K Number
K240544

Validate with FDA (Live)

Device Name
Epitomee
Date Cleared
2024-09-13

(199 days)

Product Code
Regulation Number
876.5982
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Epitomee is indicated to aid in weight management in overweight and obese adults with a Body Mass Index (BMI) of 25 - 40 kg/m², when used in conjunction with diet and exercise.

Device Description

The subject device of this 510(k) notification, Epitomee, is comprised of two components: The outer component is a capsule and the inner component contains superabsorbent hydrogel particles. Each particle is approximately 250 microns in size. The particles are enclosed in pH sensitive compartments. Once Epitomee is swallowed, the capsule disintegrates in the stomach and releases the pH sensitive compartments. By absorption of water in the stomach fluid, the superabsorbent particles within the compartments swell to form a three-dimensional matrix designed to occupy volume in the stomach that creates a sensation of fullness, enabling weight loss. Epitomee mechanism of action is purely mechanical and does not involve any chemical activity. Epitomee works directly in the gastrointestinal (GI) tract. Epitomee is provided non-sterile.

AI/ML Overview

The provided document is a 510(k) summary for the device "Epitomee," which is an ingested, transient, space-occupying device for weight management. The document describes the device, its intended use, and the performance data that supports its substantial equivalence to a predicate device.

However, the document does not contain specific acceptance criteria or an explicit study proving that the device meets those acceptance criteria in the format requested by the prompt. Instead, it discusses the outcomes of clinical studies (RESET and ELECT) in comparison to a placebo or to demonstrate continued safety and efficacy. It focuses on demonstrating substantial equivalence to a predicate device (Gelesis Inc.'s Plenity) rather than meeting predefined, quantitative acceptance criteria for a new, novel device.

The prompt specifically asks for a table of acceptance criteria and reported device performance, sample sizes, expert qualifications for ground truth, adjudication methods, MRMC studies, standalone performance, and ground truth establishment for training and test sets. This information is typically found in detailed study reports or regulatory submissions for devices that require proof of meeting specific performance endpoints, especially for software or AI/ML-based devices. Epitomee is a physical medical device, not a software or AI/ML device in the conventional sense that would require the rigorous performance evaluation methods listed in the prompt (e.g., ground truth established by experts, MRMC studies, standalone algorithm performance).

Therefore, based on the provided text, I cannot extract the specific information requested in your prompt regarding acceptance criteria and the study proving the device meets them, because such a section or study design (as understood in the context of AI/ML or software device evaluation) is not present. The document describes a clinical trial (RESET study) demonstrating efficacy and safety compared to a placebo, and an extended study (ELECT) for long-term safety, to support "substantial equivalence" to a predicate device.

To answer your prompt, I must state that the detailed information about "acceptance criteria" and a "study that proves the device meets the acceptance criteria" in the format requested (especially points 1, 3, 4, 5, 6, 7, 8, 9) is NOT available in the provided 510(k) summary. The document's purpose is to demonstrate substantial equivalence, which relies on showing similar intended use, technological characteristics, and comparable safety and effectiveness, often through non-inferiority or superiority trials against a placebo/standard of care, rather than meeting pre-specified quantitative performance metrics for a diagnostic or AI system.

I can, however, provide the closest relevant information from the document that might relate to "performance" and "study outcomes" as a proxy, but it will not directly answer your prompt's specific requirements.


Closest Related Information from the Provided Document:

The document describes the Pivotal Clinical Study (RESET Study) as the primary evidence for the device's performance.

1. A table of acceptance criteria and the reported device performance:
The document does not provide a formal table of "acceptance criteria" for its performance as would be expected for a diagnostic or AI device. Instead, it states that "both co-primary endpoints were successfully met." These endpoints function as the 'criteria' for success.

