K Number
K233979

Validate with FDA (Live)

Date Cleared
2024-07-19

(214 days)

Product Code
Regulation Number
862.1675
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VACUETTE® QUICKSHIELD Complete is intended to be used only with VACUETTE® Blood Collection Tubes as a system in routine venipuncture procedures. This device is designed with a safety shield, which can be activated to cover the needle immediately following venipuncture to aid in the protection against accidental needlestick injury. They are for single-use only and should be used by properly trained healthcare professionals only in accordance with these instructions.

The VACUETTE® QUICKSHIELD Complete PLUS is intended to be used only with VACUETTE® Blood Collection Tubes as a system in routine venipuncture procedures. This device is designed with a safety shield, which can be activated to cover the needle immediately following venipuncture to aid in the protection against accidental needlestick injury. VACUETTE® VISIO PLUS Needles are designed for use in the daily blood collection routine when delegated by a qualified practitioner. The flashback window is situated in the transparent part of the cannula hub, which assists the user to recognize successful vein penetration. They are for single- use only and should only be used by adequately trained healthcare personnel in accordance with these instructions.

Device Description

The sterile VACUETTE®QUICKSHIELD Complete is a single packed VACUETTE®QUICKSHIELD Safety Tube Holder with pre-threaded VACUETTE® Multiple Use Drawing Needle in a blister pack. It is a sterile, single-use plastic tube holder, designed with a safety shield, which can be activated to cover the needle immediately following venipuncture.

The sterile VACUETTE® QUICKSHIELD Complete PLUS is a single packed VACUETTE® QUICKSHIELD Safety Tube Holder with pre-threaded VACUETTE® VISIO PLUS Blood Collection Needle in a blister pack. It is a sterile, single-use plastic tube holder, designed with a safety shield, which can be activated to cover the needle immediately following venipuncture.

AI/ML Overview

This document is an FDA 510(k) Premarket Notification summary for a medical device, not a study describing AI algorithm performance. Therefore, most of the requested information regarding acceptance criteria for AI algorithms, sample size, ground truth expert establishment, MRMC studies, and training set details are not applicable.

However, I can extract the acceptance criteria and performance data for this physical medical device based on the provided text.

Here's the information that can be extracted:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied by the "Requirement" and "Outcome" columns in the "PERFORMANCE DATA" section. The document states that performance testing was conducted to demonstrate substantial equivalency and that the device performs as intended. All listed requirements passed.

RequirementOutcome (Pass or fail)
Force to remove shieldPass
Torque to rotate shieldPass
Torque to screw out needlePass
Spin outPass
Force to activate safety mechanismPass
Force necessary to bend safety shield with needle by 90°Pass
Puncture forcePass
Friction forcePass
LeakagePass

Additionally, the following were met:

  • Sterility: Achieved a Sterility Assurance Level (SAL) of 10⁻⁶ via e-beam irradiation, in accordance with ISO 11137-1 and ISO 11737-1.
  • Biocompatibility: Demonstrated compliance with ISO 10993 series standards and FDA guidance for:
    • Cytotoxicity (ISO 10993-5)
    • Sensitization (ISO 10993-10)
    • Irritation (ISO 10993-23)
    • Hemocompatibility (ISO 10993-4)
    • Material mediated pyrogenicity (ISO 10993-11)
    • Acute Systemic Toxicity (ISO 10993-11)

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample sizes used for each specific test (e.g., how many devices were tested for "Force to remove shield"). It refers to "Simulated use tests" and "Performance testing" in general.

Data provenance is not specified in terms of country of origin for the testing, and the testing described is prospective for demonstrating compliance with standards for the device being submitted.

3. Number of Experts Used to Establish Ground Truth and Qualifications

This information is not applicable. The document describes the testing of a physical medical device (blood collection system) against engineering and biological performance standards, not an AI algorithm requiring expert ground truth for classification or detection.

