K Number
K240035

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-08-01

(210 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TACTIX Vector Syndesmosis System is intended as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated, and as an adjunct in external and intramedullary fixation systems involving plates and rods, with fracture braces and casting.

Specifically, the TACTIC Vector Syndesmosis System is intended to provide fixation during the healing process following a syndesmotic trauma, such as fixation of syndesmosis disruptions) in connection with Weber B and C ankle fractures.

Device Description

The TACTIX Vector Syndesmosis System is intended to provide fixation during the healing process following a syndesmotic trauma using a TACTIX Vector suture loop and buttons.

The TACTIX Vector System features the following advantageous characteristics: a tensioner to apply suture tension for surgical repair, line lock suture technology without the need to tie a knot, removal of periosteum through a minimal medial button, and a small thru hole in the bone, only large enough to pass a 2.0 mm suture loop. The TACTIX Vector Syndesmosis System can be used in conjunction with Vilex fibular bone plates and screws, as deemed necessary by the surgeon.

The TACTIX Vector Syndesmosis System is composed of two titanium alloy buttons and one UHMWPE suture.

The instruments of TACTIX System are composed of stainless steel, UHMWPE, and nylon material.

The TACTIX Vector Syndesmosis System is intended for single use and provided sterile to the end user via ethylene oxide sterilization.

AI/ML Overview

This document is for a medical device called the "TACTIX Vector Syndesmosis System," which is a metallic bone fixation appliance. It is a 510(k) premarket notification to the FDA. The document focuses on establishing substantial equivalence to previously cleared predicate devices.

The information provided does not describe a study that proves the device meets acceptance criteria in the context of clinical performance or diagnostic accuracy, as would be relevant for an AI/ML-driven medical device. Instead, the document describes non-clinical (bench) testing to demonstrate the device's mechanical performance and safety characteristics.

Therefore, for aspects related to "acceptance criteria" for performance metrics like sensitivity, specificity, AUC, human reader improvement with AI, or ground truth establishment relevant to AI/ML, there is no information provided in this document.

However, I can extract the information relevant to the non-clinical testing performed:

1. Table of Acceptance Criteria and Reported Device Performance:

Test MethodAcceptance Criteria (Implicit)Reported Device Performance
Cyclic fatigue testingTACTIX Vector Syndesmosis System performs comparably to or better than the additional predicate device (K173278)."The nonclinical testing showed that the TACTIX Vector Syndesmosis System performs substantially equivalent to a predicate device."
Static pullout testingTACTIX Vector Syndesmosis System performs comparably to or better than the additional predicate device (K173278)."The nonclinical testing showed that the TACTIX Vector Syndesmosis System performs substantially equivalent to a predicate device."
Stability testing on product packagingEnsure packaging maintains device integrity and sterility over shelf life.Performed; conclusions support substantial equivalence.
Packaging performance testingEnsure packaging adequately protects the device during transit and handling.Performed; conclusions support substantial equivalence.
Biocompatibility evaluation (per ISO 10993-1)Device materials are biocompatible and do not elicit adverse biological responses.Performed; conclusions support substantial equivalence.
Bacterial endotoxin testingDevice is free of unacceptable levels of bacterial endotoxins.Performed; conclusions support substantial equivalence.

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: Not explicitly stated for each non-clinical test (e.g., number of devices tested for fatigue, pullout).
  • Data Provenance: The tests are non-clinical (bench testing) performed by the manufacturer (Vilex LLC). The document does not specify the country of origin for these tests. The tests are "retrospective" in the sense that they are done in a lab setting to assess the final device design.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. This document describes non-clinical engineering bench testing, not clinical studies or AI/ML performance evaluations requiring expert ground truth for interpretation.

4. Adjudication method for the test set:

  • Not applicable. This document describes non-clinical engineering bench testing.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, not applicable. This document describes a medical device (physical implant), not an AI/ML diagnostic or assistive tool. No MRMC study was mentioned or performed.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • No, not applicable. This document is for a physical medical device.

7. The type of ground truth used:

  • Not applicable in the AI/ML context. For the non-clinical tests, the "ground truth" is established by standard engineering test methods and their defined pass/fail criteria (e.g., meeting a certain force threshold, cycles to failure, absence of toxins). Comparison to a predicate device serves as the benchmark for "truth."

8. The sample size for the training set:

  • Not applicable. This document does not describe an AI/ML device that requires a training set.

