K Number
K233876

Validate with FDA (Live)

Date Cleared
2024-06-21

(197 days)

Product Code
Regulation Number
886.4670
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

UNITY VCS:
The UNITY VCS console, when used with compatible devices, is indicated for use during anterior segment (i.e. phacoemulsification and removal of cataracts) and posterior segment (i.e. vitreoretinal) ophthalmic surgery.

In addition, with the optional laser this system is indicated for photocoagulation (i.e. vitreoretinal and macular pathologies), iridotomy, and trabeculoplasty procedures.

UNITY CS:
The UNITY CS console, when used with compatible devices, is indicated for use during anterior segment (i.e. phacoemulsification and removal of cataracts) ophthalmic surgery.

Device Description

The UNITY VCS is a multifunctional surgical instrument for use in anterior and posterior segment ophthalmic surgeries. The capabilities of the UNITY VCS include driving a variety of handpieces that provide the ability to cut vitreous and tissues, emulsify the crystalline lens, illuminate the posterior segment of the eye, and apply diathermy to stop bleeding. Flow-controlled or vacuumcontrolled aspiration is used to remove ocular matter from the eye. Pressure-controlled irrigation/infusion capability is provided to replace fluid in the eye and enters the eye directly through either an infusion cannula or a handpiece. The graphical user interface is menu-driven. The operator provides inputs using the touchscreen panel, the remote control, and the foot controllers.

An optional, fully integrated laser module is available that can be installed in the base of the UNITY VCS configuration for vitreoretinal surgery. The laser delivers the same visible 532-nm green treatment beam designed for ophthalmic use as in the predicate device.

When used with compatible components and accessories, the system will perform anterior and posterior segment ophthalmic procedures, including irrigation, phacoemulsification, extraction, vitrectomy, diathermy coagulation, intraocular illumination, photocoagulation, viscous fluid control, tissue grabbing and cutting.

The UNITY CS is a multifunctional surgical instrument for use in anterior segment ophthalmic surgeries. The capabilities of the UNITY CS include driving a variety of handpieces that provide the ability to cut anterior vitreous and tissues, emulsify the crystalline lens, and apply diathermy to stop bleeding. Flow-controlled or pressure-controlled aspiration is used to remove ocular matter from the eye. Pressure-controlled irrigation/infusion capability is provided to replace fluid in the eye and enters the eye directly through either an infusion cannula or a handpiece. The graphical user interface is menu-driven, and the operator provides inputs using the touchscreen panel, the remote control, and the foot controllers.

When used with compatible components and accessories, the system will perform anterior segment ophthalmic procedures, including irrigation, phacoemulsification, extraction, vitrectomy, and diathermy coagulation.

The UNITY VCS and UNITY CS are intended to be used within a clinic(s)/hospital(s)/surgical practice network.

Both devices are for prescription use only.

AI/ML Overview

The provided text does not contain information about studies involving AI/ML components or their acceptance criteria. The document is an FDA 510(k) premarket notification for two ophthalmic surgical systems, UNITY VCS and UNITY CS, demonstrating their substantial equivalence to predicate devices. The studies summarized are non-clinical testing for biocompatibility, sterilization and shelf life, electromagnetic compatibility/wireless/electrical safety, software, and performance testing, all of which are standard for medical device clearance and do not involve AI/ML performance metrics, expert consensus, or MRMC studies.

Therefore, I cannot provide the requested table and information as it is not present in the given text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

June 21, 2024

Alcon Laboratories, Inc. Ganesh Balachandar Regulatory Affairs Manager 6201 South Freeway Fort Worth, Texas 76134

Re: K233876

Trade/Device Name: UNITY VCS (8065000296); UNITY CS (8065000297) Regulation Number: 21 CFR 886.4670 Regulation Name: Phacofragmentation system Regulatory Class: Class II Product Code: HQC, HQE, HQF Dated: May 17, 2024 Received: May 20, 2024

Dear Ganesh Balachandar:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Claudine H. Krawczyk -S

Claudine Krawczyk Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K233876

Device Name UNITY VCS (8065000296) ; UNITY CS (8065000297)

Indications for Use (Describe) UNITY VCS

The UNITY VCS console, when used with compatible devices, is indicated for use during anterior segment (i.e. phacoemulsification and removal of cataracts) and posterior segment (i.e. vitreoretinal) ophthalmic surgery.

In addition, with the optional laser this system is indicated for photocoagulation (i.e. vitreoretinal and macular pathologies), iridotomy, and trabeculoplasty procedures.

