(90 days)
The MAGEC Spinal Bracing and Distraction System is indicated for skeletally immature patients with severe progressive spinal deformities (e.g., Cobb angle of 30 degrees or more; thoracic spine height less than 22 cm) secondary to early-onset scoliosis associated with or at risk of Thoracic Insufficiency Syndrome(TIS). TIS is defined as the inability of the thorax to support normal respiration or lung growth.
The subject MAGEC Spinal Bracing and Distraction System has an identical design and principle of operation to the predicate design iterations cleared in the predicate Magec System (K201543, K161751, K140613). The subject system includes sterile single use MAGEC rods manufactured from Ti-6Al-4V, conforming to ASTM F136, along with various accessories including a sterile Rod Template and Manual Distractor, a non-sterile Wand Magnet Locator, and is compatible with a hand held External Remote Controller (ERC) 1 or 2. The MAGEC rod can be surgically implanted using appropriate NuVasive Reline 4.5-5.0 (Reline Small Stature or RSS) or Armada fixation components (i.e., pedicle screws, hooks and/or connectors). The titanium MAGEC rod includes an actuator portion that holds a small internal magnet. The magnet in the actuator can be turned non- invasively by use of the ERC, which is electrically powered. The hand-held non-invasive ERC is placed over the patient's spine and then manually activated, which causes the implanted magnet to rotate and either distract or retract the rod. Periodic distraction of the rod is performed to lengthen the spine and to provide adequate bracing during growth to minimize the progression of scoliosis. Once the physician determines that the implant has achieved its intended use and is no longer required, the implant is explanted.
The provided text describes a 510(k) premarket notification for the MAGEC Spinal Bracing and Distraction System, focusing on expanding its indications for use. It primarily demonstrates substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria in the typical sense of a diagnostic or AI device.
However, based on the information provided, we can infer "acceptance criteria" from the measured clinical outcomes that are presented to support the expanded indication. The device's performance is demonstrated through a retrospective registry study comparing clinical and radiographic data.
Here's a breakdown of the requested information:
1. A table of acceptance criteria and the reported device performance
The document doesn't explicitly state "acceptance criteria" with numerical thresholds as one might find for a new, novel device. Instead, the study aims to show that the performance of the MAGEC System in older skeletally immature patients (subject group) is comparable to its established performance in younger patients (predicate group). Therefore, the "acceptance criteria" are implicitly that the clinical outcomes in the subject group should not be worse than those in the predicate group, supporting substantial equivalence.
| Acceptance Criteria (Inferred from comparison) | Reported Device Performance (Subject Group 2: Age ≥10 years) | Reported Device Performance (Predicate Group 1: Age <10 years) |
|---|---|---|
| Major curve Cobb angle, Δ, mean (not worse) | 25.0% | 26.5% |
| Thoracic height (T1-T12), Δ, mean (not worse) | 15.5% | 15.1% |
| Spinal height (T1-S1), Δ mean (not worse) | 15.3% | 15.3% |
| Device-related adverse events (not worse) | 8.1% | 35.3% |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Sample Size (Test Set):
- Group 1 (Predicate: Age <10 years): N=1,080 patients
- Group 2 (Subject: Age ≥10 years): N=172 patients
- Group B (Subject/Literature: Age ≥10 years): N=57 patients
- Data Provenance: Retrospective registry study. The country of origin is not specified in the provided text.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
The document does not mention the use of experts to establish ground truth for the test set in the context of image interpretation or diagnostic accuracy. This is a spinal bracing and distraction system, and the "ground truth" for its effectiveness is based on clinical and radiographic measurements taken during the treatment process within the registry.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. The study is a retrospective registry study tracking clinical and radiographic outcomes, not an adjudication process for diagnostic accuracy.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a medical device for spinal treatment, not an AI-assisted diagnostic device that would involve human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical medical device (spinal bracing and distraction system), not an algorithm or AI system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the device's effectiveness is based on clinical and radiographic outcomes data (e.g., changes in Cobb angle, thoracic height, spinal height, and incidence of device-related adverse events) collected from patients treated with the MAGEC System.
8. The sample size for the training set
Not applicable. This is a study of a physical medical device, not a machine learning model, so there is no training set in the AI sense.
9. How the ground truth for the training set was established
Not applicable, as there is no training set for an AI model. For the clinical study, the "ground truth" (i.e., the collected patient data and outcomes) was established through existing medical records and measurements from the retrospective registry study.
{0}------------------------------------------------
February 6, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration" in a sans-serif font.
