K Number
K234134

Validate with FDA (Live)

Date Cleared
2024-09-24

(270 days)

Product Code
Regulation Number
868.5905
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AirFit F30i Mask System
The AirFit F30i mask has two product variants:
· AirFit F30i mask is intended for single-patent reuse in the home environment.
• AirFit F30i SLM (Sleep Lab Mask) variant is intended for multi-patient reuse in the hospital/institutional environment. Both masks are intended for patients weighing more than 66lb (30kg), who have been prescribed non-invasive CPAP or bi-level positive airway pressure (PAP) therapy. The Sleep Lab Mask is the only variant that is validated and intended for multi-patient reprocessing and must be reprocesssed if reused between patients.

AirFit F30i NM Mask System
The AirFit F30i Non Magnetic mask has two product variants:
· The AirFit F30i Non Magnetic variant is intended for single-patient reuse in the home environment.
· The AirFit F30i Non Magnetic SLM (Sleep Lab Mask) variant is intended for multi-patient reuse in the hospital/institutional environment.
This mask is for patients weighing more than 66 lb (30 kg), who have been prescribed non-invasive CPAP or bi-level positive airway pressure (PAP) therapy. The Sleep Lab Mask is the only variant that is validated and intended for multipatient reprocessing and must be reprocessed if reused between patients.

Arcadia Mask System
The Arcadia has two product variants:
· Arcadia mask is intended for single-patient reuse in the home environment.
· Arcadia SLM (Sleep Lab Mask) variant is intended for multi-patient reuse in the hospital/institutional environment.
These masks are intended for patients weighing more than 66 lb (30kg), who have been prescribed non-invasive CPAP or bi-level positive airway pressure (PAP) therapy. The Sleep Lab Mask is the only variant that is validated and intended for multi-patient reprocessing and must be reprocesssed if reused between patients.

Device Description

The AirFit F30i, AirFit F30i NM and Arcadia mask systems are full-face conduit masks intended for use with Continuous Positive Airway Pressure (CPAP) and/or Bi-Level devices are used for the treatment of Obstructive Sleep Apnea (OSA) and/or non-life supporting ventilatory support. Both treatments deliver pressurized air from a flow generator to the patient. Such delivery of pressurized air relies on the Patient Interface; commonly known as the mask. Full face masks are a common type of Patient Interface (Mask) and are known to efficaciously support the delivery of pressurized air by providing a seal around the patient's nose and mouth.
The AirFit F30i, AirFit F30i NM and Arcadia mask systems each have two product variants:

  • . AirFit F30i Mask/ AirFit F30i NM Mask/ Arcadia Mask: This is the home use variant that is intended for single patient re-use.
  • . AirFit F30i SLM/ AirFit F30i NM SLM/ Arcadia SLM (sleep lab mask): This is intended for multipatient re-use and must be reprocessed if reused between patients using the disinfection guide
    The AirFit F30i, AirFit F30i NM and Arcadia mask systems are made up of the following four main component assemblies: Cushion, conduit frame, elbow and headgear. The cushion and conduit frame are available in various sizes to allow for adequate mask fit over the intended patient population.
    The AirFit F30i, AirFit F30i NM and Arcadia mask systems are prescription devices supplied non-sterile.
AI/ML Overview

The provided text does not contain detailed information about the acceptance criteria or a specific study that proves the device meets those criteria in the format requested. The document is a 510(k) summary and FDA clearance letter, which outlines the device's substantial equivalence to a predicate device based on various performance and safety tests. However, it does not provide the specific "acceptance criteria" (e.g., minimum sensitivity, specificity, or accuracy thresholds) or the detailed methodology and results of a single study with the requested metrics.

Instead, the document lists several non-clinical data submitted, which include performance and safety tests in accordance with ISO standards and other bench tests. The conclusion states that "Non-clinical verification and validation assessment/testing demonstrated that AirFit F30i NM and Arcadia mask systems met all intended performance requirements." and "Verification confirmed that the new device met the predetermined acceptance criteria and the performance is substantially equivalent to the previously cleared predicate AirFit F30i mask system (K183512)."

