K Number
K233697

Validate with FDA (Live)

Date Cleared
2024-08-09

(266 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SonoMax Series Digital Color Doppler Ultrasound System is intended for diagnostic ultrasound imaging in B(2D/3D/4D), B/M, M, B+CFM, B+CPA(PD), B+DPD, B+PW, B+CW, B+CFM+D(PW)/CW, B+CPA(PD)+D(PW)/CW, TDI and Fusion Harmonic Imaging modes. The device is a general-purpose ultrasonic imaging instrument intended for use by appropriately-trained qualified healthcare professionals for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, thyroid, testes), Neonatal Cephalic, Cardiac (adult, pediatric), Musculoskeletal (Conventional, Superficial), Peripheral Vascular, Trans-esophageal, Trans-vaginal, OB/GYN and Urology, which is intended to be used in a hospital or medical clinic.

Device Description

The SonoMax Series Digital Doppler Ultrasound System is an integrated preprogrammed color doppler ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications. This system is a Track 3 device that employs a wide array of probes that include linear array, convex array and phased array.This system consists of a mobile console with keyboard control panel, power supply module, color LCD monitor and optional probes. This system is a mobile, general purpose, software controlled, color diagnostic ultrasound system. Its basic function is to acquire ultrasound echo data and to display the image in B(2D/3D/4D), B/M, M, B+CFM, B+CPA(PD), B+DPD, B+PW, B+CW, B+CFM+D(PW)/CW, B+CPA(PD)+D(PW)/CW, TDI, Fusion Harmonic lmaging modes or a combination of these mode.

AI/ML Overview

The FDA 510(k) summary for the SonoMax Series Digital Color Doppler Ultrasound System (K233697) indicates that no clinical testing was required to demonstrate substantial equivalence. Therefore, the document does not contain information about acceptance criteria or a study proving the device meets performance criteria beyond non-clinical testing and comparison to a predicate device.

Specifically, under "7. Clinical Test:" on page 6, it explicitly states: "No clinical testing was required."

Given this, I cannot extract the requested information regarding acceptance criteria, device performance, sample sizes, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, or ground truth for the training set from the provided text for this cleared device. The substantial equivalence determination was based on non-clinical tests (electrical, mechanical, thermal, electromagnetic compatibility, biocompatibility, and acoustic output) and a comparison of the device's technical specifications and indications for use to those of a predicate device (XBit 90 Digital Color Doppler Ultrasound System, K200780).

{0}------------------------------------------------

August 9, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

CHISON Medical Technologies Co., Ltd. Yingying Chen Official Correspondent No.3 Changjiang South Road, Xinwu District Wuxi. Jiangsu 214028 CHINA

Re: K233697

Trade/Device Name: SonoMax Series Digital Color Doppler Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: July 5, 2024 Received: July 8, 2024

Dear Yingying Chen:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Yanna S. Kang -S

Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K233697

Device Name

SonoMax Series Digital Color Doppler Ultrasound System

Indications for Use (Describe)

The SonoMax Series Digital Color Doppler Ultrasound System is intended for diagnostic ultrasound imaging in B(2D/3D/4D), B/M, M, B+CFM, B+CPA(PD), B+DPD, B+PW, B+CW, B+CFM+D(PW)/CW, B+CPA(PD)+D(PW)/CW, TDI and Fusion Harmonic Imaging modes. The device is a general-purpose ultrasonic imaging instrument intended for use by appropriately-trained qualified healthcare professionals for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, thyroid, testes), Neonatal Cephalic, Cardiac (adult, pediatric), Musculoskeletal (Conventional, Superficial), Peripheral Vascular, Trans-esophageal, Trans-vaginal, OB/GYN and Urology, which is intended to be used in a hospital or medical clinic.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

1. Submitter:

Submitter:CHISON Medical Technologies Co., Ltd.
Address:No.3 Changjiang South Road, Xinwu District, Wuxi, 214028 Jiangsu,P.R. China
Contact :Mrs. Chen Yingying
Tel:+86-510-8531-0019
Fax:+86-510-8531-0021
Date Prepared:November 10th, 2023

2. Device :

Trade Name: SonoMax Series Digital Color Doppler Ultrasound System(SonoMax 1/ SonoMax 2/ SonoMax 3/ SonoMax 5/ SonoMax 6/SonoMax 7/SonoMax 7 Super/ SonoMax 7 EXP/ SonoMax 8/ SonoMax 8 Super/ SonoMax 8 EXP/SonoMax 9/ SonoMax 9 Super/ SonoMax 9 EXP/SonoMax 10/ SonoMax 11/ SonoMax 22)

