K Number
K233762

Validate with FDA (Live)

Date Cleared
2024-06-27

(216 days)

Product Code
Regulation Number
880.6860
Age Range
2 - 65
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SteroScope® Sterilization Technology System is indicated for the terminal sterilization of cleaned reusable Flexible Endoscopes with up to 8 internal lumens with lumen dimensions of:

  • ID of 1.0 mm or larger and a length of 3580 mm or shorter and .
  • ID of 1.2 mm or larger and a length of 4095 mm or shorter .
Device Description

The SteroScope® Sterilization Technology System is a vaporized hydrogen peroxide based (VHP) sterilizer for the terminal sterilization of flexible endoscopes with the following characteristics: Flexible Endoscopes with up to 8 internal lumens with lumen dimensions of: · ID of 1.0 mm or larger and a length of 3580 mm or shorter and · ID of 1.2 mm or larger and a length of 4095 mm or shorter. The SteroScope® Sterilization Technology System is a Gas Plasma VHP sterilization system designed specifically for flexible endoscope terminal sterilization, not for general surgical instrument loads. It is a self-contained stand-alone system of hardware and software system that uses technology to rapidly diffuse VHP into all the channels of the flexible endoscope to facilitate the sterilization of inner device surfaces, while being occupationally and environmentally safe to use. To enable the terminal sterilization of long multi-lumen endoscopes, while minimizing oxidation damage, the SteroScope® Sterilization Technology System incorporates a proprietary gas diffusion technology which creates a pressure differential in each internal endoscope lumen channel to actively pull hydrogen peroxide vapor from the chamber of the VHP sterilizer through the distal end of each endoscope channel, while the endoscope is in a sterile container. Utilizing a separate mechanism for the rapid diffusion of VHP from the chamber into the individual SteroScope® Sterilization Technology System allows the entire endoscope to be sterilized with less overall peroxide dose exposure.

AI/ML Overview

The provided document is a 510(k) summary for the SteroScope® Sterilization Technology System. It outlines the non-clinical performance testing conducted to establish the device's safety and efficacy. Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document provides a table (Table 3) listing various testing methods, their purpose, acceptance criteria, and results. All tests listed show a "Pass" result, indicating the device met the specified acceptance criteria.

Testing MethodPurposeAcceptance CriteriaResults
System Environmental TestingSystem environmental testing under extreme conditions operating conditionsInternal SpecificationPass
SteroScope H2O2 Emissions TestH2O2 emissions compliance to required OSHA safety levelsOSHA H2O2 SEL/PELPass
System H2O2 Concentration MappingPeroxide concentration mapping during operationInternal SpecificationPass
System Temperature MappingTemperature mapping during operationInternal SpecificationPass
System Functional TestTesting of system functional requirements to ensure conformance to system specificationsInternal SpecificationPass
SteroScope System ShippingSystem shipping test for vibration, shock & drop to ensure system and packaging integrityISTA 2BPass
SteroScope Consumable ShippingConsumable shipping test for vibration, shock & drop to ensure consumable and packaging integrityISTA 2APass
EMC & Electrical SafetyThe sterilizer was tested in accordance with IEC 60601-1-2 and IEC 61010-1 standardsIEC 60601-1-2 and IEC 61010-1Pass
Qualification of Pressure LimitsValidation of sterilizer operational pressure limitsInternal SpecificationPass
Human FactorsValidation of system ergonomics and human factors with clinical usersInternal SpecificationPass
Software ValidationSoftware verification and validation of software requirementsInternal SpecificationPass
Subsystem Design VerificationsTesting of each subsystem to ensure functional performance met design requirementsInternal SpecificationPass
H2O2 Consumable Shelf-lifeLong term stability study of H2O2 consumable to establish shelf lifeInternal SpecificationPass
H2O2 Consumable StorageStability testing of H2O2 consumable in extreme conditionsInternal SpecificationPass
Endoscope Temperature MappingMapping of endoscope temperatures during cycle to verify the internal and external temperatures of a flexible endoscopeInternal SpecificationPass
Final Process QualificationOperational process qualification with multiple systemsInternal SpecificationPass
Mated Surface Connector-PortHalf cycle connector efficacy testNo growth of Geobacillus stearothermophilusPass
BI & CI FunctionalityBI & CI functionality under worst case conditionsNo Growth in BI & color change in CIPass
AOAC Sporicidal TestProcess and sterilant efficacy for multiple organismsAOAC Official Method 966.04Pass
Mated Surface SterilizationHalf cycle mated material efficacy using material couponsNo growth of Geobacillus stearothermophilusPass
Material Surface SterilizationHalf cycle surface efficacy using material couponsNo growth of Geobacillus stearothermophilusPass
Uniformity Microbial InactivationDosing study to assess uniformity of microbial inactivationUniformity in inactivation kinetics based on SLRPass
Endoscope Dose Response CurvesDosing study with various flexible endoscopesInternal SpecificationPass
Material Bacteriostasis ValidationConfirmation that there are no bacteriostatic effects that inhibit growth for material couponsNo bacteriostatic effect after sterilization with growth of defined micro-organismsPass
Bacteriostasis & Fungistasis ValidationConfirmation that there are no bacteriostatic effects that inhibit growth for flexible endoscopesNo bacteriostatic effect after sterilization with growth of defined micro-organismsPass
Simulated Use TestingWorst-case testing with unwashed flexible endoscopes inoculated with test culture challenge6-log reduction in Geobacillus stearothermophilusPass
In-Use TestingWorst-case testing with clinically used flexible endoscopes.No growth upon recovery after sterilizationPass
Functionality & Material CompatibilityWorst-case material compatibility testing of multiple flexible endoscopes with back to back cyclesInternal SpecificationPass
Biocompatibility CouponsWorst-case residual and cytotoxicity (MTT) testing with material couponsISO 10993-5Pass
Biocompatibility Flexible EndoscopesWorst-case residual and cytotoxicity (MTT) testing with flexible endoscopesISO 10993-5Pass

