K Number
K240112

Validate with FDA (Live)

Date Cleared
2024-03-15

(59 days)

Product Code
Regulation Number
872.3690
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Anterior restorations (Class III, IV) Class V restorations (cervical caries, root erosion, wedge-shaped defects) Restorations in the posterior region (Class I and II) Veneering of discolored anterior teeth Splinting of mobile teeth Repair of composite and ceramic veneers

Device Description

Light Cure Composite is a state-of-the-art light-curing, radiopaque, nano-hybrid composite for the restorative therapy. The device cures with light at the wavelength range of 400–500 nm (blue light). The device is paste contained in PP syringe.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the device:

1. Table of Acceptance Criteria and Reported Device Performance

Test ItemAcceptance CriteriaReported Device Performance
Outward appearance-ISO 4049According to visual inspection, the product should be uniform in texture and should be a paste without visible impurities and foreign matter.Meet the requirement
Sensitivity to ambient light-ISO 4049The material of sample remains physically homogeneous.Meet the requirement
Depth of Cure-ISO 4049≥1.5mm≥1.5mm
Flexural strength-ISO 4049≥80Mpa≥80Mpa
Water sorption-ISO 4049≤40µg/mm³≤40µg/mm³
solubility-ISO 4049≤7.5µg/mm³≤7.5µg/mm³
Shade-ISO 4049Without the need for a magnifying glass to observe with the naked eye, the cured material should be evenly colored.Meet the requirement
Color comparison for color stability-ISO 4049Only slight changes are allowed.Meet the requirement
Appearance of container-ISO 4049Visually observe that the surface of the container should be smooth, without scratches, cracks, burrs, small notches or macroscopic defects caused by mechanical processing.Meet the requirement
Delivery performance of container-ISO 4049Remove the syringe cap and turn the container push rod clockwise allowing resin to be smoothly squeezed out of the syringe without obstruction.Meet the requirement
Compressive Strength≥ 200 MpaMeet the requirement (Mean: 286.5)
Elastic Modulus≥ 6 GpaMeet the requirement (Mean: 8.5)
Surface Hardness≥ 60 (HV)Meet the requirement (Mean: 63.1)
Particle size0.4-2μm(Value not explicitly stated as "meet requirement," but compared in section 6 as 0.4-2μm)
Radio-opacityMean: 2.27mm (This value is provided as a characteristic, not an acceptance criterion with a pass/fail. However, it is explicitly listed as a 'target' value.)Mean: 2.27mm
BiocompatibilityMeet ISO 10993-1 - Biological evaluation of medical devices — Part 1: Evaluation and testing.Meet ISO 10993-1

2. Sample Size Used for the Test Set and Data Provenance

The provided document does not explicitly state the sample sizes used for each specific test in the "Non-clinical performance Data" section. It lists performance results but not the number of units tested.

The data provenance is from non-clinical performance testing, likely conducted by the manufacturer, Sincera Technology (Changchun) Co., Ltd. The country of origin for the testing would be P.R. China, based on the submitter's address. The data is prospective in the sense that it was generated for the purpose of demonstrating the device's performance against established standards.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This document describes a non-clinical performance study of a dental material, not a diagnostic device requiring human interpretation of images or clinical data. Therefore, there is no mention of "experts" or "ground truth" in the context of clinical interpretation for a test set. The "ground truth" for these tests is implicitly defined by the ISO standards themselves, which specify measurement methodologies and acceptance criteria.

4. Adjudication Method for the Test Set

As this is a non-clinical performance study against ISO standards, no adjudication method (like 2+1 or 3+1) is applicable or mentioned. The results are based on direct measurements and observations relative to the specified criteria.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done or is relevant to this type of device (Light Cure Composite). This is a material science and engineering performance evaluation, not a diagnostic imaging device requiring human reader input or comparison.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Not applicable. This device is a physical dental material, not a software algorithm.

7. Type of Ground Truth Used

The ground truth used for this study is based on established international standards (ISO 4049 and ISO 10993-1). These standards define the test methodologies and the acceptable performance limits for the specific physical and chemical properties of the dental composite material.

8. Sample Size for the Training Set

Not applicable. This is a hardware/material device, not an AI/ML algorithm that requires a "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable. As there is no training set for an algorithm, there's no ground truth to establish for it in this context.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 15, 2024

Sincera Technology (Changchun) Co., Ltd. % Eva Li Consultant Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave. Shanghai, China

Re: K240112

Trade/Device Name: Light Cure Composite Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: January 16, 2024 Received: January 16, 2024

Dear Eva Li:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bobak Shirmohammadi -S

For Michael E. Adjodha, M.ChE., RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices

{2}------------------------------------------------

OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K240112

Device Name Light Cure Composite

Indications for Use (Describe) Anterior restorations (Class III, IV) Class V restorations (cervical caries, root erosion, wedge-shaped defects) Restorations in the posterior region (Class I and II) Veneering of discolored anterior teeth Splinting of mobile teeth Repair of composite and ceramic veneers

Type of Use (Select one or both, as applicable)|X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) Summary

1. Information of the submitter:

Submitter's Name: Sincera Technology (Changchun) Co., Ltd. Address: No. 177 Changda Rd, Hi-tech District, Changchun City, Jilin Province, P. R. China Contact person: Tony Su

Tel: +8615143153582

Email: tony.su@sinceradental.com

Preparation Date: Jan 2nd, 2024

2. Basic information of the device

Classification Name: Material, Tooth Shade, Resin Common / Usual Name: Tooth shade resin material. Proprietary Name: Light Cure Composite Product Code: EBF Class: Class II Regulation: 21 CFR 872.3690

3. Predicate Device

K042819 TETRIC EVOCERAM VOCLAR VIVADENT, INC.

