K Number
K240712

Validate with FDA (Live)

Device Name
icobrain aria
Manufacturer
Date Cleared
2024-11-07

(237 days)

Product Code
Regulation Number
892.2090
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

icobrain aria is a computer-assisted detection (CADe) and diagnosis (CADx) software device to be used as a concurrent reading aid to help trained radiologists in the detection, assessment and characterization of Amyloid Related Imaging Abnormalities (ARIA) from a set of brain MR images. The software provides information about the presence, location, size, severity and changes of ARIA-E (brain edema or sulcal effusions) and ARIA-H (hemosiderin deposition, including microhemorrhage and superficial siderosis). Patient management decisions should not be made solely on the basis of analysis by icobrain aria.

Device Description

icobrain aria is a software-only device for assisting radiologists with the detection of amyloid-related imaging abnormalities (ARIA) on brain MRI scans of Alzheimer's disease patients under an amyloid beta-directed antibody therapy. The device utilizes 2D fluid-attenuated inversion recovery (FLAR) for the detection of ARIA-E (edema/sulcal effusion) and 2D T2* gradient echo (T2*-GRE) for the detection of ARIA-H (hemosiderin deposition).

icobrain aria automatically processes input brain MRI scans in DICOM format from two time points and generates annotated DICOM images and an electronic report.

AI/ML Overview

Here's a summary of the acceptance criteria and study that proves the device meets them, based on the provided text:

icobrain aria: Acceptance Criteria and Performance Study Summary

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly listed in a single, dedicated table with pass/fail thresholds. Instead, they are implicitly defined by the statistically significant improvements demonstrated in the clinical (MRMC) study, and the "in line with human experts" conclusion from standalone performance. The document focuses on showing the effect size of the improvement rather than pre-defined absolute thresholds for sensitivity, specificity, or AUC for human-AI combined performance. For standalone metrics, it reports specific values and concludes they are "in line with the performance of human experts," suggesting the internal acceptance criteria were met.

Therefore, the table below will summarize the reported performance results from the clinical study, which implicitly met the acceptance criteria by demonstrating significant improvement over unassisted reading.

Performance MetricAcceptance Criteria (Implicit, based on study outcomes)Reported Device Performance (Assisted)Reported Device Performance (Unassisted)Result
ARIA-E Detection (AUC)Significant improvement over unassisted reading0.873 (95% CI [0.835, 0.911])0.822Significant Improvement (+0.051 AUC, p=0.001)
ARIA-E Detection (Sensitivity)Increase over unassisted reading86.5%70.9%Significant Increase
ARIA-E Detection (Specificity)Maintain above 80% with assisted reading83.0%91.7%Maintained above 80% (slight decrease compared to unassisted, but still high)
Pooled ARIA-H Detection (AUC)Significant improvement over unassisted reading0.825 (95% CI [0.781, 0.869])0.781Significant Improvement (+0.044 AUC, p=0.001)
Pooled ARIA-H Detection (Sensitivity)Increase over unassisted reading79.0%68.7%Significant Increase
Pooled ARIA-H Detection (Specificity)Maintain above 80% with assisted reading80.3%82.8%Maintained above 80% (slight decrease compared to unassisted, but still high)
ARIA-H Microhemorrhages Detection (AUC)Significant improvement over unassisted reading0.808 (95% CI [0.760, 0.855])0.779Significant Improvement (+0.029 AUC, p=0.032)
ARIA-H Microhemorrhages Detection (Sensitivity)Increase over unassisted reading79.6%69.3%Significant Increase
ARIA-H Microhemorrhages Detection (Specificity)Maintain above 80% with assisted reading76.7%83.1%Below 80% for this specific subtype
ARIA-H Superficial Siderosis Detection (AUC)Significant improvement over unassisted reading0.784 (95% CI [0.732, 0.836])0.721Significant Improvement (+0.063 AUC, p=0.003)
ARIA-H Superficial Siderosis Detection (Sensitivity)Increase over unassisted reading59.9%49.7%Significant Increase
ARIA-H Superficial Siderosis Detection (Specificity)Maintain above 80% with assisted reading95.6%92.7%Maintained and improved
Localization PerformanceSignificant improvement in accuracy for spatial distributionSignificantly better for assisted readsN/AMet
ARIA Severity Measurement AccuracySignificantly lower absolute differences vs. ground truthSignificantly lower assisted vs. unassistedN/AMet
Inter-reader Variability (Kendall's Coeff. of Concordance)Significantly lower for assisted readsARIA-E: 0.809 (assisted) / 0.720 (unassisted); ARIA-H: 0.799 (assisted) / 0.656 (unassisted)N/ASignificant Reduction
Reading TimeFaster with assisted readingMedian 2:21min (assisted)Median 2:34min (unassisted)Faster

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: 199 cases.
  • Data Provenance: MRI datasets from subjects diagnosed with Alzheimer's disease. To guarantee independence, test data subjects were not included in the training set.
    • Country of Origin: More than 100 sites in 20 countries. Approximately half the data originated from the US and the other half from outside the US.
    • Retrospective/Prospective: The study used retrospective data from clinical trials (aducanumab clinical trials PRIME (NCT02677572), EMERGE (NCT02484547), and ENGAGE (NCT02477800)). This data provenance applies to both training and testing datasets.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: A consensus of 3 experts was used for the clinical (MRMC) study ground truth. For standalone testing, the ground truth was established by unspecified "expert neuroradiologists."
  • Qualifications of Experts:
    • Clinical Study (MRMC): Experts who performed "safety ARIA reading in clinical trials for Aβ-directed antibody therapies in AD."
    • Standalone Testing: "expert neuroradiologists (with experience performing safety ARIA reading in clinical trials for Aβ-directed antibody therapies in AD) manually segmented both ARIA-H findings." This indicates they had prior, relevant experience.

