Intended Use

The Wearable Breast Pump (Model S32) is intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.

Device Description

The Wearable Breast Pump (Model S32), is an electrically powered wearable single breast pump consisting of the following key components: a flange, pump motor, silicone diaphragm, milk collector, bra adjustment buckle, valve, and USB charging cable. It is designed to work in the user's bra and has a rechargeable battery so it can be used hands-free without any external power cords. The motor unit includes a press-button user interface, pump body, and LED display. Pumping can be performed on one breast (single pumping). The user interface allows the user to switch from stimulation, expression, massage, and auto modes and control the vacuum levels within those modes. All available modes consist of 9 vacuum levels. The S32 model is capable of providing vacuum levels from 40-120 mmHg with cycling rates from 86-110 cycles per minute in stimulation mode, vacuum levels from 120-245 mmHg with cycling rates from 33-109 cycles per minute in expression mode, vacuum levels from 40-120 mmHg with cycling rates from 90-120 cycles per minute in massage mode, and vacuum levels from 40-245 mmHg with cvcling rates from 33-120 cvcles per minute in auto mode. The model S32 Wearable Breast Pump is charged with a 5 V DC adaptor and powered by an internal rechargeable lithium-ion polymer battery. The motor unit operates on embedded software. Software updates are not supported. The subject device is for repeated use by a single user in a home environment. The device is provided not sterile. The motor unit operates on a rechargeable battery and does not function when charging. The rechargeable battery can be charged from the external USB adapter if the motor unit is not in operation. The breast pump expresses milk by creating a seal around the nipple using the flange and applying and releasing suction to the milk is collected in a milk collection container, which can be used for storage. To prevent milk from flowing into the vacuum system, a backflow protection membrane physically separates the milk-contacting pathway from the vacuum system. All other components (i.e., motor unit/housing) of the subject device are not in contact with the breast. All milk contacting components are compliant with 21 CFR 174-179.

AI/ML Overview

The provided text is a 510(k) Summary for a Wearable Breast Pump (Model S32). It outlines the device's characteristics, comparison to a predicate device, and a summary of non-clinical performance testing. However, it does not include the specific details typically found in a study proving a device meets acceptance criteria, especially for a medical device involving AI, image analysis, or complex diagnostics.

The document states: "Other performance testing was conducted to show that the device meets its design requirements and performs as intended. The performance tests include:

  • Vacuum level verification testing at each mode/cycle demonstrated that the devices meet mode/cycle specifications.
  • Backflow protection testing was conducted to verify liquid does not backflow into the tubing.
  • Use life testing was conducted to demonstrate that the device maintains its specifications ● throughout its proposed use life.
  • Battery performance testing was conducted to demonstrate that the battery remains functional during its stated battery use-life.
  • Battery status indicator testing was conducted to demonstrate that the battery status indicator ● remains functional during its stated battery life."

Based on the provided text, here's an attempt to answer your questions, highlighting what is present and what is absent:


1. A table of acceptance criteria and the reported device performance

The document broadly outlines performance tests conducted, but it does not provide explicit acceptance criteria values or the specific reported device performance results in a table format. It states that "Vacuum level verification testing at each mode/cycle demonstrated that the devices meet mode/cycle specifications." This implies that the measured vacuum levels and cycle speeds met predefined targets, but those targets (acceptance criteria) and the actual measured values (reported performance) are not detailed.

Table Placeholder (Information NOT PROVIDED in the text):

TestAcceptance CriteriaReported Device Performance
Vacuum Level Verification (Stimulation Mode)e.g., Within ±X mmHg of target rangee.g., Successfully demonstrated meeting specifications
Backflow ProtectionNo liquid backflow into tubingConfirmed no backflow
Use LifeMaintains specifications throughout proposed use life (e.g., X hours/cycles)Demonstrated maintenance of specifications
Battery PerformanceFunctional during stated battery use-life (e.g., Y hours)Demonstrated functionality
Battery Status IndicatorFunctional during stated battery lifeDemonstrated functionality

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the sample sizes for any of the performance tests (e.g., number of units tested for vacuum levels, backflow, use life, or battery performance). It also does not mention data provenance, as this is a mechanical/electrical device, not one that processes patient data. The testing is likely lab-based internal validation rather than clinical data collection.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable to the type of device and testing described. Ground truth establishment by experts (like radiologists for image analysis) is relevant for AI/ML-driven diagnostic devices. For a breast pump, performance is assessed through engineering measurements (e.g., vacuum pressure, cycle speed, battery life, physical backflow).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Adjudication methods are used in studies involving human interpretation (e.g., reading medical images) where there might be disagreement among experts. For mechanical/electrical performance testing, measurements are typically objective and repeatable.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. An MRMC study is relevant for diagnostic devices (especially those involving AI assistance to human readers/interpreters). This document is for a medical device (breast pump) that does not involve AI assistance for human interpretation or diagnosis.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. The device has "embedded software" but it's for controlling the pump's operation (modes, vacuum levels, timing), not for diagnostic or analytical purposes that would require standalone algorithm performance evaluation against a "ground truth" in the typical sense of AI/ML devices.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For this device, "ground truth" refers to the engineering specifications and expected performance, verified through direct physical measurements. For example:

  • Vacuum levels: Verified against calibrated pressure sensors.
  • Cycle speed: Verified against timing mechanisms.
  • Backflow protection: Verified visually or with sensors to confirm no liquid passes.
  • Use life/Battery performance: Verified by running the device for specified durations and checking if performance metrics remain within tolerance.

8. The sample size for the training set

Not applicable. This device's embedded software controls the pump's mechanical functions; it is not an AI/ML model trained on a dataset. Therefore, there is no "training set" in the context of machine learning.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for an AI/ML model. The "ground truth" for the device's design specifications would have been established through engineering design principles, regulatory standards (e.g., IEC standards for medical electrical equipment), and a thorough understanding of the physiological requirements for breast milk expression.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

July 1, 2024

Shenzhen TPH Technology Co., Ltd. Peter Chen General Manager Room 203, 2nd floor,29th Building, Lianchuang Technology Park, Nanwan, Longgang District Shenzhen, Guangdong 518100 CHINA

Re: K240536

Trade/Device Name: Wearable Breast Pump (Model S32) Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Received: May 29, 2024

Dear Peter Chen:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See

{2}------------------------------------------------

the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Monica D. Garcia -S

Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K240536

Device Name Wearable Breast Pump (Model S32)

Indications for Use (Describe)

The Wearable Breast Pump (Model S32) is intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K240536 Page 1 of 4

510(k) Summary - K240536

1. Submitter Information

Applicant:Shenzhen TPH Technology Co., Ltd.
Address:5th Floor, Building No.29 East side,Lianchuang 2th technology park, BulanRoad, Nanwan Community, LonggangDistrict, Shenzhen, China

2. Correspondent Information

Contact:Peter Chen
General Manager
Phone:+ 86 (755) 827-03212
Email:peter@tph-tech.com

3. Date prepared: June 24, 2024

4. Device Information

Device Name:Wearable Breast Pump (Model S32)
Common Name:Powered Breast Pump
Regulation Number:21 CFR 884.5160
Regulation Name:Powered Breast Pump
Product Code:HGX (Pump, Breast, Powered)
Regulatory Class:Class II

5. Predicate Device Information

Device Name:Wearable Breast Pump (Model S12)
510(k) Number:K212180
Manufacturer:Shenzhen TPH Technology Co., Ltd.

The predicate device has not been subject to a design-related recall.

6. Device Description

The Wearable Breast Pump (Model S32), is an electrically powered wearable single breast pump consisting of the following key components: a flange, pump motor, silicone diaphragm, milk collector, bra adjustment buckle, valve, and USB charging cable. It is designed to work in the user's bra and has a rechargeable battery so it can be used hands-free without any external power cords. The motor unit includes a press-button user interface, pump body, and LED display. Pumping can be performed on one breast (single pumping). The user interface allows the user to switch from stimulation, expression, massage, and auto modes and control the vacuum levels within those modes.

All available modes consist of 9 vacuum levels. The S32 model is capable of providing vacuum levels from 40-120 mmHg with cycling rates from 86-110 cycles per minute in stimulation mode, vacuum levels from 120-245 mmHg with cycling rates from 33-109 cycles per minute in expression mode, vacuum levels from 40-120 mmHg with cycling rates from 90-120 cycles per minute in massage mode, and vacuum levels from 40-245 mmHg with cvcling rates from 33-120 cvcles per minute in auto mode. The model S32 Wearable Breast Pump is charged with a 5 V DC adaptor and powered by an internal rechargeable lithium-ion polymer battery. The motor unit operates on embedded software. Software updates are not supported. The subject device is for repeated use by a single user in a home environment. The device is provided not sterile.

{5}------------------------------------------------

The motor unit operates on a rechargeable battery and does not function when charging. The rechargeable battery can be charged from the external USB adapter if the motor unit is not in operation.

The breast pump expresses milk by creating a seal around the nipple using the flange and applying and releasing suction to the milk is collected in a milk collection container, which can be used for storage. To prevent milk from flowing into the vacuum system, a backflow protection membrane physically separates the milk-contacting pathway from the vacuum system.

All other components (i.e., motor unit/housing) of the subject device are not in contact with the breast. All milk contacting components are compliant with 21 CFR 174-179.

7. Indications for Use

The Wearable Breast Pump (Model S32) is intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.

