K Number
K240217

Validate with FDA (Live)

Date Cleared
2025-01-17

(357 days)

Product Code
Regulation Number
866.3393
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The cobas liat CT/NG nucleic acid test is an automated, qualitative in vitro nucleic acid diagnostic test that utilizes realtime polymerase chain reaction (PCR) for the direction of Chlamydia (CT) and Neisseria gonorthoeae (NG) nucleic acid in male urine and vaginal swabs, all in cobas PCR Media (Roche Molecular Systems, Inc.).

This test is intended as an aid in the diagnosis of urogenital infections in both symptomatic individuals.

Device Description

The test is performed on the cobas® liat analyzer which automates and integrates sample purification, nucleic acid amplification, and detection of the target sequence in biological samples using real-time PCR assays. The assay targets both the Cryptic plasmid and 23S rRNA of Chlamydia trachomatis and the pivNG and NGR9 of Neisseria gonorrhoeae. An Internal Control (IC) is also included. The IC is present to control for adequate processing of the target bacteria through steps of sample purification, nucleic acid amplification, and to monitor the presence of inhibitors in the PCR processes.

AI/ML Overview

Here's a summary of the acceptance criteria and study details for the cobas® liat CT/NG nucleic acid test, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document primarily provides performance metrics rather than explicitly stated acceptance criteria with numerical targets. However, based on the demonstrated performance and the context of a 510(k) submission, the implicit acceptance criteria would be high sensitivity and specificity, indicating reliable detection of CT and NG infections.

Metric (Implicit Acceptance Criteria)Device Performance - CT (Male Urine)Device Performance - CT (Vaginal Swabs)Device Performance - NG (Overall Male Urine)Device Performance - NG (Overall Vaginal Swabs)
Sensitivity / Positive Percent Agreement (PPA)97.3% (92.4%, 99.1%)98.2% (93.6%, 99.5%)100.0% (97.7%, 100.0%)97.7% (92.0%, 99.4%)
Specificity / Negative Percent Agreement (NPA)99.9% (99.7%, 100.0%)99.8% (99.5%, 99.9%)99.9% (99.6%, 100.0%)99.8% (99.6%, 99.9%)
Reproducibility (Low Positive - 1-2x LoD)CT: 90.7%CT: 100%NG: 99.6%NG: 100%
Reproducibility (Moderate Positive - 3-5x LoD)CT: 96.3%CT: 100%NG: 100%NG: 100%
Reproducibility (Negative)CT: 100%CT: 100%NG: 100%NG: 100%
Analytical Sensitivity (LoD) - CT Serovar DUrine: 0.085 EB/mLVaginal Swab: 0.170 EB/mLN/AN/A
Analytical Sensitivity (LoD) - CT Serovar IUrine: 0.784 EB/mLVaginal Swab: 0.784 EB/mLN/AN/A
Analytical Sensitivity (LoD) - NG Strain 2948Urine: 0.250 CFU/mLVaginal Swab: 0.500 CFU/mLN/AN/A
Analytical Sensitivity (LoD) - NG Strain 891Urine: 0.200 CFU/mLVaginal Swab: 0.200 CFU/mLN/AN/A
Invalid Rate (Initial Test)0.6%0.6%0.6%0.6%
Invalid Rate (After Retesting)0.2%0.2%0.2%0.2%

2. Sample Size and Data Provenance

  • Clinical Study Test Set (Prospectively collected):
    • Total Evaluated Subjects: 4780 (2304 males, 2476 females)
    • Male Urine Specimens: 2302 (from 2302 male subjects)
    • Vaginal Swabs: 2476 (1240 clinician-collected, 1236 self-collected from 2476 female subjects)
    • Data Provenance: Multi-site, prospective study collected at 13 geographically diverse clinical sites across the US.
  • Clinical Study Test Set (Archived Specimens - Supplementation):
    • Archived Male Urine Specimens: 163
    • Archived Vaginal Swabs: 90
    • Data Provenance: Prospectively collected samples from a prior clinical trial (K173887).
  • Reproducibility Study Test Set: Total 1618 tests (811 vaginal, 807 urine) across 3 external sites. Each panel member tested in triplicate. Low positive (1-2x LoD), moderate positive (3-5x LoD), and negative panel members used.
  • Supplemental Precision Study (for CT in urine): 810 evaluable tests on urine panel members (negative, 1x-2x LoD, 3x-5x LoD).

3. Number of Experts and Qualifications for Ground Truth

The ground truth for the clinical study was established using a Patient Infected Status (PIS) or Composite Comparator Algorithm (CCA), which relied on a combination of three FDA-cleared NAATs (NAAT1, NAAT2, and NAAT3). The document does not specify the number of human experts used to establish the ground truth or their qualifications for the clinical study. The "ground truth" was algorithmically derived from the results of the comparator NAATs.

4. Adjudication Method for the Test Set

The adjudication method for the clinical study ground truth (PIS/CCA) followed a rule-based algorithm:

  • If NAAT1 and NAAT2 were concordant, that result was the final PIS/CCA.
  • If NAAT1 and NAAT2 were discordant, NAAT3 was performed as the tiebreaker.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not done. This study assesses the performance of a diagnostic test (the cobas® liat CT/NG nucleic acid test), which is an automated, qualitative in vitro nucleic acid diagnostic test. It replaced human assessment with an automated process, and the comparison was against a PIS/CCA derived from other reference NAATs, not human readers with and without AI assistance. Therefore, there is no effect size for human readers improving with AI.

6. Standalone (Algorithm Only) Performance

  • Yes, a standalone (algorithm only) performance study was done. The entire clinical performance evaluation, reproducibility studies, and analytical studies assess the performance of the cobas® liat CT/NG nucleic acid test itself, which is an automated device performing real-time PCR. It is designed to operate without human intervention beyond sample loading and results interpretation from the automated output.

7. Type of Ground Truth Used

  • Clinical Study: Patient Infected Status (PIS) or Composite Comparator Algorithm (CCA) derived from the concordant results of FDA-cleared Nucleic Acid Amplification Tests (NAATs).
  • Analytical Studies (LoD, Inclusivity, Specificity, Interference): Known concentrations of specific strains or culture subtypes of bacteria/viruses, spiked into negative clinical specimens.

8. Sample Size for the Training Set

The document does not explicitly describe a separate "training set" for an AI/ML model for the cobas® liat CT/NG nucleic acid test. As a nucleic acid diagnostic test (real-time PCR), it operates based on established biochemical principles and does not typically involve machine learning training in the same way an imaging AI algorithm would. All the data presented is for validation and performance evaluation.

9. How Ground Truth for the Training Set Was Established

Since there is no explicitly mentioned "training set" for an AI/ML model in this context, the method for establishing ground truth for such a set is not applicable or described. The clinical performance is evaluated against a PIS/CCA derived from other NAATs, and analytical performance is against known concentrations.