4. Adjudication Method for the Test Set

  • Adjudication Method: "A consensus of 3 experts" was used to establish the ground truth for the clinical (MRMC) study. The specific consensus method (e.g., majority vote, discussion to agreement) is not detailed, but the term "consensus" implies a collective agreement process.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, and Effect Size of Improvement

  • MRMC Study Done: Yes, a fully-crossed MRMC retrospective reader study was conducted.
  • Effect Size (AUC difference, Assisted vs. Unassisted):
    • ARIA-E Detection: +0.051 AUC (95% CI [0.020, 0.083]), p=0.001
    • Pooled ARIA-H Detection: +0.044 AUC (95% CI [0.017, 0.070]), p=0.001
    • ARIA-H Microhemorrhages: +0.029 AUC (95% CI [0.002, 0.055]), p=0.032
    • ARIA-H Superficial Siderosis: +0.063 AUC (95% CI [0.023, 0.102]), p=0.003

Readers also showed significant increases in sensitivity, significant decreases in inter-reader variability, and were on average faster when assisted.

6. If a Standalone (i.e. Algorithm only without human-in-the-loop performance) was Done

  • Standalone Study Done: Yes, "icometrix conducted standalone performance assessments."
    • Standalone Performance Highlights (Main Test Set on 199 cases):
      • ARIA-E Diagnosis: Sensitivity 0.94, Specificity 0.67, AUC 0.84
      • ARIA-H Diagnosis: Sensitivity 0.87, Specificity 0.66, AUC 0.81
      • ARIA-E Finding-level: True Positive Rate 69.1%, False Positive findings per case 0.7
      • ARIA-H New Microhemorrhages Finding-level: True Positive Rate 66.1%, False Positive findings per case 0.9
      • ARIA-H New Superficial Siderosis Finding-level: True Positive Rate 62.5%, False Positive findings per case 0.1
    • The document concludes that standalone performance was "in line with the performance of human experts."

7. The Type of Ground Truth Used

  • Ground Truth Type: Expert consensus for the clinical study (MRMC) and expert manual annotations for the standalone testing.
    • Details: For standalone testing, "expert neuroradiologists ... manually segmented both ARIA-E and ARIA-H findings. Ground truth ARIA measurements were derived from the expert manual annotated masks." For the MRMC study, ground truth was obtained via "a consensus of 3 experts."

8. The Sample Size for the Training Set

  • Training Set Sample Size:
    • FLAIR images (for ARIA-E): 475 image pairs from 172 subjects.
    • T2-GRE images (for ARIA-H):* 326 image pairs from 177 subjects.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth Establishment for Training Set: The data used for developing the algorithms "have been manually annotated by expert neuroradiologists with prior experience of reading ARIA in clinical trials of amyloid beta-directed antibody drugs." This implies manual annotation by experts served as the ground truth for training.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

November 7, 2024

icometrix NV Dirk Smeets CTO Kolonel Begaultlaan 1b/12 Leuven, Vlaams-Brabant 3012 Belgium

Re: K240712

Trade/Device Name: icobrain aria Regulation Number: 21 CFR 892.2090 Regulation Name: Radiological Computer Assisted Detection And Diagnosis Software Regulatory Class: Class II Product Code: QBS Dated: September 30, 2024 Received: September 30, 2024

Dear Dirk Smeets:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Signature

Daniel M. Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K240712

Device Name

icobrain aria

Indications for Use (Describe)

Intended use

icobrain aria is a computer-assisted detection (CADe) and diagnosis (CADx) software device to be used as a concurrent reading aid to help trained radiologists in the detection, assessment and characterization of Amyloid Related Imaging Abnormalities (ARIA) from a set of brain MR images. The software provides information about the presence, location, size, severity and changes of ARIA-E (brain edema or sulcal effusions) and ARIA-H (hemosiderin deposition, including microhemorrhage and superficial siderosis). Patient management decisions should not be made solely on the basis of analysis by icobrain aria.

Intended user population

The intended users of icobrain aria are trained radiologists. When a radiologist assesses the original MR images, both the icobrain aria annotated images and the original, unaltered images are used for concurrent reading.

Intended patient population

icobrain aria is intended to be used in the population of patients with Alzheimer's disease receiving treatment with an amyloid beta-directed antibody using the same ARIA radiographic severity scale as described in Aduhelm®/Legembi® prescription information, and presenting for safety monitoring, including severity assessment and monitoring changes of ARIA.

Warnings and precautions

icobrain aria is an adjunct tool and is not intended to replace a radiologist's review of the images or his or her clinical judgment. icobrain aria is not intended as an independent interpretation of MR images. Decisions should not be made solely based on analysis by icobrain aria.

The device is not intended to be used to segment macrohemorrhages with a diameter of 10 mm or more).

Type of Use (Select one or both, as applicable)

ription Use (Part 21 CFR 801 Subpart D)

e-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for icometrix. The logo consists of a stylized, abstract shape resembling a brain or a swirling pattern in shades of blue, green, and red. To the right of the shape is the word "icometrix" in a clean, sans-serif font, with all letters in lowercase and a bold, black color.

