AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Ultrasonic Scaler is used to remove dental plaque from teeth and remove stains on teeth during dental cleaning.

Device Description

A device intended for use during dental cleaning and periodontal (gum) therapy to remove adherent plaque and food debris from the teeth to reduce tooth decay, potential stains in the gingival groove, the back of the tooth and the tooth and associated gingival line. The electronic oscillation circuit generates ultrasonic frequency electric pulse wave, which is amplified to a certain intensity and conveyed to the transducer. The electrical energy is converted into mechanical energy, and the working head tool is stimulated to produce vibration of the same frequency, so as to achieve the purpose of crushing calculus and loosening tartar.

AI/ML Overview

The provided text is a 510(k) summary for an Ultrasonic Scaler (K240707). It describes the device's substantial equivalence to a predicate device and outlines performance data. However, it does not contain the acceptance criteria or a study that proves the device meets those criteria in the context of AI/ML performance.

Instead, the document focuses on:

  • Comparison of technological characteristics between the proposed device and a predicate device (K163414). This comparison highlights similarities and differences in design, operation controls, power supply, operating conditions, storage conditions, and electrical/EMC/biocompatibility standards.
  • Performance data limited to biocompatibility testing and non-clinical testing of electrical safety and electromagnetic compatibility (EMC) according to various IEC standards.

Therefore, I cannot provide the requested information regarding acceptance criteria for an AI/ML device, its reported performance, sample sizes, ground truth establishment, or specific AI/ML study details, because this information is not present in the provided document.

The document discusses acceptance criteria and performance data for a physical medical device (ultrasonic scaler), not an AI/ML component.

If this was a misunderstanding and you were expecting information related to a traditional medical device's performance, please clarify. Otherwise, the requested AI/ML specific information is absent from this document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right of the symbol, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" above the word "ADMINISTRATION". The logo is simple and professional, reflecting the FDA's role in regulating food and drugs.

July 2, 2024

Shenzhen Micro Electric Intelligence Co., Ltd. Fengmeng Han General Manager 301, Building A5, No. 13 Baonan Road, Longgang Community Longgang District Shenzhen, Guangdong 518100 CHINA

Re: K240707

Trade/Device Name: Ultrasonic Scaler (WD-JY-B002,WD-JY-W001,WD-JY-R003,WD-JY-G004.WD-JY-B003.WD-JY-C006.WD-JY-M0014. WD-JY-N0015.WD-JY-N0016.WD-JY-N0017.WD-JY-E007. WD-JY-F008.WD-JY-A005.WD-JY-H009,WD-JY-I0011, WD-JY-K0012,WD-JY-K0013,WD-JY-K0014, WD-JY-L0013,WD-JY-L0014,WD-JY-L0015, WD-JY-L0016,WD-JY-L0017) Regulation Number: 21 CFR 872.4850 Regulation Name: Ultrasonic Scaler Regulatory Class: Class II Product Code: ELC Dated: June 3, 2024 Received: June 3, 2024

Dear Fengmeng Han:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Michael E. Adjodha -S

Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K240707

Device Name

Ultrasonic Scaler (WD-JY-B002,WD-JY-R003,WD-JY-R003,WD-JY-G003,WD-JY-C006,WD-JY-M0014, WD-JY-N0015,WD-JY-N0016,WD-JY-N0017,WD-JY-E007, WD-JY-F008,WD-JY-H009,WD-JY-I0011, WD-JY-K0012,WD-JY-K0013,WD-JY-K0014, WD-JY-L0013,WD-JY-L0015, WD-JY-L0015, WD-JY-L0016,WD-JY-L0017)

Indications for Use (Describe)

Ultrasonic Scaler is used to remove dental plaque from teeth and remove stains on teeth during dental cleaning.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K240707_510(k) summary

Submitter l

Device Submitter

Shenzhen Micro Electric Intelligence Co., Ltd.

301, Building A5,No. 13 Baonan Road, Longgang Community, Longgang District, Shenzhen

Contact Person

FANGMENG HAN General Manager Phone: 86-755-23310679 E-mail: 360721873@qq.com Prepare Date: July 1, 2024

Device =

Trade Name of Device:Ultrasonic Scaler
Common NameUltrasonic Scaler
Regulation Number:21 CFR 872.4850
Regulation Name:Scaler, Ultrasonic
Regulatory Class:II
Product code:ELC
Review Panel:Dental

III Predicate Devices

510(k) NumberK163414
Device NameUltrasonic scaler
ManufacturerNANNING VV DENTAL CO., LTD
Regulation Number:21 CFR 872.4850
Regulatory Class:II
Product code:ELC
Review Panel:Dental

IV Device description

A device intended for use during dental cleaning and periodontal (gum) therapy to remove adherent plaque and food debris from the teeth to reduce tooth decay, potential stains in the gingival groove, the back of the tooth and the tooth and associated gingival line.

The electronic oscillation circuit generates ultrasonic frequency electric pulse wave, which is amplified to a certain intensity and conveyed to the transducer. The electrical energy is

{5}------------------------------------------------

converted into mechanical energy, and the working head tool is stimulated to produce vibration of the same frequency, so as to achieve the purpose of crushing calculus and loosening tartar.

> Indications for use

Ultrasonic Scaler is used to remove dental plaque from teeth and remove stains on teeth during dental cleaning.

