K Number
K240627

Validate with FDA (Live)

Date Cleared
2024-06-04

(90 days)

Product Code
Regulation Number
866.3365
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LIAISON PLEX Yeast Blood Culture (BCY) Assay is a qualitative nucleic acid multiplex in vitro diagnostic test intended for use on the LIAISON PLEX System for simultaneous detection and identification of multiple potentially pathogenic fungal organisms in positive blood culture. The LIAISON PLEX BCY Assay is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system and which contain fungal organisms as determined by Gram Stain. The LIAISON PLEX BCY Assay detects and identifies the following fungal organisms:

Candida albicans
Candida auris
Candida dubliniensis
Candida famata
Candida glabrata
Candida guilliermondii
Candida kefyr
Candida krusei
Candida lipolytica
Candida lusitaniae
Candida parapsilosis
Candida tropicalis
Candida haemulonii / duobushaemulonii
Cryptococcus neoformans / gattii

The detection and identification of specific fungal nucleic acids from individuals exhibiting signs and/or symptoms of bloodstream infection aids in the diagnosis of bloodstream infection when used in conjunction with other clinical information. The results from LIAISON PLEX BCY Assay are intended to be interpreted in conjunction with Gram stain results and should not be used as the sole basis for diagnosis, treatment management decisions.

Negative results in the setting of a suspected bloodstream infection with pathogens that are not detected by this test. Positive results do not rule out co-infection with other organisms; the organism(s) detected by LIAISON PLEX BCY Assay may not be the definite cause of disease. Additional laboratory testing (e.g. sub-culturing of positive blood cultures for identification of organisms not detected by LIAISON PLEX BCY Assay, susceptibility testing and differentiation of mixed growth) and clinical presentation must be taken into consideration in the final diagnosis of bloodstream infection.

Device Description

The LIAISON PLEX "Yeast Blood Culture Assay (BCY Assay) is performed directly on blood culture media using blood culture bottles identified as positive by a continuous monitoring blood culture system, and which contain a fungal organism, as determined by a Gram stain. The system consists of an instrument, a single-use disposable test cartridge, and a transfer pipette. The user loads the sample into the sample port of the LIAISON PLEX Yeast Blood Culture Assay Cartridge. Next, the user sets up the sample order on the LIAISON PLEX System by first entering the sample information or scanning the barcode ID located on the sample tube, then scanning the barcode ID located on the test cartridge. Last, the user inserts the test cartridge into the processing module to initiate the test. The LIAISON PLEX System identifies the assay being run and automatically initiates the proper testing protocol to process the sample, analyze the data, and generate test results.

The LIAISON PLEX System automates the BCY Assay sample analysis through the following steps: a) Sample Preparation: Nucleic acid extraction via mechanical and chemical cell lysis and magnetic bead-based nucleic acid isolation; b) Amplification: Multiplex PCR based amplification of the extracted nucleic acid to generate target specific amplicons; c) Hybridization: Amplified DNA hybridizes to specific capture DNA arrayed on a glass slide in a microarray format and the bound target DNA, in turn, hybridizes with mediator and gold-nanoparticle probes; d) Signal Analysis: Gold nanoparticle probes bound specifically to target-containing spots in the microarray are silver-enhanced, and light scatter from the spots is measured and further analyzed to determine the presence (Detected) or absence (Not Detected) of a target.

AI/ML Overview

Acceptance Criteria and Device Performance for LIAISON PLEX Yeast Blood Culture Assay

The LIAISON PLEX Yeast Blood Culture (BCY) Assay is a qualitative nucleic acid multiplex in vitro diagnostic test for the simultaneous detection and identification of multiple potentially pathogenic fungal organisms in positive blood cultures. The study summarized below aimed to demonstrate the device meets its acceptance criteria through analytical and clinical performance evaluations.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the LIAISON PLEX BCY Assay clinical study were predefined as:

  • Sensitivity: ≥ 90% for all targets
  • Specificity: ≥ 95% for each target
  • Failure rate: ≤ 15%

The clinical performance data, combining prospective and pre-selected specimens, is summarized in the table below, comparing it against the acceptance criteria.

Table 1: LIAISON PLEX BCY Assay Clinical Performance (Acceptance Criteria vs. Reported Performance)

Pathogen TargetAcceptance Criteria (Sensitivity)Reported Sensitivity/PPA (Combined)Acceptance Criteria (Specificity)Reported Specificity/NPA (Combined)Overall Failure Rate (Initial Run)Overall Failure Rate (After Retest)Acceptance Criteria (Failure Rate)
Candida albicans≥ 90%100.0% (34/34)≥ 95%99.0% (97/98)2.9% (Overall)0.2% (Overall)≤ 15%
Candida auris≥ 90%100.0% (4/4)≥ 95%100.0% (128/128)
Candida dubliniensis≥ 90%NA (0/0)≥ 95%100.0% (132/132)
Candida famata≥ 90%NA (0/0)≥ 95%100.0% (132/132)
Candida glabrata≥ 90%100.0% (46/46)≥ 95%100.0% (86/86)
Candida guilliermondii≥ 90%NA (0/0)≥ 95%100.0% (132/132)
Candida haemulonii/C. duobushaemulonii≥ 90%NA (0/0)≥ 95%100.0% (132/132)
Candida kefyr≥ 90%100.0% (1/1)≥ 95%100.0% (131/131)
Candida krusei≥ 90%100.0% (4/4)≥ 95%100.0% (128/128)
Candida lipolytica≥ 90%NA (0/0)≥ 95%100.0% (132/132)
Candida lusitaniae≥ 90%100.0% (2/2)≥ 95%100.0% (130/130)
Candida parapsilosis≥ 90%100.0% (17/17)≥ 95%99.1% (114/115)
Candida tropicalis≥ 90%100.0% (6/6)≥ 95%97.6% (123/126)
Cryptococcus neoformans/Cryptococcus gattii≥ 90%100.0% (5/5)≥ 95%100.0% (127/127)

NA = Not applicable, as there were no positive cases for this target in the combined prospective/pre-selected dataset to calculate sensitivity, or no negative cases to calculate specificity.

The reported performance demonstrates that the LIAISON PLEX BCY Assay met all defined acceptance criteria.

2. Sample Sizes Used for the Test Set and Data Provenance

The test set for the clinical performance evaluation utilized a combination of prospective, pre-selected, and contrived specimens:

  • Prospective Specimens (Clinical Study): 69 unique specimens from four geographically diverse clinical sites in the United States. These were collected prospectively between June 2023 and October 2023. One initial specimen was excluded due to an inconclusive Gram stain result.
  • Pre-selected Specimens (Clinical Study): 63 remnant, de-identified specimens sourced from 6 different sites/vendors in the United States. The data provenance is retrospective clinical specimens that were pre-characterized.
  • Contrived Specimens: 829 specimens were artificially generated. These were blinded, randomized, and tested at all four testing sites during June 2023. These are contrived data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The provided document does not explicitly state the number of experts or their specific qualifications (e.g., radiologist with 10 years of experience) used to establish the ground truth for the test set.

However, the ground truth for the clinical study was established by comparing the LIAISON PLEX BCY Assay's performance against a Standard of Care culture followed by identification by Matrix Assisted Laser Desorption/Ionization coupled to time-of-flight Mass Spectrometry (MALDI-TOF MS) for all fungal targets. For the pre-selected specimens, it is mentioned that they were "characterized by an FDA cleared molecular assay prior to enrollment in the study." This implies that the ground truth relied on established laboratory methodologies and potentially an FDA-cleared reference method.

4. Adjudication Method for the Test Set

The document does not describe an adjudication method like "2+1" or "3+1" using human experts for the clinical study. The comparison was made against a "Standard of Care culture followed by identification by MALDI-TOF MS" as the reference method, which serves as the definitive ground truth for the presence and identification of the fungal organisms. For the pre-selected specimens, an "FDA cleared molecular assay" was used for characterization prior to study entry. These are objective laboratory methods rather than subjective expert interpretations requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not reported. This device is an in vitro diagnostic (IVD) assay designed for automated detection and identification of fungal organisms directly from blood culture samples. Its performance is evaluated against laboratory reference methods (culture and MALDI-TOF MS, or existing molecular assays), not against human reader interpretation of images or other subjective assessments. Therefore, there is no discussion of how human readers improve with or without AI assistance, as AI assistance is not part of the described use case for this IVD device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, the performance presented for the LIAISON PLEX BCY Assay in the analytical and clinical performance sections is for the standalone algorithm/device performance directly on the sample, without human interpretation or intervention in the diagnostic output. The system automates sample analysis, including nucleic acid extraction, amplification, hybridization, and signal analysis, to generate detected or not detected results. The indication for use clearly states the results are "intended to be interpreted in conjunction with Gram stain results and should not be used as the sole basis for diagnosis, treatment management decisions," but this refers to the clinical application context, not a human-in-the-loop for the device's fundamental diagnostic accuracy calculation.

7. The Type of Ground Truth Used

The primary ground truth used for the clinical study was:

  • Standard of Care culture followed by identification by Matrix Assisted Laser Desorption/Ionization coupled to time-of-flight Mass Spectrometry (MALDI-TOF MS). This is a laboratory-based, objective method for identifying microorganisms.
  • For pre-selected specimens, ground truth was established by an previously FDA-cleared molecular assay.
  • For contrived specimens, the ground truth was inherently known based on how the samples were prepared (spiking known organisms).

8. The Sample Size for the Training Set

The document does not report specific sample sizes for a separate training set for this device. The information provided heavily details analytical and clinical verification/validation (test set) studies. For diagnostic devices like this (nucleic acid assays), the development process typically involves extensive analytical characterization (Limit of Detection, Inclusivity, Exclusivity, Interference, Reproducibility) and then clinical validation. It is probable that internal development, optimization, and early verification studies used various sets of samples, but these are not explicitly termed "training sets" in the context of machine learning model training as one might expect for AI/ML-based diagnostic software.

9. How the Ground Truth for the Training Set Was Established

As no explicit "training set" is documented (in the context of AI/ML), there is no description of how ground truth was established for such a set. However, for the analytical studies and development of the assay, ground truth would have been established through well-characterized reference strains and clinical isolates, quantified using standard microbiology techniques (e.g., CFU/mL for Limit of Detection). These methods involve culturing, molecular characterization, and established laboratory practices to confirm the identity and concentration of the microorganisms.

{0}------------------------------------------------

June 4, 2024

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Luminex Corporation Sarah Herzog Sr. Regulatory Affairs Associate 4088 Commercial Avenue Northbrook, Illinois 60062

Re: K240627

Trade/Device Name: LIAISON PLEX Yeast Blood Culture Assay Regulation Number: 21 CFR 866.3365 Regulation Name: Multiplex Nucleic Acid Assay For Identification Of Microorganisms And Resistance Markers From Positive Blood Cultures Regulatory Class: Class II Product Code: PEO, NSU Dated: March 5, 2024 Received: March 6, 2024

Dear Sarah Herzog:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerelv.

Noel J. Gerald -S

Noel J. Gerald, Ph.D. Branch Chief Bacterial Respiratory and Medical Countermeasures Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K240627

Device Name LIAISON PLEX Yeast Blood Culture Assay

Indications for Use (Describe)

The LIAISON PLEX Yeast Blood Culture (BCY) Assay is a qualitative nucleic acid multiplex in vitro diagnostic test intended for use on the LIAISON PLEX System for simultaneous detection and identification of multiple potentially pathogenic fungal organisms in positive blood culture. The LIAISON PLEX BCY Assay is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system and which contain fungal organisms as determined by Gram Stain. The LIAISON PLEX BCY Assay detects and identifies the following fungal organisms:

Candida albicans Candida auris Candida dubliniensis Candida famata Candida glabrata Candida guilliermondii Candida kefyr Candida krusei Candida lipolytica Candida lusitaniae Candida parapsilosis Candida tropicalis Candida haemulonii / duobushaemulonii Cryptococcus neoformans / gattii

The detection and identification of specific fungal nucleic acids from individuals exhibiting signs and/or symptoms of bloodstream infection aids in the diagnosis of bloodstream infection when used in conjunction with other clinical information. The results from LIAISON PLEX BCY Assay are intended to be interpreted in conjunction with Gram stain results and should not be used as the sole basis for diagnosis, treatment management decisions.

Negative results in the setting of a suspected bloodstream infection with pathogens that are not detected by this test. Positive results do not rule out co-infection with other organisms; the organism(s) detected by LIAISON PLEX BCY Assay may not be the definite cause of disease. Additional laboratory testing (e.g. sub-culturing of positive blood cultures for identification of organisms not detected by LIAISON PLEX BCY Assay, susceptibility testing and differentiation of mixed growth) and clinical presentation must be taken into consideration in the final diagnosis of bloodstream infection.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Luminex. The word "Luminex" is written in a bold, black font, with a red dot above the "i". Below the logo is the tagline "complexity simplified.", with a red dot at the end.

