K Number
K233687

Validate with FDA (Live)

Date Cleared
2024-05-03

(168 days)

Product Code
Regulation Number
892.1000
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ECHELON Synergy System is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spinlattice relaxation time (TI), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

Anatomical Region: Head, Body, Spine, Extremities Nucleus excited: Proton

Diagnostic uses:

  • · TI, T2, proton density weighted imaging
  • · Diffusion weighted imaging
  • · MR Angiography
  • · Image processing
  • · Spectroscopy
  • · Whole Body
Device Description

The ECHELON Synergy is a Magnetic Resonance Imaging System that utilizes a 1.5 Tesla superconducting magnet in a gantry design. Magnetic Resonance imaging (MRI) is based on the fact that certain atomic nuclei have electromagnetic properties that cause them to act as small spinning bar magnets. The most ubiquitous of these nuclei is hydrogen, which makes it the primary nuclei currently used in magnetic resonance imaging. When placed in a static maqnetic field, these nuclei assume a net orientation or alignment with the magnetic field, referred to as a net magnetization vector. The introduction of a short burst of radiofrequency (RF) excitation of a wavelength specific to the magnetic field strength and to the atomic nuclei under consideration can cause a re-orientation of the net magnetization vector. When the RF excitation is removed, the protons relax and return to their original vector. The rate of relaxation is exponential and varies with the character of the proton and its adjacent molecular environment. This re-orientation process is characterized by two exponential relaxation times, called T1 and T2. A RF emission or echo that can be measured accompanies these relaxation events. The emissions are used to develop a representation of the relaxation events in a three dimensional matrix. Spatial localization is encoded into the echoes by varving the RF excitation. applying appropriate magnetic field gradients in the x, y, and z directions, and changing the direction and strength of these gradients. Images depicting the spatial distribution of the NMR characteristics can be reconstructed by using image processing techniques similar to those used in computed tomography.

AI/ML Overview

The provided document describes the Fujifilm ECHELON Synergy V10.0 MRI system, which is an updated version of a previously cleared device. The submission focuses on demonstrating substantial equivalence to the predicate device (ECHELON Synergy MRI System K223426) by highlighting changes and providing performance evaluations.

Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of "acceptance criteria" for the overall device in a quantifiable format. Instead, it demonstrates the new features' performance through clinical image testing and phantom studies, comparing them to conventional methods or manual positioning. The acceptance criteria for "DLR Clear" are implied through achieving statistical significance for superiority in certain image quality metrics over conventional imaging and clinical acceptability. For "AutoPose," the criteria are implied through reduction or equivalence in time and steps for slice positioning.

Here's a summary of the performance results for the new features (DLR Clear and AutoPose):

FeatureAcceptance Criteria (Implied)Reported Device Performance
DLR ClearPhantom Testing: Reduce truncation artifact, improve image sharpness, improve spatial resolution (Total Validation, Relative Edge Sharpness, FWHM).Clinical Testing: Superiority or equivalence to conventional images in truncation artifact reduction, image sharpness, lesion conspicuity, and overall image quality (statistically significant if superior). Also, clinical acceptability across all images with DLR Clear.High-Resolution vs. Low-Resolution (Clinical): Superiority in overall image quality for high-resolution DLR Clear images compared to low-resolution conventional images from the same data, and clinical acceptability.Phantom Testing: Demonstrated reduction of truncation artifact, improvement of image sharpness, and improvement of spatial resolution. (Reported metrics: Total Validation, Relative Edge Sharpness, FWHM).Clinical Testing:- Truncation artifact reduction, image sharpness, and overall image quality in images with DLR Clear were superior to conventional images with statistically significant difference (p<0.05).- Lesion conspicuity in images with DLR Clear was superior or equivalent to conventional images.- All images with DLR Clear were evaluated as clinically acceptable by reviewers.High-Resolution vs. Low-Resolution (Clinical):- Overall image quality in high-resolution images with DLR Clear was superior to low-resolution conventional images with statistically significant difference.- High-resolution images with DLR Clear were clinically acceptable.
AutoPose (Spine, Breast, HipJoint)Reduce time and number of steps (or at least be equivalent) in slice positioning compared to manual slice positioning, and achieve certified radiological technologist evaluation for efficacy.- Almost all cases for AutoPose Spine, Breast, and HipJoint were able to reduce the time and number of steps in slice positioning compared to manual slice positioning.- The remaining cases showed the same time and number of steps as manual slice positioning. (Evaluated by certified radiological technologists).

