K Number
K234042

Validate with FDA (Live)

Date Cleared
2024-06-07

(169 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EFAI BONESUITE XR BONE AGE PRO ASSESSMENT SYSTEM (EFAI BAPXR) is designed to view and quantify bone age from 2D Posterior Anterior (PA) view of left-hand radiographs using deep learning techniques to aid in the analysis of bone age assessment of patients between 2 to 16 years old for pediatric radiologists. The results should not be relied upon alone by pediatric radiologists to make diagnostic decisions. The images shall be with left hand and wrist fully visible within the field of view, and shall be without any major bone destruction, deformity, fracture, excessive motion, or other major artifacts.

Device Description

The device is a software designed to aid the quantification of bone age for patients between 2 to 16 years old. The software uses deep learning techniques to analyze posterior-anterior (PA) radiographs of the left-hand according to the Greulich-Pyle (GP) method.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device's performance, based on the provided text:

EFAI Bonesuite XR Bone Age Pro Assessment System (BAP-XR-100) Performance Study

1. A table of acceptance criteria and the reported device performance

The acceptance criteria for this device are based on the intercept and slope of a Deming regression analysis between the device's output (EFAI BAPXR) and the Ground Truth (GT). The criteria are that both the intercept and slope of the regression line must fall within the range of the highest acceptable bias. The text does not explicitly state the numerical "highest acceptable bias" range, but it states that the observed results met these general criteria.

MetricAcceptance Criteria (General)Reported Device Performance (EFAI BAPXR vs. GT)
Deming Regression InterceptFall within the range of the highest acceptable bias-0.07 (95% CI: [-0.13, -0.01])
Deming Regression SlopeFall within the range of the highest acceptable bias1.00 (95% CI: [0.99, 1.00])
Percentage of cases with bone age difference < 0.5 years (EFAI BAPXR vs. GT)(Not explicitly stated as an primary acceptance criterion, but reported as an indicator of satisfactory results)> 88%
Bland-Altman 95% Limits of Agreement (EFAI BAPXR vs. GT)(Not explicitly stated as an primary acceptance criterion, but reported as an indicator of high consistency)-0.517 to 0.743 (with CIs in gray dashed lines)

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Test Set (Clinical Study): 600 cases
  • Data Provenance: Retrospectively collected from 27 locations across multiple states and multiple clinical organizations in the United States.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

  • Number of Experts: Four (4)
  • Qualifications of Experts: U.S. board-certified expert radiologists. Specific experience level (e.g., years) is not mentioned.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

The ground truth for the test set was established through a "Ground Truthing Workflow" involving multiple stages:

  • Bone Age Assessment: Individual assessments by the four expert radiologists.
  • Consensus Via Grading: Implies a process of evaluating and potentially assigning grades to assessments based on predetermined criteria (e.g., differences).
  • Majority Voting: Most likely used when assessments differed, to reach an initial consensus.
  • Final Adjudication: This step suggests a process where discrepancies or remaining disagreements after majority voting were resolved by a final decision-making body or method. The flowchart indicates a systematic process to ensure consistency and consensus, though the exact rules for "Final Adjudication" (e.g., if a lead adjudicator made a final decision or if all 4 radiologists had to agree) are not explicitly detailed beyond "consensus among all readers reviewing the radiographs."

This detailed workflow suggests a robust, multi-reader consensus approach for ground truthing, rather than a simple 'none' or majority vote without further review.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader multi-case (MRMC) comparative effectiveness study (human readers with AI vs. without AI assistance) was not explicitly described. The clinical study was a standalone performance study of the EFAI BAPXR device itself, comparing its output to ground truth established by expert radiologists, not measuring human reader improvement with AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, a standalone performance study was done. The description states: "EFAI conducted a standalone performance study with the proposed device EFAI BAPXR..." This study measured the performance of the EFAI BAPXR algorithm directly against the established ground truth.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth used was expert consensus based on assessments by four U.S. board-certified expert radiologists, following a structured "Ground Truthing Workflow" that included individual assessments, consensus via grading, majority voting, and final adjudication, comparing their findings to the Greulich-Pyle Atlas.

8. The sample size for the training set

The training set comprised 23,578 cases.

9. How the ground truth for the training set was established

For the training set, the ground truth was established as the average of the bone age assessments independently done by three board-certified radiologists.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the agency's name, "U.S. Food & Drug Administration," in blue text.