2. Sample sized used for the test set and the data provenance:

The document describes non-clinical performance testing. It doesn't explicitly state sample sizes for each test in a numerical format that would typically be used for a clinical study (e.g., N=X patients). However, it mentions "multiple systems" for "Final Process Qualification," "multiple organisms" for "AOAC Sporicidal Test," "various flexible endoscopes" for "Endoscope Dose Response Curves," and "multiple flexible endoscopes" for "Functionality & Material Compatibility."

The data provenance is from non-clinical laboratory testing of the device, not from human subjects. The country of origin for the data is not specified, nor is whether the data is retrospective or prospective. Given it's non-clinical testing for a new device, it is inherently prospective data generation.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided in the document. The tests performed are primarily performance-based tests against established standards and internal specifications, not expert-driven diagnostic interpretations. For example, for sterilization efficacy, the ground truth is "no growth" of specified microorganisms or a "6-log reduction," which are objective laboratory measurements, not expert consensus.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

This information is not applicable as the tests are non-clinical performance evaluations, not human reader-based assessments requiring adjudication of interpretations.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable as this is a non-clinical 510(k) submission for a sterilization system, not a diagnostic imaging device utilizing AI that would involve human reader performance studies.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

The device is a physical sterilization system, not an algorithm or AI. Therefore, standalone algorithm performance is not applicable. The tests performed are essentially "standalone" device performance evaluations in a controlled environment.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The ground truth for the non-clinical performance testing is based on established scientific and engineering principles, industry standards (e.g., ISO, IEC, AAMI, ASTM, AOAC, USP-NF, ISTA, OSHA), and internal specifications. For sterilization efficacy tests, the ground truth is a specific microbial reduction (e.g., 6-log reduction) or absence of microbial growth, which is a definitively measurable biological outcome. For other tests, it's adherence to specified physical, electrical, or chemical parameters.

8. The sample size for the training set:

This information is not applicable. The device is a physical sterilization system, not a machine learning model that requires a "training set" in the context of AI/ML.

