K Number
K233611

Validate with FDA (Live)

Date Cleared
2024-03-26

(134 days)

Product Code
Regulation Number
866.3920
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SPOTFIRE® RSP Positive Control is intended for use (as applicable) as an external positive assayed quality control to monitor performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Bordetella parapertussis, Bordetella pertussis, Chlamydia pneumoniae, Mycoplasma pneumoniae, Streptococcus dysgalactiae (Group C/G Strep), Streptococcus pyogenes (Group A Strep), Adenovirus SARS-CoV-2, Coronavirus (seasonal), Human Metapneumovirus, Human Rhinovirus, Influenza A Subtype H1-2009, Influenza A Subtype H3, Influenza B, Parainfluenza Virus and Respiratory Syncytial Virus using the BIOFIRE® SPOTFIRE Respiratory (R) Panel, BIOFIRE SPOTFIRE Respiratory (R) Panel Mimi, and BIOFIRE SPOTFIRE Respiratory/Sore Throat (R/ST) Panel on the BIOFIRE SPOTFIRE System. SPOTFIRE RSP Positive Control is comprised of in vitro RNA transcripts and stabilizing solution and is designed for and intended to be used solely with the BIOFIRE SPOTFIRE R Panel, BIOFIRE SPOTFIRE R Panel Mini, and BIOFIRE SPOTFIRE R/ST Panel assays. This product is not intended to replace manufacturer internal controls provided with these devices.

Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements.

SPOTFIRE® RSP Negative Control is intended for use (as applicable) as an external negative assayed quality control to monitor performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Bordetella parapertussis, Bordetella pertussis, Chlamydia pneumoniae, Streptococcus dysgalactiae (Group C/G Strep), Streptococcus pyogenes (Group A Strep), Adenovirus SARS-CoV-2, Coronavirus (seasonal), Human Metapneumovirus, Human Rhinovirus, Influenza A Subtype H1-2009, Influenza A Subtype H3, Influenza B, Parainfluenza Virus and Respiratory Syncytial Virus using the BIOFIRE® SPOTFIRE Respiratory (R) Panel, BIOFIRE SPOTFIRE Respiratory (R) Panel Mini, and BIOFIRE SPOTFIRE Respiratory/Sore Throat (R/ST) Panel on the BIOFIRE System. SPOTFIRE RSP Negative Control is comprised of a solution that does not contain target and is designed for and intended to be used solely with the BIOFIRE SPOTFIRE R Panel, BIOFIRE SPOTFIRE R Panel Mini, and BIOFIRE SPOTFIRE R/ST Panel assays. This product is not intended to replace manufacturer internal controls provided with these devices.

Device Description

SPOTFIRE RSP Pos & Neg Controls, P/N M425, is a quality control panel consisting of 2 readyto-use, liquid controls, SPOTFIRE RSP Positive Control (Positive Control), P/N M42638 and SPOTFIRE RSP Negative Control (Negative Control), P/N M42738. The Positive Control contains non-infectious surrogate control material; a solution of synthetic RNA transcripts in buffer, stabilizers and preservatives. The RNA in the Positive Control carries RNA segments of all the respiratory and sore throat pathogens detected by the BIOFIRE SPOTFIRE Respiratory (R) Panel, BIOFIRE SPOTFIRE Respiratory (R) Panel Mini, and BIOFIRE SPOTFIRE Respiratory/Sore Throat (R/ST) Panel assays, performed on the BIOFIRE System. The Negative Control contains buffer and preservatives with no RNA. Each liquid control of SPOTFIRE RSP Pos & Neg Controls is processed separately according to the SPOTFIRE Panel assay manufacturer's Instructions for Use for patient samples (nasopharyngeal swabs or throat swabs placed in transport media) obtained from individuals with signs and symptoms of respiratory tract infection or pharyngitis.

AI/ML Overview

The provided text is a 510(k) Summary for a medical device called "SPOTFIRE RSP Pos & Neg Controls," which are quality control materials for molecular diagnostic assays. The submission is a change to an existing device (K230868) to include compatibility with a new panel, the "BIOFIRE SPOTFIRE Respiratory/Sore Throat (R/ST) Panel."

