K Number
K233986

Validate with FDA (Live)

Date Cleared
2024-03-15

(88 days)

Product Code
Regulation Number
866.1645
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, and Streptococcus, This premarket notification is for the BD Phoenix™ Automated Microbiology System with Ciprofloxacin at a concentration of 0.0156-4 ug/mL. Ciprofloxacin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.

Device Description

This submission is for a range extension of a single antimicrobial cleared for use on BD Phoenix ID/AST or AST only panels. The ID portion of the ID/AST combination panel was not subject to review in this submission.

The Phoenix AST method is a broth-based microdilution test. The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. The ID/AST combination panel includes an ID side (51 wells) with dried substrates for bacterial identification and an AST side (85 wells). The AST panel contains a wide range of two-fold doubling dilution concentrations of antimicrobial agents and growth and fluorescent controls at appropriate well locations. The AST panel does not include wells for isolate identification.

The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. The organism to be tested must be a pure culture and be preliminarily identified as Gram-positive or Gram-negative. Colonies are then suspended in ID broth and equated to a 0.5 McFarland suspension using a nephelometer device. A further dilution is made into AST broth (a cation-adjusted formulation of Mueller-Hinton broth containing 0.010% Tween 80), to which the redox-buffered oxidation-reduction AST indicator solution is added producing a blue color in the wells. The concentration of organisms in the final AST broth suspension is approximately 5 X 105 CFU/mL.

The Phoenix AST Broth is poured into the inoculation port of the AST panel and the inoculum flows into the panel, filling panel wells. Polyethylene caps are applied to seal the inoculation ports. An air admittance port is located in the panel lid to ensure adequate oxygen tension in the panel for the duration of the test. Inoculated panels are barcode scanned and loaded into the BD Phoenix Automated Microbiology System instrument where panels are continuously incubated at 35 ℃ ± 1 ℃.

Continuous measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. The instrument takes readings every 20 minutes. Organisms growing in the presence of a given antimicrobial agent reduce the indicator (changing it to a pink color). This signals organism growth and resistance to that antimicrobial agent. Organisms killed or inhibited by the antimicrobial agent do not cause reduction of the indicator and therefore do not produce a color change. The Phoenix instrument reads and records the results of the antimicrobial tests contained in the panel and interprets the reactions (based on the organism identification) to give a minimal inhibitory concentration (MIC) value and category interpretations (susceptible, intermediate, resistant, or not susceptible). AST results are available within 16 hours. This is an auto read result; no manual readings are possible with this system.

Additional comments concerning specific organism/antimicrobial combinations are provided from the software-driven expert system (BDXpert), using rules derived from CLSI documentation and/or the FDA-approved drug labeling.

AI/ML Overview

The BD Phoenix Automated Microbiology System - GN Ciprofloxacin (0.0156-4 ug/mL) is an antimicrobial susceptibility testing system. The provided text describes the performance characteristics of this device, particularly for Salmonella species, and the study conducted to demonstrate its performance against acceptance criteria.

1. Table of Acceptance Criteria and Reported Device Performance:

Performance MetricAcceptance Criteria (General for AST devices based on FDA guidance)Reported Device Performance (Salmonella Species)
Essential Agreement (EA)Overall EA > 90% (when compared to reference method within ±1 serial two-fold dilution)Combined Clinical and Challenge: 98.4% (127/129)
Category Agreement (CA)Overall CA > 90% (when reference method agrees exactly with device interpretation)Combined Clinical and Challenge: 94.6% (122/129)
Minor Discrepancies (Min)Acceptable level generally specified in guidance; often tied to reaching target EA/CA7 (Combined Clinical and Challenge)
Major Discrepancies (Maj)0% or very low percentage (often < 3%)0 (Combined Clinical and Challenge)
Very Major Discrepancies (Vmj)0% or very low percentage (often < 1.5%)0 (Combined Clinical and Challenge)
Reproducibility> 95% agreement (±1 dilution)Manual Inoculation: 100% (378/378)
Phoenix AP Inoculation: 99.7% (377/378)

(Note: The document states "The BD Phoenix Ciprofloxacin performance met the acceptance criteria for Salmonella with overall EA and CA greater than 90%." The specific numerical acceptance criteria for Min, Maj, Vmj are usually found in the referenced guidance documents but are not explicitly detailed as numerical percentages in the provided text for the device itself, other than the observed performance.)