2. Sample Sizes and Data Provenance

DLR Clear:

  • Test Set Sample Size: 53 unique subjects (patients and volunteers).
  • Data Provenance: U.S. and Japan.
  • Retrospective/Prospective: Not explicitly stated, but the description "scanned in the anatomical regions... for the test datasets separately from the training and validation datasets" suggests a prospective collection for the test set or careful selection from retrospective archives to act as a distinct test set.

AutoPose (Spine, Breast, HipJoint):

  • Test Set Sample Size:
    • Spine: 177 cases
    • Breast: 66 cases
    • HipJoint: 65 cases
  • Data Provenance: FUJIFILM Healthcare Corporation and clinical site (country not specified for clinical site, but assuming Japan given the company origin and DLR Clear's data provenance).
  • Retrospective/Prospective: Not explicitly stated.

3. Number of Experts and Qualifications for Ground Truth for the Test Set

DLR Clear:

  • Number of Experts: Three (3) US board-certified radiologists.
  • Qualifications of Experts: US board-certified radiologists. (Specific experience levels not provided).

AutoPose:

  • Number of Experts: "Certified radiological technologists." (Number not specified, specific experience not provided). Their evaluation was on the reduction/equivalence of time and steps for slice positioning.

4. Adjudication Method for the Test Set

DLR Clear:

  • The images were randomized and blinded to the reviewers.
  • Reviewers compared image quality metrics (truncation artifact reduction, image sharpness, lesion conspicuity, and overall image quality) using a 3-point scale.
  • The results indicate statistical analysis (p<0.05), implying that the individual ratings were aggregated or compared statistically to determine superiority. It does not explicitly state a consensus-based adjudication method (like 2+1 or 3+1). It seems individual reviewer scores were used.

AutoPose:

  • Evaluated by "certified radiological technologists." The method of combining their evaluations or achieving consensus is not explicitly described. It states "They evaluated that almost cases..." suggesting a collective assessment or majority opinion, but no formal adjudication process is detailed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

There is no explicit MRMC comparative effectiveness study directly comparing human reader performance with and without AI assistance (i.e., effect size of human improvement) described in this document.

The study for DLR Clear involved radiologists reviewing images reconstructed with DLR Clear versus conventional methods. This evaluates the impact of the AI-powered reconstruction on image quality, which in turn facilitates human reading, but it does not measure the human reader's diagnostic performance improvement (e.g., AUC, sensitivity, specificity) when using the AI-processed images compared to conventional images.

6. Standalone Performance Study

DLR Clear:
Yes, a form of standalone performance was done for DLR Clear. The phantom testing ("Total Validation, Relative Edge Sharpness, and FWHM") demonstrates the algorithm's direct performance on objective metrics before human interpretation. The clinical image review also evaluates the quality of the algorithm's output directly, without explicitly involving an "AI-assisted reading" scenario. The radiologists were evaluating the images produced by the algorithm, not acting as diagnosticians making decisions with or without AI output.

AutoPose:
Yes, the AutoPose feature's performance (reduction in time/steps for positioning) was evaluated by radiological technologists, which can be considered a standalone performance assessment for the positioning automation.

