K Number
K240506

Validate with FDA (Live)

Date Cleared
2024-10-22

(244 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The KARL STORZ Holders are manually operated surgical devices intended to hold endoscopes (such as arthroscopes, etc) and accessories during general and neurologic diagnostic and therapeutic procedures.

Device Description

The KARL STORZ Holder is a manually operated surgical instrument intended to hold scopes (such as endoscopes, arthroscopes, etc) in a desired position during diagnostic, therapeutic, and surgical procedures (including neurologic procedures). The holder is manufactured from Stainless Steel and functions by the user manually tightening or loosening a knob, which opens and closes around a surgical scope. The device is intended to attach to the several KARL STORZ holding systems, which are a Class 1, table-mounted, mechanical and electromechanical holding arms that have many general surgical functions in addition to scope holding; it can also be used for instrument holding, retraction, and positioning.

AI/ML Overview

The provided text is a 510(k) summary for the KARL STORZ Holders, which are manually operated surgical devices intended to hold endoscopes and accessories. This document primarily focuses on demonstrating substantial equivalence to a predicate device rather than providing extensive details about a clinical study with acceptance criteria for an AI/CAD-like device.

Therefore, the requested information regarding acceptance criteria, study design, sample sizes, expert involvement, ground truth establishment, and MRMC studies for an AI/CAD device is not available in the provided text. The document describes a traditional medical device (a mechanical holder) and its non-clinical performance testing.

However, based on the information that is available in the document, here's a summary of the non-clinical performance data and the demonstration of substantial equivalence:

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not detail specific quantitative acceptance criteria or reported values in a table format for a clinical outcome beyond demonstrating functional equivalence. The performance data is described qualitatively as meeting the needs for substantial equivalence.

Performance CategoryAcceptance Criteria (Implied for Substantial Equivalence)Reported Device Performance
Holding PerformanceSufficient to securely hold intended devices (scopes, accessories) during procedures and maintain position.Testing was performed (Non-Clinical Performance Data: "Holding Performance testing"). Implied to be acceptable for intended use. Max loading weight of 3kg reported.
Direction of View/Image StabilityMaintain stable direction of view/image during use.Testing was performed (Non-Clinical Performance Data: "Direction of view/Image stability testing"). Implied to be acceptable for intended use, as it contributes to substantial equivalence.
Material of ConstructionBiocompatible and durable for surgical use.Stainless Steel, same as predicate device.
Mode of OperationManual, user-friendly.Manual, same as predicate device.
Cleaning & ReprocessingSterilizable and cleanable without degradation.Manual cleaning and Steam Sterilization, same as predicate device.

2. Sample Size Used for the Test Set and Data Provenance:

  • The document describes non-clinical performance testing rather than a clinical test set involving human subjects or AI algorithm evaluation.
  • The exact sample size for the "Holding Performance testing" and "Direction of view/Image stability testing" is not specified.
  • Data Provenance: The testing was performed on the "subject device" (KARL STORZ Holders). The country of origin of the data is not explicitly stated beyond the submitter being KARL STORZ SE & Co. KG in Tuttlingen, Germany. The data would be considered prospective as it's testing of the new device for regulatory submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

  • This information is not applicable as the document describes non-clinical performance testing of a mechanical device, not the evaluation of an AI algorithm requiring expert ground truth for image interpretation.

4. Adjudication Method for the Test Set:

  • This is not applicable for the same reason as point 3.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

  • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of AI on human reader performance, which is not the subject of this 510(k) submission for a mechanical endoscope holder.

6. If a Standalone (algorithm only without human-in-the-loop performance) was done:

  • No, a standalone algorithm performance study was not done. This device is a mechanical surgical instrument, not an AI algorithm.

7. The Type of Ground Truth Used:

  • The concept of "ground truth" as typically used for AI/CAD devices (e.g., expert consensus, pathology, outcome data) is not applicable. For this mechanical device, performance is evaluated against engineering specifications and functional requirements (e.g., ability to hold weight, maintain stability).

8. The Sample Size for the Training Set:

  • This is not applicable. The device is a mechanical instrument, not an AI algorithm that requires a training set.

9. How the Ground Truth for the Training Set was Established:

  • This is not applicable for the same reason as point 8.

