K Number
K240139

Validate with FDA (Live)

Date Cleared
2024-03-18

(60 days)

Product Code
Regulation Number
882.4560
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

REAL INTELLIGENCE CORI is indicated for use in surgical procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be determined. These procedures include:
· unicondylar knee replacement (UKR),
· total knee arthroplasty (TKA),
· revision knee arthroplasty, and
· total hip arthroplasty (THA).

Device Description

The subject of this Traditional 510(k) is REAL INTELLIGENCE CORI (CORI), a robotic-assisted orthopedic surgical navigation and burring system. CORI uses established technologies of navigation via a passive infrared tracking camera. For knee applications, CORI aids the surgeon in planning the surgical implant location and in executing the surgical plan by controlling the cutting engagement of the surgical bur.

For knee applications, CORI software controls the cutting engagement of the surgical bur based on its proximity to the planned target surface. The cutting control is achieved with two modes:

  • Exposure control adjusts the bur's exposure with respect to a guard. If the surgeon encroaches on a portion of bone that is not to be cut, the robotic system retracts the bur inside the guard, disabling cutting.
  • o Speed control regulates the signal going to the tool control unit itself and limits the speed of the drill if the target surface is approached.

Alternatively, the surgeon can disable both controls and operate the robotic drill as a standard navigated surgical drill.

AI/ML Overview

This document, K240139, describes a 510(k) premarket notification for a modification to the REAL INTELLIGENCE™ CORI™ device. The modification integrates an image-based planning option into the Total Knee Arthroplasty (TKA) and Unicondylar Knee Replacement (UKR) applications. The provided text, however, does not include detailed acceptance criteria or a specific study proving the device meets those criteria with performance data in the format requested.

The document primarily focuses on demonstrating substantial equivalence to a predicate device (K231963) by outlining similarities in intended use, indications for use, technological characteristics, and environmental use. It mentions "Design verification and validation testing demonstrated that CORI meets all design requirements" and "Comprehensive testing demonstrated that the system meets required design inputs," but does not provide the specific quantitative acceptance criteria or performance results from these tests.

Therefore, the following points address the requested information based only on what is available in the provided text. Many points will be marked as "Not provided" or "Not applicable" due to the nature of the submission (a 510(k) for a modification focusing on substantial equivalence rather than a full de novo study with detailed performance metrics).


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriterionReported Device Performance
System Meets Design Requirements"CORI meets all design requirements." (No specific quantitative criteria or metrics provided)
Meets Required Design Inputs"Comprehensive testing demonstrated that the system meets required design inputs." (No specific quantitative criteria or metrics provided)
Image-Based Planning Registration is Safe and Effective"The key determining factor in establishing substantial equivalence is whether CORI can register the pre-operative plans safely and effectively." "Comprehensive verification testing demonstrated that the system meets required design inputs." (No specific quantitative metrics for safety and effectiveness of registration, e.g., accuracy, precision, or success rates, are provided.)
Usability (Safe & Effective Use in Simulated Environment)"Summative usability validation testing demonstrating that representative users were able to use the subject device safely and effectively in a simulated use environment." "The study demonstrated that the surgeon users were able to safely and effectively perform tasks necessary to execute TKA and UKA procedures using the CORI system." (No specific quantitative usability metrics, like task completion rates, error rates, or time to completion, are provided.)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not specified for the software verification, design verification, or image-based planning registration testing. For the summative usability validation testing, it mentions "representative users," but the number is not provided.
  • Data Provenance: Not specified. (e.g., country of origin, retrospective/prospective). This is a modification of an existing device, so the testing mainly focuses on the modified functionalities.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

  • Number of Experts: Not specified.
  • Qualifications of Experts: For the usability study, "surgeon users" are mentioned, implying qualified medical professionals, but specific qualifications (e.g., years of experience, specialty) are not detailed. For other testing, the method of establishing ground truth for technical performance (e.g., registration accuracy) is not explicitly described as relying on "experts" in the sense of clinical reviewers, but rather on engineering verification.