7. Type of Ground Truth Used

DLR Clear:

  • Phantom Testing: The ground truth for phantom testing is the physical properties of the cylindrical and ACR phantoms themselves, measured by objective metrics (e.g., FWHM for spatial resolution).
  • Clinical Testing: The ground truth for clinical image quality assessment relies on expert consensus/opinion (ratings by 3 US board-certified radiologists) regarding image quality metrics (truncation artifact, sharpness, lesion conspicuity, overall image quality) and clinical acceptability. There's no mention of pathology or outcomes data as ground truth for these image quality assessments.

AutoPose:

  • The ground truth for AutoPose is implicitly the objective measurement of time and steps required for slice positioning, compared against manual methods, as evaluated by certified radiological technologists.

8. Sample Size for the Training Set

DLR Clear: Not explicitly stated. The document mentions "test datasets separately from the training and validation datasets" for DLR Clear, indicating that training data existed, but its size is not provided.

AutoPose: Not explicitly stated.

9. How the Ground Truth for the Training Set Was Established

DLR Clear: Not explicitly stated. Since it's a deep learning reconstruction (DLR), the ground truth for training would typically involve pairs of raw MRI data and high-quality, potentially "gold standard" reconstructed images (e.g., using more extensive acquisition or advanced reconstruction techniques) or specific labels for image quality characteristics.

AutoPose: Not explicitly stated. For an automated positioning algorithm, training ground truth would likely involve expert-defined optimal slice prescriptions on a large dataset of diverse anatomies.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services - USA logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 3, 2024

Fujifilm Healthcare Americas Corporation Chaitrali Kulkarni Sr. Regulatory Affairs Specialist 81 Hartwell Ave Suite 300 Lexington, Massachusetts 02421

Re: K233687

Trade/Device Name: ECHELON Synergy V10.0 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH Dated: April 9, 2024 Received: April 9, 2024

Dear Chaitrali Kulkarni:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

D. R. X.

Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K233687

Device Name

ECHELON Synergy V10.0

Indications for Use (Describe)

The ECHELON Synergy System is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spinlattice relaxation time (TI), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

Anatomical Region: Head, Body, Spine, Extremities Nucleus excited: Proton

Diagnostic uses:

  • · TI, T2, proton density weighted imaging
  • · Diffusion weighted imaging
  • · MR Angiography
  • · Image processing
  • · Spectroscopy
  • · Whole Body

Type of Use (Select one or both. as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K233687

Section 5

510(k) Summary

{4}------------------------------------------------

Submitter:FUJIFILM Healthcare Corporation2-1, ShintoyofutaKashiwa-shi, Chiba, 277-0804 Japan
Contact:Ms. Kulkarni Chaitrali
Telephone number:704-517-4886
Telephone number:704-517-4886
E-mail:chaitrali.kulkarni@fujifilm.com
Date:November 16, 2023

Subject Device Name

Trade/Proprietary Name:ECHELON Synergy V10.0 MRI system
Regulation Number:21 CFR 892.1000
Regulation Name:System, Nuclear Magnetic Resonance Imaging
Product CodeLNH
Class2
PanelRadiology

Predicate Device Name

Predicate Device(s):ECHELON Synergy MRI system (K223426)
Regulation Number:21 CFR 892.1000
Regulation Name:System, Nuclear Magnetic Resonance Imaging
Product CodeLNH
Class2
PanelRadiology

Device Intended Use

The ECHELON Synergy System is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

Anatomical Region: Head, Body, Spine, Extremities

Nucleus excited: Proton

Diagnostic uses:

  • . T1, T2, proton density weighted imaging
  • Diffusion weighted imaging ●
  • MR Angiography
  • Image processing
  • Spectroscopy ●
  • Whole Body

{5}------------------------------------------------

Device Description

Function

The ECHELON Synergy is a Magnetic Resonance Imaging System that utilizes a 1.5 Tesla superconducting magnet in a gantry design.