4. Adjudication Method for the Test Set

This information is not applicable as the device performance is measured against objective engineering and biological standards, not requiring human adjudication of subjective data.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

This information is not applicable. MRMC studies are typically performed for imaging diagnostics or AI-assisted diagnostic tools to assess the impact on human reader performance. This document is for a physical blood collection device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

This information is not applicable. The device is a physical product, not an algorithm. The performance tests ("Torque to rotate shield," "Puncture force," etc.) are inherent to the device's design and function in a standalone manner.

7. The Type of Ground Truth Used

The "ground truth" for this device's performance is established by recognized consensus standards (e.g., ISO 23908, ISO 7864, ISO 9626, ISO 11137-1, ISO 11737-1, ISO 10993 series) and internal specifications. This means that the device was tested to meet predefined physical, mechanical, and biological properties, not a "ground truth" derived from expert consensus, pathology, or outcomes data in the context of an AI study.

8. The Sample Size for the Training Set

This information is not applicable. This is not an AI algorithm; there is no training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

July 19, 2024

Greiner Bio-One North America Inc Manfred Abel Regulatory Affairs & Quality Systems Manager 4238 Capital Drive Monroe, North Carolina 28110

Re: K233979

Trade/Device Name: VACUETTE® QUICKSHIELD Complete (Plus) Regulation Number: 21 CFR 862.1675 Regulation Name: Blood specimen collection device Regulatory Class: Class II Product Code: JKA, FMI Dated: June 20, 2024 Received: June 20, 2024

Dear Manfred Abel:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

David Walloschek

David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors

{2}------------------------------------------------

OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K233979

Device Name VACUETTE® QUICKSHIELD Complete (Plus)

Indications for Use (Describe)

The VACUETTE® QUICKSHIELD Complete is intended to be used only with VACUETTE® Blood Collection Tubes as a system in routine venipuncture procedures. This device is designed with a safety shield, which can be activated to cover the needle immediately following venipuncture to aid in the protection against accidental needlestick injury. They are for single-use only and should be used by properly trained healthcare professionals only in accordance with these instructions.

The VACUETTE® QUICKSHIELD Complete PLUS is intended to be used only with VACUETTE® Blood Collection Tubes as a system in routine venipuncture procedures. This device is designed with a safety shield, which can be activated to cover the needle immediately following venipuncture to aid in the protection against accidental needlestick injury. VACUETTE® VISIO PLUS Needles are designed for use in the daily blood collection routine when delegated by a qualified practitioner. The flashback window is situated in the transparent part of the cannula hub, which assists the user to recognize successful vein penetration. They are for single- use only and should only be used by adequately trained healthcare personnel in accordance with these instructions.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Greiner Bio-One. The logo features a stylized blue wave-like shape above the word "greiner" in a bold, sans-serif font. Below "greiner" is the text "BIO-ONE" in a smaller, sans-serif font.

K233979 - Premarket Notification 510(K) Summary

1. SUBMITTER

Applicant name:Greiner Bio-One GmbH.Bad Haller Strasse 324550 Kremsmünster,Austria
Contact person:Manfred AbelGreiner Bio-One NA Inc.4238 Capital DriveMonroe, NC 28110Phone +1 704 261 7800manfred.abel@gbo.com
Establishment registration number:8020040
Date prepared:July 19, 2024
2. DEVICE
Trade name:VACUETTE® QUICKSHIELD Complete (Plus)
Common name:Safety tube holder pre-assembled with the needle
Regulation:862.1675, Blood Specimen Collection Device
Primary classification product code:JKA – Blood specimen collection device
Subsequent classification product code:FMI – Needle, hypodermic, single lumen
Medical device class:II
Review panel:General Hospital
3. PREDICATE DEVICES
Primary predicate device:Blood Collection Needle with/without Holder,Safety Blood Collection Needle with/without Holder,Luer Access Device-holder with Pre-attached Multiple SampleAdapter (K200027)

4. DEVICE DESCRIPTION

Secondary predicate device:

The sterile VACUETTE®QUICKSHIELD Complete is a single packed VACUETTE®QUICKSHIELD Safety Tube Holder with pre-threaded VACUETTE® Multiple Use Drawing Needle in a blister pack. It is a sterile, single-use plastic tube holder, designed with a safety shield, which can be activated to cover the needle immediately following venipuncture.

