K Number
K240158

Validate with FDA (Live)

Date Cleared
2024-09-04

(229 days)

Product Code
Regulation Number
880.5725
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Modular Medical MODD1 Insulin Delivery System is indicated for the subcutaneous delivery of insulin at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, for individuals 18 years of age and greater.

Device Description

The Modular Medical MODD1 Insulin Delivery System (i.e., MODD1 System) consists of the following:

  • A reusable, software-controlled programmable Pump,
  • A single-use, disposable 3.0mL (300 unit) Insulin Cartridge with an integrated battery to supply power to the Pump, infusion tubing, and Tubing Cap for attaching to the Infusion Set.
  • . A single-use, disposable Adhesive Pad for affixing the MODD1 System to the user's body and allowing for temporary removal of the MODD1 System,
  • A single-use, disposable Infusion Set inserted into the patient's subcutaneous tissue, and
  • The MMI App, a smartphone application used to program the basal delivery schedule on the Pump.

Accessories to the MODD1 System are FDA-cleared and include the following:

  • A single-use disposable syringe (K110771) and needle (K021475) used to fill the Insulin Cartridge reservoir, and
  • A reusable Inserter (K163400), optionally used to insert the Infusion Set into the patient's subcutaneous tissue.

The MODD1 System has been demonstrated to be compatible with Humalog U-100 insulin.

The MODD1 System can deliver insulin at basal rates between 0.5 - 4 units per hour (programmable in 0.1-unit increments) and user-selected bolus doses of between 2 and 20 units (in 2-unit increments). Basal insulin delivery may also be suspended for a set period of 30 minutes when required by the user.

The delivery of insulin is achieved by gear motor rotation of a single fixed-position camshaft (components of the Pump controlled by software) that interfaces with the Insulin Cartridge pistons to control the movement of insulin from the insulin reservoir, into an interim chamber, and into the infusion tubing towards the Infusion Set inserted into the patient's subcutaneous tissue. Software monitors dispense output via feedback provided by multiple sensors, providing an alarm when issues with delivery (e.g., malfunctions, low reservoir volume, occlusions) are detected.

The Pump serves as the primary user interface of the MODD1 System. A Control Button is used to prime the fluid path, program bolus deliveries, check system status, and to temporarily suspend insulin delivery. The multi-color LED and piezo buzzer of the Pump provide audio-visual feedback to the user during the set-up process, programming of a bolus delivery, and changing of delivery modes (e.g., bolus, basal, suspend). Audio-visual feedback is also used to inform the user of system alarms (automatically generated) and system status (when requested by the user).

The MMI App is downloaded directly onto an Apple® iPhone 12 Pro Max running iOS version 16. The MMI App allows for the configuration of a basal delivery schedule on the Pump, consisting of up to two basal rates in a 24-hour period. The MMI App and Pump communicate using Bluetooth® Low Energy (BLE) and Near Field Communication (NFC). After the basal delivery schedule has been programmed on the Pump, all insulin delivery functions are controlled using the Pump user interface. Optional features of the MMI App include a dashboard display of current and historical Pump information and delivery history.

AI/ML Overview

The provided text is a 510(k) Summary for the Modular Medical MODD1 Insulin Delivery System. It details the device, its indications for use, comparison to a predicate device, and a summary of non-clinical performance data.

However, the document does not provide information on an AI-driven medical device, nor does it conduct a study involving human readers or expert ground truth. It describes the regulatory submission for an insulin pump.

Therefore, I cannot extract the requested information regarding acceptance criteria, device performance, sample sizes, expert ground truth establishment, adjudication methods, MRMC studies, standalone AI performance, or training set details from the provided text.

The text describes compliance with various medical device standards (e.g., ISO 14971, IEC 62366-1, ISO 10993-1, IEC 60601-1, IEC 62304) and includes bench testing for infusion delivery accuracy and occlusion detection, but these are part of standard medical device validation, not specific to AI performance as requested in the prompt.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 4, 2024

Modular Medical, Inc. Kelsie DiPerna Sr. Regulatory Affairs Specialist 10740 Thornmint Road San Diego, California 92127

Re: K240158

Trade/Device Name: Modular Medical MODD1 Insulin Delivery System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: LZG Dated: January 19, 2024 Received: January 19, 2024

Dear Kelsie DiPerna:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a

{1}------------------------------------------------

change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joshua Balsam -S

Joshua M. Balsam, Ph.D. Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K240158

Device Name

Modular Medical MODD1 Insulin Delivery System

Indications for Use (Describe)

The Modular Medical MODD1 Insulin Delivery System is indicated for the subcutaneous delivery of insulin at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, for individuals 18 years of age and greater.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows a logo for Modular Medical. The logo consists of two overlapping letter 'M's, one in orange and the other in blue. To the right of the 'M's, the words 'MODULAR MEDICAL' are written in a simple, sans-serif font, with 'MODULAR' on top of 'MEDICAL'.