UNITY CS

The UNITY CS console, when used with compatible devices, is indicated for use during anterior segment (i.e. phacoemulsification and removal of cataracts) ophthalmic surgery.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

In accordance with 21 CFR 807.92, Alcon hereby provides the 510(k) summary for the UNITY VCS and UNITY CS.

1. SUBMITTER

ApplicantAlcon Laboratories, Inc.6201 South FreewayFort Worth, TX 76134-2099
Primary CorrespondentGanesh Balachandar, M.S.Regulatory Affairs ManagerAlcon Research LLC,on behalf of Alcon Laboratories, Inc.20511 Lake Forest DriveLake Forest, CA 92630-7741+1 (949) 322-5101ganesh.balachandar@alcon.com
Secondary CorrespondentOphelia BiggsDirector, Regulatory AffairsAlcon Research LLC,on behalf of Alcon Laboratories, Inc.20511 Lake Forest DriveLake Forest, CA 92630-7741+1 (949) 217-3854ophelia.biggs@alcon.com
Date PreparedJune 21, 2024

{5}------------------------------------------------

2. SUBJECT DEVICES

Device Trade NameUNITY VCS (8065000296);UNITY CS (8065000297)
ClassificationNumber and Name(Product Code inParentheses)For UNITY VCS:• 21 CFR 886.4150, Vitreous aspiration and cutting instrument(HQE)• 21 CFR 886.4670, Phacofragmentation system (HQC)• 21 CFR 886.4690, Ophthalmic laser (HQF)For UNITY CS:• 21 CFR 886.4670, Phacofragmentation system (HQC)
Regulatory ClassII

3. PREDICATE DEVICES

FOR UNITY VCS:
Predicate DeviceCONSTELLATION Vision System
510(k) NumberK141065
ClassificationNumber and Name(Product Code inParentheses)21 CFR 886.4150, Vitreous aspiration and cutting instrument (HQE) 21 CFR 886.4670, Phacofragmentation system (HQC)
Regulatory ClassII
FOR UNITY VCS (Laser Functionality):
Predicate DeviceNext-Generation Laser
510(k) NumberK062624
ClassificationNumber and Name(Product Code inParentheses)• 21 CFR 886.4390, Laser, Ophthalmic (HQF)
Regulatory ClassII
FOR UNITY CS:
Predicate DeviceCENTURION Vision System with Active Sentry
510(k) NumberK161794
Classification21 CFR 886.4670, Phacofragmentation system (HQC)

{6}------------------------------------------------

Number and Name(Product Code inParentheses)
Regulatory ClassII

{7}------------------------------------------------

DEVICE DESCRIPTION 4.

The UNITY VCS is a multifunctional surgical instrument for use in anterior and posterior segment ophthalmic surgeries. The capabilities of the UNITY VCS include driving a variety of handpieces that provide the ability to cut vitreous and tissues, emulsify the crystalline lens, illuminate the posterior segment of the eye, and apply diathermy to stop bleeding. Flow-controlled or vacuumcontrolled aspiration is used to remove ocular matter from the eye. Pressure-controlled irrigation/infusion capability is provided to replace fluid in the eye and enters the eye directly through either an infusion cannula or a handpiece. The graphical user interface is menu-driven. The operator provides inputs using the touchscreen panel, the remote control, and the foot controllers.

An optional, fully integrated laser module is available that can be installed in the base of the UNITY VCS configuration for vitreoretinal surgery. The laser delivers the same visible 532-nm green treatment beam designed for ophthalmic use as in the predicate device.

When used with compatible components and accessories, the system will perform anterior and posterior segment ophthalmic procedures, including irrigation, phacoemulsification, extraction, vitrectomy, diathermy coagulation, intraocular illumination, photocoagulation, viscous fluid control, tissue grabbing and cutting.

The UNITY CS is a multifunctional surgical instrument for use in anterior segment ophthalmic surgeries. The capabilities of the UNITY CS include driving a variety of handpieces that provide the ability to cut anterior vitreous and tissues, emulsify the crystalline lens, and apply diathermy to stop bleeding. Flow-controlled or pressure-controlled aspiration is used to remove ocular matter from the eye. Pressure-controlled irrigation/infusion capability is provided to replace fluid in the eye and enters the eye directly through either an infusion cannula or a handpiece. The graphical user interface is menu-driven, and the operator provides inputs using the touchscreen panel, the remote control, and the foot controllers.