The document details the device's intended use, its description, and a summary of performance data collected to demonstrate that the modified device remains substantially equivalent to its predicate.

Here's the breakdown of the acceptance criteria and study information, as requested:

Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied)Reported Device Performance
Overall Correct Result Rate for the ControlsOverall Correct Result Rate: 99.7% (351 out of 351 valid results were correct). This includes both positive and negative controls across all sites (internal and external).
Reproducibility for Positive Control (Percent Correct)Positive Control: 99.3% correct. (171 out of 172 valid positive control results were correct. One positive control failed due to false negative results for 4 analytes.)
Reproducibility for Negative Control (Percent Correct)Negative Control: 100% correct. (179 out of 179 valid negative control results were correct.)
Performance at External Sites (Overall Correct Result Rate)External Sites Overall: 99.6% correct. (224 out of 225 valid external site results were correct.)
Performance at Internal Site (Overall Correct Result Rate)Internal Site Overall: 100% correct. (126 out of 126 valid internal site results were correct.)
Robustness across various conditions (pouch lots, operators, instruments, sites)The study incorporated: - 3 control lots - 27 pouch lots (across all sites, including one shared between internal and external) - Multiple operators (specified as "multiple" for external sites, 4 for internal) - Multiple instruments (specified as "multiple" for external sites, 1 for internal) - 6 testing sites (5 external clinical sites + 1 internal MMQCI site) - Testing over time (21 different days for internal testing)

Study Information

  1. Sample Size Used for the Test Set and Data Provenance:

    • Total Test Results: 361 initially, with 10 invalid results that were repeated, leading to 351 valid test results used for performance analysis.
    • Data Provenance:
      • External Study: 232 test results collected from 5 near-patient clinical sites. The document does not specify the country of origin of these sites, but given the FDA submission, it's presumed to be in the USA. This was a prospective study as the sites were asked to follow specific protocols for testing controls during normal operations (e.g., training new operators, new kit shipments, new/repaired instruments).
      • Internal Study: 129 test results collected from MMQCI's facility in Saco, Maine (USA). This was also a prospective study.
  2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

    • This study evaluates the performance of quality control materials for in vitro laboratory nucleic acid testing. The "ground truth" here is objective: either the control produced the expected positive result (detecting the specified analytes) or the expected negative result (no target analytes detected).
    • Therefore, no human experts (like radiologists) were establishing a subjective "ground truth" diagnosis. The performance is based on whether the automated instrument correctly identifies the presence or absence of the target analytes in the control material as designed.
  3. Adjudication Method for the Test Set:

    • Not applicable in the conventional sense of human expert discrepancies. The performance is measured against the pre-defined composition of the positive and negative control materials.
    • Invalid results (instrument errors, failed internal pouch controls) were re-tested according to BioFire instructions and excluded from the "Percent Correct" analysis. This acts as a form of "adjudication" for technical failures rather than diagnostic disagreement.
  4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No, an MRMC comparative effectiveness study was not done. This study is not evaluating a diagnostic AI algorithm that assists human readers. It's evaluating the performance of quality control materials for a molecular diagnostic assay. The "users" are laboratory professionals (or non-lab professionals in CLIA waived settings) operating the BIOFIRE SPOTFIRE System. The study assessed the reproducibility of the control materials, not the improvement of human readers with AI assistance.
  5. If a Standalone (algorithm only without human-in-the-loop performance) was done:

    • This is not an AI algorithm. The performance evaluation is inherently "standalone" in the sense that the control materials are processed by the BIOFIRE SPOTFIRE System, and the system's output (positive or negative detection of specific analytes) is compared against the known composition of the control. The "human-in-the-loop" would be the technician performing the test, but their role is to follow instructions, not to interpret complex images or data subjectively.
  6. The Type of Ground Truth Used:

    • The ground truth is the known, manufactured composition of the control materials.
      • Positive Control: Contains synthetic RNA transcripts for all target respiratory and sore throat pathogens. The ground truth is that these specific targets should be detected by the assay.
      • Negative Control: Contains buffer and preservatives with no RNA targets. The ground truth is that no target analytes should be detected.
  7. The Sample Size for the Training Set:

    • Not applicable. This device is a quality control material, not a machine learning or AI model. Therefore, there is no "training set" in the context of algorithm development. The control materials are manufactured to a specific design.
  8. How the Ground Truth for the Training Set was Established:

    • Not applicable, as there is no training set for an AI/ML model. The ground truth for the control materials is inherent in their design and manufacturing process.

