K Number
K240431

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-07-24

(161 days)

Product Code
Regulation Number
892.5050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ExacTrac Dynamic is intended to position patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures, to monitor the patient position and to provide a beam hold signal in case of a deviation in order to treat lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Device Description

ExacTrac Dynamic (ETD) is a patient positioning and monitoring device used in a radiotherapy environment as an add-on system to standard linear accelerators (linacs). It uses radiotherapy treatment plans and the associated computed tomography (CT) data to determine the patient's planned position and compares it via oblique X-ray images to the actual patient position. The calculated correction shift will then be transferred to the treatment machine to align the patient correctly at the machine's treatment position. During treatment, the patient is monitored with a thermal-surface camera and X-ray imaging to ensure that there is no misalignment due to patient movement. Positioning and monitoring are also possible in combination with implanted markers. By defining the marker positions, ExacTrac Dynamic can position the patient by using X-rays and thereafter monitor the position during treatment.

Additionally, ExacTrac Dynamic features a breath-hold (BH) functionality to serve as a tool to assist respiratory motion management. This functionality includes special features and workflows to correctly position the patient at a BH level and thereafter monitor this position using surface tracking. Regardless of the treatment indication, a correlation between the patient's surface and internal anatomy must be evaluated with Image-Guided Radiation Therapy. The manually acquired X-ray images support a visual inspection of organs at risk (OARs). The aim of this technique is to treat the patient only during breath hold phases where the treatment target is at a certain position to reduce respiratory-induced tumor motion and to ensure a certain planned distance to OARs such as the heart. In addition to the X-ray based positioning technique, the system can also monitor the patient after external devices such as Cone-Beam CT (CBCT has been used to position the patient).

The ExacTrac Dynamic Surface (ETDS) is a camera-only platform without the X-ray system and is available as a configuration which enables surface-based patient monitoring. This system includes an identical thermal-surface camera, workstation, and interconnection hardware to the linac as the ETD system. The workflows supported by ETDS are surface based only and must be combined with an external IGRT device (e.g., CBCT).

AI/ML Overview

The FDA 510(k) summary for Brainlab AG's ExacTrac Dynamic (2.0) and ExacTrac Dynamic Surface provides information regarding its performance testing to demonstrate substantial equivalence to its predicate device, ExacTrac Dynamic 1.1 (K220338).

Here is a breakdown of the requested information based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state a table of "acceptance criteria" side-by-side with "reported device performance" values for all aspects of the device. However, it does reference the AAPM Task Group 1472 guidelines as a reference for the accuracy of surface-based monitoring and indicates that specific tests aimed to verify that accuracy specifications were not negatively affected. From the "Bench Tests" section, we can infer the objectives of the tests and how performance was evaluated.

Feature/TestAcceptance Criteria (Inferred/Referenced)Reported Device Performance
Rigid Body Surface Monitoring Accuracy TestFeasibility of surface-based patient monitoring for radiotherapy and adherence to AAPM Task Group 1472 guidelines for non-radiographic radiotherapy localization and positioning systems.The accuracy of the surface-based monitoring functionality was "checked using this new camera revision and an in-house phantom" and the goal was to "prove the feasibility". The document implies successful demonstration of feasibility and adherence.
Workflow & Accuracy Test ExacTrac DynamicAccuracy specifications for patient positioning and monitoring at phantom treatment with ExacTrac Dynamic are not affected by relevant conditions, settings, and workflows.The test was conducted to "verify that accuracy specifications... are not affected". The conclusion of substantial equivalence implies these specifications were met.
Response Time MeasurementImplicit: To measure the time between phantom movement and the "Beam-off" signal, and the "out of tolerance" signal appearance. No explicit numerical threshold is given in the provided text.The test "measures the time" and "is tracked." The conclusion of substantial equivalence implies acceptable response times.
Verification of the Radiation Isocenter Calibration in ETD 2.0Not inferior to the previous, well-established Radiation Isocenter Calibration in ETD 1.1 by more than a given threshold.The test was intended to "demonstrate that the Radiation Isocenter Calibration in ETD 2.0 is not inferior" within the specified threshold. The conclusion of substantial equivalence implies this was demonstrated.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The document mentions the use of an "in-house phantom" for the Rigid Body Surface Monitoring Accuracy Test and "phantom treatment" for the Workflow & Accuracy Test. It does not provide specific numerical sample sizes (e.g., number of phantom instances, number of trials).
  • Data Provenance: All testing appears to be retrospective (bench tests, phantom studies) and conducted internally (in-house phantom). There is no mention of data from human subjects or specific countries of origin.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This information is not provided in the document. The testing described focuses on device performance against physical standards (e.g., phantom, previous system performance) and referenced guidelines (AAPM Task Group 1472), rather than expert-established ground truth on clinical images.