Ever Fortune.AI Co., Ltd. Ti-Hao Wang Chief Technology Officer Rm. D, 8F. No. 573, Sec. 2 Taiwan Blvd. West Dist. Taichung City, 403020 Taiwan

June 7, 2024

Re: K234042

Trade/Device Name: EFAI Bonesuite XR Bone Age Pro Assessment System (BAP-XR-100) Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH Dated: May 24, 2024 Received: May 24, 2024

Dear Ti-Hao Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jessica Lamb

Jessica Lamb Assistant Director Imaging Software Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K234042

Device Name

EFAI BONESUITE XR BONE AGE PRO ASSESSMENT SYSTEM (BAP-XR-100)

Indications for Use (Describe)

EFAI BONESUITE XR BONE AGE PRO ASSESSMENT SYSTEM (EFAI BAPXR) is designed to view and quantify bone age from 2D Posterior Anterior (PA) view of left-hand radiographs using deep learning techniques to aid in the analysis of bone age assessment of patients between 2 to 16 years old for pediatric radiologists. The results should not be relied upon alone by pediatric radiologists to make diagnostic decisions. The images shall be with left hand and wrist fully visible within the field of view, and shall be without any major bone destruction, deformity, fracture, excessive motion, or other major artifacts.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Ever Fortune AI. The logo consists of a teal-colored icon that resembles a person with a globe as their head, followed by the text "EVER FORTUNE.AI" in a matching teal color. The globe icon is made up of interconnected dots, suggesting a network or connection.

510(k) Summary

General Information 1.

510(k) SponsorEver Fortune.AI Co., Ltd.
AddressRm. D, 8F. No. 573, Sec. 2 Taiwan Blvd.West Dist.Taichung City 403020TAIWAN
ApplicantJoseph Chang
Contact Information886-04-23213838 #216joseph.chang@everfortune.ai
Correspondence PersonTi-Hao Wang
Contact Information886-04-23213838 #168thothwang@gmail.comtihao.wang@everfortune.ai
Date PreparedOctober, 2023

2. Proposed Device

Proprietary NameEFAI Bonesuite XR Bone Age Pro Assessment System (BAP-XR-100
Common NameEFAI BAPXR
Classification NameSystem, Image Processing, Radiological
Regulation Number21 CFR 892.2050
Regulation NameMedical Image Management and Processing System
Product CodeQIH
Regulatory ClassII

3. Predicate Device

Proprietary NameMedo Aria
Premarket NotificationK200356

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains a logo for a company called "EVER FORTUNE.AI". The logo consists of two parts: a stylized human figure on the left and the company name on the right. The human figure is teal and has a green circle with a network of white dots for a head. The company name is written in teal, with "EVER" on top and "FORTUNE.AI" below.

Classification NameSystem, Image Processing, Radiological
Regulation Number21 CFR 892.2050
Regulation NameMedical Image Management and Processing System
Product CodeQIH
Regulatory ClassII

Device Description 4.

The device is a software designed to aid the quantification of bone age for patients between 2 to 16 years old. The software uses deep learning techniques to analyze posterior-anterior (PA) radiographs of the left-hand according to the Greulich-Pyle (GP) method.

Intended User Population

The intended users of EFAI BAPXR are pediatric radiologists.

When a clinician accesses the patient radiographs in a picture archiving and communication system (PACS) client or a workstation, both the EFAI BAPXR quantification and the original unaltered radiographs are available in the same patient study, allowing for concurrent reading.

EFAI BAPXR is an adjunct tool and is not intended to replace a clinician's review of the radiograph or his or her clinical judgment. Clinicians must not use the software generated output as the primary interpretation.

Intended Patient Population

The target population of the device are patients between 2 and 16 years old that require a bone age assessment.

Description of output

The algorithm outputs a standard JSON message, to ensure consistent interpretation and simple integration across systems. The result contains structured information in regards to the estimated bone age in years and the data can be transferred and processed by RIS/HIS/PACS and other workstations.

5. Intended Use

EFAI BONESUITE XR BONE AGE PRO ASSESSMENT SYSTEM (EFAI BAPXR) is designed to view and quantify bone age from 2D Posterior (PA) view of left-hand radiographs using deep learning techniques to aid in the analysis of bone age assessment of

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains a logo for a company called "EVER FORTUNE.AI". The logo consists of a stylized human figure in teal, with a green circle containing a network of white dots as its head. To the right of the figure, the company name is written in teal, with "FORTUNE.AI" appearing below "EVER" in a slightly smaller font size. The overall design is clean and modern, suggesting a focus on technology and innovation.

patients between 2 to 16 years old for pediatric radiologists. The results should not be relied upon alone by pediatric radiologists to make diagnostic decisions. The images shall be with left hand and wrist fully visible within the field of view, and shall be without any major bone destruction, deformity, fracture, excessive motion, or other major artifacts.