510(k) Summary

K240712

1 Submitter

Nameicometrix NV
Address:Kolonel Begaultlaan 1b / 123012 LeuvenBELGIUM
Contact Person:Dirk Smeets
Telephone number:+32 16 369 000
Fax Number:N.A.
E-mail:dirk.smeets@icometrix.com
Date Prepared:12 Mar 2024

2 Device

Device Trade Name:icobrain aria
Regulation Number21 CFR 892.2090
Regulation NameRadiological Computer Assisted Detection And Diagnosis Software
Regulatory ClassClass II
Product Code:QBS
Classification Panel:Radiology

3 Predicate Device

DeviceOsteoDetect
De Novo NumberDEN180005
ManufacturerImagen Technologies
Regulation Number21 CFR 892.2090
Regulation NameRadiological Computer Assisted Detection And Diagnosis Software
Regulatory ClassClass II
Product Code:QBS

4 Device Description

icobrain aria is a software-only device for assisting radiologists with the detection of amyloid-related imaging abnormalities (ARIA) on brain MRI scans of Alzheimer's disease patients under an amyloid beta-directed antibody therapy. The device utilizes 2D fluid-attenuated inversion recovery (FLAR) for the detection of ARIA-E (edema/sulcal effusion) and 2D T2* gradient echo (T2*-GRE) for the detection of ARIA-H (hemosiderin deposition).

icobrain aria automatically processes input brain MRI scans in DICOM format from two time points and generates annotated DICOM images and an electronic report. The following main measurements are included in the icobrain aria electronic report:

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the icometrix logo. The logo consists of a stylized brain shape on the left and the word "icometrix" on the right. The brain shape is made up of three curved lines in different colors: blue, green, and red. The word "icometrix" is written in a sans-serif font, with the letters in black.

  • ARIA-E: the length of the longest axis computed from the segmented ARIA-E abnormalities, and the number of brain sites affected by ARIA-E;
  • ARIA-H: the count of stable and new T2*-GRE hypointensities indicated as microhemorrhages or superficial siderosis.

From these measurements, ARIA radiographic severities for ARIA-E, ARIA-H microhemorrhages and ARIA-H superficial siderosis are automatically derived and included in the icobrain aria electronic report. The ARA radiographic severity definitions implement the suggested categorization that has been incorporated in U.S. prescribing information for currently FDA-approved amyloid betadirected antibody therapies, such as Aduhelm® and also published more generally (Cogswell et al, 2022; doi: 10.3174/ ainr.A7586).

Table: ARIA Radiographic Severity scale.
TypeARIA Radiographic Severity
MildModerateSevere
ARIA-Eone location with longestaxis < 5 cmone location with longest axis 5 to 10cm, or multiple locations, each withlongest axis < 10 cmat least one location withlongest axis > 10 cm
ARIA-H microhemorrhage1 to 4 new incidentmicrohemorrhages5 to 9 new incidentmicrohemorrhages10 or more new incidentmicrohemorrhages
ARIA-H superficialsiderosis1 new focal area ofsuperficial siderosis2 new focal areas of superficialsiderosis> 2 new focal areas ofsuperficial siderosis

The image processing implemented in icobrain aria for ARIA-E and ARIA-H detection and segmentation is based on deep learning technology. Hyperintensities in FLAIR brain suggestive of ARIA-E, and/or hypointensities in T2 -GRE brain scans, suggestive of ARIA-H, are segmented and the brain is parcellated into 10 regions (left and right frontal, parietal, temporal, occipital lobes, cerebellum and the rest of the brain) in order to provide location-specific ARIA presence. Representative data was used for developing the algorithms and consisted of brain MRI scans of patients from the aducanumab clinical trials PRIME (NCT02677572), EMERGE (NCT02484547) and ENGAGE (NCT02477800), which have been manually annotated by expert neuroradiologists with prior experience of reading ARIA in clinical trials of amyloid beta-directed antibody drugs. In total, 475 FLAR image pairs from 172 subjects aged 51 to 85, and 326 T2* -GRE image pairs from 177 subjects aged 51 to 86, were used for the model development. Among the selected data for training, 81 cases were ARIA-free cases, for which 76.5% had FLAIR hyperintensities that were not labeled as ARIA by the experts.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for icometrix. The logo consists of a stylized, abstract shape resembling a brain or a curved leaf in gradient colors of teal, green, and red. To the right of the shape is the word "icometrix" in a simple, sans-serif font, with the letters in black.

4.1 Outputs produced by the software

The software outputs are annotated DICOM images and an eletronic report summarizing the automated finding. These outputs are reviewed by radiologists concurrently alongside the raw input image review software. Finally, the radiologist summarizes all findings (according to his/her best interpretation after reviewing all images and software outputs) in a structured report, which is then forwarded to the PACS.

Annotated DICOM images

The following color-coded images are generated from the image processing results. These annotated images are made available as an output to the user in DICOM format.

    1. ARIA-E overlay: original FLAIR image from the second time point overlaid with the ARIA-E segmentation in 3D. The stable and increasing regions are marked in different colors (Orange = New or enlarging FLAIR ARIA-E hyperintensities; Green = Stable FLAIR ARIA-E hyperintensities).
    1. ARIA-H overlay: original T2*-GRE image from the second time point overlaid with the ARIA-H (microhemorrhages and superficial siderosis, in different colors) segmentation in 3D. The stable and new ARIA-H regions are marked in different colors (Red = New ARIA-H microhemorrhages; Orange = New ARIA-H superficial siderosis; Green = Stable ARIA-H microhemorrhages; Cyan = Stable ARIA-H superficial siderosis).

FLAIR NEW-ENLARGING ARIA-E

  • Orange = New or enlarging FLAIR hyperintensities suggestive of ARIA-E
  • Green = Stable FLAIR hyperintensities . suggestive of ARIA-E

T2 *- GRE ARIA-H

  • Red = New T2 *-GRE hypointensity suggestive of microhemorrhage .
  • Orange = New T2 *-GRE hypointensity suggestive of superficial siderosis .
  • Green = Stable T2 *-GRE hypointensity suggestive of microhemorrhage .
  • Cyan = Stable T2*-GRE hypointensity suggestive of superficial siderosis

Image /page/6/Picture/15 description: This is a medical scan of a brain. The scan is an axial view, meaning it is a cross-section of the brain. There are two highlighted regions on either side of the brain. The highlighted regions are yellow and green.