2. Sample Size Used for the Test Set and Data Provenance:

  • Clinical Isolates: 47 (fresh: 7 (14.9%), stock: 40 (85.1%))
    • Organisms: Salmonella species (43 isolates), Salmonella enterica ssp. enterica serovar Typhi (4).
    • Provenance: "Clinical testing was conducted at three U.S. sites." (Suggests prospective and retrospective clinical data from U.S. sources).
  • Challenge Isolates: 82 stock isolates
    • Organisms: Salmonella species (79), Salmonella enterica ssp. enterica serovar Typhi (3).
    • Provenance: "Additional stock challenge isolates were tested at each study site." (Implying laboratory-controlled challenge strains, likely geographically diverse or from culture collections, but no specific country of origin is mentioned beyond "U.S. sites" for the clinical part which also conducted challenge testing).
  • Reproducibility Study Isolates: 14 isolates of non-fastidious Gram-negative organisms (including Pseudomonas aeruginosa (1), Salmonella enterica ssp. enterica serovar Paratyphi A (1), and Salmonella species (11)).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

The ground truth was established using the CLSI frozen broth microdilution reference panel, prepared according to CLSI M07 guidelines. This is a standardized laboratory reference method, implicitly implying experts in microbiology and antimicrobial susceptibility testing were involved in its execution and interpretation, rather than individual "experts" reviewing each case for a ground truth panel. The document does not specify the number or qualifications of "experts" as individuals but relies on the standardized, recognized reference method.

4. Adjudication Method for the Test Set:

Not applicable in the typical sense of expert adjudication for diagnostic imaging or similar devices. The ground truth is determined by a universally accepted reference laboratory method (CLSI frozen broth microdilution). The comparison involves measuring agreement between the device and this reference method.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. The study focuses on the standalone performance of the automated system against a reference method, not on human reader performance with or without AI assistance.

6. Standalone (Algorithm Only) Performance:

Yes, a standalone performance study was conducted. The "BD Phoenix Automated Microbiology System" is an automated device, and its performance was evaluated against a reference method (CLSI frozen broth microdilution) without human intervention in the MIC determination or category interpretation on the device side. The reported Essential Agreement (EA) and Category Agreement (CA) metrics represent this standalone algorithmic performance.

7. Type of Ground Truth Used:

The ground truth used was expert consensus methodology/reference standard, specifically the CLSI frozen broth microdilution reference panel prepared according to CLSI M07 guidelines. This is the gold standard for antimicrobial susceptibility testing.

8. Sample Size for the Training Set:

The document does not explicitly state the sample size for a training set. This type of device (AST system) is typically developed and validated against extensive datasets during its initial creation and subsequent range extensions. However, the provided text describes the validation/test set used for this specific premarket notification (47 clinical isolates and 82 challenge isolates for Salmonella). This suggests a re-evaluation of performance based on updated breakpoints rather than a new algorithm development requiring a separate training set description in this 510(k) summary.

9. How the Ground Truth for the Training Set Was Established:

As no explicit training set sample size is provided, the method for establishing its ground truth is also not detailed. However, if a training phase was involved in the initial development of the BD Phoenix system, it would have traditionally relied on similar reference methodologies like CLSI broth microdilution to establish accurate MIC values and corresponding susceptibility categories.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 15, 2024

Becton, Dickinson and Company Kamisha Gray Senior Regulatory Affairs Specialist 7 Loveton Circle Sparks, Maryland 21152

Re: K233986

Trade/Device Name: BD Phoenix Automated Microbiology System - GN Ciprofloxacin (0.0156-4 ug/mL) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System Regulatory Class: Class II Product Code: LON Dated: December 15, 2023 Received: December 18, 2023

Dear Kamisha Gray:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an

{1}------------------------------------------------

established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ribhi Shawar -S

Ribhi Shawar, Ph.D. (ABMM) Branch Chief General Bacteriology and Antimicrobial Susceptibility Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K233986

Device Name

BD Phoenix™ Automated Microbiology System - GN Ciprofloxacin (0.0156-4 ug/mL)

Indications for Use (Describe)

Indications for Use:

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, and Streptococcus, This premarket notification is for the BD Phoenix™ Automated Microbiology System with Ciprofloxacin at a concentration of 0.0156-4 ug/mL. Ciprofloxacin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.

