Intended Use

The Difiney Anti Snoring Device is intended to help reduce snoring in adults 18 years of age or older.

Device Description

The Difiney Anti Snoring device is an intraoral device that is composed of an upper and lower intraoral trays with a handle that is used during the process of customizing the trays. The Difiney Anti Shoring device repositions the jaw anteriorly relative to the maxilla in order to increase the patient's pharyngeal space to improve the ability to exchange air and decrease air turbulence, a causative factor in snoring.

AI/ML Overview

The provided text is a 510(k) summary for the Difiney Anti Snoring Device. It outlines the regulatory classification, intended use, and a comparison to a predicate device (SnoreRx). Crucially, the document states "Performance Testing - Clinical: Not Applicable" and "Performance Testing - Animal: Not Applicable."

This means the submission does not include data from a clinical study or animal study to demonstrate the device's performance based on specific acceptance criteria. Instead, the substantial equivalence determination is based on technological characteristics and non-clinical testing (biocompatibility and physical properties).

Therefore, I cannot provide the information requested in your prompt regarding acceptance criteria, device performance from a study, sample sizes, expert involvement, or MRMC studies, because this information is explicitly stated as "Not Applicable" in the provided document.

The document's conclusion is that the device is substantially equivalent to the predicate based on technological characteristics and non-clinical testing, not on clinical performance metrics.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA U.S. FOOD & DRUG" on the top line and "ADMINISTRATION" on the bottom line.

April 18, 2024

Fissiontech LLC % Libray Chang Manager Shanghai Spica Management Consulting Co., Ltd. 609 Room, No. 133 Shengang Avenue, Pudong New District Shanghai, 200120 China

Re: K233850

Trade/Device Name: Difiney Anti Snoring Device Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK Dated: April 8, 2024 Received: April 8, 2024

Dear Libray Chang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bobak Shirmohammadi -S

For Michael E. Adjodha, M.ChE., RAC, CQIA Assistant Director

{2}------------------------------------------------

DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K233850

Device Name Difiney Anti Snoring Device

Indications for Use (Describe)

The Difiney Anti Snoring Device is intended to help reduce snoring in adults 18 years of age or older.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) Summary K233850

Type of submissionTraditional
Date preparedApril 8, 2024
Submission sponsor
Manufacturer NameFISSIONTECH LLC
Address1 LIBERTY STREET, FLOOR 3, NEW YORK CITYNEW YORK 10006
Tel314 475 4711
Emailmike@everymarket.com
Contact PersonMike Lee
Device identification
Trade NameDifiney Anti Snoring Device
Regulation Number21 CFR 872.5570
Regulation NameIntraoral devices for snoring and intraoral devices forsnoring and obstructive sleep apnea
Device ClassificationClass II
Product CodeLRK
PanelDental
Application correspondent
Company NameShanghai Spica Management Consulting Co., Ltd.
Address609 Room, No.133 Shengang Avenue, Pudong NewDistrict, Shanghai, China
Tel86-13020102321

Page 1 of 6

Email

Libray@spicagloble.com

{5}------------------------------------------------

Summary

Contact Person

Libray Chang

Predicate device information

SponsorApnea Sciences Corporation
Trade/Device NameSnoreRx
510(K) numberK170285
Regulation Number21 CFR 872.5570

Indications for use

The Difiney Anti Snoring device is intended to help reduce snoring in adults 18 years of age or older.

Device description

The Difiney Anti Snoring device is an intraoral device that is composed of an upper and lower intraoral trays with a handle that is used during the process of customizing the trays. The Difiney Anti Shoring device repositions the jaw anteriorly relative to the maxilla in order to increase the patient's pharyngeal space to improve the ability to exchange air and decrease air turbulence, a causative factor in snoring.

Performance Testing - Clinical

Not Applicable.

Performance Testing - Animal

Not Applicable.

{6}------------------------------------------------

Technological Characteristic Comparison

Provided below is a comparison of the subject device with the predicate device.

Table 14A: General Comparison
ItemSubject DevicePredicate Device (K170285)Comparison
Product CodeLRKLRKSame
ClassClass IIClass IISame
ClassificationNameAnti-Snoring DeviceAnti-Snoring DeviceSame
ProprietaryNameDifiney Anti Snoring DeviceSnoreRx/
TechnologicalFeaturesMandibular repositioning to advance the mandiblerelative to the maxilla to increase pharyngeal space.Mandibular repositioning device that advancesthe lower jaw to increase pharyngeal space.Similar
Intended UseThe Difiney Anti Snoring device is intended tohelp reduce snoring in adults 18 years of age orolder.Minimize air turbulence that causes snoring.Device is intended for OverThe-Counter use.Similar

{7}------------------------------------------------

Materials- Polycarbonate resin- Ethylene vinyl acetate copolymer- Polycarbonate resin- Ethylene vinyl acetate copolymerSame
DesirableCharacteristicsHome use, heat sensitive / moldable, adjustablejaw advancement positionHome use, heat sensitive / moldable, adjustablejaw advancement positionSame
Specifications:- Physical:- Mechanical:- Single use:- Custom-fitted intraoral device- Repositions mandible anteriorly up to 4 mm- Reusable- Custom-fitted intraoral device- Repositions mandible anteriorly up to 6 mm- ReusableSimilar
SterilityNon-sterileNon-sterileSame
BiocompatibilityISO 10993ISO 10993Same

Summary

Summary of Non-Clinical Testing

Non-clinical tests were conducted to verify that the met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

Biocompatibility test

  • Bio-compatibility assessment was conducted in accordance with ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.
    Physical properties

  • ASTM D790-17 for flexural strength and flexural modulus.

{8}------------------------------------------------

  • ASTM D638-22 for Tensile Strength and Elongation at break.
  • ASTM D570-22 for water absorption.
  • ASTM D3418-21 for melting temperature.
  • ASTM D792-20 for Density.

{9}------------------------------------------------

Conclusion

Based on the comparison of the indications for use, technical characteristics and performance, we find that the Difiney Anti Snoring Device is substantially equivalent to the predicate SnoreRx with 510(K) number K170285.

§ 872.5570 Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea.

(a)
Identification. Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.(b)
Classification. Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”