{0}------------------------------------------------

January 17, 2025

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with an abstract design. To the right of the seal, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG" are written in blue, with the word "ADMINISTRATION" written in a smaller font size below.

Roche Molecular Systems, Inc. Deborah Leu Regulatory Affairs Project Manager 4300 Hacienda Drive Pleasanton, California 94588

Re: K240217

Trade/Device Name: cobas liat CT/NG nucleic acid test Regulation Number: 21 CFR 866.3393 Regulation Name: Device To Detect Nucleic Acids From Non-Viral Microorganism(S) Causing Sexually Transmitted Infections And Associated Resistance Marker(S) Regulatory Class: Class II Product Code: QEP, LSL, MKZ Dated: January 25, 2024 Received: January 26, 2024

Dear Deborah Leu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Himani Bisht -S

Himani Bisht, Ph.D. Assistant Director Viral Respiratory and HPV Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K240217

Device Name cobas liat CT/NG nucleic acid test

Indications for Use (Describe)

The cobas liat CT/NG nucleic acid test is an automated, qualitative in vitro nucleic acid diagnostic test that utilizes realtime polymerase chain reaction (PCR) for the direction of Chlamydia (CT) and Neisseria gonorthoeae (NG) nucleic acid in male urine and vaginal swabs, all in cobas PCR Media (Roche Molecular Systems, Inc.).

This test is intended as an aid in the diagnosis of urogenital infections in both symptomatic individuals.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

cobas® liat CT/NG nucleic acid test 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

Submitter NameRoche Molecular Systems, Inc.
Address4300 Hacienda Drive,Pleasanton, CA 94588-2722
ContactDeborah LeuPhone: 925-523-8362Email: deborahleu@roche.com
Date PreparedJanuary 15, 2025
Proprietary Namecobas® liat CT/NG nucleic acid test
Common Namecobas® liat CT/NG
Classification NameNucleic Acid Detection System For Non-Viral Microorganism(S) CausingSexually Transmitted InfectionsDNA probe, Nucleic Acid Amplification, ChlamydiaNeisseria spp. direct serological test reagents
Product CodesQEPMKZLSL
Predicate Devicescobas® 6800/8800 CT/NG
Establishment RegistrationRoche Molecular Systems, Inc. (2243471)

{5}------------------------------------------------

DEVICE DESCRIPTION 1.

The test is performed on the cobas® liat analyzer which automates and integrates sample purification, nucleic acid amplification, and detection of the target sequence in biological samples using real-time PCR assays. The assay targets both the Cryptic plasmid and 23S rRNA of Chlamydia trachomatis and the pivNG and NGR9 of Neisseria gonorrhoeae. An Internal Control (IC) is also included. The IC is present to control for adequate processing of the target bacteria through steps of sample purification, nucleic acid amplification, and to monitor the presence of inhibitors in the PCR processes.

2. INDICATIONS FOR USE

The cobas® liat CT/NG nucleic acid test is an automated, qualitative in vitro nucleic acid diagnostic test that utilizes real-time polymerase chain reaction (PCR) for the direct detection of Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (NG) nucleic acid in male urine and vaginal swabs, all in cobas® PCR Media (Roche Molecular Systems, Inc.).

This test is intended as an aid in the diagnosis of urogenital infections in both symptomatic and asymptomatic individuals.

TECHNOLOGICAL CHARACTERISTICS 3.

The primary technological characteristics and intended use of the RMS cobas® liat CT/NG nucleic acid test are substantially equivalent to other legally marketed nucleic acid amplification tests intended for the qualitative detection of CT and NG.

As indicated in Table 1, the RMS cobas® liat CT/NG nucleic acid test is substantially equivalent to significant characteristics of the identified predicate device, the currently cleared cobas® 6800/8800 CT/NG (K173887) for use on cobas® 6800/8800 Systems.

{6}------------------------------------------------

Submitted Device:cobas® liat CT/NG nucleic acid testPredicate Device:cobas® 6800/8800 CT/NG for use on cobas®6800/8800 Systems.
Regulation Name866.3393866.3120866.3390866.3390866.3120862.2570
Product CodeQEPMKZLSLLSLMKZOOI
Intended UseThe cobas® liat CT/NG nucleic acid testis an automated, qualitative in vitro nucleicacid diagnostic test that utilizes real-timepolymerase chain reaction (PCR) for thedirect detection of Chlamydia trachomatis(CT) and Neisseria gonorrhoeae (NG)nucleic acid in male urine and vaginalswabs, all in cobas® PCR Media (RocheMolecular Systems, Inc.).This test is intended as an aid in thediagnosis of urogenital infections in bothsymptomatic and asymptomaticindividuals.The cobas® CT/NG on the cobas® 6800/8800system is an automated, qualitative in vitronucleic acid diagnostic test, that utilizes real-timepolymerase chain reaction (PCR), for the directdetection of Chlamydia trachomatis (CT) and/orNeisseria gonorrhoeae (NG) DNA in male andfemale urine, clinician-instructed self-collectedvaginal swab specimens (collected in a clinicalsetting), clinician-collected vaginal swabspecimens, and endocervical swab specimens,all collected in cobas® PCR Media (RocheMolecular Systems, Inc.), and cervical specimenscollected in PreservCyt® solution. This test isintended as an aid in the diagnosis of chlamydialand gonococcal disease in both symptomatic andasymptomatic individuals.
Sample TypeMale and female urine, vaginal swabsMale and female urine,Self-collected/clinician-collected vaginal swabspecimens in cobas® PCR Media,Endocervical swab specimens in cobas® PCRMedia,Cervical specimens in PreservCyt® solution.
Analyte TargetsChlamydia trachomatis (CT)Neisseria gonorrhoeae (NG)Chlamydia trachomatis (CT),Neisseria gonorrhoeae (NG)
Ancillary CollectionKitscobas® PCR Urine Sample Kitcobas® PCR Media Uni Swab Sample Kitcobas® PCR Media Dual Swab Sample Kitcobas® PCR Media Uni Swab Sample Kitcobas® PCR Urine Sample Kit
Sample PreparationAutomatedSame
AmplificationTechnologyReal-time PCRSame
DetectionChemistryAssay using different reporter dyes fortarget and controlPaired reporter and quencher fluorescencelabeled probes (TaqMan Technology) usingfluorescence resonance energy transfer (FRET)
Controls UsedSample processing control (IC) Positiveand negative controlSame
Results AnalysisPCR Cycle threshold analysisSame

Comparison of the cobas® liat CT/NG nucleic acid test and the Predicate Device Table 1:

{7}------------------------------------------------

NON-CLINICAL PERFORMANCE EVALUATION 4.