510(k) Summary

This Summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

Preparation date: 05 March 2024

A. 510(k) Number:

K240627

B. Purpose for Submission:

Traditional 510(k), New Device

C. Measurand:

Candida albicans, Candida auris, Candida dublinensis, Candida qlabrata, Candida guilliermondii, Candida kefyr, Candida krusei, Candida lipolytica, Candida lusitaniae, Candida parapsilosis, Candida tropicalis, Candida haemulonii/duobushaemulonii, and Cryptococcus neoformans/gatti

D. Type of Test:

Qualitative Real Time Polymerase Chain Reaction (PCR)

E. Applicant:

Sarah Herzog, Luminex Corporation 4088 Commercial Avenue Northbrook, IL 60062 (847) 400-9000

F. Proprietary and Established Names:

LIAISON PLEX® Yeast Blood Culture Assay

G. Regulatory Information:

ProductCodeClassificationRegulation SectionPanel
PEOII21 CFR 866.3365 – Multiplex Nucleic AcidAssay for Identification of Microorganismsand Resistance Markers from PositiveBlood Cultures83 (Microbiology)

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Luminex logo. The word "Luminex" is written in a bold, sans-serif font, with the letters in black. There is a red dot above the "i" in Luminex. A small registered trademark symbol is located to the right of the "x".

complexity simplit

LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission

H. Intended Use:

1. Intended use(s):

The LIAISON PLEX® Yeast Blood Culture (BCY) Assay is a qualitative nucleic acid multiplex in vitro diagnostic test intended for use on the LIAISON PLEX® System for simultaneous detection and identification of multiple potentially pathogenic fungal organisms in positive blood culture. The LIAISON PLEX® BCY Assay is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system and which contain fungal organisms as determined by Gram Stain.

The LIAISON PLEX® BCY Assay detects and identifies the following fungal organisms:

Candida albicans
Candida auris
Candida dubliniensis
Candida famata
Candida glabrata
Candida guilliermondii
Candida kefyr
Candida krusei
Candida lipolytica
Candida lusitaniae
Candida parapsilosis
Candida tropicalis
Candida haemulonii/duobushaemulonii
Cryptococcus neoformans/gattii

The detection and identification of specific fungal nucleic acids from individuals exhibiting signs and/or symptoms of bloodstream infection aids in the diagnosis of bloodstream infection when used in conjunction with other clinical information. The results from LIAISON PLEX® BCY Assay are intended to be interpreted in conjunction with Gram stain results and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

Negative results in the setting of a suspected bloodstream infection may be due to infection with pathogens that are not detected by this test. Positive results do not rule out co-infection with other organisms; the organism(s) detected by LIAISON PLEX® BCY Assay may not be the definite cause of disease. Additional laboratory testing (e.g. sub-culturing of positive blood cultures for identification of organisms not detected by LIAISON PLEX® BCY Assay, susceptibility testing and differentiation of mixed growth) and clinical presentation must be taken into consideration in the final diagnosis of bloodstream infection.

    1. Indication(s) for use:
      Same as intended use.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Luminex. The word "Luminex" is written in a bold, sans-serif font. The letters are black, except for a red dot above the "i". A registered trademark symbol is next to the x.

  • LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission 3. Special conditions for use statement(s):
    For prescription use only.

For in vitro diagnostic use only.

    1. Special instrument requirements:
      For use with LIAISON PLEX ® Systems.

l. Device Description:

The LIAISON PLEX "Yeast Blood Culture Assay (BCY Assay) is performed directly on blood culture media using blood culture bottles identified as positive by a continuous monitoring blood culture system, and which contain a fungal organism, as determined by a Gram stain. The system consists of an instrument, a single-use disposable test cartridge, and a transfer pipette. The user loads the sample into the sample port of the LIAISON PLEX Yeast Blood Culture Assay Cartridge. Next, the user sets up the sample order on the LIAISON PLEX System by first entering the sample information or scanning the barcode ID located on the sample tube, then scanning the barcode ID located on the test cartridge. Last, the user inserts the test cartridge into the processing module to initiate the test. The LIAISON PLEX System identifies the assay being run and automatically initiates the proper testing protocol to process the sample, analyze the data, and generate test results.

The LIAISON PLEX System automates the BCY Assay sample analysis through the following steps: a) Sample Preparation: Nucleic acid extraction via mechanical and chemical cell lysis and magnetic bead-based nucleic acid isolation; b) Amplification: Multiplex PCR based amplification of the extracted nucleic acid to generate target specific amplicons; c) Hybridization: Amplified DNA hybridizes to specific capture DNA arrayed on a glass slide in a microarray format and the bound target DNA, in turn, hybridizes with mediator and gold-nanoparticle probes; d) Signal Analysis: Gold nanoparticle probes bound specifically to target-containing spots in the microarray are silver-enhanced, and light scatter from the spots is measured and further analyzed to determine the presence (Detected) or absence (Not Detected) of a target.

Substantial Equivalence Information: J.

    1. Predicate device name(s): GenMark ePlex Blood Culture Identification Fungal Pathogen (BCID-FP)
    1. Predicate 510(k) number(s): K182690
    1. Comparison with predicate:

The following tables compare Luminex's LIAISON® Yeast Blood Culture Assay to the ePlex Blood Culture Identification Fungal Pathogen (BCID-FP).

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Luminex. The logo is black text with a red dot above the i. The text is bold and slightly slanted.

LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission

Comparison to Predicate

Predicate Device:Candidate Device:
Comparisonto PredicateDeviceGenMark ePlex Blood Culture IdentificationFungal Pathogen (BCID-FP), K182690LIAISON PLEX® Yeast Blood Culture Assay
Product CodePEOPEO
Regulation Number21 CFR 866.336521 CFR 866.3365
Organism DetectedCandida albicans, Candida auris, Candidadubliniensis, Candida famata, Candidaglabrata, Candida guilliermondii, Candidakefyr, Candida krusei, Candida lusitaniae,Candida parapsilosis, Candida tropicalis,Cryptococcus gattii, Cryptococcusneoformans, Fusarium, RhodotorulaCandida albicans, Candida auris, Candidadublinensis,Candida famata, Candida glabrata,Candida guilliermondii, Candida kefyr,Candida krusei, Candida lipolytica,Candida lusitaniae, Candida parapsilosis,Candida tropicalis, Candida haemulonii/duobushaemulonii, Cryptococcusneoformans/gatti
MeasurandNucleic acid from Organisms detectedNucleic acid from Organisms detected
Intended UseThe GenMark ePlex Blood CultureIdentification Fungal Pathogen (BCID-FP)Panel is a qualitative nucleic acid multiplexin vitro diagnostic test intended for use onGenMark's ePlex Instrument forsimultaneous detection and identification ofmultiple potentially pathogenic fungalorganisms in positive blood culture. TheePlex BCID-FP Panel is performed directly onblood culture samples identified as positiveby a continuous monitoring blood culturesystem and which contains fungalorganisms.The following fungal organisms are identifiedusing the ePlex BCID-FP Panel:The LIAISON PLEX Yeast Blood Culture(BCY) Assay is a qualitative nucleic acidmultiplex in vitro diagnostic test intendedfor use on the LIAISON PLEX System forsimultaneous detection and identificationof multiple potentially pathogenic fungalorganisms in positive blood culture. TheLIAISON PLEX BCY Assay is performeddirectly on blood culture samplesidentified as positive by a continuousmonitoring blood culture system andwhich contain fungal organisms asdetermined by Gram Stain.The LIAISON PLEX BCY Assay detects andidentifies the following fungal organisms:
Candida albicans Candida auris Candida dubliniensis Candida famata Candida glabrata Candida guilliermondii Candida kefyr Candida krusei Candida lusitaniae Candida parapsilosis Candida tropicalis Cryptococcus gattii Cryptococcus neoformans Fusarium RhodotorulaCandida albicans Candida auris Candida dublinensis Candida famata Candida glabrata Candida guilliermondii Candida kefyr Candida krusei Candida lipolytica Candida lusitaniae Candida parapsilosis Candida tropicalis Candida haemulonii/duobushaemulonii Cryptococcus neoformans/gatti
The detection and identification of specificfungal nucleic acids from individualsexhibiting signs and/or symptoms ofbloodstream infection aids in the diagnosisof bloodstream infection when used inconjunction with other clinical information.The results from the ePlex BCID-FP Panelare intended to be interpreted inconjunction with Gram stain results andshould not be used as the sole basis fordiagnosis, treatment, or other patientmanagement decisions.Negative results in the setting of asuspected bloodstream infection may bedue to infection with pathogens that are notdetected by this test. Positive results do notrule out co-infection with other organisms;the organism(s) detected by the ePlex BCID-FP Panel may not be the definite cause ofdisease. Additional laboratory testing (e.g.sub-culturing of positive blood cultures foridentification of organisms not detected byePlex BCID-FP Panel, susceptibility testingand differentiation of mixed growth) andclinical presentation must be taken intoconsideration in the final diagnosis ofbloodstream infection.The detection and identification ofspecific fungal nucleic acids fromindividuals exhibiting signs and/orsymptoms of bloodstream infection aidsin the diagnostic of bloodstreaminfection when used in conjunction withother clinical information. The resultsfrom the LIAISON PLEX BCY Assay areintended to be interpreted inconjunction with Gram stain results andshould not be used as the sole basis fordiagnosis, treatment, or other patientmanagement decisions.Negative results in the setting of asuspected bloodstream infection may bedue to infection with pathogens that arenot detected by this test. Positive resultsdo not rule out co-infection with otherorganisms; the organism(s) detected byLIAISON PLEX BCY Assay may not be thedefinite cause of disease. Additionallaboratory testing (e.g. sub-culturing ofpositive blood cultures for identificationof organisms not detected by LIAISONPLEX BCY Assay, susceptibility testingand differentiation of mixed growth) andclinical presentation must be taken intoconsideration in the final diagnosis ofbloodstream infection.
Automated System(Sample to Answer)AutomatedSame
InstrumentationePlex InstrumentLIAISON PLEX®
Sample TypesBlood CultureSame

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Luminex. The logo is black, with a red dot above the "i" in Luminex. The logo is simple and modern, and it is likely used on the company's website and marketing materials. The logo also has a registered trademark symbol on the bottom right.

LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Luminex. The word "Luminex" is written in a bold, sans-serif font, with the letters in black. There is a red dot above the "i" in Luminex. A registered trademark symbol is to the right of the "x".

LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission

K. Standards/Guidance Documents Referenced:

  • . Class II Special Controls Guideline: Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures (May 2015)
  • . Electronic Submission Template for Medical Device 510(k) Submissions - Guidance for Industry and Food and Drug Administration Staff (October 2, 2023).
  • . Content of Premarket Submissions for Device Software Functions - Guidance for Industry and Food and Drug Administration Staff (June 14, 2023).
  • . Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions - Guidance for Industry and Food and Drug Administration Staff (September 27, 2023).
  • . Statistical Guidance on Reporting Results from Studies Evaluating Diagnostic Tests - Guidance for Industry and FDA Staff (March 13, 2007).
  • . CLSI. User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline - Second Edition. CLSI document EP12-A2. Wayne, PA: Clinical and Laboratory Standards Institute; 2008.
  • CLSI. Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline. CLSI document EP25-A. Wayne, PA: Clinical and Laboratory Standards Institute; 2009.
  • . CLSI. Interference Testing in Clinical Chemistry. 3rd ed. CLSI guideline EP07. Wayne, PA: Clinical and Laboratory Standards Institute; 2018.
  • ISO 14971:2019 Medical devices - Application of risk management to medical devices
  • . IEC 62366-1:2015 Medical devices - Part 1: Application of usability engineering to medical devices
  • . ISO 62304:2006 Medical device software - Software life-cycle processes
  • . ISO 15223-1:2016: Medical Devices - Symbols to be used with medical device labels, labeling and information to be supplied - Part 1: General requirements
  • . IEC 61010-1 Ed. 3.0 2010: Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements
  • . EN 61010-2-101:2002/IEC 61010-2-101:2015: Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment.
  • . IEC 60601-1-2:2014 (Edition 4.0): Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
  • . ISO 13485:2016/EN ISO 13485:2016; Medical devices - Quality Management System -Requirements for regulatory purposes
  • . ISO 20916:2019; In vitro diagnostic medical devices. Clinical performance studies using specimens from human subjects. Good study practice
  • EN ISO 18113-1:2011; In vitro diagnostic medical devices - Information supplied by the manufacturer (labeling). Terms, definition and general requirements

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the Luminex logo. The word "Luminex" is written in a bold, sans-serif font, with the "i" dotted with a red circle. The logo has a registered trademark symbol to the right of the word.

LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission

  • . EN ISO 18113-2:2011; In vitro diagnostic medical devices - Information supplied by the manufacturer (labeling) – Part 2: In vitro diagnostic reagents for professional use
  • . EN ISO 18113-3:2011; In vitro diagnostic medical devices - Information supplied by the manufacturer (labeling) – Part 3: In vitro diagnostic instruments for professional use
  • EN ISO 23640:2015; In vitro diagnostic medical devices - Evaluation of stability of in vitro diagnostic reagents
  • . IEC 61326-1:2012; Electrical equipment for measurement control and laboratory use -EMC requirements - Part 1: General requirements
  • . EN 61326-2-6:2006/IEC 61326-2-6:2012; Electrical equipment for measurement control and laboratory use - EMC requirements - Part 2-6: Particular requirements - In vitro diagnostic (IVD) medical equipment

L. Test Principle:

The LIAISON PLEX® Yeast Blood Culture Assay (BCY Assay) is performed directly on blood culture media using blood culture bottles identified as positive by a continuous monitoring blood culture system, and which contain a fungal organism, as determined by a Gram stain. The system consists of an instrument and a single-use, disposable test cartridge and a transfer pipette. The user loads the sample into the sample port of the LIAISON PLEX® Yeast Blood Culture Assay Cartridge. Next, the user sets up the sample order on the LIAISON PLEX System by first entering the sample information or scanning the barcode ID located on the sample tube, then scanning the barcode ID located on the test cartridge. Last, the user inserts the test cartridge into the processing module to initiate the test. The LIAISON PLEX® System identifies the assay being run and automatically initiates the proper testing protocol to process the sample, analyze the data, and generate test results.

The LIAISON PLEX® System automates the LIAISON PLEX® BCY Assay sample analysis through the following steps: a) Sample Preparation: Nucleic acid extraction via mechanical and chemical cell lysis and magnetic bead-based nucleic acid isolation; b) Amplification: Multiplex PCR based amplification of the extracted nucleic acid to generate target specific amplicons; c) Hybridization: Amplified DNA hybridizes to specific capture DNA arrayed on a glass slide in a microarray format and the bound target DNA, in turn, hybridizes with mediator and goldnanoparticle probes; d) Signal Analysis: Gold nanoparticle probes bound specifically to target-containing spots in the microarray are silver-enhanced, and light scatter from the spots is measured and further analyzed to determine the presence (Detected) or absence (Not Detected) of a target.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the Luminex logo. The word "Luminex" is written in a bold, black font, with a red dot above the "i". Below the logo, the words "complexity simplified." are written in a smaller, thinner font.

M. Performance Characteristics:

1. Analytical performance:

a. Analytical Sensitivity (Limit of Detection)

A limit of detection study (LoD) was performed to evaluate the analytical sensitivity of the LIAISON Plex® BCY Assay. Twenty-eight (28) strains and isolates that represent the 14 reportable targets of the LIAISON Plex BCY Assay were tested individually by serially diluting each target in a simulated blood culture sample matrix. Testing was broken into two parts; Preliminary LoD and Confirmation LoD. The Preliminary LoD concentrations were evaluated using a 6 point 3-fold dilution series and testing of at least four replicates per dilution. The Preliminary LoD for each target was defined as the lowest concentration at which 100% of four replicates were positive for the intended reportable target. With the preliminary LoD results, a 3-point 3 fold dilution series was made. Twenty replicates of each applicable dilution series were tested. Two concentrations were required with 20 replicates each to help determine the confirmed LoD for each strain. The confirmed LoD for each organism was defined as the lower concentration at which ≥ 95% of the 20 replicates were positive for the intended reportable target. The confirmed LoD for each target tested is listed in the following Table 1.

LIAISON PLEX TargetOrganismSource & IDConfirmedLoD (CFU/mL)
Candida albicansCandida albicansATCC 102313.14E+04
Candida albicansCandida albicansATCC 140532.83E+05
Candida kefyrCandida kefyrATCC 85533.33E+03
Candida kefyrCandida kefyrATCC 41353.33E+03
Candida famataCandida famataATCC 202787.77E+02
Candida famataCandida famataATCC 602296.99E+03
Candida dubliniensisCandida dubliniensisATCC MYA-5783.16E+03
Candida dubliniensisCandida dubliniensisATCC MYA-5773.16E+03
Candida glabrataCandida glabrataATCC 155459.99E+03
Candida glabrataCandida glabrataATCC 151261.00E+04
Candida kruseiCandida kruseiATCC 62589.83E+03
Candida kruseiCandida kruseiATCC 288702.95E+04
Candida guilliermondiiCandida guilliermondiiATCC 220173.33E+03
Candida guilliermondiiCandida guilliermondiiATCC 341341.00E+04
Candida lipolyticaCandida lipolyticaATCC 204603.18E+04
Candida lipolyticaCandida lipolyticaATCC 201773.19E+04
Candida lusitaniaeCandida lusitaniaeATCC 427201.11E+04
Candida lusitaniaeCandida lusitaniaeATCC 344493.33E+04

Table 1. LIAISON® PLEX BCY Assay Limit of Detection Results Summary

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for Luminex. The word "Luminex" is written in a bold, black font. Above the "i" in Luminex is a red dot. Below the logo is the phrase "complexity simplified."

LIAISON PLEX TargetOrganismSource & IDConfirmedLoD (CFU/mL)
Candida haemulonii/duobushaemuloniiCandida haemuloniiCBS 73753.51E+03
Candida duobushaemuloniiCBS 77983.51E+03
Candida aurisCandida aurisCBS 109133.50E+03
Candida aurisCBS 127663.50E+03
Candida parapsilosisCandida parapsilosisATCC 284742.85E+04
Candida parapsilosisATCC 284759.50E+03
Candida tropicalisCandida tropicalisATCC 138033.51E+03
Candida tropicalisATCC 2013803.51E+03
Cryptococcus neoformans/gattiiCryptococcus neoformansATCC 2088219.01E+04
Cryptococcus gattiiATCC MYA-48719.00E+04

b. Precision/Reproducibility:

Precision/Repeatability

Within-laboratory precision/repeatability for the LIAISON PLEX® BCY Assay was evaluated across a minimum of five non-consecutive days utilizing two operators, two LIAISON PLEX® Systems and one lot of the LIAISON PLEX® BCY Assay cartridges. The targets were randomized so that the target being tested was blinded to the operators performing the testing. Each target panel (Bottle/Ring Positive (RP), Bottle/Ring Positive + 8 hours (RP+8), Negative Blood Matrix (N), and Contrived-negative containing Escherichia coli + 8 hours (CN)) was run in triplicate by each operator for five non-consecutive days, totaling 15 replicates per target per operator.

The results of the within-laboratory precision / repeatability study are shown in Table 2.The results show the repeatability of the LIAISON PLEX® BCY Assay with an overall percent agreement of 100%, meaning the expected results were obtained for all analytes at all test levels across all samples, instruments, operators, and days.

ReportableTargetPanelCartridgeLotTarget PositivityOperator 1Target PositivityOperator 295% C.I.Overall95% C.I.Lower95% C.I.Upper
Instrument 1Instrument 2Instrument 1Instrument 2
C. albicansRing Positive1100% (7/7)100% (8/8)100% (6/6)100% (9/9)100% (30/30)88.7%100%
C. albicansRing Positive+8 hours1100% (8/8)100% (7/7)100% (5/5)100% (10/10)100% (30/30)88.7%100%
C. albicansNegativeBlood Matrix10% (0/7)0% (0/8)0% (0/8)0% (0/7)0% (0/30)88.7%100%

Table 2. Within Laboratory Precision / Repeatability Confidence Interval

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for Luminex. The word "Luminex" is written in a bold, sans-serif font, with a red dot above the "i". Below the logo, the words "complexity simplified." are written in a smaller, sans-serif font.

ReportableTargetPanelCartridgeLotOperator 1Operator 2Overall95% C.I.
Instrument1Instrument2Instrument1Instrument2LowerUpper
C. tropicalisRing Positive100% (7/7)100% (8/8)100% (6/6)100% (9/9)100%(30/30)88.7%100%
Ring Positive+8 hours100% (8/8)100% (7/7)100% (5/5)100%(10/10)100%(30/30)88.7%100%
NegativeBlood Matrix0% (0/7)0% (0/8)0% (0/8)0% (0/7)0%(0/30)88.7%100%
C.neoformansRing Positive100% (7/7)100% (8/8)100% (6/6)100% (9/9)100%(30/30)88.7%100%
Ring Positive+8 hours100% (8/8)100% (7/7)100% (5/5)100%(10/10)100%(30/30)88.7%100%
NegativeBlood Matrix0% (0/7)0% (0/8)0% (0/8)0% (0/7)0%(0/30)88.7%100%
ContrivedNegative(E. coli)N/A0% (0/6)0% (0/9)0% (0/9)0% (0/6)0%(0/30)88.7%100%

Lot-to-lot Reproducibility

Lot-to-lot reproducibility of the LIAISON PLX BCY Assay was evaluated by testing three different lots of LIAISON PLEX BCY Assay cartridges with one operator over five non-consecutive days, and with two runs per lot. The same target panels used for Precision/Repeatability were used for Lot-to-Lot reproducibility. The blinded reproducibility panels were tested in triplicate for each sample type by operator. Results of the Lot-to-Lot reproducibility are summarized below in Table 3. Overall percent agreement for Lot-to-Lot reproducibility was 100%.

Table 3 Lot-to-Lot Reproducibility Confidence Intervals

ReportablePanelCartridge LotTarget95% confidence interval
TargetPositivityLowerUpper
C. albicansRPLot 1 (051823106ADEVIUO)100% (15/15)79.6%100%
Lot 2 (032823106ADEVIUO)100% (15/15)79.6%100%
Lot 3 (061923106ADEVIUO)100% (15/15)79.6%100%
Overall100% (45/45)92.1%100%
RP+8Lot 1 (051823106ADEVIUO)100% (15/15)79.6%100%Lot 2 (032823106ADEVIUO)100% (15/15)79.6%100%Lot 3 (061923106ADEVIUO)100% (15/15)79.6%100%Overall100% (45/45)92.1%100%
RP+8Lot 1 (051823106ADEVIUO)100% (15/15)79.6%100%
Lot 2 (032823106ADEVIUO)100% (15/15)79.6%100%
Lot 3 (061923106ADEVIUO)100% (15/15)79.6%100%
Overall100% (45/45)92.1%100%

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the Luminex logo. The logo is black, except for a red dot above the "i" in Luminex. The logo is a sans-serif font and has a registered trademark symbol.

complexity sim

LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission

ReportableTargetPanelCartridge LotTargetPositivity95% confidence interval
TargetLowerUpper
C. tropicalisRPLot 1 (051823106ADEVIUO)100% (15/15)79.6%100%
Lot 2 (032823106ADEVIUO)100% (15/15)79.6%100%
Lot 3 (061923106ADEVIUO)100% (15/15)79.6%100%
Overall100% (45/45)92.1%100%
C. neoformansRP+8Lot 1 (051823106ADEVIUO)100% (15/15)79.6%100%
Lot 2 (032823106ADEVIUO)100% (15/15)79.6%100%
Lot 3 (061923106ADEVIUO)100% (15/15)79.6%100%
Overall100% (45/45)92.1%100%
NegativeRPLot 1 (051823106ADEVIUO)100% (15/15)79.6%100%
Lot 2 (032823106ADEVIUO)100% (15/15)79.6%100%
Lot 3 (061923106ADEVIUO)100% (15/15)79.6%100%
Overall100% (45/45)92.1%100%
RP+8Lot 1 (051823106ADEVIUO)100% (15/15)79.6%100%
Lot 2 (032823106ADEVIUO)100% (15/15)79.6%100%
Lot 3 (061923106ADEVIUO)100% (15/15)79.6%100%
Overall100% (45/45)92.1%100%
CNLot 1 (051823106ADEVIUO)0% (0/15)79.6%100%
Lot 2 (032823106ADEVIUO)0% (0/15)79.6%100%
Lot 3 (061923106ADEVIUO)0% (0/15)79.6%100%
Overall0% (0/45)92.1%100%
NLot 1 (051823106ADEVIUO)0% (0/15)79.6%100%
Lot 2 (032823106ADEVIUO)0% (0/15)79.6%100%
Lot 3 (061923106ADEVIUO)0% (0/15)79.6%100%
Overall0% (0/45)92.1%100%

Site-to-site Reproducibility

Site-to-site reproducibility of the LIAISON PLEX BCY Assay was evaluated by testing one lot of LIAISON PLEX BCY Assay cartridges with two operators at each of three sites over five nonconsecutive days. Four target panels were prepared and tested across all sites and operators to evaluate site-to-site reproducibility. The same target panels used for Precision/Repeatability and Lot-to-Lot reproducibility were used for Site-to-Site reproducibility. The blinded reproducibility panels were tested in triplicate for each sample type by each operator on each testing day.

Results of the site-to-site reproducibility study are summarized in Table 4. Reproducibility of

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the Luminex logo. The word "Luminex" is in bold, black font, with a red dot above the "i". Below the logo, the words "complexity simplified." are written in a smaller, thinner font, with a red dot at the end.

the LIAISON PLEX® BCY Assay across three sites was 99.7%.

Panel TypeSiteResults (% Agreement)95% C.I.
Ring PositiveSite 1100% (30/30)94.0% - 100%
Site 296.7% (29/30)83.3% - 99.4%
Site 3100% (30/30)88.6% - 100%
All Sites98.9% (89/90)94.0% - 99.8%
Ring Positive + 8 hoursSite 1100% (30/30)88.6% - 100%
Site 2100% (30/30)88.6% - 100%
Site 3100% (30/30)88.6% - 100%
All Sites100% (90/90)95.9% - 100%
Negative Blood MatrixSite 1100% (30/30)88.6% - 100%
Site 2100% (30/30)88.6% - 100%
Site 3100% (30/30)88.6% - 100%
All Sites100% (90/90)95.9% - 100%
Contrived NegativeSite 1100% (30/30)88.6% - 100%
Site 2100% (30/30)88.6% - 100%
Site 3100% (30/30)88.6% - 100%
All Sites100% (90/90)95.9% - 100%
Overall99.7% (359/360)98.4% - 100%

Table 4. LIAISON® PLEX BCY Assay Site to Site Reproducibility Results

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the logo for Luminex. The word "Luminex" is written in a bold, sans-serif font, with a red dot above the "i". Below the word "Luminex" is the tagline "complexity simplified.", written in a smaller, thinner font. The logo is simple and modern, and the red dot adds a pop of color.