When used with compatible components and accessories, the system will perform anterior segment ophthalmic procedures, including irrigation, phacoemulsification, extraction, vitrectomy, and diathermy coagulation.

The UNITY VCS and UNITY CS are intended to be used within a clinic(s)/hospital(s)/surgical practice network.

Both devices are for prescription use only.

INDICATIONS FOR USE ડાં

UNITY VCS:

The UNITY VCS console, when used with compatible devices, is indicated for use during anterior segment (i.e. phacoemulsification and removal of cataracts) and posterior segment (i.e.

{8}------------------------------------------------

vitreoretinal) ophthalmic surgery.

In addition, with the optional laser this system is indicated for photocoagulation (i.e. vitreoretinal and macular pathologies), iridotomy and trabeculoplasty procedures.

UNITY CS:

The UNITY CS console, when used with compatible devices, is indicated for use during anterior segment (i.e. phacoemulsification and removal of cataracts) ophthalmic surgery.

{9}------------------------------------------------

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS TO PREDICATE 6.

CharacteristicCONSTELLATIONVision System(Predicate Device forSubject Device 1)UNITY VCS(Subject Device 1)CENTURION VisionSystem with ActiveSentry (Predicate Devicefor Subject Device 2)UNITY CS(Subject Device 2)
Administrative
510(k)K141065K233876K161794K233876
Intended UseAnterior and posteriorsegment ophthalmicsurgerySame(The subject device,consisting of the consoleand compatible devices, isintended to facilitatemanagement of fluid andgases, as well as removal,grasping, cutting,illumination, andcoagulation of ocularmaterials)1Anterior segmentophthalmic surgerySame(The subject device,consisting of the consoleand compatible devices, isintended to facilitatemanagement of fluid aswell as removal, cutting,and coagulation of ocularmaterials)1
Indications for UseThe CONSTELLATION®Vision System is indicatedfor the following:The CONSTELLATION®Vision System is anophthalmic microsurgicalsystem that is indicated forboth anterior segment (i.e.phacoemulsification andThe UNITY VCS console,when used withcompatible devices, isindicated for use duringanterior segment (i.e.phacoemulsification andremoval of cataracts) andposterior segment (i.e.vitreoretinal) ophthalmicThe Centurion VisionSystem is indicated foremulsification, separation,irrigation, and aspirationof cataracts, residualcortical material and lensepithelial cells, vitreousaspiration and cuttingassociated with anteriorThe UNITY CS console,when used withcompatible devices, isindicated for use duringanterior segment (i.e.phacoemulsification andremoval of cataracts)ophthalmic surgery.
CharacteristicCONSTELLATIONVision System(Predicate Device forSubject Device 1)UNITY VCS(Subject Device 1)CENTURION VisionSystem with ActiveSentry (Predicate Devicefor Subject Device 2)UNITY CS(Subject Device 2)
removal of cataracts) andposterior segment (i.e.,vitreoretinal) ophthalmicsurgery in addition to theindications included withthe optional Next-Generation laser. TheAutoSert® IOL InjectorHandpiece is intended todeliver qualifiedACRYSOF® intraocularlenses into the eyefollowing cataractremoval.The AutoSertⓇIOL Injector Handpieceachieves the functionalityof injection of intraocularlenses. The AutoSert isindicated for use withACRYSOF® lensesSN60WF, SN6AD1,SN6AT3 throughSN6AT9, as well asapproved ACRYSOF®lenses that are specificallyindicated for use with thisinserter as indicated in thesurgery.In addition, with theoptional laser this systemis indicated forphotocoagulation (i.e.vitreoretinal and macularpathologies), iridotomyand trabeculoplastyprocedures.2vitrectomy, bipolarcoagulation, andintraocular lens injection.The AutoSert IOL InjectorHandpiece is intended todeliver qualified AcrySofintraocular lenses into theeye following cataractremoval.The AutoSert IOL InjectorHandpiece achieves thefunctionality of injectionof intraocular lenses. TheAutoSert IOL InjectorHandpiece is indicated foruse with the AcrySoflenses SN60WF,SNWAD1, SN6AT3through SN6AT9,as well as approvedAcrySof lenses that arespecifically indicated foruse with this inserter, asindicated in the approvedlabeling of those lenses.
