K Number
K233534

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-08-02

(273 days)

Product Code
Regulation Number
866.2390
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Hardy Diagnostics' Viral Transport Medium (VTM) is intended for the collection and transport of clinical specimens for the preservation of viral agents influenza A, Influenza B, Adenovirus, and Echovirus from the collection site to the testing laboratory. Hardy Diagnostics' VTM is a culture-based media that is intended to be used in standard laboratory procedures for virus culture and diagnostic assays that utilize stable recoverable infectious viral particles.

Device Description

Hardy Diagnostics' Viral Transport Medium (VTM) is a non-propagating culture-based transport media used for the collection and transport of specimens suspected of containing viruses including Influenza A, Influenza B, Adenovirus, and Echovirus for downstream laboratory test methods. The VTM includes a screw-cap polypropylene tube with skirted conical bottom containing 3mL of transport medium. VTM tubes can be supplied alone, or in a kit format with a mini-tip flocked swab in a sterile peel-pouch. Hardy Diagnostics' VTM is not claimed to be sterile nor is it intended to be sterilized by the end user. Hardy Diagnostics' VTM vials are single use devices.

The product is supplied in multiple configurations described in more detail in table 1 below: tubes alone, or in a kit format with a swab.

AI/ML Overview

The provided text is a 510(k) Premarket Notification summary for a Viral Transport Medium (VTM), a Class I device. It describes the design, intended use, and studies conducted to demonstrate its substantial equivalence to a predicate device.

It's important to note that this document is not for an AI/ML-based medical device. Therefore, many of the requested elements for describing the acceptance criteria and study that proves an AI device meets acceptance criteria (such as MRMC studies, ground truth establishment by experts, and training set details) are not applicable to this type of medical device submission.

However, I can extract the relevant information regarding the device's performance studies and acceptance criteria as provided for this specific product, which focuses on viral recovery performance and shelf-life stability.

Here's a breakdown based on the provided document:


Acceptance Criteria and Device Performance for Viral Transport Medium (VTM)

The studies presented focus on demonstrating the VTM's ability to preserve viral agents over time and under various storage conditions.

1. Table of Acceptance Criteria and Reported Device Performance

For Viral Recovery Performance:

Acceptance CriteriaReported Device Performance (20-25°C storage)Reported Device Performance (2-8°C storage)
Average viral recovery for each time point and storage condition demonstrates any percent changes within ±90% (i.e., 1 log change) from baseline (T=0).Influenza A:24 hrs: -92.24%* (Accepted as 48h was <90%)48 hrs: -59.76%Influenza A:24 hrs: -5.98%48 hrs: -58.59%
Influenza B:24 hrs: -23.86%48 hrs: -3.45%Influenza B:24 hrs: -36.32%48 hrs: -44.27%
Echovirus:24 hrs: -20.55%48 hrs: -25.23%Echovirus:24 hrs: 21.91%48 hrs: -3.98%
Adenovirus:24 hrs: 18.11%48 hrs: -7.71%Adenovirus:24 hrs: -20.19%48 hrs: 23.11%

*The note states that -92.24% for Influenza A at 24 hours (20-25°C) was "Considered acceptable because subsequent timepoints, i.e., 48 h time point showed < 90% increase." This implies a tolerance for transient excursions if the overall trend at later, crucial time points remains within the acceptance range.

For Shelf-life Stability:

Acceptance CriteriaReported Device Performance
Shelf-life of 12 months when stored at 2-25°C, with functional and physical characteristics (virus viability, pH, appearance, fill volume) maintained.All results met the study acceptance criteria, supporting a shelf-life stability of Hardy Diagnostics' VTM for up to 12 months (365 days) when stored at 2-25°C.
Microbial contamination check: No growth on blood agar and SabDex plates incubated at 35°C for a minimum of 48 hours.No growth was observed on any of the plates tested.

2. Sample Size Used for the Test Set and Data Provenance

  • Viral Recovery Study: "A minimum of three lots of Viral Transport Medium" were inoculated with known virus concentrations. Aliquots of each lot were pulled at 0, 24, and 48 hours, serially diluted, and inoculated in triplicate into susceptible host cell lines.
  • Shelf-life Stability Study: "Three lots of VTM were assessed qualitatively at each time point."
  • Data Provenance: The studies appear to be prospective laboratory studies conducted by Hardy Diagnostics. The document does not specify the country of origin of the data beyond being part of a US FDA submission. The "pooled negative clinical matrix" suggests human-derived samples were used as the base, but the testing itself was laboratory-controlled.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

  • Not applicable in the context of an AI/ML device. For this VTM, the "ground truth" (or reference standard) for viral recovery was established by laboratory measurements of viral titer (TCID50/mL) using the Reed-Muench method, which is a quantitative, objective measure. There's no indication of human expert consensus building for the "ground truth" in this type of product.