9. How the ground truth for the training set was established:
This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

June 27, 2024

IDEATE Medical, Inc. % Kevin Corrigan Regulaory Consultant Corrigan Regulatory Consulting P.O. Box 6284 Anaheim, California 92816

Re: K233762

Trade/Device Name: SteroScope® Sterilization Technology System Regulation Number: 21 CFR 880.6860 Regulation Name: Ethylene Oxide Gas Sterilizer Regulatory Class: Class II Product Code: MLR Dated: November 22, 2023 Received: November 24, 2023

Dear Kevin Corrigan:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Christopher K. Dugard -2

Christopher K. Dugard, M.S. Assistant Director DHT4B: Division of Infection Control

{2}------------------------------------------------

and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K233762

Device Name

SteroScope® Sterilization Technology System

Indications for Use (Describe)

The SteroScope® Sterilization Technology System is indicated for the terminal sterilization of cleaned reusable

  • Flexible Endoscopes with up to 8 internal lumens with lumen dimensions of:
  • ID of 1.0 mm or larger and a length of 3580 mm or shorter and ●
  • . ID of 1.2 mm or larger and a length of 4095 mm or shorter
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY K233762

26 June 2024

I. SUBMITTER

IDEATE Medical, Inc. 2821 N. Ballas Road C55 St. Louis, MO 63131

Phone: 760-402-8322

Contact Person: William Wong, CEO bill.wong@ideatemedical.com

II. DEVICE

Name of Device:SteroScope® Sterilization Technology System
Common Name:Vaporized Hydrogen Peroxide Sterilizer
Classification Name:Ethylene oxide gas sterilizer (21 CFR 888.6860)
Regulatory:Class II
Product Code:MLR, Sterilizer, Chemical

III. PREDICATE DEVICE

STERRAD 100NX DUO Cycle (K111377)

Reference Device: STERRAD 100NX EXPRESS Cycle (K092622)

IV. DEVICE DESCRIPTION

The SteroScope® Sterilization Technology System is a vaporized hydrogen peroxide based (VHP) sterilizer for the terminal sterilization of flexible endoscopes with the following characteristics:

Flexible Endoscopes with up to 8 internal lumens with lumen dimensions of:

  • · ID of 1.0 mm or larger and a length of 3580 mm or shorter and
  • · ID of 1.2 mm or larger and a length of 4095 mm or shorter.

The SteroScope® Sterilization Technology System is a Gas Plasma VHP sterilization system designed specifically for flexible endoscope terminal sterilization, not for general surgical instrument loads. It is a self-contained stand-alone system of hardware and software system that uses technology to rapidly diffuse VHP into all the channels of the flexible endoscope to facilitate the sterilization of inner device surfaces, while being occupationally and environmentally safe to use. To enable the terminal sterilization of long multi-lumen endoscopes, while minimizing oxidation damage, the SteroScope® Sterilization Technology System incorporates a proprietary gas diffusion technology which creates a pressure differential

{5}------------------------------------------------

in each internal endoscope lumen channel to actively pull hydrogen peroxide vapor from the chamber of the VHP sterilizer through the distal end of each endoscope channel, while the endoscope is in a sterile container. Utilizing a separate mechanism for the rapid diffusion of VHP from the chamber into the individual SteroScope® Sterilization Technology System allows the entire endoscope to be sterilized with less overall peroxide dose exposure.

V. INDICATIONS FOR USE

The SteroScope® Sterilization Technology System is indicated for the terminal sterilization of cleaned reusable Flexible Endoscopes with up to 8 internal lumens with lumen dimensions of:

  • ID of 1.0 mm or larger and a length of 3580 mm or shorter and .
  • ID of 1.2 mm or larger and a length of 4095 mm or shorter .

VI. TECHNOLOGICAL CHARACTERISTICS COMPARISON

Table 1 (below) provides a summary of the device's technological characteristics comparing the subject device and the predicate device and reference device(s).

CharacteristicSubject DevicePredicate DeviceReference DeviceComments
STEROSCOPESYSTEM (K233762)STERRAD100NX DUO(K111377)STERRAD 100NXEXPRESS (K092622)
Intend UseSteroScope Sterilizer isintended for the terminalsterilization of properlyprepared (cleaned,rinsed, and dried) flexibleendoscopes in HealthcareFacilities. Thepreprogrammedsterilization cycleoperates at low pressureand temperature, suitablefor processingendoscopes withoutleaving toxic residues.The STERRAD Sterilizersare designed forsterilization of both metaland nonmetal medicaldevices at lowtemperatures. Because thecycle operates within a dryenvironment and at lowtemperatures, it isespecially suitable forinstruments sensitive toheat and moisture.The STERRAD Sterilizersare designed for sterilizationof both metal and nonmetalmedical devices at lowtemperatures. Because thecycle operates within a dryenvironment and at lowtemperatures, it is especiallysuitable for instrumentssensitive to heat andmoisture.SIMILARSubject topredicate &reference
Indications forUseThe SteroScopeSterilization System isindicated for the terminalsterilization of cleanedreusable flexibleendoscopes with up to 8internal lumens withlumen dimensions of:The STERRAD®100NXTM DUOCycle is an additionaloptional cycle designed forsterilization of medicaldevices including mostflexible endoscopes, withup to 1 internal lumen withlumen dimension of:1.0 mm or larger and alength of 875mm or shorterThe STERRAD® 100NXTMEXPRESS Cycle is anadditional optional cycledesigned for surfacesterilization of both metaland nonmetal medicaldevices at low temperatures.It can sterilize instrumentsurfaces and it can sterilizerigid and semi-rigidendoscopes without lumensSIMILARSubject topredicate &reference

