive
Material(Conductivearea)HydrogelHydrogel
Single useSingle useSingle use
Electrode sterilizationEOEO

{9}------------------------------------------------

Substantial equivalence Comparison table of Radiofrequency Generator System

Comparison ElementsSubject DevicePredicate Device
Classification Regulation878.4400878.4400
Product CodeGEIGEI
510(k) NumberK240758K163450
Indications for useRadiofrequency Generator System is used inconjunction with Cura RF Electrode forcoagulation and ablation of tissue duringpercutaneous, laparoscopic, and intraoperativesurgical procedures.The VIVA combo RF System is intended for use inpercutaneous and intraoperative coagulation andablation of tissue.
PrincipleThe working frequency of the RadiofrequencyGenerator System is 480kHz±10%. When the radiofrequency current flows through the human tissue,the rapid change of the electromagnetic field causesthe positive and negative ions in the cell to moverapidly, so the friction between them, also betweenthem and other molecules and ions causes thelesion to heat up, causing the water inside andoutside the cell to evaporate, dry, shrink and falloff, resulting in aseptic necrosis, thereby achievingthe purpose of treatment.The RF generator works at 480kHz±10%. Thefrequency flows to the electrode's tip and then isapplied to the tissue. Frictional heat occurs andcauses the ions to move from the negative pole tothe positive pole and from the positive pole to thenegative pole forty to fifty thousand times persecond. Tissue necrosis is the principle that occursby using heat generated from the tissue impedance.
Type of usePrescriptionPrescription
Energy UsedRadiofrequencyRadiofrequency
ComponentRadiofrequency Generator System consists of a generator, a peristaltic pump for electrode cooling, a foot switch and a respiratory locator.The VIVA combo RF System consists of RF generator, active electrode, grounding pad and peristaltic pump for electrode cooling.
Bipolar or monopolarBipolar and MonopolarMonopolar
Temperature MonitoringAvailableAvailable
Impedance monitorAvailableAvailable
Voltage Supply100-240VAC 50/60 Hz100-240VAC 50/60 Hz
Continuity monitorYesYes
Output frequency480kHZ± 10 %480kHZ± 10 %
Maximum Power OutputUp to 250 watts at 50 ohmUp to 200 watts at 50 ohm
Drive on timeUp to 15 minutesUp to 30 minutes
Crest factor1.41.4
Physical dimension360(L) x 334(W) x 185(H) mm260mm (W) x 348mm (L) x 115 mm (H)
FootFunctionControl the starting and stopping of the energyControl the starting and stopping of the energy
switchoutput.output.
PerformancespecificationSingle trigger, IPX 8Single trigger, IPX 8
PeristalticpumpPump typePeristaltic pumpPeristaltic pump
Flow rateWith load : 80ml or higher(no load: 120 ml or higher)With load: 90ml or higher(no load: 120 ml or higher)
Electrical testsComplied with IEC 60601-1, IEC 60601-2-2 andIEC60601-2-18Complied with IEC 60601-1 and IEC 60601-2-2
EMCComplied with IEC 60601-1-2Complied with IEC 60601-1-2

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

7 SUMMARY OF NON-CLINICAL DATA

The following performance data were provided in support of the substantial equivalence determination.

Software testing

The software system of Radiofrequency Generator System is embedded software. The power, time, impedance and temperature are monitored by embedded software. Software verification and validation testing were conducted as recommended by IEC 62304:2015 and FDA guidance "Content of Premarket Submissions for Device Software Functions" to support enhanced level.

Ex-vivo tissue testing

The test on thermal effects was performed to evaluate the width and depth of thermally damaged zone in relation to the active part length, gauge, working mode and power setting. The test was performed on liver, kidney, and muscle tissue according to the FDA Guidance 'Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery'.

The test results demonstrated that the proposed device complies with the following standards:

  • ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment -Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]
  • A IEC60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
  • IEC/TR 60601-4-2 Edition 1.0 2016-05 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests

  • A IEC 60601-2-2 Edition 6.0 2017-03 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
  • IEC 60601-2-18: Edition 3.0 2009-08 Medical electrical equipment - Part 2-18:Particular requirements for the basic safety and essential performance of endoscopic equipment

  • IEC 60601-1-6 Edition 3.1 2013-10 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability

{13}------------------------------------------------

  • ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices -A Part 5: Tests for in vitro cytotoxicity
  • ISO 10993-10 Third Edition 2010-08-01 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization

  • ISO 10993-11 Third edition 2017-09 Biological evaluation of medical devices -Part 11: Tests for systemic toxicity

  • ISO 10993-11 Third edition 2017-09 Biological evaluation of medical devices-Part 11: Test for systemic toxicity--Pyrogen test

  • ISO 10993-7:2008 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals [Including: Technical Corrigendum 1 (2009)]- EO residual test

  • ISO 11135 Second edition 2014-07-15 Sterilization of health-care products -Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2018)]

  • ISO 11137-2 Third edition 2013-06 [Including AMD1:2022] Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose [Including Amendment 1 (2022)]

  • A ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices

8 CONCLUSION

The subject device is substantially equivalent to the currently marketed and predicate devices in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness. The standards testing and nonclinical performance testing demonstrates the subject device technological characteristics are equivalent to the predicate device for coagulation and ablation of tissue during percutaneous, laparoscopic, and intraoperative. In conclusion, the subject device is substantially equivalent to the predicate device.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.