8. Comparison of Intended Use and Technological Characteristics with the Predicate Device

The table below compares the intended use and technological characteristics of the subject and predicate device.

Wearable Breast Pump(Model S32)K240536Wearable Breast Pump(Model S12)K212180Comparison
Subject DevicePredicate Device
Product NameWearable Breast Pump (Model S32)Wearable Breast Pump (Model S12)N/A
Product CodeHGXHGXSame
Regulation Number21 CFR 884.516021 CFR 884.5160Same
Regulatory ClassClass IIClass IISame
Patient PopulationLactating WomenLactating WomenSame
Indications for UseThe Wearable Breast Pump(Model S32) is intended toexpress milk from lactatingwomen in order to collect milkfrom their breasts. The device isintended for a single user.The Wearable Breast Pump(Model S12) is intended toexpress milk from lactatingwomen in order to collect milkfrom their breasts. The device isintended for a single user.Same
Pump OptionsSingleSingleSame
Cycling controlmechanismMicrocontrollerMicrocontrollerSame
Backflow ProtectionYesYesSame
Suction ModesStimulation Mode andExpression ModeStimulation Mode andExpression ModeSame
Suction levels99Same
Adjustable suctionlevelsYesYesSame

Table 1: Comparator Table for Subject and Predicate Devices

{6}------------------------------------------------

Vacuum range:Stimulation-40 to -120 (±5) mmHg-40 to -105 (±5) mmHgDifferent
Vacuum range:Expression-120 to -245 (±5) mmHg-40 to -245 (±5) mmHgDifferent
Vacuum range:Massage-40 to -120 (±5) mmHgN/ADifferent
Vacuum range:Auto-40 to -245 (±5) mmHgN/ADifferent
Cycle Speed:Stimulation86 to 110 (±2) cycles/minute70 to 114 cycles/minuteDifferent
Cycle Speed:Expression33 to 109 (±2) cycles/minute23 to 90 cycles/minuteDifferent
Cycle Speed:Massage90 to 120 (±2) cycles/minuteN/ADifferent
Cycle Speed:Auto33 to 120 (±2) cycles/minuteN/ADifferent
ControlsOn-Off switch, vacuumadjustment, modechange/powerOn/Off button;Mode selectionIncrease/decrease vacuumbutton;Similar
Power SupplyLi-Ion BatteryLi-Ion BatterySame
IndicatorsYes, LEDYes, LEDSimilar
MaterialsMilk Container: PolypropyleneFlange: SiliconePump Outer Housing:Acrylonitrile ButadieneStyrene (ABS) plasticMilk Container: PolypropyleneFlange: SiliconePump Outer Housing:Acrylonitrile ButadieneStyrene (ABS) plasticSame

The indications for use of the subject and predicate device are identical, and they have the same intended use (i.e., for collection of breast milk from the breasts of lactating women).

The subject and predicate devices have similar technological features, including wearable operation, power supply, and user interface. However, as shown in the table above, there are technological differences between the subject and predicate device, including different overall vacuum/cycle specifications and available modes. The different technological characteristics of the subject device, as compared to the predicate device, do not raise different questions of safety and effectiveness.

9. Summary of Non-Clinical Performance Testing

Biocompatibility

Biocompatibility information was provided in accordance with Attachment G of the 2023 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process."

Electrical Safety

Testing was conducted in accordance with ANSI/AAMI ES60601- 1:2005/A2:2010 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance), IEC 62133-2:2017, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems, and IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.

{7}------------------------------------------------

Electromagnetic Compatibility

Testing was conducted in accordance with IEC 60601-1-2:2014 Medical Electrical Equipment - Part 1-2: "General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests. "

Software

Software was evaluated at the Basic Documentation level as recommended in the 2023 FDA guidance document "Content of Premarket Submissions for Device Software Functions. "

Performance Testing

Other performance testing was conducted to show that the device meets its design requirements and performs as intended. The performance tests include:

  • . Vacuum level verification testing at each mode/cycle demonstrated that the devices meet mode/cycle specifications.
  • Backflow protection testing was conducted to verify liquid does not backflow into the tubing.
  • Use life testing was conducted to demonstrate that the device maintains its specifications ● throughout its proposed use life.
  • . Battery performance testing was conducted to demonstrate that the battery remains functional during its stated battery use-life.
  • Battery status indicator testing was conducted to demonstrate that the battery status indicator ● remains functional during its stated battery life.

10. Conclusion

The subject and predicate devices have the same intended use and comparable technological characteristics. The differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness. The performance data demonstrate that the subject device is substantially equivalent to the predicate device.

§ 884.5160 Powered breast pump.

(a)
Identification. A powered breast pump in an electrically powered suction device used to express milk from the breast.(b)
Classification. Class II (performance standards).