Therefore, I can extract the types of tests performed and the comparison to the predicate device, but I cannot fill in the table or answer all sub-questions as the specific numerical acceptance criteria and a single "study" proving it with the requested detail are not present.

Here's what can be extracted based on the provided text, with an explanation for missing information:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally qualitative (e.g., "met all intended performance requirements") and tied to ISO standards or established performance of the predicate device. Specific numerical thresholds for acceptance are not explicitly listed in the 510(k) summary, except for some comparative parameters.

Acceptance Criteria (General)Reported Device Performance (as compared to predicate)
CO2 rebreathing (<20% for normal, <60% for single fault) (ISO 17510:2015 Annex F)Predicate (AirFit F30i): Normal Condition: 8.1% (4 cm H2O), 5.8% (5 cm H2O), 1.9% (10 cm H2O). Single Fault: 32.6% (Fault 1), 45.3% (Fault 2). AirFit F30i Mask System (Modified): Normal Condition: 7.7%, 5.2%, 1.5%. Single Fault: 42.5% (Fault 1), 55.1% (Fault 2). AirFit F30i NM Mask System (Modified - not listed, assumed similar to AirFit F30i Mask System for these parameters unless specified otherwise): Normal Condition: 7.7%, 5.2%, 1.5%. Single Fault: 42.5% (Fault 1), 55.1% (Fault 2). Arcadia Mask System (New): Normal Condition: Arcadia Mask 9.9%, 8.5%, 6.3%; Arcadia SLM 9.7%, 8.4%, 6.2%. Single Fault: Arcadia Mask 42.8% (Fault 1), 54.8% (Fault 2); Arcadia SLM 41.9% (Fault 1), 54.3% (Fault 2).
Exhaust flow (ISO 17510:2015 Annex B)Predicate (AirFit F30i): 4 L/min @ 4 cm H2O, 31 L/min @ 9 cm H2O, 41 L/min @ 15 cm H2O, 49 L/min @ 20 cm H2O, 55 L/min @ 25 cm H2O. AirFit F30i Mask System (Modified): (Not explicitly listed, assumed similar to predicate or shown as equivalent). AirFit F30i NM Mask System (Modified): 22 L/min @ 4 cm H2O, 38 L/min @ 11 cm H2O, 48 L/min @ 17 cm H2O, 59 L/min @ 24 cm H2O, 67 L/min @ 30 cm H2O. Arcadia Mask System (New): 22 L/min @ 4 cm H2O, 39 L/min @ 11 cm H2O, 49 L/min @ 17 cm H2O, 60 L/min @ 24 cm H2O, 69 L/min @ 30 cm H2O.
Resistance to flow (Pressure drop across mask in cmH₂O) (ISO 17510:2015 Annex C)Predicate (AirFit F30i): 0.2 @ 50 L/min, 1.0 @ 100 L/min. AirFit F30i Mask System (Modified): 0.2 @ 50 L/min, 1.0 @ 100 L/min. AirFit F30i NM Mask System (Modified): (Not explicitly listed, assumed similar to AirFit F30i Mask System). Arcadia Mask System (New): 0.3 @ 50 L/min, 1.2 @ 100 L/min.
Anti-Asphyxia valve operating pressures (Open/Closed-to-atmosphere pressures <= 4 cm H2O) (ISO 17510:2015 Annex D)Predicate (AirFit F30i): <= 4 cm H2O. AirFit F30i Mask System (Modified): <= 4 cm H2O. AirFit F30i NM Mask System (Modified): < 4 cm H2O. Arcadia Mask System (New): (Not explicitly listed, stated as Equivalent to predicate).
Breathing resistance (Inspiratory/Expiratory) (ISO 17510:2015 Annex E)Predicate (AirFit F30i): Inspiratory 0.2, Expiratory 0.3. AirFit F30i Mask System (Modified): Inspiratory 0.2, Expiratory 0.4. AirFit F30i NM Mask System (Modified): (Not explicitly listed, assumed similar to AirFit F30i Mask System). Arcadia Mask System (New): Inspiratory 0.2, Expiratory (Not explicitly listed).
Vibration and noise (Sound power/pressure level) (ISO 17510:2015 Annex G)Predicate (AirFit F30i): Sound power level: 25 dBA, Sound pressure level: 18 dBA. AirFit F30i Mask System (Modified): Sound power level: 25 dBA, Sound pressure level: 18 dBA. AirFit F30i NM Mask System (Modified): Sound power level: 25 dBA, Sound pressure level: 18 dBA. Arcadia Mask System (New): Multi-hole vent, QuietAir vent, Sound power level: 22 dBA, Sound pressure level: 15 dBA.
Mechanical IntegrityDemonstrated after home cleaning, transportation and storage, operation environment, free fall and sit test, reprocessing. (Passed, details not provided).
BiocompatibilityEvaluated as per ISO 18562-1, ISO 10993-1 and applicable regulatory guidance. (Passed, details not provided).
Reprocessing claimsValidated in accordance with ISO 17664-2, ST98 and AAMI TIR12, including cleaning efficacy, disinfection efficacy, residual toxicity, and mechanical integrity. (Passed, details not provided).
Magnetic interferenceRisk assessment activities, including tests of magnetic field strength (<400mT on surface, <0.5mT at 50mm distance for devices with magnets), were conducted. (Passed, details not provided).