Metric / EndpointAcceptance Criteria (Met/Achieved)Reported Device Performance (Epitomee Group)Control Group (Placebo) PerformanceStatistical Significance
Co-primary Endpoint 1:Significant reduction in total body weight (TBW) compared to placebo at 24 weeks.Mean TBW reduction: 6.6% (SD 6.5)Mean TBW reduction: 4.6% (SD 4.7)P<0.0001
Co-primary Endpoint 2:Rate of subjects with ≥5% TBW reduction at 24 weeks significantly exceeding a threshold.55.5% (CI: 46.1-64.6) of subjects achieved ≥5% TBW reduction(Not explicitly stated for placebo in this context, but implies lower)P<0.0001 (>35% threshold met)
Secondary Endpoint:Higher percentage of responders losing between 7.5-12.5% of TBW compared to placebo.≥ 7.5% TBW loss: 38.7% ≥ 10% TBW loss: 26.9% ≥ 12.5% TBW loss: 16.8%≥ 7.5% TBW loss: 21.5% ≥ 10% TBW loss: 10.7% ≥ 12.5% TBW loss: 6.6%Implied significance (stated "significantly higher percentage")
Safety:Favorable safety profile, low incidence of adverse events, no serious adverse device effects (SADEs).Fewer patient dropouts (13.8%) No SADEs Overall incidence of treatment-emergent AEs similar to placebo (mild/moderate for >95%) No difference in gastrointestinal AEs.Patient dropouts (14.2%)Met

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: The "test set" in this context refers to the subjects enrolled in the clinical trial.
    • Total Enrolled: 279 men and non-pregnant, non-lactating women.
    • Epitomee Group: 138 subjects (13.8% dropouts, so ~119 completed)
    • Placebo Group: 141 subjects (14.2% dropouts, so ~121 completed)
  • Data Provenance:
    • Country of Origin: All 9 study sites were in the US, located across 9 different states.
    • Retrospective or Prospective: Prospective, randomized, double-blind, placebo-controlled, multicenter trial.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This information is not applicable and not provided. "Ground truth" established by experts is typically for diagnostic interpretation (e.g., radiologist reads). For a weight management device, the "ground truth" for efficacy is objective clinical measurements (body weight, BMI) and for safety, it's reported adverse events and monitoring by investigators.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • This is not applicable and not provided. Adjudication methods are typically used for interpreting diagnostic images or clinical events by multiple readers/experts. For a device like Epitomee, efficacy is measured directly (weight change), and safety is based on adverse event reporting by subjects and assessment by clinical investigators, not adjudicated panel reviews for "ground truth".

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • This is not applicable. MRMC studies are specific to evaluating AI/software performance in diagnostic tasks by comparing human reader performance with and without AI assistance. Epitomee is a physical device, not an AI or diagnostic software.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • This is not applicable. Standalone performance refers to the performance of an algorithm or software system without human interaction. Epitomee is a physical device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The "ground truth" for efficacy was objective outcomes data (measured Total Body Weight reduction, percentage of subjects achieving specific weight loss thresholds, BMI reduction, improvements in quality of life scores).
  • The "ground truth" for safety was adverse event reporting by subjects and assessment by the investigational site staff (physicians, nurses).

8. The sample size for the training set:

  • This information is not provided because Epitomee is a physical device, not an AI/ML model that requires a "training set." Device development typically involves bench testing, non-clinical studies (e.g., animal studies), and clinical trials, not "training" in the AI sense.

9. How the ground truth for the training set was established:

  • This is not applicable as there is no "training set" in the context of an AI/ML model for this physical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 13, 2024

Epitomee Medical Ltd. % Janice Hogan Regulatory Counsel Hogan Lovells US LLP 1735 Market Street Suite 2300 Philadelphia, Pennsylvania 19103

Re: K240544

Trade/Device Name: Epitomee Regulation Number: 21 CFR 876.5982 Regulation Name: Ingested, Transient, Space Occupying Device For Weight Management And/Or Weight Loss Regulatory Class: Class II Product Code: QFQ Dated: February 27, 2024 Received: August 16, 2024

Dear Janice Hogan:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Anthony Lee -S

Anthony C. Lee, Ph.D., M.B.A. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K240544

Device Name

Epitomee

Indications for Use (Describe)

Epitomee is indicated to aid in weight management in overweight and obese adults with a Body Mass Index (BMI) of 25 - 40 kg/m2, when used in conjunction with diet and exercise.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) SUMMARY