Scientific Concepts

Magnetic Resonance imaging (MRI) is based on the fact that certain atomic nuclei have electromagnetic properties that cause them to act as small spinning bar magnets. The most ubiquitous of these nuclei is hydrogen, which makes it the primary nuclei currently used in magnetic resonance imaging. When placed in a static maqnetic field, these nuclei assume a net orientation or alignment with the magnetic field, referred to as a net magnetization vector. The introduction of a short burst of radiofrequency (RF) excitation of a wavelength specific to the magnetic field strength and to the atomic nuclei under consideration can cause a re-orientation of the net magnetization vector. When the RF excitation is removed, the protons relax and return to their original vector. The rate of relaxation is exponential and varies with the character of the proton and its adjacent molecular environment. This re-orientation process is characterized by two exponential relaxation times, called T1 and T2. A RF emission or echo that can be measured accompanies these relaxation events.

The emissions are used to develop a representation of the relaxation events in a three dimensional matrix. Spatial localization is encoded into the echoes by varving the RF excitation. applying appropriate magnetic field gradients in the x, y, and z directions, and changing the direction and strength of these gradients. Images depicting the spatial distribution of the NMR characteristics can be reconstructed by using image processing techniques similar to those used in computed tomography.

Physical and Performance Characteristics

MRI is capable of producing high quality anatomical images without the associated risks of ionizing radiation. The biological properties that contribute to MR image contrast are different from those responsible for x-ray image contrast. In MR imaging, difference in proton density, blood flow, and T1 and T2 relaxation times can all contribute to image contrast. By varying the pulse sequence characteristics, the resulting images can emphasize T1, T2, proton density, or the molecular diffusion of water or other proton containing molecules. And MR system has the Function of measuring spectroscopy.

Performance Evaluation

The ECHELON Synergy V10.0 MRI System is equivalent to the ECHELON Synergy MRI System (K223426) with following exceptions:

  • Camera unit is added for Synergy Vision and Visual StillShot.
  • . Alternative super-conducting magnet and units related to alternative super-conducting magnet are added.
  • . Application software is changed to V10.0A.

A rationale analysis was then conducted, and the results are contained in Table 1.

Testing TypeRationale Analysis
Performance Testing - BenchPerformance bench testing was conducted on the applicable new feature.Test data confirmed that new feature perform as intended for diagnostic use.
Performance Testing - ClinicalClinical image examples are provided for applicable new feature and that we judged to be sufficient to evaluate clinicalusability.

Table 1 Performance Analysis

{6}------------------------------------------------

Device Technological Characteristics

The control and image processing hardware and the base elements of the system software are identical to the predicate device. The ECHELON Synergy V10.0 MRI system software is substantially equivalent to the ECHELON Synergy MRI System (K223426). See tables below. The technological characteristics in regard to hardware of the ECHELON Synergy V10.0 MRI system and the predicate are listed in Table 2.

Table 2 Comparison: Hardware
ITEMPREDICATE DEVICEECHELON SYNERGY MRI SYSTEM(K223426)SUBJECT DEVICEECHELON Synergy V10.0DIFFERENCE
SystemStandards MetNEMA: MS 1, MS 2, MS 3, MS 4, MS 5, MS 8, MS 14,IEC: 60601-1, 60601-1-2, 60601-2-33, 62304NEMA: MS 1, MS 2, MS 3, MS 4, MS 5, MS 8, MS 14,IEC: 60601-1, 60601-1-2, 60601-2-33, 62304No
Magnet and GantryType and Field StrengthSuper-conducting magnet, horizontal bore, 1.5 TeslaSuper-conducting magnet, horizontal bore, 1.5 TeslaNo
Resonant Frequency63.86MHz63.86MHzNo
Bore dimensionCircle shape with diameter 70cmCircle shape with diameter 70cmNo
Gradient SystemGradient Strength33mT/m33mT/mNo
Slew Rate130 T/m/sec130 T/m/secNo
Rise Time254µsec to 33mT/m254µsec to 33mT/mNo
Audible Noise (MCAN)Ambient59.9 dBA59.9 dBANo
Lpeak122.7 dBA122.7 dBANo
Leq116.5 dBA116.5 dBANo
RF SystemTransmitter channels11No
Peak Envelop Power18 kW18 kWNo
Duty Cycle85% (Gating max), 10% at full power85% (Gating max), 10% at full powerNo
RF receiver channel3232No

Table 2 Comparison: Hardware

The hardware differences from the predicate device to the ECHELON Synerqy V10.0 MRI System are analyzed in Table 3.