Detailed numerical acceptance criteria for each test (e.g., "CO2 rebreathing must be less than X%") are not explicitly provided, but the document confirms that the devices "met all intended performance requirements" and "predetermined acceptance criteria." The performance characteristics are compared to the predicate device, often indicating "Equivalent" or "Identical."

2. Sample size used for the test set and the data provenance

  • Sample Size: Not specified for any of the tests. The document refers to "non-clinical verification and validation assessment/testing."
  • Data Provenance: Not specified. These appear to be laboratory-based bench tests or in-house engineering tests. There is no indication of country of origin for data or if it's retrospective/prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • This information is not applicable as the tests described are non-clinical, bench-top performance and safety evaluations, not typically requiring expert-driven ground truth establishment in the way clinical studies with diagnostic devices might.

4. Adjudication method for the test set

  • This information is not applicable for the non-clinical, bench-top tests described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This information is not applicable. The device is a CPAP mask system, not an AI-assisted diagnostic or interpretive tool. Therefore, MRMC studies involving human readers and AI are not relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This information is not applicable. The device is a physical medical device (CPAP mask system), not a software algorithm or AI.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • For the non-clinical tests described, the "ground truth" is typically defined by established engineering standards (e.g., ISO 17510:2015 parameters, pressure measurements, flow rates, CO2 concentration readings) and reference methods. For biocompatibility and reprocessing, it's based on validated test methods and safety thresholds defined by relevant ISO and AAMI standards.

8. The sample size for the training set

  • This information is not applicable. The device is a physical medical device; there is no indication of a "training set" in the context of an algorithm or AI.

9. How the ground truth for the training set was established

  • This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 24, 2024

ResMed Pty Ltd % Sheila Bruschi Senior Director ReMed Corp 9001 Spectrum Center Boulevard San Diego, California 92123

Re: K234134

Trade/Device Name: AirFit F30i Mask System; AirFit F30i NM Mask System; Arcadia Mask System Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: Class II Product Code: BZD Dated: December 29, 2023 Received: August 21, 2024

Dear Sheila Bruschi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Rachana Visaria -S

Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K234134

Device Name

AirFit F30i Mask System, AirFit F30i NM Mask System, Arcadia Mask System

Indications for Use (Describe)

AirFit F30i Mask System

The AirFit F30i mask has two product variants:

· AirFit F30i mask is intended for single-patent reuse in the home environment.