{0}------------------------------------------------

March 26, 2024

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized eagle and text encircling it. To the right is the FDA logo, with the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

Maine Molecular Quality Controls, Inc. Joan Gordon President 23 Mill Brook Rd Saco, Maine 04072

Re: K233611

Trade/Device Name: SPOTFIRE RSP Pos & Neg Controls (M425), SPOTFIRE RSP Positive Control (M42638), SPOTFIRE RSP Negative Control (M42738) Regulation Number: 21 CFR 866.3920 Regulation Name: Assayed Quality Control Material For Clinical Microbiology Assays Regulatory Class: Class II Product Code: PMN Dated: November 3, 2023 Received: November 13, 2023

Dear Joan Gordon:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely. Joseph Briggs -S

Joseph Briggs Deputy Branch Chief Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K233611

Device Name SPOTFIRE RSP Pos & Neg Controls (M425); SPOTFIRE RSP Positive Control (M42638); SPOTFIRE RSP Negative Control (M42738)

Indications for Use (Describe) SPOTFIRE RSP Positive Control:

SPOTFIRE® RSP Positive Control is intended for use (as applicable) as an external positive assayed quality control to monitor performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Bordetella parapertussis, Bordetella pertussis, Chlamydia pneumoniae, Mycoplasma pneumoniae, Streptococcus dysgalactiae (Group C/G Strep), Streptococcus pyogenes (Group A Strep), Adenovirus SARS-CoV-2, Coronavirus (seasonal), Human Metapneumovirus, Human Rhinovirus, Influenza A Subtype H1-2009, Influenza A Subtype H3, Influenza B, Parainfluenza Virus and Respiratory Syncytial Virus using the BIOFIRE® SPOTFIRE Respiratory (R) Panel, BIOFIRE SPOTFIRE Respiratory (R) Panel Mimi, and BIOFIRE SPOTFIRE Respiratory/Sore Throat (R/ST) Panel on the BIOFIRE SPOTFIRE System. SPOTFIRE RSP Positive Control is comprised of in vitro RNA transcripts and stabilizing solution and is designed for and intended to be used solely with the BIOFIRE SPOTFIRE R Panel, BIOFIRE SPOTFIRE R Panel Mini, and BIOFIRE SPOTFIRE R/ST Panel assays. This product is not intended to replace manufacturer internal controls provided with these devices.

Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements.

SPOTFIRE RSP Negative Control:

SPOTFIRE® RSP Negative Control is intended for use (as applicable) as an external negative assayed quality control to monitor performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Bordetella parapertussis, Bordetella pertussis, Chlamydia pneumoniae, Streptococcus dysgalactiae (Group C/G Strep), Streptococcus pyogenes (Group A Strep), Adenovirus SARS-CoV-2, Coronavirus (seasonal), Human Metapneumovirus, Human Rhinovirus, Influenza A Subtype H1-2009, Influenza A Subtype H3, Influenza B, Parainfluenza Virus and Respiratory Syncytial Virus using the BIOFIRE® SPOTFIRE Respiratory (R) Panel, BIOFIRE SPOTFIRE Respiratory (R) Panel Mini, and BIOFIRE SPOTFIRE Respiratory/Sore Throat (R/ST) Panel on the BIOFIRE System. SPOTFIRE RSP Negative Control is comprised of a solution that does not contain target and is designed for and intended to be used solely with the BIOFIRE SPOTFIRE R Panel, BIOFIRE SPOTFIRE R Panel Mini, and BIOFIRE SPOTFIRE R/ST Panel assays. This product is not intended to replace manufacturer internal controls provided with these devices.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for MMQCI, which stands for Maine Molecular Quality Controls, Inc. The logo features the letters "MMQCI" in a combination of maroon and yellow colors. The "MM" is in maroon, while the "QCI" is in yellow. The full company name is written in smaller font below the letters. A yellow swoosh design is behind the letters.

