K Number
K240247

Validate with FDA (Live)

Date Cleared
2024-07-12

(164 days)

Product Code
Regulation Number
876.5885
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MoFi is a liquid culture medium intended for human ex vivo tissue and cell culture processing applications.

Device Description

MoFi is a basal medium based on Iscove's Modified Dulbecco's Medium (IMDM) and cell-degradabe polysaccharide, made of water, carbohydrate, amino acid and salt. It can be mixed with various serums or additional additives to form a complete medium, which can be used to culture suspension/adherent primary cultured cells or cell lines. MoFi provides cells with energy to maintain survival and promote growth.

AI/ML Overview

The provided document describes the FDA 510(k) premarket notification for "MoFi" Cell Culture Basal Medium. This K-submission focuses on demonstrating substantial equivalence to a predicate device, rather than providing details on a study performed by the applicant (DuoGenic StemCells Corporation) to prove performance against specific acceptance criteria for an AI/ML powered device. The document describes a biologically-based product (cell culture medium), not an AI/ML-powered medical device.

Therefore, the requested information regarding acceptance criteria, AI/ML study design (sample sizes, ground truth establishment, expert adjudication, MRMC studies, standalone performance), and training set details for an AI/ML device cannot be extracted from this document as it does not pertain to such a device.

The document discusses the following performance characteristics for the MoFi Cell Culture Basal Medium:

1. Acceptance Criteria and Reported Device Performance (for a cell culture medium, not an AI/ML device):

Acceptance Criteria (for MoFi Medium)Reported Device Performance (for MoFi Medium)
Appearance: Transparent yellowish liquidMet (as implied by successful testing)
pH: 7.0-8.0Met (as implied by successful testing)
Cell viability: Good cell viabilityGood cell viability of four kinds of cells (suspension primary culture cells, suspension cell lines, adherent primary culture cells, and adherent cell lines)
Sterility: Negative or not detectedNegative or not detected
Endotoxin: < 1 EU/mL< 1 EU/mL (well controlled)
Mycoplasma: Negative or not detectedNegative or not detected
Residues of plasticizers, heavy metals and particlesNot detected or can be safely ignored
Product Performance: Good cell viability and normal cell morphology of four kinds of cellsGood cell viability and normal cell morphology of four kinds of cells (suspension primary culture cells, suspension cell lines, adherent primary culture cells and adherent cell lines)

2. Sample Size and Data Provenance (for evaluating cell culture medium, not an AI/ML device):

  • The document states "All batches tested met the acceptance criteria of performance testing." It does not specify the exact number of batches or the sample size (e.g., number of cell cultures) per test.
  • Data provenance is not explicitly stated in terms of country of origin or whether it was retrospective/prospective, but it is implied to be internal testing conducted by DuoGenic StemCells Corporation for regulatory submission.

3. Number of Experts and Qualifications:

  • This information is not applicable as the evaluation is for a cell culture medium's biological performance, not an AI/ML algorithm requiring expert interpretation.

4. Adjudication Method:

  • Not applicable for this type of product evaluation.

5. Multi-reader Multi-case (MRMC) Comparative Effectiveness Study:

  • Not applicable as this is not an AI/ML-powered device.

6. Standalone Performance (Algorithm Only):

  • Not applicable as this is not an AI/ML algorithm.

7. Type of Ground Truth Used:

  • For the cell culture medium, the "ground truth" for performance is based on established biological and chemical assays, such as pH measurement, sterility testing, endotoxin detection, and direct observation of cell viability and morphology compared to established controls and the predicate device.

8. Sample Size for Training Set:

  • Not applicable as this is not an AI/ML device requiring a training set.

9. How Ground Truth for Training Set Was Established:

  • Not applicable as this is not an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.