• AirFit F30i SLM (Sleep Lab Mask) variant is intended for multi-patient reuse in the hospital/institutional environment. Both masks are intended for patients weighing more than 66lb (30kg), who have been prescribed non-invasive CPAP or bi-level positive airway pressure (PAP) therapy. The Sleep Lab Mask is the only variant that is validated and intended for multi-patient reprocessing and must be reprocesssed if reused between patients.

AirFit F30i NM Mask System

The AirFit F30i Non Magnetic mask has two product variants:

· The AirFit F30i Non Magnetic variant is intended for single-patient reuse in the home environment.

· The AirFit F30i Non Magnetic SLM (Sleep Lab Mask) variant is intended for multi-patient reuse in the hospital/institutional environment.

This mask is for patients weighing more than 66 lb (30 kg), who have been prescribed non-invasive CPAP or bi-level positive airway pressure (PAP) therapy. The Sleep Lab Mask is the only variant that is validated and intended for multipatient reprocessing and must be reprocessed if reused between patients.

Arcadia Mask System

The Arcadia has two product variants:

· Arcadia mask is intended for single-patient reuse in the home environment.

· Arcadia SLM (Sleep Lab Mask) variant is intended for multi-patient reuse in the hospital/institutional environment.

These masks are intended for patients weighing more than 66 lb (30kg), who have been prescribed non-invasive CPAP or bi-level positive airway pressure (PAP) therapy. The Sleep Lab Mask is the only variant that is validated and intended for multi-patient reprocessing and must be reprocesssed if reused between patients.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for ResMed. The logo consists of a curved line of dots that transition in color from blue to red. Below the line of dots is the word "ResMed" in a bold, sans-serif font.

510(k) Summary[As required by 21 CFR 807.92(c)]
Date Prepared:24 September 2024
Company Name/Owner:ResMed Pty Ltd1 Elizabeth Macarthur DriveBella Vista, NSW, 2153 Australia
Prepared and Submitted by:Ms. Shu-Ying HuangRegulatory Affairs ManagerShuying.huang@resmed.com.sg
Official Contact:Ms. Sheila BruschiSenior Director, Regulatory AffairsResMed Corp.9001 Spectrum Center BlvdSan Diego CA 92123 USATel +1 858 836 5934sheila.bruschi@resmed.com
Device Trade Name:AirFit F30i Mask System, AirFit F30i NM Mask System, Arcadia Mask System
Device Common Name:Vented Full Face Mask
Classification and Classification Name:21 CFR 868.5905, 73 BZD (Class II)Accessory to Noncontinuous Ventilator (IPPB)
Product Code:BZD
Predicate Device:Moore Park Mask (K183512)
Submission Reason:Changes being effectedNew device

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for ResMed. The logo features the company name "ResMed" in a bold, sans-serif font, positioned below a stylized graphic. The graphic consists of a series of connected circles that form a curved line, transitioning in color from blue on the left to red on the right, with shades of purple in the middle.

Device description:

The AirFit F30i, AirFit F30i NM and Arcadia mask systems are full-face conduit masks intended for use with Continuous Positive Airway Pressure (CPAP) and/or Bi-Level devices are used for the treatment of Obstructive Sleep Apnea (OSA) and/or non-life supporting ventilatory support. Both treatments deliver pressurized air from a flow generator to the patient. Such delivery of pressurized air relies on the Patient Interface; commonly known as the mask. Full face masks are a common type of Patient Interface (Mask) and are known to efficaciously support the delivery of pressurized air by providing a seal around the patient's nose and mouth.

The AirFit F30i, AirFit F30i NM and Arcadia mask systems each have two product variants:

  • . AirFit F30i Mask/ AirFit F30i NM Mask/ Arcadia Mask: This is the home use variant that is intended for single patient re-use.
  • . AirFit F30i SLM/ AirFit F30i NM SLM/ Arcadia SLM (sleep lab mask): This is intended for multipatient re-use and must be reprocessed if reused between patients using the disinfection guide

The AirFit F30i, AirFit F30i NM and Arcadia mask systems are made up of the following four main component assemblies: Cushion, conduit frame, elbow and headgear. The cushion and conduit frame are available in various sizes to allow for adequate mask fit over the intended patient population.