Section 5: 510 (k) Summary

510(k) Number:

Purpose for submission: Change to existing device (reference K230868) to include the BIOFIRE® SPOTFIRE® Respiratory/Sore Throat (R/ST) Panel (SPOTFIRE R/ST Panel) in the intended use for SPOTFIRE RSP Pos & Neg Controls.

Applicant Information:

Applicant:Maine Molecular Quality Controls, Inc
Address:23 Mill Brook RoadSaco, Maine 04072
Contact Person:Joan Gordon, President MMQCI
Phone:207-885-1072 extension 201
Fax:207-885-1079
Email Address:jgordon@mmqci.com

Preparation Date: March 19, 2024

Device

Device Trade Name:SPOTFIRE RSP Pos & Neg Controls, model M425SPOTFIRE RSP Positive Control, model M42638SPOTFIRE RSP Negative Control, model M42738
Device Common Name:Quality Control Material for Microbiology Assays
Device Type:Assayed quality control material for clinical microbiologyassays
Class:Class II (Special controls)
Regulation:21 CFR 866.3920
Panel:Microbiology - 83
Product code:PMN

Predicate Device

K230868 SPOTFIRE RSP Pos & Neg Controls SPOTFIRE RSP Positive Control SPOTFIRE RSP Negative Control

Proposed Change:

The specific change to the existing device is the inclusion of the SPOTFIRE R/ST Panel in the intended use for SPOTFIRE RSP Pos & Neg Controls, with subsequent updates of the Indications of Use (FDA Form 3881), Package Inserts for SPOTFIRE RSP Positive Control and

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Maine Molecular Quality Controls, Inc. The logo features the letters "MMQCI" in a combination of maroon and yellow. The "MM" is in maroon, while the "QCI" is in yellow. The text "MAINE MOLECULAR QUALITY CONTROLS, INC." is written in a smaller font below the letters. There is also a yellow swirl design on the left side of the logo.

SPOTFIRE RSP Negative Control, and the addition of data for SPOTFIRE R/ST Panel analytes Streptococcus dysgalactiae (Group C/G Strep) and Streptococcus pyogenes (Group A Strep) to this submission. No other changes have been made to the original device referenced in K230868.

Description of Modified Device (inclusion of use with SPOTFIRE R/ST Panel)

SPOTFIRE RSP Pos & Neg Controls, P/N M425, is a quality control panel consisting of 2 readyto-use, liquid controls, SPOTFIRE RSP Positive Control (Positive Control), P/N M42638 and SPOTFIRE RSP Negative Control (Negative Control), P/N M42738. The Positive Control contains non-infectious surrogate control material; a solution of synthetic RNA transcripts in buffer, stabilizers and preservatives. The RNA in the Positive Control carries RNA segments of all the respiratory and sore throat pathogens detected by the BIOFIRE SPOTFIRE Respiratory (R) Panel, BIOFIRE SPOTFIRE Respiratory (R) Panel Mini, and BIOFIRE SPOTFIRE Respiratory/Sore Throat (R/ST) Panel assays, performed on the BIOFIRE System. (see Tables 1-3 below) SPOTFIRE RSP Pos & Neg Controls are specifically designed for and intended to be used solely with the BIOFIRE SPOTFIRE R Panel, BIOFIRE SPOTFIRE R Panel Mini, and BIOFIRE SPOTFIRE R/ST Panel assays on the BIOFIRE SPOTFIRE System. The Negative Control contains buffer and preservatives with no RNA. Each liquid control of SPOTFIRE RSP Pos & Neg Controls is processed separately according to the SPOTFIRE Panel assay manufacturer's Instructions for Use for patient samples (nasopharyngeal swabs or throat swabs placed in transport media) obtained from individuals with signs and symptoms of respiratory tract infection or pharyngitis.