Image /page/6/Picture/16 description: The image shows two axial slices of a brain MRI. The left image shows the brain at a lower level, with the temporal lobes and brainstem visible. A small red dot is visible in the left temporal lobe. The right image shows the brain at a higher level, with the frontal and parietal lobes visible. A small yellow area is visible in the left frontal lobe.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for icometrix. The logo consists of a stylized brain shape in shades of blue, green, and red, followed by the word "icometrix" in black, sans-serif font. The brain shape is positioned to the left of the text.

Report

The icobrain aria report has 2 pages that correspond to the two time-point analysis of FLAIR images (for findings suggestive of ARA-E) and T2* GRE images (for findings suggestive of ARIA-H). Patient identification and scan dates are clearly printed in the headers.

The ARIA-E page includes a visual representation of brain slices from the FLAIR images at two time points, with the ARIA-E segmentation overlaid in color on the image of the second time point. If ARIA-E is detected, the slice corresponding the ARIA-E finding with the longest axis is also represented as a line segment on the FLAR image slice corresponding to the second time point.

SAMPLEARIA-ENAMEicobrain ariaDATE OF BIRTH1952-01-01STUDY DATES2018-02-06 - 2018-11-21icobrain ariaby icometrixIDICO-ID
This report is a summary of the automated findings. Please review the original images and annotated images by the software.
2018-02-062018-11-212018-11-21
IMAGING FINDINGS SUGGESTIVE OF ARIA-E
LONGEST AXIS99 mmSITES OF INVOLVEMENT2SUGGESTED SEVERITY*Moderate
Left (ml)Right (ml)
Frontal Lobe00 (+2)
Parietal Lobe0 (+20)0
Occipital Lobe0 (+4)0
Temporal Lobe0 (+28)0
Total (ml)
Cerebellum0
Other0
Whole Brain0 (+54)
*Suggested severity is derived from the automated findings• None: No ARIA-E• Mild: ARIA-E in 1 location axis < 5 cm• Moderate: ARIA-E 5 - 10 cm or more than 1 site of involvement• Severe: ARIA-E > 10 cmVolumes at Time Point 1Volume change+ Increasing: > +9.2 mlDecreasing: < -2.1 mlNote:In benchmark tests, automated measurements for patients with the same suggestedseverity (moderate) had, on average, 0.7 missed, and 1.3 false positive detected sites ofARIA-E. 10% of cases had > 23 mm overestimation of longest ARIA-E axis. 10% of cases had> 27 mm underestimation of longest ARIA-E axis. 10% of cases had > 5 ml overestimation ofvolume. 10% of cases had > 28 ml underestimation of volume.+/- within measurement error: -2.1 - +9.2 ml

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for icometrix. The logo consists of a stylized brain shape on the left, with the word "icometrix" in black font to the right of the brain shape. The brain shape is colored with a gradient, transitioning from blue at the top to green in the middle and red at the bottom.

The ARIA-H page includes a visual representation of brain slices from the T2*-GRE images at two time points, with the ARIA-H segmentation overlaid in colors on the image of the second time point. If ARIA-H is detected, the slice corresponding to the highest proportion of ARIA-H findings (microhemorrhages and/or superficial siderosis).

Image /page/8/Figure/2 description: This image is a medical report from icobrain aria, detailing imaging findings suggestive of ARIA-H, specifically Microhemorrhages and Superficial Siderosis. For Microhemorrhages, the new count is 6, with a suggested severity of Moderate, and for Superficial Siderosis, the new count is 0, with a suggested severity of None. The report also includes a breakdown of counts in different lobes of the brain, such as the Frontal, Parietal, Occipital, and Temporal Lobes, as well as the Cerebellum and other areas.

In case any of the input images is missing or does not satisfy the required acquisition protocol, the report contains a warning stating that a required image is missing, followed by a list of image acquisition requirements for that type of image. The other page, if input images are correct, is still provided.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for icometrix. The logo consists of a stylized, abstract shape on the left and the word "icometrix" on the right. The shape is a curved, semi-circular form with a gradient of colors, transitioning from blue at the top to green in the middle and red at the bottom. The word "icometrix" is written in a clean, sans-serif font, with all letters in lowercase and black.

5 Intended use

Intended use

icobrain aria is a computer-assisted detection (CADe) and diagnosis (CADx) software device to be used as a concurrent reading aid to help trained radiologists in the detection, assessment and characterization of Amyloid Related Imaging Abnormalities (ARMA) from a set of brain MR images. The software provides information about the presence, location, size, severity and changes of ARIA-E (brain edema or sulcal effusions) and ARIA-H (hemosiderin deposition, including microhemorrhage and superficial siderosis). Patient management decisions should not be made solely on the basis of analysis by icobrain aria.

Intended user population

The intended users of icobrain aria are trained radiologist. When a radiologist assesses the original MR images, both the icobrain aria annotated images and the original, unaltered images are used for concurrent reading.

Intended patient population

icobrain aria is intended to be used in the population of patients with Alzheimer's disease receiving treatment with an amyloid betadirected antibody using the same ARIA radiographic severity scale as described in Aduhelm®/Leqembi® prescription information, and presenting for safety monitoring, including severity assessment and monitoring changes of ARIA.

Warnings and precautions

icobrain aria is an adjunct tool and is not intended to replace a radiologist's review of the images or his or her clinical iudgment. icobrain aria is not intended as an independent interpretation of MR images. Decisions should not be made solely based on analysis by icobrain aria.

The device is not intended to be used to segment macrohemorrhages with a diameter of 10 mm or more).

Similar to the predicate device (OsteoDetect, DEN180005), icobrain aria is intended to be used by radiologists interpreting radiological images, to help them with localizing anormalities. The devices are intended to be used concurrently with the reading of images and are not intended as a replacement for the review of a clinicial judgement. The indications for use of icobrain aria and the predicate device differ in the disease-specific findings the devices detect, the type of medical images the devices process, and the intended patient population. However, these differences should not raise new questions regarding safety and effectiveness of the device when used as labeled, under the same general and special controls as the predicate device.