Table 3 Hardware Comparison Analysis
FDARequirementsAnalyze why any differences between the subject device and predicate(s) do not render the device NSE (e.g., does notconstitute a new intended use; and any differences in technological characteristics are accompanied by information thatdemonstrates the device is as safe and effective as the predicate and do not raise different questions of safety andeffectiveness than the predicate ), affect safety or effectiveness, or raise different questions of safety and effectiveness (seesection 513(i)(1)(A) of the FD&C Act and 21 CFR 807.87(f)).
DeviceModificationSummary• Camera unit is added for Synergy Vision and Visual StillShot.• Alternative super-conducting magnet and units related to alternative super-conducting magnet are added. The deviceis changed but there are no significant changes in technology, engineering and performance.
SignificantChanges□ Manufacturing Process□ Labeling□ Technology□ Performance
□ Engineering□ Materials□ Others☑ None (See rationale statement)
FUJIFILMRationaleStatementModified specification doesn't constitute a new intended use. There are no significant changes in technologicalcharacteristics. For safety, gradient system and RF system is controlled according to same regulation as ECHELONSynergy MRI System (K223426). So, safety and effectively of the device are same as ECHELON Synergy MRI System(K223426).

{7}------------------------------------------------

The technological characteristics in regards to coils of the ECHELON Synergy V10.0 MRI System and the predicate are listed in Table 4.

ITEMPREDICATE DEVICESUBJECT DEVICEDIFFERENCE
ECHELON SYNERGY MRI SYSTEM(K223426)ECHELON Synergy V10.0
RF CoilsTransmit CoilT/R BodyT/R BodyNo
Receiver CoilsFlexFit Neuro CoilFlexFit Neuro CoilNo
FlexFit Blanket Coil A,FlexFit Blanket Coil BFlexFit Blanket Coil A,FlexFit Blanket Coil BNo
Extremity CoilExtremity CoilNo
Hand/Wrist CoilHand/Wrist CoilNo
Breast CoilBreast Support Kit 2Breast CoilBreast Support Kit 2No
Micro Coil A,Micro Coil BMicro Coil A,Micro Coil BNo
Shoulder CoilShoulder CoilNo
Spine CoilSpine CoilNo
Foot/Ankle CoilFoot/Ankle CoilNo
Flex M Coil, Flex S CoilFlex M Coil, Flex S CoilNo
Table 4 Comparison: RF Coils
----------------------------------

The coil differences from the predicate device to the ECHELON Synergy MRI system are analyzed in Table 5.

Table 5 Coil Comparison Analysis
FDARequirementsAnalyze why any differences between the subject device and predicate(s) do not render the device NSE (e.g., does notconstitute a new intended use; and any differences in technological characteristics are accompanied by information thatdemonstrates the device is as safe and effective as the predicate and do not raise different questions of safety andeffectiveness than the predicate ), affect safety or effectiveness, or raise different questions of safety and effectiveness (seesection 513(i)(1)(A) of the FD&C Act and 21 CFR 807.87(f)).
DeviceModificationSummaryThere are no differences from the predicate device.
SignificantChanges□ Manufacturing Process□ Labeling□ Technology□ Performance
□ Engineering□ Materials□ Others☑ None (See rationale statement)
FUJIFILMRationaleStatementThere are no significant changes in technological characteristics. Therefore, safety, intended use and effectively of the RF coils are same as ECHELON Synergy MRI System (K223426).