Common Name: Diagnostic Ultrasound System with Transducers

Classfication: Requlatory Class: II Review Category: Tier II

Classification Name21 CFRSectionProduct Code
Ultrasonic Pulsed Doppler Imaging System892.155090-IYN
Ultrasonic Pulsed Echo Imaging System892.156090-IYO
Diagnostic Ultrasound Transducer892.157090-ITX

3. Predicate Device(s):

DeviceModelProduct Code510(k)Number
Predicate deviceXBit 90 Digital ColorDoppler Ultrasound SystemIYN, IYO, ITXK200780

4. Device Description:

The SonoMax Series Digital Doppler Ultrasound System is an integrated preprogrammed color doppler ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications.

This system is a Track 3 device that employs a wide array of probes that include linear array, convex array and phased array.This system consists of a mobile console with keyboard control panel, power supply module, color LCD monitor and optional probes. This system is a mobile, general purpose, software controlled, color diagnostic ultrasound system. Its basic function is to acquire ultrasound echo data

{4}------------------------------------------------

CHISON Medical Technologies Co., Ltd. SonoMax Series Digital Color Doppler Ultrasound System and to display the image in B(2D/3D/4D), B/M, M, B+CFM, B+CPA(PD), B+DPD, B+PW, B+CW, B+CFM+D(PW)/CW, B+CPA(PD)+D(PW)/CW, TDI, Fusion Harmonic lmaging modes or a combination of these mode.

5. Indications for Use:

The SonoMax Series Digital Color Doppler Ultrasound System is intended for diagnostic ultrasound imaging in B(2D/3D/4D), B/M, M, B+CFM, B+CPA(PD), B+DPD, B+PW, B+CW, B+CFM+D(PW)/CW, B+CPA(PD)+D(PW)/CW, TDI and Fusion Harmonic Imaging modes. The device is a general-purpose ultrasonic imaging instrument intended for use by appropriately-trained qualified healthcare professionals for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, thyroid, testes), Neonatal Cephalic, Adult Cephalic, Cardiac (adult, pediatric), Musculo-skeletal (Conventional, Superficial), Peripheral Vascular, Trans-esophageal, Trans-rectal, Trans-vaginal, OB/GYN and Urology, which is intended to be used in a hospital or medical clinic.

6. Summary of Non-Clinical Tests:

The SonoMax Series Digital Color Doppler Ultrasound System has been evaluated for electrical, mechanical, thermal and electromagnetic compatibility safety, biocompatibility and acoustic output.

The device has been found to conform to applicable medical device safety standards in regards to thermal, mechanical and electrical safety as well as biocompatibility.

ANSI AAMI ES60601-1:2015 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance.

IEC 60601-1-2: 2014 Medical Electrical Equipment - Part 1-2: General Requirements For Safety - Collateral Standard: Electromagnetic Compatibility -- Requirements and Tests.

IEC 60601-2-37:2015 Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment.

Output Indices on Diagnostic Ultrasound Equipment

ISO 10993-1:2018 Biological Evaluation of Medical Devices -- Part 1: Evaluation And Testing Within A Risk Management Process

The following quality assurance measures are applied to the development of the system:

Risk Management

Requirement review and Design reviews

Testing on unit level (Module verification)

Integration testing (system verification)

Performance testing (Verification)

Safety testing (Verification)

The biocompatibility was evaluated and meets the ISO10993 series standard and FDA quidance.

{5}------------------------------------------------

7. Clinical Test:

No clinical testing was required.

Software Documentation for a Basic Documentation Level, per the FDA guidance document,"Guidance for Content of Premarket Submissions for Device Software Functions Document issued on June 14, 2023", is also included as part of this submission.