The AirFit F30i, AirFit F30i NM and Arcadia mask systems are prescription devices supplied non-sterile.

Compared to the predicate:

  • AirFit F30i mask system labeling was revised with magnet-related contraindications and warnings,
  • . AirFit F30i NM mask system has no magnetic clips to enable usage of the device by patients who are now excluded from the intended users of the predicate due to new magnet-related contraindications and warnings,
  • . Arcadia mask system is a new design with magnetic clips.

Indications for use:

AirFit F30i Mask System

The AirFit F30i mask has two product variants:

  • AirFit F30i mask is intended for single-patent reuse in the home environment.
    AirFit F30i SLM (Sleep Lab Mask) variant is intended for multi-patient reuse in the hospital/institutional environment. ● Both masks are intended for patients weighing more than 66lb (30kg), who have been prescribed non-invasive CPAP or bilevel positive airway pressure (PAP) therapy. The Sleep Lab Mask is the only variant that is validated and intended for multipatient reprocessing and must be reprocessed if reused between patients.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains the logo for ResMed. The logo features a curved line made up of a series of circles that transition in color from blue to red. Below the line of circles is the word "ResMed" in a bold, sans-serif font.

AirFit F30i NM Mask System

The AirFit F30i Non Magnetic mask has two product variants:

  • The AirFit F30i Non Magnetic variant is intended for single-patient reuse in the home environment. .
  • . The AirFit F30i Non Magnetic SLM (Sleep Lab Mask) variant is intended for multi-patient reuse in the hospital/institutional environment.

This mask is for patients weighing more than 66 lb (30 kg), who have been prescribed non-invasive CPAP or bi-level positive airway pressure (PAP) therapy. The Sleep Lab Mask is the only variant that is validated and intended for multi-patient reprocessing and must be reprocessed if reused between patients.

Arcadia Mask System

The Arcadia has two product variants:

  • Arcadia mask is intended for single-patient reuse in the home environment.
  • Arcadia SLM (Sleep Lab Mask) variant is intended for multi-patient reuse in the hospital/institutional environment.

These masks are intended for patients weighing more than 66 lb (30kg), who have been prescribed non-invasive CPAP or bi-level positive airway pressure (PAP) therapy. The Sleep Lab Mask is the only variant that is validated and intended for multi-patient reprocessing and must be reprocessed if reused between patients.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for ResMed. The logo consists of a curved line of colored dots that transition from blue to red. Below the dots, the word "ResMed" is written in a bold, sans-serif font.

Comparison table

Predicate deviceSubject device
Design parameter orfeatureK183512K234134
Moore Park Mask:(Commercial name AirFit F30i)Modified device:AirFit F30i MaskSystemModified device:AirFit F30i NM Mask SystemNew device:Arcadia Mask SystemComments
Indications for UseThe Moore Park mask is intendedto be used by patients weighingmore than 66 lb (30 kg) whohave been prescribed non-invasive positive airway pressure(PAP) therapy such as CPAP or bi-level therapy. The mask isintended for single patient re-usein the home and multi-patient re-use in the hospital/institutionalenvironment.The AirFit F30i mask hastwo product variants:AirFit F30i mask isintended for single-patent reuse in thehome environment. AirFit F30i SLM(Sleep Lab Mask)variant is intendedfor multi-patientreuse in thehospital/institutionalenvironment. Both masks are intendedfor patients weighingmore than 66lb (30kg),who have beenprescribed non-invasiveCPAP or bi-level positiveairway pressure (PAP)therapy. The Sleep LabMask is the only variantthat is validated andintended for multi-patient reprocessing andThe AirFit F30i Non Magnetic mask hastwo product variants:The AirFit F30i Non Magneticvariant is intended for single-patient reuse in the homeenvironment. The AirFit F30i Non Magnetic SLM(Sleep Lab Mask) variant isintended for multi-patient reuse inthe hospital/institutionalenvironment. This mask is for patients weighing morethan 66 lb (30 kg), who have beenprescribed non-invasive CPAP or bi-levelpositive airway pressure (PAP) therapy.The Sleep Lab Mask is the only variantthat is validated and intended for multi-patient reprocessing and must bereprocessed if reused between patients.The Arcadia has two productvariants:Arcadia mask is intended forsingle-patient reuse in thehome environment. Arcadia SLM (Sleep LabMask) variant is intended formulti-patient reuse in thehospital/institutionalenvironment. These masks are intended forpatients weighing more than 66 lb(30kg), who have been prescribednon-invasive CPAP or bi-levelpositive airway pressure (PAP)therapy. The Sleep Lab Mask is theonly variant that is validated andintended for multi-patientreprocessing and must bereprocessed if reused betweenpatients.Equivalent