4.1. Analytical sensitivity (Limit of Detection)

Analytical sensitivity was determined by analyzing a dilution series of two representative strains/serovars of Chlamydia trachomatis (CT, Serovar D and I) and Neisseria gonorrhoeae (NG, Strains 2948 and 891). The CT and NG cultures were diluted in pooled negative urine (UR) or pooled negative vaginal swab (VS) clinical specimens to 7 concentration levels. All levels were tested with at least 20 replicates per concentration tested across 3 unique lots of reagents. LoD for each specimen type is shown in Table 2 and Table 3 for CT and NG respectively as the target concentration which can be detected in ≥ 95% of the replicates for all lots.

Table 2: CT concentration levels with at least 95% observed hit rate for all lots tested

Specimen TypesCTSerovar D LoD(EB/mL)CTSerovar D Mean CtValueCTSerovar I LoD(EB/mL)CTSerovar I Mean CtValue
Urine in cobas® PCRMedia0.08536.20.78436.0
Vaginal Swab incobas® PCR Media0.17035.30.78435.7

EB = Elementary Bodies

Table 3: NG concentration levels with at least 95% observed hit rate for all lots tested
Specimen TypesNGStrain 2948 LoD(CFU/mL)NGStrain 2948 Mean CtValueNGStrain 891 LoD(CFU/mL)NGStrain 891 Mean CtValue
Urine in cobas® PCRMedia0.25034.70.20034.5
Vaginal Swab incobas® PCR Media0.50034.20.20034.5

CFU = Colony Forming Units

4.2. Inclusivity

Inclusivity was performed for an additional 15 CT serovars and 43 NG strains using one lot of reagents. Testing was performed using CT and NG cultures that were spiked into pools of negative clinical specimens. Three replicates per dilution level were tested for each subtype per

{8}------------------------------------------------

specimen type. The lowest level at which all three replicates tested as positive are reported in Table 4 and Table 5 for CT and NG respectively.

Serovar Type or VariantUrine Specimens (EB/mL)Vaginal Swab Specimens (EB/mL)
A0.10.2
B0.40.2
Ba0.41
C0.70.7
E236
F0.40.04
G0.40.4
H0.43
J0.10.2
K0.10.04
LGV Type 10.10.04
LGV Type 21600200
LGV Type 30.10.7
nvCT0.10.7
Finnish-nvCT1:100 of Patient Sample1:100 of Patient Sample

Table 4: Inclusivity testing for CT serovars

Inclusivity testing for NG strains Table 5:

Strain IDUrine Specimens(CFU/mL)Vaginal Swab Specimens (CFU/mL)
ATCC 276330.20.5
ATCC 4922610.006
ATCC 7008250.010.001
Clinical Isolate SS1690.060.02
NBL 16060.30.08
NBL 19520.20.1
NBL 20120.20.3
NRL 19770.020.02
NRL 8042 - Belgium0.020.02
NRL 134770.090.1
NRL 138190.0060.004
NRL 33155 - Atlanta0.090.001
NRL 336410.010.07
Strain IDUrine Specimens(CFU/mL)Vaginal Swab Specimens (CFU/mL)
NRL 354950.010.07
NRL DAN 096120.020.03
NRL DN 7896 - DENMARK0.90.3
NRL DN 7901 - DENMARK0.020.02
NRL DOM 362 - Dominican Republic0.090.09
NRL DOM 1271 - Dominican Republic0.40.1
NRL KPO 1148 - KENYA (KPO)0.20.07
NRL KPO 1161 - KENYA (KPO)0.020.02
NRL Peru 330.070.07
NRL Peru 830.020.02
NRL PITT 94-4833 - PITTSBURGH (PITT)0.020.02
NRL PITT 94-8561 - PITTSBURGH (PITT)0.090.1
NRL PP 132 - PHILLIPINES0.090.1
NRL SEN 97 P-292 - SENEGAL (SEN)0.0060.02
NRL SEN 97 P-301 - SENEGAL (SEN)0.0060.07
Roche Diagnostics K.K.,Japan RDN001-001930.020.03
Roche Diagnostics, Australia 04D125:Darwin Northern Territory, Australia0.090.1
Roche Diagnostics, Australia 04D127:Darwin Northern Territory, Australia0.090.1
Roche Diagnostics, Australia 04D129:Darwin Northern Territory, Australia0.090.1
Roche Diagnostics, Australia 04D130:Darwin Northern Territory, Australia0.40.1
Roche Diagnostics, Australia 04D132:Darwin Northern Territory, Australia0.090.09
Roche Diagnostics, Australia 05D003:Darwin Northern Territory, Australia0.020.03
Roche Diagnostics, Australia 05D004:Darwin Northern Territory, Australia0.0060.004
Roche Diagnostics, Australia 4551 -Western Australia0.020.02
Statens Serum Institut 223/060.0060.006
Statens Serum Institut 1498/460.020.02
Statens Serum Institut 2170/460.020.02
Statens Serum Institut 2222/460.40.09
Statens Serum Institut 6973/450.090.09
UCSF580.060.07

{9}------------------------------------------------

{10}------------------------------------------------

4.3. Analytical specificity/cross reactivity

A panel of 181 strains of bacteria, fungi and viruses, including those commonly found in patient specimens, as well as 52 representative strains of non-gonorrhoeae Neisseria species and other phylogenetically unrelated organisms, were tested to assess analytical specificity. The organisms listed in Table 6 were spiked at concentrations of ≥ 1 x 106 units/mL* for bacteria or fungi and ≥ 1 x 105 units/mL for viruses into pools of negative vaginal swab specimens collected in cobas® PCR Media and negative urine specimens stabilized in cobas® PCR Media. Testing was performed with each potential interfering organism in the absence of, as well as mixed with. CT and NG cultures at ~3x LoD. Results indicated that 180 of the non-target organisms tested did not generate any false positive or false negative results due to cross-reactivity or interference. One strain of Neisseria lactamica (CCUG 26479), at concentrations greater than 1 x 104 CFU/mL, interfered with detection of NG at ~3x LoD. At 1 x 104 CFU/mL, this N. lactamica strain did not interfere with detection of NG at ~3x LoD, nor did 8 additional strains of N. lactamica when tested at concentrations ≥ 1 x 10° CFU/mL.

*Four bacteria could only be tested at a concentration below 1 x 106 units/mL and above 7 x 104 units/mL due to low stock titers.