Table 2: TECHNOLOGY Comparison

{6}------------------------------------------------

CharacteristicSubject DeviceSTEROSCOPESYSTEM (K233762)ID of 1.0 mm or largerand a length of 3580 mmor shorter andID of 1.2 mm or largerand a length of 4095 mmor shorterPredicate DeviceSTERRAD100NX(K111377)Reference DeviceSTERRAD 100NXEXPRESS (K092622)Comments
SterilizationTechnologyVHP Gas PlasmaVHP Gas PlasmaVHP Gas PlasmaSAME
Sterilant59% H2O259% H2O259% H2O2SAME
H2O2/cycle5.8 mL3.1 mL10.8 mLSIMILARSubject topredicate &reference
ChamberTemperature57° C57° C57° CSAME
ChamberVolume91.0 Liters93.4 Liters93.4 LitersSIMILARSubject topredicate &reference
Load Type(1) flexible endoscope(8) internal lumens(1.0mm ID x 3585mm)(2) flexible endoscopes(1) internal lumen(1.0mm ID x 850mm)(2) Semi-rigid/rigidendoscopes(no lumens up to 500mm)SIMILARSubject topredicate
Load Weight17.8 lbs.13.2 lbs.10.7 lbs.SIMILARSubject topredicate
H2O2 DeliveryExtracted from H2O2consumable and meteredExtracted from H2O2consumable and meteredExtracted from H2O2consumable with no meteringSAME
Cycle Time43 mins.51 mins.26 minsSIMILARSubject topredicate

VII. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING

Table 3 (below) provides a summary of the non-clinical performance testing conducted on the subject device, SteroScope Sterilization System, to establish safety and efficacy. Table 4 below provides a list of standards used in the testing of the SteroScope Sterilization System