July 12, 2024

DuoGenic StemCells Corporation Hong-Lin Su Chairman No. 18, Ln. 10, Taiping 21st St., Taiping Dist. Taichung City, Taiwan 41161 Taiwan

Re: K240247

Trade/Device Name: "MoFi" Cell Culture Basal Medium Regulation Number: 21 CFR 876.5885 Regulation Name: Tissue Culture Media For Human Ex Vivo Tissue And Cell Culture Processing Applications Regulatory Class: Class II Product Code: NDS Dated: June 14, 2024 Received: June 14, 2024

Dear Hong-Lin Su:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Maura Rooney -S

Maura Rooney Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K240247

Device Name

"MoFi" Cell Culture Basal Medium

Indications for Use (Describe)

MoFi is a liquid culture medium intended for human ex vivo tissue and cell culture processing applications.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

"MoFi" Cell Culture Basal Medium

510(k) Premarket Notification

Ch. 5 – 510(k) Summary

Date: 03/07/2024

In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CRF 872, this information serves as a Summary of Safety and Effectiveness for the use of the "MoFi" Cell Culture Basal Medium (hereinafter referred to as MoFi).

Submitted byDuoGenic StemCells corporation
Date2024/07/03
Contact PersonChih-Yao Lin
Applicant AddressNo. 18, Ln. 10, Taiping 21st St., Taiping Dist.,Taichung City, Taiwan (R.O.C.)
Telephone Number+886-4-2276-6193
Proprietary Name"MoFi" Cell Culture Basal Medium
Common NameMoFi
Device Trade NameMoFi
Review PanelGastroenterology/Urology
Classification Name and Reference876.5885 Tissue culture media for human ex vivotissue and cell culture processing applications. ClassII (special controls).
Device Product Code and Panel CodeNDS
Predicate DeviceKnockoutTM SR Medium and KnockoutTM SRXenofree Medium
Predicate Submission NumberK100616

1. Intended Use and Indications for Use

MoFi is a liquid culture medium intended for human ex vivo tissue and cell culture processing applications.

2. Device Description

2.1 Description:

MoFi is a basal medium based on Iscove's Modified Dulbecco's Medium (IMDM) and cell-degradabe polysaccharide, made of water, carbohydrate, amino acid and salt. It can be mixed with various serums or additional additives to form a complete medium, which can be used to culture suspension/adherent primary cultured cells or cell lines. MoFi provides cells with energy to maintain survival and promote growth.

  • 2.2 Storage condition: 2 - 6°C
  • 2.3 Shelf life: 9 months (from the manufactured date)
  • 2.4 Volume: 80 mL/bottle

{5}------------------------------------------------

2.5 Instruction:

  • 2.5.1 The preparation of full medium is recommended according to the cell type. Please see the table below. Antibiotics may be added if needed.
  • 2.5.2 Temperature and time for pre-warming can be adjusted according to user's need, while it's recommended not to exceed 37°C.
  • 2.5.3 This product is recommended to be handled aseptically.
ConcentrationReagentPeripheral Blood Cells(suspension cells)Mesenchymal Stem Cells(adherent cells)
MoFi1X1X
Serum (human or animal)10%20%
L-glutamine2 mM2 mM
Sodium pyruvate1 mM1 mM
Non-essential amino acidsNot applicable (N/A)0.1 mM
Fibroblast growth factor 2N/A4 ng/mL

2.6 Warning and Precautions:

  • 2.6.1. Protective glasses, clothing and gloves should be worn during operation.
  • 2.6.2. This product is for single use, and should not be re-sterilized and reused.
  • 2.6.3. This product should be handled aseptically, and should follow biosafety instructions.
  • 2.6.4. This product should be protected from light.
  • 2.6.5. This product should not be stored or operated at over 37°C.
  • 2.6.6. Please do NOT inject this product into the human body, or allergic reactions may occur.
  • 2.6.7. When used as a medical device, Federal Law restricts this device to sale by or on the order of a physician.

3. Device Specification

  • Appearance: Transparent yellowish liquid 3.1
  • 3.2 pH: 7.0-8.0
  • 3.3 Cell viability
  • 3.4 Sterility: Negative or not detected
  • 3.5 Endotoxin: < 1 EU/mL
  • 3.6 Mycoplasma: Negative or not detected