4. Indication for Use

Anterior restorations (Class III, IV) Class V restorations (cervical caries, root erosion, wedge-shaped defects) Restorations in the posterior region (Class I and II) Veneering of discolored anterior teeth Splinting of mobile teeth Repair of composite and ceramic veneers

5. Device Description

Light Cure Composite is a state-of-the-art light-curing, radiopaque, nano-hybrid composite for the restorative therapy. The device cures with light at the wavelength range of 400–500 nm (blue light). The device is paste contained in PP syringe.

ItemSubject devicePredicated deviceComparison
Device NameLight Cure CompositeTETRIC VOCERAMN/A
510(K) No.K042819N/A
ClassificationClass IIClass IISame
Regulation21 CFR 872.369021 CFR 872.3690Same
Indication for useAnterior restorationsAnterior restorations (Class)Same
No. 177 Changda Rd, Hi-tech District, Changchun City, Jilin Province, P. R. China
(Class III, IV)Class V restorations(cervical caries, rooterosion, wedge-shapeddefects)Restorations in theposterior region (Class Iand II)Veneering of discoloredanterior teethSplinting of mobile teethRepair of composite andceramic veneers(Class III, IV)Class V restorations(cervical caries, rooterosion, wedge-shapeddefects)Restorations in theposterior region (Class Iand II)Veneering of discoloredanterior teethSplinting of mobile teethRepair of composite andceramic veneers
Product CodeEBFEBFSame
Patient PopulationAll the groupsAll the groupsSame
Prescription UseRx onlyRx onlySame
EnvironmentDental prosthetics andauthorized laboratoriesand clinicsDental prosthetics andauthorized laboratoriesand clinicsSame
Applicable StandardsISO 4049; ISO 10993-1ISO 4049; ISO 10993-1Same
Device SterilizationNoNoSame
Shelf life2 years3 yearsDifferent*1
Light curing unitPower: ≥ 400mW/cm2Wavelength: 400-500nmCuring time:20sPower: ≥500mW/cm2Wavelength: 400-500nmCuring time:20sDifferent*2
particle size0.4-2μm0.4-3μmDifferent*3
Elastic Modulus(GPa)Mean:8.5Mean:7.4Different*3
CompressiveStrength(MPa)Mean:286.5Mean:248.4Different*3
SurfaceMean:63.1Mean:53.8Different*3
Hardness(HV)
Radio-opacityMean:2.27mmNot providedN/A

6. Technological Characteristics Comparison:

{5}------------------------------------------------

7. Non-clinical performance Data:

Test ItemAcceptance criterialResult
Outward appearance-ISO 4049According to visual 1nspect1on, the product should be uniform in texture and should be a paste without visible impurities and foreign matter.Meet the requirement
Sensitivity to ambient light-ISO 4049The material of sample remains physically homogeneous.Meet the requirement
Depth of Cure-ISO 4049≥1.5mm≥1.5mm
Flexural strength-ISO 4049≥80Mpa≥80Mpa
Water sorption-ISO 4049≤40µg/mm³≤40µg/mm³
solubility-ISO 4049≤7.5µg/mm³≤7.5µg/mm³

{6}------------------------------------------------

Sincera Technology (Changchun) Co., Ltd. No. 177 Changda Rd, Hi-tech District, Changchun City, Jilin Province, P. R. China

Shade-ISO 4049Without the need for a magnifying glass to observe with the naked eye, the cured material should be evently coloredkMeet the requirement
Color comparison for color stability-ISO 4049Only slight changes are allowedMeet the requirement
Appearance of container-ISO 4049Visually observe that the surface of the container should be smooth, without scratches, cracks, burrs, small notches or macroscopic defects caused by mechanical processing.Meet the requirement
Delivery performance of container-ISO 4049Remove the syringe cap and turn the container push rod clockwise allowing resin to be smoothly squeezed out of the syringe without obstructionMeet the requirement
Compressive Strength$\ge$ 200 MpaMeet the requirement
Elastic Modulus$\ge$ 6 GpaMeet the requirement
Surface Hardness$\ge$ 60(HV)Meet the requirement
particle size0.4-2μm
Radio-opacityMean: 2.27mm

8. Biocompatibility

Meet ISO 10993-1 - Biological evaluation of medical devices — Part 1: Evaluation and testing.

9. Conclusion

The results of the testing described above provide reasonable assurance that the subject device is as safe and effective as the predicate device(s) which cleared under K042819 and supports a determination of substantial equivalence.

§ 872.3690 Tooth shade resin material.

(a)
Identification. Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.(b)
Classification. Class II.