510(k) Summary

K240158

Table 1: Submitter and Device Information

Submitter NameModular Medical, Inc.
Submitter Address10740 Thornmint RoadSan Diego, CA 92127
Contact PersonKelsie DiPernaSenior Regulatory Affairs Specialist(808) 315-6427kelsie@modular-medical.com
Device Trade / Proprietary NameModular Medical MODD1 Insulin Delivery System
Device Common NamePump, Infusion, Insulin
Regulation Medical SpecialtyGeneral Hospital
Review PanelClinical Chemistry
Product CodeLZG (Class II) - Pump, Infusion, Insulin
Regulation21 CFR 880.5725 - Infusion pump
Submission TypeTraditional 510(k)
Predicate Devicet:slim® Insulin Delivery System (K160056)
Date PreparedSeptember 4, 2024

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows a logo for Modular Medical. The logo consists of two overlapping letter M's, one in orange and one in blue. To the right of the M's are the words "MODULAR MEDICAL" in a simple sans-serif font, with "MODULAR" on top of "MEDICAL".

1.0 Indications for Use

The Modular Medical MODD1 Insulin Delivery System is indicated for the subcutaneous delivery of insulin at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, for individuals 18 years of age and greater.

2.0 Device Description

The Modular Medical MODD1 Insulin Delivery System (i.e., MODD1 System) consists of the following:

  • A reusable, software-controlled programmable Pump, ●
  • A single-use, disposable 3.0mL (300 unit) Insulin Cartridge with an integrated ● battery to supply power to the Pump, infusion tubing, and Tubing Cap for attaching to the Infusion Set.
  • . A single-use, disposable Adhesive Pad for affixing the MODD1 System to the user's body and allowing for temporary removal of the MODD1 System,
  • A single-use, disposable Infusion Set inserted into the patient's subcutaneous ● tissue, and
  • The MMI App, a smartphone application used to program the basal delivery ● schedule on the Pump.

Accessories to the MODD1 System are FDA-cleared and include the following:

  • A single-use disposable syringe (K110771) and needle (K021475) used to fill the ● Insulin Cartridge reservoir, and
  • A reusable Inserter (K163400), optionally used to insert the Infusion Set into the . patient's subcutaneous tissue.

The MODD1 System has been demonstrated to be compatible with Humalog U-100 insulin.

The MODD1 System can deliver insulin at basal rates between 0.5 - 4 units per hour (programmable in 0.1-unit increments) and user-selected bolus doses of between 2 and 20 units (in 2-unit increments). Basal insulin delivery may also be suspended for a set period of 30 minutes when required by the user.

The delivery of insulin is achieved by gear motor rotation of a single fixed-position camshaft (components of the Pump controlled by software) that interfaces with the Insulin Cartridge pistons to control the movement of insulin from the insulin reservoir, into an interim chamber, and into the infusion tubing towards the Infusion Set inserted into the patient's subcutaneous tissue. Software monitors dispense output via feedback provided by multiple sensors, providing an alarm when issues with delivery (e.g., malfunctions, low reservoir volume, occlusions) are detected.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows a logo for a company called "Modular Medical". The logo consists of two overlapping letter "M"s, one in orange and one in blue. The word "MODULAR" is written in a smaller font size above the word "MEDICAL" to the right of the letter "M"s.

The Pump serves as the primary user interface of the MODD1 System. A Control Button is used to prime the fluid path, program bolus deliveries, check system status, and to temporarily suspend insulin delivery. The multi-color LED and piezo buzzer of the Pump provide audio-visual feedback to the user during the set-up process, programming of a bolus delivery, and changing of delivery modes (e.g., bolus, basal, suspend). Audio-visual feedback is also used to inform the user of system alarms (automatically generated) and system status (when requested by the user).