{8}------------------------------------------------

The technological characteristics in regard to changes in functionality of the ECHELON Synergy V10.0 MRI System as compared to the predicate are listed in Table 6.

ITEMDIFFERENCESANALYSIS
Operating SystemNoneNo
CPU PlatformNoneNo
Application SoftwareGoing from V9.0A to V10.0ASee Table 7
Scan TasksFollowing function is updated.- Auto VoiceFollowing positioning applications are added.- AutoPose Breast, AutoPose HipJointFollowing positioning application is updated.- AutoPose SpineFollowing function is added.- Series SaveSee Table 7
2D Processing TasksNoneNo
3D Processing TasksNoneNo
Analysis TasksNoneNo
Maintenance TasksNoneNo
Viewport ToolsNoneNo
Film, Archive ToolsNoneNo
Network ToolsFollowing Network Tools are added.- AutoProtocolSee Table 7
Protocol EnhancementsFollowing function is updated.- Delayed Enhancement Imaging- Navi slice positioning- BeamNaviFollowing variation of Protocol Enhancements are added in Auto Table Centering.- Chest, Breast, General Abdomen, Liver/Kidney, Pelvis, WholeBodyFollowing Protocol Enhancements are added.- Navigated StillShot- Visual StillShot- MROC- IQ Retouch- Exp.RAPID- DWI HD- DLR Clear2D RADAR sequence is added to the applicable sequence of DLR Rise.3D sequences and 2D RADAR are added to the applicable sequence of IterativeRAPID.VASC-FSE is modified.See Table 7
Pulse SequencesNoneNo
Monitoring ToolsFollowing function is added.See Table 7

Table 6 Comparison: Functionality

The functionality differences from the predicate device to the ECHELON Synergy V10.0 MRI System are analyzed in Table 7.

Table 7 Functionality Comparison Analysis

FDARequirementsAnalyze why any differences between the subject device and predicate(s) do not render the device NSE (e.g., does notconstitute a new intended use; and any differences in technological characteristics are accompanied by information thatdemonstrates the device is as safe and effective as the predicate and do not raise different questions of safety andeffectiveness than the predicate ), affect safety or effectiveness, or raise different questions of safety and effectiveness (seesection 513(i)(1)(A) of the FD&C Act and 21 CFR 807.87(f)).
DeviceModificationSummary· Following function is updated in Scan Tasks.- AutoVoice· Following positioning applications are added in Scan Tasks.- AutoPose Breast, AutoPose HipJoint

{9}------------------------------------------------

- Following positioning application in Scan Tasks is updated.- AutoPose Spine- Following function is added.- Series Save- Following Network Tool is added.- AutoProtocol- Following variation of Protocol Enhancements are added in Auto Table Centering.- Chest, Breast, General Abdomen, Liver/Kidney, Pelvis, WholeBody- Following function of Protocol Enhancements is updated.- Delayed Enhancement Imaging- Navi slice positioning- BeamNavi- Following Protocol Enhancements are added.
- Navigated StillShot- Visual StillShot- MROC- IQ Retouch- Exp.RAPID- DWI HD- DLR Clear- Following task is added as "Monitoring Tools".- Synergy Vision- Function name of Deep Learning Reconstruction (DLR) added in ECHELON Synergy MRI System (K223426) is changedDLR Rise. DLR is an abbreviation for Deep Learning Reconstruction.- 2D RADAR is added in "DLR Rise" of Protocol Enhancements.- 3D sequences and 2D RADAR are added in "IterativeRAPID" of Protocol Enhancements.- VASC-FSE of Protocol Enhancements is modified.Application software is changed in V10.0A.
SignificantChanges□ Manufacturing Process□ Labeling□ Technology□ Performance
□ Engineering□ Materials□ Others☑ None (See rationale statement)
FUJIFILMRationaleStatementModified functions do not constitute a new intended use. There are no significant changes in technological characteristics.For safety, pulse sequences are controlled according to the same safety limits as ECHELON Synergy MRI System(K223426). Therefore, safety and effectiveness of the device are the same as ECHELON Synergy MRI System (K223426).