6 Comparison of technological characteristics with the predicate device

At a high level, the subject and predicate devices are based on the following similar design and technological characteristics elements:

  • · software-only;
  • · the input consists of native DICOM images;
  • the end user only interacts with the electronic output of the device (including annotated DICOM images);
  • both devices employ machine learning (more specifically, supervised deep learning) methodology for detecting and diagnosing abnormalities in medical images.

The following technological differences exist between the subject and predicate devices:

  • different type of medical images the devices process;
  • annotated DICOM image output is a bounding box for OsteoDetect and color-coded 3D regions of interest for icobrain aria.

Both devices rely on highly similar forms of standalone performance testing and clinical performance testing to demonstrate effectiveness. Despite the different disease domains in which both devices are applied, icobrain aria was found to enable a comparable improvement in clinical performance to the intended users.

7 Performance data

7.1 Software verification and validation testing

Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions" (issued on June 14, 2023). Basic documentation level was applied to this submission.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Icometrix. The logo consists of a stylized brain-like shape on the left, with the word "icometrix" in black, sans-serif font to the right of the shape. The brain shape is made up of three curved lines, colored in shades of blue, green, and red.

The device's software development, verification, and validation have been carried out following FDA-recognised standards and guidance documents, including:

  • IEC 62304:2006/Amd1:2015 Medical device software Software life cycle processes.
  • ISO 14971:2019 Medical devices Application of risk management to medical devices.
  • IEC 62366-1:2015/Amd1:2020 Medical devices Application of usability engineering to medical devices.
  • Computer-Assisted Detection Devices Applied to Radiology Device Data Premarket Notification [510(k)] Submissions, Document issued on September 28, 2022.
  • · Guidance for Industry and FDA Staff Clinical Performance Assessment: Considerations for Computer-Assisted Detection Devices Applied to Radiology Images and Radiology Device Data - Premarket Approval (PMA) and Premarket Notification [510(k)] Submissions, Document issued on September 28, 2022.
  • Off-The-Shelf Software Use in Medical Devices, Guidance for Industry and Food and Drug Administration Staff, Document issued on September 27, 2019.
  • Applying Human Factors and Usability Engineering to Medical Devices, Guidance for Industry and Food and Drug Administration Staff, Document issued on: February 3, 2016.
  • General Principles of Software Validation; Final Guidance for Industry and FDA Staff, Document issued on: January 11, 2002.
  • · DEN180005 Evaluation of automatic class III designation for OsteoDetect Decision summary with special controls.

The standalone software was tested against software design specific test plans to assure the device meets performance targets, as intended. The device risk management file was completed and risk controls were implemented to mitigate identified hazards. The testing results support that all the software specifications have met the individual modules and at system level. As such, icobrain aria passes all the testing and substantial equivalence with the predicate.

Validation activities also included a usability study (as an add-on questionnaire to the MRMC reader study described in section 7.6 Clinical Testing below), which demonstrated that intended users consider the device usable and useful, and the HCP manual provided with the device as comprehensive.

7.2 Biocompatibility testing

There are no direct or indirect patient components. Therefore, patient contact information is not needed for this device.

7.3 Electrical safety and electromagnetic compatibility (EMC)

icobrain aria is a software-only device, therefore electrical safety and EMC testing are not applicable.

7.4 Bench testing (Standalone Performance Testing)

icometrix conducted standalone performance assessments on cases that are representative for the indication for use in terms of patient demographics, diversity of ARIA findings and MRI scanner variability. Acceptance criteria were established based on statistical analyses of inter-expert variability for ARIA case-level and finding-level detection, which was complemented by a literature study on the radiological reading of ARA, and a review of CADe/CADx devices for other clinical applications and intended patient populations.

Data characteristics: The study data population consists of MRI datasets from subjects diagnosed with Alzheimer's disease. To guarantee independence of test data from training data, MR images used for testing were acquired from subjects that were not included in the training set. MRI scans originated from more than 100 sites in 20 countries, with approximately half the data originating from the US and the other half from outside the US. The protocols used to collect images match the required acquisition parameters of the finished software as intended for clinical use (cfr. HCP manual).

The main test set on which standalone performance results are reported coincides with the testing (see section below, where a direct comparison is made between performance of radiologists assisted by the software, and the standalone software itself). For testing the ARIA-H models, the dataset consisted of 199 pairs of MRI scans from 199 individual subjects. Mean (SD) age was 70.4 (7.2) years, ranging from 52 to 89; 105 (52.8%) were Asian, 1 (0.5%) was Black, 157 (78.9%) were White, and 18 (9.0%) were other or unreported race and ethnicity. The scans were acquired in 33 distinct MRI scanners manufactured by GE (26%), Philips (17%), at 1.5 Tesla (44%) and 3 Tesla (56%).

An extended test set was also available for additional standalone performance testing. This largerabhics characteristics as the main test set: mean (SD) age was 70.6 (7.57), ranging from 52 to 89; 53% female; 75% White, 12% Asian, 1% Black or African-American, and 12% not reported. The scans were acquired in MRI scanners manufactured by GE (24%), Philips (19%) and Siemens (57%), at 1.5 Tesla (42%) and 3 Tesla (58%). For testing the dataset consisted of 1015 pairs of MRI scans from 385 individual subjects. For testing the ARA-H models, the dataset consisted of 449 pairs of MRI scans from 369 individual subjects.

Methods: The standalone testing consisted of comparing software outputs against expert manual annotations. To establish the ground truth, expert neuroradiologists (with experience performing safety ARIA reading in clinical trials for Aβ-directed antibody

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Icometrix. The logo consists of a stylized, curved shape in shades of blue, green, and red on the left, followed by the word "icometrix" in a sans-serif font. The word is written in lowercase letters and is black.

therapies in AD) manually segmented both ARIA-H findings. Ground truth ARIA measurements were derived from the expert manual annotated masks.