Acholeplasma laidlawiiEikenella corrodensMobiluncus curtisiiPeptostreptococcusanaerobius
Acholeplasma oculi1,3Enterobacter aerogenes(Klebsiella aerogenes)Moraxella catarrhalisPlesiomonas shigelloides
Acinetobacter calcoaceticusEnterobacter cloacaeMoraxella lacunataPrevotella bivia
Acinetobacter lwoffiiEnterococcus aviumMoraxella osloensisCutibacterium acnes
Actinomyces israelii1,3Enterococcus faecalis (2strains)Morganella morganiiProteus mirabilis
Actinomyces pyogenes(Trueperella pyogenes)Enterococcus faecium (2strains)Mycobacterium smegmatisProteus vulgaris
Aerococcus viridansErwinia herbicola(Pantoea agglomerans)Mycoplasma faucium1,3Providencia stuartii
Aeromonas hydrophilaErysipelothrix rhusiopathiaeMycoplasma fermentansPseudomonas aeruginosa
Alcaligenes faecalisEscherichia coliMycoplasma hominisPseudomonas fluorescens
Atopobium vaginae(Fannyhessea vaginae)Flavobacteriummeningosepticum(Elizabethkingiameningoseptica)Mycoplasma oralePseudomonas putida
Bacillus subtilisFusobacterium nucleatumMycoplasma penetransRahnella aquatilis
Bacteroides fragilisGardnerella vaginalisMycoplasma pirumRhizobium radiobacter(Agrobacteriumtumefaciens)
Bacteroides ureolyticus(Campylobacterureolyticus)Gemella haemolysansMycoplasma pneumoniaeRhodospirillum rubrum
BifidobacteriumadolescentisGiardia IntestinalisMycoplasma primatumSaccharomyces cerevisiae
Bifidobacterium breveHaemophilus ducreyiMycoplasma salivariumSalmonella minnesota
Blautia productaHaemophilus influenzaeMycoplasmaspermatophilumSalmonella typhimurium
Brevibacterium linensHerpes simplex virus INeisseria cinerea (4strains)Serratia marcescens
Campylobacter jejuniHerpes simplex virus IINeisseria denitrificans(Bergeriella denitrifican)Staphylococcus aureus
Candida albicans (2 strains)HIV-1Neisseria elongata (3strains)Staphylococcus epidermidis
Candida glabrata(Nakaseomyces glabratus)Human papilloma virus 16(CaSki cells)Neisseria flavaStaphylococcussaprophyticus
Candida parapsilosisKingella denitrificansNeisseria flavescens (2strains)Streptococcus agalactiae
Candida tropicalisKingella kingaeNeisseria lactamica (9strains)2Streptococcus bovis
Chlamydia pneumoniaeKlebsiella oxytocaNeisseria macacaeStreptococcus mitis
Chlamydia psittaciKlebsiella pneumoniaeNeisseria meningitidisSerogroup AStreptococcus mutans
ChromobacteriumviolaceumLactobacillus acidophilusNeisseria meningitidisSerogroup BStreptococcus pneumoniae
Citrobacter braakiiLactobacillus brevis(Levilactobacillus brevis)Neisseria meningitidisSerogroup C(4 strains)Streptococcus pyogenes
Citrobacter freundiiLactobacillus crispatusNeisseria meningitidisSerogroup DStreptococcus salivarius
Clostridium difficile(Clostridioides difficile)Lactobacillus jenseniiNeisseria meningitidisSerogroup W135Streptococcus sanguinis
Clostridium perfringensLactobacillus lactisNeisseria meningitidisSerogroup YStreptomyces griseinus
CorynebacteriumgenitaliumLactobacillus vaginalis(Limosilactobacillusvaginalis)Neisseria mucosa (3strains)Trichomonas tenax
Corynebacterium xerosisLegionella pneumophila (2strains)Neisseria perflavaUreaplasma parvum
Cryptococcus neoformansLeptotrichia buccalisNeisseria polysacchareaUreaplasma urealyticum1,3
CytomegalovirusLeuconostocmesenteroidesNeisseria sicca (3 strains)Veillonella parvula
Deinococcus radioduransLeuconostocparamesenteroides(Weissellaparamesenteroides)Neisseria subflava (14strains)Vibrio parahaemolyticus
Derxia gummosaListeria monocytogenesParacoccus denitrificansYersinia enterocolitica
Dientamoeba fragilisMicrococcus luteusPentatrichomonas hominis-

Microorganisms tested for analytical specificity/cross reactivity Table 6:

{11}------------------------------------------------

{12}------------------------------------------------

1 Organism was tested at a concentration of < 1.0e+6 units/mL and > 7.0e+4 units/mL.

2 One strain of organism was tested at a concentration of < 1.0e+6 units/mL and > 1.0e+4 units/mL. 3Tested at highest concentration possible per stock concentration.

Interference 4.4.

The effects of over-the-counter or prescription products that may be present in urine or vaginal swab clinical specimens were evaluated at the concentration listed in Table 7. Testing was executed using pooled clinical specimens spiked with potential interferents at levels expected from normal patient usage. Interferents were tested in CT/NG negative specimen pools as well as in positive specimen pools spiked with CT/NG at ~3x LoD for each specimen type using one lot of reagents. Five replicates each of CT/NG negative sample and CT/NG positive sample (for each of two culture subtypes per microorganism) were tested with each exogenous substance in each specimen type, except for Azo Urinary Pain Relief, which was tested in urine only.

Of the products tested, no interference was observed in 15 substances when tested at concentrations of 1.5 mg/mL. Azo Urinary Pain Relief and carbomer-containing Replens™ Long-Lasting Vaginal Moisturizer resulted in false negative results in at least one replicate when tested at higher concentrations. Azo Urinary Pain Relief and Replens™ Long-Lasting Vaginal Moisturizer at concentrations greater than 0.5 mg/mL, respectively, may interfere with the assay performance. Levels of substances tolerated by the assay for all specimen types are shown in Table 7.

{13}------------------------------------------------

Product NameUrine (mg/mL)Vaginal Swabs (mg/mL)
Azo Urinary Pain Relief (urine only)0.5*-
Clindamycin Phosphate Vaginal Cream1.51.5
Equate tioconazole 1 Day1.51.5
Equate Vagicaine Anti-Inch Cream1.51.5
Estradiol Vaginal Cream1.51.5
7 Day vaginal cream1.51.5
K-Y® UltraGel1.51.5
Metronidazole Vaginal Gel1.51.5
Monistat Miconazole Nitrate Vaginal Cream (2%)1.51.5
Monistat® Instant Itch Relief Cream1.51.5
Norforms Deodorant Suppositories1.51.5
Premarin Vaginal Cream1.51.5
Replens™ Long-Lasting Vaginal Moisturizer1.0*1.5
Summer's Eve Ultra Freshening Spray1.51.5
VCF - Vaginal Contraceptive Gel1.51.5
Yeast Gard Gel Treatment1.51.5
RepHresh™ Vaginal Gel1.51.5

Table 7: List of products tested for interference

*Note: Concentrations above this level may cause interference in clinical samples.