Active In Vitro and in Clinical Infections Against:

Citrobacter koseriCitrobacter freundii
Enterobacter cloacaeEscherichia coli
Klebsiella pneumoniaeMorganella morganii
Proteus mirabilisProteus vulgaris,
Providencia rettgeriProvidencia stuartii
Pseudomonas aeruginosaSalmonella typhi
Serratia marscescensShigella boydii
Shigella dysenteriaeShigella flexneri
Shigella sonnei

Active In Vitro but clinical significance is unknown

Edwardsiella tardaKlebsiella aerogenes (formerly Enterobacter aerogenes)
Klebsiella oxytocaSalmonella enteritidis

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

Summary Preparation Date: March 15, 2024

Background Information: I

A 510(k) Number

K233986

B Applicant

BD Diagnostic Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, Maryland 21152 Establishment Registration Number: 1119779 Contact: Kamisha Gray Telephone: 410-316-4000

C Proprietary and Established Names

BD Phoenix™ Automated Microbiology System - GN Ciprofloxacin (0.0156-4 µg/mL)

D Regulatory Information

ProductCode(s)ClassificationRegulationSectionPanel
LONClass II21 CFR 866.1645 - Fully Automated Short-Term Incubation Cycle AntimicrobialSusceptibility SystemMI - Microbiology

II Submission/Device Overview:

A Purpose for Submission:

Range extension of Ciprofloxacin to the BD Phoenix Gram negative ID/AST and AST only Phoenix panels for Salmonella species to accommodate the updated FDA-recognized breakpoints for Salmonella as published in the FDA STIC website.

Update the BD Phoenix Gram negative ID/AST and AST only Phoenix panels (0.25 - 4 ug/mL) for the current reporting range to include updated FDA-recognized breakpoints for Enterobacterales and Pseudomonas aeruginosa, as published in the FDA STIC website.

Establish a Pre-Determined Change Control Plan (PCCP) to address future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage.

{5}------------------------------------------------

B Measurand:

Ciprofloxacin (0.0156-4 ug/mL)

C Type of Test:

Antimicrobial Susceptibility Test (Quantitative) colorimetric, oxidation-reduction, growth based.

III Intended use/Indications for Use:

A Intended Use(s):

The BD Phoenix Automated Microbiology System is intended for the in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of Gram Negative aerobic and facultative anaerobic bacteria belonging to the family Enterobacterales and non-Enterobacterales.

B Indication(s) for Use:

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus. and Streptococcus, and Streptococcus.

This premarket notification is for the BD Phoenix™ Automated Microbiology System with Ciprofloxacin at a concentration of 0.0156-4 ug/mL. Ciprofloxacin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDAapproved package insert for this antimicrobial agent.

Active In Vitro and in Clinical Infections Against:

Citrobacter koseri Citrobacter freundii Enterobacter cloacae Escherichia coli Klebsiella pneumoniae Morganella morganii Proteus mirabilis Proteus vulgaris Providencia rettgeri Providencia stuartii Pseudomonas aeruginosa Salmonella typhi Serratia marscescens Shigella boydii

{6}------------------------------------------------

Shigella dysenteriae Shigella flexneri Shigella sonnei

Active In Vitro but clinical significance is unknown:

Edwardsiella tarda Klebsiella aerogenes (formerly Enterobacter aerogenes) Klebsiella oxytoca Salmonella enteritidis

C Special Conditions for Use Statements(s):

Rx - For Prescription Use Only

D Special Instrument Requirements:

BD Phoenix™ Automated Microbiology System and software (V2.20.0.0 or higher) PhoenixSpec™ Nephelometer BD Phoenix™ AP Instrument

IV Device/System Characteristics:

A Device Description:

This submission is for a range extension of a single antimicrobial cleared for use on BD Phoenix ID/AST or AST only panels. The ID portion of the ID/AST combination panel was not subject to review in this submission.