Table 1. Respiratory pathogens detected by BIOFIRE SPOTFIRE R Panel
VirusesBacteria
AdenovirusChlamydia pneumoniae
Coronavirus SARS-CoV-2Mycoplasma pneumoniae
Coronavirus (seasonal)Bordetella parapertussis
Human MetapneumovirusBordetella pertussis
Human Rhinovirus/ Enterovirus
Influenza A Virus
Influenza A virus A/H1-2009
Influenza A virus A/H3
Influenza B Virus
Parainfluenza
Respiratory Syncytial Virus

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Maine Molecular Quality Controls, Inc. The logo features the acronym "MMQCI" in a stylized font, with the "MM" in a dark red color and the "QCI" in yellow. A golden, abstract design element curves around the letters, adding a visual flair to the logo. Below the acronym, the full name of the company, "MAINE MOLECULAR QUALITY CONTROLS, INC.," is written in a smaller, dark font.

Table 2. Respiratory viruses detected by theBIOFIRE SPOTFIRE R Panel Mini
Coronavirus SARS-CoV-2
Human Rhinovirus
Influenza A Virus
Influenza B Virus
Respiratory Syncytial Virus
Table 3. Respiratory and sore throat pathogens detected by BIOFIRE SPOTFIRE R/ST Panel
Viruses (Respiratory and Sore Throat)Bacteria (Respiratory and Sore Throat)
AdenovirusChlamydia pneumoniae
Coronavirus SARS-CoV-2Mycoplasma pneumoniae
Coronavirus (seasonal)Bacteria (Respiratory Only)
Human MetapneumovirusBordetella parapertussis
Human Rhinovirus/ EnterovirusBordetella pertussis
Influenza A VirusBacteria (Sore Throat Only)
Influenza A virus A/H1-2009Streptococcus dysgalactiae (group C/G Strep)
Influenza A virus A/H3Streptococcus pyogenes (group A Strep)
Influenza B Virus
Parainfluenza
Respiratory Syncytial Virus

Intended Use of SPOTFIRE RSP Pos & Neg Controls modified to include the addition SPOTFIRE R/ST Panel

SPOTFIRE RSP Positive Control:

SPOTFIRE RSP Positive Control is intended for use (as applicable) as an external positive assayed quality control to monitor performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Bordetella parapertussis, Bordetella pertussis, Chlamydia pneumoniae, Mycoplasma pneumoniae, Streptococcus dysgalactiae (Group C/G Strep), Streptococcus pyogenes (Group A Strep), Adenovirus, Coronavirus SARS-CoV-2, Coronavirus (seasonal), Human Metapneumovirus, Human Rhinovirus/Enterovirus, Influenza A Subtype H1-2009, Influenza A Subtype H3, Influenza B, Parainfluenza Virus, and Respiratory Syncytial Virus using the BIOFIRE SPOTFIRE Respiratory (R) Panel, BIOFIRE SPOTFIRE Respiratory (R)

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Maine Molecular Quality Controls, Inc. The logo features the acronym "MMQCI" in bold, with "MM" in a dark red color and "QCI" in a yellow color. Below the acronym, the full name of the company, "MAINE MOLECULAR QUALITY CONTROLS, INC.," is written in a smaller, dark red font. A stylized, curved yellow shape is on the left side of the logo.

Panel Mini, and BIOFIRE SPOTFIRE Respiratory/Sore Throat (R/ST) Panel on the BIOFIRE SPOTFIRE System. SPOTFIRE RSP Positive Control is comprised of in vitro RNA transcripts and stabilizing solution and is designed for and intended to be used solely with the BIOFIRE SPOTFIRE R Panel, BIOFIRE SPOTFIRE R Panel Mini, and BIOFIRE SPOTFIRE R/ST Panel assays. This product is not intended to replace manufacturer internal controls provided with these devices.

Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements.

