Intended Use

Dental Barrier and Sleeves are intended to be used as a disposable barrier for dental instruments and equipment. This device is non-sterile and intended for single patient use only.

Device Description

Dental Barrier and Sleeves consist of various sizes and shapes of polyethylene covers which are positioned on various small hand-held dental instruments such as hand pieces, curing lights, air/water syringes and similar hand instruments. In other forms, they are used to cover various devices such as dental chairs, dental instrument trays, x-ray heads, and others. The products are sold non-sterile, prepackaged, and are disposable, single use only.

AI/ML Overview

This document is a 510(k) premarket notification for a medical device called "Dental Barrier and Sleeves." It outlines the device's characteristics and compares them to a legally marketed predicate device to demonstrate substantial equivalence.

Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

Important Note: This submission is for a Class II medical device, "Dental Barrier and Sleeves," which are physical barriers (polyethylene covers) for dental instruments and equipment. The document does not describe an AI/Software as a Medical Device (SaMD) and therefore doesn't contain information related to AI model performance, expert consensus on image interpretation, or multi-reader multi-case studies. The performance criteria and studies described are for the physical properties and biocompatibility of the barrier sleeves themselves.


Acceptance Criteria and Reported Device Performance

The acceptance criteria are primarily based on demonstrating equivalence in physical properties and biocompatibility with the predicate device.

  1. Table of Acceptance Criteria and Reported Device Performance:
Acceptance Criteria CategorySpecific Acceptance Criteria (Method)Reported Device Performance and Outcome
Intended UseSame intended use as predicate deviceSame: "disposable barrier for dental instruments and equipment. This device is non-sterile and intended for single patient use only."
ClassificationClass II (21 CFR 878.4370)Same: Class II (21 CFR 878.4370)
Product CodePEMSame: PEM
MaterialPolyethylene film (same composition as predicate or justified)Same: Polyethylene film (LLDPE 80%, LDPE 20%)
Specifications & TolerancesFilm Thickness: 0.02-0.06mm, Tolerance: <0.01mm (similar to predicate)Film Thickness: 0.02-0.06mm, Tolerance: <0.01mm (Deemed "Similar" by sponsor)
SterilityNon-sterileNon-sterile
Performance PropertiesSynthetic Blood Penetration test (ASTM F1670) - PassPass
Viral Penetration test (ASTM F1671) - PassPass
Tear Resistance (ASTM D1004)Performed (Result implied to be acceptable for equivalence)
Tensile Strength (ASTM D882)Performed (Result implied to be acceptable for equivalence)
Puncture Resistance (ASTM F1342)Performed (Result implied to be acceptable for equivalence)
Effectiveness on X-ray devices (no impact on function)Performed to demonstrate no impact on dental x-ray devices' function.
BiocompatibilityCompliance with ISO 10993-5 (Cytotoxicity)Test results indicate non-cytotoxic.
Compliance with ISO 10993-10 (Sensitization)Test results indicate non-sensitizing.
Compliance with ISO 10993-23 (Irritation)Test results indicate non-irritating.
LabelingSingle use deviceSame: Single use device

  1. Sample Sizes Used for the Test Set and Data Provenance:

    • Test Set Sample Size: The document does not specify a specific "sample size" in terms of number of units for the performance tests (e.g., blood penetration, viral penetration, physical properties). It implies that standard testing procedures typical for medical devices of this type were followed.
    • Data Provenance: The studies are "Non-Clinical Testing" which means they were likely conducted in a laboratory setting. The country of origin for the tests or data is not explicitly stated. It is considered retrospective as the tests were performed to support the 510(k) submission.
  2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • This question is not applicable to this device. As a physical barrier, there is no "ground truth" to be established by experts in the context of image interpretation or diagnostic accuracy. The performance is based on measurable physical and biological properties.
  3. Adjudication Method for the Test Set:

    • This question is not applicable to this device. Adjudication methods like "2+1" or "3+1" are relevant for expert review of medical images or clinical outcomes, which is not the nature of the testing for this device.
  4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size:

    • No. This is a physical device, not an AI/SaMD. MRMC studies are not relevant to its evaluation.
  5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • No. This is a physical device, not an algorithm. Standalone performance is not relevant.
  6. The Type of Ground Truth Used:

    • The "ground truth" for this device's performance is based on standardized test methods and material properties.
      • For performance properties (blood/viral penetration, tear, tensile, puncture), the ground truth is defined by the passing criteria of the respective ASTM standards (ASTM F1670, ASTM F1671, ASTM D882, ASTM F1342, ASTM D1004).
      • For biocompatibility, the ground truth is defined by the "non-cytotoxic, non-sensitizing, and non-irritating" outcomes determined by the ISO 10993 suite of standards (ISO 10993-5, ISO 10993-10, ISO 10993-23).
      • For X-ray device effectiveness, the ground truth is the continued proper function of dental X-ray devices when covered by the barriers.
  7. The Sample Size for the Training Set:

    • Not applicable. This is a manufactured physical device, not an AI model. There is no "training set" in the context of machine learning.
  8. How the Ground Truth for the Training Set was Established:

    • Not applicable. As there is no training set, there is no ground truth to be established for it.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".

May 9, 2024

Yongqing Huaguan Dental Instruments Factory % Jarvis Wu Senior Consultant Shanghai SUNGO Management Consulting Co., Ltd. 14th Floor, Dongfang Building, 1500# Century Ave Shanghai, Shanghai 200122 CHINA

Re: K233918

Trade/Device Name: Dental Barrier and Sleeves Regulation Number: 21 CFR 878.4370 Regulation Name: Surgical Drape And Drape Accessories Regulatory Class: Class II Product Code: PEM Dated: April 10, 2024 Received: April 10, 2024

Dear Jarvis Wu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

that may require a new premarket notification are provided in the FDA

Page

2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Michael E. Adjodha -S

Michael Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K233918

Device Name

Dental Barrier and Sleeves

Indications for Use (Describe)

Dental Barrier and Sleeves are intended to be used as a disposable barrier for dental instruments and equipment. This device is non-sterile and intended for single patient use only.

See the attached table for the model numbers and descriptions in 510(k) Summary.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Intersection Of Ronghua Road And Siwei Street, Yongoing Industrial Area, Langfang City,065600,Hebei, China

510(K) Summary

K233918

Document prepared date: 2024/04/10

A. Applicant:

Yongqing Huaguan Dental Instruments Factory Address: Intersection Of Ronghua Road And Siwei Street. Yongqing Industrial Area. Langfang City,065600,Hebei, China Contact Person: Ms. ZengYan Wu Title: Sales Manager Tel: +0086-316-6659888 Fax: +0086-316-6659111 Email: hbhls1@126.com

Submission Correspondent: Primary contact: Mr. Jarvis Wu Title: Senior Consultant Shanghai SUNGO Management Consulting Co., Ltd. 14th Floor, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: jiawei.wu@sungoglobal.com

Secondary contact: Mr. Raymond Luo Title: Technical Director Shanghai SUNGO Management Consulting Co., Ltd. 14th Floor, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: zxfda@sungoglobal.com

B. Device:

Trade Name: Dental Barrier and Sleeves Common Name: Dental Barrier and Sleeves

Regulatory Information Classification Name: Surgical Drape And Drape Accessories Classification: Class II Product code: PEM Regulation Number: 21 CFR 878.4370

C. Predicate device:

Predicate Device: K190484 Product name: BH Medical Dental Barrier Sleeves and Barrier Film Manufacturer: BH Medical Products Co., Ltd.

D. Intended use of the device:

Dental Barrier and Sleeves are intended to be used as a disposable barrier for dental instruments and equipment. This device is non-sterile and intended for single patient use only

{5}------------------------------------------------

Intersection Of Ronghua Road And Siwei Street, Yongoing Industrial Area, Langfang City,065600,Hebei, China

E. Device Description:

Dental Barrier and Sleeves consist of various sizes and shapes of polyethylene covers which are positioned on various small hand-held dental instruments such as hand pieces, curing lights, air/water syringes and similar hand instruments. In other forms, they are used to cover various devices such as dental chairs, dental instrument trays, x-ray heads, and others. The products are sold non-sterile, prepackaged, and are disposable, single use only.