Endogenous substances that may be present in urine or vaginal swab clinical specimens were evaluated at the concentration listed in Table 8. Testing was executed using pooled clinical specimens spiked with potential endogenous interferents at levels expected in a typical clinical sample. Endogenous substances were tested in CT/NG negative specimen pools as well as in positive specimen pools spiked with CT/NG at ~3x LoD for each relevant specimen type using one lot of reagents. Five replicates each of CT/NG negative sample and CT/NG positive sample (for each of two culture subtypes per microorganism) were tested with each endogenous substance in each relevant sample type.

For all endogenous substances tested, no interference was observed. Levels of endogenous substances tolerated by the assay for each specimen types are shown in Table 8.

{14}------------------------------------------------

Endogenous SubstanceUrineVaginal Swab
Human cells (PBMCs) cells/mL1.0E+061.0E+06
Mucus1 swab dipped into mucus1 swab dipped into mucus
Whole blood (v/v)10%10%
Semen (v/v) (vaginal swab only)-1.5%
Albumin (w/v) (urine only)5%-
Bilirubin (w/v) (urine only)1% (w/v)-
Glucose (w/v) (urine only)1% (w/v)-
Acidic pH (urine only)pH 4-

Summary of endogenous substance concentrations that do not show Table 8: interforonco

Competitive inhibition 4.5.

Alkaline pH (urine only)

To assess competitive inhibition between CT and NG, a total of six different combinations of low concentration of target (~2x LoD) were mixed with high concentrations of the other targets in both urine and vaginal swab clinical specimen matrices. Each combination was tested in replicates of 10 using one lot of reagents.

pH 9

Testing results indicated that when one or two target microorganisms were present at high concentrations, no interference was observed for microorganisms that were present at low concentrations (~2x LoD), when tested in both urine and vaginal swab clinical specimen matrices.

REPRODUCIBILITY STUDIES 5.

A reproducibility study was performed across different sites, lots, days, operators, instruments for cobas® liat CT/NG panels prepared from vaginal swabs and urine in cobas® PCR Media. Testing was performed at three external sites with a minimum of 3 cobas® liat analyzers per site. Operators at the CLIA-waived sites that met the definition of intended use operators were considered for this study. Selected operators were provided with the assay's IFU, Quick Reference Instructions, and the cobas® liat system User Guide. Operators were asked to read the materials before beginning any study testing. No assay or instrument training was provided to the operators.

{15}------------------------------------------------

Two operators at each site each tested 1 panel per specimen type per day (1 complete panel consists of 3 panel members each tested in triplicate) for a total of 15 days. All replicates for each panel member were always tested on the same analyzer. Each panel, per specimen type, consisted of a negative panel member (negative for all 3 analytes), a low positive panel member, and a moderate positive panel member with each positive panel member being co-formulated with all 3 analytes. For each panel member, approximately 270 results were produced.

The Reproducibility Study was executed with a total of 1618 tests consisting of 811 tests for the vaginal specimen type and 807 tests for the urine specimen type.

Table 9 and Table 10 show the site-to-site reproducibility study results for cobas® liat CT/NG by sample type and panel member concentration, respectively for CT and NG.

SpecimenTypePanel MemberConcentrationSite 1*Site 2*Site 3*Overall*
Vaginal1-2x LoD100%(90/90)(95.9% - 100.0%)100%(89/89)(95.9% - 100.0%)100%(90/90)(95.9% - 100.0%)100%(269/269)(98.6% - 100.0%)
Vaginal3-5x LoD100%(90/90)(95.9% - 100.0%)100%(90/90)(95.9% - 100.0%)100%(90/90)(95.9% - 100.0%)100%(270/270)(98.6% - 100.0%)
VaginalNegative100%(90/90)(95.9% - 100.0%)100%(83/83)(95.6 - 100.0%)100%(90/90)(95.9% - 100.0%)100%(263/263)(98.6% - 100.0%)
Urine1-2x LoD87.8%(79/90)(79.4%- 93.0%)93.3%(83/89)(86.1% - 96.9%)91.1%(82/90)(83.4% - 95.4%)90.7%(244/269)(86.6%- 93.6%)
Urine3-5x LoD95.6%(86/90)(89.1% - 98.3%)98.9%(88/89)(93.9% - 99.8%)94.4%(85/90)(87.6% - 97.6%)96.3%(259/269)(93.3%- 98.0%)
UrineNegative100%(90/90)(95.9% - 100.0%)100%(80/80)(95.6% - 100.0%)100%(90/90)(95.9% - 100.0%)100%260/260(98.5% - 100.0%)

Summary CT of site-to-site reproducibility results with cobas® liat CT/NG Table 9:

Note: LoD : limit of detection

*Percent Agreement with Expected Results (n/N) (95% Confidence Interval)

{16}------------------------------------------------

SpecimenTypePanelMemberConc.Site 1*Site 2*Site 3*Overall*
Vaginal1-2x LoD100%(90/90)(95.9% - 100.0%)100%(89/89)(95.9% - 100.0%)100%(90/90)(95.9% - 100.0%)100%(269/269)(98.6% - 100.0%)
Vaginal3-5x LoD100%(90/90)(95.9% - 100.0%)100%(90/90)(95.9% - 100.0%)100%(90/90)(95.9% - 100.0%)100%(270/270)(98.6% - 100.0%)
VaginalNegative100%(90/90)(95.9% - 100.0%)100%(83/83)(95.6 - 100.0%)100%(90/90)(95.9% - 100.0%)100%(263/263)(98.6% - 100.0%)
Urine1-2x LoD100%(90/90)(95.9% - 100.0%)98.9%(88/89)(93.9% - 99.8%)100%(90/90)(95.9% - 100.0%)99.6%(268/269)(97.9% - 99.9%)
Urine3-5x LoD100%(90/90)(95.9% - 100.0%)100%(89/89)(95.9% - 100.0%)100%(90/90)(95.9% - 100.0%)100%(269/269)(98.6% - 100.0%)
UrineNegative100%(90/90)(95.9% - 100.0%)100%(80/80)(95.6% - 100.0%)100%(90/90)(95.9% - 100.0%)100%(260/260)(98.5% - 100.0%)

Table 10: Summary NG of site-to-site reproducibility results with cobas® liat CT/NG

Note: LoD : limit of detection

*Percent Agreement with Expected Results (n/N) (95% Confidence Interval)

Table 11 and Table 12 present the total SD, and total percent CV (%) for Cycle Threshold Values from the Reproducibility Study for each specimen panel type run in cobas® liat CT/NG, respectively for CT and NG.