VI Comparison of technological characteristics with the predicate devices

The Ultrasonic Scaler has the same intended use, technology, design and performance specifications are either identical or substantially equivalent to existing legally marketed predicate devices. The differences between the Ultrasonic Scaler and predicate devices do not alter suitability of the proposed device for its intended use.

ItemProposed DevicePredicate Device(K163414)Comparison
ManufacturerShenzhen Micro ElectricIntelligence Co., Ltd.NANNING VV DENTALCO., LTD./
Regulation No.21 CFR 872.485021 CFR 872.4850Identical
ClassIIIIIdentical
Product CodeELCELCIdentical
Indication foruseUltrasonic Scaler is usedto remove dental plaquefrom teeth and removestains on teeth duringdental cleaning.The ultrasonic scaler isintended to:1. Remove supra and subgingival calculusdeposits and stain fromthe teeth;2. Carry out periodontalpocket lavage withsimultaneous ultrasonic tipmovement;3. Clean and irrigate rootcanalsIdentical
OperationControlsCleaning and disinfectionmethod of tooth cleaningtip: after each use, scaldScaler tip: autoclavedunder high temperature andhigh pressureDifferent
ItemProposed DevicePredicate Device(K163414)Comparison
the tooth cleaning tip withboiling water for fiveminutes, then wipe it dryand store it for next use
Display ModesCorresponding signal ofexternal display drivercircuit lights upcorresponding LED.Corresponding signal ofexternal display drivercircuit lights upcorresponding LED.Identical
MeasurementItemsremove adherent plaquefrom the teeth andpotential stains in thegingival groove, the backof the tooth and the toothand associatedgingival line.Remove calculus or plaqueon surface of teeth in oralcavity, cleaning andirrigation of root canals.Similar
Power SupplyInput Voltag: 3.7VDC 50HZOutput Power: 2.8WInput Voltage: 30VDC 50HZOutput Power: 3-20WDifferent
OperatingconditionsTemperature: 0-35°CRelative Humidity:<80%R.HPressure: 70-106kpaTemperature: 5°C-40°CRelative Humidity: 30%--80%Pressure: 50kpa-106kpaDifferent
StorageconditionsTemperature: 0°C - +40°CRelative Humidity:<80%R.H.Pressure: 70-106kpaTemperature: -10°C -+50°CRelativehumidity: ≤ 80%Pressure: 50kpa-106kpaDifferent
Tips ofultrasonicscalerMechanical vibrationtransmitted from thehandpiece is converted tothe vibration of the scalertip, then removes calculusor plaque from the teeth.Mechanical vibrationtransmitted from thehandpiece is converted tothe vibration of the scalertip, thenremoves calculus or plaquefrom theteethDifferent
ItemProposed DevicePredicate Device(K163414)Comparison
ElectricalSafetyIEC 60601-1IEC 60601-1Identical
EMCIEC 60601-1-2IEC 60601-1-2Identical
PerformnceIEC 60601-1-11IEC 61205Different
BiocompatibilityISO 10993-5, ISO 10993-10ISO 10993-5, ISO 10993-10Identical
Software Levelof ConcernModerateModerateIdentical

{6}------------------------------------------------

510(k) Summary

{7}------------------------------------------------

Discussion

Similarities:

Measurement Items is similar.

Proposed Device: remove adherent plaque and food debris from the teeth to reduce tooth decay, potential stains in the gingival groove, the back of the tooth and the tooth and associated gingival line.

Predicate Device: Remove calculus or plaque on surface of teeth in oral cavity, cleaning and irrigation of root canals.

Difference:

Design:Operation Controls, Power Supply, Operating conditions and Storage conditions is slight different. However, Proposed Device has passed the Electrical Safety and EMC tests. Performance: the requirement of performance standards is slight different. However, Proposed Device has passed the performance test.

  • The difference does not raise additional questions for safety and effectiveness. . Performance testing including biocompatibility evaluation has been performed on the final finished device which includes all construction materials and color additives.
    In conclusion, there is no substantial difference between our products and similar products currently marketed in the United States.

VII Performance data

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

Biocompatibility of the Ultrasonic Scaler was evaluated in accordance with ISO 10993-1:2018 for the body contact category of "Surface Device - Mucosal Membrane" with a contact duration of "Limited (< 24 hours)". The following tests were performed, as recommended: Cytotoxicity ISO 10993-5: 2009

{8}------------------------------------------------

Skin sensitization Intracutaneous reactivity

ISO 10993-10: 2021 ISO 10993-23: 2021

Non-Clinical Testing:

A battery of tests was performed to verify that the proposed device met all design specification. The test result demonstrated that the proposed device complies with the following standards: Electrical safety and electromagnetic compatibility

IEC 60601-1:2005 + AMD1:2012 + AMD2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

IEC 60601-1-2:2015 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance-Collateral Standard: Electromagnetic disturbances -Requirements and tests

IEC 60601-1-11:2015+AMD1:2020 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

IEC 62471:2006 Photobiological safety of lamps and lamp systems

IEC 80601-2-60:2019 Medical electrical equipment - Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment

VIII Conclusion

Conclusions drawn from non-clinical trials indicate that the subject device is as safe and effective as the legally marketed predicted K163414 ultrasonic scale and performs as well or better than the K163414 ultrasonic scaler.

§ 872.4850 Ultrasonic scaler.

(a)
Identification. An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.(b)
Classification. Class II.