VACUETTE® QUICKSHIELD Complete (K072320)

The sterile VACUETTE® QUICKSHIELD Complete PLUS is a single packed VACUETTE® QUICKSHIELD Safety Tube Holder with pre-threaded VACUETTE® VISIO PLUS Blood Collection Needle in a blister pack. It is a sterile, single-use plastic tube holder, designed with a safety shield, which can be activated to cover the

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Greiner Bio-One. The logo features a stylized blue swoosh above the word "greiner" in a bold, sans-serif font. Below "greiner" is the text "BIO-ONE" in a smaller, sans-serif font.

needle immediately following venipuncture.

5. INDICATION FOR USE

The VACUETTE® QUICKSHIELD Complete is intended to be used only with VACUETTE® Blood Collection Tubes as a system in routine venipuncture procedures. This device is designed with a safety shield, which can be activated to cover the needle immediately following venipuncture to aid in the protection against accidental needlestick injury. They are for single-use only and should be used by properly trained healthcare professionals only in accordance with these instructions.

The VACUETTE® QUICKSHIELD Complete PLUS is intended to be used only with VACUETTE® Blood Collection Tubes as a system in routine venipuncture procedures. This device is designed with a safety shield, which can be activated to cover the needle immediately following venipuncture to aid in the protection against accidental needlestick injury. VACUETTE® VISIO PLUS Needles are designed for use in the daily blood collection routine when delegated by a qualified practitioner. The flashback window is situated in the transparent part of the cannula hub, which assists the user to recognize successful vein penetration. They are for singleuse only and should only be used by adequately trained healthcare personnel in accordance with these instructions.

6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Greiner Bio-One NA Inc. believes that the VACUETTE® QUICKSHIELD Complete (Plus) is substantially equivalent in intended use, indications for use, principles of operation, and technological characteristics to the predicate devices. Any differences between VACUETTE® QUICKSHIELD Complete (Plus) and the predicate devices do not raise new types of safety or effectiveness issues, as further discussed below.

Candidate DevicePrimaryPredicateDeviceSecondaryPredicateDeviceRemark
NameVACUETTE®QUICKSHIELDComplete (Plus)Blood CollectionNeedle with/withoutHolder, SafetyBlood CollectionNeedle with/withoutHolder, LuerAccess Device-VACUETTE®QUICKSHIELDCompleten/a
510(k) NumberK233979K200027K072320n/a
ClassificationClass IIClass IIClass IIn/a
Regulation862.1675862.1675880.5570n/a
Product CodePrimary product code:JKASubsequent product code:FMIPrimary product code:JKASubsequent productcode: FMIFMIn/a

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains the logo for Greiner Bio-One. The logo features a stylized blue arc above the word "greiner" in a bold, sans-serif font. Below "greiner" is the text "BIO-ONE" in a smaller, sans-serif font.