{10}------------------------------------------------

Substantial Equivalence

A summary decision was based on analysis of Table 8.

Table 8 Rationale Analysis: ECHELON Synergy V10.0 vs. Predicate
ITEMOverall Rationale Analysis
HardwareModified specification doesn't constitute a new intended use. There are no significant changes in technologicalcharacteristics. For safety, gradient system and RF system is controlled according to same regulation as ECHELON SynergyMRI System (K223426). So, safety and effectively of the device are same as ECHELON Synergy MRI System (K223426).
CoilsThere are no differences regarding receiver coils.
FunctionalityModified functions do not constitute a new intended use. There are no significant changes in technological characteristics.For safety, pulse sequences are controlled according to the same safety limits as ECHELON Synergy MRI System(K223426). Therefore, safety and effectiveness of the device are the same as ECHELON Synergy MRI System (K223426)

Therefore, based on a thorough analysis and comparison of the functions, scientific concepts, physical and performance characteristics, performance comparison and technological characteristics, the proposed ECHELON Synergy V10.0 is considered substantially equivalent to the currently marketed predicate device (ECHELON Synergy MRI System (K223426)) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.

{11}------------------------------------------------

Summary of Non-Clinical Testing

The ECHELON Synergy MRI System was subjected to the following laboratory testing.

  • ANSI / AAMI ES60601-1:2005/(R) 2012 and A1:2012. c1:2009/(r) 2012 and A2:2010/(R) 2012 . (consolidated text) medical electrical equipment - part 1: general requirements for basic safety and essential performance (IEC 60601-1:2005, mod).
  • IEC 60601-2-33 Edition 3.2 b:2015, medical electrical equipment part 2-33: particular . requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnostic.
  • . IEC 62304 Edition 1.1 2015-06, CONSOLIDATED VERSION medical device software - software life cycle processes.
  • . IEC 60601-1-2 Edition 4.0:2014. medical electrical equipment - part 1-2: general requirements for basic safety and essential performance - collateral standard: electromagnetic disturbances requirements and tests.

The revisions to the ECHELON Synergy V10.0 MRI System will have no effect on the standards tests, which were conducted on the ECHELON Synergy MRI System (K223426) and included in the original submission.

Therefore, ECHELON Synergy V10.0 MRI System is in conformance with the applicable parts of the following standards:

  • . NEMA MS 1-2008, Determination of Signal-to-noise Ratio (SNR) in Diagnostic Magnetic Resonance Images
  • NEMA MS 2-2008, Determination of Two-Dimensional Geometric Distortion in Diagnostic . Magnetic Resonance Images
  • NEMA MS 3-2008, Determination of Image Uniformity in Diagnostic Magnetic Resonance Images
  • NEMA MS 4-2010, Acoustic Noise Measurement Procedure for Diagnostic Magnetic Resonance ● Imaging Devices
  • NEMA MS 5-2018, Determination of Slice Thickness in Diagnostic Resonance Imaging .
  • NEMA MS 8-2016, Characterization of the Specific Absorption Rate for Magnetic Resonance Imaging Systems
  • NEMA MS 14-2019, Characterization of Radiofrequency (RF) Coil Heating in Magnetic . Resonance Imaging Systems
  • . IEC 60825-1:2014, safety of laser products – part 1: equipment classification and requirements.

Summary of Clinical Testing

Clinical images were collected and analyzed, to ensure that images from the new feature meet user needs.

The validation results of the new features using machine learning (DLR Clear and AutoPose) were described below.

