K Number
K240007

Validate with FDA (Live)

Date Cleared
2024-05-06

(125 days)

Product Code
Regulation Number
876.1725
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Solar Compact is intended to record, store, view, and analyze pressure, impedance, and EMG data, gathered from anywhere in the gastrointestinal tract (pharynx, esophagus, stomach, and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Designated catheters and accessories are required for measurement in each specific area of the gastrointestinal tract.

An optional perfusion pump can also be used for automated balloon filling (for anorectal manometry studies). The filling lumen of the catheter can be connected to the perfusion pump.

The Solar Compact is indicated for use in adult to infants with common gastrointestinal conditions, for the monitoring and diagnosis of motility disorders of the GI tract.

Device Description

The Solar Compact is a medical device for measuring gastrointestinal motility characteristics. The system consists of an electromechanical hardware device which connects to pressure catheters (air-charged or solid-state), and a PC software. Catheters provide measurements of pressure and impedance to the device from various sites along the gastrointestinal tract, which are then displayed and analyzed by the Laborie PC software. This provides the clinician with data that may aid in the diagnosis or management of various gastrointestinal disorders. Additionally, the system connects to provide EMG measurements of the muscles along the pelvic floor. Devices such as pressure catheters and surface electrodes are required to complete a gastrointestinal motility investigation, but they are not considered part of the Solar Compact device and not included standard with the system.

The device hardware consists of an external housing, pneumatics to inflate and deflate rectal balloons, and electronics to readout data from the pressure catheters and to control all other functions of the system. The device connects to a PC via Bluetooth to display measurements within the Laborie software. The Laborie software contains a patient database module, and an analysis module. The software displays measurements from the system as time-tracing based plots, Clouse Countour Plots (High Resolution Manometry), and as 3D plots. Clinicians can analyze the data provided by placing markers and using calculations within the software. The device does not provide an independent diagnosis.

The Solar Compact device is compatible with a number of of catheters for these aforementioned parameter measurements. For a full list of these compatible catheters please consult the instructions for use.

AI/ML Overview

The provided text is an FDA 510(k) clearance letter and summary for a medical device called "Solar Compact (G4-1)". It contains information about the device's indications for use, technological characteristics, and comparison to predicate devices, but it explicitly states that "Clinical testing was not conducted and is not applicable for demonstrating substantial equivalence."

Therefore, I cannot provide the requested information about acceptance criteria and a study proving the device meets those criteria, as no such clinical study was performed or needed for this specific 510(k) clearance based on the provided document.

To clarify, the document focuses on demonstrating substantial equivalence to existing legally marketed devices (predicates) through non-clinical testing and comparison of technological characteristics. This pathway to market typically does not require a full clinical study with acceptance criteria like those outlined in the request, especially for devices that are considered well-understood or incremental improvements.

The document does mention "System verifications were conducted, including functional, electrical, lifetime verification testing, calibration and volume measurement, to ensure that the Solar Compact demonstrates by meeting requirements defined for the device." These "requirements defined for the device" are internal engineering and design specifications, not clinical acceptance criteria derived from a human study.

Therefore, since the document explicitly states "Clinical testing was not conducted and is not applicable," it's impossible to extract:

  1. A table of acceptance criteria and reported device performance from a clinical study.
  2. Sample size used for the test set and data provenance from a clinical study.
  3. Number of experts used to establish ground truth or their qualifications from a clinical study.
  4. Adjudication method from a clinical study.
  5. MRMC study details or effect size from a clinical study.
  6. Standalone performance details from a clinical study.
  7. Type of ground truth used from a clinical study.
  8. Sample size for the training set from a clinical study.
  9. How the ground truth for the training set was established from a clinical study.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

May 6, 2024

Laborie Medical Technologies Corp. Ingrid Dirtzu Regulatory Affairs Specialist II 180 International Drive Portsmouth, New Hampshire 03801

Re: K240007

Trade/Device Name: Solar Compact (G4-1) Regulation Number: 21 CFR 876.1725 Regulation Name: Gastrointestinal Motility Monitoring System Regulatory Class: Class II Product Code: FFX Dated: April 5, 2024 Received: April 5, 2024

Dear Ingrid Dirtzu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Shanil P. Haugen -S

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K240007

Device Name

Solar Compact (G4-1)

Indications for Use (Describe)

The Solar Compact is intended to record, store, view, and analyze pressure, impedance, and EMG data, gathered from anywhere in the gastrointestinal tract (pharynx, esophagus, stomach, and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Designated catheters and accessories are required for measurement in each specific area of the gastrointestinal tract.

An optional perfusion pump can also be used for automated balloon filling (for anorectal manometry studies). The filling lumen of the catheter can be connected to the perfusion pump.

The Solar Compact is indicated for use in adult to infants with common gastrointestinal conditions, for the monitoring and diagnosis of motility disorders of the GI tract.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) #:K240007K240007 Page 1 of 2 Prepared on: 2024-04-18
-------------------------------------------------------------------

510(k) Summary
Contact Details

21 CFR 807.92(a)(1)
Applicant NameLaborie Medical Technologies Corp.
Applicant Address180 International Drive Portsmouth NH 03801 United States
Applicant Contact Telephone952-564-3948
Applicant ContactMs. Ingrid Dirtzu
Applicant Contact Emailidirtzu@laborie.com

Device Name

21 CFR 807.92(a)(2)
Device Trade NameSolar Compact (G4-1)
Common NameGastrointestinal motility monitoring system
Classification NameSystem, Gastrointestinal Motility (Electrical)
Regulation Number876.1725
Product CodeFFX