CharacteristicCONSTELLATIONVision System(Predicate Device forSubject Device 1)UNITY VCS(Subject Device 1)CENTURION VisionSystem with ActiveSentry (Predicate Devicefor Subject Device 2)UNITY CS(Subject Device 2)
approved labeling of thoselenses.
Console Features
MonolithYesYesYesYes
TrayYesYesYesYes
Touch Panel DisplayYesYesYesYes
RFID for accessoryconnectionsYesYesNoYes
e-ConnectivityYesYesYesYes
User InterfaceTouch panel display, footcontrollers, remote controlSameSameSame
Electrical PowerSpecifications100-240 V AC, 50/60 HzSameSameSame
OperatingTemperature Range10-35°CSameSameSame
Console Height,Width and DepthHeight: 62.4"Width: 30.3"Depth: 36"Height: 64.6"Width: 25.4"Depth:28.3"Height:64.6"Width:19.6'Depth:31.6"Height: 64.6"Width: 25.4"Depth:28.3"
Maximum Humidity(RH = RelativeHumidity)95% RH non-condensingSameSameSame
Fluidics Features
Push PrimeYesYesNoYes
Infusion/IrrigationSource ChangeYesYesYesYes
Aspiration PressureControlYesYesYesYes
Aspiration FlowControlYesYesYesYes
CharacteristicCONSTELLATIONVision System(Predicate Device forSubject Device 1)UNITY VCS(Subject Device 1)CENTURION VisionSystem with ActiveSentry (Predicate Devicefor Subject Device 2)UNITY CS(Subject Device 2)
Suction FlowControlYesYesNoNo
Flow Control –IrrigationNoYesYesYes
Flow Control –InfusionNoYesNoNo
IOP-ControlledInfusion/IrrigationYesYesYesYes
Pressurized InfusionIrrigationYesYesNoNo
YesYesYesYes
Proportional RefluxYesYesNoYes
Continuous RefluxYesYesYesYes
Micro RefluxYesYesNoYes
Pneumatics Features
Anterior VitrectomyYesYesYesYes
Posterior VitrectomyYesYesNoNo
Proportional ForcepsYesYesNoNo
Proportional ScissorsYesYesNoNo
Multi-Cut ScissorsYesYesNoNo
VFC InjectYesYesNoNo
VFC ExtractYesYesNoNo
Fluid-Air ExchangeYesYesNoNo
Auto-Gas FillingYesNoNoNo
Ultrasound Features
Longitudinal PhacoYesYesYesYes
Longitudinal FragYesYesNoNo
Torsional PhacoYesYesYesYes
CharacteristicCONSTELLATIONVision System(Predicate Device forSubject Device 1)UNITY VCS(Subject Device 1)CENTURION VisionSystem with ActiveSentry (Predicate Devicefor Subject Device 2)UNITY CS(Subject Device 2)
Torsional FragNoYesNoNo
Active Sentry PhacoNoYesYesYes
4D PhacoNoYesNoYes
Diathermy Features
Frequency1.5 MHzSameSameSame
Rated Load75 ΩSameSameSame
Maximum Power10 WSameSameSame
Illumination Features
Illumination sourceXenon BulbRGB LEDNoneNone
Number ofSimultaneousOutputs23NoneNone
Illumination ColorFixedAdjustableNoneNone
Laser Features
Therapeutic laserwavelength532 nmSameNoNo
Single-spot poweroutput range30 mW to 2 WSameNoNo
Laser IndirectOphthalmoscopesupportYesYesNoNo
Includes Multi-SpotLaser Probe (MSLP)NoYesNoNo
IntegratedIllumination (MSLPonly)NoYesNoNo
FMS and Accessory Device Packs
Single-UseYesYesYesYes
CharacteristicCONSTELLATIONVision System(Predicate Device forSubject Device 1)UNITY VCS(Subject Device 1)CENTURION VisionSystem with ActiveSentry (Predicate Devicefor Subject Device 2)UNITY CS(Subject Device 2)
Sterilization methodEthylene OxideSameSameSame
Includes AnteriorPackYesYesYesYes
Includes PosteriorPackYesYesNoNo
IncludesCombination PackYesYesNoNo
Surgical Suite Integration
NGENUITY 3DVisualization SystemYesYesYesYes
LuxOR microscopeNoYesYesYes
Verion ImageGuided SystemNoYesYesYes

' The proposed language for the subject devices' Intended Use statements are in parenthe manner of describing the medical pupose of the subject devices. Despite the difference in worder the same between the language in parentheses and the intended use language of the predicate devices (anterior and posterior segment ophthalmic surgery), because the actions of filud and gas management, removal, grasping, cutting, illumination, and coagulation of ocular materials can occur only during anterior or posterior segment ophthalmic surgery.