The following main testing scenarios were considered:

    1. Diagnostic performance testing for disting (on the case-level) between no ARIA-E and (mild, moderate or severe) ARIA-E, for distinguishing between no ARIA-H and (mild, moderate or severe) ARIA-H, and for distinguishing between severity levels for ARIA-E and ARIA-H, respectively, specificity and area under the receiver operating characteristic curve (AUC) are reported as mean and bootstrapped 95% confidence intervals. ARIA severity discrimination performance is evaluated in terms of weighted agreement to experts.
    1. Detection performance, measured as (finding-level) detection rate, false positive rate and misclassification rate (the latter only for ARIA-H) for individual findings (ARIA-E sites of involvement, ARIA-H new microhemorrhages, ARIA-H new areas of superficial siderosis).

Results:

    1. ARIA-E diagnostic performance in terms of case-level detection of no vs (mild, moderate or severe) ARIA-E was evaluated on a sample of 199 cases with 123 positive cases. The standalone software had a sensitivity of 0.94 (95% Cl [0.90, 0.98]], specificity 0.67 (95% Cl [0.56, 0.77]), and AUC 0.84 (95% Cl [0.77, 0.89]).
      ARIA-H diagnostic performance in terms of case-level detection of no vs (mild, moderate or severe) ARIA-H was evaluated on a sample of of 199 cases with 120 positive for ARIA-H microhemorrhages, 78 positive for ARIA-H superficial siderosis). The standalone software had a sensitivity of 0.87 (95% Cl [0.80, 0.93]], specificity 0.66 (95% Cl [0.55, 0.76]], and AUC 0.81 (95% Cl [0.75, 0.86]). Analogously, for ARA-H microhemorrhages, sensitivity was 0.89 (95% Cl [0.83, 0.92]), specificity was 0.62 (95% Cl [0.54, 0.67]), and AUC was 0.80 (95% Cl [0.74, 0.85]). For ARIA-H superficial siderosis, sensitivity was 0.67 (95% Cl [0.57, 0.76]), specificity was 0.95 (95% Cl [0.91, 0.98]), and AUC was 0.82 (95% Cl [0.76, 0.87]).

ARIA severity discrimination was sufficiently interchangeable with the severity ratings of ARIA experts, taking into account their mutual variability.

    1. ARIA-E finding-level detection was evaluated on a sample of 199 cases with 123 positive cases. The standalone software had a finding-level true positive rate of 69.1% (95% Cl [62.4%, 75.5%]) with 0.7 (95% Cl [0.5, 0.8]) false positive findings per case.
      ARIA-H new microhemorrhages finding-level detection was evaluated on a sample of 199 cases with 82 positive cases. The standalone software had a finding-level true positive rate of 66.1% (95% Cl [57.7%, 74.1%)] with 0.9 (95% Cl [0.7, 1.1]) false positive findings per case.

ARIA-H new superficial siderosis finding-level detection was evaluated on a sample of 199 cases with 78 positive cases. The standalone software had a finding-level true positive rate of 62.5% (95% Cl [51.2%, 73.5%]) with 0.1 (95% Cl [0.1, 0.2]) false positive findings per case.

Misclassification between individual new microhemorrhages and new superficial siderosis findings was of the order of 0.2-0.3 findings per ARIA-H-positive case.

Additional performance testing: The additional performance testing on the extended test set and on both main and extended sets combined showed similar performance results as those evaluated on the main test set only. Generalizability was tested by subgroup analyses with respect to country of origin, expert ground truther, assessed ARA severity, MRI scanner (mis)match, etc.

Conclusions: The standalone performance testing results demonstrated that the software detects ARIA-E and ARIA-H at case level and individual finding level in line with the performance of human experts.

7.5 Animal testing

Not applicable. Animal studies are not necessary to establish the substantial equivalence of this device.

7.6 Clinical testing

icometrix conducted a fully-crossed multiple-case (MRMC) retrospective reader study in line with the special controls of 21 CFR 892.2090, and following recommendations from the FDA guidance document "Clinical Performance Assessment: Considerations for Computer-Assisted Devices Applied to Radiology Images and Radiology Device Data" (issued Sept 2022).

Methods: A multiple-reader multiple-case reader study was conducted to compare the diagnostic performance of radiologists assessing ARIA on MRI scans from patients treated with aducanumab in two situations: unassisted by the icobrain aria software.

The study encompassed 199 cases, and 16 U.S. Board of Radiologists with a range of experience. Radiologists reported the presence, severity and location of ARIA-H (microhemorrhages and superficial siderosis). Cases with ARA were over-represented in the sample for study design purposes. The co-primary study endpoints were the difference between assisted and unassisted detection of ARIA-H (considering in turn either microhemorrhages, superficial siderosis, or both) independently, assessed with the area under the empirical receiver operating characteristic curve (AUC), where the reader assessments for ARIA severity were compared against ground truth obtained via a consensus of 3 experts. Secondary endpoints evaluated the difference in diagnostic performance of detecting mild ARIA in the subgroup of cases with no and mild ARIA, and in discriminating moderate-or-severe ARA versus no-or-mild ARA, for ARIA-H (considering in turn either microhemorrhages,

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for Icometrix. The logo consists of a stylized, abstract shape resembling a brain or a connected network, with a gradient of colors ranging from teal to green to red. To the right of the shape is the word "icometrix" in a clean, sans-serif font, with the "ico" portion in a lighter weight and the "metrix" portion in a bolder weight.

superficial siderosis, or both), independently. Accuracy in ARIA localization performance was also assessed. Finally, sensitivity and specificity for all endpoints, inter-reader variability, and reading time were also computed.