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the ResMed logo. The logo features a curved line made of connected circles that transition in color from blue to red. Below the line is the word "ResMed" in a bold, sans-serif font.

Design parameter orfeaturePredicate deviceK183512Subject deviceK234134Comments
Moore Park Mask:(Commercial name AirFit F30i)Modified device:AirFit F30i MaskSystemModified device:AirFit F30i NM Mask SystemNew device:Arcadia Mask System
must be reprocesssed ifreused between patients.
Intended UseThe mask is intended toprovide an interface for CPAPor bi-level devices.The mask is intended to provide an interface for CPAP or bi-level devices.Identical
FDA Product CodeBZDBZDIdentical
Patient populationPatients weighing more than66lb (30kg) for whom positiveairway pressure therapy hasbeen prescribed.Patients weighing more than 66lb (30kg) for whom positive airway pressure therapy has beenprescribed.Identical
Environment of UseHome orhospital/institutionalenvironment.Home or hospital/institutional environment.Identical
Reprocessing claimsSingle patient re-use or multi-patient re-use.Single patient re-use or multi-patient re-use.Identical
Sterility state as providedNon-sterileNon-sterileIdentical
Validated reprocessingmethodsHigh-Level ThermaldisinfectionHigh-Level Thermal disinfectionIdentical
Cushion typeOral nasal cradle cushion thatseals under the nose andaround the mouthOral nasal cradle cushion that seals under the nose and aroundthe mouthOral pillows cushions that sealsunder the nose and aroundthe mouthEquivalent
Frame typeTubular conduitTubular conduitIdentical

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image contains the ResMed logo. The logo features a curved line made of connected circles that transition in color from blue to red. Below the line is the word "ResMed" in a bold, sans-serif font.

Design parameter or featurePredicate deviceK183512Subject deviceK234134New device:Arcadia Mask SystemComments
Moore Park Mask:(Commercial name AirFit F30i)Modified device:AirFit F30i Mask SystemModified device:AirFit F30i NM Mask System
PAP tubing connection point22mm conical connector as per ISO 5356-1 over the head connection22mm conical connector as per ISO 5356-1 over the head connectionIdentical
Exhaust port locationElbow and cushionElbow and cushionIdentical
SizesCushion available in four sizesFrame available in three sizesHeadgear available in one sizeCushion available in four sizesFrame available in three sizesHeadgear available in one sizePillows cushion available in three sizesFrame available in three sizesHeadgear available in one sizeEquivalent
Mask exhaust flow(Nominal)Pressure (cm H2O) Flow (L/min) 'Full Face' curve 4 20 9 31 15 41 20 49 25 55Pressure (cm H2O) Flow (L/min) 'Full Face' curve 4 22 11 38 17 48 24 59 30 67Pressure (cm H2O) Flow (L/min) 'Full Face' curve 4 22 11 39 17 49 24 60 30 69ISO 17510:2015 Annex B
CO2 re-breathing performance – Normal Condition (<20%)Pressure (cm H2O) Relative CO2 increase 4 8.1% 5 5.8% 10 1.9%Relative CO2 increase 7.7% 5.2% 1.5%Relative CO2 increase Arcadia Mask Arcadia SLM 9.9% 9.7% 8.5% 8.4% 6.3% 6.2%ISO 17510:2015 Annex F

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the ResMed logo. The logo consists of a series of colored dots arranged in a curved line, transitioning from blue to red. Below the dots, the word "ResMed" is written in a bold, sans-serif font.