  • ﻥ Linearity/assay reportable range:
    Not applicable. The LIAISON® PLEX Yeast Blood Culture Assay is a qualitative assay.

  • d. Traceability, Stability, Expected values (controls, calibrators, or methods):

Controls:

Several controls are built into the assay and system to ensure identification of processing errors and to establish validity of test results.

Internal Controls

Each LIAISON® PLEX BCY Assay cartridge includes internal controls to ensure performance of sample preparation, amplification, and detection. Internal control is automatically added to the sample prior to initiation of sample preparation and assesses extraction, nucleic acid recovery, amplification, and detection. Finally, a post-amplification hybridization control serves as an indicator of successful hybridization. Internal control results are reported as Pass or Fail on the printed reports (see the following Table 5 for detailed explanations of each control result). Internal controls must generate a signal above the threshold in each internal reaction for the system to report a valid test result.

Internal ControlResultExplanationSuggested Action
PassTest was completed and internal controls weresuccessful, indicating that valid results weregenerated.Review and report result
FailOne or more internal controls failed.Repeat test with a newcartridge

Table 5. Interpretation of Controls on the LIAISON® PLEX BCY Assay Report

External Controls

Positive and negative external controls should be tested with each new lot or shipment of reagents, or monthly, (whichever occurs first), or in accordance with updated local, regional, state, and/or federal guidelines. Verified negative blood matrix can be used as the negative control. Previously characterized positive samples or verified negative blood matrix spiked with well characterized organisms may be used as the external positive control. External controls should be used in accordance with laboratory protocols and in accordance with local, state, and federal accrediting organizations, as applicable.

Stability:

Specimen Stability

Contrived specimen stability at incubated storage (33°C - 37°C), refrigerated (2°C - 8°C), and frozen (-80°C ± 15) storage was evaluated for use with the LIAISON" PLEX BCY Assay. Two reference panels, each made of three organisms, were tested at concentrations representing bottle positivity. The panels were then stored at incubated, refrigerated, or frozen temperature conditions. Testing occurred at various time points up to 1 month for

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the Luminex logo. The word "Luminex" is written in a bold, sans-serif font, with the letters in black. There is a red dot above the "i" in Luminex. To the right of the word is a registered trademark symbol.

LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission the frozen storage, up to 1 month for refrigerated storage, and up to 72 hours for incubated storage. Specimen stability was also evaluated for specimens grown in a continuous blood monitoring system by testing fresh blood bottle ring positive (RP) and blood bottle ring positive plus 12 hours (RP + 12) samples.

The results of this study demonstrated that specimens stored frozen (< -80°C) are stable for up to 1 month, specimens stored refrigerated (2°C - 8°C) are stable for up to 1 month and specimens stored at incubation (33°C - 37°C) are stable for up to 72 hours. Specimens incubated up to 12 hours after bottle ring positive were 100% detected.

Device Stability

A shelf-life study was conducted to evaluate the real-time stability of the LIAISON PLEX BCY Assay at the recommended storage conditions of room temperature (15°C - 30°C). Real-time stability was assessed using three Positive Control panels which each consisted of representative on-panel fungal organisms, and one Negative Control panel which consisted of negative blood matrix. Results of real-time stability demonstrated the LIAISON PLEX BCY Assay is stable for at least 3 months when stored at 15°C - 30°C.

Open-box stability of the LIAISON PLEX BCY Assay was also evaluated by testing four lots of LIAISON PLEX BCY Assay after removal from their foil pouches and stored at room temperature. The same Positive Control panels and Negative Control panel used for real-time stability were used for open-box stability. Results of open-box stability indicate the cartridges are stable for up to eight hours after cartridges are removed from their foil pouches and stored at room temperature.

Fresh vs. Frozen Specimen Stability

A study was performed to assess tolerance of the LIAISON® PLEX BCY Assay to correctly detect specimens containing fungal organisms that were tested fresh (i.e. unfrozen) which then underwent one or two freeze-thaw cycles and were tested after each freeze-thaw cycle. This was executed by testing three replicates of each target organism at each of the following concentrations: Blood bottle ring positive (RP), Blood bottle ring positive + 8 hours (RP+8) and Blood bottle ring positive at a ten-fold dilution concentration (RP - 10).

Each target demonstrated an overall positive agreement of 100% at all three testing conditions of fresh, 1 freeze-thaw, and 2 freeze-thaws. Data supports the stability of specimens for up to two freeze-thaw cycles.

{18}------------------------------------------------

Image /page/18/Picture/0 description: The image shows the logo for Luminex. The word "Luminex" is written in a bold, sans-serif font, with the letters in black. There is a red dot above the "i" in "Luminex". The logo is simple and modern.

LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission

  • e. Growth and Detection Study:
    A Growth and Detection study was performed to evaluate expected organism concentrations in blood cultures at bottle positivity (Ring Positive) and eight hours after bottle positivity (RP+8), using one representative strain for each organisms.

A minimum of one ring positive bottle and one 8 hour plus ring positive bottle from each BCY organism were tested in triplicate on the LIAISON PLEX® BCY assay fresh, meaning the sample was retrieved from the automated blood culture system within the specified timeframe and was tested the same day. A minimum of two ring positive bottles and two 8 hour plus ring positive bottles from each BCY organism were tested in triplicate on the LIAISON PLEX® BCY assay either fresh or after being stored frozen at -65°C to -95°C. In addition, one negative blood bottle was tested fresh in triplicate on the LIAISON PLEX® BCY assay. Concentrations shown below in Table 6 represent approximate levels that may be observed in a clinical setting.

OrganismRepresentativeStrainTesting Condition(RP/ RP+8)Concentration(CFU/mL)TargetPositivity
Candida albicansCandidaalbicansRP Fresh2.21E+04100%
RP Frozen4.30E+05100%
RP Frozen8.60E+05100%
RP + 8 Fresh1.83E+06100%
RP + 8 Frozen1.54E+07100%
RP + 8 Frozen6.90E+06100%
Candida aurisCandida aurisRP Fresh1.88E+07100%
RP Fresh1.45E+07100%
RP Fresh2.02E+07100%
RP + 8 Fresh5.30E+07100%
RP + 8 Frozen4.20E+07100%
RP + 8 Frozen4.80E+07100%
CandidadubliniensisCandidadubliniensisRP Fresh1.47E+07100%
RP Fresh1.98E+07100%
RP Fresh3.20E+07100%
RP + 8 Fresh4.20E+07100%
RP + 8 Fresh4.20E+07100%
RP + 8 Fresh5.10E+07100%
Candida famataDebarymoycesfabryiRP Fresh7.30E+06100%
RP Frozen4.90E+06100%
RP Frozen3.20E+06100%
RP + 8 Fresh1.20E+07100%
RP + 8 Frozen1.80E+07100%
RP + 8 Frozen5.90E+06100%
Candida glabrataCandidaglabrataRP Fresh8.00E+06100%
RP Frozen3.80E+06100%
RP Frozen6.10E+06100%
RP + 8 Fresh1.44E+08100%
RP + 8 Frozen1.36E+08100%
RP + 8 Frozen1.60E+08100%

{19}------------------------------------------------

Image /page/19/Picture/0 description: The image shows the Luminex logo. The word "Luminex" is written in a bold, sans-serif font, with the letters in black. There is a red dot above the "i" in Luminex. A small registered trademark symbol is located to the right of the "x".

OrganismRepresentativeStrainTesting Condition(RP/ RP+8)Concentration(CFU/mL)TargetPositivity
CandidaguilliermondiiMeyerozymaguilliermondiiRP Fresh1.91E+07100%
RP Fresh3.60E+07100%
RP Fresh1.86E+07100%
RP + 8 Fresh3.60E+07100%
RP + 8 Fresh3.80E+07100%
RP + 8 Fresh3.40E+07100%
Candidahaemulonii/duobushaemuloniiCandidahaemulonisvar.haemulonisRP Fresh9.70E+06100%
RP Fresh9.40E+06100%
RP Fresh1.07E+07100%
RP + 8 Fresh3.20E+07100%
RP + 8 Fresh3.50E+07100%
RP + 8 Fresh4.30E+07100%
Candida kefyrCandida kefyrRP Fresh8.20E+06100%
RP Fresh6.30E+06100%
RP Fresh1.20E+07100%
RP + 8 Fresh2.34E+07100%
RP + 8 Fresh2.07E+07100%
RP + 8 Fresh2.15E+07100%
Candida kruseiCandida kruseiRP Fresh1.24E+07100%
RP Frozen1.66E+07100%
RP Frozen2.09E+07100%
RP + 8 Fresh4.00E+07100%
RP + 8 Fresh4.20E+07100%
RP + 8 Fresh3.50E+07100%
RP Fresh2.44E+06100%
RP Frozen2.50E+06100%
Candida lipolyticaYarrowialipolyticaRP Frozen2.42E+06100%
RP + 8 Frozen3.90E+06100%
RP + 8 Fresh4.60E+06100%
RP + 8 Fresh3.80E+06100%
RP Fresh2.23E+07100%
RP Frozen5.10E+07100%
RP Frozen3.50E+07100%
Candida lusitaniaeCandidalusitaniaeRP + 8 Fresh1.33E+08100%
RP + 8 Fresh1.22E+08100%
RP + 8 Fresh1.32E+08100%
RP Fresh1.54E+07100%
RP Fresh1.61E+07100%
RP Fresh1.76E+07100%
CandidaparapsilosisCandidaparapsilosisRP + 8 Fresh4.10E+07100%
RP + 8 Fresh3.90E+07100%
RP + 8 Fresh5.50E+07100%
RP Fresh1.31E+07100%
RP Fresh1.20E+07100%
RP Fresh1.35E+07100%
Candida tropicalisCandidatropicalisRP + 8 Fresh3.00E+07100%

{20}------------------------------------------------

Image /page/20/Picture/0 description: The image shows the Luminex logo. The word "Luminex" is written in a bold, sans-serif font, with the letters in black. There is a red dot above the "i" in Luminex. A registered trademark symbol is located to the right of the "x".

complexity simplit

LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission
OrganismRepresentativeStrainTesting Condition(RP/ RP+8)Concentration(CFU/mL)TargetPositivity
RP + 8 Fresh3.90E+07100%
RP + 8 Fresh3.30E+07100%
Cryptococcusneoformans/gattiiCryptococcusneoformansvar. grubiiRP Fresh1.84E+06100%
RP Fresh3.40E+06100%
RP Fresh2.09E+06100%
RP + 8 Fresh5.00E+06100%
RP + 8 Frozen5.50E+06100%
RP + 8 Frozen4.40E+06100%
NBMNAFreshNA0%

f. Analytical Reactivity (Inclusivity)

The analytical reactivity (inclusivity) of the LIAISON PLEX BCY Assay was performed to evaluate inclusive detection of multiple strains of each organism, representing all 14 reportable targets of 16 yeast species, for a total of 80 strains. Assay performance at the representative concentration of bottle positivity was assessed with five strains of each organism diluted in Negative Blood Matrix and tested in triplicate. Bottle positivity represents clinically relevant titers of each organism strain and was determined in execution of the Growth & Detection study for the LIAISON PLEX® BCY Assay. The results, presented in the Table 7, demonstrated 100% positivity for each strain.

