K Number
K240595

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-08-14

(163 days)

Product Code
Regulation Number
884.1100
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Utepreva Endometrial Sampler (UP01) is used to obtain endometrial cytological and histological samples.

Device Description

The Utepreva Endometrial Sampler (UP01) is a single-use, sterile device designed to obtain endometrial cytological and histological samples. The device is an Rx device that functions as follows: The Utepreva Endometrial Sampler is gently inserted though the cervix into the uterus. The skirt (guard) will reach the cervix indicating to the clinician to stop inserting the device. The brush is then extended by pushing the handle until resistance is felt when the tip meets the fundus of the uterus. The sponge on the tip is intended to minimize the risk of piercing of the fundus by the tip of the brush and enables additional tissue sampling by absorption of the fluid and the cells from the uterus. The Utepreva Endometrial Sampler is rotated in a clockwise or counterclockwise manner for four or five 360° rotations. Upon completion of the rotations, the handle is pulled back so that the brush and tip are withdrawn into the outer sheath. The suction created by the plunger located inside of the sheath provides aspiration of the cells and prevents loss of the sample from the brush, sponge and any fluid that was aspirated can then be immersed into the preservative solution to be used by the pathology lab.

The device materials consist of polydimethylsiloxane, high density polyethylene, cyanoacrylate adhesive, Loctite, Permabond polyolefin primer, stainless steel, Nylon, polycarbonate, and polypropylene. The device is packaged in a Tyvek® pouch. One pouch is then packaged in a unit box. Fifteen unit boxes are packaged within a single master box to create the final shipped product.

AI/ML Overview

I am sorry, but the provided text does not contain the detailed information necessary to answer your request about acceptance criteria and the study that proves the device meets them. The document is a 510(k) premarket notification letter from the FDA, and it focuses on the substantial equivalence of the "Utepreva Endometrial Sampler (UP01)" to a predicate device.

While it mentions non-clinical performance testing (biocompatibility, visual & dimensional inspection, tensile strength, torque, bristle retention, fluid retention, fluid transfer) and states that these tests "passed," it does not:

  1. Provide a table of specific acceptance criteria and reported device performance values. It only states "Pass" for each test.
  2. Detail the sample size used for the test set.
  3. Specify the data provenance (e.g., country of origin, retrospective/prospective).
  4. Mention the number or qualifications of experts used to establish ground truth.
  5. Describe any adjudication method.
  6. Indicate if a multi-reader multi-case (MRMC) comparative effectiveness study was done.
  7. Discuss standalone performance (algorithm only without human-in-the-loop).
  8. Elaborate on the type of ground truth used (expert consensus, pathology, outcomes data, etc.).
  9. Provide the sample size for the training set.
  10. Explain how the ground truth for the training set was established.

The document's purpose is to establish substantial equivalence based on indications for use, technological characteristics, and general performance testing, not to provide a detailed clinical or performance study breakdown with specific metrics and methodologies.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 14, 2024

Utepreva LLC % Melissa Paffenroth Senior Regulatory Consultant MEDIcept, LLC. 200 Homer Avenue Ashland, Massachusetts 01721

Re: K240595

Trade/Device Name: Utepreva Endometrial Sampler (UP01) Regulation Number: 21 CFR 884.1100 Regulation Name: Endometrial Brush Regulatory Class: II Product Code: HFE Dated: March 4, 2024 Received: July 1, 2024

Dear Melissa Paffenroth:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drue, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdth/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Ouality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/trainingand-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory- assistance/contact-usdivision-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jason Roberts -S

Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K240595

Device Name

Utepreva Endometrial Sampler (UP01)

Indications for Use (Describe)

The Utepreva Endometrial Sampler (UP01) is used to obtain endometrial cytological and histological samples.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary Utepreva, LLC Utepreva Endometrial Sampler (UP01)

I. SUBMITTER

510(k) Holder:Utepreva, LLC100 Jericho QuadrangleSuite 333Jericho, NY 11753
Contact Person:Melissa Paffenroth, PhD, RAC-DevicesSenior Consultant, RegulatoryMEDIcept, Inc.Tel: (617) 512-9943Email: mpaffenroth@medicept.com

Date Prepared: March 4, 2024

II. DEVICE

Name of Device:Utepreva Endometrial Sampler (UP01)
Common Name:Endometrial brush
Classification Name:Brush, Endometrial
Regulation Number:21 CFR 884.1100
Regulatory Class:Class II
Product Code:HFE

III. PREDICATE DEVICE

K170603, Tao BrushTM I.U.M.C Endometrial

The predicate device identified above has not been subject to design-related recalls.