The Phoenix AST method is a broth-based microdilution test. The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. The ID/AST combination panel includes an ID side (51 wells) with dried substrates for bacterial identification and an AST side (85 wells). The AST panel contains a wide range of two-fold doubling dilution concentrations of antimicrobial agents and growth and fluorescent controls at appropriate well locations. The AST panel does not include wells for isolate identification.

The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. The organism to be tested must be a pure culture and be preliminarily identified as Gram-positive or Gram-negative. Colonies are then suspended in ID broth and equated to a 0.5 McFarland suspension using a nephelometer device. A further dilution is made into AST broth (a cation-adjusted formulation of Mueller-Hinton broth containing 0.010% Tween 80), to which the redox-buffered oxidation-reduction AST indicator solution is added producing a blue color in the wells. The concentration of organisms in the final AST broth suspension is approximately 5 X 105 CFU/mL.

The Phoenix AST Broth is poured into the inoculation port of the AST panel and the inoculum flows into the panel, filling panel wells. Polyethylene caps are applied to seal the

CONFIDENTIAL AND PROPRIETARY

{7}------------------------------------------------

inoculation ports. An air admittance port is located in the panel lid to ensure adequate oxygen tension in the panel for the duration of the test. Inoculated panels are barcode scanned and loaded into the BD Phoenix Automated Microbiology System instrument where panels are continuously incubated at 35 ℃ ± 1 ℃.

Continuous measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. The instrument takes readings every 20 minutes. Organisms growing in the presence of a given antimicrobial agent reduce the indicator (changing it to a pink color). This signals organism growth and resistance to that antimicrobial agent. Organisms killed or inhibited by the antimicrobial agent do not cause reduction of the indicator and therefore do not produce a color change. The Phoenix instrument reads and records the results of the antimicrobial tests contained in the panel and interprets the reactions (based on the organism identification) to give a minimal inhibitory concentration (MIC) value and category interpretations (susceptible, intermediate, resistant, or not susceptible). AST results are available within 16 hours. This is an auto read result; no manual readings are possible with this system.

Additional comments concerning specific organism/antimicrobial combinations are provided from the software-driven expert system (BDXpert), using rules derived from CLSI documentation and/or the FDA-approved drug labeling.

B Principle of Operation:

The BD Phoenix Automated Microbiology System is a broth-based microdilution method that utilizes a redox indicator (colorimetric oxidation-reduction) to enhance detection of organism growth. The MIC is determined by comparing growth in wells containing serial two-fold dilutions of an antibiotic to the growth in "growth control wells" that contain no antibiotic.

V Substantial Equivalence Information:

A Predicate Device Names(s):

BD Phoenix™ Automated Microbiology System - Moxifloxacin-0.125-8 and Ciprofloxacin-0.25-4 µg/mL

B Predicate 510(k) Numbers(s):

K060217

C Comparison with Predicate(s):