Epitomee Medical's Epitomee

Submitter

Epitomee Medical Ltd. 17 Hatochen St. Caesarea lsrael

Contact Person: Ruthie Amir Date Prepared: September 12, 2024

Name of Device: Epitomee

Common Name: Weight management device

Regulation Number: 21 CFR 876.5982

Classification Name: Ingested, transient, space occupying device for weight management and/or weight loss

Product Code: QFQ

Regulatory Class: II

Predicate Device: Gelesis Inc.'s Plenity (DEN180060)

Device Description

The subject device of this 510(k) notification, Epitomee, is comprised of two components: The outer component is a capsule and the inner component contains superabsorbent hydrogel particles. Each particle is approximately 250 microns in size. The particles are enclosed in pH sensitive compartments. Once Epitomee is swallowed, the capsule disintegrates in the stomach and releases the pH sensitive compartments. By absorption of water in the stomach fluid, the superabsorbent particles within the compartments swell to form a three-dimensional matrix designed to occupy volume in the stomach that creates a sensation of fullness, enabling weight loss. Epitomee mechanism of action is purely mechanical and does not involve any chemical activity. Epitomee works directly in the gastrointestinal (GI) tract. Epitomee is provided non-sterile.

Intended Use / Indications for Use

Epitomee is intended to aid in weight management in overweight and obese adults with a Body Mass Index (BMI) of 25 - 40 kg/m², when used in conjunction with diet and exercise.

{5}------------------------------------------------

Summary of Technological Characteristics

Epitomee has very similar technological characteristics compared to the predicate device. Both devices are composed of two major components: an outer capsule and an inner expendable component occupying space in the stomach. Both devices are ingestible, transient, space occupying devices designed as a standard capsule for oral intake. For both devices the oral capsule encloses self-expandable particles that form a three-dimensional matrix designed to occupy volume in the stomach, to create a sensation of fullness. The capsule of both devices disintegrates in the stomach and releases the enclosed self-expanded particles. For both devices the hydrogel particles increase their weight and dimensions after absorbing water from the stomach and both devices interact with the stomach by occupving volume. Both devices are self-disintegrated to enable natural evacuation through gastric emptying.

Although the subject device consists of a triangular arrangement and has a smaller net volume than the predicate, the triangular shape of the compartments was designed to attain relatively large dimensions, creating a space occupying tree-dimensional structure and the difference does not raise new questions of safety or effectiveness as both devices achieve their intended use by occupying volume in the stomach, thereby creating sensation of fullness. Clinical study results showed that the subject device is as safe and effective as the predicate device. In addition, despite the different materials used by the two devices utilize hydrogel materials to expand with water. Both devices have been demonstrated to be biocompatible. Furthermore, both devices are intended to degrade within the user's GI tract and be eliminated through normal bowel movements emptying.

A table comparing the key features of the subject and predicate device is provided below.

