K Number
K233772

Validate with FDA (Live)

Date Cleared
2024-03-19

(116 days)

Product Code
Regulation Number
876.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Dual Action Tissue Closure Device is intended for use in flexible endoscopy for the compression of tissue in the gastrointestinal tract for adult patient only.

The Disposable Dual Action Tissue Closure Device is indicated for clip placement within the gastrointestinal tract for the purpose of:

  1. Endoscopic marking:

  2. Hemostasis for:

a.Mucosal/sub-mucosal defects < 5 cm,

b.Bleeding ulcers,

c.Polyps < 1.5 cm in diameter,

d.Diverticula in the colon,

e. Arteries < 2 mm,

  1. As a supplementary method. closure of GI tract luminal perforations < 20 mm that can be treated conservatively.
Device Description

The proposed device Disposable Dual Action Tissue Closure Device is a sterile, single-use endoscopic clipping device, intended to be used in flexible endoscopy for the compression, manipulation of tissue in the gastrointestinal tract. It consists of two main components, delivery system and clip assembly. The proposed devices are EO sterilized to achieve the Sterility Assurance Level (SAL) of 106 and placed in a sterility maintenance package to ensure a shelf life of 3 years.

AI/ML Overview

This document outlines the acceptance criteria and supporting studies for the Micro-Tech (Nanjing) Co., Ltd. Disposable Dual Action Tissue Closure Device (K233772).

1. Table of Acceptance Criteria and Reported Device Performance

The provided document describes the device as substantially equivalent to a predicate device (K212748) except for one indication and shelf-life, and therefore, directly stated acceptance criteria are not explicitly presented as separate quantitative values for each performance test. Instead, the document relies on demonstrating that the proposed device performs the same or better than the predicate device for all shared characteristics, and that the changes in indications and shelf-life are supported by additional testing or literature.

Below is a table summarizing the performance data presented, emphasizing the comparison to the predicate device.

Test Item / CharacteristicAcceptance Criteria (Implied by Predicate Equivalence or Specific Testing)Reported Device Performance (Proposed Device K233772)
Form Factor / DimensionsSame as predicate (K212748)Same as predicate (K212748), including: - Open width (15 mm) - Working Length (1650, 1950, 2350 mm) - Minimal working channel of endoscopy (3.2 mm)
Material / ConfigurationSame as predicate (K212748)Same as predicate (K212748)
Principles of OperationSame as predicate (K212748)Same as predicate (K212748)
Endoscopic Marking IndicationSame as predicate (K212748)Same as predicate (K212748)
Hemostasis for Bleeding UlcersSame as predicate (K212748)Same as predicate (K212748)
Hemostasis for Polyps (< 1.5 cm)Same as predicate (K212748)Same as predicate (K212748)
Hemostasis for Diverticula in the ColonSame as predicate (K212748)Same as predicate (K212748)
Hemostasis for Arteries (< 2 mm)Same as predicate (K212748)Same as predicate (K212748)
GI Tract Luminal Perforation Closure (< 20 mm)Same as predicate (K212748)Same as predicate (K212748)
Hemostasis for Mucosal/Sub-mucosal Defects< 5 cm (Expanded from predicate's < 3 cm)Supported by clinical literature to include up to 5 cm.
Device Performance (Mechanical, Functional)Same performance specifications as predicate (K212748)- Dimension: Same as predicate - Clip Assembly Repeated Open/Close: Same performance as predicate. - Scope Compatibility/Usability: Same performance as predicate. - Endoscope Damage: Same performance as predicate. - Biopsy Valve compatibility: Same performance as predicate. - Clip approach: Same performance as predicate. - Clip Open and Close Force: Same performance as predicate. - Release Force: Same performance as predicate. - Tensile Strength: Same performance as predicate. - Clamping Strength: Same performance as predicate. - Coil to handle Tensile: Same performance as predicate. - Mechanical Integrity of Clip Assembly: Same performance as predicate.
Shelf Life3 years (Increased from predicate's 1 year)3-years accelerated aging tests performed using the same test method and acceptance criteria as predicate device.
Sterilization (Method, SAL)EO Sterilized, SAL: 10^-6 (Same as predicate)EO Sterilized, SAL: 10^-6 (Same as predicate)
BiocompatibilityConform to ISO 10993-1 (Same as predicate)Conform to ISO 10993-1. Biocompatibility tests (Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Material Mediated Pyrogenicity, Chemical Characterization) are the same as predicate.
MRI CompatibilityComply with ASTM F 2503, ASTM F 2052, ASTM F2119, ASTM F2182, ASTM F2213 (Same as predicate)Comply with ASTM F 2503, ASTM F 2052, ASTM F2119, ASTM F2182, ASTM F2213 (Same as predicate).