ItemModel#Model DescriptionDesigned for
Models for intra-orally use (Subject to this 510k submission)
1HL-6562/HL-6563Air/Water Syringe SleevesAir/Water Syringe, HVE Universal 2-1/2 " ×10 "
2HL-6564Month Mirror Covers orSleevesDental Mirror 1-1/2 " ×8 "
3HL-6565Low Speed HandpieceSleevesDental Low-Speed Handpiece Universal 1-1/2 " ×8 "
4HL-6566High Speed HandpieceSleevesDental High-Speed Handpiece Universal 1 " ×8 "
5HL-6577X-Ray CoversX-Ray Universal 15 " ×26 "
6HL-65783 - function sprayer Sleeves3-Fuction Sprayer 7-3/10 " ×5 "
7HL-6579Scaler CoversDental Scaler 1-3/5 " ×8 "
8HL-6580Intraoral Camera SheathIntraoral Camera 9-4/5 " ×2 "
9HL-6581Digital X-ray Sensor CoversDigital X-Ray Sensor, Universal 2-1/2 " ×10 "
10HL-6582-1Digital X-ray Sensor CoversDigital X-Ray Sensor, Universal 8 " ×1-5/8 "
11HL-6582-2Digital X-ray Sensor CoversDigital X-Ray Sensor, Universal 8 " ×1-3/8 "
12HL-6585Syringe Sleeves (Roll bag)Syringe 2.2 " ×17.7 "
13HL-6586Turbine Motor Sleeves (Rollbag)Turbine 1.8 " ×17.7 "
14HL-6588SCuring Light SleevesDental Curing Light 4.33"X2.95"
15HL-6588LCuring Light SleevesDental Curing Light 4.33"X2.95"
16HL-6597Curing Light SleevesDental Curing Light 4.7" ×1.2"
17HL-6592Curing Light SleevesDental Curing Light 13 " ×2.1 "
18HL-6593Curing Light SleevesDental Curing Light 13 " ×3.3 "
19HL-6594Intraoral Camera SleeveIntraoral Camera 10.2 " ×2 "
20HL-6595Curing Light SleevesDental Curing Light 9.9"X1.85"
21HL-6596Curing Light SleevesDental Curing Light 12.4"X4.13"
22HL-6604Disposable Barrier SleevesBarrier Sleeve 9.25 " ×3.15 "
23HL-6605Disposable Barrier SleevesBarrier Sleeve 11.2 " ×2.2 "
24HL-6606Disposable Barrier SleevesBarrier Sleeve 11.2 " ×2.2 "

Table for the model numbers and descriptions :

{6}------------------------------------------------

Intersection Of Ronghua Road And Siwei Street, Yongqing Industrial Area, Langfang City,065600,Hebei, China