SPOTFIRE RSP Negative Control:

SPOTFIRE RSP Negative Control is intended for use (as applicable) as an external negative assayed quality control to monitor performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Bordetella parapertussis, Bordetella pertussis, Chlamydia pneumoniae, Mycoplasma pneumoniae, Streptococcus dysgalactiae (Group C/G Strep), Streptococcus pyogenes (Group A Strep), Adenovirus, Coronavirus SARS-CoV-2, Coronavirus (seasonal), Human Metapneumovirus, Human Rhinovirus/Enterovirus, Influenza A Subtype H1-2009, Influenza A Subtype H3, Influenza B, Parainfluenza Virus, and Respiratory Syncytial Virus using the BIOFIRE SPOTFIRE Respiratory (R) Panel, BIOFIRE SPOTFIRE Respiratory (R) Panel Mini, and BIOFIRE SPOTFIRE Respiratory/Sore Throat (R/ST) Panel on the BIOFIRE SPOTFIRE System. SPOTFIRE RSP Negative Control is comprised of a solution that does not contain target analytes and is designed for and intended to be used solely with the BIOFIRE SPOTFIRE R Panel, BIOFIRE SPOTFIRE R Panel Mini, and BIOFIRE SPOTFIRE R/ST Panel assays. This product is not intended to replace manufacturer internal controls provided with these devices.

Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements.

Substantial Equivalence

Maine Molecular Quality Controls, Inc. (MMQCI) proposes that the changes made to the labeling of SPOTFIRE RSP Pos & Neg Controls to include the additional intended use of SPOTFIRE R/ST Panel do not impact substantial equivalence to SPOTFIRE RSP Pos & Neg Controls (K230868). There have been no changes to the predicate device, other than labeling. The rationale for equivalence is listed in Table 4 below.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Maine Molecular Quality Controls, Inc. The logo features the acronym "MMQCI" in a combination of maroon and yellow colors. The full name of the company, "Maine Molecular Quality Controls, Inc.," is written in a smaller font size below the acronym. A stylized yellow graphic element is placed to the left of the text, adding a visual accent to the logo.

Table 4. Substantial Equivalence of Predicate Devices

CharacteristicCandidate Device:SPOTFIRE RSP Pos &Neg ControlsPredicate Device: SPOTFIRERSP Pos & Neg Controls(K230868)
Intended UseExternal assayed qualitycontrol to monitor in vitrolab nucleic acid test usingthe SPOTFIRE R Panel,SPOTFIRE R Panel Mini,and SPOTFIRE R/ST Panelperformed on theSPOTFIRE System.Original 510(k) IFU didnot include theSPOTFIRE R/ST Panel
LabelingSPOTFIRE R/ST Paneladded to the Package Insert,no changes to Kit or Tubelabels.SPOTFIRE R/ST Panel notincluded in Package Insert
Physical formatReady-to-Use LiquidSame
Directions for UseProcess like patient sampleSame
CompositionSynthetic RNA transcriptsSame
Assay stepsmonitoredReverse transcription,amplification, detection,identificationSame
Test SystemSPOTFIRESame
Number of targetsmonitored in oneassayMultipleOriginal 510(k) data did notinclude the sore throat targetsdetected by SPOTFIRE R/STPanel
Pathogenscontained in thecontrol panelRespiratory and Sore ThroatRespiratory
UserNon-lab (CLIA Waived)and lab professionals (CLIAcertified)Same

Summary of Performance Data

Three lots of SPOTFIRE RSP Pos & Neg Controls were manufactured by MMQCI at MMQCI's facility in Saco, Maine. The lots were manufactured and tested such that routine variables,

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Maine Molecular Quality Controls, Inc. (MMQCI). The logo features the letters "MMQCI" in a combination of maroon and yellow colors. The word "MAINE MOLECULAR QUALITY CONTROLS, INC." is written in a smaller font below the letters. A yellow, abstract design element is placed to the left of the text.

including lots of key manufacturing components, different operators, pouch lots, instruments, test sites, and testing over time were incorporated to challenge manufacture processes and product performance. External and internal studies were performed by testing the 3 lots with the SPOTFIRE R/ST Panel assay on the SPOTFIRE System (BioFire Diagnostics) to generate a total of 361 test results.

For the external performance study, 5 near-patient clinical sites tested samples of each of the 3 lots of SPOTFIRE RSP Positive Controls and SPOTFIRE RSP Negative Controls as they would test controls during normal operations, using 25 SPOTFIRE R/ST Panel pouch lots, incorporating multiple operators and instruments for a total of 232 tests, including 7 Invalid results which were repeated. The internal study was performed by testing samples of each of the 3 SPOTFIRE RSP Pos & Neg Controls lots at MMQCI (Saco, Maine) on different days with 3 pouch lots by 3 operators for a total of 129 tests (3 Invalid results were repeated). One of the pouch lots used at the MMOCI site was also used at a clinical site.