Indications for UseThe VACUETTE® QUICKSHIELD CompleteThe Blood Collection Needle with/without holderThe QUICKSHIELD CompleteDifferent. S&E evaluated by external and internal performance testing, see further information below.
The VACUETTE® QUICKSHIELD Complete is intended to be used only with VACUETTE® Blood Collection Tubes as a system in routine venipuncture procedures. This device is designed with a safety shield, which can be activated to cover the needle immediately following venipuncture to aid in the protection against accidental needlestick injury. They are for single-use only and should be used by properly trained healthcare professionals only in accordance with these instructions. The VACUETTE® QUICKSHIELD Complete PLUS is intended to be used only with VACUETTE® Blood Collection Tubes as a system in routine venipuncture procedures. This device is designed with a safety shield, which can be activated to cover the needle immediately following venipuncture to aid in the protection against accidental needlestick injury VACUETTE® VISIO PLUS Needles are designed for use in the daily blood collection routine when delegated by a qualified practitioner. The flashback window is situated in the transparent part of the cannula hub, which assists the user to recognize successful vein penetration. They are for single-use only and should only be used by adequately trained healthcare personnel in accordance with these instructions.The Blood Collection Needle with/without holder is intended to be used with vacuum blood collection tube for the collection of venous blood. The Safety Blood Collection Needle with/without holder is intended to be used with vacuum blood collection tube for the collection of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury. The Luer access device-holder with preattached multiple sample adapter is a sterile, non-invasive device used to connect devices with male or female luer connectors to blood collection tubes for the collection of blood.The QUICKSHIELD Complete is intended to be used only with VACUETTE® Blood Collection Tubes as a system in routine venipuncture procedures. This device is to be used by properly trained healthcare professionals only in accordance with these instructions. This device is designed with a safety shield, which can be activated to cover the needle immediately following venipuncture to aid in the protection against accidental needlestick injury. VACUETTE® VISIO PLUS Blood Collection Needles are designed for use in the daily blood collection routine when delegated by a qualified practitioner. The flashback window is situated in the transparent part of the cannula hub, which assists the user to recognize successful vein penetration. They are for single-use only and should only be used by adequately trained healthcare personnel in accordance with these instructions.Different. S&E evaluated by external and internal performance testing, see further information below.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Greiner BIO-ONE. The logo features a blue, curved shape above the word "greiner" in a bold, black sans-serif font. Below "greiner" is the text "BIO-ONE" in a smaller, black sans-serif font.

ManufacturerGreiner Bio-One (GBO)Jiangsu CainaMedical Co., LtdGreiner Bio-One (GBO)n/a
SafetymechanismActive safetymechanism. One-handactivation.The safety shield isintended to preventneedle sticks. Only onthe SafetyActive safetymechanism. One-hand activation.Same
Manual operation foractivation.ManualManual operationfor activation.
Cannulavariants• MultipleSampling Needle(K973620)• Visio Plus Needle(K061483)Various gages andlength• Multiple SamlingNeedle (K973620)• Visio Plus Needle(K061483)SE
CannulaGauges21G or 22G27G, 25G, 23G, 22G,21G, 20G, 18G21G or 22GSE
CannulaLength1" or 1.25" or 1.5"1" or 1.25" or 1.5"1" or 1.25" or 1.5"SE
SterileYesYesYesSE
SAL10-610-610-6SE
SterilizationE-Beam IrradiationETOETODifferent,evaluationaccording to ISO11137-1 and ISO11137-2.
Non-pyrogenicYesYesYesSE
Materials -complywith ISO10993-1YesYesYesSE
UseSingle use onlySingle use onlySingle use onlySE
Materials:
Needle CapPPSameSameSE
NeedleStainless Steel AISI 304SameSameSE
Rubber sleeveRubber SleeveRubber SleeveRubber SleeveSE
HolderPPSameSameSE
Safety shieldPPSameSameSE
Connection of:
Needle andHolderPre-attachedPre-attachedPre-attachedSE
Holder andshieldRotational shieldn/aNon-rotational shieldDifferent, S&Eevaluated byinternal testmethod

Comparison of Technological Characteristics

As can be seen in the above table a comparison of the VACUETTE® QUICKSHIELD Complete (Plus) and its features show that both the candidate device and the predicate devices have the same fundamental scientific technological characteristics. The VACUETTE® QUICKSHIELD Complete (Plus) and the predicate devices include the same safety mechanism covering the needle after the blood collection. The used safety mechanisms need to be activated manual activation of the safety mechanism of candidate device as well as the predicate devices the safe and locked position of the safety