25HL-6607Low Speed Handpiece CoverLow Speed Handpiece 7.5 "x1.53 "
26HL-6608X-ray sensors CoverX-Ray Sensor 8 " x1-5/8 "
27HL-6610Digital Sensor EnvelopesDigital Sensor 2"X4.5"
28HL-6611Digital Sensor EnvelopesDigital Sensor 1-5/8"X4"
29HL-6612Digital Sensor EnvelopesDigital Sensor 1.6"X4.5"
30HL-6613Digital Sensor EnvelopesDigital Sensor 1.4"X4.3"
31HL-6614Bite Block CoversBite Block 1.4"X2.5"
32HL-6615Intraoral Camera SleeveIntraoral Camera 35mm*110mm
33HL-6579SScaler CoversDental Scaler 1.4 " x7 "
34HL-6701Fitted Digital Sensor SleevesDigital Sensor 2" ×9.124"
35HL-20963Intraoral Camera SheathIntraoral Camera 220mm*48mm
36HL-21013Intraoral Camera SheathIntraoral Camera 210*57mm
37HL-20855Intraoral Camera SheathIntraoral Camera 228mm*57mm
38HL-20999X-ray Sensor SheathDigital X-Ray Sensor(Dexis) Compatible TIDI:20999225mm*57mm
39HL-20978X-ray Sensor SheathDigital X-Ray Sensor(Kodak 6100 Size 1)CompatibleTIDI:20978 226mm*57mm
40HL-20824X-ray Sensor SheathFor SCHICK Size 1 Compatible TIDI:20824220mm*47mm
41HL-20861X-ray Sensor SheathDigital Sensor 223mm*57mm
42HL-20825X-ray Sensor SheathFor SCHICK Size 2 Compatible TIDI:20825220mm*57mm
43HL-20979X-ray Sensor SheathDigital X-Ray Sensor(Kodak 6100 Size 2)223mm*56mm
44HL-20831Intraoral Camera SheathIntraoral Camera 225mm*55mm
45HL-6577AX-ray BarrierX-Ray 23" X 31 "
46HL-6577BX-ray BarrierX-Ray 14" X 13" X 24.5"
47HL-100Barrier FilmSurface Area That Might Be Touched DuringProcedure 4"X6"
48HL-101Barrier Film BlueSurface Area That Might Be Touched DuringProcedure 4"X6"
49HL- A Size 0X-ray Barrier EnvelopesPhosphor Plates Size 0 Inner: 22×35mm
50HL- A Size 1X-ray Barrier EnvelopesPhosphor Plates Size 1 Inner:24mmx40mm
51HL- A Size 2X-ray Barrier EnvelopesPhosphor Plates Size Inner:31mmx41mm
52HL- A Size 3X-ray Barrier EnvelopesPhosphor Plates Size 3 Inner: 27mmx54mm
53HL- A Size 4X-ray Barrier EnvelopesPhosphor Plates Size 4 Inner: 57mmx76mm
54HL- B Size 0X-ray Barrier EnvelopesPhosphor Plates Size 0 Inner: 22×35mm
55HL- B Size 1X-ray Barrier EnvelopesPhosphor Plates Size 1 Inner:24mmx40mm
56HL-B Size 2X-ray Barrier EnvelopesPhosphor Plates Size Inner:31mmx41mm
57HL-B Size 3X-Ray Barrier EnvelopesPhosphor Plates Size 3 Inner: 27mmx54mm
58HL-B Size 4X-Ray Barrier EnvelopesPhosphor Plates Size 4 Inner: 57mmx76mm
59IC-001Intraoral Camera SleevesIntraoral Camera 1" HEAD X 2" OPENING X8.25" LENGTH
60IC-085Intraoral Camera SleevesIntraoral Camera 0.85" HEAD X 1.75" OPENING X8.25" LENGTH
61PS-00/01Digital Sensor CoverDental Sensor 9"L * 1-3/4"W
62PS-002Comfort Sensor SleevesDental Sensor 9"L*2"W
63PS-5500Perforated Bite Block CoversBite Block 1"W * 2"L
64PS-5501Perforated Bite Block CoversBite Block 1 1/2"W X 3"L
65PS-5502Large Digital Sensor CoversDigital Sensor 2-3/4"W * 8"L
66PS-5505Universal Digital SensorCoversDigital Sensor 1-1/2"W *7"L
67PS-5511Large Comfort Sensor CoversDental Sensor 2"W *4-1/2"L
68PS-5522Comfort Sensor Covers, #2Dental Sensor 1-5/8"W * 4-1/8"L
69PS-5533Comfort Sensor Covers, #0/#1Dental Sensor 1-3/8"W * 4-1/4"L
70PS-CYGNUS-1Digital Sensor SleevesCygnus#1Digital Dental Sensor 24*3.5
71PS-DEXIS-UDigital Sensor Sleeves DexisUniversalDigital Dental Sensor 9.5"*1.77"