{8}------------------------------------------------

Device & PredicateDevice(s):DeviceK233986PredicateK060217
Ciprofloxacin Extended Range(0.0156-4 μg/mL)Ciprofloxacin (0.25-4 μg/mL)
Device Trade NameBD Phoenix™ Automated Microbiology System – GN Ciprofloxacin (0.0156-4 μg/mL)BD Phoenix™ AutomatedMicrobiology System – GNCiprofloxacin (0.25-4 μg/mL)
General Device Characteristic Similarities
Antimicrobial AgentCiprofloxacinSame
IntendedUse/Indications forUseDetermination of in vitro antimicrobialsusceptibility testing of aerobic and facultativeanaerobic Gram- negative, Gram-positive, andStreptococcus bacteria.Same
Source ofMicroorganisms forTestingBacterial colonies isolated from culture.Same
TechnologyAutomated growth-based detectionSame
MethodologyDetermination of MIC using serial two-folddilution formatSame
Read MethodAutomatedSame
Inoculation MethodsManual: BD PhoenixSpec nephelometerAutomated: BD Phoenix AP InstrumentSame
Result ReportedReport results as minimum inhibitoryconcentration (MIC) and categoricalinterpretation (S, I, R)Same
Incubation Time< 16 hoursSame
General Device Characteristic Differences
Indicated OrganismsActive In Vitro and in Clinical InfectionsAgainst:Citrobacter koseriCitrobacter freundiiEnterobacter cloacaeEscherichia coliKlebsiella pneumoniaeMorganella morganiiProteus mirabilisProteus vulgaris ,Providencia rettgeriProvidencia stuartiiPseudomonas aeruginosaSalmonella typhiSerratia marscescensShigella boydiiShigella dysenteriaeShigella flexneriShigella sonneiGram negative aerobic andfacultative anaerobic bacteriabelonging to Enterobacterales andnon-Enterobacterales
Device & PredicateDeviceK233986PredicateK060217
Device(s):Ciprofloxacin Extended Range(0.0156-4 µg/mL)Ciprofloxacin (0.25-4 µg/mL)
Active In Vitro but clinical significance isunknown:
Edwardsiella tardaKlebsiella aerogenes (formerly Enterobacter
aerogenes)Klebsiella oxytocaSalmonella enteritidis
BreakpointsEnterobacterales:(S/I/R) ≤0.25 / 0.5 / ≥1Pseudomonas aeruginosa:(S/I/R) ≤0.5 / 1 / ≥2Salmonella species:(S/I/R) ≤0.0625 / 0.125-0.5 / ≥1Enterobacterales:(S/I/R) ≤1 / 2 / ≥4Pseudomonas aeruginosa:(S/I/R) ≤1 / 2 / ≥4Salmonella species:(S/I/R) ≤1 / 2 / ≥4
Reporting RangeSalmonella spp.0.0156-4 µg/mLEnterobacterales and P. aeruginosa0.25 - 4 µg/mL0.25-4 µg/mL
Breakpoint ChangeEvaluation ProcedureProcedure addedNone

Table 1. Comparison with the Predicate

{9}------------------------------------------------

Standards/Guidance Documents Referenced VI

  • Guidance for Industry and FDA, Class II Special Controls Guidance Document: 1. Antimicrobial Susceptibility Test (AST) Systems, August 28, 2009.
    1. CLSI. Performance Standards for Antimicrobial Susceptibility Testing. 33rd ed. CLSI supplement M100. Clinical and Laboratory Standards Institute; 2023.
    1. CLSI. Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically. 11th ed. CLSI supplement M07. Clinical Laboratory Standards Institute; 2018.

VII Performance Characteristics (if/when applicable):

A Analytical Performance

    1. Precision/Reproducibility:
      Reproducibility was conducted at three clinical sites using 14 isolates of non-fastidious Gram-negative organisms. The isolates were tested at each site in triplicate over three different days using both inoculation methods (i.e., manual, BD Phoenix AP) resulting in 378 data points (14 strains x 3 replicates x 3 sites x 3 days = 378). The isolates tested in the reproducibility study included Pseudomonas aeruginosa (1). Salmonella enterica sso. enterica serovar Paratyphi A (1), and Salmonella species (11). The reproducibility was calculated based on MIC values falling within ±1 dilution of the predetermined mode of

{10}------------------------------------------------

the reference MIC values. There were no "off-scale" MIC results with manually prepared inocula and one "off-scale" result with inocula prepared using the BD Phoenix AP. The best- and worst-case reproducibility was calculated as described in the AST Special Controls Guidance document. The results of the study demonstrate that for this antimicrobial agent and the Gram-negative organisms tested, there was an overall reproducibility across test sites of greater than 95% (± 1 dilution) agreement when compared to the test mode. The reproducibility results for each inoculation method are shown in Table 2.

Note: The testing for the BD Phoenix AP instrument was performed at three internal BD sites.

Table 2. Summary of Reproducibility Studies - BD Phoenix Ciprofloxacin Inoculation Method Best Case Worst Case 100% (378/378)

  • Manual BD PhoenixSpec Nephelometer 100% (378/378) 99.7% (377/378) 99.7% (377/378) BD Phoenix AP Instrument
    1. Linearity:

Not applicable

    1. Analytical Specificity/Interference:
      Not applicable
    1. Assay Reportable Range:
      Not applicable
    1. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

Quality Control Testing:

The CLSI recommended QC organisms (E. coli ATCC 25922 and P. aeruginosa ATCC 27853) were tested a sufficient number of times (i.e., at least 20/site) at each of three testing sites. It was tested using both manual and Phoenix AP inoculation methods and read by the BD Phoenix instrument. The results are summarized in Table 3. Results were acceptable for greater than 95% of tests performed using both inoculation methods.