Performance tests of DLR Clear were conducted by using cylindrical phantom and ACR phantom in terms of truncation artifact reduction, improvement of image sharpness, and improvement of spatial resolution with the sampling patterns of 2D Cartesian, 2D Radial, 3D Rectanqular, and 3D Elliptic. Using metrics of Total Validation, Relative Edge Sharpness, and FWHM (Full Width at Half Minimum) of the line-profile of the holes in the ACR phantom, the phantom testing demonstrated that DLR Clear could reduce the truncation artifact in the image, improve image sharpness, and improve the spatial resolution of image. In addition, three US board certified radiologists reviewed the image quality with DLR Clear in terms of the truncation artifact reduction, image sharpness, spatial resolution of image. The

{12}------------------------------------------------

images reconstructed with either the conventional method or DLR Clear were randomized, blinded to the reviewers, and compared by the reviewers in terms of image quality metrics (truncation artifact reduction, image sharpness, lesion conspicuity, and overall image quality) using 3-point scale. All images used for this comparison were also evaluated by the reviewers in terms of clinical acceptability. The 53 unique subjects (patients and volunteers) from U.S. and Japan (Male: 33, Female: 20, Age: 9 - 92 years, BMI:18.9 - 50.5) were scanned in the anatomical regions from brain to ankle in order to provide the test datasets separately from the training and validation datasets using FUJIFILM 1.5T MRI Scanners (ECHELON Smart, ECHELON OVAL, and ECHELON Synergy). The total of 169 images in multiple orientations (axial, sagittal and coronal), multiple dimensions (2D and various contrast weightings (T1-/T2-/T2*-/PD-weighted image with/without Fat saturation. FLAIR, DWI, MRA, MRCP, STIR, and Cine) were obtained for the test dataset by pulse sequences of SE, GE, FSE, FIR, BASG, RSSG, EPI, TOF, and PC. The review results indicated that the truncation artifact reduction, image sharpness, and overall image quality in the images with DLR Clear were superior to those in the conventional images with statistically significant difference (p<0.05). The review results also indicated that the lesion conspicuity in the images with DLR Clear was superior or equivalent to that in the conventional images. All of the images with DLR Clear were evaluated as clinically acceptable by the reviewers. The comparison of the overall image quality between the high-resolution images with DLR Clear and the low-resolution images without DLR Clear was also performed by the radiologists. The pairs of the high-resolution images with DLR Clear and the low-resolution images without DLR Clear, reconstructed from the same acquired data, were prepared from the test dataset. All images used for this comparison were evaluated by the radiologists in terms of clinical acceptability. The review results indicated that the overall image quality in the high-resolution images with DLR Clear were superior to that in the low-resolution images without DLR Clear with statistically significant difference and were clinically acceptable. In conclusion, the phantom testing and the clinical Image testing demonstrated that DLR Clear could reduce the truncation artifact in the image, improve image sharpness, and generate higher spatial resolution images from lower resolution images.

The performance tests of AutoPose Spine, Breast and HipJoint, were conducted by the certified radiological technologists. They evaluated that almost cases in AutoPose Spine, Breast, and HipJoint, were able to reduce the time and number of steps in the slice positioning compared to the manual slice positioning. They also evaluated that the remaining cases of AutoPose Spine, Breast, and HipJoint were able to show the same time and number of steps as the manual slice positioning. The information on the data in the above evaluations is shown below.

SpineBreastHipJoint
Data acquisition siteFUJIFILM Healthcare Corporation and clinical site
Subject typeHealthy volunteer and patients
Number of cases1776665

As a result of the analysis:

Testing TypeRationale Analysis
Performance Testing –ClinicalClinical image examples are provided for applicable new feature and that we judged to be sufficient to evaluate clinical usability.

{13}------------------------------------------------

Conclusions

The ECHELON Synergy V10.0 MRI system is substantially equivalent with respect to hardware, base elements of the software, safety, effectiveness, and functionality to the ECHELON Synergy MRI System (K223426).

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.