Reportable TargetOrganism StrainSource IDConcentration(CFU/mL)TargetPositivity
Candida albicansCandida albicansATCC 900284.4E+05100% (3/3)
ATCC 244334.4E+05100% (3/3)
ATCC 900294.4E+05100% (3/3)
ATCC MYA-47804.4E+05100% (3/3)
Zeptometrix 8015044.4E+05100% (3/3)
Candida aurisCandida aurisCBS-KNAW 123731.5E+07100% (3/3)
CBS-KNAW 127671.5E+07100% (3/3)
CBS-KNAW 127681.5E+07100% (3/3)
CDC AR-03831.5E+07100% (3/3)
Zeptometrix 8043861.5E+07100% (3/3)
Candida dubliniensisCandida dubliniensisATCC MYA-5821.5E+07100% (3/3)
ATCC MYA-5791.5E+07100% (3/3)
ATCC 445081.5E+07100% (3/3)
ATCC MYA-29751.5E+07100% (3/3)
Zeptometrix 8019151.5E+07100% (3/3)
Candida famataDebaryomyces hansenii¹ATCC 93653.2E+06100% (3/3)
ATCC 106193.2E+06100% (3/3)
ATCC 362393.2E+06100% (3/3)
CBS-KNAW 19613.2E+06100% (3/3)
CBS-KNAW 142663.2E+06100% (3/3)
Candida glabrataCandida glabrataATCC MYA-29503.8E+06100% (3/3)
Reportable TargetOrganism StrainSource IDConcentration(CFU/mL)TargetPositivity
ATCC 341383.8E+06100% (3/3)
ATCC 660323.8E+06100% (3/3)
ATCC 908763.8E+06100% (3/3)
Zeptometrix 8015353.8E+06100% (3/3)
Candida guilliermondiiMeyerozyma guilliermondiiATCC 568221.9E+07100% (3/3)
Candida guilliermondiiMeyerozyma guilliermondiiATCC 142421.9E+07100% (3/3)
Candida guilliermondiiMeyerozyma guilliermondiiATCC 62601.9E+07100% (3/3)
Candida guilliermondiiMeyerozyma guilliermondiiATCC 901971.9E+07100% (3/3)
Candida guilliermondiiCandida guilliermondiiZeptometrix 8016021.9E+07100% (3/3)
Candida haemulonii/duobushaemuloniiCandida haemuloniiCDC AR-03959.4E+06100% (3/3)
Candida haemulonii/duobushaemuloniiCandida haemulonisCBS-KNAW 109709.4E+06100% (3/3)
Candida haemulonii/duobushaemuloniiCandida haemulonisCBS-KNAW 109739.4E+06100% (3/3)
Candida haemulonii/duobushaemuloniiCandida haemulonii var. vulneraCBS-KNAW 124379.4E+06100% (3/3)
Candida haemulonii/duobushaemuloniiCandida haemulonii var. vulneraCBS-KNAW 124389.4E+06100% (3/3)
Candida haemulonii/duobushaemuloniiCBS-KNAW 69159.4E+06100% (3/3)
Candida haemulonii/duobushaemuloniiCBS-KNAW 70989.4E+06100% (3/3)
Candida haemulonii/duobushaemuloniiCandida duobushaemuloniiCDC AR-03929.4E+06100% (3/3)
Candida haemulonii/duobushaemuloniiCandida duobushaemuloniiCDC AR-03949.4E+06100% (3/3)
Candida haemulonii/duobushaemuloniiCandida duobushaemuloniiCDC AR-03919.4E+06100% (3/3)
Candida kefyrCandida kefyrATCC 25126.3E+06100% (3/3)
Candida kefyrCandida kefyrATCC 660286.3E+06100% (3/3)
Candida kefyrCandida kefyrATCC 2040936.3E+06100% (3/3)
Candida kefyrCandida kefyrATCC 41356.3E+06100% (3/3)
Candida kefyrCandida kefyrZeptometrix 8018376.3E+06100% (3/3)
Candida kruseiCandida kruseiATCC 321961.2E+07100% (3/3)
Candida kruseiCandida kruseiATCC 142431.2E+07100% (3/3)
Candida kruseiCandida kruseiATCC 229851.2E+07100% (3/3)
Candida kruseiCandida kruseiATCC 341351.2E+07100% (3/3)
Candida kruseiCandida kruseiZeptometrix 8015361.2E+07100% (3/3)
Candida lipolyticaCandida lipolyticaATCC 86622.4E+06100% (3/3)
Candida lipolyticaYarrowia lipolyticaATCC 2012422.4E+06100% (3/3)
Candida lipolyticaYarrowia lipolyticaATCC 908122.4E+06100% (3/3)
Candida lipolyticaYarrowia lipolyticaATCC MYA-26132.4E+06100% (3/3)
Candida lipolyticaYarrowia lipolyticaATCC 323382.4E+06100% (3/3)
Candida lusitaniaeCandida lusitaniaeATCC 2009532.2E+07100% (3/3)
Candida lusitaniaeCandida lusitaniaeCBS-KNAW 48702.2E+07100% (3/3)
Candida lusitaniaeCandida lusitaniaeATCC 660352.2E+07100% (3/3)
Candida lusitaniaeClavispora lusitaniaeCBS-KNAW 69362.2E+07100% (3/3)
Candida lusitaniaeClavispora lusitaniaeZeptometrix 8016031.9E+072100% (3/3)
Candida parapsilosisCandida parapsilosisATCC 341361.5E+07100% (3/3)
Candida parapsilosisCandida parapsilosisATCC 961371.5E+07100% (3/3)
Reportable TargetOrganism StrainSource IDConcentration(CFU/mL)TargetPositivity
ATCC 2009541.5E+07100% (3/3)
ATCC 382911.5E+07100% (3/3)
Zeptometrix 8015371.5E+07100% (3/3)
Candida tropicalisCandida tropicalisATCC MYA-27341.2E+07100% (3/3)
ATCC 13691.2E+07100% (3/3)
ATCC 908741.2E+07100% (3/3)
ATCC 188071.2E+07100% (3/3)
Zeptometrix 8015381.2E+07100% (3/3)
Cryptococcusneoformans/gattiiCryptococcus neoformansvar. grubiiATCC 141161.8E+06100% (3/3)
Cryptococcus neoformansATCC 901121.8E+06100% (3/3)
ATCC MYA-45641.8E+06100% (3/3)
ATCC MYA-45661.8E+06100% (3/3)
Zeptometrix 8018031.8E+06100% (3/3)
Cryptococcus gattiiATCC MYA-45631.8E+06100% (3/3)
ATCC MYA-48731.8E+06100% (3/3)
ATCC MYA-45601.8E+06100% (3/3)
ATCC MYA-40941.8E+06100% (3/3)
Cryptococcus gattiiZeptometrix 8019171.8E+06100% (3/3)

Table 7. LIAISON PLEX® BCY Inclusivity Summary

{21}------------------------------------------------

Image /page/21/Picture/0 description: The image shows the logo for Luminex. The logo is black, with a red dot above the "i". The word "Luminex" is written in a bold, sans-serif font.

LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission

{22}------------------------------------------------

Image /page/22/Picture/0 description: The image shows the logo for Luminex. The word "Luminex" is written in a bold, sans-serif font. There is a red dot above the "i" in Luminex. The logo is simple and modern.

complexity simplified.

¹Debaryomyces hansenii is a synonymous taxon name for Candida famata.

2 The stock concentration of Clavispora lusitaniae (Zeptometrix 801603) was not concentrative bottle positivity concentration. This prepared as stock material diluted 1:1 in Negative Blood Matrix which is allowed per protocol directive.

Predicted (in silico) Reactivity (Inclusivity) Results

In-silico inclusivity analyses of the assay oligo sequences were performed against sequences available in the GenBank nt database as of December 30, 2023. The inclusivity analyses involved assessing the precent homology of each oligo sequence to its binding region on each target sequence retrieved from the public databases. The predicted inclusivity based on sequence homology is the percentage of sequences with at least 90% oligo identity. A summary of in-silico inclusivity analysis results are summarized below in Table 8.

{23}------------------------------------------------

Image /page/23/Picture/0 description: The image shows the logo for Luminex. The logo is black, except for a red dot above the "i" in Luminex. The word Luminex is in a bold, sans-serif font. There is a registered trademark symbol to the right of the "x".

complexity simplified.

Reportable TargetInclusive Organism/TargetTotal # SequencesinAlignment# Sequences withPercent OligoIdentity≥ 90%PredictedInclusivityPercentage(%)
Candida albicansCandida albicans181 *181100
Candida aurisCandida auris286286100
Candida dubliniensisCandida dubliniensis13 *13100
Candida famataCandida famata32 *32100
Candida glabrataCandida glabrata206206100
Candida guilliermondiiCandida guilliermondii24 *24100
Candida haemulonii/duobushamuloniiCandida haemulonii17 *17100
Candida duobushamulonii9 *9100
Candida kefyrCandida kefyr30 *30100
Candida kruseiCandida krusei48 *48100
Candida lipolyticaCandida lipolytica82 *82100
Candida lusitaniaeCandida lusitaniae75 *75100
Candida parapsilosisCandida parapsilosis59 *59100
Candida tropicalisCandida tropicalis86 *8599
Cryptococcusneoformans/gattiiCryptococcus neoformans32433243100
Cryptococcus gattii12361236100

Table 8. In-silico Analysis Results

  • Includes WGS (whole genome shotgun) sequences.

g. Analytical Specificity (Exclusivity)

Cross-Reactivity:

A Cross-Reactivity study was performed to evaluate the LIAISON PLEX® BCY Assay's analytical specificity through the detection of off-panel and potentially cross-reactive organisms, including those phylogenetically related to the on-panel organisms and organisms likely to be present in blood culture samples.

Out of the 40 off-panel bacterial species tested, there was no cross-reactivity with any of the on-panel organisms. Out of the 37 off-panel yeast species tested, 35 organisms had no crossreactivity with on-panel organisms. Yarrowia deformans (NRRL Y-63659) cross-reacted and had 100% detection for the BCY target Candida lipolytica when tested at a concentration of 1.0E+07 CFU/mL. Candida pseudohaemulonii (CBS-KNAW 10004) cross-reacted and had 100% detection for the BCY target Candida haemulonii/duobushaemulonii when tested at 1.0E+07 CFU/mL.

One bacterial organism, Klebsiella pneumoniae (ATCC-BAA 2146), had 11% positivity (1 of 9) for the on-panel organisms Candida albicans and Candida dubliniensis at a concentration of 1.0E+08 CFU/mL. Candida pseudohaemulonii (CBS-KNAW 10004) also had 11% detection (1 of 9) for the BCY target Candida lusitaniae when tested at 1.0E+07 CFU/mL. Trichosporon asteroides (CBS-KNAW 6183) had 50% positivity (3 of 6) for Candida glabrata and 33% (2 of 6) positivity for Candida haemulonii/duobushaemulonii at a concentration of 1.0E+07 CFU/mL. In-silico analysis and wet testing support that cross-reactivity is not expected to occur with these targets.

{24}------------------------------------------------

Image /page/24/Picture/0 description: The image shows the logo for Luminex. The word "Luminex" is in bold, black font, with a red dot above the "i". Below the logo is the tagline "complexity simplified." in a smaller, black font.

Refer to the Tables below which summarizes cross reactivity testing performed with Bacteria strains (Table 9) and Yeast strains (Table 10).

OrganismConcentrationTested (CFU/mL)Source &Source IDPositivity % (Organism)
Acinetobacter baumannii1.0E+08IHMA 1283070% (0/3)
Acinetobacter Iwoffii1.0E+08ATCC 153090% (0/3)
Aerococcus viridans1.0E+08ATCC 7004060% (0/3)
Bacillus cereus1.0E+08ATCC 145790% (0/3)
Bacteroides fragilis1.0E+08ATCC 252850% (0/3)
Bordetella pertussis1.0E+08ATCC 97970% (0/3)
Cutibacterium (Propionibacterium) acnes1.0E+08ATCC 331790% (0/3)
Citrobacter freundii1.0E+08ATCC 80900% (0/3)
Clostridium perfringens1.0E+08ATCC 131240% (0/3)
Corynebacterium striatum1.0E+08ATCC 437350% (0/3)
Enterobacter aerogenes1.0E+08ATCC 350290% (0/3)
Enterococcus avium1.0E+08ATCC 140250% (0/3)
Enterobacter cloacae1.0E+08ATCC 350300% (0/3)
Escherichia coli1.0E+08NCTC 138460% (0/3)
Enterococcus faecalis1.0E+08ATCC 292120% (0/3)
Enterococcus faecium1.0E+08ATCC 7002210% (0/3)
Eggerthella lenta1.0E+08ATCC 255590% (0/3)
Klebsiella oxytoca1.0E+08ATCC 431650% (0/3)
Klebsiella pneumoniae1.0E+08ATCC-BAA214611% (C. albicans) (1/9)11% (C. dubliniensis) (1/9)
Listeria monocytogenes1.0E+08ATCC 191150% (0/3)
Lactobacillus rhamnosus1.0E+08ATCC 531030% (0/3)
Micrococcus luteus8.9E+06ATCC 100540% (0/3)
Morganella morganii1.0E+08ATCC 258300% (0/3)
Pseudomonas aeruginosa1.0E+08IHMA 5766020% (0/3)
Prevotella denticola1.0E+08BEI ResourcesHM-11730% (0/3)
Proteus mirabilis1.0E+08ATCC 124530% (0/3)
Streptococcus agalactiae1.0E+08ATCC 124010% (0/3)
Streptococcus anginosus1.0E+08ATCC 333970% (0/3)
Staphylococcus aureus1.0E+08ATCC 259230% (0/3)
Salmonella enterica1.0E+08ATCC 99930% (0/3)
Staphylococcus epidermidis1.0E+08ATCC 7005670% (0/3)
Staphylococcus hominis1.0E+08ATCC 278440% (0/3)
Staphylococcus intermedius1.0E+08ATCC 296630% (0/3)
Staphylococcus lugdunensis1.0E+08ATCC 438090% (0/3)
Stenotrophomonas maltophilia8.9E+07ATCC 176660% (0/3)
Serratia marcescens1.0E+08NCTC 139200% (0/3)
OrganismConcentrationTested (CFU/mL)Source &Source IDPositivity % (Organism)
Streptococcus mitis1.0E+08ATCC 159140% (0/3)
Streptococcus pneumoniae1.0E+08ATCC 63150% (0/3)
Streptococcus pyogenes1.0E+08ATCC 129640% (0/3)
Staphylococcus saprophyticus1.0E+08ATCC 153050% (0/3)

Table 9. LIAISON PLEX® BCY Assay Analytical Specificity – Cross Reactivity Bacteria Summary

Confidential & Restricted

{25}------------------------------------------------

Image /page/25/Picture/0 description: The image shows the logo for Luminex. The word "Luminex" is written in a bold, sans-serif font, with the letters in black. There is a red dot above the "i" in "Luminex". The logo is simple and modern.

LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission

Table 10. LIAISON PLEX® BCY Assay Analytical Specificity – Cross Reactivity Yeast Summary

OrganismConcentrationTested (CFU/mL)Source & SourceIDPositivity % (Organism)
Aspergillus fumigatus1.0E+07ATCC 10220% (0/3)
Acremonium kiliense1.0E+07ATCC 203370% (0/3)
Candida bracarensis1.0E+07JH 470% (0/3)
Candida carpophila1.0E+07CBS-KNAW 52560% (0/3)
Candida inconspicua1.0E+07NRRL Y-20290% (0/3)
Candida intermedia1.0E+07ATCC 144390% (0/3)
Candida metapsilosis1.0E+07NRRL Y-484700% (0/3)
Candida nivariensis1.0E+07CBS-KNAW 99830% (0/3)
Candida orthopsilosis1.0E+07ATCC 205040% (0/3)
Candida pseudohaemulonii1.0E+07CBS-KNAW 10004100% (C. haemulonii) (9/9)11% (C. lusitaniae) (1/9)
Candida pseudohaemulonii1.0E+06CBS-KNAW 10004100% (C. haemulonii) (3/3)
Candida pseudohaemulonii1.0E+05CBS-KNAW 10004100% (C. haemulonii) (3/3)
Candida pseudohaemulonii1.0E+04CBS-KNAW 10004100% (C. haemulonii) (3/3)
Candida pseudohaemulonii1.0E+03CBS-KNAW 1000433% (C. haemulonii) (1/3)
Candida pseudohaemulonii1.0E+02CBS-KNAW 100040% (0/3)
Candida rugosa1.0E+07NRRL Y-950% (0/3)
Candida sake1.0E+07NRRL Y-16220% (0/3)
Candida solani1.0E+07NRRL Y-22240% (0/3)
Candida utilis1.0E+07ATCC 99500% (0/3)
Exophiala lecanii-corni1.0E+07ATCC 127340% (0/3)
Filobasidium elegans1.0E+07NRRL Y-64860% (0/3)
Filobasidium globisporum1.0E+07NRRL Y-178280% (0/3)
Kluyveromyces lactis1.0E+07ATCC 242070% (0/3)
Kodamaea ohmeri1.0E+07NRRL Y-19320% (0/3)
Meyerozyma caribbica1.0E+07NRRL Y-272740% (0/3)
Malassezia furfur1.0E+07ATCC 145210% (0/3)
Malassezia globosa1.0E+07ATCC-MYA 46120% (0/3)
Metschnikowia pulcherrima1.0E+07ATCC 220320% (0/3)
Malassezia restricta1.0E+07ATCC-MYA 46110% (0/3)
Malassezia sympodialis1.0E+07ATCC 968030% (0/3)
Mucor velutinosus1.0E+07ATCC-MYA 47660% (0/3)
Pichia fermentans1.0E+07NRRL Y-16190% (0/3)
Pichia norvegensis1.0E+07NRRL YB-39040% (0/3)

Confidential & Restricted

{26}------------------------------------------------

Image /page/26/Picture/0 description: The image shows the logo for Luminex. The word "Luminex" is written in a bold, sans-serif font, with the letters in black. There is a red dot above the "i" in Luminex. The logo has a registered trademark symbol next to the "x".

complexity simplified.

OrganismConcentrationTested (CFU/mL)Source & SourceIDPositivity % (Organism)
Rhodotorula mucilaqinosa1.0E+07ATCC 660340% (0/3)
Saccharomyces cerevisiae1.0E+07ATCC 188240% (0/3)
Scedosporium prolificans1.0E+07ATCC 649130% (0/3)
Sporidiobolus salmonicolor1.0E+07NRRL Y-54830% (0/3)
Sporothrix schenckii1.0E+07ATCC 582510% (0/3)
Trichosporon asteroides1.0E+07CBS-KNAW 618350% (C. glabrata) (3/6)33% (C. haemulonii) (2/6)
Talaromyces marneffei1.45E+06spores/mLATCC 182240% (0/3)
Wickerhamomyces anomalus1.0E+07ATCC 102620% (0/3)
Yarrowia deformans1.0E+07NRRL Y-63659100% (C. lipolytica) (9/9)
Yarrowia deformans1.0E+06NRRL Y-63659100% (C. lipolytica) (3/3)
Yarrowia deformans1.0E+05NRRL Y-63659100% (C. lipolytica) (3/3)
Yarrowia deformans1.0E+04NRRL Y-636590% (0/3)

In silico cross-reactivity

In silico exclusivity assessment was performed using sequences available in the GenBank nt database as of December 28, 2023. Based on analysis of the sequences available in the GenBank nt database, the following potential cross-reactivity scenarios are predicted:

  • The Candida lipolytica oligo designs are predicted to detect Yarrowia deformans strains ● resulting in false positive results.
  • . The Candida haemulonii/duobushamulonii oligo designs are predicted to detect Candida pseudohaemulonii strains resulting in false positive results.

The potential cross-reactive organisms assessed for in silico exclusivity analysis are provided below in Table 11.

Off-Panel Organisms
On-Panel OrganismsBacteriaFungi
Candida albicansAcinetobacter baumaniiAcremonium kiliense
Candida aurisAcinetobacter IwoffiiAspergillus fumigatus
Candida dubliniensisAerococcus viridansCandida bracarensis
Candida duobushamuloniiBacillus cereusCandida carpophila
Candida famataBacteroides fragilisCandida inconspicua
Candida glabrataBordetella pertussisCandida intermedia
Candida guilliermondiiCitrobacter freundiiCandida lambica
Candida haemuloniiClostridium perfringensCandida metapsilosis
Candida kefyrCorynebacterium striatumCandida nivariensis
Candida kruseiCutibacterium (Propionibacterium) acnesCandida norvegensis
Candida lipolyticaEggerthella lentaCandida orthopsilosis

Table 11. Potential Cross-Reactive Organisms assessed in the In Silico Exclusivity Analysis

{27}------------------------------------------------

Image /page/27/Picture/0 description: The image shows the logo for Luminex. The word "Luminex" is written in a bold, sans-serif font, with the letters in black. There is a red dot above the "i" in Luminex. The logo is simple and modern.

complexity simplif

LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission

Off-Panel Organisms
On-Panel OrganismsBacteriaFungi
Candida lusitaniaeEnterobacter aerogenesCandida pelliculosa
Candida parapsilosisEnterobacter cloacaeCandida pseudohaemulonii
Candida tropicalisEnterococcus aviumCandida sake
Cryptococcus gattiiEnterococcus faecalisCandida utilis
Cryptococcus neoformansEnterococcus faeciumDiutina (Candida) rugosa
Escherichia coliExophiala jeanselmei
Klebsiella oxytocaFilobasidium elegans
Klebsiella pneumoniaeFilobasidium globisporum
Lactobacillus rhamnosusFusarium solani
Listeria monocytogenesKluyveromyces lactis
Micrococcus luteusKodamaea ohmeri
Morganella morganiiMalassezia furfur
Prevotella denticolaMalassezia globose
Proteus mirabilisMalassezia restricta
Pseudomonas aeruginosaMalassezia sympodialis
Salmonella enterica (Typhi)Metschnikowia (Candida) pulcherrima
Serratia marcescensMeyerozyma caribbica (Candida fermentati)
Staphylococcus aureusMucor velutinosus
Staphylococcus epidermidisPenicillium marneffei
Staphylococcus hominisRhodotorula mucilaginosa
Staphylococcus intermediusSaccharomyces cerevisiae
Staphylococcus lugdunensisScedosporium prolificans
Staphylococcus saprophyticusSporidiobolus salmonicolor
Stenotrophomonas maltophiliaSporothrix schenckii
Streptococcus agalactiaeTrichosporon spp.
Streptococcus anginosusYarrowia deformans
Streptococcus mitisProtozoa
Streptococcus pneumoniaePlasmodium falciparum
Streptococcus pyogenesTrypanosoma cruzi

h. Interference

Microbial Interference

The impact of nine potentially interfering microbial organisms commonly found in blood culture samples were tested in the presence of on-panel organisms. C. glabrata, and C. parapsilosis were selected as the three representative on-panel organisms used throughout this study. Each potentially interfering organism was tested at a high concentration of ≥ 1.00+E08 CFU/mL for bacteria, in the presence of the on-panel organisms which were tested at low concentrations (ring positive), which were determined during the Growth and Detection Study. Table 12 summarizes the results from the Microbial Interference studies.

{28}------------------------------------------------

Image /page/28/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. The word is black, except for a red dot above the "i". There is a registered trademark symbol to the right of the "x".

LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission Table 12. LIAISON PLEX® BCY Microbial Interference Summary

Table 12. LIAISON PLEX® BCY Microbial Interference Summary
Off-Panel High ConcentrationInterfering MicrobeOn-Panel LowConcentration TargetTargetPositivity
Escherichia coli(1.0E+08 CFU/mL)Candida albicans(4.4E+05 CFU/mL)100% (3/3)
Candida glabrata(3.8E+06 CFU/mL)100% (3/3)
Candida parapsilosis(1.5E+07 CFU/mL)100% (3/3)
Staphylococcus aureus(1.0E+08 CFU/mL)Candida albicans(4.4E+05 CFU/mL)100% (3/3)
Candida glabrata(3.8E+06 CFU/mL)100% (3/3)
Candida parapsilosis(1.5E+07 CFU/mL)100% (3/3)
Klebsiella pneumoniae(1.0E+08 CFU/mL)Candida albicans(4.4E+05 CFU/mL)100% (3/3)
Candida glabrata(3.8E+06 CFU/mL)100% (3/3)
Candida parapsilosis(1.5E+07 CFU/mL)100% (3/3)
Pseudomonas aeruginosa(1.0E+08 CFU/mL)Candida albicans(4.4E+05 CFU/mL)100% (3/3)
Candida glabrata(3.8E+06 CFU/mL)100% (3/3)
Candida parapsilosis(1.5E+07 CFU/mL)100% (3/3)
Enterococcus faecalis(1.0E+08 CFU/mL)Candida albicans(4.4E+05 CFU/mL)100% (3/3)
Candida glabrata(3.8E+06 CFU/mL)100% (3/3)
Candida parapsilosis(1.5E+07 CFU/mL)100% (3/3)
Staphylococcus epidermidis(1.0E+08 CFU/mL)Candida albicans(4.4E+05 CFU/mL)100% (3/3)
Candida glabrata(3.8E+06 CFU/mL)100% (3/3)
Candida parapsilosis(1.5E+07 CFU/mL)100% (3/3)
Enterococcus faecium(1.0E+08 CFU/mL)Candida albicans(4.4E+05 CFU/mL)100% (3/3)
Candida glabrata(3.8E+06 CFU/mL)100% (3/3)
Candida parapsilosis(1.5E+07 CFU/mL)100% (3/3)
Off-Panel High ConcentrationInterfering MicrobeOn-Panel LowConcentration TargetTargetPositivity
(1.0E+08 CFU/mL)(4.4E+05 CFU/mL)
(1.0E+08 CFU/mL)Candida glabrata(3.8E+06 CFU/mL)100% (3/3)
(1.0E+08 CFU/mL)Candida parapsilosis(1.5E+07 CFU/mL)100% (3/3)
Streptococcus pneumoniae(1.0E+08 CFU/mL)Candida albicans(4.4E+05 CFU/mL)100% (3/3)
Streptococcus pneumoniae(1.0E+08 CFU/mL)Candida glabrata(3.8E+06 CFU/mL)100% (3/3)
Streptococcus pneumoniae(1.0E+08 CFU/mL)Candida parapsilosis(1.5E+07 CFU/mL)100% (3/3)

{29}------------------------------------------------

Image /page/29/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. The word is black, except for a red dot above the "i". A small registered trademark symbol is to the right of the "x".

LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission

{30}------------------------------------------------

Image /page/30/Picture/0 description: The image shows the logo for Luminex. The logo is black text with a red dot above the i. The text is bold and slightly slanted to the right. There is a registered trademark symbol next to the x.

LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission

Interfering Substances:

An Interfering Substance study was performed to evaluate non-microbial interference of 18 substances which may be present during collection or within clinical blood samples using the LIAISON PLEX® BCY Assay. Two positive panels containing three (3) on-panel organisms each (C. albicans, C. tropicalis, C. neoformans and C. glabrata, C. guilliermondii, and C. kefyr) and negative blood matrix were used to test each interfering substance at clinically relevant concentrations. The potentially interfering substances and their testing concentrations are listed in Table 15, the substances do not interfere with the LIAISON PLEX BCY Assay. All positive panel results were 100% positive. However, Fluconazole at the testing concentration of 25 mg/L showed interference and resulted in a false negative for C. albicans.At the testing concentrations listed in Table 13, the substances do not interfere with the LIAISON PLEX BCY Assay. All positive panel results were 100% positive. However, Fluconazole at the testing concentration of 25 mg/L showed interference and resulted in a false negative for C. albicans.