8. Determination of Substantially Equivalent:

ItemsPredicate DeviceSubmission DeviceRemark
ItemsXBit 90 Digital Color DopplerUltrasound System(K200780)SonoMax Series Digital ColorDoppler Ultrasound SystemRemark
Indicationsfor UseFetalAbdominalPediatricSmall Organ (breast,thyroid ,testes)Neonatal CephalicAdult CephalicTrans-rectalTrans-vaginalMusculo-skeletal( Conventional,Superficial)Cardiac(adult ,pediatric)Peripheral VascularOB/GYN, UrologyTrans-esophagealFetalAbdominalPediatricSmall Organ (breast,thyroid ,testes)Neonatal CephalicAdult CephalicTrans-rectalTrans-vaginalMusculo-skeletal( Conventional,Superficial)Cardiac(adult ,pediatric)Peripheral VascularOB/GYN, UrologyTrans-esophagealSame
DesignAutocorrelation for colorprocessing and FFT for pulseand CW Doppler processing.Supporting Linear, Curve ,Phase array and Volume probes .Cine play back capabilityImage file archiveAutocorrelation for colorprocessing and FFT for pulseand CW Doppler processing.Supporting Linear, Curve andPhase array probes.Cine play back capabilityImage file archiveSame
TGC 8 sliderTGC 10 sliderSEAnalysis I
Depth Range: 0-45 cmDepth Range:0- 45 cmSame
OperatingControlsItems256 shades of gray256 shades of graySame
OperatingControlsItemsB Dynamic range control: 20-180dBB Dynamic range control: 20-180dBSame
OperatingControlsItemsGain:0-255, 1/stepGain:0-255,1/stepSame
OperatingControlsItemsFocal Number: adjustableFocal Number: adjustableSame
OperatingControlsItemsFocus position: adjustableFocus position: adjustableSame
B steer: available on lineartransducersB steer: available on lineartransducersSame
B Persistence: 7 stepsB Persistence: 7 stepsSame
ROI size/position: adjustableROI size/position: adjustableSame
Color Wall Filter settings: 8 stepsColor Wall Filter settings: 8 stepsSame
Color Baseline:16 stepsColor Baseline: 16 stepsSame
Color Maps: 21 mapsColor Maps: 21 mapsSame
PW sweeping speed: 6 stepsPW sweeping speed: 6 stepsSame
Color Invert: on/offColor Invert: on/offSame
PW Wall Filter: 7 stepsPW Wall Filter: 7 stepsSame
PW sample volume: 0.5-30mm(PW only)PW sample volume: 0.5-30mm(PW only)Same
PW angle correction:89~89degrees,1/stePW angle correction:89~89degrees, 1/stepSame
Baseline: 8stepsBaseline: 8stepsSame
Cine control:step, play backward,play continuouslyCine control:step, play backward,play continuouslySame
Doppler Auto TraceDoppler Auto TraceSame
Freeze control:Toggling freezekeyFreeze control: Toggling freezekeySame
SafetyComplianceANSI AAMI ES60601-1IEC 60601-1-2ISO 10993-1ISO 10993-5ISO 10993-10ISO 10993-23IEC 60601-2-37ANSI AAMI ES60601-1IEC 60601-1-2ISO 10993-1ISO 10993-5ISO 10993-10ISO 10993-23IEC 60601-2-37Same
OperationModeB ModeB Mode
FHIFHISame
B/M ModeB/M ModeSame
M ModeM ModeSame
Dual modeDual modeSame
Quad modeQuad modeSame
CFM modeCFM modeSame
CPA modeCPA modeSame
DPD modeDPD modeSame
PW modePW modeSame
CW modeCW modeSame
B/BC modeB/BC modeSame
2D Steer2D SteerSame
TriplexTriplexSame
QuadplexQuadplexSame
Free Steer MFree Steer MSame
HPRF ModeHPRF ModeSame
TDITDISame
Color M ModeColor M ModeSame
TSSTSSSame
Curved PanoramicCurved PanoramicSame
Sono ContrastSono ContrastSame
SoundFlowSoundFlowSame
SonoVectorSonoVectorSame
Auto TGCAuto TGCSame
LGCLGCSame
HD 3DHD 3DSame
Stress EchoStress EchoSame
Strain and Strain RateStrain and Strain RateSame
Trapezoidal ImagingTrapezoidal ImagingSame
CompoundCompoundSame
SRASRASame
ECGECGSame
Human BodymarkHuman BodymarkSame
Auto IMTAuto IMTSame
Free NTFree NTSame
BiopsyBiopsySame
Super NeedleSuper NeedleSame
general measurement packagegeneral measurement packageSame
OB measurement packageOB measurement packageSame
GYN measurement packageGYN measurement packageSame
URO measurement packageURO measurement packageSame
cardiac measurement packagecardiac measurement packageSame
vascular measurement packagevascular measurement packageSame