4. Adjudication Method for the Test Set

This information is not provided in the document. Given the nature of the bench and phantom tests, an adjudication method by experts is not described as it would be for clinical image interpretation studies.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, an MRMC comparative effectiveness study was not reported. The document states, "No clinical testing was required for the subject device." The testing described focuses on the device's physical performance, accuracy, and workflow.

6. Standalone Performance (Algorithm Only without Human-in-the-Loop)

The described tests, specifically the "Rigid Body Surface Monitoring Accuracy Test," "Workflow & Accuracy Test ExacTrac Dynamic," "Response Time Measurement," and "Verification of the Radiation Isocenter Calibration," appear to evaluate the device's inherent performance characteristics, often in a controlled phantom environment. This implies a focus on the standalone capabilities of the system, even though its ultimate use is in assisting human operators in a radiotherapy setting. The "Beam-off signal" in response to movement is an automated system response, indicating standalone algorithmic functioning.

7. Type of Ground Truth Used

The ground truth for the performance tests appears to be:

  • Physical Phantoms: An "in-house phantom" for surface monitoring accuracy and "phantom treatment" for workflow and accuracy.
  • Established Reference System Performance: Comparison to the "previous, well-established Radiation Isocenter Calibration in ETD 1.1."
  • Industry Guidelines: Reference to "quality assurance guidelines for non-radiographic radiotherapy localization and positioning systems, that were defined by AAPM Task Group 1472."
  • Expected System Behavior: Verification of expected responses (e.g., "Beam-off signal") to phantom movement.

8. Sample Size for the Training Set

This information is not provided in the document. The 510(k) summary focuses on verification and validation testing, not the development or training of specific algorithms that would require a "training set" in the context of machine learning.

9. How the Ground Truth for the Training Set was Established

As no training set is mentioned (see point 8), this information is not applicable/provided.

{0}------------------------------------------------

July 24, 2024

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.

Brainlab AG % Sadwini Suresh QM Consultant Olof-Palme Str.9 Munich, 81829 GERMANY

Re: K240431

Trade/Device Name: ExacTrac Dynamic (2.0); ExacTrac Dynamic Surface Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: IYE Dated: February 13, 2024 Received: February 14, 2024

Dear Sadwini Suresh:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming

{1}------------------------------------------------

product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Loran Werther

Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K240431

Device Name

ExacTrac Dynamic (2.0); ExacTrac Dynamic Surface

Exactrac Dynamic Surface

Indications for Use (Describe)

ExacTrac Dynamic is intended to position patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures, to monitor the patient position and to provide a beam hold signal in case of a deviation in order to treat lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a stylized, abstract symbol in a bright pink color. The symbol appears to be composed of three stacked, curved shapes that resemble stylized snakes or ribbons. The shapes are arranged vertically, with each one slightly overlapping the one below it, creating a sense of depth and continuity.

510(k) Summary

July 24, 2024

General Information
ManufacturerBrainlab AGOlof-Palme-Str.9Munich, 81829Germany
Establishment Registration8043933
Device NameExacTrac Dynamic 2.0
Trade NameExacTrac Dynamic; ExacTrac Dynamic Surface
Classification NameMedical charged-particle radiation therapy system
Product CodeIYE
Regulation Number892.5050
Regulatory ClassII
PanelRadiology
Predicate Device and KNumberK220338; ExacTrac Dynamic 1.1
Contact Information
Primary ContactAlternate Contact
Sadwini SureshRegulatory Affairs Brainlab
QM ConsultantPhone: +49 89 99 15 68 0
Phone: +49 89 99 15 68 0Fax: +49 89 99 15 68 5033
Email: regulatory.affairs@brainlab.comEmail: regulatory.affairs@brainlab.com

1. Indications for Use

ExacTrac Dynamic is intended to position patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures, to monitor the patient position and to provide a beam hold signal in case of a deviation in order to treat lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

2. Device Description

ExacTrac Dynamic (ETD) is a patient positioning and monitoring device used in a radiotherapy environment as an add-on system to standard linear accelerators (linacs). It uses radiotherapy treatment plans and the associated computed tomography (CT) data to determine the patient's planned position and compares it via oblique X-ray images to the actual patient position. The calculated correction shift will then be transferred to the treatment machine to align the patient correctly at the machine's treatment position. During treatment, the patient is monitored with a thermal-surface camera and X-ray imaging to

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a stylized, abstract symbol in a vibrant pink color. The symbol appears to be composed of three interconnected, flowing shapes, possibly resembling stylized letters or decorative elements. The shapes are stacked vertically, creating a unified design that is eye-catching and distinctive.

ensure that there is no misalignment due to patient movement. Positioning and monitoring are also possible in combination with implanted markers. By defining the marker positions, ExacTrac Dynamic can position the patient by using X-rays and thereafter monitor the position during treatment.