Results: Radiologists assisted by icobrain aria were significantly better in detecting ARA, with an average assisted AUC 0.873 (95% Cl [0.835, 0.911]) for ARIA-E detection, 0.825 (95% C [0.781, 0.869]) for ARIA-H (pooled microhemorrhages and superficial siderosis) detection (ARIA-H microhemorrhages: assisted AUC 0.808 (95% Cl [0.760, 0.855]); ARIA-H superficial siderosis: assisted AUC 0.784 (95% Cl [0.732, 0.836])). As such, icobrain aria-assisted reading performance was superior (vs unassisted) for all ARIA detection coprimary endpoints, with an AUC difference of 0.051 (95% Cl [0.020, 0.083]) for ARIA-E (p=0.001), 0.044 (95% Cl [0.017, 0.070]) for pooled ARIA-H (p=0.001), 0.029 (95% Cl [0.002, 0.055]) for ARIA-H microhemorrhages (p=0.032), and 0.063 (95% Cl [0.023, 0.102]) for ARIA-H superficial siderosis (p=0.003). Sensitivity increased significantly from 70.9% (unassisted) for ARIA-E detection, from 68.7% to 79.0% for pooled ARIA-H detection, from 69.3% to 79.6% for ARIA-H microhemorrhages detection, and from 49.7% to 59.9% for ARIA-H superficial siderosis detection, while specificity remained above 80% for the detection of all ARA types, except for a specificity of 76.7% for the detection of (mild) ARIA-H microhemorrhages.

Table: Primary endpoint results for ARIA detection performance. Results are reported for assisted and standalone software. Effect size is defined as the difference between assisted and unassisted AUC, and the p-value corresponds to the hypothesis test on the AUC difference.

typewithout:withconditionratioeffect size[95% CI]assistedunassistedp-value(MRMC Ustatistictest)softwarestandalone
ARIA detection performance
no ARIA-E vs. (mild, moderate or severe) ARIA-EAUC76:123 cases0.051 [0.020, 0.083]0.8730.8220.0010.838
sensitivity0.8650.7090.943
specificity0.8300.9170.671
no ARIA-H vs. (mild, moderate or severe) ARIA-HAUC79:120 cases0.044 [0.017, 0.070]0.8250.7810.0010.811
sensitivity0.7900.6870.867
specificity0.8030.8280.658
no ARIA-H microhemorrhages vs. (mild, moderate or severe) ARIA-H microhemorrhagesAUC117:82 cases0.029 [0.002, 0.055]0.8080.7790.0320.802
sensitivity0.7960.6930.890
specificity0.7670.8310.624
no ARIA-H superficial siderosis vs. (mild, moderate or severe) ARIA-H superficial siderosisAUC121:78 cases0.063 [0.023, 0.102]0.7840.7210.0030.816
sensitivity0.5990.4970.667
specificity0.9560.9270.950

ARIA detection performance

The highest gain in performance for the assisted reads was an increase from 47.2% to 70.2% in sensitivity for detecting mild ARIA-E (secondary endpoint).

Accuracy in localization performance for spatial distribution of both ARIA-H was also significantly better for the assisted reads compared to unassisted reads. The absolute differences between the individual ARIA severity measurements (i.e. ARIA-E longest axis, count of brain regions affected by ARA-E, count of new microhemorrhages and count of new superficial siderois sites), performed by the readers and the corresponding measurements derived from the expert ground truth masks were all significantly lower assisted versus unassisted.

The inter-reader variability was also significantly lower for assisted reads (Kendall's coefficient of concordance of 0.720 for unassisted and 0.809 for assisted ARIA-E severity assessments, and of 0.656 for unassisted ARIA-H severity assessments), and readers were on average faster (median unassisted reading 2:34min; median assisted reading 2:21min) when performing assisted reading with icobrain aria.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for icometrix. The logo consists of a stylized, curved shape in shades of blue, green, and red, resembling a brain. To the right of the shape is the word "icometrix" in a sans-serif font, with the letters in black.

Conclusions: Radiological reading performance for ARA detection and diagnosis is significantly improved when using icobrain aria. This demonstrates icobrain aria if used in clinical practice may enhance the sensitivity of safety monitoring of patients under amyloid beta-directed antibody therapies.

8 Conclusions

Non-clinical and clinical performance that icobrain aria is substantially equivalent to the chosen predicate. icobrain aria differs from the predicate device in the indications for use, because it detects different disease specific findings in different radiological images. However, both the proposed device (icobrain aria) and the predicate device (OsteoDetect) are computer-assisted detection and diagnostic devices that accept as input medical images in DICOM format and use machine learning techniques to identify and highlight image abnormalities. The overall design of the basic functionality that it provides to the end-user are the same. The differences in technological characteristics do not raise different questions of safety and effectiveness.

The risks associated with use of the device are comparable because they are both intended to be used in a similar same way as an concurrent reading aid for the radiologist, not as a replacement for their clinical judgement. Main risks are related to false positives and false negatives of the devices and these are mitigated by special controls defined in DEN180005. While the risks for both the subject and the predicate device are very similar, the icobrain aria device targets a different clinical condition compared to the predicate, potentially leading to different clinical harms. For this reason, a benefit/risk determination was conducted for icobrain aria:

Summary of thebenefitsThe clinical performance assessment multiple-reader multiple-case (MRMC) study demonstrated astatistically significant improvement in reader performance in detecting ARIA-E and ARIA-H in the intendedpatient population, as measured by the co-primary endpoints of the difference in Area Under the Curve forassisted radiological reading with icobrain aria as compared to the same task without icobrainaria according to clinical standard of care:
ARIA-E detection: AUCassisted- AUCunassisted = 0.873 - 0.822 = 0.051 (two sided 95% confidence interval [0.020,0.083])
ARIA-H detection: AUCassisted- AUCunassisted = 0.825 - 0.781= 0.044 (two sided 95% confidence interval[0.017, 0.070]).
The MRMC study also demonstrated that mean sensitivity increased significantly from 70.9% (unassisted)to 86.5% (assisted) for ARIA-E detection, and from 68.7% to 79.0% for ARIA-H detection, while the device-assisted specificity remained above 80% for the detection of both ARIA types.
Large diagnostic gains included an increase in assisted versus unassisted performance for mild ARIA-E andmild ARIA-H detection (AUCassisted- AUCunassisted = 0.071 for ARIA-E and AUCassisted- AUCunassisted = 0.042 forARIA-H), notably a much higher sensitivity, at no downside in performance for the easier tasks of detectingmoderate-or-severe ARIA-E and ARIA-H. Earlier detection of subtle ARIA findings will allow earlierintervention and mitigate ARIA progression through appropriate decisions on dosing continuation orsuspension.
The inter-reader variability in ARIA severity determination was also significantly lower for assisted versusunassisted reads (Kendall's coefficient of concordance of 0.720 for unassisted and 0.809 for assisted ARIA-Eseverity assessments, and of 0.656 for unassisted and 0.799 for assisted ARIA-H severity assessments), andreaders were on average faster (median unassisted reading 2:34min; median assisted reading 2:21min) whenperforming assisted reading with icobrain aria.
Summary of therisksThere are potential risks associated with use of the device, including:• The device could provide false positive results (i.e., falsely detecting ARIA-E if there is no ARIA-E, falsedetecting ARIA-H if there is no ARIA-H), which could contribute to the radiologist using this informationto make a false positive diagnosis (i.e., diagnosing ARIA-E when there is no ARIA-E, diagnosing ARIA-Hwhen there is no ARIA-H, overestimating mild ARIA-E or mild ARIA-H as moderate-or-severe ARIA-E orARIA-H, respectively). Such a false positive diagnosis can result in unnecessary additional MRI follow-up,unnecessary suspension of amyloid beta-directed antibody treatment, delay in treatment (if falsepositive findings correspond to symptoms caused by some other condition), HCP/patient concern,inconvenience, discomfort from additional MRI, delay to potential benefit due to suspension of theamyloid beta-directed antibody treatment,• The device could provide false negative results (i.e., not detecting actual ARIA-E or ARIA-H findings),which could contribute to the radiologist using this information to make a false negative diagnosis (e.g.,not diagnosing ARIA-E when there is ARIA-E, not diagnosing ARIA-H when there is ARIA-H,underestimating moderate-or-severe ARIA-E or ARIA-H as mild ARIA-E or ARIA-H, respectively). A falsenegative diagnosis could lead to delayed diagnosis, delayed suspension of treatment when necessary,symptoms exacerbating, no follow-up MRI,• The device could be misused to analyze brain images from other patient populations or incompatibleimage acquisition parameters, leading to inappropriate information regarding the presence/location ofimaging abnormalities provided to the end user.• The device could fail and lead to absence of results, delay of results, or incorrect results, which can leadto delayed or inaccurate patient diagnosis.

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for icometrix. The logo consists of a stylized brain-like shape on the left, with the word "icometrix" in black sans-serif font to the right of the shape. The brain shape is colored with a gradient, transitioning from blue at the top to green in the middle and red at the bottom.

Benefit-riskconsiderations(Do the probablebenefits outweighthe probablerisks?)The assistive icobrain aria software aims to support radiologists in detecting ARIA and grading its severityfor reporting to referral clinicians. Such tool contributes to improved patient wellbeing and outcomethrough improved clinical decisions and workflow. icobrain aria is not a replacement of the standard clinicalpractice but rather a supplement to it. In case of clinical decisions based on incorrect data inputs orincorrect data interpretation by the practitioners, there will be no direct significant impact on the patient'shealth, as potential harm can be handled by regular clinical and MRI monitoring.
The benefited population would include Alzheimer's disease patients receiving an amyloid beta-directedantibody treatment and monitored for symptomatic or asymptomatic ARIA. Based on the performance dataand the application of general controls and special controls in accordance to the device classification and inalignment with the predicate device, it is unlikely that the aforementioned risks would result in worse ARIAmonitoring compared with the current clinical standard of practice. Clinical performance assessmentreveals that there is little to no risk of worse health outcome due to automation bias (i.e., users accepting thesuggested findings by the software without critical review). This software device potentially brings generalradiologists, who have little to no experience with ARIA monitoring, on par with experts that participated inARIA safety monitoring in clinical trials of amyloid beta-directed antibody therapies. Possible misuse of thedevice does not present additional risks compared with the misuse of other types of radiological imageprocessing devices. Use of incompatible image acquisition parameters is mitigated by not providing theicobrain aria outputs to the end user if non-compliant protocols are used.
The device has been developed and tested on brain scans acquired from Alzheimer's disease patients whoparticipated in the clinical trials of aducanumab, and has not yet been extensively tested on other patientpopulations. Nevertheless, ARIA is a radiographic finding that, when it occurs, has identical appearance forother amyloid beta-directed antibody treatments, which should not limit applicability of the device to onlyaducanumab-treated patients. However, the sample population on which the device was tested includedonly patients who qualified to the inclusion criteria of clinical trials and might not represent all possibleconcomitant pathologies that may occur in parallel to Alzheimer's disease. Since more data is currentlyunavailable for product development outside clinical trials, post-market vigilance criteria will be applied tomonitor potentially higher rates of false positives or false negatives in real world intended population.
Conclusionicometrix has applied a risk management process in accordance with FDA recognized standards to identify,evaluate, and mitigate all known hazards related to icobrain aria. These hazards may occur when accuracyof diagnosis is potentially affected, causing either false positives or false negatives. All identified risks areeffectively mitigated and it can be concluded that the residual risk is outweighed by the benefits.

Because of similar intended use and technological characteristics, icobrain aria is substantially equivalent to a device (OsteoDetect, DEN180005) that has been granted its De Novo request in the United and clinical and clinical performance tests demonstrate that icobrain aria is as safe and effective as the predicate device.

N/A