2. Sample Size Used for the Test Set and Data Provenance

  • Mechanical and Functional Performance Tests (Dimension, Open/Close, Compatibility, Forces, Strength, etc.): The document states that "The bench performance testing that was previously conducted on the predicate device and not re-conducted on the proposed device." This implies that the sample sizes for these tests were those used for the K212748 predicate device. The specific sample sizes for these tests are not provided in this summary. The data provenance would be from the original studies for K212748, likely prospective bench testing. The country of origin of this data is not specified, but given the manufacturer's location (China), it's highly probable the testing was conducted there or by affiliated labs.
  • Shelf-life Testing: "Three-years accelerated aging tests were performed." The sample size for these accelerated aging tests is not specified. This would be prospective testing. The country of origin is not specified.
  • Sterilization Validation: "The sterilization testing that was previously conducted on the predicate device and not re-conducted on the proposed device." The sample size for this is not provided but would refer to the tests for K212748.
  • Biocompatibility Testing: "The biocompatibility performance testing that was previously conducted on the predicate device and not re-conducted on the proposed device." The sample size for this is not provided but would refer to the tests for K212748.
  • MRI Compatibility Testing: "The MR compatibility performance testing that was previously conducted on the predicate device and not re-conducted on the proposed device." The sample size for this is not provided but would refer to the tests for K212748.
  • Indication Change (Mucosal/sub-mucosal defects < 5 cm): "clinical literature was used to support changing the IFU." This suggests the test set for this particular change relied on retrospective data from published clinical studies, not new testing by the manufacturer. The specific sample sizes and provenance of these literature studies are not detailed in this summary.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • For all mechanical, functional, and material tests (including shelf-life, sterilization, biocompatibility, MRI compatibility), the "ground truth" is established by adherence to engineering specifications, regulatory standards (e.g., ISO, ASTM), and comparison to the predicate device's established performance. There is no mention of external human experts establishing ground truth for these technical tests in the sense of clinical interpretation.
  • For the indication change regarding mucosal/sub-mucosal defects, the "ground truth" was established by clinical literature. The document does not specify the number of experts in these literature studies or their qualifications, as it references existing publications rather than new expert review panels for this submission.

4. Adjudication Method for the Test Set

  • For the technical and performance testing (mechanical, functional, material, shelf-life, sterilization, MRI compatibility), no specific adjudication method is mentioned. Adherence to established test protocols and comparison to predicate device performance would be the primary method.
  • For supporting the indication change with clinical literature, no adjudication method is described for the selection or interpretation of the literature. It's implicitly assumed that the referenced literature itself has undergone peer review and scientific scrutiny.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The submission explicitly states "No animal study is included in this submission. No clinical study is included in this submission."

Therefore, there is no reported effect size of how much human readers improve with AI vs. without AI assistance, as this device is a physical medical instrument, not an AI-assisted diagnostic tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This question is not applicable as the device is a physical tissue closure device and does not involve an algorithm or human-in-the-loop performance in the sense of AI-driven medical imaging or diagnostics. The performance described is purely mechanical and functional.

