K Number
K240568

Validate with FDA (Live)

Date Cleared
2024-04-29

(60 days)

Product Code
Regulation Number
882.4560
Age Range
All
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DuraPro™ Oscillating System is indicated for drilling, burring, removing, and otherwise manipulating hard and soft tissue, bone, and other bone related tissue during spinal and orthopedic procedures.

ExcelsiusGPS® is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy. The system is indicated for the placement of spinal and orthopedic bone screws and interbody fusion devices.

ExcelsiusHub™ is intended for use as an aid for precisely locating anatomical structures to be used by surgeons for navigating compatible surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy. The system is indicated for the placement of spinal and orthopedic bone screws and interbody fusion devices.

Device Description

DuraPro™ Oscillating System consists of electrical The drill handpieces/attachments powered by an oscillating motor with a foot switch, and may be navigated using ExcelsiusGPS® or ExcelsiusHub™. The navigated components are provided non-sterile and are reusable. Attachments are provided sterile and are single-use.

AI/ML Overview

The provided text does not contain information about acceptance criteria and a study proving device performance in the context of an AI/ML medical device. The document is a 510(k) clearance letter for a device called "DuraPro™ Oscillating System," which is a surgical tool.

The document discusses:

  • Device Name: DuraPro™ Oscillating System
  • Regulation Number: 21 CFR 882.4560 (Stereotaxic Instrument)
  • Indications for Use: Drilling, burring, manipulating hard and soft tissue, bone during spinal and orthopedic procedures. It also mentions its use with ExcelsiusGPS® and ExcelsiusHub™ for navigation and guiding surgical instruments for screw and interbody fusion device placement.
  • Technological Characteristics: States it has similar characteristics to predicate instruments in design, intended use, material composition, function, and range of sizes.
  • Performance Testing: "The navigation and guidance accuracy of the DuraPro™ Oscillating System was evaluated using intra-operative imaging and accuracy verification testing. Testing confirmed that accuracy values meet the product requirement specification."
  • Electrical Safety/EMC: Confirmed to meet all standards.

However, none of this information outlines specific acceptance criteria for performance metrics usually associated with AI/ML devices (e.g., sensitivity, specificity, AUC), nor does it describe a study using human readers, ground truth establishment by experts, or training/test set details that would be relevant to an AI/ML performance evaluation. The "Performance Testing" section is very high-level and does not provide the granular details requested in your prompt regarding AI/ML device evaluation.

Therefore, I cannot provide the requested table and details because the source document does not pertain to the evaluation of an AI/ML medical device's performance in the way your prompt describes.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the date April 29, 2024. The month is spelled out, followed by the day and year. The text is in a simple, sans-serif font and is easy to read. The date is likely part of a document or calendar.

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" below it.

Globus Medical, Inc. Jennifer Antonacci Senior Group Manager, Regulatory Affairs Valley Forge Business Center 2560 General Armistead Avene Audubon, Pennsylvania 19403

Re: K240568

Trade/Device Name: DuraPro™ Oscillating System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: February 29, 2024 Received: February 29, 2024

Dear Jennifer Antonacci:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@)fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali -S

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices

{2}------------------------------------------------

OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K240568

Device Name

DuraPro™ Oscillating System

Indications for Use (Describe)

The DuraPro™ Oscillating System is indicated for drilling, burring, and otherwise manipulating hard and soft tissue, bone, and other bone related tissue during spinal and orthopedic procedures.

ExcelsiusGPS® is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or quiding compatible surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy. The system is indicated for the placement of spinal and orthopedic bone screws and interbody fusion devices.

ExcelsiusHub™ is intended for use as an aid for precisely locating anatomical structures to be used by surgeons for navigating compatible surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy. The system is indicated for the placement of spinal and orthopedic bone screws and interbody fusion devices.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary: DuraPro™ Oscillating System

Company: Globus Medical Inc. 2560 General Armistead Ave. Audubon, PA 19403 610-930-1800

  • Primary Contact: Jennifer Antonacci, Ph.D. Senior Group Manager, Regulatory Affairs
  • Date Prepared: February 29, 2024
  • Device Name: DuraPro™ Oscillating System
  • Common Name: Orthopedic Stereotaxic Instrument
  • Classification: Per 21 CFR as follows: &882.4560 Stereotaxic instrument Product Code(s): OLO Regulatory Class: II, Panel Code: 87