Note: The lower end range extension only applies to the Salmonella species. Therefore, the BD Phoenix instrument will not report an MIC below 0.25 ug/mL for E. coli or Pseudomonas aeruginosa. The expected range for Escherichia coli remains ≤0.25 µg/mL and Pseudomonas aeruginosa remains <0.25-1 ug/mL.

{11}------------------------------------------------

OrganismConcentration(µg/mL)ReferenceBD Phoenix
Manual Inoculation(PhoenixSpec)Phoenix APInoculation
≤ 0.0156128
0.03134
Escherichia coliATCC® 259220.06251
0.125
≤0.2518982
Expected Range:0.004-0.016 µg/mL0.5
1
2
≤ 0.0156
0.0313
PseudomonasaeruginosaATCC® 278530.0625
0.125
0.25180
Expected Range:0.125-1 µg/mL0.5538780
112
2
Table 3. Quality Control Results - Ciprofloxacin
--------------------------------------------------

1 The expected range for Escherichia coli ATCC 25922 remains ≤0.25 µg/mL and Pseudomonas aeruginosa ATCC 27853 remains ≤0.25-1 µg/mL for the BD Phoenix System since the lower range extension only applies to Salmonella.

Inoculum Density Check:

The BD PhoenixSpec Nephelometer was used to prepare the inocula for testing of the clinical, challenge, reproducibility, and QC isolates. The same inoculum suspension was used for both the Phoenix System and the reference method testing. The BD Phoenix AP instrument was used to standardize the inocula for challenge, QC, and reproducibility isolates. Validation data for both the PhoenixSpec and the Phoenix AP instrument was provided and found to be acceptable.

Growth Failure Rate:

The growth rate for both inoculation methods was 100%.

Purity Check:

Purity check plates were performed on all isolates from each inoculum preparation.

    1. Detection Limit:
      Not applicable
    1. Assay Cut-Off:
      Not applicable

{12}------------------------------------------------

B Comparison Studies:

1. Method Comparison with Predicate Device:

Results obtained with the BD Phoenix Automated Microbiology System - GN Ciprofloxacin (0.0156 - 4 µg/mL) panel were compared to results obtained with the CLSI frozen broth microdilution reference panel. Reference panels were prepared according to CLSI M07 guidelines. The range of dilutions evaluated with the reference panels was the same as that used for the BD Ciprofloxacin panel. The BD Phoenix Spec Nephelometer, the primary inoculation method, was used to obtain a 0.50 - 0.60 McFarland for all challenge, clinical, QC, and reproducibility isolates. The BD Phoenix AP instrument, the secondary inoculation method, was used to test challenge, OC, and reproducibility isolates. It is designed to standardize the ID broth inoculum equivalent to the BD Phoenix Spec Nephelometer, add the preset amount of AST indicator broth to the AST broth tube, and transfer the required aliquot of ID broth inoculum to AST broth tubes.

Clinical:

Clinical testing was conducted at three U.S. sites using 7 fresh (14.9%) and 40 stock (85.1%) Salmonella organisms for a total of 47 clinical isolates. These consisted of Salmonella species (43 isolates), and Salmonella enterica ssp. enterica serovar Typhi (4).

Challenge:

Additional stock challenge isolates were tested at each study site. These isolates consisted of organisms with known resistance mechanisms to challenge the ability of AST system to correctly identify the susceptibility category. Challenge testing was conducted using 82 Salmonella organisms including Salmonella species (79), and Salmonella enterica ssp. enterica serovar Typhi (3).

Results for clinical and challenge isolates were evaluated separately and combined. Table 4 below illustrates the performance of testing ciprofloxacin using the manual inoculation method only.