Interfering SubstanceInterfering SubstanceConcentration
Unconjugated Bilirubin20 mg/dL
Conjugated Bilirubin20 mg/dL
Hemoglobin14 g/L
y-globulin6 g/dL
SodiumPolyanetholsulfonate0.25% w/v
Amoxicillin clavulanate3.5 µg/mL
Amphotericin B2 µg/mL
Caspofungin5 µg/mL
Ceftriaxone0.23 mg/mL
Table 13 - List of Potentially Interfering Substances
Interfering SubstanceInterfering Substance Concentration
Ciprofloxacin3 mg/L
Fluconazole25 mg/L
8.3 mg/L
Flucytosine90 µg/mL
Gentamicin sulfate3 µg/mL
Heparin0.9 U/mL
Imipenem83 µg/mL
Tetracycline5 mg/L
Vancomycin30 mg/L
Intralipid/Triglycerides3000 mg/dL
ControlsN/A
0.9% NaCl Solvent BlankN/A

{31}------------------------------------------------

Image /page/31/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. The letters are black, except for a red dot above the "i". The word appears to be a logo for a company or product.

LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission

{32}------------------------------------------------

Image /page/32/Picture/0 description: The image shows the logo for Luminex. The word "Luminex" is written in a bold, black font. Above the "i" in Luminex is a red dot. Below the logo is the phrase "complexity simplified."

Competitive Interference/Co-infection:

Competitive inhibition of the LIAISON PLEX® BCY was assessed by testing six pairings of clinically prevalent co-infections. A single pairing consisted of one target at a high concentration with another target at a low concentration targets were tested at a concentration representative of Ring Positivity, as determined during the Growth and Detection study. All high concentration targets were tested at a concentration representative of Ring Positivity + 8 hours, as determined during the Growth and Detection study. Testing of each combination was performed in triplicate.

On-Panel HighConcentration TargetHigh ConcentrationTarget PositivityOn-Panel LowConcentration TargetLow ConcentrationTarget Positivity
Candida parapsilosis(5.5E+07 CFU/mL)100% (3/3)Candida albicans(4.4E+05 CFU/mL)100% (3/3)
Candida parapsilosis(5.5E+07 CFU/mL)100% (3/3)Candida glabrata(3.8E+06 CFU/mL)100% (3/3)
Candida glabrata(1.6E+08 CFU/mL)100% (3/3)Candida parapsilosis(1.5E+07 CFU/mL)100% (3/3)
Candida glabrata(1.6E+08 CFU/mL)100% (3/3)Candida albicans(4.4E+05 CFU/mL)100% (3/3)
Candida albicans(1.5E+07 CFU/mL)100% (3/3)Candida glabrata(3.8E+06 CFU/mL)100% (3/3)
Candida albicans(1.5E+07 CFU/mL)100% (3/3)Candida parapsilosis(1.5E+07 CFU/mL)100% (3/3)

Table 14. LIAISON PLEX® BCY Co-Infection Summary

Carry-Over/Cross-Contamination:

Carry-over and cross contamination for the LIAISON PLEX® BCY Assay was evaluated by testing positive and negative samples in an alternating series. Two operators tested 30 high concentration C. lusitaniae positive samples and 30 negative samples across two instruments over the course of two days. The samples were loaded into cartridges alternating between positive and negative samples, six cartridges at a time using sample prep trays. The six cartridges were loaded into a LIAISON PLEX® instrument in the same order they were loaded with sample. The instruments set-ups made it so the sample loaded and ran on each blade alternated between positive samples and negative samples for all runs executed in this study. The results demonstrated 100% agreement between expected and observed results, indicating that no carry-over or cross contamination occurred between and during runs.

i. Assay Cut-off

The specific assay parameters for the LIAISON PLEX® BCY Assay are considered confidential and proprietary.

{33}------------------------------------------------

Image /page/33/Picture/0 description: The image shows the logo for Luminex. The word "Luminex" is written in a bold, sans-serif font. There is a red dot above the "i" in Luminex. The logo also includes the registered trademark symbol.

2. Comparison Studies:

  • Method comparison with predicate device: a.
    Refer to Section 3 Clinical Performance.

  • b. Matrix Comparison: Testing of Blood Culture Bottle Types / Matrix Equivalency
    The LIAISON PLEX " BCY Assay was evaluated to confirm equivalent performance for target detection in twelve different blood bottle matrixes from three different blood culture systems (BacT/Alert, BACTEC, and VersaTREK). The matrix was tested with negative blood matrix, two positive control panels, and with representative bacteria strains. Positive Control Panel 1 was comprised of Candida albicans, Cryptococcus neoformans, and Candida tropicalis. Positive Control Panel 2 was comprised of Candida glabrata, Candida guilliermondii, and Candida kefyr. The Positive Control Panels were tested at concentration representing blood culture bottle positivity, as determined during the Growth and Detection study. The bacteria strains used to test the assay were Acinetobacter baumannii, Escherichia coli, Klebsiella pneumonia, Straphylococcus aureus, and Pseudomonas aeruginosa. Bacteria strains were all tested at the concentration of 1.00E+08 CFU/mL. A negative sample, Negative Blood Matrix, was also tested. A summary of blood bottle equivalency results is explained in Table 15.

ManufacturerBottleManufacturerPart NumberBottleNameBottle LotPercent Positivity
NBMPositiveControl11PositiveControl22OffPanelTargets3
ThermoScientific7106-44VersaTrekRedox 15938150%(0/5)100%(5/5)100%(5/5)0%(0/10)
5938170%(0/5)100%(5/5)100%(5/5)0%(0/10)
ThermoScientific7107-44VersaTrekRedox 25764190%(0/5)100%(5/5)100%(5/5)0%(0/10)
6268230%(0/5)100%(5/5)100%(5/5)0%(0/10)
Becton,Dickinson,and Company(BD)442022Bactec PlusAnaerobic/F30104060%(0/5)100%(5/5)100%(5/5)0%(0/10)
30411610%(0/5)100%(5/5)100%(5/5)0%(0/10)

4 This percent positivity reflects the percent positive control 1, which includes Candida albicans, Candida tropicalis, and Cryptococcus neoformans.

2 This percent positivity reflects the percent positive control 2, which includes Condida globrato, Candida guilliermondii, and Candida kefyr.

3 This percent positivity reflects the percent positivity of all replicates from the 5 bacteria strains tested.

{34}------------------------------------------------

Image /page/34/Picture/0 description: The image shows the word "Luminex" in a bold, sans-serif font. The letters are black, except for a red dot above the "i". There is a registered trademark symbol to the right of the "x".

LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission

ManufacturerBottleManufacturerPart NumberBottleNameBottle LotPercent Positivity
NBMPositiveControl11PositiveControl22OffPanelTargets3
442024Bactec PlusStandardAnaerobic/F21869920%(0/5)100%(5/5)100%(5/5)0%(0/10)
Bactec PlusStandardAnaerobic/F232655220%(1/5)100%(5/5)100%(5/5)0%(0/10)
442021BactecLytic/10Anaerobic/F22278610%(0/5)100%(5/5)100%(5/5)0%(0/10)
BactecLytic/10Anaerobic/F30460480%(0/5)100%(5/5)100%(5/5)0%(0/10)
442023Bactec PlusAerobic/F21787100%(0/5)100%(5/5)100%(5/5)0%(0/10)
Bactec PlusAerobic/F30418790%(0/5)100%(5/5)100%(5/5)0%(0/10)
442020Bactec PedsPlus22632550%(0/5)100%(5/5)100%(5/5)0%(0/10)
Bactec PedsPlus30675760%(0/5)100%(5/5)100%(5/5)0%(0/10)
442027BactecStandard/10Aerobic22279330%(0/5)100%(5/5)100%(5/5)0%(0/10)
BactecStandard/10Aerobic30608930%(0/5)100%(5/5)100%(5/5)0%(0/10)
BioMérieux259789BACT/AlertSA00010594000%(0/5)100%(5/5)100%(5/5)0%(0/10)
00010597080%(0/5)100%(5/5)100%(5/5)0%(0/10)
410852BACT/AlertFN Plus00040591600%(0/5)100%(5/5)100%(5/5)0%(0/10)
00040598440%(0/5)100%(5/5)100%(5/5)0%(0/10)
259790BACT/AlertSN00010598600%(0/5)100%(5/5)100%(5/5)0%(0/10)
00010599000%(0/5)100%(5/5)100%(5/5)0%(0/10)
410853BACT/AlertPE Plus00041013000%(0/5)100%(5/5)100%(5/5)0%(0/10)

{35}------------------------------------------------

Image /page/35/Picture/0 description: The image shows the logo for Luminex. The word "Luminex" is written in a bold, sans-serif font, with the letters in black. There is a red dot above the "i" in Luminex. A small registered trademark symbol is next to the "x".

complexity simplified.

ManufacturerBottleManufacturerPart NumberBottleNameBottle LotPercent Positivity
NBMPositiveControl11PositiveControl22OffPanelTargets3
00041015320%(0/5)100%(5/5)100%(5/5)0%(0/10)

3. Clinical Performance:

A multi-site clinical study established the diagnostic accuracy of the LIAISON PLEX® BCY Assay for the detection and identification of multiple potentially pathogenic fungal organisms in positive blood culture from blood culture media identified as positive by a continuous monitoring blood culture system and which contain fungal organisms as determined by Gram Stain. The clinical performance of the LIAISON PLEX BCY Assay was evaluated using a combination of prospectively collected de-identified remnant samples, preselected positive samples, and contrived samples.

A total of 3447 unique prospectively collected specimens were evaluated for their eligibility across four geographically diverse US clinical sites. Of these, 69 (1 specimen was excluded due to inconclusive gram stain result) prospective specimens that met the pre-determined inclusion criteria and exhibited yeast morphology following gram stain were enrolled in the study. Clinical runs and re-runs using the LIAISON PLEX® BCY Assay were tested on the LIAISON PLEX® System by trained operators at four sites. Prospective specimen testing occurred between June 2023 and October 2023.

As all targets exhibited low prevalence rates in the prospective specime specimen set was supplemented with pre-selected and contrived specimens. The 63 pre-selected specimens were remnant, de-identified specimens sourced from 6 different sites/vendors in the United States. The pre-selected specimens were characterized by an FDA cleared molecular assay prior to enrollment in the study. The pre-selected specimens were tested in a randomized, blinded manner with negative specimens at one site during October 2023.

Out of the 132 clinical specimens included in the prospective and pre-selected study analysis, 130 (98.5%) generated valid BCY Assay results (i.e., Detected or Not Detected) on the first attempt. Two (1.5%) specimens with initial invalid results generated valid BCY results after a single retest. 132 specimens generated valid BCY results for a final success rate of 100.0% (132/132).

Table 16 provides a summary of the general demographic information of the 69 prospectively collected specimens that were included in the clinical study.

{36}------------------------------------------------

Image /page/36/Picture/0 description: The image shows the logo for Luminex. The word "Luminex" is written in a bold, sans-serif font, with the "i" dotted by a red circle. The logo is simple and modern, and the red dot adds a pop of color. The logo also has a registered trademark symbol.

LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission Table 16 - Demographic Information for Prospectively collected specimens

GenderSite 01Site 02Site 03Site 04All Sites
Male24(57.1%)6(54.5%)5(45.5%)1(20.0%)36(52.2%)
Female17(40.5%)5(45.5%)6(54.5%)4(80.0%)32(46.4%)
Gender Unknown1(2.4%)0(0.0%)0(0.0%)0(0.0%)1(1.4%)
Total42(100.0%)11(100.0%)11(100.0%)5(100.0%)69(100.0%)
Age (years)
0-12(4.8%)0(0.0%)0(0.0%)1(20.0%)3(4.3%)
>1-50(0.0%)0(0.0%)0(0.0%)0(0.0%)0(0.0%)
>5-210(0.0%)1(9.1%)0(0.0%)0(0.0%)1(1.4%)
>21-6521(50.0%)9(81.8%)8(72.7%)3(60.0%)41(59.4%)
>6518(42.9%)1(9.1%)3(27.3%)1(20.0%)23(33.3%)
Age Unknown1(2.4%)0(0.0%)0(0.0%)0(0.0%)1(1.4%)
Total42(100.0%)11(100.0%)11(100.0%)5(100.0%)69(100.0%)
Subject Status
Emergency Room0(0.0%)1(9.1%)4(36.4%)0(0.0%)5(7.2%)
Hospitalized0(0.0%)10(90.9%)7(63.6%)5(100.0%)22(31.9%)
Status Unknown42(100.0%)0(0.0%)0(0.0%)0(0.0%)42(60.9%)
Total42(100.0%)11(100.0%)11(100.0%)5(100.0%)69(100.0%)
Blood CultureBottle type
BD BACTEC LyticAnaerobic0(0.0%)3(27.3%)0(0.0%)0(0.0%)3(4.3%)
BD BACTEC PlusAerobic0(0.0%)8(72.7%)6(54.5%)0(0.0%)14(20.3%)
BD BACTECStandard Aerobic0(0.0%)0(0.0%)5(45.5%)0(0.0%)5(7.2%)
BacT/ALERT FAPlus40(95.2%)0(0.0%)0(0.0%)0(0.0%)40(58.0%)
BacT/ALERT FNPlus1(2.4%)0(0.0%)0(0.0%)0(0.0%)1(1.4%)
BacT/ALERT PFPlus1(2.4%)0(0.0%)0(0.0%)1(20.0%)2(2.9%)
BacT/ALERT SAStandard Aerobic0(0.0%)0(0.0%)0(0.0%)3(60.0%)3(4.3%)
BacT/ALERT SNStandard Anaerobic0(0.0%)0(0.0%)0(0.0%)1(20.0%)1(1.4%)
Bottle Type Total42(100.0%)11(100.0%)11(100.0%)5(100.0%)69(100.0%)

Table 17 provides a summary of the general demographic information of the 63 pre-selected collected specimens that were included in the clinical study.