7. The Type of Ground Truth Used

The ground truth used depends on the aspect being evaluated:

  • Mechanical, Functional, Material, Sterilization, MRI Compatibility Performance: Established through objective physical measurements against specifications and regulatory standards (e.g., ISO 11135, ISO 10993-1, ASTM standards) and direct comparison to the predicate device's previously established performance.
  • Shelf-life: Established through accelerated aging test data (ASTM F1980-16).
  • Expanded Indication (Mucosal/Sub-mucosal defects < 5 cm): Established through clinical literature.

8. The Sample Size for the Training Set

This question is not applicable. The device is a physical medical device, not an AI/ML algorithm that requires a "training set" of data.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA seal. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

March 19, 2024

Micro-Tech (Nanjing) Co., Ltd. Sally He Regional RA Manager No.10 Gaoke Third Road Nanjing National Hi-Tech, Industrial Development Zone Nanjing, Jiangsu 210032 China

Re: K233772

Trade/Device Name: Disposable Dual Action Tissue Closure Device Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal Ligator Regulatory Class: Class II Product Code: PKL Dated: January 19, 2024 Received: January 19, 2024

Dear Sally He:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Shanil P. Haugen -S

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K233772

Device Name

Disposable Dual Action Tissue Closure Device

Indications for Use (Describe)

The Disposable Dual Action Tissue Closure Device is intended for use in flexible endoscopy for the compression of tissue in the gastrointestinal tract for adult patient only.

The Disposable Dual Action Tissue Closure Device is indicated for clip placement within the gastrointestinal tract for the purpose of:

  1. Endoscopic marking:

  2. Hemostasis for:

a.Mucosal/sub-mucosal defects < 5 cm,

b.Bleeding ulcers,

c.Polyps < 1.5 cm in diameter,

d.Diverticula in the colon,

e. Arteries < 2 mm,

  1. As a supplementary method. closure of GI tract luminal perforations < 20 mm that can be treated conservatively.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are connected and appear to be made of solid blocks. A registered trademark symbol is located in the upper right corner of the logo.

510(k) summary

510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: K233772

1. Date of Preparation: 2024-03-14

2. Sponsor Identification

Micro-Tech (Nanjing) Co., Ltd.

No.10 Gaoke Third Road, Nanjing National Hi-Tech, Industrial Development Zone, Nanjing, Jiangsu

Province, PRC

Establishment Registration Number: 3004837686

Contact Person: Sally He

Position: Regional RA Manager

Tel: +86-25-58646395

Fax: +86-25-58350006

Email: ra.micro-tech@outlook.com

3. Identification of Proposed Device

Trade Name: Disposable Dual Action Tissue Closure Device

Common Name: Hemostasis Clip

Regulatory Information

Classification Name: Hemostatic Metal Clip For The GI Tract

Classification: 2

Product Code: PKL

Regulation Number: 876.4400

Review Panel: Gastroenterology/Urology

{5}------------------------------------------------

K233772 Page 2 of 7

510(k) summary

Image /page/5/Picture/2 description: The image shows a blue logo with the letters 'MT' in a stylized, three-dimensional font. The letters are connected and appear to be angled slightly to the right, giving the logo a sense of depth. A registered trademark symbol is located in the upper right corner of the logo.

Identification of Predicate Device 4.

510(k) Number: K212748

Product Name: Disposable Dual Action Tissue Clip

Manufacturer: Micro-Tech (Nanjing) Co., Ltd.

5. Indications for Use

The Disposable Dual Action Tissue Closure Device is intended for use in flexible endoscopy for the compression of tissue in the gastrointestinal tract for adult patient only.

The Disposable Dual Action Tissue Closure Device is indicated for clip placement within the gastrointestinal tract for the purpose of:

  • Endoscopic marking,
  • Hemostasis for
    • O Mucosal/sub-mucosal defects < 5 cm,
    • O Bleeding ulcers.
    • O Polyps < 1.5 cm in diameter,
    • O Diverticula in the colon,
    • O Arteries < 2 mm,
  • As a supplementary method, closure of GI tract luminal perforations < 20 mm that can be treated conservatively.