Epitomee Medical'sEpitomee(Subject Device)Gelesis Inc/'s Plenity(DEN180060)(Predicate Device)
Regulatory ClassIIII
CFR Regulation21 CFR 876.598221 CFR 876.5982
Product CodeQFQQFQ
Intended UseEpitomee is indicated to aid inweight management inoverweight and obese adultswith a Body Mass Index (BMI)of 25 - 40 kg/m², when used inconjunction with diet andexercise.Plenity is indicated to aid inweight management inoverweight and obese adultswith a Body Mass Index (BMI)of 25 - 40 kg/m², when used inconjunction with diet andexercise.
Epitomee Medical'sEpitomee(Subject Device)Gelesis Inc/'s Plenity(DEN180060)(Predicate Device)
Treated area/Site of application in the bodyNon-systemic and works directly in the gastrointestinal (GI) tract (stomach)Non-systemic and works directly in the gastrointestinal (GI) tract (stomach)
System componentsAn outer capsule that is swallowed by the patients and an inner expendable device occupying space in the stomachAn outer capsule that is swallowed by the patients and an inner expendable device occupying space in the stomach
AdministrationStandard capsule for oral intake;2 capsules per day (one capsule before lunch and one capsule before dinner) with waterStandard capsule for oral intake;6 capsules per day (3 capsules before lunch and 3 capsules before dinner) with water
Duration of Use6 months6 months
Device Design/Mode of ActionThe device is an ingestible, transient, space occupying device designed as a standard capsule for oral intake.The oral capsule contains thousands of particles. The particles are contained in compartments and can each hydrate up to 100 times their original weight, to form a three-dimensional matrix designed to occupy volume in the stomach..The device is an ingestible, transient, space occupying device designed as a standard capsule for oral intake.The oral capsule contains thousands of Plenity particles. The particles are non-clustering and can each hydrate up to 100 times their original weight, to form a three-dimensional matrix designed to occupy volume in the stomach.
Capsule MaterialPharmaceutical grade hydroxypropyl methylcellulose (HPMC) capsulePorcine gelatin capsule
BiocompatibilityBiocompatibleBiocompatible
Epitomee Medical'sEpitomee(Subject Device)Gelesis Inc/'s Plenity(DEN180060)(Predicate Device)
Compartment SizeDeployedParticles contained in threecompartments that form athree-dimensional triangularmatrix once hydrated in thestomach. The total net volumein the stomach is 17cc, witheach edge of triangular matrixhaving a dimension of 62 mmThe Plenity device does notcontain a compartment. Itcontains thousands of particlesthat form a non-clusteringthree-dimensional matrix in thestomach once hydrated. Thetotal volume is approximatelyone quarter of the stomachvolume
Self-ExpandingSelf-expanding when hydratedSelf-expanding when hydrated
Principles ofOperationThe capsule disintegrates in thestomach, allowing water topenetrate the envelope andhydrate the gel particles. Theexpansion of the particlesresults in unfolding of the deviceinto a three-dimensional matrixcomposed of 3 identicalcompartments occupyingvolume in the stomach.The hydrogel particles increasetheir weight and volume afterabsorbing the water from thestomach.When fully hydrated, theEpitome structure occupiesonly about 17cc net volume inthe stomach, with each edge ofthe triangular matrix having adimension of 62 mm,generating a space occupyingtree-dimensional structure tocreate a sensation of fullness.The capsule disintegrates inthe stomach allowing water tohydrate the gel particles. Thegel particles spread and form athree-dimensional matrixoccupying volume in thestomach.The hydrogel particles increasetheir weight and volume afterabsorbing the water from thestomach.When fully hydrated, thePlenity particles occupy aboutquarter of average stomachvolume to create a sensation offullness.
Release characteristics ofsubstancesThe capsule disintegrates in thestomach and releases theparticles enclosed in thecompartmentsThe capsule disintegrates inthe stomach and releases thecontained particles.
Epitomee Medical's Epitomee(Subject Device)Gelesis Inc/'s Plenity(DEN180060)(Predicate Device)
Degradation mechanismand profileEpitomee self-disintegrates.
Epitomee passes through thepylorus and device breaksdown in the small intestinewhere the hydrogel particlesare free to flow with the mixoutside the body.Plenity passes through thedigestive system maintainingits three-dimensional structurein the stomach and smallintestine before breaking downin the colon and flowing freelywith the mix outside the body.The water is then released andreabsorbed by the body.Plenity particles are eliminatedthrough normal bowelmovements.

Epitomee Substantial Equivalence Chart

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

Performance Data

Nonclinical and clinical studies have been conducted to demonstrate the performance of Epitomee.