{17}------------------------------------------------

Table 11: CT - Overall mean estimate, standard deviations, and coefficients of variation (%) for cycle threshold values by sample type and expected concentration for cobas® liat CT/NG by sample type and positive panel member concentration

Bet-weenSiteBet-weenSiteBet-weenLotBet-weenLotBet-weenDayBet-weenDayBet-weenOperator/RunBet-weenOperator/RunWithin-RunWithin-RunTotal
SampleTypePanelMemberConcen-trationn/NaMeanCtSDCV%SDCV%SDCV%SDCV%SDCV%
Vaginal1x-2xLoD269/26933.40.000.000.531.600.220.670.000.000.842.521.023.06
Vaginal3x-5xLoD270/27032.10.210.640.581.820.300.930.000.001.003.131.223.79
Urine1x-2xLoD244/26934.80.150.440.842.410.310.880.000.000.912.611.283.69
Urine3x-5xLoD259/26934.00.150.450.702.070.230.680.000.000.982.891.243.65

Ct: cycle threshold; CV%: percent coefficient of variation; LoD: Limit of Detection; SD: standard deviation. ªn is the number of tests in agreement with expected results. N is the total number of valid tests for the panel member.

Table 12: NG - Overall mean estimate, standard deviations, and coefficients of variation (%) for cycle threshold values by sample type and expected concentration for cobas® liat CT/NG by sample type and positive panel member concentration

Between SiteBetween LotBetween DayBetween Operator/ RunWithin RunTotal
Sample TypePanel Member Concentrationn/NaMean CtSDCV%SDCV%SDCV%SDCV%SDCV%SDCV%
Vaginal1x-2x LoD269/26932.20.110.340.591.830.290.890.140.420.591.830.902.79
Vaginal3x-5x LoD270/27030.90.100.330.150.500.180.570.000.000.411.330.481.56
Urine1x-2x LoD268/26932.90.160.470.702.120.260.780.461.410.742.251.163.51
Urine3x-5x LoD269/26931.40.070.230.250.800.160.510.000.000.561.790.642.04

Ct: cycle threshold; CV%: percent coefficient of variation; LoD: Limit of Detection; SD: standard deviation.

ªn is the number of tests in agreement with expected results. N is the total number of valid tests for the panel member.

{18}------------------------------------------------

In the Reproducibility Study, the PPA for CT in urine panel members was less than the expected 95%. Therefore, a supplemental Precision Study was performed at one site across different lots, days, operators and instruments for cobas® liat CT/NG for the detection of CT in urine from urine panels prepared at negative, 1x-2x and 3x-5xLoD concentration levels. There were six total untrained operators and the level of instructional material were the same for this supplemental Precision study. Each operator tested 1 panel per day for 5 non-consecutive days for each lot (1 complete panel consisted of 3 panel members). This supplemental Precision Study was executed with a total of 810 evaluable tests on urine panel members.

Table 13 shows the supplemental between operator Precision Study for cobas® liat CT/NG by panel member concentration for CT in urine.

Panel MemberConcentrationOperatorn/N aAgreement with ExpectedResults (%)
1-2x LoD144/4597.8%
1-2x LoD244/44100.0%
1-2x LoD345/45100.0%
1-2x LoD444/44100.0 %
1-2x LoD545/45100.0%
1-2x LoD645/45100.0%
3-5x LoD145/45100.0%
3-5x LoD245/45100.0%
3-5x LoD345/45100.0%
3-5x LoD445/45100.0%
3-5x LoD545/45100.0%
3-5x LoD644/44100.0%
Negative143/4497.7%
Negative245/45100.0%
Negative345/45100.0%
Negative445/45100.0%
Negative545/45100.0%
Negative644/44100.0%

Table 13: Summary of CT Precision/Repeatability study results

a n is the number of tests with expected results. N is the total number of valid tests.

{19}------------------------------------------------

Table 14 shows the supplemental Reproducibility Study for cobas® liat CT/NG standard deviation (SD) and coefficient of variation (CV) of Cycle Threshold Values for each factor as well as the total SD and total CV (%) for each positive panel member.

Table 14: CT - Overall mean estimate, standard deviations, and coefficients of variation (%) for cycle threshold values and expected concentration for cobas® liat CT/NG by positive panel member concentration in urine

-BetweenInstrumentBetweenInstrumentBetweenLotBetweenLotBetweenDayBetweenDayBetweenOperator/RunBetweenOperator/RunWithin-RunWithin-RunTotalTotal
PanelMemberConcentrationn/NaMeanCtSDCV%SDCV%SDCV%SDCV%SDCV%SDCV%
1x-2xLOD267/26835.30.000.000.030.080.000.000.000.000.852.410.852.41
3x-5xLOD269/26933.70.000.000.000.000.000.000.461.351.073.191.173.47

Note: Ct = cycle threshold, CT=Chlamydia trachomatis, CV(%) = percent coefficient of variation, LoD = Limit of Detection, NG=Neisseria gonorrhoeae, SD =standard deviation.

ªn is the number of tests in agreement with expected results. N is the total number of valid tests for the panel member.

CLINICAL PERFORMANCE EVALUATION 6.

6.1. Clinical study

The clinical utility and performance of cobas® liat CT/NG was established in a multi-site, prospective study by comparing the results to a Patient Infected Status (PIS) or a Composite Comparator Algorithm (CCA) derived from a combination of FDA-cleared NAATs for the 2 analytes. A result for PIS (for male urine) or a CCA (for vaginal swabs) was generated for CT or NG. Male urine, and vaginal swabs were collected and tested at 13 geographically diverse intended use clinical sites across the US. There were 48 operators that took part in cobas® liat CT/NG testing, of which, 43 represented CLIA-waived operators. Five of the 48 operators represented experienced laboratorians in a moderate complexity laboratory.

A total of 4852 subjects (2512 females and 2340 males) were enrolled in the study and provided specimens for collection. Note, two subjects, declared male at birth, provided vaginal swab specimens. Of these subjects, 72 were non-evaluable due to protocol deviations and incidents

{20}------------------------------------------------

(18), invalid cobas and/or final comparator result (45), or sample collection incidents (9). Of the evaluable subjects, 2304 male subjects provided 2302 male urine specimens (2 subjects provided vaginal swab specimens) and 2476 females provided 1240 clinician-collected vaginal swabs and 1236 self-collected vaginal swabs for evaluation in the clinical study.

Prospectively enrolled female subjects provided 4 vaginal swab specimens, three for comparator tests and one for the cobas® liat CT/NG/MG nucleic acid test. Vaginal swab specimen for the cobas® liat CT/NG/MG nucleic acid test was either collected by clinician or self-collected.

Prospectively enrolled male subjects provided a urine specimen that was aliquoted into the respective manufacturers' collection devices and cobas® PCR Media.

Specimens were tested for CT and NG with the investigational and the reference comparator NAATs. All tests were run according to the respective IFU.