Additionally, ExacTrac Dynamic features a breath-hold (BH) functionality to serve as a tool to assist respiratory motion management. This functionality includes special features and workflows to correctly position the patient at a BH level and thereafter monitor this position using surface tracking. Regardless of the treatment indication, a correlation between the patient's surface and internal anatomy must be evaluated with Image-Guided Radiation Therapy. The manually acquired X-ray images support a visual inspection of organs at risk (OARs). The aim of this technique is to treat the patient only during breath hold phases where the treatment target is at a certain position to reduce respiratory-induced tumor motion and to ensure a certain planned distance to OARs such as the heart. In addition to the X-ray based positioning technique, the system can also monitor the patient after external devices such as Cone-Beam CT (CBCT has been used to position the patient).

The ExacTrac Dynamic Surface (ETDS) is a camera-only platform without the X-ray system and is available as a configuration which enables surface-based patient monitoring. This system includes an identical thermal-surface camera, workstation, and interconnection hardware to the linac as the ETD system. The workflows supported by ETDS are surface based only and must be combined with an external IGRT device (e.g., CBCT).

3. Substantial Equivalence

For the Substantial Equivalence determination, comparison of the Subject Device features with the following predicate device(s) was carried out:

ExacTrac Dynamic 1.1; K220338

The predicate was chosen since it's the predecessor and similar to the subject device with respect to the indications for use, technological characteristics and use cases.

The main differences compared to the predicate device is providing the 'Surface Only' Workflow. The ExacTrac Dynamic Surface (ETDS) is a camera-only platform without the Xray system and is available as a configuration which enables surface-based patient monitoring. This system includes an identical thermal-surface camera, workstation, and interconnection hardware to the linac as the ETD system. The workflows supported by ETDS are surface based only and must be combined with an external IGRT device (e.g., CBCT).

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Brainlab logo. The logo consists of a stylized caduceus symbol on the left and the word "BRAINLAB" in capital letters on the right. The color of the logo is pink.

Topic/ FeaturePredicate DeviceSubject Device
Indications for useExacTrac Dynamic is intended to position patients at anaccurately defined point within the treatment beam of a medicalaccelerator for stereotactic radiosurgery or radiotherapyprocedures, to monitor the patient position and to provide abeam hold signal in case of a deviations in order to treatlesions, tumors and conditions anywhere in the body whenradiation treatment is indicated.ExacTrac Dynamic is intended to position patients atan accurately defined point within the treatment beamof a medical accelerator for stereotactic radiosurgeryor radiotherapy procedures, to monitor the patientposition and to provide a beam hold signal in case ofa deviations in order to treat lesions, tumors andconditions anywhere in the body when radiationtreatment is indicated.
Localizationtechnique basedon:The camera detects both the patient surface and the patientthermal surface which together, can be used to track thepatient geometries.Stereo X-ray is acquired and compared with the plannedposition (room based).CBCT data are imported from a from 3rd party CBCT Deviceand compared with the planned position.ExacTrac Dynamic (ETD) :The camera detects boththe patient surface and the patient thermal surfacewhich together, can be used to track the patientgeometries.Stereo X-ray is acquired and compared with theplanned position (room based).CBCT data are imported from a from 3rd party CBCTDevice and compared with the planned position.ExacTrac Dynamic Surface (ETDS):The camera detects both the patient surface and thepatient thermal surface which together, can be usedto track the patient geometries/ position.No X-Ray acquisition required.
Topic/ FeaturePredicate DeviceSubject Device
ETD Preparationand ReviewworkflowPreparation:• Import of RT Dicom• Define patient related settings• Define TemplateReview• Review of treatment• Approval of treatmentPreparation:• Import of RT Dicom• Define patient related settings• Define TemplateReview• Review of treatment• Approval of treatmentAdditional minor Bugfixes
User ManagementsetupDone via Windows user management according to ourinstructionsDone via Windows user management according toour instructionsAdditional minor Bugfixes

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a stylized, abstract symbol in a bright pink color. The symbol appears to be composed of three stacked elements, each resembling a simplified, curved shape. The overall design has a flowing, organic feel, and the bright pink color makes it stand out against a white background.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows a pink symbol that appears to be a stylized representation of three snakes stacked on top of each other. The snakes are facing to the left and have a curved shape. The symbol is simple and abstract, with no other details or background elements present.