Nu Vasive Specialized Orthopedics, Incorporated Miriam Cervantes Senior Regulatory Affairs Specialist 101 Enterprise Suite 100 Aliso Viejo, California 92656
Re: K233593
Trade/Device Name: MAGEC Spinal Bracing and Distraction System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: PGN Dated: November 8, 2023 Received: November 8, 2023
Dear Miriam Cervantes:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin
O'neill -S
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K233593
Device Name
MAGEC Spinal Bracing and Distraction System
Indications for Use (Describe)
The MAGEC Spinal Bracing and Distraction System is indicated for skeletally immature patients with severe progressive spinal deformities (e.g., Cobb angle of 30 degrees or more; thoracic spine height less than 22 cm) secondary to earlyonset scoliosis associated with or at risk of Thoracic Insufficiency Syndrome(TIS). TIS is defined as the inability of the thorax to support normal respiration or lung growth.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Nuvasive Specialized Orthopedics, Inc. The logo features a stylized purple and silver graphic to the left of the company name. The company name is written in silver, with "SPECIALIZED ORTHOPEDICS, INC." in smaller purple font below "NUVASIVE".
K233593 Page 1 of 5
MAGEC Spinal Bracing and Distraction System 510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:
A. Submitted bv:
Miriam Cervantes Senior Regulatory Affairs Specialist NuVasive Specialized Orthopedics, Inc. 101 Enterprise, Suite 100 Aliso Viejo, CA 92656 Telephone: (909) 229-7836
Date Prepared: November 7, 2023
B. Device Name
Trade or Proprietary Name: Common or Usual Name: Classification Name: Device Class: Regulation Number: Product Code:
MAGEC Spinal Bracing and Distraction System Non Fusion Growing Rod System Growing Rod System- Magnetic Actuation Class II 21 CFR § 888.3070 PGN
C. Predicate Devices
The subject MAGEC Spinal Bracing and Distraction System is substantially equivalent to the primary predicate MAGEC® System (K201543) and additional predicates MAGEC® Spinal Bracing and Distraction System (K161751) and Magec Spinal Bracing And Distraction System (K140613).
D. Device Description
The subject MAGEC Spinal Bracing and Distraction System has an identical design and principle of operation to the predicate design iterations cleared in the predicate Magec System (K201543, K161751, K140613). The subject system includes sterile single use MAGEC rods manufactured from Ti-6Al-4V, conforming to ASTM F136, along with various accessories including a sterile Rod Template and Manual Distractor, a non-sterile Wand Magnet Locator, and is compatible with a hand held External Remote Controller (ERC) 1 or 2. The MAGEC rod can be surgically implanted using appropriate NuVasive Reline 4.5-5.0 (Reline Small Stature or RSS) or Armada fixation components (i.e., pedicle screws, hooks and/or connectors). The titanium MAGEC rod includes an actuator portion that holds a small internal magnet. The magnet in the actuator can be turned non- invasively by use of the ERC, which is electrically powered. The hand-held non-invasive ERC is placed over the patient's spine and then manually activated, which causes the implanted magnet to rotate and either distract or retract the rod. Periodic distraction of the rod is performed to lengthen the spine and to provide adequate bracing during growth to
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a logo for Nuvasive Specialized Orthopedics, Inc. The logo features a stylized, abstract shape on the left, which is partially purple and partially gray. To the right of the shape, the word "NUVASIVE" is written in a gray, sans-serif font. Below "NUVASIVE", the words "SPECIALIZED ORTHOPEDICS, INC." are written in a smaller, purple font.
minimize the progression of scoliosis. Once the physician determines that the implant has achieved its intended use and is no longer required, the implant is explanted.
The purpose of this premarket notification is to expand the indications for use to include treatment of skeletally immature patients of any age with early onset scoliosis.
E. Indications for Use
The MAGEC Spinal Bracing and Distraction System is indicated for skeletally immature patients with severe progressive spinal deformities (e.g., Cobb angle of 30 degrees or more; thoracic spine height less than 22 cm) secondary to early-onset scoliosis associated with or at risk of Thoracic Insufficiency Syndrome (TIS). TIS is defined as the inability of the thorax to support normal respiration or lung growth.
Technological Characteristics F.
As was established in this submission, the subject MAGEC System is substantially equivalent to its predicate devices cleared by the FDA for commercial distribution in the United States. The subject device has been shown to be substantially equivalent and have equivalent technological characteristics to the predicate through comparison in areas including clinical use, design, material composition, intended use, and function.
The following table describes the summary comparison of technological characteristics of the subject device with the predicate device:
| Subject Device | Predicate Device(K201543/K161571/K140613) | |
|---|---|---|
| Indications forUse | The MAGEC Spinal Bracing andDistraction System is indicated forskeletally immature patients withsevere progressive spinal deformities(e.g., Cobb angle of 30 degrees ormore; thoracic spine height less than22 cm) secondary to early-onsetscoliosis associated with or at risk ofThoracic Insufficiency Syndrome(TIS). TIS is defined as the inabilityof the thorax to support normalrespiration or lung growth. | The Magec System is indicated forskeletally immature patients less than10 years of age with severeprogressive spinal deformities (e.g.,Cobb angle of 30 degrees or more;thoracic spine height less than 22 cm)associated with or at risk of ThoracicInsufficiency Syndrome (TIS). TIS isdefined as the inability of the thorax tosupport normal respiration or lunggrowth. |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Nuvasive Specialized Orthopedics, Inc. The logo features a stylized purple and gray graphic to the left of the company name, "NUVASIVE", in gray, block letters. Below the company name is the text "SPECIALIZED ORTHOPEDICS, INC." in a smaller font size.