{6}------------------------------------------------

4. Substantial Equivalence Information

Predicate Device: Knockout™ SR Medium and Knockout™ SR Xenofree Medium

Manufacturer: Life Technologies Corporation

ItemNew device (applicant)Predicate device
Regulationnumber &product code876.5885NDS876.5885NDS
510(k)NumberK240247K100616
Intended UseA liquid culture medium intended forhuman ex vivo tissue and cell cultureprocessing applicationsA liquid tissue culture media productintended for human ex vivo tissue and cellculture processing applications
Comparison: Both MoFi and the predicate are used for human ex vivo cell and tissue processing applications.
Container/packagingBottle: polyethylene terephthalatecopolyester, glycol modified (PETG)2/O2Closure: high-density polyethylene (HDPE)Bottle: polyethylene terephthalate (PET)Closure: high-density polyethylene (HDPE)
Comparison: PET and PETG are both suitable containers for cell culture and media storage applications.
ProductConfigurationAmino acids, salts, vitamins andcarbohydratesAmino acids, salts, vitamins andcarbohydrates
Comparison: Both MoFi and the predicate are chemically-defined culture media and composed of amino acids, salts, vitamins and carbohydrates.
AsepticallyProcessedCompliance with GMP requirementsregarding aseptic processingCompliance with GMP requirementsregarding aseptic processing
Comparison: The process validation and environmental monitoring of the MoFi process support aseptic processing. Both MoFi and the predicate comply with GMP requirements regarding aseptic processing.
StorageTemperature2-6°C2-8°C
Comparison: The difference in storage temperature between MoFi and the predicate does not affect the safety of the product.
Shelf-life9 months14 months
Comparison: According to the stability data, shelf life of MoFi is recommended to be 9 months.
ProductPerformanceGood cell viability and normal cellmorphology of four kinds of cells(suspension primary culture cells,suspension cell lines, adherent primaryculture cells and adherent cell lines)Good cell viability and normal cellmorphology of four kinds of cells(suspension primary culture cells,suspension cell lines, adherent primaryculture cells and adherent cell lines)
Comparison: Cells are roughly categorized into four groups, suspension primary culture cells, suspension cell lines, adherent primary culture cells and adherent cell lines. The results showed that both KSR and MoFi passed the acceptance criteria of cell viability, and the cell morphology of four kinds of cells cultured in both MoFi and the predicate is normal and typical. This indicates that MoFi possess an equivalent function as the predicate and no safety issues occur

{7}------------------------------------------------

"MoFi" Cell Culture Basal Medium

510(k) Premarket Notification

5. Shelf Life

According to the stability test's result, shelf life for MoFi is recommended to be 9 months.

6. Biocompatibility

The biocompatibility evaluation was conducted in accordance with the FDA Guidance "Use of International Standard ISO 10993-1,"Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.

7. Performance Testing

All batches tested met the acceptance criteria of performance testing. The residual endotoxin is well controlled. No bacteria and mycoplasma were detected. The residues of plasticizers, heavy metals and particles are not detected or can be safely ignored.

8. Conclusion

MoFi and Knockout™ SR Medium and Knockout™ SR Xenofree Medium are used for human ex vivo tissue and cell culture processing applications and have the same/similar container, storage temperature, efficacy (generic cellular growth and maintenance) and safety (consistency in chemical content and formulation, biocompatibility with cells, and purity). Their efficacy in supporting the survival, growth, development, and maintenance of human cells or tissue culture systems has been well established in scientific publications included in this submission. These products (MoFi and Knockout™ SR Medium and Knockout™ SR Xenofiree Medium) are manufactured in accordance with QSR requirements and are labeled as aseptically processed. Thus, MoFi are substantially equivalent to the legally marketed device (K100616) intended for the human ex vivo tissue and cell culture processing applications.

§ 876.5885 Tissue culture media for human ex vivo tissue and cell culture processing applications.

(a)
Identification. Tissue culture media for human ex vivo tissue and cell culture processing applications consist of cell and tissue culture media and components that are composed of chemically defined components (e.g., amino acids, vitamins, inorganic salts) that are essential for the ex vivo development, survival, and maintenance of tissues and cells of human origin. The solutions are indicated for use in human ex vivo tissue and cell culture processing applications.(b)
Classification. Class II (special controls): FDA guidance document, “Class II Special Controls Guidance Document: Tissue Culture Media for Human Ex Vivo Processing Applications; Final Guidance for Industry and FDA Reviewers.”