4. Adjudication Method for the Test Set

  • Not applicable/Not specified. The document refers to "design verification and validation testing," "software verification testing," and "summative usability validation testing." There is no mention of an adjudication process typically associated with reviewer consensus on medical images or clinical outcomes.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not done/Not applicable. This submission is for a surgical navigation system, not an AI-assisted diagnostic tool that would typically involve a multi-reader, multi-case study to assess human reader performance improvement. The "image-based planning option" implies the import and use of pre-operative imaging data, but the device itself is not described as an AI intended to assist human readers in image interpretation.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Not applicable in the conventional sense of a diagnostic algorithm. The device, REAL INTELLIGENCE™ CORI™, is described as a "robotic-assisted orthopedic surgical navigation and burring system" with an integrated image-based planning option. Its function inherently involves human interaction (a surgeon) and mechanical components. The "standalone" performance would relate to the accuracy of its navigation and burring control, which is implied by "Comprehensive testing demonstrated that the system meets required design inputs," but specific standalone performance metrics are not given.

7. The Type of Ground Truth Used

  • The type of ground truth for general "design requirements" and "design inputs" would typically be engineering specifications, physical measurements, and computational models. For the "image-based planning option," the ground truth for registration accuracy would likely be established metrologically against known anatomical landmarks or fiducials in the pre-operative image data. For usability, the ground truth is simply the ability of users to safely and effectively complete tasks, assessed through observation and user feedback. No mention of pathology or outcomes data in this context.

8. The Sample Size for the Training Set

  • Not applicable/Not specified. The document does not describe the development of a machine learning or AI model with a distinct "training set." The modification is described as integrating an "image-based planning option," implying a capability to import and use image data, rather than a system trained on image data in an AI development pipeline.

9. How the Ground Truth for the Training Set was Established

  • Not applicable. As noted in point 8, no training set for an AI model is described.

{0}------------------------------------------------

March 18, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Herlinger Corrine Senior Principal Regulatory Affairs Specialist 2828 Railroad Street Pittsburgh, Pennsylvania 15222

Re: K240139

Trade/Device Name: Real Intelligence™ CORI™ Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: January 17, 2024 Received: January 18, 2024

Dear Herlinger Corrine:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

K240139 - Herlinger Corrine

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100. Corrective action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali-S

Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K240139

Device Name REAL INTELLIGENCE CORI

Indications for Use (Describe)

REAL INTELLIGENCE CORI is indicated for use in surgical procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be determined. These procedures include: · unicondylar knee replacement (UKR),

  • · total knee arthroplasty (TKA),
  • · revision knee arthroplasty, and
  • total hip arthroplasty (THA).
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for Smith+Nephew. The logo is in orange and features the company name in a bold, sans-serif font. The plus sign is slightly raised and connects the two words.

510(k) Summary

510(k) OwnerBlue Belt Technologies, Inc.2875 Railroad StreetPittsburgh, PA 15222 USATel: (412) 683-3844
Contact PersonCorrine HerlingerSenior Principal Regulatory Affairs SpecialistTel: 412.552.6428Email: corrine.herlinger@smith-nephew.com
Date PreparedMarch 14, 2024
Classification Reference21 CFR 882.4560
Product CodeOLO
Supported CodesHSX, JWH, MBH, NJD
Common/Usual NameOrthopedic Stereotaxic Instrument
Trade/Proprietary NameREAL INTELLIGENCE™ CORI™ (CORI)
Predicate Device(s)REAL INTELLIGENCE™ CORI™ (K231963)
Reason for SubmissionThe purpose of this Traditional 510(k) submission isto seek clearance for a modification to CORI tointegrate an image-based planning option into theTotal Knee Arthroplasty (TKA) and UnicondylarKnee Replacement (UKR) applications.

{4}------------------------------------------------

Smith+Nephew

Intended Use

REAL INTELLIGENCE CORI is intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.

Indications for Use

REAL INTELLIGENCE CORI is indicated for use in surgical procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be determined. These procedures include:

  • unicondylar knee replacement (UKR),
  • . total knee arthroplasty (TKA),
  • revision knee arthroplasty, and
  • . total hip arthroplasty (THA).