In summary, the provided FDA 510(k) summary (K240506) pertains to a mechanical endoscope holder, not an AI/CAD-enabled medical device. Therefore, the detailed requirements for acceptance criteria and study designs related to AI performance are not present in this document. The purpose of this 510(k) is to demonstrate substantial equivalence to a predicate mechanical device through non-clinical performance testing.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. Food & Drug Administration".

October 22, 2024

KARL STORZ Endoscopy America, Inc Mario Trujillo Regulatory Affairs Specialist 2151 E. Grand Avenue El Segundo, California 90245

Re: K240506

Trade/Device Name: KARL STORZ Holders Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: OCV, GCJ Dated: September 23, 2024 Received: September 23, 2024

Dear Mario Trujillo:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Digitally signed by Long H. Long H. Chen -Chen -S S Date: 2024.10.22 13:32:12 -04'00'

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K240506

Device Name

KARL STORZ Holders

Indications for Use (Describe)

The KARL STORZ Holders are manually operated surgical devices intended to hold endoscopes (such as arthroscopes, etc) and accessories during general and neurologic diagnostic and therapeutic procedures.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image is a logo for STORZ. The logo is blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller font. The logo is simple and professional.

510(k) Summary

This 510(i) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92 and the FDA guidance document titled "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" issued on July 28, 2014. All data included in this document is accurate and complete to the best of KARL STORZ SE & Co. KG knowledge.

Submitter:KARL STORZ SE & Co. KGDr.-Karl-Storz-Straße 3478532 Tuttlingen, Germany
Contact:Mario TrujilloRegulatory Affairs SpecialistTel.: (424) 218-8481Email : Mario.Trujillo@karlstorz.com
Date ofPreparation:October 21, 2024
DeviceIdentification:Trade Name: KARL STORZ HoldersCommon Name: Endoscope HolderClassification Name: Endoscope Holder (21 CFR Part 876.1500);
RegulatoryClass:2
Product Code:Primary: OCVSecondary: GCJ
PredicateDevices:Walterlorenz Surgical Assist Arm Scope Holder (K190576)
DeviceDescription:The KARL STORZ Holder is a manually operated surgical instrument intended to hold scopes (such as endoscopes,arthroscopes, etc) in a desired position during diagnostic, therapeutic, and surgical procedures (including neurologicprocedures). The holder is manufactured from Stainless Steel and functions by the user manually tightening or loosening aknob, which opens and closes around a surgical scope. The device is intended to attach to the several KARL STORZ holdingsystems, which are a Class 1, table-mounted, mechanical and electromechanical holding arms that have many generalsurgical functions in addition to scope holding; it can also be used for instrument holding, retraction, and positioning.
Indications ForUse:The KARL STORZ Holders are manually operated surgical devices intended to hold endoscopes (such as arthroscopes, etc)and accessories during general and neurologic diagnostic and therapeutic procedures.
SubstantialequivalenceThe similarities of the subject device to the predicate devices are as follows:
Subject Device – K240506KARL STORZ HoldersPredicate Device - K190576WalterLorenz Surgical Assist Arm ScopeHolder
Indications for UseThe KARL STORZ Holders are manuallyoperated surgical devices intended to holdendoscopes (such as arthroscopes, etc)and accessories during general andneurologic diagnostic and therapeuticprocedures.The WalterLorenz Surgical Assist Arm ScopeHolder is a manually operated surgical deviceintended to hold scopes (such as endoscopes,arthroscopes, etc) and accessories duringgeneral diagnostic and therapeutic procedures.The device is also intended to holdendoscopes during diagnostic and therapeuticneurologic procedures.
Material of constructionStainless SteelStainless Steel
Devices to be heldEndoscopes and other surgical instrumentsEndoscopes and other surgical instruments
Mode of operationManualSame as subject device
Range of scope diametersUp to 23mmUp to 10mm
Scratch Protection padsYesNo
Maximum loading weight3kgInformation not available
CleaningManual cleaningManual cleaning
Sterilization/ ReprocessingMethodsSteam SterilizationSteam Sterilization
Non-ClinicalPerformanceData:The following non-clinical testing were performed on the subject device:1) Holding Performance testing2) Direction of view/Image stability testing

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.