Biocompatibility

Epitomee underwent biocompatibility assessment according to ISO 10993-1 requirements for a device having permanent (> 30 days) surface contact with mucosal membranes. All the biological endpoints have been addressed by biocompatibility studies and/or rationales as the device components have long and documented history of safe use. Endpoints considered are:

  • Cytotoxicity
  • Sensitization ●
  • Irritation ●
  • Acute Systemic Toxicity ●
  • Genotoxicity
  • Sub-chronic Toxicity ●
  • Chronic Toxicity
  • Carcinogenicity ●
  • Material Mediated Pyrogenicity ●
  • Implantation

{9}------------------------------------------------

Nonclinical Testing

Bench testing has been conducted to demonstrate that the Epitomee meets its specification as functions as intended. The testing includes:

  • Device dimensional verification ●
  • Device inflation .
  • Device functional performance ●
  • Device mechanical Integrity ●
  • Device loss of rigidity ●
  • Device inflation with different beverages and food mediums .
  • Shape retention ●
  • Device clearance from stomach
  • Device degradation in intestine ●
  • Device irregular surface test ●
  • . Device interaction with medication

The in vitro tests of inflation and degradation were verified by ex vivo studies. In addition, potential interaction between the Epitomee and Metformin was evaluated. GLP animal studies were conducted to demonstrate the device safety in vivo.

Clinical Data

The performance of Epitomee is supported by multiple clinical studies.

Pivotal Clinical Study (RESET Study)

The RESET study was a prospective, randomized, double-blind, placebo-controlled, multicenter trial to support the use of Epitomee for weight reductions in patients. The study enrolled 279 men and non-pregnant, non-lactating women, 18 years of age or older, who are overweight (body mass index [BMI] of 25-30 kg/m²) or obese (BMI of 30-40 kg/m²), with and without prediabetes. All 9 sites enrolled in the study were in the US, located across 9 different states.

The subjects were evenly matched for gender and age groups with females constituting 80.4% and 79.4% of the Epitomee and placebo groups, respectively, and the majority of subjects being between 40 and 65 years of age (66.7% and 67.4% in the Epitomee and placebo groups, respectively). Mean BMI was 34.1 in Epitomee group and 33.7 in the placebo group. The majority of the study subjects were Caucasian, with 70.3% and 65.2% of the recruited in the Epitomee and placebo groups, respectively. Black or African American subjects constituted 21.0% and 24.1% of the total sample in the Epitomee and placebo groups, respectively.

Eligible subjects were randomized to test or placebo arms. Subjects were to take one capsule of the investigational device or placebo twice daily, with 2 cups of water half-hour before the main meals, for 24 weeks. Subjects in both groups received the same program of diet, physical activity, and behavior therapy. Efficacy and safety assessments were conducted up to week 24 since start of the study treatment. An additional safety assessment was conducted at week 28.

{10}------------------------------------------------

The study demonstrated that both co-primary endpoints were successfully met. Epitomee treatment group total body weight (TBW) reduction at 24-weeks post randomization was significantly greater (mean, 6.6%; SD, 6.5) than the placebo group (mean, 4.6%; SD, 4.7) (P<0.0001). The co-primary endpoint of treatment responders was also achieved: the rate of Epitomee treatment group subjects whose total body weight was reduced by at least 5% at 24-weeks post randomization was 55.5% (Cl: 46.1-64.6), significantly exceeding the threshold of >35% (P<0.0001).

Subgroup analyses were performed to evaluate total body weight loss at week 24 in clinically relevant subgroups of patients (gender, baseline BMI categories, subjects with %TBW≥0) and showed higher TBW reduction in the Epitomee arm than the placebo arm. The differences between the arms ranged from approximately 2%-5%.

Epitomee capsule treatment efficacy is further supported by the analysis of seconday endpoints. Epitomee capsule treatment resulted in a significantly higher percentage of responders losing between 7.5-12.5% of TBW as compared to the placebo treatment. Specifically, 38.7% of Epitomee treated patients achieved ≥ 7.5%, 26.9% achieved ≥ 10%, and, 16.8%, ≥ 12.5% of TBW loss compared to 21.5%, 10.7% and 6.6% in the placebo group. Consistent with the weight results, Epitomee treatment achieved a greater reduction in Excess Weight Loss (EWL) and higher BMI reduction over time resulting in more subjects shifting from obesity Class II to a lower obesity category.

Finally, subjects treated with Epitomee showed better improvement in quality of life in several items of the IWQOL-Lite-CT questionnaire as well as the physical function relative to the placebo group, with physical function score being statistically meaningful and both physical score and total score being marqinally significance.