72PS308Air/Water Syringe CoverAir Water Syringe 2 ½" X 8 ¼"
73PS321Air/Water Syringe Cover (¾"Seal)Air Water Syringe 2 ½" X10"
74PS3767-WOAir/Water Syringe Cover(Notched)Air Water Syringe 2 ½" X10"
75PS4500Curing Light Tip SleevesCuring Light FOR 7-8mm TIP STICK HOLD
76PS4511Curing Light Tip SleevesCuring Light FOR 11mm TIP STICK HOLD
77PS4700Cordless Curing Light CoverCuring Light 1 ¾" X 9 ¾"
78PS4701Cordless Curing Light CoverCuring Light 2" X 12 ¾"
79PS4707Cordless Curing Light CoverCuring Light 4" X 12 ½"
80PS4708Cordless Curing Light CoverCuring Light 3 ¼" X 13"
81PS730Scaler Sleeves-Stick HoldDental Scaler Stick Hold 1-1/2"W*7"L
82PS735Scaler SleevesDental Scaler 1-⅝" X 8"
83XS-6100-0Digital Sensor Sheath ForKodak Size 0Digital Sensor 24*3.5
84XS-6100-1Digital Sensor Sheath ForKodak Size 1Digital Sensor 24*4.3
85XS-6100-2Digital Sensor Sheath ForKodak Size 2Digital Sensor 24*4.3
86XS-CYGNUS-1Digital Sensor Sheath ForCygnus Size 1Digital Sensor 24*3.5
87XS-DEXIS-2Digital Sensor Sheath ForDexis Size 2Digital Sensor 24*3.9
88XS-DEXIS-UDigital Sensor Sheath ForDexis UniversalDigital Sensor
89XS-GXDR-1Digital Sensor Sheath ForGendex/Xdr Size 1Digital Sensor 24*3.4
90XS-GXDR-2Digital Sensor Sheath ForGendex/Xdr Size 2Digital Sensor 24*3.9
91XS-SHICK-1Digital Sensor Sheath ForSchick Size 1Digital Sensor 24*3.5
92XS-SHICK-2Digital Sensor Sheath ForSchick Size 2Digital Sensor 24*4.3
93SS100Single Syringe SleevesSingle Syringe 210mm51mm0.0275
94SS200Automix Syringe SleevesAutomix Syringe 235mm65mm0.0275
Models for non intra-orally use
95HL-6567T Light Handle CoversLight Handle For T-Style 4 " L×5-3/4 " W
96HL-6568Tray SleevesDental Instrument Tray 11-5/8 " ×16 "
97HL-6569Tray SleevesDental Instrument Tray 10-1/2 " ×14 "
98HL-6569STray SleevesDental Instrument Tray 7 1/2 " ×10 1/2 "
99HL-6571Headrest CoversDental Chair Headrest 11"X10"
100HL-6572Headrest CoversDental Chair Headrest 14"X10"
101HL-6573-1Half Chair CoversHalf Dental Chair 27-1/2 " ×24 "
102HL-6573-2Half Chair CoversHalf Dental Chair 32 " ×32 "
103HL-6574/HL-6576Full Chair CoversFull Dental Chair 29 " ×80 "
104HL-6575Full Chair CoversWide Full Dental Chair 48 " ×56 "
105HL-6583T Light Handle CoversLight Handle Cover 6 " ×2-9/10 "
106HL-6587Headrest Covers (Roll bag)Dental Chair Headrest 11.8 " ×9 "
107HL-6601PC mouse CoversPC Mouse 6.25 " ×3.5 "
108HL-6702Keyboard CoversKeyboard 22"X13.78"
109HL-6568-2Tray BarriersDental Instrument Tray 11-5/8" X 14" Size A
110HL-6569-2Tray BarriersDental Instrument Tray 11.5" X 16" Size E
111HL-7510Tray BarriersDental Instrument Tray 7.5" X 10.5" Size F
112HL-7521Keyboard CoverKeyboard 22" X 14"
113PS101Small Half Chair CoverDental Chair 20''W X 30''L
114PS105Full Chair CoverDental Chair 33''W X 54''L
115PS203Size E Tray Cover W/LockTopPlastice/Metal Trays 11-1/2"*16"
116PS328TT-Shape Lite Handle CoverT-Shape Lite Handle 6"W X 3"L
117PS408Wireless Keyboard CoversKeyboard 24"Lx 10"W
118PS410-STWDWide Lcd CoverMonitor And Keyboard 25"WX16"L W/3" POCKET
119PS410-WDWide Lcd+Keyboard CoverMonitor And Keyboard 25"W*26"Lw/3"POCKET
120PS670Chair Armrest CoversChair Armrest 24"Lx6"W
121PS680Jumbo Headrest CoverChair Headrest 14 1/2"L X 14 3/4"W

