K Number
K240670

Validate with FDA (Live)

Date Cleared
2024-10-30

(233 days)

Product Code
Regulation Number
862.3100
Age Range
2 - 12
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DRI™ Ecstasy Plus Assay is a homogeneous enzyme immunoassay intended for the qualitative or semiquantitative determination of ecstasy drugs in human urine. The assay provides a simple and rapid analytical screening procedure for detecting ecstasy drugs at a cutoff level of 500 ng/mL. The product is intended for use in clinical laboratories only.

This assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/GC/MS) or Liquid chromatography with tandem mass spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Device Description

DRI Ecstasy Plus Assay is a homogeneous enzyme immunoassay intended for the qualitative or semiquantitative determination of ecstasy drugs in human urine. The assay provides a simple and rapid analytical screening procedure for detecting ecstasy drugs at a cutoff level of 500 ng/mL.

DRI Ecstasy Plus Assay is a class assay. Ecstasy drugs represent a group of ring substituted methylenedioxy analogues of amphetamine including 3, 4 methylenedioxyamphetamine (MDA), 3, 4 - methylenedioxymethamphetamine (MDMA) and 3, 4 - methylenedioxyethylamphetamine (MDEA). They are central nervous system (CNS) stimulants popularly abused for their psychotropic effects and are listed by the U.S. Drug Enforcement Administration as Schedule | (no accepted medical application with great abuse potential).

AI/ML Overview

The provided document is a 510(k) Pre-market Notification from the FDA regarding the DRI™ Ecstasy Plus Assay. This is a medical device for in-vitro diagnostics, specifically a homogeneous enzyme immunoassay for detecting ecstasy drugs in human urine. The document mainly focuses on demonstrating substantial equivalence to a predicate device.

Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The document describes the performance characteristics and states the acceptance criteria (stated as "met the acceptance criteria" or specific percentage/range values).

Performance CharacteristicAcceptance Criteria (Stated or Implied)Reported Device Performance
Precision (Qualitative)Assay within run ≤ 2% CV; Total run precision ≤ 5.0 % CVAssay within run is ≤ 2% CV and total run precision is ≤ 5.0 % CV.
Precision (Semi-Qualitative)Assay within run ≤ 5% CV; Total run precision ≤ 7.5 % CVAssay within run is ≤ 5% CV and total run precision is ≤ 7.5 % CV.
Method Comparison (Negative Sample Agreement)100% agreement100% in both Qualitative and Semi-Quantitative modes.
Method Comparison (Positive Sample Agreement)100% agreement100% in both Qualitative and Semi-Quantitative modes.
Method Comparison (Overall Sample Agreement)100% agreement100% for DRI Ecstasy Plus assay in both Qualitative and Semi-Quantitative mode.
Specificity (Structurally Related Compounds - Positive Result)Produced a positive result above the cut-off calibrator at lowest concentrations.Reported at lowest concentrations.
Specificity (Structurally Related Compounds - Negative Result)Produced a negative result below the cut-off calibrator at highest concentrations.Reported at the highest concentrations.
Specificity (Structurally Unrelated Compounds)Shall produce a negative result below the 500 ng/mL MDMA sample in both Qualitative and Semi-quantitative modes at the concentrations tested.Produced a negative result below the 500 ng/mL MDMA cutoff when structurally unrelated compounds were spiked into negative pool urine at the tested concentrations.
Interference (Qualitative - Low Control)Rates of low control samples spiked with interfering substances and pH adjusted were negative.Were negative.
Interference (Qualitative - High Control)Rates of high control samples spiked with interfering substances and pH adjusted were positive.Were positive.
Interference (Semi-quantitative - Low Control Recovery)Recovery of low control level sample spiked with interfering substances and pH adjusted within 80 - 120% of nominal value of 375 ng/mL.Within 80 - 120% of the nominal value of 375 ng/mL.
Interference (Semi-quantitative - High Control Recovery)Recovery of high control level sample spiked with interfering substances and pH adjusted within 80 - 120% of nominal value of 625 ng/mL.Within 80 - 120% of the nominal value of 625 ng/mL.
Stability (Open-Vial / In-Use)Met acceptance criteria.Demonstrated that three lots met the acceptance criteria.
Stability (Unopened Shelf Life)Stable for 36 months when stored unopened at 2-8°C.Supports the claim that the kits are stable for 36 months when stored unopened at 2-8°C.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size used for the test set. It mentions "Negative sample agreement" and "Positive sample agreement" in the method comparison and "structurally unrelated compounds were spiked into negative pool urine at the tested concentrations" for specificity, but no specific numbers of samples are provided.