6. Comparison of Technological Characteristics with Predicate Device

Feature/FunctionProposed Device:EFAI BAPXR(K234042)Predicate Device:Medo Aria(K200356)
IntendedUse/Indication forUseEFAI BONESUITE XR BONE AGEASSESSMENT SYSTEM PRO(EFAI BAPXR) is designed to viewand quantify bone age from 2DPosterior Anterior (PA) view ofleft-hand radiographs using deeplearning techniques to aid in theanalysis of bone age assessment ofpatients between 2 to 16 years old forpediatric radiologists. The resultsshould not be relied upon alone bypediatric radiologists to makediagnostic decisions. The imagesshall be with left hand and wrist fullyvisible within the field of view, andshall be without any major bonedestruction, deformity, fracture,excessive motion, or other majorartifacts.MEDO ARIA is designed to viewand quantify ultrasound image datausing machine learning techniquesto aid trained medical professionalsin diagnosis of developmentaldysplasia of the hip (DDH). Thedevice is intended to be used onneonates and infants, aged 0 to 12months.
Environment of UseHealthcare facility/HospitalHealthcare facility/Hospital
Intended userPediatric radiologistRadiologist
Clinical conditionBone age assessmentdevelopmental dysplasia of the hip(DDH)
Image InputComplies with DICOM standardComplies with DICOM standard
Scan TypeX-ray2D and 3D Ultrasound
Body PartLeft hand and wristHip
Image display modeStaticStatic
Artificial IntelligenceAlgorithmYesYes
Image navigation andmanipulationtoolsNoAdjust image brightness andcontrast, slice-scroll, panelayout, reset
2D image reviewNoYes, capable of reviewing allframes of multi-frame(multi-slice) image
Manual landmarkplacementNoYes
Semi-automaticlandmarkplacementNoYes, user-modifiable
Quantitative analysisBone age assessment (years)● Angle (alpha angle)● Distance ratio (coverage)
Report creationNoYes

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Ever Fortune AI. The logo consists of a stylized human figure in teal, with a green circle containing a network of white dots above its head. To the right of the figure, the word "EVER" is written in large, teal letters. Below "EVER", the words "FORTUNE.AI" are written in smaller, teal letters, with a green circle containing a network of white dots replacing the "O" in "FORTUNE".

Performance Data 7.

Performance of the EFAI BAPXR has been evaluated and verified in accordance with software specifications and applicable performance standards through software verification and validation testing. Additionally, the software validation activities were performed in accordance with IEC 62304:2006/A1:2016 - Medical device software - Software life cycle processes, in addition to the FDA Guidance documents, "Content of Premarket Submissions for Device Software Functions" and "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions."

To establish the performance of EFAI BAPXR, the performance was validated by nonclinical and clinical tests.

Nonclinical Tests

During the process of model development, a total of 26,222 cases were retrospectively collected from Taiwan and the United States (U.S.). These cases were subsequently divided into training and testing datasets, consisting of 23.578 and 2.644 cases, respectively.

An internal validation test was conducted to assess the performance of the EFAI BAPXR. Ground-truthing was the average of the bone age assessment independently done by three board-certified radiologists.

Overall, the Deming regression analysis result showed a good correlation between GT and EFAI BAPXR's output. The intercept of the regression line was -0.06 with a 95% CI of [-0.09, -0.03]

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for Ever Fortune AI. The logo consists of a stylized human figure with a green circle containing a network of white dots above its head. To the right of the figure, the words "EVER" and "FORTUNE.AI" are written in a sans-serif font, with the word "EVER" stacked above "FORTUNE.AI". The color scheme is primarily teal and green.

and the slope was 1.00 with a 95% CI of [1.00. 1.01]. By confirming that the Cls of both measures fall within the highest acceptable bias.