IV. DEVICE DESCRIPTION

The Utepreva Endometrial Sampler (UP01) is a single-use, sterile device designed to obtain endometrial cytological and histological samples. The device is an Rx device that functions as follows: The Utepreva Endometrial Sampler is gently inserted though the cervix into the uterus. The skirt (guard) will reach the cervix indicating to the clinician to stop inserting the device. The brush is then extended by pushing the handle until resistance is felt when the tip meets the fundus of the uterus. The sponge on the tip is intended to minimize the risk of piercing of the fundus by the tip of the brush and enables additional tissue sampling by absorption of the fluid and the cells from the uterus. The Utepreva Endometrial Sampler is rotated in a clockwise or counterclockwise manner for four or five 360° rotations. Upon completion of the rotations, the handle is pulled back so that the brush and tip are withdrawn into the outer sheath. The suction

{4}------------------------------------------------

created by the plunger located inside of the sheath provides aspiration of the cells and prevents loss of the sample from the brush, sponge and any fluid that was aspirated can then be immersed into the preservative solution to be used by the pathology lab.

The device materials consist of polydimethylsiloxane, high density polyethylene, cyanoacrylate adhesive, Loctite, Permabond polyolefin primer, stainless steel, Nylon, polycarbonate, and polypropylene. The device is packaged in a Tyvek® pouch. One pouch is then packaged in a unit box. Fifteen unit boxes are packaged within a single master box to create the final shipped product.

V. INDICATIONS FOR USE

The Utepreva Endometrial Sampler (UP01) is used to obtain endometrial cytological and histological samples.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE

In accordance with the 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] issued July 28, 2014, the comparison between the predicate and the subject devices is shown to be substantially equivalent by comparing the indications for use, principles of operation, technological characteristics, and performance testing similarities and differences (Table 1). The technological characteristic differences rationale demonstrates that the subject device is safe and effective as a legally marketed device and does not raise different questions of safety and effectiveness than the predicate device.

CharacteristicSubject DeviceUtepreva EndometrialSampler (UP01)(K240595)Predicate DeviceTao Brush™ I.U.M.C.Endometrial Sampler(K170603)Comparison
Regulatory ClassClass IIClass IISame
Regulatory Number884.1100884.1100Same
Regulation NameEndometrial BrushEndometrial BrushSame
Product CodeHFEHFESame
ManufacturerUtepreva LLCCook Incorporated-
Rx/OTCRxRxSame
SterileYesYesSame
SterilizationEtOEtOSame
Shelf-Life3 months3 yearsDifferent
Insertion DepthIndicatorSkirt (Guard) on SheathBlack Ink Mark on SheathDifferent
Dimensions
Working Length26.8 cm26.9 cmSimilar
Ball Tip10.2 Fr9.5 FrSimilar

Table 1. Comparison Table for Subject Device Utepreva Endometrial Sampler and Predicate Device Tao Brush™ I.U.M.C. Endometrial Sampler

{5}------------------------------------------------

CharacteristicSubject DeviceUtepreva EndometrialSampler (UP01)(K240595)Predicate DeviceTao Brush™ I.U.M.C.Endometrial Sampler(K170603)Comparison
Sheath OD11.43 Fr9.3 FrSimilar
Outer Sheath Length22.9 cm21.5 cmSimilar
Brush Length3.2 cm3.5 cmSimilar
Brush Diameter3 mm6 mmDifferent
Insertion DepthIndicator Distance fromDistal Tip6.58 cm7 cmSimilar
Materials
Insertion DepthIndicatorSkirt: Polycarbonate,acrylic adhesiveBlack ink mark on sheathDifferent
Wire CoreStainless steelStainless steelSame
PlungerPolydimethylsiloxaneNA-
Silicone FluidPolydimethylsiloxaneNA-
Brush BristlesNylonNylonSame
BackstopNylon 11NA-
Ball TipPolyurethane,PolypropyleneAcrylonitrile butadienestyrene (ABS)Different

PERFORMANCE DATA VII.

Non-clinical performance testing was conducted to verify the Utepreva Endometrial Sampler performance. The following performance data was provided in support of the substantial equivalence determination (Table 2).

Table 2. Non-Clinical Performance Testing

TestConclusion
BiocompatibilityPass
Cytotoxicity Testing (ISO 10993-5:2009) Sensitization (ISO 10993-10:2010) Vaginal Irritation (ISO 10993-10:2010)
Visual & Dimensional InspectionPass
Tensile StrengthPass
TorquePass
Bristle RetentionPass
Fluid RetentionPass
Fluid TransferPass

VIII. CONCLUSIONS

The information provided and summarized in the substantial equivalence table and performance testing results demonstrate that the Utepreva Endometrial Sampler is as safe and effective as the predicate device and support a determination of substantial equivalence.

§ 884.1100 Endometrial brush.

(a)
Identification. An endometrial brush is a device designed to remove samples of the endometrium (the mucosal lining of the uterus) by brushing its surface. This device is used to study endometrial cytology (cells).(b)
Classification. Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ” and
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(2) Labeling:
(i) Indication: Only to evaluate the endometrium, and
(ii) Contraindications: Pregnancy, history of uterine perforation, or a recent cesarean section, and
(3) Design and testing:
(i) The sampling component is covered within the vagina, and
(ii) For adherence of the bristles and brush head.