K Number
K240497

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-03-12

(21 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Incision, excision, ablation, vaporization, and coagulation of body soft tissue including intraoral tissue, in medical specialities including aesthetic (dermatology and plastic surgery), otolaryngology (ENT), gynaecology, neurosurgery, dental and oral surgery and genitourinary surgery. The use with the scanning unit is indicated for ablative skin resurfacing.

Device Description

The SMARTXIDE TETRA PRO DEVICE consists of a 10,600nm carbon dioxide laser (CO2) laser with 40W of maximum power. The laser energy is delivered to the treatment area via an articulated arm and a delivery accessory connected to its distal end. The articulated arm is an optical assembly that delivers free beam laser radiation. It is made up of seven mirrors placed on rotating knuckles: the mechanical accuracy of the articulated arm allows the CO2 laser beam to travel inside it and along its axis regardless of the arm orientation. The device may be used in combination with scanning units in order to achieve greater predictability and reproducibility. The use with the scanning units is indicated for layer-by-layer char-free ablation, enhancing the safety of the treatment with more uniform, accurate and controllable impact such as ablative skin resurfacing.

The overall weight is approximately 62 kg and the sizes are 42 x 122 x 54cm (W x H x D) with folded articulated arm.

Electrical requirements: 100-230V ~ 50/60Hz, 1200VA (max).

AI/ML Overview

This document is a 510(k) summary for a medical device called SMARTXIDE TETRA PRO, a CO2 laser surgical instrument. As such, it does not describe acceptance criteria and a study that proves the AI/ML device meets the acceptance criteria. The document explicitly states "Clinical Performance Data: None".

Instead, it focuses on demonstrating substantial equivalence to a predicate device (DEKA Smartxide family) through non-clinical performance data, primarily electrical safety, electromagnetic compatibility, and software validation.

To directly answer your request based on the provided text, the relevant information regarding acceptance criteria and device performance is as follows:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (from standards)Reported Device Performance (from testing)
Compliance with general requirements for basic safety and essential performance (ANSI AAMI ES60601-1)Electrical safety testing conducted and device meets requirements
Compliance with electromagnetic disturbances requirements (IEC 60601-1-2)EMC testing conducted and device meets requirements
Compliance with particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment (IEC 60601-2-22)Compliance testing conducted and device meets requirements
Compliance with safety of laser products - Equipment classification and requirements (IEC 60825-1)Compliance testing conducted and device meets requirements
Software verification and validation (as per FDA Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions")Software verification and validation testing conducted and documented

2. Sample size used for the test set and the data provenance: Not applicable. This document refers to non-clinical testing of the device hardware and software, not a study involving patient data or test sets in the context of an AI/ML device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for an AI/ML device is not relevant for the type of non-clinical testing described.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is not an AI-assisted device for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable to this type of device and testing. The "ground truth" for the non-clinical tests would be the specifications and requirements defined by the relevant electrical safety, EMC, and laser safety standards.

8. The sample size for the training set: Not applicable. This is not an AI/ML device, so there is no training set in the context of machine learning.

9. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square are the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" below it.

March 12, 2024

El.En. S.p.A. Paolo Peruzzi Regulatory Affairs Manager Via Baldanzese 17 Calenzano, FI 50141 Italy

Re: K240497

Trade/Device Name: Smartxide Tetra Pro Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: February 20, 2024 Received: February 20, 2024

Dear Paolo Peruzzi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).

{2}------------------------------------------------

Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Tanisha L. Hithe -S Tanisha L. Hithe -S 2024.03.12 09:07:47 -04'00'

Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K240497

Device Name

SMARTXIDE TETRA PRO

Indications for Use (Describe)

Incision, excision, ablation, vaporization, and coagulation of body soft tissue including intraoral tissue, in medical specialities including aesthetic (dermatology and plastic surgery), otolaryngology (ENT), gynaecology, neurosurgery, dental and oral surgery and genitourinary surgery. The use with the scanning unit is indicated for ablative skin resurfacing.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary SMARTXIDE TETRA PRO- Special 510(k)

510(k) Submission Number: K240497

Submitter:

El.En. S.p.A. Via Baldanzese, 17 50041 Calenzano (FI), Italy

Contact:

Paolo Peruzzi Regulatory Affairs Manager & Official Correspondent Phone: +39.055.8826807 E-mail: p.peruzzi@elen.it

Date Summary Prepared:

February 20th, 2024

Device Trade Name: SMARTXIDE TETRA PRO

Common Name: Medical Laser

Regulation Number: 21 CFR 878.4810

Regulation Name: Laser Surgical Instrument for use in General and Plastic and in Dermatology

Laser Surgical Instrument for use in General and Plastic and in Derma

Regulatory Class: Class II

Product Code: GEX

Predicate Device:

DEKA Smartxide family (Smartxide Touch, Smartxide Punto) (K180193)

Device Description:

The SMARTXIDE TETRA PRO DEVICE consists of a 10,600nm carbon dioxide laser (CO2) laser with 40W of maximum power. The laser energy is delivered to the treatment area via an articulated arm and a delivery accessory connected to its distal end. The articulated arm is an optical assembly that delivers free beam laser radiation. It is made up of seven mirrors placed on rotating knuckles: the mechanical accuracy of the articulated arm allows the CO2 laser beam to travel inside it and

{5}------------------------------------------------

along its axis regardless of the arm orientation. The device may be used in combination with scanning units in order to achieve greater predictability and reproducibility. The use with the scanning units is indicated for layer-by-layer char-free ablation, enhancing the safety of the treatment with more uniform, accurate and controllable impact such as ablative skin resurfacing.

The overall weight is approximately 62 kg and the sizes are 42 x 122 x 54cm (W x H x D) with folded articulated arm.

Electrical requirements: 100-230V ~ 50/60Hz, 1200VA (max).

Indications for Use:

Incision, excision, ablation, vaporization, and coagulation of body soft tissue including intraoral tissue, in medical specialities including aesthetic (dermatology and plastic surgery), otolaryngology (ENT), gynaecology, neurosurgery, dental and oral surgery and genitourinary surgery. The use with the scanning unit is indicated for ablative skin resurfacing.

{6}------------------------------------------------

Comparison with The Predicate Device:

Device Trade NameSubject DeviceSMARTXIDE TETRA PROPredicate DeviceK180193DEKA Smartxide family (Smartxide Touch,Smartxide Punto)Comment
Laser TypeCO2CO2Identical
Wavelength10.6μm10.6μmIdentical
Max Power40W30W, 50W (Smartxide Punto)60W (Smartxide Touch)In the range ofpredicate device
Handpieces Spot Sizes0.2 mm, 0.4 mm0.2 mm, 0.4 mmIdentical
Pulse Duration0.02 to 70 ms0.02 to 70 msIdentical
Pulse Rep Rate5 to 100 Hz5 to 100 HzIdentical
Scanner Spot size350μm350μmIdentical
Scanner Focal length (EFL)100 mm100 mmIdentical

{7}------------------------------------------------

Clinical Performance Data:

None

Non-Clinical Performance Data:

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the SMARTXIDE TETRA PRO device, according to the following standards:

. ANSI AAMI ES60601-1Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.

. IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests.

. IEC 60601-2-22: Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment

  • IEC 60825-1: Safety of laser products - Part 1: Equipment classification and requirements.

Software Validation and Verification Testing

Software verification and validation testing were conducted and documented as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions".

Conclusion:

On the basis of the comparison with the predicate device and on the non-clinical performance data, we can conclude that SMARTXIDE TETRA PRO is deemed to be substantially equivalent to the predicate device for the proposed indications for use.

Additional Information:

None

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.