Results and Conclusions:

Of the total 361 tests performed, there were 10 Invalid results which repeated for 351 valid test results. The Invalid results included instrument errors (3) and failed internal pouch controls (7). The Invalid samples were re-tested according to BioFire instructions and were not included in the Percent Correct analysis in Table 5 below. Of the 351 valid results were correct. One SPOTFIRE RSP Positive Control failed due to false negative results for 4 analytes.

Reproducibility of SPOTFIRE RSP Pos & Neg Controls at 6 sites, across 27 pouch lots, 3 control lots, on multiple instruments by multiple operators is acceptable according to predetermined criteria with an overall correct result rate of 99.7%. Test results demonstrate robust performance across 6 testing sites.

Summary of Reproducibility Test Results: External and Internal Sites
TotalTestsInvalid*CorrectPositiveControlResultIncorrectPositiveControlResultPercentCorrectPositiveControlCorrectNegativeControlResultIncorrectNegativeControlResultPercentCorrectNegativeControlTotalPercentCorrect
36110171199.3%1790100%99.7%

Table 5. SPOTFIRE RSP Pos & Neg Controls: Reproducibility results, all sites.

*The Invalid samples were not included in the Percent Correct analysis.

External Site Testing Details

A total of 232 samples of SPOTFIRE RSP Positive and RSP Negative Controls were tested at 5 near-patient clinical sites. The sites were asked to follow the SPOTFIRE R/ST Panel IUO Quick Guide to run the controls as they would during normal operations, e.g. training a new operator, receiving new shipments of R/ST Panel test kits, receiving new or repaired instrumentation, and

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for MMQCI, which stands for Maine Molecular Quality Controls, Inc. The logo features the letters "MMQCI" in a bold, serif font, with "MM" in a larger size and a darker red color, and "QCI" in a smaller size and a lighter red color. Below the letters, the full name of the company, "MAINE MOLECULAR QUALITY CONTROLS, INC.", is written in a smaller, sans-serif font. A stylized, golden yellow swirl design is placed to the left of the letters, adding a visual element to the logo.

whenever the test kits have been stored at an incorrect temperature. It was recommended a negative quality control sample be run at least monthly to monitor for environmental contamination. The amount of QC a site performed was based on the number of operators they had and the amount of SPOTFIRE R/ST Panel testing their site performed.

Three lots of SPOTFIRE RSP Positive Control (A02NOV20, B06NOV20, C10NOV20) and 4 lots of SPOTFIRE RSP Negative Control (F09JUL19, D11MAR20, H05MAY20, C13MAY20) were tested on 25 SPOTFIRE R/ST Panel pouch lots across the 5 sites, incorporating multiple operators and instruments. Of the 232 samples tested, there were 7 Invalid results which were repeated for a total of 225 valid results. Of the 7 Invalid results, 4 were failed internal pouch controls and 3 were due to instrument errors.

Results and Conclusion:

Out of 225 tests with valid results, the correct analytes were detected (for positive controls) or not detected (for negative controls) in 224 tests, for an overall success rate of 99.6% (Table 5). One SPOTFIRE RSP Positive Control failed due to false negative results for 4 analytes. The predetermined acceptance criteria for reproducibility were met. SPOTFIRE RSP Positive Control and RSP Negative Control performed robustly at the 5 external test sites across 25 pouch lots on the SPOTFIRE System, incorporating multiple operators.

Summary of External Results for 3 Lots of SPOTFIRE RSP Pos & Neg Controls
SiteTotalTestsInvalidCorrectPositiveControlResultIncorrectPositiveControlResultPercentCorrectPositiveControlCorrectNegativeControlResultIncorrectNegativeControlResultPercentCorrectNegativeControlTotalPercentCorrect
1260130100%130100%100%
2331130100%190100%100%
3120453198.1%620100%99.1%
4441210100%220100%100%
59180100%NANANANA
AllSites2327*108199.1%1160100%99.6%

Table 5. Summary of External Reproducibility Results for 3 Control Lots at 5 Clinical Sites

*Invalid results were not included in percent correct.