Clinical Tests

EFAI conducted a standalone performance study with the proposed device EFAI BAPXR with a pre-determined primary endpoint and performance goal to evaluate the software's performance on left-hand X-ray images in PA view. The test data comprised 600 cases aged 2-16 years, acquired from 27 locations across multiple states and multiple clinical organizations in the United States. The study population contained 50.0% males and 50.0% females, with a mean chronological age of cases was 9.2 years old and 120 cases in each three-vear age interval.The race or ethnicities of cases include White, Hispanic, Black, Asian, Multiple, etc. The X-Ray scanner manufacturers of images include Samsung Electronics, Carestream Health, Kodak, GE Healthcare, Siemens, Konica Minolta, etc. None of the cases was used as part of the EFAI BAPXR model training or development.

The study design measured the performance of EFAI BAPXR against the ground truth (GT) from four U.S. board-certified expert radiologists. As shown in the following figure A) for the ground truthing workflow, the ground truthing was generated through the truthing process based on the current standard of care, with the addition of multiple checkpoints to ensure consistency and consensus among all readers reviewing the radiographs when comparing them to the Greulich-Pyle Atlas.

Image /page/7/Figure/5 description: The image shows a flowchart describing a ground truthing workflow. The workflow is divided into four phases: Bone Age Assessment, Consensus Via Grading, Majority Voting, and Final Adjudication. Each phase involves a decision-making process based on criteria such as differences in bone age assessments, grades, and votes. The flowchart outlines the steps to determine the ground truth (GT) based on the readers' assessments.

Figure A. Ground Truthing Workflow

EFAI BAPXR Traditional 510(k)

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains a logo for a company called "EVER FORTUNE.AI". The logo consists of a stylized human figure with a green globe-like head, connected by lines and dots, and a teal body. To the right of the figure, the word "EVER" is written in large, teal letters, with "FORTUNE.AI" written in smaller letters below it, also in teal.

The primary endpoint was the intercept and slope of a Deming regression between GT and EFAI BAPXR's output. The acceptance criteria were met if both the intercept and slope of the regression line had fallen within the range of the highest acceptable bias.

The observed results of the standalone performance validation study demonstrated that EFAI BAPXR can provide bone age assessments with satisfactory results including over 88% of the cases with a difference less than 0.5 years between GT and EFAI BAPXR's output. The intercept of the Deming regression result was -0.07 with a 95% Confidence Interval (95% CI) of [-0.13, -0.01] and slope of 1.00 with a 95% CI of [0.99, 1.00]. Overall, the device performance met the performance goal compared with the GT, thus reaching the primary endpoint. In addition, as shown in the following Bland-Altman plot (Figure B), the 95% limits of agreement (red solid lines) for bone age assessment between GT and EFAI BAPXR's output ranged from -0.517 to 0.743 (with corresponding CIs in gray dashed lines), indicating high consistency.

Image /page/8/Figure/3 description: This image is a Bland-Altman plot of the EFAI BAPXR. The x-axis represents the average of GT and EFAI BAPXR's output in years, ranging from 0 to 20. The y-axis represents the difference between GT and EFAI BAPXR's output in years, ranging from -4 to 4. The plot shows a scatter of points around the zero line, with two dashed lines indicating the limits of agreement.

As shown in the table below, we also evaluated the performance of EFAI BAPXR in various clinical subgroups to see if these factors systematically affect the device performance, including gender, age, race or ethnicity, clinical organization, X-ray manufacturer, input image with radiologic findings (including epiphyseal, finger, carpal/ metacarpal, radius/ ulna, others, and multiple), and image quality issues (including poor image quality, poor position, foreign body, artifact, and multiple).

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Ever Fortune AI. The logo consists of a stylized human figure with a green globe as its head, and the text "EVER FORTUNE.AI" in a teal color. The globe has a network pattern on it, suggesting a connection or global reach.