The study demonstrated a favorable safety profile of Epitomee treatment throughout the study duration. Epitomee treatment was well tolerated, with fewer patient dropouts in Epitomee group than the control group 19 (13.8%) and 20 (14.2%), respectively. There were no serious adverse device effects (SADEs) in the study.

There were no differences in the incidence of treatment-emergent AEs between the study groups. Adverse events were equally distributed between the groups when assessed by severity. There were no differences in the incidence of treatment-emergent AEs between the groups in relation to causality, AE outcome, and action taken. In both groups, almost all adverse events were assessed by the investigator as mild or moderate in intensity (>95%). The overall incidence of treatment-emergent gastrointestinal adverse events was no different in the Epitomee group vs. the control group.

Notably, the safety and efficacy performance of Epitomee as shown in the RESET study was very similar to that of the predicate device.

Extended Study (ELECT Study)

The ELECT study assessed the effect of Epitomee on body weight after an additional exposure of 24 weeks in subjects from the RESET pivotal study. Subjects from Epitomee and Placebo

{11}------------------------------------------------

groups in the RESET study who completed the 24-week treatment period and had at least 3% weight loss were offered to participate in this extended study. The ELECT study is an openlabel, one arm extension study and all subjects were treated with Epitomee, twice daily. Sixteen (16) subjects who were assigned to the placebo during RESET crossed over to the Epitomee group (Placebo-Epitomee) and 17 subjects who were assigned to the Epitomee group during the RESET continued Epitomee for an additional 24 weeks (Epitomee-Epitomee).

Safety results from the ELECT study showed that the overall incidence of AEs was no different between the Epitomee subjects who were exposed for an entire year (27 AEs in 12 [70.6%] subjects) versus the Placebo subjects who were switched to Epitomee and were exposed to Epitomee for the first time for a period of 24 weeks (17 AEs in 10 [62.5%] subjects). All AEs in the ELECT study were mild gastrointestinal (GI) events, and all recovered. There were no deaths or SAEs.

Gastric Emptying Study

Epitomee Medical has conducted a prospective, single center, open-label, single-arm study to examine the effect of the Epitomee device on gastric emptying rate in healthy subjects. Ten (10) healthy subjects were enrolled in the study. Subjects completed a gastric emptying test using gastric scintigraphy before taking Epitomee (baseline). Following the baseline scintigraphy, the subjects received the Epitomee device treatment for 10 consecutive days. and underwent another gastric emptying test with capsule intake prior to the test meal.

The results showed that Epitomee was well-tolerated and had a favorable safety profile with no SAEs. There was no significant effect on gastric emptying with Epitomee.

Conclusions

Epitomee is as safe and effective as its predicate device. Epitomee has the same intended use and indications for use, and similar technological characteristics as its predicate device. The minor technological differences between Epitomee and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that Epitomee is as safe and effective as the predicate device. Thus, Epitomee is substantially equivalent.

§ 876.5982 Ingested, transient, space occupying device for weight management and/or weight loss.

(a)
Identification. This device is an ingested material that transiently occupies space in the stomach. The device passes from the body via the natural gastrointestinal tract.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible for its intended use.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions for use, as follows:
(i) Performance bench testing in a simulated use model must evaluate device disintegration and device hydration state throughout the gastrointestinal tract;
(ii) Bioburden and moisture content assessments must evaluate device infection risk throughout the labeled shelf life; and
(iii) Performance data must support the shelf life of the device by demonstrating continued package integrity and device functionality over the labeled shelf life.
(3) Clinical performance testing must demonstrate the device performs as intended and evaluate the following:
(i) Weight change;
(ii) All adverse events, including obstruction, dilation, diarrhea, constipation, and dehydration; and
(iii) Interaction with representative medications.
(4) Physician and patient device labeling must state:
(i) The clinical benefit of the device as assessed by using percent total body weight loss;
(ii) Treatment must be offered in combination with diet and exercise;
(iii) Instructions on how to use the device as intended including how to avoid interaction with medication; and
(iv) The shelf life of the device.