{37}------------------------------------------------

Image /page/37/Picture/0 description: The image shows the word "Luminex" in bold, black font. There is a red dot above the "i" in "Luminex". A registered trademark symbol is located to the right of the "x" in "Luminex".

complexity simplified.

Table 17 - Demographic Information for Pre-Selected Clinical Specimens
# Samples (%)
Gender
Male24(38.1%)
Female25(39.7%)
Gender Unknown14(22.2%)
Total63(100.0%)
Age (years)
0-10(0.0%)
>1-51(1.6%)
>5-210(0.0%)
>21-6527(42.9%)
>6519(30.2%)
Age Unknown16(25.4%)
Total63(100.0%)
Subject Status
Hospitalized2(3.2%)
Status Unknown61(96.8%)
Total63(100.0%)
Blood Culture Bottle Type
BD BACTEC Lytic Anaerobic2(3.2%)
BD BACTEC Plus Aerobic17(27.0%)
BD BACTEC Standard Aerobic29(46.0%)
Bottle Type Unknown15(23.8%)
Total63(100.0%)

A total of 829 specimens were contrived and tested. To minimize bias, contrived specimens were blinded, randomized, and tested along with negative clinical specimens at all four testing sites during June 2023. Results from contrived specimens were analyzed separately from the prospective and pre-selected data sets.

Out of the 829 specimens included in the contrived study analysis, 803 specimens (96.9%) generated valid BCY Assay results (i.e., Detected or Not Detected) on the first attempt. There were 26 specimens (3.1%) with an invalid result on the initial run. Of the 26 specimens retested, 24 specimens generated a valid result after a single retest for a final success rate of 99.7% (827/829).

The invalid rate for prospective, pre-selected, and contrived specimens combined was 2.9% (28/961) after the initial run. After allowing for a single retest of any initial invalid results, 26 (2.7%) specimens generated valid BCY results. Two (0.2%) specimens remained invalid on repeat for an overall success rate of 99.8% (959/961).

For each target in the BCY Assay Panel, the diagnostic performance of the BCY Assay was determined using Sensitivity/PPA and Specificity/NPA, along with the associated 95% confidence intervals, (95% Cl as compared to the reference method). The results of the combined prospective and pre-selected specimen analysis are summarized in Table 18. The performance of the BCY Assay for contrived specimens is presented separately in Table 19.

{38}------------------------------------------------

Image /page/38/Picture/0 description: The image shows the Luminex logo. The word "Luminex" is in bold, black font, with a red dot above the "i". Below the logo is the tagline "complexity simplified." in a smaller, black font with a red dot at the end.

Sensitivity / PPA1Specificity / NPA2
Pathogen TargetTP /(TP+FN)Sensitivity/ PPA195% CITN /(TN+FP)Specificity/ NPA295% CI
Prospective17/17100.0%81.6%-100.0%51/5298.1%89.9%-99.7%
Candida albicansPre-selected17/17100.0%81.6%-100.0%46/46100.0%92.3%-100.0%
Combined34/34100.0%89.8%-100.0%97/9899.0%94.4%-99.8%
Prospective4/4100.0%51.0%-100.0%65/65100.0%94.4%-100.0%
Candida aurisPre-selected0/0NANA63/63100.0%94.3%-100.0%
Combined4/4100.0%51.0%-100.0%128/128100.0%97.1%-100.0%
Prospective0/0NANA69/69100.0%94.7%-100.0%
CandidadubliniensisPre-selected0/0NANA63/63100.0%94.3%-100.0%
Combined0/0NANA132/132100.0%97.2%-100.0%
Prospective0/0NANA69/69100.0%94.7%-100.0%
Candida famataPre-selected0/0NANA63/63100.0%94.3%-100.0%
Combined0/0NANA132/132100.0%97.2%-100.0%
Prospective25/25100.0%86.7%-100.0%44/44100.0%92.0%-100.0%
Candida glabrataPre-selected21/21100.0%84.5%-100.0%42/42100.0%91.6%-100.0%
Combined46/46100.0%92.3%-100.0%86/86100.0%95.7%-100.0%
Prospective0/0NANA69/69100.0%94.7%-100.0%
CandidaguilliermondiiPre-selected0/0NANA63/63100.0%94.3%-100.0%
Combined0/0NANA132/132100.0%97.2%-100.0%
Candidahaemulonii/C.duobushaemuloniiProspective0/0NANA69/69100.0%94.7%-100.0%
Pre-selected0/0NANA63/63100.0%94.3%-100.0%
Combined0/0NANA132/132100.0%97.2%-100.0%
Prospective1/1100.0%20.7%-100.0%68/68100.0%94.7%-100.0%
Candida kefyrPre-selected0/0NANA63/63100.0%94.3%-100.0%
LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission
Sensitivity / PPA1Specificity / NPA2
Pathogen TargetTP /(TP+FN)Sensitivity/ PPA195% CITN /(TN+FP)Specificity/ NPA295% CI
Combined1/1100.0%20.7%-100.0%131/131100.0%97.2%-100.0%
Prospective3/3100.0%43.9%-100.0%66/66100.0%94.5%-100.0%
Candida kruseiPre-selected1/1100.0%20.7%-100.0%62/62100.0%94.2%-100.0%
Combined4/4100.0%51.0%-100.0%128/128100.0%97.1%-100.0%
Prospective0/0NANA69/69100.0%94.7%-100.0%
Candida lipolyticaPre-selected0/0NANA63/63100.0%94.3%-100.0%
Combined0/0NANA132/132100.0%97.2%-100.0%
CandidalusitaniaeProspective2/2100.0%34.2%-100.0%67/67100.0%94.6%-100.0%
Pre-selected0/0NANA63/63100.0%94.3%-100.0%
Combined2/2100.0%34.2%-100.0%130/130100.0%97.1%-100.0%
Prospective11/11100.0%74.1%-100.0%57/5898.3%90.9%-99.7%
CandidaparapsilosisPre-selected6/6100.0%61.0%-100.0%57/57100.0%93.7%-100.0%
Combined17/17100.0%81.6%-100.0%114/115499.1%95.2%-99.8%
CandidatropicalisProspective6/6100.0%61.0%-100.0%60/6395.2%86.9%-98.4%
Pre-selected0/0NANA63/63100.0%94.3%-100.0%
Combined6/6100.0%61.0%-100.0%123/126597.6%93.2%-99.2%
Prospective0/0NANA69/69100.0%94.7%-100.0%
Cryptococcusneoformans/Cryptococcus gattiiPre-selected5/5100.0%56.6%-100.0%58/58100.0%93.8%-100.0%
Combined5/5100.0%56.6%-100.0%127/127100.0%97.1%-100.0%

Table 18. LIAISON PLEX BCY Assay Performance with Prospective and Pre-Selected Specimens

{39}------------------------------------------------

Image /page/39/Picture/0 description: The image shows the logo for Luminex. The word "Luminex" is written in a bold, sans-serif font. There is a red dot above the "i" in Luminex. The logo is simple and modern.

1 Sensitivity is designated for prospective sample testing, while PPA is designated for pre-selected sample testing. 2 Specificity is designated for prospective sample testing, while NPA is designated for pre-selected sample testing.

3The one Candida albicans False Positive was positive by a cleared molecular assay.

4The one Candida parapsilosis False Positive was positive by a cleared molecular assay.

5One out of three Candida tropicalis False Positive by a cleared molecular assay.

{40}------------------------------------------------

Image /page/40/Picture/0 description: The image shows the logo for Luminex. The word "Luminex" is written in a bold, sans-serif font, with the letters in black. There is a red dot above the "i" in Luminex. The logo is simple and modern.

LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission Table 19 – LIAISON PLEX BCY Assay Performance with Contrived Specimens

Pathogen TargetPositive Percent AgreementNegative Percent Agreement
AnalyteTP / (TP+FN)PPA95% CITN / (TN+FP)NPA95% CI
Candida albicans50/50100.0%92.9%- 100.0%777/777100.0%99.5%- 100.0%
Candida auris50/50100.0%92.9%- 100.0%777/777100.0%99.5%- 100.0%
Candida dubliniensis50/50100.0%92.9%- 100.0%777/777100.0%99.5%- 100.0%
Candida famata50/50100.0%92.9%- 100.0%776/777199.9%99.3%- 100.0%
Candida glabrata49/49100.0%92.7%- 100.0%778/778100.0%99.5%- 100.0%
Candida guilliermondii50/50100.0%92.9%- 100.0%777/777100.0%99.5%- 100.0%
Candida haemulonii/C. duobushaemulonii50/50100.0%92.9%- 100.0%777/777100.0%99.5%- 100.0%
Candida kefyr50/50100.0%92.9%- 100.0%777/777100.0%99.5%- 100.0%
Candida krusei50/50100.0%92.9%- 100.0%777/777100.0%99.5%- 100.0%
Candida lipolytica50/50100.0%92.9%- 100.0%777/777100.0%99.5%- 100.0%
Candida lusitaniae50/50100.0%92.9%- 100.0%777/777100.0%99.5%- 100.0%
Candida parapsilosis50/50100.0%92.9%- 100.0%776/777299.9%99.3%- 99.9%
Candida tropicalis49/49100.0%92.7%- 100.0%775/778399.6%98.9%- 100.0%
Cryptococcus neoformans/Cryptococcus gattii100/100100.0%96.3%- 100.0%727/727100.0%99.5%- 100.0%

The clinical performance of the LIAISON PLEX® BCY Assay was compared to the standard of care culture followed by identification by Matrix Assisted Laser Desorption/lonization coupled to time-of-flight Mass Spectrometry (MALDI-TOF MS) for all targets, as summarized below in Table 20.

{41}------------------------------------------------

Image /page/41/Picture/0 description: The image shows the logo for Luminex. The word "Luminex" is written in a bold, sans-serif font, with the letters slightly slanted to the right. There is a red dot above the "i" in "Luminex". The logo is simple and modern.

LIAISON PLEX® Yeast Blood Culture Assay Traditional 510(k) Submission

LIAISON PLEX® BCY Assay TargetComparator Method
Candida albicansStandard of Care culture followed byidentification by MALDI-TOF MS
Candida auris
Candida dubliniensis
Candida famata
Candida glabrata
Candida guilliermondii
Candida kefyr
Candida krusei
Candida lipolytica
Candida lusitaniae
Candida parapsilosis
Candida tropicalis
Candida haemulonii/duobushaemulonii (1)
Cryptococcus neoformans / gattii

Table 20 - Reference Method for the LIAISON PLEX BCY Assay Clinical Study

The pre-defined acceptance criteria for the clinical study were defined as follows:

  • The assay shall achieve a target Sensitivity of ≥ 90% for all targets. .
  • The Specificity for each target should be established at a level of ≥95%. .
  • Failure rate shall be ≤15%. .

Acceptance Criteria are based on Arms 1, 2, and 3 combined. All acceptance criteria were met successfully.

The study results demonstrate that the diagnostic accuracy of the BCY Assay is acceptable for the safe and effective detection and identification of fungal pathogens from blood culture media identified as positive by a continuous monitoring blood culture system and which contain fungal organisms as determined by Gram Stain from patients exhibiting clinical signs and symptoms of bloodstream infection.

{42}------------------------------------------------

Image /page/42/Picture/0 description: The image shows the Luminex logo. The word "Luminex" is in bold, black font, with a red dot above the "i". Below the logo is the tagline "complexity simplified." in a smaller, lighter font.

M. Proposed Labeling:

The labeling provided in the submission satisfies the requirements of 21 CFR 809.10.

N. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 866.3365 Multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures.

(a)
Identification. A multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures is a qualitative in vitro device intended to simultaneously detect and identify microorganism nucleic acids from blood cultures that test positive by Gram stain or other microbiological stains. The device detects specific nucleic acid sequences for microorganism identification as well as for antimicrobial resistance. This device aids in the diagnosis of bloodstream infections when used in conjunction with other clinical and laboratory findings. However, the device does not replace traditional methods for culture and susceptibility testing.(b)
Classification. Class II (special controls). The special control for this device is FDA's guideline document entitled “Class II Special Controls Guideline: Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures.” For availability of the guideline document, see § 866.1(e).