| Explanation ofdifferences inIndications forUse | The subject device is indicated forskeletally immature patients of anyage with early-onset scoliosis. Thepurpose of specifying early-onsetscoliosis is to explicitly identify thetreated condition for patients,surgeons, hospitals, and payers. It issolely to clarify the indications. Thisdoes not change the intended patientpopulation or treated conditions. | The predicate device is indicated foruse in skeletally immature patientsless than 10 years of age. |
|---|---|---|
| Summary ofthe technologysimilarities tothe predicatedevice | • Principle of Operation: Brace thespine during growth to minimizethe progression of scoliosis• Material Composition: Titanium• Design: Magnetic mechanism• Accessories: Magec ManualDistractor, Magec Rod Template,and Magec Wand Magnet Locator | • Principle of Operation: Brace thespine during growth to minimizethe progression of scoliosis• Material Composition: Titanium• Design: Magnetic mechanism• Accessories: Magec ManualDistractor, Magec Rod Template,and Magec Wand Magnet Locator |
| Summary ofthe technologydifferences tothe predicatedevice | Identical to predicate | No change in technology |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for "NUVASIVE SPECIALIZED ORTHOPEDICS, INC.". The logo features a stylized, abstract shape in purple and gray on the left. To the right of the shape, the word "NUVASIVE" is written in gray, and below that, the words "SPECIALIZED ORTHOPEDICS, INC." are written in purple.
C. Performance Data
MAGEC System The subject to the predicate Magec System (K201543/K161571/K140613). There have been no design changes introduced as part of this submission. The purpose of this submission is to expand the indications for use of the MAGEC System to include treatment of skeletally immature patients of any age with early onset scoliosis.
The following non-clinical testing was previously provided and reviewed by FDA:
- Biocompatibility evaluation per ISO 10993-1, including cytotoxicity per ISO 10993-5, chemical characterization per ISO 10993-18 and toxicological risk assessment per ISO 10993-17
- Wear Debris Testing and Post-Wear Debris Testing Device Analysis
As the subject MAGEC System device is identical to the predicate Magec System (K201543/K161571/K140613) device in design, material, and manufacturing, there are no new risks related to biocompatibility of the subject device when compared to the predicate.
Additionally, a large retrospective registry study of pediatric early-onset scoliosis patients treated with the subject device was performed. The study population consisted of patient groups:
-
- Group 1: Age at index MAGEC implantation at less than 10 years (N= 1,080 patients)
-
- Group 2: Age at index MAGEC implantation at 10 years or greater (N= 172 patients)
All patients were diagnosed to have early-onset scoliosis before 10 years of age. All patients were treated with the MAGEC System. In total, there were 1,252 patients treated with the MAGEC System, including 1,080 patients treated before 10 years of age, and 172 patients who were treated with the MAGEC System at 10 years of age or older. The clinical and radiographic data demonstrates that the MAGEC System device, when used in patients 10 years of age or older, is substantially equivalent to the predicate Magec System device indicated for use in patients younger than 10 years of age. Any potential hazards of the changes introduced as part of this submission have been evaluated and assessed to have risk mitigators, and any relevant information has been addressed in the subject device labeling, after all control measures have been implemented.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for Nuvasive Specialized Orthopedics, Inc. The logo features a stylized, abstract shape in purple and gray on the left side. To the right of the shape is the word "NUVASIVE" in a sans-serif font, with the words "SPECIALIZED ORTHOPEDICS, INC." in a smaller font below it.
| Data Source | Registry Study | Literature | |
|---|---|---|---|
| Study Group | Group 1 | Group 2 | Group B |
| (Predicate) | (Subject) | (Subject) | |
| Age at Magec implantation | <10 years | ≥10 years | ≥10 years |
| Demographic Information | |||
| N | 1,080 | 172 | 57 |
| Age, mean (range) | 6.9 (1-9) | 11.3 (10-16) | 11.4 (10.0-18.4) |
| Clinical Outcomes | |||
| Major curve Cobb angle, Δ, mean | 26.5% | 25.0% | 36.6% |
| Thoracic height (T1-T12), Δ, mean | 15.1% | 15.5% | 16.4% |
| Spinal height (T1-S1), Δ mean | 15.3% | 15.3% | 15.4% |
| Device-related adverse events | 35.3% | 8.1% | 10.5% |
The results demonstrate that the subject MAGEC System is substantially equivalent to the predicate.
H. Conclusions
The subject MAGEC Spinal Bracing and Distraction System has been shown to be substantially equivalent to the legally marketed predicate device for its intended use.
N/A