Design parameter or featurePredicate deviceK183512Subject deviceK234134Comments
Moore Park Mask:(Commercial name AirFit F30i)Modified device:AirFit F30i Mask SystemModified device:AirFit F30i NM Mask SystemNew device:Arcadia Mask System
CO2 re-breathingperformance- Single FaultCondition(<60%)FaultRelative CO2 increaseRelative CO2 increaseRelative CO2 increase
Arcadia MaskArcadia SLM
Fault 132.6%42.5%42.8%41.9%ISO17510:2015Annex F
Fault 245.3%55.1%54.8%54.3%
Resistance to flow(Pressure drop acrossmask in cmH₂O)@50 L/min @100 L/min0.2 1.0@50 L/min @100 L/min0.2 1.0@50 L/min @100 L/min0.3 1.2ISO17510:2015Annex C
BreathingResistance(cmH₂O)Inspiratory0.20.20.2ISO17510:2015Annex E
Expiratory0.30.4
AAVOperatingPressuresOpen-to-atmospherepressure≤ 4 cm H2O< 4 cm H2OEquivalentISO17510:2015Annex D
Closed-to-atmospherepressure≤ 4 cm H2O< 4 cm H2OISO17510:2015Annex D
Flow generator setting oncompatible ResMed CPAP"Full face" (Air 10 devices)"Full face" (Air 10 devices)Identical
Design parameter orfeaturePredicate deviceK183512Subject deviceK234134Comments
Moore Park Mask:(Commercial name AirFit F30i)Modified device:AirFit F30i MaskSystemModified device:AirFit F30i NM Mask SystemNew device:Arcadia Mask System
and Bi-level flowgenerators.
Operating pressure range(cmH2O)4 - 254 - 304 - 304 - 30Equivalent
SoundSound power level: 25 dBASound pressure level: 18 dBASound power level: 25 dBASound pressure level: 18 dBASound power level: 25 dBASound pressure level: 18 dBAMulti-hole ventQuietAir ventSound power level: 22 dBASound pressure level: 15 dBAEquivalent
MagneticClipsNumber ofmagnets4404Equivalent
Strength(mT) onthesurface<400mT<400mTN/A<400mT
Strength(mT) at 50mmdistance<0.5 mT<0.5 mTN/A<0.5 mT
Operating and storagetemperatureOperating temperature: 5°Cto 40°CStorage temperature: -20°C to+60°COperating temperature: 5°C to 40°CStorage temperature: -20°C to +60°CIdentical

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image contains the ResMed logo. The logo features a curved line of dots that transition in color from blue to red. Below the line of dots is the word "ResMed" in a bold, sans-serif font.

{12}------------------------------------------------

Image /page/12/Picture/1 description: The image shows the ResMed logo. The logo consists of the word "ResMed" in a bold, sans-serif font, with the "R" slightly larger than the other letters. Above the word "ResMed" is a curved line made up of a series of colored dots that transition from blue to purple to red. The dots are arranged in a wave-like pattern, creating a visual representation of airflow or breathing.

Non-clinical data submitted:

Non-clinical verification and validation assessment/testing demonstrated that AirFit F30i NM and Arcadia mask systems met all intended performance requirements. These included:

Applicable performance and safety tests in accordance with ISO 17510:2015 Medical devices -- Sleep apnoea breathing therapy -- Masks and application accessories:

  • CO2 rebreathing (ISO 17510:2015 Annex F) ●
  • . Exhaust flow (ISO 17510:2015 Annex B)
  • Resistance to flow (ISO 17510:2015 Annex C)
  • Anti-Asphyxia valve operating pressures (ISO 17510:2015 Annex D)
  • Breathing resistance (ISO 17510:2015 Annex E) ●
  • . Vibration and noise (ISO 17510:2015 Annex G)

Other bench tests:

  • Pressure accuracy and pressure swing performance
  • Mechanical Integrity of the mask system before and after the following environmental tests:
    • Home cleaning
    • l Transportation and Storage
    • 트 Operation environment
    • Free fall and sit test
    • Reprocessing

Biocompatibility evaluation was assessed and/or conducted in accordance with ISO 18562-1, ISO 10993-1 and applicable regulatory guidance.