{10}------------------------------------------------

2 The laser functionality of UNITY VCS is substantially equivalent to Alcon's Next-Generation Laser (K062624), cleared November 2007.

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

Page 10 of 13

{14}------------------------------------------------

{15}------------------------------------------------

SUMMARY OF STUDIES 7.

7.1 Non-Clinical Testing

Biocompatibility

The UNITY VCS and UNITY CS consoles are not intended to have patient contact. The materials used for the device consoles are common and widely used for ophthalmic and similar applications without reported heath concerns. Per ISO 10993-1, no further biocompatibility testing is required.

Alcon conducted biocompatibility testing on the patient-contacting accessories and met ISO 10993 requirements for chemical characterization, cytotoxicity, sensitization, and acute systemic toxicity. Packaging materials passed cytotoxicity requirements from ISO 11607-1 and ASTM F2475.

Sterilization and Shelf Life

The UNITY VCS and CS system consoles are provided non-sterile and are not intended to be sterilized during routine use. The system consoles are intended for use with sterile accessories.

Alcon performed sterilization testing on the accessories provided sterile to the user. Test results passed the predetermined acceptance criteria and met the appropriate sterilization standards.

Shelf life testing on the sterile accessories yielded successful test results for the latest pull point for the consumable accessories covered by this premarket notification.

Electromagnetic Compatibility (EMC )/ Wireless / Electrical Safety

Alcon assessed the EMC, wireless, and electrical / mechanical safety of the UNITY VCS and UNITY CS according to IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-6, IEC 60601-2-2, and IEC 60601-4-2, and the final FDA guidance documents "Electromagnetic Compatibility (EMC) of Medical Devices," issued June 6, 2022, and "Radio Frequency Wireless Technology in Medical Devices," issued on August 14, 2013. The subject devices met all requirements from the aforementioned standards and followed FDA recommendations where applicable from said guidance documents.

Software

Alcon developed and tested the UNITY VCS and UNITY CS device software functions according to ANSI / AAMI / IEC 62304, "Medical Device Software – Software Life Cycle Processes" and internal Alcon procedures that follow this standard. Premarket submission documentation was prepared in accordance with the final FDA guidance "Content of Premarket Submissions for Device Software Functions," issued on June 14, 2023. Testing met all requirements from IEC 62304, and all FDA recommendations for documentation were met.

Alcon also fulfilled FDA's cybersecurity recommendations for the subject devices in accordance

{16}------------------------------------------------

with the final FDA guidance "Cybersecurity in Medical Devices: Ouality System Considerations and Content of Premarket Submissions," dated September 2023.

Performance Testing

Alcon conducted performance tests on the UNITY VCS and UNITY CS devices to demonstrate safety, effectiveness, and substantial equivalence to predicate devices, in accordance with IEC 80601-2-58.

Safety tests of the UNITY VCS and UNITY CS demonstrated the compliance of the subject devices to the appropriate standards. Biocompatibility evaluations of materials coming into contact with the patient or patient fluid path demonstrated compliance with the appropriate standard from the ISO 10993 series or with ISO 16671:2015.

Sterile, single-use UNITY VCS and UNITY CS accessories have been shown to be sterilizable to a sterility assurance level (SAL) of 10-, and the sterilization process for these accessories has been validated per the required standard.

Technological characteristics affecting clinical performance are similar to those of the predicate devices previously listed. Like the predicate devices have been developed and manufactured in compliance with 21 CFR 820 and ISO 14971. Non-clinical testing noted above has demonstrated that the functional requirements have been met and that the subject devices are equivalent to the predicate devices.

7.2

The risk assessment for the subject devices did not require any animal or clinical testing as mitigation against any identified risks.

CONCLUSION 8.

The subject devices share the same intended use as the predicate devices. Their indications for use are equivalent with each other. The technological characteristics affecting clinical performance are similar to those of the predicate devices previously listed. Risk profiles are equivalent between the subject devices and the predicate devices. The predicate devices and the subject devices have been developed and manufactured in compliance with 21 CFR 820 and ISO 14971. Non-clinical testing noted above demonstrated that the functional requirements were met, and that the subject devices are equivalent to the predicate devices.

Therefore, the UNITY VCS and UNITY CS are deemed substantially equivalent to their predicate devices, the CONSTELLATION Vision System, Alcon Next-Generation Laser (for the UNITY VCS laser functionality), and the CENTURION Vision System, respectively.

§ 886.4670 Phacofragmentation system.

(a)
Identification. A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.(b)
Classification. Class II.