Device Description

The subject of this Traditional 510(k) is REAL INTELLIGENCE CORI (CORI), a robotic-assisted orthopedic surgical navigation and burring system. CORI uses established technologies of navigation via a passive infrared tracking camera. For knee applications, CORI aids the surgeon in planning the surgical implant location and in executing the surgical plan by controlling the cutting engagement of the surgical bur.

For knee applications, CORI software controls the cutting engagement of the surgical bur based on its proximity to the planned target surface. The cutting control is achieved with two modes:

  • Exposure control adjusts the bur's exposure with respect to a guard. If the surgeon encroaches on a portion of bone that is not to be cut, the robotic system retracts the bur inside the guard, disabling cutting.
  • o Speed control regulates the signal going to the tool control unit itself and limits the speed of the drill if the target surface is approached.

Alternatively, the surgeon can disable both controls and operate the robotic drill as a standard navigated surgical drill.

{5}------------------------------------------------

ew

Blue Belt Technologies, Inc. is a subsidiary of Smith & Nephew, Inc.

Currently Supported Knee Implants

The following Smith+Nephew knee implants are supported on CORI:

Table 1: Currently Supported Smith+ Nephew Knee Implants

Implant Model Name510(k) NumberClassificationProduct Code
STRIDE Unicondylar KneeK123380HSX
ZUK Select Knee SystemK160738HSX
JOURNEY II Unicompartmental Knee SystemK191211HSX
JOURNEY UNIK102069HSX
JOURNEY II CRK121443JWH
JOURNEY II BCSK111711JWH
JOURNEY II XRK141471, K152726JWH
LEGION CR/PSK951987, K962557, K093746JWH
LEGION Porous CR Femoral ComponentsK073325, K091543MBH
LEGION Porous CR Narrow Femoral ComponentsK210566MBH
LEGION Porous TibiaK100897MBH
Porous Tibia BaseplateK211221MBH
GENESIS II CR/PSK951987, K962557JWH
ANTHEMK142807JWH
SMITH & NEPHEW, INC. REVISION KNEE SYSTEMK043440JWH
REVISION KNEE SYSTEMK041106JWH
LEGION RK TIBIAL WEDGES (Hemi-Step & Full-Step)K953274JWH
LEGION COBALT CHROME REVISION KNEE SYSTEMK060742JWH
LEGION Knee SystemK180334JWH, MBH
ENGAGE Partial Knee SystemK190439NJD, HSX
JOURNEY II UK™ and ENGAGE™ Cementless PartialKnee SystemK222653NJD, HSX

{6}------------------------------------------------

Smith+Nephew

Discussion of Similarities and Differences

This Traditional 510(k) submission supports an update to integrate an image-based planning option into the TKA and UKR applications. The proposed modification to CORI will allow the CORI 3.0 software to import a pre-operative plan if desired by the surgeon.

The modifications made to support the change do not impact the system's intended use, indications for use, or fundamental scientific technology.

Blue Belt Technologies believes that CORI is subject to premarket notification requirements under Section 510(k) of the Federal Food, Drug and Cosmetic Act ("FFDCA" or "the Act") and is substantially equivalent to the previously cleared REAL INTELLIGENCE CORI (K221224).

Table 1: Predicate Device

ManufacturerDescriptionSubmission NumberClearance Date
Blue Belt Technologies, Inc.REAL INTELLIGENCE CORIK23196308/01/2023
FeatureSubject DeviceCORIPrimary PredicateCORI – K231963
Intended UseSame as predicate.REAL INTELLIGENCE CORI (CORI) is intended toassist the surgeon in providing software-defined spatial boundaries for orientation andreference information to anatomical structuresduring orthopedic procedures.
Indications forUseSame as predicate.CORI is indicated for use in surgicalprocedures, in which the use of stereotacticsurgery may be appropriate, and wherereference to rigid anatomical bony structurescan be determined. These procedures include:• unicondylar knee replacement (UKR),• total knee arthroplasty (TKA),• revision knee arthroplasty, and• total hip arthroplasty (THA).
Knee ImplantProduct CodesSupportedSame as predicate.HSX, JWH, MBH, NJD
Environment ofUseSame as predicate.CORI is intended to be used by trained medicalprofessionals in a hospital or clinical settingequivalent to an orthopedic surgery suite.