Topic/ FeaturePredicate DeviceSubject Device
Deep InspirationBreath-Hold (DIBH /Breath Hold (BH)workflowDeep Inspiration Breath-Hold (DIBH) functionality to assistrespiratory motion management. This functionality includesspecial features and workflows to correctly position thepatient to breath-hold level and then to monitor this positionusing the ExacTrac surface tracking and x-ray positioningtechnology. During breath-Hold the Linac Beam is enabled.While the patient is not in position the beam of the LINAC willbe hold.The aim of this technology is to treat the patient only duringbreath-hold phases where the breast is at a defined positionwith a maximum distance to critical structures like the heart.ExacTrac Dynamic:Additionally, ExacTrac Dynamic 2.0 extended thedeep inspiration breath hold to Breath-Hold (BH)functionality to treat breast and thorax cancerpatients.Additional minor BugfixesExacTrac Dynamic Surface:Workflow adaption to make a BH with externalimaging for positioning and Surface Tracking Unitfor monitoringAdditional minor Bugfixes
Topic/ FeaturePredicate DeviceSubject Device
Surface OnlyworkflowThe ExacTrac Dynamic Surface Only workflow enablespatient monitoring by surface tracking after 3rd partypositioning, e.g. by CBCT, without using X-ray images asreference or for monitoring. X-ray imaging is disabled withinthis workflow.ExacTrac Dynamic:N/A, no change to ETD 1.1Additional minor BugfixesExacTrac Dynamic SurfaceThe Surface Only workflow is provided as forExacTrac Dynamic 1.1 with the followingmodifications:- Surface tracking during 3rd Party Positioning isprovided to set the reference point for thetreatment surface monitoring- Non-coplanar treatment angles are supporteduntil a limited range (Only up to +-30°)- Prepositioning can be based on surfacereference from previous treatment sessionsThe workflow can be executed on a HW platform notproviding X-ray imaging components. For this setupa specific calibration function is introduced.
Patient FeedbackGlassesN/APatient Feedback Glasses provide an alternativesolution to the Patient Feedback System to allow thepatient to see the room monitor and on it the liverespiratory status and the BH Gating Window.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows a stylized, abstract symbol in a vibrant pink color. The symbol appears to be composed of three identical, curved shapes stacked vertically on top of each other. The shapes resemble a stylized letter or a decorative flourish, with a flowing, calligraphic quality. The overall impression is one of elegance and artistic design.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows a stylized symbol in a bright pink color. The symbol appears to be composed of three curved shapes stacked vertically on top of each other. The shapes are abstract and do not resemble any specific object or letter, giving the symbol a unique and potentially logo-like appearance. The background is plain white, which makes the pink symbol stand out prominently.

4. Performance Data

The Subject Device has been verified and validated according to Brainlab processes for product design and development. A high level explanation of the testing provided in this submission for the Subject Device is provided below.

Software verification

Software verification and validation testing has been conducted and documentation is provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Device Software Functions". These include successful implementation of product specifications, incremental testing for different release candidates, testing of risk control measures, compatibility testing or cybersecurity tests. The documentation submitted is for enhanced level.

Bench Tests

  • -RIGID BODY SURFACE MONITORING ACCURACY TEST:
    As ExacTrac Dynamic 2.0 uses a new revision of the ETD 3D Camera A5060 (45401B) with a new LED, the accuracy of the surface based monitoring functionality of ExacTrac Dynamic 2.0 was checked using this new camera revision and an in-house phantom. The goal of the test was to prove the feasibility of the surface based patient monitoring of ExacTrac Dynamic 2.0 for use in radiotherapy. The quality assurance guidelines for nonradiographic radiotherapy localization and positioning systems, that were defined by AAPM Task Group 1472, shall be used as reference.

  • । WORKFLOW & ACCURACY TEST EXACTRAC DYNAMIC:
    This test is to verify that accuracy specifications for patient positioning and monitoring at phantom treatment with ExacTrac Dynamic are not affected by relevant conditions, settings and workflows.

  • -RESPONSE TIME MEASUREMENT:
    The test measures the time between phantom movement and the correspondingly created Beam-off signal. The time between phantom movement and the appearance of the "out of tolerance" signal on the monitor is tracked.

  • -VERIFICATION OF THE RADIATION ISOCENTER CALIBRATION IN ETD 2.0:
    This test is intended to demonstrate that the Radiation Isocenter Calibration in ETD 2.0 is not inferior to the previous, well-established Radiation Isocenter Calibration in ETD 1. * by more than a given threshold.

No clinical testing was required for the subject device.

5. Conclusion

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows a stylized, abstract symbol in a vibrant pink color. The symbol appears to be composed of three stacked elements, each resembling a curved or flowing shape, possibly inspired by calligraphy or a stylized letter. The overall design is compact and eye-catching, with a sense of movement and fluidity conveyed through the curved lines.

The comparison of the Subject Device with the predicate device shows that it has similar functionality, intended use and technological characteristics as the predicate device. Based on the comparison to the predicate and the performance testing conducted, the Subject Device is considered substantially equivalent to the predicate device.

§ 892.5050 Medical charged-particle radiation therapy system.

(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.