Intercept (95% CI)Slope (95% CI)
Gender
Female0.01 (-0.07, 0.09)0.99 (0.98, 1.00)
Male-0.14 (-0.23, -0.04)1.00 (0.99, 1.01)
Age
Pre-puberty0.03 (-0.07, 0.13)0.98 (0.96, 0.99)
Early & Mid-puberty-0.06 (-0.25, 0.13)0.99 (0.98, 1.01)
Late-puberty-1.10 (-1.46, -0.74)1.07 (1.04, 1.09)
Race or Ethnicity
White-0.05 (-0.14, 0.04)0.99 (0.99, 1.00)
Hispanic0.05 (-0.12, 0.22)0.98 (0.97, 1.00)
Black or African-0.17 (-0.55, 0.20)1.01 (0.98, 1.04)
Asian-0.12 (-0.28, 0.04)0.99 (0.98, 1.01)
Multiple-0.19 (-0.36, -0.01)1.00 (0.98, 1.02)
Other Races or Ethnicities-0.07 (-0.31, 0.17)0.99 (0.97, 1.01)
Clinical Organization
Org A0.12 (-0.13, 0.37)0.97 (0.94, 1.01)
Org B-0.24 (-0.69, 0.21)1.01 (0.97, 1.05)
Org C-0.11 (-0.24, 0.02)0.99 (0.98, 1.00)
Org D-0.07 (-0.15, 0.01)1.00 (0.99, 1.01)
X-ray Manufacturer
Samsung Electronics-0.17 (-0.33, -0.01)1.00 (0.99, 1.01)
Carestream Health-0.03 (-0.16, 0.10)0.99 (0.98, 1.00)
Kodak-0.17 (-0.28, -0.07)1.01 (1.00, 1.02)
GE Healthcare0.01 (-0.23, 0.25)0.99 (0.96, 1.01)
Siemens0.05 (-0.10, 0.20)1.00 (0.99, 1.02)
Konica Minolta0.12 (-0.10, 0.35)0.97 (0.94, 1.00)
Other Manufacturers0.05 (-0.29, 0.39)0.99 (0.96, 1.02)
Cases with Radiologic Findings-0.03 (-0.14, 0.08)0.99 (0.98, 1.01)
Cases with Image Quality Findings-0.04 (-0.19, 0.11)1.00 (0.98, 1.01)

Table. Subgrouping Analysis Results of EFAI BAPXR

Pre-puberty: female ages [2 - 7) & male ages [2 - 9).

Early & Mid-puberty: female ages [7 - 13) & male ages [9 - 14).

Late-puberty: female ages [13 - 16] & male ages [14 - 16].

Other races or ethnicities include American Indian/Alaska Native, Pacific Islander/Native Hawaiian, others and unreported.

Other manufacturers include Fujifilm Corporation, Canon Inc., and Hologic, Inc..

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Ever Fortune AI. The logo consists of a teal-colored icon on the left, resembling a person with a globe as their head. To the right of the icon, the words "EVER" and "FORTUNE.AI" are written in teal. The logo appears to be for a company that specializes in artificial intelligence.

Overall, we found the device performs consistently and reliably under these circumstances, this concludes the performance demonstrated that the EFAI BAPXR device is as safe and effective as the predicate device Medo Aria.

Safety & Effectiveness 8.

EFAI BAPXR has been designed, verified and validated in compliance with 21 CFR, Part 820.30 requirements. The device has been designed to meet the requirements associated with ISO 14971:2019 Medical devices - Application of risk management to medical devices. The EFAI BAPXR performance has been validated using retrospective data from case data and through the use of Reader comparison analysis.

Substantial Equivalence 9.

The EFAI BAPXR has minor differences compared to the predicate device. These differences are related to the Intended Use and to the Quantitative Analysis features.

The Medo Aria predicate is intended to quantify ultrasound images to support diagnosis of development dysplasia of the hip (DDH) based on Graf. The EFAI BAPXR is intended to quantify bone age based on an image atlas. Although the intended uses are different, the principle of operation is the same as both devices use artificial intelligence algorithms to view and quantify radiograph images and provide results based on an established reference or standard. Both devices have similar risks and are mitigated by similar special controls. Both devices rely on labeling and clinical performance testing as mitigations to address risks related to the use of the devices. Therefore, although the intended uses are not exactly the same, the risks and technology used by both devices are the same and are mitigated similarly and thus the proposed device does not raise new questions of safety or efficacy as no new risks are introduced.

The predicate device consists of a user interface in which end users are able to interact with images using typical image navigation and manipulation tools, including landmark placement. The proposed device does not consist of a typical user interface which allows users to interact with images using typical image navigation and manipulation tools, including landmark placement. However, this difference does not raise new questions of safety or efficacy and does not add new or significant risk to the device as the differences are workflow related and the end users are able to still access the original image to interpret the image according to standard practices.

10. Conclusion

Based on the information submitted in this premarket notification, and based on the indications for use, technological characteristics, and performance testing, the EFAI BAPXR raises no new questions of safety and effectiveness and is substantially equivalent to the predicate device in terms of safety, efficacy, and performance.

N/A