The MMI App is downloaded directly onto an Apple® iPhone 12 Pro Max running iOS version 16. The MMI App allows for the configuration of a basal delivery schedule on the Pump, consisting of up to two basal rates in a 24-hour period. The MMI App and Pump communicate using Bluetooth® Low Energy (BLE) and Near Field Communication (NFC). After the basal delivery schedule has been programmed on the Pump, all insulin delivery functions are controlled using the Pump user interface. Optional features of the MMI App include a dashboard display of current and historical Pump information and delivery history.

3.0 Summary of Technological Characteristics Compared to Predicate Device

Similarities and differences between the predicate device (t:slim® Insulin Delivery System) and the subject device (Modular Medical MODD1 Insulin Delivery System) are illustrated in Table 2.

The subject device has the same intended use (subcutaneous delivery of insulin at set and variable rates), and similar indications for use as the predicate device, with the sole difference being the indicated age of users in the patient population. This difference does not introduce a new intended use.

Minor differences in the technological characteristics exist but do not raise new questions of safety and effectiveness. The information and testing evidence provided in this submission demonstrate that the subject device is as safe and effective as the predicate device.

Element ofComparisonPredicate DeviceSubject Device
Indications for UseThe t:slim® Insulin Delivery Systemis indicated for the subcutaneousdelivery of insulin at set and variablerates, for the management of diabetesmellitus in persons requiring insulin,for individuals 6 years of age andThe Modular Medical MODD1Insulin Delivery System is indicatedfor the subcutaneous delivery ofinsulin at set and variable rates, forthe management of diabetes mellitusin persons requiring insulin, for
Element ofPredicate DeviceSubject Device
Comparison
greater. The device is indicated for usewith NovoLog or Humalog U-100insulin.individuals 18 years of age andgreater.
Intended UseInfusion System intended for thetreatment of diabetes mellitus inpersons requiring insulin.Same
Prescription orOver the CounterPrescription onlySame
Use EnvironmentAt homeSame
SpecificDrug/BiologicalUseNovoLog or Humalog U-100 insulinHumalog U-100 insulin
Pump TypeMotor-driven (linear piston)Motor-driven (multiple piston)
Use TimeUp to 72 hoursSame
ComponentsInfusion pump,Sterile disposable insulin cartridgeInfusion pump,Sterile disposable insulin cartridge,Disposable adhesive pad,Sterile disposable infusion set,MMI App (smartphone application)
AccessoriesUnoMedical Comfort™ disposableinsulin infusion set (or equivalent),3mL sterile syringe and 26-gaugeneedle (for filling cartridge),AC power supply and DC car adaptorpower supply with USB3 mL sterile syringe (K110771),26-gauge sterile needle (K021475),Inserter (K163400)
SterilizationMethod (fordisposable insulincartridge)Gamma sterilizationSame
Dimensions3.13" x 2.0" x 0.6" (L x W x H)Pump only: 1.51" x 2.38" x 0.56" (Lx W x H)
Weight3.95 ounces (112 grams) (with fulldisposable attached)1 ounce (28 grams) (with fulldisposable attached)
Reservoir Capacity3.0 mL custom disposable insulincartridgeSame
User InterfaceTouch screen and key input (Pump)Touch screen (MMI App) and buttoninput (Pump)
WearabilityBelt / pocket mountedBody worn
CommunicationBLEBLE and NFC
Element ofComparisonPredicate DeviceSubject Device
Pump Informationand HistoryVisible on the t:slim pump screenCurrent pump information visibleand audible on the pump;Pump history optionally visible onthe MMI App
Alarm TypeVisual, audible, and vibratoryVisual and audible
Power SourcesLithium Polymer battery(rechargeable)Internal coin cell battery (non-rechargeable)
Pump WaterResistanceIPX7IP24
Basal SuspendModeYesSame
Quick BolusUser selectable dose through singlebuttonSame
Flow Rates andProfilesBasal: 0.1 - 15 units/hour (in 0.001-unit increments)Bolus: 0.05 - 25 units (in 0.01-unitincrements)Basal: 0.5-4 units/hour (in 0.1-unitincrements)Bolus: 2 - 20 units (in 2-unitincrements)
Delivery AccuracyBasal Delivery Accuracy atintermediate rate: ±5%Bolus Delivery Accuracy: ±5%Typical (mean) Basal DeliveryAccuracy at All Flow Rates: ±5%Typical (mean) Bolus DeliveryAccuracy at All Bolus Volumes:+5%
Maximum Time toOcclusion AlarmBolus (3U or greater): 23 secondsIntermediate Basal (2U/hr): 1 hour 10minutesMinimum Basal (0.1U/hr): 25 hoursBolus (2U or greater): 2 minutes 19secondsIntermediate Basal (2U/hr): 31minutes 31 secondsMinimum Basal (0.5U/hr): 26 hours

Table 2: Substantial Equivalence Comparison

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Modular Medical. The logo consists of two overlapping letter M's, one in orange and one in blue. To the right of the M's is the text "MODULAR MEDICAL" in a sans-serif font, with "MODULAR" stacked on top of "MEDICAL".