Legally Marketed Predicate Devices

21 CFR 807.92(a)(3)
-----------------------
Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K052338Solar Gl with CIM-AUXFFX
K121014MPP PlusFFX

Device Description Summary

21 CFR 807.92(a)(4)
-----------------------

The Solar Compact is a medical device for measuring gastrointestinal motility characteristics. The system consists of an electromechanical hardware device which connects to pressure catheters (air-charged or solid-state), and a PC software. Catheters provide measurements of pressure and impedance to the device from various sites along the gastrointestinal tract, which are then displayed and analyzed by the Laborie PC software. This provides the clinician with data that may aid in the diagnosis or management of various gastrointestinal disorders. Additionally, the system connects to provide EMG measurements of the muscles along the pelvic floor. Devices such as pressure catheters and surface electrodes are required to complete a gastrointestinal motility investigation, but they are not considered part of the Solar Compact device and not included standard with the system.

The device hardware consists of an external housing, pneumatics to inflate and deflate rectal balloons, and electronics to readout data from the pressure catheters and to control all other functions of the system. The device connects to a PC via Bluetooth to display measurements `within the Laborie software. The Laborie software contains a patient database module, and an analysis module. The software displays measurements from the system as time-tracing based plots, Clouse Countour Plots (High Resolution Manometry), and as 3D plots. Clinicians can analyze the data provided by placing markers and using calculations within the software. The device does not provide an independent diagnosis.

The Solar Compact device is compatible with a number of of catheters for these aforementioned parameter measurements. For a full list of these compatible catheters please consult the instructions for use.

Intended Use/Indications for Use

{5}------------------------------------------------

The Solar Compact is intended to record, store, view, and analyze pressure, impedance, and EMG data, gathered from anywhere in the gastrointestinal tract (pharynx, esophagus, and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Designated catheters and accessories are required for measurement in each specific area of the gastrointestinal tract.

An optional perfusion pump can also be used for automated balloon filling (for anorectal manometry studies). The filling lumen of the catheter can be connected to the perfusion pump.

The Solar Compact is indicated for use in adult to infants with conditions, for the monitoring and diagnosis of motility disorders of the GI tract.

Indications for Use Comparison

The indications for use of the proposed Solar Compact are similar to that of the primary predicate device, Solar Gl with ClM-AUX (K052338). While the number of sites for investigation by way of example has been limited, the Solar Compact device retains the statement that it can be used to analyze data anywhere in the gastrointestinal tract, and in this regard the devices are similar.

The signals analyzed by the proposed Solar Compact are limited relative to the Solar Gl with CIM-AUX and do not specify swallow or respiration. The proposed Solar Compedance data within the indications as a form of auxiliary device input data which the system analyzes. However, the predicate device, Solar Gl with CM-AUX, has incorporated impedance measurement since the Solar GI with CIM-AUX (K052338) clearance.

The MPP Plus (K121014) is an accessory to a system whose primary intended use is as an accessory to a complete gastrointestinal motility system, whereas the proposed is a complete gastrointestinal motility system. The proposed Solar Compact incorporates the same functionality of automated balloon filling that specified for the predicate MPP Plus (K121014), in order to achieve the system's intended use of gastrointestinal motility investigations.

Technological Comparison

The principles of operation, energy source, measuring range, and materials used in the proposed Solar Compact are similar to those used in the predicate devices.

The proposed Solar Compact features a smaller device form factor relative to primary predicate Solar Gl with CM-AUX (K052338). Additionally, Bluetooth functionality is incorporated drectly within the main housing of the than as a accessory module.

For the proposed Solar Compact, the pump is incorporated within the body of the than as an accessory item as on the MPP Plus predicate (K121014). The pump system in the proposed Solar Compact is an air compressor and does not perfuse water, whereas the pump in the MPP Plus pressurizes a water-perfusion system for compatibility with water-perfused catheters.

The proposed Solar Compact is compatible with solid-state and air-charged catheters, which is identical to the primary predicate, Solar GI with CIM-AUX (K052338). The Solar Gl with MPP Plus (K121014) is compatible with water catheters only; the Solar Gl with MPP Plus is a secondary predicate and is considered not applicable in this regard.

The pump in the proposed Solar Compact is able to control both inflation and deflation, whereas the pump in the predicate MPP Plus (K121014) controls inflation only and does not have a controlled deflation.

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

System verifications were conducted, including functional, electrical, lifetime verification testing, calibration and volume measurement, to ensure that the Solar Compact demonstrates by meeting requirements defined for the device. Additionally, software verfitcations were conducted to verify system requirements during device development and an additional software verification was performed to confirm the Solar Compact software into the Laborie software 10.1 release. Package integrity testing was performed to the ISTA 3A standard and electrical/mechanical safety testing was performed to demonstrate compliance with IEC 60601-1 and 60601-1-2.

Clinical testing was not conducted and is not applicable for demonstrating substantial equivalence.

The Solar Compact device passed all verification testing, confirming the device performance and demonstrating that the device is as safe, as effective, and perform as as well or better than the legally marketed device predicates.

21 CFR 807.92(a)(5)

21 CFR 807.92(a)(6)

§ 876.1725 Gastrointestinal motility monitoring system.

(a)
Identification. A gastrointestinal motility monitoring system is a device used to measure peristalic activity or pressure in the stomach or esophagus by means of a probe with transducers that is introduced through the mouth into the gastrointestinal tract. The device may include signal conditioning, amplifying, and recording equipment. This generic type of device includes the esophageal motility monitor and tube, the gastrointestinal motility (electrical) system, and certain accessories, such as a pressure transducer, amplifier, and external recorder.(b)
Classification. Class II (performance standards).