small parts measurementpackagesmall parts measurementpackageSame
Pediatric measurement packagePediatric measurement packageSame
TCD measurement packageTCD measurement packageSame
4D software package4D software packageSame
Virtual HDVirtual HDSame
SonoFusionSonoFusionSame
HelpfunctionSonoHelpSame
Q-imageQ-imageSame
Q-flowQ-flowSame
Q-beamQ-beamSame
SonoColorSonoColorSame
ElastographyElastographySame
SonoBeamSonoBeamSame
AIOAIOSame
DisplayAnnotationsLogo; Hospital Name;Exam date;Exam time; AcousticPower ;Mechanical index;Thermalindes;Probe model;ECG ico;TGCCorve;Focus position;Imagingparameters;Dynamic Trackballindices; Systemstatus;Gray/Color barLogo; Hospital Name;Exam date;Exam time; AcousticPower :Mechanical index;Thermalindes;Probe model; TGCCorve;Focus position;Imagingparameters;TTouch pad; Systemstatus:Gray/Color barSame
Measurements2D mode: Depth,Distance,Area:Ellipse, Trace, Spline,Trace, Length ,Volume :Distance,Ellipse, Ellipse + Distance,Distance Ratio, Area Ratio , IMT,Volume Flow,Color Velocity;M mode: Distance,Time, Slope,Heart Rate,Velocity;Doppler mode: D Velocity ,Time,Heart Rate, Acceleration,DTrace,ED/PS, Volume Flow;2D mode: Depth,Distance,Area:Ellipse, Trace, Spline,Trace, Length ,Volume :Distance,Ellipse, Ellipse + Distance,Distance Ratio, Area Ratio , IMT,Volume Flow,Color Velocity;M mode: Distance,Time, Slope,Heart Rate,Velocity;Doppler mode: D Velocity ,Time,Heart Rate, Acceleration,DTrace,ED/PS, Volume Flow;Same
TransducerTypes &ConnectorsConvex Array, Phased Array,Linear Array, Volume probe4portsConvex Array, Phased Array,Linear Array, Volume probe5portsSEAnalysis 2
Users /SitesHospitals, clinics usageHospitals, clinics usageSame
AcousticOutputTrack 3; MI, TIS, TIC, TIBDerated Ispta: 720mW/cm²maximum,TIS/TIB/TIC:0.1-4.0 Range,Mechanical Index: 1.9 Maximum,or Derated Isppa: 190 W/cm²maxTrack 3; MI, TIS, TIC, TIBDerated Ispta: 720mW/cm2maximum,TIS/TIB/TIC:0.1-4.0 Range,Mechanical Index: 1.9 Maximum,or Derated Isppa: 190 W/cm²maxSame
PowerRequirementsPower requirements:AC :100V-240V,Frequenzy:50-60Hz;Operating temperature:10-40°C;Relative humidity 30-75%;Barometric pressure:700 to 1060hPaPower requirements:AC :100V-240V,Frequenzy:50-60Hz;Operating temperature:10-40°C;Relative humidity 30-75%;Barometric pressure:700 to 1060 hPaSame

{6}------------------------------------------------

SonoMax Series Digital Color Doppler Ultrasound System

{7}------------------------------------------------

SonoMax Series Digital Color Doppler Ultrasound System

{8}------------------------------------------------

SonoMax Series Digital Color Doppler Ultrasound System

{9}------------------------------------------------

SE Analysis 1:

Operation Controls, compared with the predicate device, the submission device employs the same operation controls design ,but has different TGC slider. But both of them comply with the requirements of ANSI AAMI ES60601-1 & IEC60601-2-37 and meet clinical requirements. Therefore they can be considered Substantially Equivalent in safety and effectiveness, and no new risk is raised, so the SE is not affected.

SE Analysis 2:

Compared with the predicate device, the submission device has different number of probe ports. But both them comply with the requirements of ANSI AAMI ES60601-1. Therefore they can be considered Substantially Equivalent in safety, and no new risk is raised, so the SE is not affected.

9. Conclusion

In accordance with the Act. 21 CFR Part 807 and based on the information provided in this premarket notification, CHISON Medical Technologies Co., Ltd. concludes that the SonoMax Series Digital Color Doppler Ultrasound System is substantially equivalent to the predicate device with regard to safety and effectiveness.

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.