9. How the ground truth for the training set was established:

  • Not applicable. This document does not describe an AI/ML device that requires a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 1, 2024

Vilex LLC Brock Johnson President 111 Moffitt Street McMinnville, Tennessee 37110

Re: K240035

Trade/Device Name: TACTIX Vector Syndesmosis System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HTN Dated: June 28, 2024 Received: July 11, 2024

Dear Mr. Johnson:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4. Subpart B) for combination products (see https://www.fda.gov/combinationproducts/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-reportmedical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuingeducation/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumereducation-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Christopher Ferreira -S

for

Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K240035

Device Name

TACTIX Vector Syndesmosis System

Indications for Use (Describe)

The TACTIX Vector Syndesmosis System is intended as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated, and as an adjunct in external and intramedullary fixation systems involving plates and rods, with fracture braces and casting.

Specifically, the TACTIC Vector Syndesmosis System is intended to provide fixation during the healing process following a syndesmotic trauma, such as fixation of syndesmosis disruptions) in connection with Weber B and C ankle fractures.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) #:K240035510(k) SummaryK240035Prepared on: 2024-07-29
----------------------------------------------------------------------------

21 CFR 807.92(a)(1)

Contact Details
Applicant NameVilex LLC
Applicant Address111 Moffitt Street McMinnville TN 37110 United States
Applicant Contact Telephone8019164157
Applicant ContactMr. Brock Johnson
Applicant Contact Emailbrock.johnson@vilex.com

21 CFR 807.92(a)(2)

Device Name
Device Trade NameTACTIX Vector Syndesmosis System
Common NameSingle/multiple component metallic bone fixation appliances and accessories
Classification NameWasher, Bolt Nut
Regulation Number888.3030
Product CodeHTN

21 CFR 807.92(a)(3)

Legally Marketed Predicate Devices
Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K201522Arthrex Syndesmosis TightRope XP Buttress Plate Implant SysteHTN
K173278ToggleLoc System (Only ToggleLoc with ZipTight Fixation DeviceMBI
K230204Alphalok Plating System (reference device)HRS

21 CFR 807.92(a)(4)

Device Description Summary
----------------------------

The TACTIX Vector Syndesmosis System is intended to provide fixation during the healing process following a syndesmotic trauma using a TACTIX Vector suture loop and buttons.

The TACTIX Vector System features the following advantageous characteristics: a tensioner to apply suture tension for surgical repair, line lock suture technology without the need to tie a knot, removal of periosteum through a minimal medial button, and a small thru hole in the bone, only large enough to pass a 2.0 mm suture loop. The TACTIX Vector Syndesmosis System can be used in conjunction with Vilex fibular bone plates and screws, as deemed necessary by the surgeon.

Intended Use/Indications for Use

21 CFR 807.92(a)(5)

The TACTIX Vector Syndesmosis System is intended as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated, and as an adjunct in external and intramedullary fixation systems involving plates and rods, with fracture braces and casting.

Specifically, the TACTIC Vector Syndesmosis System is intended to provide fixation during the healing process following a syndesmotic trauma, such as fixation of syndesmosis disruptions) in connection with Weber B and C ankle fractures.

{4}------------------------------------------------

Indications for Use Comparison

The indications for use of the TACTIX Vector Syndesmosis System are identical to the primary predicate device.

Technological Comparison

The TACTIX Vector Syndesmosis System is composed of two titanium alloy buttons and one UHMWPE suture. The predicate devices are also composed of two titanium alloy buttons and one UHMWPE suture. The TACTIX Vector suture and buttons have the same basic design and intended use as the suture and buttons of the predicate devices.

The TACTIX Vector Syndesmosis System uses a tension on the implanted suture loop. The predicate devices also utilize an instrument for the same intended use. The TACTIX Vector System uses a threader rod to install the suture. The predicate devices also use an instrument for the same intended use. The instruments of TACTIX System are composed of stainless steel, UHMWPE, and nylon material. The instruments of the predicate devices are composed of stainless materials.

The TACTIX Vector Syndesmosis System is intended for single use and provided sterile to the end user via ethylene oxide sterilization. The predicate devices are intended for single use and provided sterile to the end user via ethylene oxide sterilization.

Therefore, the technological characteristics of the TACTIX Vector System are substantially equivalent to those of the predicate device. The differences in the technological characteristics between the TACTIX Vector System and the predicate device are considered minor and raise no questions of safety or effectiveness.

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

The following nonclinical tests were performed for the TACTIX Vector Syndesmosis System:

  • Cyclic fatigue testing compared against the additional predicate device (K173278)

  • Static pullout testing compared against the additional predicate device (K173278)

  • Stability testing on product packaging

  • Packaging performance testing

  • Biocompatibility evaluation per ISO 10993-1

  • Bacterial endotoxin testing

No clinical data was used used in the support of this submission.

The noncinical testing showed that the TACTIX Vector Syndesmosis System performs substantially equivalent to a predicate device. Therefore, the TACTIX Vector Syndesmosis System raises no new questions of safety or effectiveness and is substantially equivalent to the predicate devices.

21 CFR 807.92(a)(5)

21 CFR 807.92(a)(6)

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.