K Number
K233556

Validate with FDA (Live)

Date Cleared
2024-01-04

(59 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LED Eye mask (Model: EY-20R, A20, EY-20N) is an Over-the-Counter (OTC) device intended for use in treating wrinkles within the periorbital region.

Device Description

The LED Eye Mask, model: EY-20R, A20, EY-20N is an over-the-counter light-emitting diode (LED) device that emits energy for dermatology to treat wrinkles within the periorbital region. The device uses four types of LEDs is 605nm, 633nm, 660nm, 830nm. The LED Eye Mask components contain the main unit. USB cable cord and velcro straps and removable eve protection. There is a total of 40 LEDs to provide a power intensity of about 65 mW/cm². The device has only one button to turn on/off. The user wears the device on their eye area for the treatment, and the device will shut down automatically after a 3-minute after finishing treatment. There are no differences between the three models other than the appearance.

AI/ML Overview

I am sorry, but the provided text from the FDA 510(k) summary for the LED Eye Mask (K233556) does not contain any information regarding acceptance criteria or a study that proves the device meets those criteria for its intended use of treating wrinkles.

The document primarily focuses on demonstrating substantial equivalence to a predicate device (K221444) by comparing design principles, intended use, functions, materials, and applicable standards.

Here's what the document does include, and why it doesn't answer your specific questions:

  • Non-Clinical Tests: The document lists several non-clinical tests performed, such as electrical safety, electromagnetic compatibility, biocompatibility, software verification/validation, and usability validation. These tests confirm the device's electrical safety, material compatibility, and user-friendliness, but they do not directly assess the clinical effectiveness of reducing wrinkles.
  • No Clinical Study Mentioned: Crucially, Section 7.2 explicitly states: "Clinical testing was not needed for this 510(k). The non-clinical performance testing described above is sufficient to support that the device can be used safely and effectively." This means no clinical study was conducted for this specific 510(k) submission to prove its effectiveness in treating wrinkles. The substantial equivalence argument relies on the predicate device having demonstrated effectiveness, and the new device being sufficiently similar not to raise new safety or effectiveness concerns.

Therefore, I cannot provide the information you requested about acceptance criteria and a study proving clinical effectiveness, as it is not present in the provided FDA 510(k) summary. The document focuses on regulatory compliance through substantial equivalence, rather than directly showing new clinical efficacy data.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.

January 4, 2024

Shenzhen Kaiyan Medical Equipment Co., Ltd Alain Dijkstra, General Manager Building#3 and Building#5, 40th of Fuxin Street Huaide Community Fuyong Town, Baoan District Shenzhen, Guangdong 510530. China

Re: K233556

Trade/Device Name: LED Eye mask (EY-20R, A20, EY-20N) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS Dated: November 2, 2023 Received: November 6, 2023

Dear Alain Dijkstra:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/6 description: The image shows a digital signature. The signature is for Jianting Wang -S. The date of the signature is 2024.01.04, and the time is 15:15:05 -05'00'.

Digitally signed by
Jianting Wang -S
Date: 2024.01.04
15:15:05 -05'00'

For Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K233556

Device Name

LED Eye mask (EY-20R, A20, EY-20N)

Indications for Use (Describe)

The LED Eye mask (Model: EY-20R, 20A, EY-20N) is an Over-the-Counter (OTC) device intended for use in treating wrinkles within the periorbital region.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary of K233556

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.

1. Submitter's Information

Sponsor Name: Shenzhen Kaiyan Medical Equipment Co., Ltd Establishment Registration Number: 3011644607 Address: Building#3 and Building#5, 40th of Fuxin Street, Huaide Community Fuyong Town, Baoan District, Shenzhen, Guangdong 518103, China Contact Person (including title): Alain Dijkstra (Manager) Tel: +86-135-10378748 Fax: +86-755-25024651 E-mail: regulation@kaiyanmedical.com

Application Correspondent:

Contact Person: Alain Dijkstra Company: Shenzhen Kaiyan Medical Equipment Co., Ltd Address: Building#3 and Building#5, 40th of Fuxin Street, Huaide Community Fuyong Town, Baoan District, Shenzhen, Guangdong 518103, China Tel: +86 755 82129361 Fax: +86 755 25024651 Email: requlation@kaiyanmedical.com

2. Subject Device Information:

Trade Name: LED Eye Mask, model: EY-20R, A20, EY-20N Classification Name: Light based OTC for winkle reduction Review Panel: General & Plastic Surgery Product Code: OHS Regulation Number:21 CFR 878.4810 Regulation Class: II

Predicate Device Information 3. Predicate Device (K221444)

Sponsor: Light Tree Ventures Europe B.V. Trade Name: LED Eye Perfector, model: EY-36A, EY-36B Classification Name: Light based over the counter wrinkle reduction Review Panel: General & Plastic Surgery Product Code: OHS Requlation Number: 21 CFR 878.4810

4. Device Description

The LED Eye Mask, model: EY-20R, A20, EY-20N is an over-the-counter light-emitting diode (LED) device that emits energy for dermatology to treat wrinkles within the periorbital region. The device uses four types of LEDs is 605nm, 633nm, 660nm, 830nm. The LED Eye Mask components contain the main unit. USB cable cord and velcro straps and removable eve protection. There is a total of 40 LEDs to provide a power intensity of about 65 mW/cm². The device has only one button to turn on/off. The user wears the device on their eye area for the treatment, and the device will shut down automatically

{4}------------------------------------------------

after a 3-minute after finishing treatment.

There are no differences between the three models other than the appearance.