{13}------------------------------------------------

Table 4. Combined (Clinical and Challenge) Performance Summary of BD Phoenix with Clinical and Challenge Salmonella Isolates – Manual Inoculation Method

CiprofloxacinEA TotalEA N%EA TotalEval EA TotEval EA N%EA EvalCA TotalCA N%CA#RMinMajVmj
Salmonella ≤0.0625 (Susceptible), 0.125-0.5 (Intermediate), ≥1 (Resistant)
Clinical474595.7343294.1474710012000
Challenge82821005555100827591.53700
Combined12912798.4898797.812912294.615700

EA - Essential Agreement

Maj - major discrepancies Vmj - very major discrepancies

CA - Category Agreement R - resistant isolates

Min - minor discrepancies

Essential Agreenent (E.A) occurs when the result of the reference method and that of BD Phoenix within plus or ninus one serial two-fold dilution of the antibiotic. Evalus are those that are on scale for both the reference nethod or those in witch an off scale result is at least two doubling dilutions from the on scale result. Category Agreement (C.A) occurs when the result of the result of the reference method agrees exactly with the interpretation of the BD Phoenix result.

{14}------------------------------------------------

The BD Phoenix Ciprofloxacin performance met the acceptance criteria for Salmonella with overall EA and CA greater than 90%. There were 7 minor errors and no major errors or very major errors. Performance was calculated as described in the AST guidance, with the exception of evaluable EA. While the AST guidance indicates that all results with a reference value of "less than" or "greater than" a certain dilution are not considered evaluable, evaluable EA was calculated in line with the FDA 's current thinking where evaluable results for EA can be defined as any result that either (i) has on-scale MIC values for both the reference and test, OR (ii) has an off-scale value for the reference and/or test that is at least two doubling dilutions from the other result.

Inoculum Preparation Methods:

The challenge organisms were also tested by one clinical site using suspensions prepared by the BD Phoenix AP instrument. The comparison between the manual (PhoenixSpec) preparation method and the BD Phoenix AP preparation method is shown in Table 5. The overall % EA and % CA consistently met the acceptance criteria of greater than or equal to 90%. There were no very major or major discrepancies with either inoculation method.

{15}------------------------------------------------

Table 5. Comparison of Inoculation Methods with Challenge Isolates Only

CiprofloxacinTotalEA N%EA TotalEval EA TotEval EA N%EA EvalCA TotalCA N%CA#RMinMajVmj
Salmonella ≤0.0625 (Susceptible), 0.125-0.5 (Intermediate), ≥1 (Resistant)
Manual(PhoenixSpec)82821005555100827591.53700
Phoenix AP82821004040100827895.11400
EA - Essential Agreement Maj – major discrepancies

EA - Essential Agreement CA - Category Agreement

Vmj - very major discrepancies

R - resistant isolates

Min – minor discrepancies

{16}------------------------------------------------

2. Matrix Comparison:

Not applicable

C Clinical Studies:

    1. Clinical Sensitivity:
      Not applicable
    1. Clinical Specificity:
      Not applicable
    1. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
      Not applicable

D Clinical Cut-Off:

Not applicable

E Expected Values/Reference Range:

The FDA-recognized susceptibility interpretive criteria for ciprofloxacin are as listed in Table 6.

Table 6. FDA-Recognized Interpretive Criteria for Ciprofloxacin*
Minimum Inhibitory Concentrations (µg/mL)
PathogenSIR
Enterobacterales≤0.250.5≥1
Salmonella spp.≤0.060.12-0.5≥1
Pseudomonas aeruginosa≤0.51≥2

According to the FDA STIC Website

S = Susceptible; I = Intermediate; R = Resistant

Proposed Labeling: VIII

The labeling supports the finding of substantial equivalence for this device when evaluated with the current FDA-recognized ciprofloxacin breakpoints.

{17}------------------------------------------------

IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission included a predetermined change control plan (PCCP) that was reviewed and accepted by FDA, as described in the Antimicrobial Susceptibility Test (AST) System Devices - Updating Breakpoints in Device Labeling guidance. This PCCP addresses future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage (https://www.fda.gov/drugs/development-resources/antibacterial-susceptibility-test-interpretivecriteria). The PCCP outlined the specific procedures and acceptance criteria that BD intends to use to evaluate the cleared BD Phoenix antimicrobial Ciprofloxacin when revised breakpoints are published on the FDA STIC webpage. The PCCP included with the submission indicated that if specific criteria are met, BD will update the device label to include (1) the new breakpoints, (2) an updated performance section after re-evaluation of data in this premarket notification with the new breakpoints, and (3) any new limitations as determined by their evaluation.

§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.

(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”