Internal Site Testing Details

Run-to-run Precision Testing at MMOCI: Of the total 361 reproducibility tests. 129 SPOTFIRE RSP Positive Control and SPOTFIRE RSP Negative Control samples were tested on 21 different days at MMQCI's facility in Saco, Maine. Three lots of Positive Control (A02NOV20, B06NOV20, C10NOV20) and 3 lots of Negative Control (F09JUL19, D11MAR20, H05MAY20)

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for MMQCI, which stands for Maine Molecular Quality Controls, Inc. The logo features the letters "MMQCI" in a stylized font, with the "MM" in a darker color and the "QCI" in a lighter color. The full name of the company is written in smaller letters below the main logo. There is a yellow abstract design to the left of the letters.

were tested on 3 SPOTFIRE R/ST pouch lots, incorporating 4 operators and 1 SPOTFIRE instrument (Table 6). One of the pouch lots was also used for the external site testing.

Results and conclusions:

Of the 129 results, 3 were Invalid which were repeated for a total of 126 valid results. All 126 of the valid results were correct for an overall correct result rate of 100%. Predetermined acceptance criteria for reproducibility were met. SPOTFIRE RSP Positive Control and SPOTFIRE RSP Negative Control performed robustly on 21 different days at MMQCI across 3 pouch lots on 1 SPOTFIRE instrument, incorporating 4 operators.

ControlControl Lot #No. of TestsInvalid*Correct ResultsIncorrect ResultsPercent Correct
SPOTFIRE RSP Positive ControlA02NOV20232210100%
SPOTFIRE RSP Positive ControlB06NOV20210210100%
SPOTFIRE RSP Positive ControlC10NOV20221210100%
SPOTFIRE RSP Negative ControlF09JUL19210210100%
SPOTFIRE RSP Negative ControlD11MAR20210210100%
SPOTFIRE RSP Negative ControlH05MAY20210210100%
TOTAL12931260100%

Table 6. SPOTFIRE RSP Pos & Neg Controls: Reproducibility results at MMQCL

Overall Conclusion

The above information supports the substantial equivalence of the SPOTFIRE RSP Positive and Negative Controls to the predicate device.

§ 866.3920 Assayed quality control material for clinical microbiology assays.

(a)
Identification. An assayed quality control material for clinical microbiology assays is a device indicated for use in a test system to estimate test precision or to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. This type of device consists of single or multiple microbiological analytes intended for use with either qualitative or quantitative assays.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed device description documentation and information concerning the composition of the quality control material, including, as appropriate:
(i) Analyte concentration;
(ii) Expected values;
(iii) Analyte source;
(iv) Base matrix;
(v) Added components;
(vi) Safety and handling information; and
(vii) Detailed instructions for use.
(2) Premarket notification submissions must include detailed documentation, including line data as well as detailed study protocols and a statistical analysis plan used to establish performance, including:
(i) Description of the process for value assignment and validation.
(ii) Description of the protocol(s) used to establish stability.
(iii) Line data establishing precision/reproducibility.
(iv) Where applicable, assessment of matrix effects and any significant differences between the quality control material and typical patient samples in terms of conditions known to cause analytical error or affect assay performance.
(v) Where applicable, identify or define traceability or relationship to a domestic or international standard reference material and/or method.
(vi) Where applicable, detailed documentation related to studies for surrogate controls.
(3) Premarket notification submissions must include an adequate mitigation (e.g., real-time stability program) to the risk of false results due to potential modifications to the assays specified in the device's 21 CFR 809.10 compliant labeling.
(4) Your 21 CFR 809.10 compliant labeling must include the following:
(i) The intended use of your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following:
(A) Assayed control material analyte(s);
(B) Whether the material is intended for quantitative or qualitative assays;
(C) Stating if the material is a surrogate control; and
(D) The system(s), instrument(s), or test(s) for which the quality control material is intended.
(ii) The intended use in your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following statement: “This product is not intended to replace manufacturer controls provided with the device.”
(iii) A limiting statement that reads “Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements.”