5. Intended Use / Indications for Use

The LED Eye Mask (Model: EY-20R, A20, EY-20N) is an Over-the-Counter (OTC) device intended for use in treating wrinkles within the periorbital region.

Comparison to predicate devices 6.

Compare with the predicate devices, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between the subject device and predicate devices do not raise new questions of safety or effectiveness.

Elements ofComparisonSubject devicePredicate device (K221444)Remark
CompanyShenzhen Kaiyan MedicalEquipment Co., LtdLight Tree Ventures Europe B.V.--
Trade NameLED Eye maskLED Eye Perfector--
ModelEY-20R, A20, EY-20NEY-36A, EY-36B--
510 (K) NumberApplyingK221444--
Classification NameLight Based Over-The CounterWrinkle ReductionLight Based Over-The CounterWrinkle Reduction--
Regulation ClassClass IIClass IISame
Product CodeOHSOHSSame
Indications for Use /Intended useThe LED Eye mask (Model:EY-20R, A20, EY-20N) is anOver-the-Counter (OTC)device intended for use intreating wrinkles within theperiorbital region.The LED Eye Perfector (Model:EY-36A, EY 36B) is an Over-theCounter (OTC) device intendedfor use in treating wrinkles withinthe periorbital region.Same
Power SourceMain unit: 3.7V, 300mAhlithium battery, 1.11WhAdapter Input: 100 - 240Va.c.,50/60Hz Adapter Output:5Vd.c, 2AMain unit: 3.7V, 420mAh lithiumbattery, 1.55Wh Adapter Input:100 - 240Va.c., 50/60Hz AdapterOutput: 5Vd.c, 1ADifferent,note 1
Wavelengths605nm, 633nm, 660nm, 830nmEY-36A: 605nm, 633nm, 660nm,830nmEY-36B: 605nm, 625nm, 660nm,880nmSame
ModeOn/OffOn/OffSame
Irradiance sourceLEDLEDSame
Visiblle light LEDsYesYesSame
Treatment Area36.5cm²40cm²Similar, note2
LED number605nm: 20633nm: 20660nm: 20830nm: 20Total: 40 (Double lamp beads)605nm: 20633nm: 20660nm: 20830nm: 20Total: 40 (Double lamp beads)Same
Power flux (mW/cm²)605nm: 26.2±3605nm: 24.5±3Similar, note
Elements ofComparisonSubject devicePredicate device (K221444)Remark
633nm: 13.8±3660nm: 20.3±3830nm: 8.5±3Total: 65633nm: 16.2±3660nm: 18.2±3830nm: 6.7±3Total: 652
LED distributionUniform distributionUniform distributionSame
Treatment Time3minutes per treatment3minutes per treatmentSame
Target PopulationIndividuals with wrinkles ontheir face within the periorbitalregion.Individuals with wrinkles on theirface within the periorbital region.Same
Location for USEOTCOTCSame
EMCIEC60601-1-2IEC60601-1-2Same
SafetyIEC60601-1, IEC60601-2-57,IEC60601-1-11, IEC62471IEC60601-1, IEC60601-2-57,IEC60601-1-11, IEC62471Same
BiocompatibilityISO10993-5, ISO10993-10ISO10993-1, ISO10993-5, ISO10993-10Same

{5}------------------------------------------------

Comparison in Detail(s):

Note 1:

Although the "Power Source" and "Power flux" are different from the predicate devices, they all complied with the IEC 60601-1, IEC 60601-1-2 and IEC 62133-2 safety standards' requirements. So, these differences will not raise any safety or effectiveness issues.

Note 2:

Although the "Treatment Area" is different from the predicate devices, the values are very similar. So, these differences will not raise any safety or effectiveness issues.

7. Test Summary

7.1 Non-Clinical Tests Performed

1) Electrical safety, and electromagnetic compatibility Test

Non-clinical tests were performed on the subject device in order to validate the design and to assure conformance with the following voluntary design standards in connection with medical device electrical safety, and electromagnetic compatibility:

  • IEC 60601-1 2020-08 Edition 3.2 Medical electrical equipment Part 1: General requirements for basic � safety and essential performance - Note: This standard is recognized with relevant US national differences applied, see references #1 and #2 in the Relevant FDA Guidance and/or Supportive Publication section below.
  • � IEC 60601-1-11 Edition 2.1 2020-07 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance --Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment.
  • � IEC 60601-2-57 Edition 1.0 2011-01 Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use.
  • � IEC 60601-1-2 Edition 4.1 2020-09 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests.
  • � IEC 62471 First edition 2006-07 Photobiological safety of lamps and lamp systems.

{6}------------------------------------------------

  • � IEC 62133-2 Edition 1.0 2017-02 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems.

2) Biocompatibility Test

The component of the LED Eye Mask (Models: EY-20R, A20, EY-20N) has been conformed to ISO 10993-5 and ISO 10993-10.

3) Software verification and validation

Software verification and validation testing were conducted and documentation was provided as recommended by FDA'S Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level concern, since a malfunction of, or a latent design flaw in, the Software Device leads to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury.

4) Usability validation

Usability testing was conducted on the LED Eye Mask (Models: EY-20R, A20, EY-20N), which complies with IEC 62366-1 and IEC 60601-1-6.

7.2 Summary of Clinical Performance

Clinical testing was not needed for this 510(k). The non-clinical performance testing described above is sufficient to support that the device can be used safely and effectively.

8. Date of the summary prepared: January 4, 2024

9. Final Conclusion

The subject device is as safe, as effective, and performs as well as or better than the legally marketed predicated devices K221444.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.