6. Device Description

The proposed device Disposable Dual Action Tissue Closure Device is a sterile, single-use endoscopic clipping device, intended to be used in flexible endoscopy for the compression, manipulation of tissue in the gastrointestinal tract. It consists of two main components, delivery system and clip assembly. The proposed devices are EO sterilized to achieve the Sterility Assurance Level (SAL) of 106 and placed in a sterility maintenance package to ensure a shelf life of 3 years.

7. Comparison of Technological Characteristics

The Disposable Dual Action Tissue Closure Device incorporates substantially equivalent device's

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are interconnected, with the "M" on the left and the "T" on the right. A registered trademark symbol is located in the upper right corner of the logo. The logo has a modern and corporate appearance.

materials, design, configuration, packaging fundamental technology, manufacturing processes including

sterilization process as those featured in the predicate device Disposable Dual Action Tissue Clip

cleared under K212748.
--------------------------
Proposed DevicePredicate Device
ITEMDisposable Dual Action TissueClosure DeviceDisposable Dual Action TissueClip (K212748)Remark
Product CodePKLPKLSame
Regulation No.876.4400876.4400Same
Class22Same
Indications for UseThe Disposable Dual ActionTissue Closure Device is intendedfor use in flexible endoscopy forthe compression of tissue in thegastrointestinal tract for adultpatient only.The Disposable Dual ActionTissue Closure Device is indicatedfor clip placement within thegastrointestinal tract for thepurpose of:• Endoscopic marking,• Hemostasis for○Mucosal/sub-mucosal defects < 5 cm,○Bleeding ulcers,○Polyps < 1.5 cm in diameter,○Diverticula in the colon,○Arteries < 2 mm,• As a supplementary method,closure of GI tract luminalperforations < 20 mm that can betreated conservatively.The Disposable Dual Action TissueClosure Device is intended for usein flexible endoscopy for thecompression of tissue in thegastrointestinal tract for adultpatient only.The Disposable Dual Action TissueClosure Device is indicated for clipplacement within thegastrointestinal tract for thepurpose of:• Endoscopic marking,• Hemostasis for○Mucosal/sub-mucosal defects < 3 cm,○Bleeding ulcers,○Polyps < 1.5 cm in diameter,○Diverticula in the colon,○Arteries < 2 mm,• As a supplementary method,closure of GI tract luminalperforations < 20 mm that can betreated conservatively.Oneindicationis changed
Configuration2 delivery units of the deliverysystem and 2 separate clips of theclip assembly2 delivery units of the deliverysystem and 2 separate clips of theclip assemblySame
Principles ofOperationPull the steel wire through theoperating handle to drive theopening and closing of the clip toclamp the tissue. It is a metallicPull the steel wire through theoperating handle to drive theopening and closing of the clip toclamp the tissue. It is a metallicSame
510(k) summary
Proposed DevicePredicate Device
ITEMDisposable Dual Action TissueClosure Devicedevice with a clasping mechanism.Disposable Dual Action TissueClip (K212748)device with a clasping mechanism.Remark
Open width (mm)1515Same
Working Length(mm)1650,1950,23501650,1950,2350Same
Minimal workingchannel ofendoscopy (mm)3.23.2Same
Supplied in SterileYesYesSame
Single UseYesYesSame
PackagingSingle-use EO sterilized pouchwith one device per pouchSingle-use EO sterilized pouchwith one device per pouchSame
Shelf LifeThree yearsOne yearDifferent
BiocompatibilityConform to ISO 10993-1Conform to ISO 10993-1Same
SterilizationEO Sterilized, SAL:10-6EO Sterilized, SAL:10-6Same
LabelingConform to 21 CFR part 801Conform to 21 CFR part 801Same
MRI informationComply with ASTM F 2503,ASTM F 2052, ASTM F2119,ASTM F2182, ASTM F2213Comply with ASTM F 2503,ASTM F 2052, ASTM F2119,ASTM F2182, ASTM F2213Same

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows a logo with the letters "MT" in a stylized, three-dimensional font. The letters are predominantly blue, with the "M" appearing to the left and slightly behind the "T", creating a sense of depth. A registered trademark symbol is visible in the upper right corner of the logo.