Table 2: Summary of Technological Similarities with Predicate

{7}------------------------------------------------

Image /page/7/Picture/2 description: The image shows the logo for Smith+Nephew. The logo is in orange font. The plus sign is slightly raised between the two names.

FeatureSubject DeviceCORIPrimary PredicateCORI - K231963
TechnologicalCharacteristicsFor knee applications, CORI uses establishedtechnologies to prepare bone for attachmentof UKR and TKA implant components. In thecase of a total knee arthroplasty, the bonesurface may also be prepared to receive thefemoral and tibial cutting guides.For knee applications, CORI uses establishedtechnologies to prepare bone for attachmentof UKR and TKA implant components. In thecase of a total knee arthroplasty, the bonesurface may also be prepared to receive thefemoral and tibial cutting guides.
CORI uses either pre-operative data (forimage-based cases) or intraoperative datacollection (for image-free or non-CT datageneration cases) to create a model of thepatient's femur and tibia in CORI trackingspace and allows the surgeon toprepare/modify the surgical plan.CORI uses intraoperative data collection(image-free or non-CT data generation) tocreate a model of the patient's femur and tibiain CORI tracking space and allows the surgeonto prepare/modify a surgical plan.
The system uses predefined boundariesgenerated during the planning process tocontrol the motion of the surgical bur andlimit the amount of bone removed to shapethe operative condyle or tibial plateau inpreparation for placement of the surgicalimplant.The system uses predefined boundariesgenerated during the planning process tocontrol the motion of the surgical bur and limitthe amount of bone removed to shape thecondyles or tibial plateau in preparation forplacement of the surgical implant.
Bur cutting is controlled either by retractingthe bur in a guard, or by controlling the speedof the bur as the target surface isapproached.Bur cutting is controlled either by retractingthe bur in a guard, or by controlling the speedof the bur as the target surface is approached.To support the hip application, CORI uses aVirtual Machine with hypervisor to enable the
To support the hip application, CORI uses aVirtual Machine with hypervisor to enable theWindows-based HIP7 software to run onCORI.Windows-based HIP7 software to run on CORI.

{8}------------------------------------------------

Smith+Nephew

Non-Clinical Testing (Bench)

Design verification and validation testing demonstrated that CORI meets all design requirements and is as safe and effective as its predicate device (K231963). Comprehensive testing demonstrated that the system meets required design inputs. Additionally, the following evidence was provided:

  • . Software verification testing, including software integration and workflow testing, was completed. Software was developed in accordance with IEC 62304 Medical device software -Software life cycle processes, and this submission contains documentation per the requirements of FDA's Guidance for the Content of Premarket Submissions for Device Software Functions.
  • . Summative usability validation testing demonstrating that representative users were able to use the subject device safely and effectively in a simulated use environment. Human factors and usability engineering processes were followed per IEC 62366 - Application of Usability Engineering to Medical Devices.

Per Human Factors evaluation, updates made to the graphical user interface and software workflows for TKA and UKA in support of the RI.KNEE-v3.0.0 software were found to be safe and effective for the intended users, uses, and use environments. The study demonstrated that the surgeon users were able to safely and effectively perform tasks necessary to execute TKA and UKA procedures using the CORI system.

Conclusion

The subject device, CORI, described in this submission has the same intended use, indications for use, and fundamental scientific technology as the predicate device, CORI (K231963). The difference between the two systems is an update to CORI to integrate an image-based planning option into the Total Knee Arthroplasty (TKA) and Unicondylar Knee Replacement (UKR) applications. The proposed modification to CORI will allow the CORI 3.0 software to import a pre-operative plan if desired by the surgeon.

The key determining factor in establishing substantial equivalence is whether CORI can register the pre-operative plans safely and effectively. Comprehensive verification testing demonstrated that the system meets required design inputs. Summative usability validation testing established that representative users can use the subject device safely and effectively in a simulated use environment.

The information presented in this 510(k) premarket notification demonstrates that CORI with image-based planning is as safe and effective as the predicate device (K231963). Blue Belt Technologies believes that FDA can find CORI to be substantially equivalent to the predicate device.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).