The data provenance is also not stated. It does not specify the country of origin of the data or whether the studies were retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

For this type of in-vitro diagnostic device (immunoassay for drug detection), the ground truth is typically established by alternative chemical methods (e.g., GC/MS or LC-MS/MS), not by expert human readers. Therefore, the concept of "experts used to establish the ground truth" in the traditional sense of clinical image review is not applicable here. The document states: "A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) or Liquid chromatography with tandem mass spectrometry (LC-MS/MS) is the preferred confirmatory method."

4. Adjudication Method for the Test Set

Since the ground truth is established by alternative chemical methods like GC/MS or LC-MS/MS, an adjudication method for reconciling expert opinions is not applicable. The result from the confirmatory chemical method serves as the definitive ground truth.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. This device is an in-vitro diagnostic immunoassay, not an AI-powered diagnostic tool requiring human interpretation or a multi-reader multi-case study. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not relevant to this submission.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, in a sense. The "performance testing" described (Precision, Method Comparison, Specificity, Interference, Stability) represents the standalone performance of the assay itself. The immunoassay provides a direct analytical result (qualitative or semi-quantitative) without human interpretation affecting the primary outcome of detection. The device's output is an objective measurement (e.g., optical density change leading to a positive/negative determination), rather than an image requiring human interpretation.

7. The Type of Ground Truth Used

The type of ground truth used is confirmatory analytical methods, specifically Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography with Tandem Mass Spectrometry (LC-MS/MS). This is explicitly stated as the "preferred confirmatory method" to obtain a "confirmed analytical result."

8. The Sample Size for the Training Set

The document does not mention a training set or its sample size. Immunoassays are typically developed and validated using a series of laboratory experiments with known samples (spiked urine, controls, clinical samples confirmed by reference methods) rather than being "trained" on a dataset in the way an AI algorithm is. The development process would involve optimizing reagents and assay conditions but doesn't usually involve a "training set" in the machine learning context.

9. How the Ground Truth for the Training Set was Established

As there is no mention of a "training set" in the context of an immunoassay, the establishment of its ground truth is not applicable as outlined for AI/ML devices. The "ground truth" for developing and validating such an assay would be established by preparing samples with known concentrations of the target analytes and confirming those concentrations using gold-standard analytical techniques like GC/MS or LC-MS/MS.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "FDA U.S. FOOD & DRUG ADMINISTRATION".

Microgenics Corporation Amrit Takhar Regulatory Affairs Specialist III 46500 Kato Road Fremont, California 94538

Re: K240670

Trade/Device Name: DRI™ Ecstasy Plus Assay Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine test system Regulatory Class: Class II Product Code: DKZ Dated: October 4, 2024 Received: October 4, 2024

Dear Amrit Takhar:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Joseph A. Joseph A. Kotarek =S Kotarek -S Date: 2024.10.30 09:41:12 -04'00" Joseph Kotarek, Ph.D. Branch Chief for Toxicology Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K240670

Device Name DRI™ Ecstasy Plus Assay

Indications for Use (Describe)

The DRI™ Ecstasy Plus Assay is a homogeneous enzyme immunoassay intended for the qualitative or semiquantitative determination of ecstasy drugs in human urine. The assay provides a simple and rapid analytical screening procedure for detecting ecstasy drugs at a cutoff level of 500 ng/mL. The product is intended for use in clinical laboratories only.

This assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/GC/MS) or Liquid chromatography with tandem mass spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510K Summary K240670

l. Device Information

Contact Details
Sponsor:Microgenics Corporation Thermo FisherScientific 46500 Kato RoadFremont, CA 94538Phone: 510-979-5000FAX: 510-979-5002
Correspondent Contact Information:Amrit TakharRegulatory Affairs Specialist III,Regulatory AffairsEmail: Amrit.Takhar@thermofisher.comPhone: 510- 771-5935FAX: 510-979-5002
Device Name
Device Trade Name:DRI TM Ecstasy Plus Assay
Common Name:Amphetamine Test System
Classification Name:Enzyme Immunoassay, Amphetamine
Regulation Number:862.3100
Product Code:DKZ
Legally Marketed Predicate Device
Predicate Premarket Notification Number:K012110
Predicate Trade Name:DRI Ecstasy Enzyme Immunoassay
Predicate Common Name:Amphetamine Test System
Predicate Classification Name:Enzyme Immunoassay, Amphetamine
Predicate Regulation Number:862.3100
Predicate Product Code:DKZ

II. Date Summary Prepared

March 8, 2024

lll. Description of Device

DRI Ecstasy Plus Assay is a homogeneous enzyme immunoassay intended for the qualitative or semiquantitative determination of ecstasy drugs in human urine. The assay provides a simple and rapid analytical screening procedure for detecting ecstasy drugs at a cutoff level of 500 ng/mL.