The clinical performance of cobas® liat CT/NG was evaluated by comparing the results from collected specimen types to a pre-specified PIS algorithm. The PIS/CCA result for each analyte was derived from a combination of 3 reference NAATs (NAAT1, NAAT2, and NAAT3). If NAAT1 and NAAT2 are concordant, then the final PIS/CCA result for the respective analyte is the concordant result obtained from NAAT1 and NAAT2. If NAAT1 and NAAT2 are discordant, then NAAT3 is performed to be the tiebreaker between the first 2 discordant results. Table 15 below shows the PIS and CCA algorithm for each analyte.

NAAT 1NAAT 2NAAT 3 (if needed)Patient InfectedStatusªCompositeComparatorAlgorithm
++N/AInfectedPositive
+-+InfectedPositive
-++InfectedPositive
--N/ANot InfectedNegative
+--Not InfectedNegative
-+-Not InfectedNegative
-Invalid+IndeterminateIndeterminate
-Invalid-Not InfectedNegative
Invalid-+IndeterminateIndeterminate
Invalid--Not InfectedNegative
+Invalid-IndeterminateIndeterminate

Table 15: Determination of the PIS/CCA result for CT and NG, respectively

{21}------------------------------------------------

NAAT 1NAAT 2NAAT 3 (if needed)Patient InfectedStatusaCompositeComparatorAlgorithm
Invalid+-IndeterminateIndeterminate
+Invalid+InfectedPositive
Invalid++InfectedPositive
InvalidInvalidN/AIndeterminateIndeterminate

N/A: not applicable; NAAT: nucleic acid amplification test.

a The results from NAAT1 and NAAT2 determined if NAAT3 needed to be performed. The "Infected" or "Not Infected" patient infected status was derived from the total combination of results obtained from the reference NAATs.

The sample types of male urine and vaginal swab were used to create the PIS and CCA results, respectively, for men and women. The cobas® liat CT/NG results of each analyte from each sample type (male urine and vaginal swab) were compared to the PIS/CCA resultto determine the clinical performance of the assay. Sensitivity (SENS), specificity (SPEC), positive percent agreement (PPA), and negative percent agreement (NPA) of cobas® liat CT/NG were calculated separately for CT and NG.

Supplementation with archived specimens was included in this study due to the expected low NG prevalence for male urine and vaginal swabs. The archived specimens were prospectively collected samples from a prior clinical trial study (K173887).

6.1.1. Performance results

Sensitivity, specificity, and predictive values of cobas® liat CT/NG as defined by the PIS/CCA results are presented by gender, sample type, and symptom status in Table 16, and Table 17, respectively for CT and NG prospectively collected specimens. NG archived specimens and NG for prospective and archived specimens combined.

Upon initial testing, the cobas® liat CT/NG invalid rate was 0.6% and after retesting the final invalid rate was 0.2%.

{22}------------------------------------------------

SpecimenTypeSymptomStatusNSensitivityEstimate(95% CI)Sensitivityn/NSpecificityEstimate(95% CI)Specificityn/N
Male UrineSymptomatic80898.2%(90.6%, 99.7%)55/5699.9%(99.3%, 100.0%)751/752
Male UrineAsymptomatic148896.4%(87.7%, 99.0%)53/5599.9%(99.6%, 100.0%)1432/1433
Male UrineTotal229697.3%(92.4%, 99.1%)108/11199.9%(99.7%, 100.0%)2183/2185
SpecimenTypeSymptomStatusNPositive PercentAgreementEstimate(95% CI)PositivePercentAgreementn/NNegative PercentAgreementEstimate(95% CI)Negative PercentAgreementn/N
VaginalSwabsSymptomatic111698.4%(91.3%, 99.7%)60/6199.7%(99.2%, 99.9%)1052/1055
VaginalSwabsAsymptomatic135797.9%(89.1%, 99.6%)47/4899.8%(99.4%,100.0%)1307/1309
VaginalSwabsTotal247398.2%(93.6%, 99.5%)107/10999.8%(99.5%, 99.9%)2359/2364
SpecimenTypeSymptomStatusNSensitivityEstimate(95% CI)Sensitivityn/NSpecificityEstimate(95% CI)Specificityn/N
Male UrineSymptomatic813100.0%(94.7%, 100.0%)68/68100.0%(99.5%, 100.0%)745/745
Male UrineAsymptomatic148100.0%(74.1%, 100.0%)11/1199.8%(99.4%, 99.9%)1474/1477
Male UrineTotal231100.0%(95.4%, 100.0%)79/7999.9%(99.6%, 100.0%)2219/2222
ArchivedMale UrineSymptomatic125100.0%(95.2%, 100.0%)77/77100.0%(92.6%, 100.0%)48/48
ArchivedMale UrineAsymptomatic38100.0%(56.6%, 100.0%)5/5100.0%(89.6%, 100.0%)33/33
ArchivedMale UrineTotal163100.0%(95.5%, 100.0%)82/82100.0%(95.5%, 100.0%)81/81
Overall MaleUrineSymptomatic938100.0%(97.4%, 100.0%)145/145100.0%(99.5%, 100.0%)793/793
Overall MaleUrineAsymptomatic1526100.0%(80.6%, 100.0%)16/1699.8%(99.4%, 99.9%)1507/1510
Overall MaleUrineTotal2464100.0%(97.7%, 100.0%)161/16199.9%(99.6%, 100.0%)2300/2303
SpecimenTypeSymptomStatusNPositive PercentAgreementPositive PercentAgreementNegative PercentAgreementNegativePercentAgreement
Estimate(95% CI)n/NEstimate(95% CI)n/N
VaginalSwabsSymptomatic111591.7%(74.2%, 97.7%)22/2499.8%(99.3%, 99.9%)1089/1091
VaginalSwabsAsymptomatic1357100.0%(82.4%, 100.0%)18/1899.9%(99.5%, 100.0%)1337/1339
VaginalSwabsTotal247295.2%(84.2%, 98.7%)40/4299.8%(99.6%, 99.9%)2426/2430
ArchivedVaginalSwabsSymptomatic42100.0%(83.9%, 100.0%)20/20100.0%(85.1%, 100.0%)22/22
ArchivedVaginalSwabsAsymptomatic48100.0%(86.7%, 100.0%)25/25100.0%(85.7%, 100.0%)23/23
ArchivedVaginalSwabsTotal90100.0%(92.1%, 100.0)45/45100.0%(92.1%, 100.0%)45/45
OverallVaginalSwabsSymptomatic115795.5%(84.9%, 98.7%)42/4499.8%(99.3%, 100.0%)1111/1113
OverallVaginalSwabsAsymptomatic1405100.0%(91.8%, 100.0%)43/4399.9%(99.5%, 100.0%)1360/1362)
OverallVaginalSwabsTotal256297.7%(92.0%, 99.4%)85/8799.8%(99.6%, 99.9%)2471/2475

Table 16: CT - Clinical performance of cobas® liat CT/NG compared with PIS/CCA by specimen type and symptom status

CI: confidence interval

{23}------------------------------------------------

Table 17: NG - Clinical performance of cobas® liat CT/NG compared with PIS/CCA by specimen type and symptom status

CI: confidence interval

{24}------------------------------------------------

Expected values for urogenital specimens 6.2.