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows a logo for Modular Medical. The logo consists of two overlapping letter M's, one in orange and the other in blue. To the right of the M's are the words "MODULAR MEDICAL" stacked on top of each other in a sans-serif font. The word "MODULAR" is on top and "MEDICAL" is on the bottom.

Summary of Non-Clinical Performance Data 4.0

Appropriate testing was performed to confirm the subject device met specified requirements and performed as intended. The following performance and safety testing have confirmed the Modular Medical MODD1 Insulin Delivery System to be safe, effective, and substantially equivalent to the predicate device:

  • Risk Management: Risk management was conducted in accordance with ISO ● 14971:2019. Verification of risk control measures demonstrates that all risk associated with the subject device has been mitigated to an acceptable level and that the subject device is safe for its intended use.
    • o Safety Assurance: A safety assurance case has been provided as part of this submission per FDA guidance "Infusion Pumps Total Product Life Cycle"

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows a logo for a company called "Modular Medical". The logo consists of two overlapping letters "M", one in orange and the other in blue. The word "MODULAR" is stacked on top of the word "MEDICAL" to the right of the letters.

and supports the determination that the subject device has adequately addressed hazards associated with its intended use within its environment of use.

  • Human Factors: Human factors validation testing was conducted on the subject device to assess comprehension and usability of the device for critical device tasks per IEC 62366-1:2020, ANSI AAMI HE75:2009/(R)2018, and FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices". Results of validation testing performed demonstrate that the subject device has been found to be safe and effective for the intended users, uses, and use environments.
  • . Insulin Compatibility: Testing was conducted on the subject device to demonstrate compatibility with Humalog U-100 insulin.
  • Biocompatibility: A biological evaluation was conducted on the subject device and . included appropriate biocompatibility testing based on the nature and duration of patient-contacting materials to meet the requirements of ISO 10993-1:2018 and FDA guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"".
  • Sterility: Sterilization validation testing was performed on the subject device per applicable sterility standards.
  • Bench Testing: Verification testing has demonstrated that the subject device meets design requirements. Infusion delivery accuracy at various flow rates and occlusion detection performance were verified.
  • Electrical Safety and Electromagnetic Compatibility: Testing was conducted on . the subject device to demonstrate safety and performance per IEC 60601-1:2020, IEC 60601-1-2:2020, and applicable collateral standards. Additionally, testing for electromagnetic compatibility in the aircraft environment was conducted per RTCA DO-160G.
  • Wireless Coexistence: Testing was conducted per AAMI TIR69:2017/(R)2020 for ● the risk management of wireless coexistence and IEEE/ANSI C63.27:2021 for the evaluation of wireless coexistence.
  • Software: Software verification and validation testing were conducted on the subject device per IEC 62304:2015 and FDA guidance "General Principles of Software Validation". Software documentation and testing evidence have been provided in accordance with FDA guidance "Content of Premarket Submissions for Device Software Functions".
  • Cybersecurity: A cybersecurity risk analysis and testing were performed on the ● subject device per FDA guidance "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions", and guidelines of "Postmarket Management of Cybersecurity in Medical Devices". Cybersecurity documentation and testing evidence are provided in this submission to support the determination of substantial equivalence for the subject device.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows a logo for a company called "Modular Medical". The logo consists of two overlapping letter "M"s, one in orange and one in blue. The word "MODULAR" is written in blue to the right of the letter M's, and the word "MEDICAL" is written below it.

5.0 Substantial Equivalence Conclusion

After analyzing the intended use/indications for use, technological characteristics, and performance data, Modular Medical, Inc. concludes that the Modular Medical MODD1 Insulin Delivery System is substantially equivalent to the legally marketed Tandem Diabetes Care Inc. t:slim® Insulin Delivery System (K160056). While the subject device's technological characteristics differ slightly from the predicate, the differences do not raise different questions of safety and effectiveness as substantiated by testing and information supplied in this submission. Therefore, the Modular Medical MODD1 Insulin Delivery System is substantially equivalent to the predicate device.

§ 880.5725 Infusion pump.

(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).