Validation of reprocessing claims (in accordance with ISO 17664-2, ST98 and AAMI TIR12) included a combination of cleaning efficacy, disinfection efficacy, residual toxicity and mechanical integrity assessment/testing.

Verification confirmed that the new device met the predetermined acceptance criteria and the performance is substantially equivalent to the previously cleared predicate AirFit F30i mask system (K183512).

Risk assessment activities, including tests of magnetic field strength, were conducted to assess the risk of magnetic interference from static magnetic fields for ResMed masks containing magnets with medical implants/devices and metallic implanted objects.

The AirFit F30i, AirFit F30i NM and Arcadia mask systems were assessed and/or tested in accordance with the applicable requirements in relevant FDA consensus standards including:

StandardsTitle
ISO 17510:2015Medical devices -- Sleep apnoea breathing therapy -- Masks and application accessories
ISO 18562-1:2017Biocompatibility evaluation of breathing gas pathways in healthcare applications -- Part 1: Evaluationand testing within a risk management process
ISO 18562-2:2017Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 2: Tests foremissions of particulate matter

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for ResMed. The logo consists of the word "ResMed" in a bold, sans-serif font, with a stylized wave-like design above it. The wave is made up of a series of circles that transition in color from blue to red. The logo is simple, modern, and easily recognizable.

ISO 18562-3:2017Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 3: Tests foremissions of volatile organic compounds (VOCs)
ISO 18562-4:2017Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 4: Tests forleachables in condensate
ISO 10993-1:2018Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk managementprocess
ISO 10993-3:2014Biological evaluation of medical devices -- Part 3: Tests for genotoxicity, carcinogenicity and reproductivetoxicity
ISO 10993-5:2009Biological evaluation of medical devices - Part 5: Test for in vitro cytotoxicity
ISO 10993-10:2021Biological evaluation of medical devices - Part 10: Tests for skin sensitization
ISO 10993-11:2017Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity
ISO10993-12:2021Biological evaluation of medical devices -- Part 12: Sample preparation and reference materials
ISO 10993-17:2002Biological evaluation of medical devices - Part 17: Establishment of allowable limits for leachablesubstances
ISO 17664-1:2021Processing of health care products - Information to be provided by the medical device manufacturer forthe processing of medical devices - Part 1: Critical and semi-critical medical devices
ISO17664-2:2021Processing of health care products - Information to be provided by the medical device manufacturer forthe processing of medical devices - Part 2: Non-critical medical devices.
ISO 5356-1:2015Anaesthetic and respiratory equipment -- Conical connectors -- Part 1: Cones and sockets
ISO 15223-1:2021Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1:General requirements
ISO 20417:2021Medical devices — Information to be supplied by the manufacturer
ST98:2022Cleaning validation of health care products - Requirements for development and validation of a cleaningprocess for medical devices
AAMI TIR12:2020Designing, Testing, And Labeling Medical Devices Intended For Processing By Health Care Facilities: AGuide For Device Manufacturers

Substantial Equivalence Conclusion:

The AirFit F30i, AirFit F30i NM and Arcadia mask systems are substantially equivalent to the predicate K183512:

  • They have equivalent intended use ●
  • . They have similar technological characteristics
  • They have similar performance characteristics
  • . The differences do not raise any new questions of safety or effectiveness
  • They are as safe and as effective as the predicate device .

§ 868.5905 Noncontinuous ventilator (IPPB).

(a)
Identification. A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.(b)
Classification. Class II (performance standards).