The positivity rate of the cobas® liat CT/NG nucleic acid assay test for CT and NG observed during the study is shown for each specimen type, by collection site in Table 18 below

CollectionSiteCTNG
MaleUrineVSMaleUrineVS
18.7%(30/343)9.2%(14/152)11.0%(38/346)3.3%(5/152)
22.6%(9/346)6.2%(15/241)1.7%(6/346)3.7%(9/241)
311.9%(18/151)8.2%(30/366)6.6%(10/151)0.55%(2/364)
411.2%(12/107)0.63%(1/160)9.3%(10/107)1.25%(2/160)
50.0%(0/4)NC0.0%(0/4)NC
60.9%(1/117)1.2%(1/85)0.0%(0/118)1.2%(1/85)
75.3%(3/57)5.7%(2/35)1.8%(1/57)2.9%(1/35)
80.0%(0/80)0.0%(0/19)2.5%(2/80)0.0%(0/19)
91.3%(6/468)1.9%(10/527)0.4%(2/469)2.3%(12/528)
100.5%(1/198)1.4%(5/347)1.0%(2/198)0.86%(3/347)
1117.1%(18/105)5.6%(17/305)4.8%(5/105)2.0%(6/305)
123.5%(10/289)8.8%(12/136)1.7%(5/289)1.5%(2/136)
136.5%(2/31)5.0%(5/100)3.2%(1/31)1.0%(1/100)

Table 18: Positivity of CT/NG/MG as Determined by the cobas liat CT/NG/MG nucleic acid test by Specimen Type and Clinical Site

The hypothetical PPVs and NPVs of cobas® liat CT/NG derived from disease prevalences of 1% to 50% are shown in Table 19 and Table 20 respectively, for CT and NG.

{25}------------------------------------------------

Specimen type for CNMAtestingHypothetical Prevalence(%)PPV (%)NPV (%)
Male Urine191.5100.0
Male Urine397.099.9
Male Urine598.299.9
Male Urine1099.299.7
Male Urine1599.599.5
Male Urine2099.699.3
Male Urine3099.898.9
Male Urine5099.997.4
Vaginal Swabs182.4100.0
Vaginal Swabs393.599.9
Vaginal Swabs596.199.9
Vaginal Swabs1098.199.8
Vaginal Swabs1598.899.7
Vaginal Swabs2099.199.5
Vaginal Swabs3099.599.2
Vaginal Swabs5099.898.2

Table 19: CT - Positive predictive value and negative predictive value for hypothetical CT prevalence

Note: NPV: negative predictive value; PPV: positive value; PPA: positive percent agreement; NPA: negative percent agreement;

The PPV and NPV were calculated using the sensitivity/PPA and specificity/NPA of cobas® liat CT/NG/MG from the prospectively collected population.

{26}------------------------------------------------

Specimen type for CNMAtestingHypothetical Prevalence(%)PPV (%)NPV (%)
Male Urine188.2100.0
Male Urine395.8100.0
Male Urine597.5100.0
Male Urine1098.8100.0
Male Urine1599.2100.0
Male Urine2099.5100.0
Male Urine3099.7100.0
Male Urine5099.9100.0
Vaginal Swabs185.4100.0
Vaginal Swabs394.799.9
Vaginal Swabs596.899.7
Vaginal Swabs1098.599.5
Vaginal Swabs1599.099.2
Vaginal Swabs2099.398.8
Vaginal Swabs3099.698.0
Vaginal Swabs5099.895.4

Table 20: NG - Positive predictive value and negative predictive value for hypothetical NG prevalence

Note: NPV: negative predictive value; PPV: positive value; PPA: positive percent agreement; NPA: negative percent agreement;

The PPV and NPV were calculated using the sensitivity/PPA and specificity/NPA of cobas® liat CT/NG/MG from the prospectively collected population.

7. CONCLUSIONS

A comparison of the intended use, technological characteristics, and the results of non-clinical analytical and clinical performance studies demonstrate that cobas® liat CT/NG nucleic acid test is substantially equivalent to the predicate devices.

§ 866.3393 Device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s).

(a)
Identification. A device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s) is an in vitro diagnostic device intended for the detection and identification of nucleic acids from non-viral microorganism(s) and their associated resistance markers in clinical specimens collected from patients suspected of sexually transmitted infections. The device is intended to aid in the diagnosis of non-viral sexually transmitted infections in conjunction with other clinical and laboratory data. These devices do not provide confirmation of antibiotic susceptibility since mechanisms of resistance may exist that are not detected by the device.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The intended use for the labeling required under § 809.10 of this chapter must include a detailed description of targets the device detects, the results provided to the user, the clinical indications appropriate for test use, and the specific population(s) for which the device is intended.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed discussion of the performance characteristics of the device for all claimed specimen types based on analytical studies, including Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, within lab precision, and reproducibility, as appropriate;
(iii) Detailed descriptions of the test procedure, the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) Limiting statements indicating that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) Reliable results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(D) If appropriate (
e.g., recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer reviewed research), that the clinical performance is inferior in a specific clinical subpopulation or for a specific claimed specimen type; and(v) If the device is intended to detect antimicrobial resistance markers, limiting statements, as appropriate, indicating that:
(A) Negative results for claimed resistance markers do not indicate susceptibility of detected microorganisms, as resistance markers not measured by the assay or other potential mechanisms of antibiotic resistance may be present;
(B) Detection of resistance markers cannot be definitively linked to specific microorganisms and the source of a detected resistance marker may be an organism not detected by the assay, including colonizing flora;
(C) Detection of antibiotic resistance markers may not correlate with phenotypic gene expression; and
(D) Therapeutic failure or success cannot be determined based on the assay results, since nucleic acid may persist following appropriate antimicrobial therapy.
(4) Design verification and validation must include:
(i) Detailed device description documentation, including methodology from obtaining sample to result, design of primer/probe sequences, rationale for target sequence selection, and computational path from collected raw data to reported result (
e.g., how collected raw signals are converted into a reported result).(ii) Detailed documentation of analytical studies, including, Limit of Detection, inclusivity, cross-reactivity, microbial interference, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, within lab precision, and reproducibility, as appropriate.
(iii) Detailed documentation and performance results from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, when determined to be appropriate by FDA, additional characterized clinical samples. The study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained from FDA accepted comparator methods. Documentation from the clinical studies must include the clinical study protocol (including a predefined statistical analysis plan) study report, testing results, and results of all statistical analyses.
(iv) A detailed description of the impact of any software, including software applications and hardware-based devices that incorporate software, on the device's functions.