DRI Ecstasy Plus Assay is a class assay. Ecstasy drugs represent a group of ring substituted methylenedioxy analogues of amphetamine including 3, 4 methylenedioxyamphetamine (MDA), 3, 4 - methylenedioxymethamphetamine (MDMA) and 3, 4 - methylenedioxyethylamphetamine (MDEA). They are central nervous system

{4}------------------------------------------------

(CNS) stimulants popularly abused for their psychotropic effects and are listed by the U.S. Drug Enforcement Administration as Schedule | (no accepted medical application with great abuse potential). At low dose, both MDMA and MDA produce euphoria, increased self-awareness, and an increased sense of trust. At higher doses, they are thought to be hallucinoqenic. Toxic effects are similar to those of other CNS stimulants and include anxiety, depression, tachycardia, elevated blood pressure, cardiac arrhythmias, pupil dilation, and sleep disorders.

The length of time following drug use for which a positive result may occur is dependent upon several factors including the frequency and amount of drug, metabolic rate, excretion rate, drug half-life and the drug user's age, weight, activity and diet. Within the body, MDMA is known to metabolize to MDA by N-demethylation. Urinary excretion accounts for 65% of the dose as parent drug and 7% as MDA within 3 days. Urinary MDMA concentrations following a 1.5 mg/kg oral dose may exceed 17 mg/L. Other urinary metabolites include mono- and dihydroxy-derivatives of MDMA and MDA, resulting from fission of the methylene bridge, which are eliminated as conjugates. The human metabolism of MDA has not been studied. Urine concentrations in fatal cases of up to 160 mg/L have been recorded and are indicative of excretion of substantial portions of unchanged drugs.

Therefore, DRI Ecstasy Plus Assay is a first-line device, which may be used by medical personnel, along with clinical observations, as an aid to diagnosis of ecstasy abuse.

IV. Intended Use

A. Indications for Use:

See indications for use below.

B. Intended Use:

The DRI™ Ecstasy Plus Assay is a homogeneous enzyme immunoassay intended for the qualitative or semiquantitative determination of ecstasy drugs in human urine. The assay provides a simple and rapid analytical screening procedure for detecting ecstasy drugs at a cutoff level of 500 ng/mL. The product is intended for use in clinical laboratories only.

This assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) or Liquid chromatography with tandem mass spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

V. Comparison to Predicate Device

{5}------------------------------------------------

CharacteristicsPredicate DeviceDRI Ecstasy EnzymeImmunoassay (K012110)Candidate DeviceDRI Ecstasy Plus AssayComparison
Indications forUseSee intended use below forindications useSee intended use below forindications useIdentical
Intended UseThe DRI Ecstasy EnzymeImmunoassay is a homogeneousenzyme immunoassay intendedfor the qualitative orsemiquantitative determination ofecstasy drugs in human urine.The assay provides a simple andrapid analytical screeningprocedure for detecting ecstasydrugs at a Cutoff level of500ng/mL. The product isintended for use in clinicallaboratories only.This assay provides only apreliminary analytical test result.A more specific alternatechemical method must be usedin order to obtain a confirmedanalytical result. Gaschromatography/ massspectrometry (GC/MS) is thepreferred confirmatory method.Clinical consideration andprofessional judgment should beapplied to any drug of abuse testresult, particularly whenpreliminary positive results areused.The DRITM Ecstasy Plus Assayis a homogeneous enzymeimmunoassay intended for thequalitative or semiquantitativedetermination of ecstasy drugs inhuman urine. The assay providesa simple and rapid analyticalscreening procedure fordetecting ecstasy drugs at acutoff level of 500 ng/mL. Theproduct is intended for use inclinical laboratories only.This assay provides only apreliminary analytical test result.A more specific alternativechemical method must be usedin order to obtain a confirmedanalytical result. Gaschromatography/massspectrometry (GC/MS) or Liquidchromatography with tandemmass spectrometry (LC-MS/MS)is the preferred confirmatorymethod. Clinical considerationand professional judgmentshould be applied to any drug ofabuse test result, particularlywhen preliminary positive resultsare used.Identicalwithexpectationto additionalconfirmatorymethodoptionadded. Thisdoes notimpact theintendeduse ofdevice.
FDA ProductCodeDKZDKZIdentical
DeviceClassificationand Name21 CFR 862.3100Amphetamine Test System21 CFR 862.3100Amphetamine Test SystemIdentical
OperatingPrinciple(Technology)DRIDRIIdentical
AnalyteEcstasyEcstasyIdentical
MeasuredAnalyte3, 4 -methylenedioxyamphetamine(MDA)3, 4 -methylenedioxymethamphetamine (MDMA)3, 4 -methylenedioxyamphetamine(MDA)3, 4 -methylenedioxymethamphetamine (MDMA)Identical
Test MatrixHuman UrineHuman UrineIdentical
CharacteristicsPredicate DeviceDRI Ecstasy EnzymeImmunoassay (K012110)Candidate DeviceDRI Ecstasy Plus AssayComparison
Cut-off Levels500 ng/mL500 ng/mLIdentical
MethodologyHomogeneous EnzymeImmunoassayHomogeneous EnzymeImmunoassayIdentical
MaterialsAntibody/Substrate Reagent(R1) contains mouse monoclonalanti-MDMA antibody, glucose-6-phosphate (G6P), andnicotinamide adeninedinucleotide (NAD) in Tris bufferwith sodium azide as apreservative.Enzyme Conjugate Reagent(R2) contains MDMA labeledwith glucose-6-phosphatedehydrogenase (G6PDH) in Trisbuffer with sodium azide as apreservative.Antibody/Substrate Reagent(R1) contains mousemonoclonal anti-MDMAantibody, glucose-6-phosphate(G6P), and nicotinamideadenine dinucleotide (NAD) inTris buffer with sodium azide asa preservative.Enzyme Conjugate Reagent(R2) contains MDMA labeledwith glucose-6-phosphatedehydrogenase (G6PDH) inTris buffer with sodium azide asa preservative.Identical
Reagent FormReagents are sold in two sizes,100 mL and 500 mL kits.Reagents are sold in one size,500 mL kit.Identical
AntibodyMouse Monoclonal AntibodiesMouse Monoclonal AntibodiesIdentical
Storage2-8°C until expiration date2-8°C until expiration dateIdentical
PrincipalOperatorTrained professionalsTrained professionalsIdentical
InstrumentHitachi 717Abbott Architect C4000Different
Package InsertHeader and FooterAdditional MaterialPrecautious and WarningSpecimen Collection andPreparationReferenceHeader and FooterAdditional MaterialPrecautious and WarningSpecimen Collection andPreparationReferenceDifferent
ProductLabelingKit LabelingComponent LabelingKit LabelingComponent LabelingDifferent
PerformanceCharacteristicsPrecisionAccuracySpecificity (structurally related)SensitivityPrecisionAccuracySpecificity (structurally related)Different