Table 3:
Testing MethodPurposeAcceptance CriteriaResults
System EnvironmentalTestingSystem environmental testing underextreme conditions operatingconditionsInternalSpecificationPass
SteroScope H2O2Emissions TestH2O2 emissions compliance torequired OSHA safety levelsOSHA H2O2SEL/PELPass
Testing MethodPurposeAcceptance CriteriaResults
System H2O2Concentration MappingPeroxide concentration mapping during operationInternalSpecificationPass
System TemperatureMappingTemperature mapping during operationInternalSpecificationPass
System Functional TestTesting of system functional requirements to ensure conformance to system specificationsInternalSpecificationPass
SteroScope SystemShippingSystem shipping test for vibration, shock & drop to ensure system and packaging integrityISTA 2BPass
SteroScope ConsumableShippingConsumable shipping test for vibration, shock & drop to ensure consumable and packaging integrityISTA 2APass
EMC & Electrical SafetyThe sterilizer was tested in accordance with IEC 60601-1-2 and IEC 61010-1 standardsIEC 60601-1-2 andIEC 61010-1Pass
Qualification of PressureLimitsValidation of sterilizer operational pressure limitsInternalSpecificationPass
Human FactorsValidation of system ergonomics and human factors with clinical usersInternalSpecificationPass
Software ValidationSoftware verification and validation of software requirementsInternalSpecificationPass
Subsystem DesignVerificationsTesting of each subsystem to ensure functional performance met design requirementsInternalSpecificationPass
H2O2 Consumable Shelf-lifeLong term stability study of H2O2 consumable to establish shelf lifeInternalSpecificationPass
H2O2 ConsumableStorageStability testing of H2O2 consumable in extreme conditionsInternalSpecificationPass
Endoscope TemperatureMappingMapping of endoscope temperatures during cycle to verify the internal and external temperatures of a flexible endoscopeInternalSpecificationPass
Final ProcessQualificationOperational process qualification with multiple systemsInternalSpecificationPass
Mated Surface Connector-PortHalf cycle connector efficacy testNo growth ofGeobacillusstearothermophilusPass
BI & CI FunctionalityBI & CI functionality under worst case conditionsNo Growth in BI &color change in CIPass
AOAC Sporicidal TestProcess and sterilant efficacy for multiple organismsAOAC OfficialMethod 966.04Pass
Mated SurfaceSterilizationHalf cycle mated material efficacy using material couponsNo growth ofGeobacillusstearothermophilusPass
Testing MethodPurposeAcceptance CriteriaResults
Material SurfaceSterilizationHalf cycle surface efficacy usingmaterial couponsNo growth ofGeobacillusstearothermophilusPass
Uniformity MicrobialInactivationDosing study to assess uniformity ofmicrobial inactivationUniformity ininactivation kineticsbased on SLRPass
Endoscope DoseResponse CurvesDosing study with various flexibleendoscopesInternalSpecificationPass
Material BacteriostasisValidationConfirmation that there are nobacteriostatic effects that inhibitgrowth for material couponsNo bacteriostaticeffect aftersterilization withgrowth of definedmicro-organismsPass
Bacteriostasis &Fungistasis ValidationConfirmation that there are nobacteriostatic effects that inhibitgrowth for flexible endoscopesNo bacteriostaticeffect aftersterilization withgrowth of definedmicro-organismsPass
Simulated Use TestingWorst-case testing with unwashedflexible endoscopes inoculated withtest culture challenge6-log reduction inGeobacillusstearothermophilusPass
In-Use TestingWorst-case testing with clinically usedflexible endoscopes.No growth uponrecovery aftersterilizationPass
Functionality & MaterialCompatibilityWorst-case material compatibilitytesting of multiple flexible endoscopeswith back to back cyclesInternalSpecificationPass
Biocompatibility CouponsWorst-case residual and cytotoxicity(MTT) testing with material couponsISO 10993-5Pass
Biocompatibility FlexibleEndoscopesWorst-case residual and cytotoxicity(MTT) testing with flexibleendoscopesISO 10993-5Pass

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

Table 4 Standards Used:

OrganizationDesignationTitle
IEC60601-1-2 Edition 4.0 2014-02Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
IEC61010-1 Edition 3.1 2017-01CONSOLIDATED VERSIONSafety requirements for electrical equipment for measurement control and laboratory use - Part 1: General requirements [Including: Corrigendum 1 (2019)]
AAMITIR 12:2020Design, Testing and Labeling of Reusable Medical Devices
ANSI AAMIISO11737-1:2018Microbial Methods Determination of the population of microorganisms
ASTME1837-96 (2014)Simulated Use Test
AAMI ANSIST77:2013(R)2018Containment Devices
ANSI AAMIISO17664:2017Info To Be Provided for the processing of medical devices
AAMITIR39:2009(R)2017Selecting a Microbial Challenge and Inoculation sites
ANSI AAMIISO11138-1:2017BIs part 1
ISO10993-5 Third edition 2009-06-01Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
ISO10993-12 Fourth edition 2012-07-01Biological evaluation of medical devices - Part 12: Sample preparation and reference materials
USP-NF1035BIs for Sterilization
AOAC966.04Sporicidal Activity of Disinfectants
AAMI ANSIST14:2008EtO Sterilization in Health Care Facilities
IEC61000Electromagnetic Compatibility
ISTA2A and 2BInternational Safe Transit Association
OSHACAS 7722-84-1H2O2 Handling Guidance
AOACOfficial Method 966.04Sporicidal Activity of Disinfectants. Official Methods of Analysis of the AOAC, 2013 Edition

Conclusion

The conclusion drawn from the non-clinical testing demonstrates that the subject device, SteroScope® Sterilization Technology System submitted under K233762, is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K131407, the STERRAD 100NX DUO Cycle, and the legally marketed reference device cleared under K092622, the STERRAD 100NX EXPRESS Cycle.

§ 880.6860 Ethylene oxide gas sterilizer.

(a)
Identification. An ethylene gas sterilizer is a nonportable device intended for use by a health care provider that uses ethylene oxide (ETO) to sterilize medical products.(b)
Classification. Class II (performance standards).