Primary Predicate: VERZA™ High Speed Drills (K233638)

Spacer (K222270)

Additional

Predicates:Medtronic Stealth-Midas MR8 System (K183644)ExcelsiusGPS® High Speed Drill Instruments (K190653)ExcelsiusGPS® (K171651, K191100)ExcelsiusHub™ (K211616)
Reference Devices:Electrical Drill System [XPS4000, Midas Rex EHS systemIntegrated Power Console (IPC)] (K081475)HEDRON® Cervical & Lumbar Spacers, SABLE® Expandable

Purpose:

The purpose of this submission is to request clearance of DuraPro™ Oscillating System for use with ExcelsiusGPS® and ExcelsiusHub™.

Device Description:

DuraPro™ Oscillating System consists of electrical The drill handpieces/attachments powered by an oscillating motor with a foot switch, and may be navigated using ExcelsiusGPS® or ExcelsiusHub™. The navigated components are provided non-sterile and are reusable. Attachments are provided sterile and are single-use.

{5}------------------------------------------------

Indications for Use:

The DuraPro™ Oscillating System is indicated for drilling, burring, removing, and otherwise manipulating hard and soft tissue, bone, and other bone related tissue during spinal and orthopedic procedures.

ExcelsiusGPS® is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy. The system is indicated for the placement of spinal and orthopedic bone screws and interbody fusion devices.

ExcelsiusHub™ is intended for use as an aid for precisely locating anatomical structures to be used by surqeons for navigating compatible surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy. The system is indicated for the placement of spinal and orthopedic bone screws and interbody fusion devices.

Technological Characteristics:

The subject DuraPro™ Oscillating System has the same technological characteristics as predicate instruments including design, intended use, material composition, function, and range of sizes.

FeaturesSubjectDuraPro™ Oscillating SystemPredicate &ReferenceDevices
IndicationsThe DuraPro™ Oscillating System is indicated fordrilling, burring, removing, and otherwisemanipulating hard and soft tissue, bone, and otherbone related tissue during spinal and orthopedicprocedures.Similar
ExcelsiusGPS® is intended for use as an aid forprecisely locating anatomical structures and for thespatial positioning and orientation of an instrumentholder or guide tube to be used by surgeons fornavigating and/or guiding compatible surgicalinstruments in open or percutaneous proceduresprovided that the required fiducial markers and rigidpatient anatomy can be identified on CT scans orfluoroscopy. The system is indicated for theplacement of spinal and orthopedic bone screws andinterbody fusion devices.
ExcelsiusHub™ is intended for use as an aid forprecisely locating anatomical structures to be used

{6}------------------------------------------------

FeaturesSubjectDuraPro™ Oscillating SystemPredicate &ReferenceDevices
by surgeons for navigating compatible surgicalinstruments in open or percutaneous proceduresprovided that the required fiducial markers and rigidpatient anatomy can be identified on CT scans orfluoroscopy. The system is indicated for theplacement of spinal and orthopedic bone screws andinterbody fusion devices.
InstrumentsNavigated and guided instrumentsSame
MaterialsStainless steel, PEEK, Radel, Titanium alloySame
BiocompatibilityStainless steel, PEEK, Radel, and titanium alloyhave a history of biocompatibility as instrumentmaterials per ASTM specifications and predicatedevices.Same
PerformanceThe navigation and guidance accuracy of the DuraPro™ OscillatingSystem was evaluated using intra-operative imaging and accuracyverification testing. Testing confirmed that accuracy values meet theproduct requirement specification.
ElectricalSafety/EMCElectrical safety and EMC testing were conducted on the DuraPro™Oscillating System. Testing confirmed the system meets allstandards.

Performance Testing:

The navigation and guidance accuracy of the DuraPro™ Oscillating System was evaluated using intra-operative imaging and accuracy verification testing. Testing confirmed that accuracy values meet the product requirement specification. Verification and validation testing and comparison to the predicate devices demonstrate that the subject DuraPro™ Oscillating System can be used in accordance with their indications.

Basis of Substantial Equivalence:

The DuraPro™ Oscillating System has been found to be substantially equivalent to the predicate devices with respect to technical characteristics, performance, and intended use. The information provided within this premarket notification supports substantial equivalence of the subject device to the predicate devices.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).