8. Performance Data

The performances of Disposable Dual Action Tissue Closure Device are as follows:

  • Dimension
  • . Clip Assembly Repeated Open/Close;
  • Scope Compatibility/Usability;
  • Endoscope Damage;
  • Biopsy Valve compatibility;

{8}------------------------------------------------

K233772 Page 5 of 7

510(k) summary

Image /page/8/Picture/2 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional design. The letters are bold and appear to be interconnected, with the "M" on the left and the "T" on the right. A small "R" in a circle, indicating a registered trademark, is positioned in the upper right corner of the logo.

  • Clip approach; O
  • Clip Open and Close Force ●
  • Release Force;
  • Tensile Strength:
  • Clamping Strength;
  • Coil to handle Tensile.
  • Mechanical Integrity of Clip Assembly

Compared with the predicate device Disposable Dual Action Tissue Clip cleared under K212748, the proposed devices have the same design, dimension, configuration, material, principles of operation, control mechanism as the predicate device, the performances and performance specification are also the same as the predicate device, no change. The performance of proposed device is substantially equivalent to the predicate devices Disposable Dual Action Tissue Clip (K212748).

The bench performance testing that was previously conducted on the predicate device and not reconducted on the proposed device.

Shelf-life testing was conducted based on an accelerated aging test in accordance with ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices. Compared with the predicate device, the shelf life was changed to three years. Three-years accelerated aging tests were performed using the same test method and acceptance criteria as predicate device to demonstrate the stability and support the results of the accelerated aging test.

Sterilization validation was carried out in accordance with ISO 11135:2014 "Sterilization of Health Care products - Ethylene Oxide - Part 1: Requirements for Development, Validation, and Routine Control of Sterilization processes for Medical Devices". Compared with the predicate device, the sterilization method is not changed. The sterilization testing that was previously conducted on the predicate device and not re-conducted on the proposed device.

Compared with the predicate device, the material of proposed device is not changed, therefore, the biocompatibility of proposed device is also the same. The tests that were submitted for the predicate

{9}------------------------------------------------

510(k) summary

Image /page/9/Picture/2 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional design. The letters are bold and appear to be interconnected, creating a sense of depth and solidity. A small "R" in a circle, indicating a registered trademark, is positioned in the upper right corner of the logo.

device are as followed:

  • Cytotoxicity a)
  • b) Sensitization
  • Intracutaneous Reactivity c)
  • d) Acute Systemic Toxicity
  • Material Mediated Pyrogenicity e)
  • f) Chemical Characterization

The biocompatibility performance testing that was previously conducted on the predicate device and not re-conducted on the proposed device.

MR compatibility was evaluated in accordance with ASTM F 2503, ASTM F 2052, ASTM F2119, ASTM F2182, ASTM F2213 and FDA guidance on Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment issued on May 20, 2021. Compared with the predicate device, the material and MR environment of proposed device are not changed. The MR compatibility performance testing that was previously conducted on the predicate device and not re-conducted on the proposed device.

For the indication change, clinical literature was used to support changing the IFU to include hemostasis for mucosal/sub-mucosal defects up to 5 cm.

For the shelf life change, shelf-life testing was conducted to support a longer shelf-life of 3 years.

9. Animal Study

No animal study is included in this submission.

10. Clinical Study

No clinical study is included in this submission.

11. Substantially Equivalent (SE) Conclusion

Based on the indications for use, technological characteristics, safety and performance testing,

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are interconnected, with the "M" on the left and the "T" on the right. A registered trademark symbol is located in the upper right corner of the logo. The logo appears to be for a company or organization with the initials "MT".

510(k) summary

Disposable Dual Action Tissue Closure Device has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the currently cleared predicate device Disposable Dual Action Tissue Clip cleared under K212748.

§ 876.4400 Hemorrhoidal ligator.

(a)
Identification. A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.(b)
Classification. Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.