{6}------------------------------------------------

{7}------------------------------------------------

VI. Summary of Performance Testing

A. Precision:

In Qualitative, assay within run is ≤ 2% CV and total run precision is ≤ 5.0 % CV. In Semi-Quantitative, assay within run is ≤ 5% CV and total run precision is ≤ 7.5 % CV

B. Method Comparison:

Negative sample agreement is 100% in both Qualitative and Semi-Quantitative modes. Positive sample agreement is 100% in both Qualitative and Semi-Quantitative modes. Overall sample agreement is 100%for DRI Ecstasy Plus assay in both Qualitative and Semi-Quantitative mode.

C. Specificity:

Structurally related compounds that produced a positive result above the cut-off calibrator were reported at lowest concentrations.

Structurally related compounds that produced a negative result below the cut-off calibrator were reported at the highest concentrations.

Structurally unrelated compounds shall produce a negative result below the 500 ng/mL MDMA sample in both Qualitative and Semi-quantitative modes at the concentrations tested.

In both Qualitative and Semi-quantitative modes, assay produced a negative result below the 500 ng/mL MDMA cutoff when structurally unrelated compounds were spiked into negative pool urine at the tested concentrations.

Interference

In Qualitative mode, the rates of the low control samples spiked with interfering substances and pH adjusted were negative, and the rates of the high control samples spiked with interference substances and pH adjusted were positive.

In Qualitative mode, the rates of the low control level samples were negative, and the rates of the high control level samples were positive.

In Semi-quantitative mode, the recovery of the low control level sample spiked with interfering substances and pH adjusted were within 80 - 120% of the nominal value of 375 ng/mL, and recovery of the high control level sample spiked with interfering substances and pH adjusted were within 80 - 120% of the nominal value of 625 ng/mL ng/mL.

D. Stability:

The results from the open-vial condition (in-use) demonstrated that three lots of DRI Ecstasy Plus Reagent Kit, 500 mL kit met the acceptance criteria. The results demonstrate that the product is stable over the shelf life once opened and stored tightly

{8}------------------------------------------------

capped at 2-8°C. These results support the claim that the DRI Ecstasy Plus Reagent Kit, 500mL kits are stable for 36 months when stored unopened at 2-8°C.

VII. Conclusion

The information supports a determination of substantial equivalence between the candidate device DRI Ecstasy Plus and the predicate device DRI Ecstasy Enzyme lmmunoassay (K012110).

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).