{8}------------------------------------------------

Premarket Notification VACUETTE®QUICKSHIELD Complete (Plus) Greiner Bio-One

Image /page/8/Picture/1 description: The image shows the logo for Greiner Bio-One. The logo features a blue, curved shape above the word "greiner" in bold, black letters. Below "greiner" is the text "BIO-ONE" in smaller, black letters. The blue curved shape is reminiscent of a stylized wave or swoosh.

mechanism is reached by the Quickshield-needle-cover. The used materials for all products are the same or at least comparable and the differences do not have an influence on the safety and effectiveness of the candidate device. The candidate device and the predicate devices are provided sterile and single use only.

Comparison of Intended Use

The above comparison table shows the indications for use of the candidate device as well as predicate devices. The indications are comparable in their content so that all products are intended to be used for venous blood collection.

7. PERFORMANCE DATA

The VACUETTE® QUICKSHIELD Complete (Plus) is manufactured and verified by testing in conformity with applicable recognized consensus standards like ISO 23908, ISO 7864, and ISO 9626 but not limited to those standards. Simulated use tests were performed for venous blood collection procedure as well as the functionality of the safety mechanism was tested.

The VACUETTE® QUICKSHIELD Complete (Plus) is manufactured and tested in conformity with internal specifications and recognized consensus standards. Performance testing was conducted to demonstrate and support our substantial equivalency claim that the device performs as intended and in an equivalent manner as the predicate devices.

RequirementOutcome(Pass or fail)
Force to remove shieldPass
Torque to rotate shieldPass
Torque to screw out needlePass
Spin outPass
Force to activate safety mechanismPass
Force necessary to bend safety shield with needle by 90°Pass
Puncture forcePass
Friction forcePass
LeakagePass

Sterility testing:

VACUETTE® QUICKSHIELD Complete (Plus) are sterilized by e-beam irradiation for a final Sterility Assurance Level (SAL) of 10° per ISO 11137-1 Sterilization of health care products – Radiation – Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices. Bioburden testing for determination of verification dose and subsequently the processing dose was performed in accordance with ISO 11737-1 - Sterilization of medical devices - microbiological methods - Part 1: Determination of a population of microorganisms on products.

Biocompatibility testing:

Biocompatibility for the VACUETTE® QUICKSHIELD Complete (Plus) was performed in accordance with the standards of the ISO 10993 series and under consideration of the Guidance for industry and Food and Drug Administration Stuff - Use of the International Standard ISO 10993-1. "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" (Attachment A: Table A. 1 Endooints to be addressed in a biological risk assessment). The holder was classified as "surface device with short term contact to intact skin", and the cannula was classified as "external communicating device with short term contact to the blood path." Based on the classification the following biological endpoints were evaluated, and it was demonstrated that the candidate device meets the following standards:

{9}------------------------------------------------

Premarket Notification VACUETTE®QUICKSHIELD Complete (Plus) Greiner Bio-One

Image /page/9/Picture/1 description: The image shows the logo for Greiner Bio-One. The logo features the word "greiner" in a bold, sans-serif font, with the letters in black. Above the word "greiner" is a blue, curved shape that resembles a stylized "G". Below the word "greiner" is the text "BIO-ONE" in a smaller, sans-serif font, also in black.

  • Cytotoxicity (ISO 10993-5) ●
  • . Sensitization (ISO 10993-10)
  • . Irritation (ISO 10993-23)
  • Hemocompatibility (ISO 10993-4) .
  • Material mediated pyrogenicity (ISO 10993-11) ●
  • Acute Systemic Toxicity (ISO 10993-11) .

8. CONCLUSION

The test results and the comparison table above show the VACUETTE® QUICKSHIELD Complete (Plus) is substantially equivalent to its predicate devices with respect to intended use, technological and performance characteristics and does not raise new safety or effectiveness concerns.

§ 862.1675 Blood specimen collection device.

(a)
Identification. A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.(b)
Classification. Class II.