{7}------------------------------------------------

Intersection Of Ronghua Road And Siwei Street, Yongqing Industrial Area, Langfang City,065600,Hebei, China

{8}------------------------------------------------

Intersection Of Ronghua Road And Siwei Street, Yongqing Industrial Area, Langfang City,065600,Hebei, China

{9}------------------------------------------------

Intersection Of Ronghua Road And Siwei Street, Yongqing Industrial Area, Langfang City,065600,Hebei, China

F. Comparison with predicate device

DeviceSubject DevicePredicate DeviceComment
ManufacturerYongqing Huaguan DentalInstruments FactoryBH Medical Products Co., Ltd../
510K numberK233918K190484/
Product NameDental Barrier and SleevesBH Medical Dental Barrier Sleevesand Barrier Film/
ClassificationClass II Device (21 CFR 878.4370)Class II Device (21 CFR 878.4370)Same
Intended useDental Barrier and Sleeves areintended to be used as a disposablebarrier for dental instruments andequipment. This device is non-sterileand intended for single patient useonlyBH Medical Dental Barrier Sleevesand Barrier Film are intended to beused as a disposable barrier for dentalinstruments and equipment. Thisdevice is non-sterile and intended forsingle patient use only.Same
DeviceDescriptionDental Barrier and Sleeves consist ofvarious sizes and shapes ofpolyethylene covers which arepositioned on various small hand-helddental instruments such as handpieces, curing lights, air/watersyringes and similar handinstruments. In other forms, they areused to cover various devices such asdental chairs, dental instrument trays,x-ray heads, and others. The productsare sold non-sterile, prepackaged, andare disposable, single use only.BH Medical Dental Barrier Sleevesand Barrier Film consist of varioussizes and shapes of clear polyethylenecovers which are positioned onvarious small hand-held dentalinstruments such as hand pieces,curing lights, air/water syringes andsimilar hand instruments. In otherforms, they are used to cover variousdevices such as dental chairs, dentalinstrument trays, x-ray heads, andothers. The products are sold non-sterile, prepackaged, and aredisposable, single use only.Same
ClassificationProduct CodePEMPEMSame
MaterialPolyethylene filmPolyethylene filmSame
MaterialCompositionLLDPE(80%),LDPE(20%)LLDPE(80%),LDPE(20%),Bluepigment(#992482),Adhesive(#992482,#992483)Same
Specificationsand TolerancesPaper Backing: Some of the model;Film Thickness: 0.02-0.06mmTolerance:<0.01mmPaper Backing: NoFilm Thickness: 0.025mmTolerance:<0.01mmSimilar
LabelingSingle use deviceSingle use deviceSame
SterilityNon-sterileNon-sterileSame
PerformancePropertiesSynthetic Blood Penetration test(ASTM F1670) -PassViral Penetration test (ASTM F1671)-PassSynthetic Blood Penetration test(ASTM F1670) -PassViral Penetration test (ASTM F1671)-PassSame
BiocompatibilityBiocompatibility tests wereconducted in compliance with ISO10993-5, ISO 10993-10 and ISO10993-23Test results indicate the device isnon-cytotoxic, non-sensitizing andnon-irritating.Biocompatibility tests wereconducted in compliance with ISO10993-5 and ISO 10993-10.Test results indicate the device isnon-cytotoxic, non-sensitizing andnon-irritating.Same
DimensionsDifference in dimensions is due tothe size of instruments and equipmentthey coverDifference in dimensions is due tothe size of instruments and equipmentthey coverSame
FDA-RecognizedStandardsASTM F1670ASTM F1671ASTM D882ASTM F1342ASTM D1004ISO 10993-5ISO 10993-10ISO 10993-23ASTM F1670ASTM F1671ASTM D882ASTM F1342ASTM D1004ISO 10993-5ISO 10993-10Same

{10}------------------------------------------------

Intersection Of Ronghua Road And Siwei Street, Yongoing Industrial Area, Langfang City,065600,Hebei, China

Substantial Equivalence Discussion:

The subject and predicate devices have the same intended use. Both are single-use devices, and material and performance characteristic are identical. The minor differences in specifications and tolerances between the subject and predicate devices do not raise questions of safety and effectiveness.

G. Non-Clinical Testing

Performance test including resistance to penetration, tear resistance, tensile strength, puncture resistance was performed. Biocompatibility test in accordance with ISO 10993-10 and ISO 10993-23 has been performed. Effectiveness of x-ray devices covered with Dental Barrier and Sleeves were performed to demonstrate the subject device does not impact the function of dental x-ray devices.

The biocompatibility test result indicates that the barrier sleeves do not have the potential cytotoxicity, sensitization and irritation.

H. Clinical Test Conclusion

No clinical study is included in this submission.

I. Conclusion

Based on the comparison and analysis above, the subject device, Dental Barrier and Sleeves, is as safe, as effective, and performs as well as or better than the legally marketed predicate devices.

§ 878.4370 Surgical drape and drape accessories.

(a)
Identification. A surgical drape and drape accessories is a device made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination. The device includes a plastic wound protector that may adhere to the skin around a surgical incision or be placed in a wound to cover its exposed edges, and a latex drape with a self-retaining finger cot that is intended to allow repeated insertion of the surgeon's finger into the rectum during performance of a transurethral prostatectomy.(b)
Classification. Class II (special controls). The device, when it is an ear, nose, and throat surgical drape, a latex sheet drape with self-retaining finger cot, a disposable urological drape, a Kelly pad, an ophthalmic patient drape, an ophthalmic microscope drape, an internal drape retention ring (wound protector), or a surgical drape that does not include an antimicrobial agent, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.