4. Adjudication Method for the Test Set

  • Not applicable. This is not an AI/ML device requiring human adjudication of performance. The performance was determined by quantitative laboratory assays.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size

  • Not applicable. This is not an AI/ML device, and no human reader studies (MRMC or otherwise) would be relevant for evaluating its performance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is a physical transport medium, not an algorithm. Its performance is inherent in its chemical composition and ability to preserve viral viability.

7. The Type of Ground Truth Used

  • For Viral Recovery Performance: The ground truth was laboratory-derived viral titers (TCID50/mL) determined at baseline (T=0 hours) using the Reed-Muench method, which serves as the reference for percentage change calculations.
  • For Shelf-life Stability: The ground truth for functional and physical characteristics included quantitative measurements (pH, fill volume) and qualitative assessments (appearance, growth/no growth for contamination) against predetermined specifications.

8. The Sample Size for the Training Set

  • Not applicable. This device is not an AI/ML algorithm that requires a "training set."

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As there is no training set for this device, a ground truth for it was not established.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 2, 2024

Hardy Diagnostics Anna Klavins Senior R&D and Performance Studies Manager 1430 West McCoy Lane Santa Maria, California 93455

Re: K233534

Trade/Device Name: Viral Transport Medium Regulation Number: 21 CFR 866.2390 Regulation Name: Transport Culture Medium Regulatory Class: Class I, reserved Product Code: JSM Dated: June 25, 2024 Received: June 25, 2024

Dear Anna Klavins:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming

{1}------------------------------------------------

K233534 - Anna Klavins

product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ribhi Shawar -S

Ribhi Shawar, Ph.D., D(ABMM) Branch Chief General Bacteriology and Antimicrobial Susceptibility Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K233534

Device Name Viral Transport Medium

Indications for Use (Describe)

Hardy Diagnostics' Viral Transport Medium (VTM) is intended for the collection and transport of clinical specimens for the preservation of viral agents influenza A, Influenza B, Adenovirus, and Echovirus from the collection site to the testing laboratory. Hardy Diagnostics' VTM is a culture-based media that is intended to be used in standard laboratory procedures for virus culture and diagnostic assays that utilize stable recoverable infectious viral particles.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summarv

July 31, 2024

The following information is provided in accordance with 21 CFR 807.92 for the Premarket 510(k) Summary:

I. General Information

Applicant Name: Hardy Diagnostics 1430 W. McCoy Lane Santa Maria, CA 93455

Contact Person: Anna Klavins Senior Technical Services and R&D Manager Phone: 800-266-2222 x5752 E-mail: KlavinsA@hardydiagnostics.com

II. Device Information

Device Trade Name: Viral Transport Medium Common Name: Transport culture medium Classification Name: Culture Media, Non-Propagating Transport Regulation Number: 21 CFR 866.2390 Class: Class I Product Code: JSM

III. Predicate Device

Copan Universal Transport Medium (UTM-RT) System (K042970)

IV. Device Description

Hardy Diagnostics' Viral Transport Medium (VTM) is a non-propagating culture-based transport media used for the collection and transport of specimens suspected of containing viruses including Influenza A, Influenza B, Adenovirus, and Echovirus for downstream laboratory test methods. The VTM includes a screw-cap polypropylene tube with skirted conical bottom containing 3mL of transport medium. VTM tubes can be supplied alone, or in a kit format with a mini-tip flocked swab in a sterile peel-pouch. Hardy Diagnostics' VTM is not claimed to be sterile nor is it intended to be sterilized by the end user. Hardy Diagnostics' VTM vials are single use devices.

The product is supplied in multiple configurations described in more detail in table 1 below: tubes alone, or in a kit format with a swab.

{4}------------------------------------------------

CatalogNumberDescriptionQuantity
R99VTM, 16×100mm polypropylene tube, 3mL fill20 tubes/box
R64BXVTM, 13×80mm polypropylene tube, 3mL fill100 tubes/box
TPV50TransPRO VTM System, single 16×100mm polypropylenetube, 3mL fill with individually wrapped, sterile, mini-tip,flocked swab with 80mm breakpoint.50 each/box

Table 1. List of Configurations: Hardy Diagnostics' VTM

V. Composition of Hardy Diagnostics' VTM

The VTM consists of Hank's Balanced Salt Solution, fetal bovine serum, sucrose to stabilize viral agents, along with Amphotericin B and Gentamicin Sulfate to inhibit bacterial and fungal contaminants. Hardy Diagnostics' VTM also contains a pH indicator (phenol red) to provide a visual check on the medium's pH. VTM appears translucent and light peach in color.

VI. Intended Use/Indications for Use

Hardy Diagnostics' Viral Transport Medium (VTM) is intended for the collection and transport of clinical specimens for the preservation of viral agents including, Influenza A, Influenza B, Adenovirus, and Echovirus from the collection site to the testing laboratory. Hardy Diagnostics VTM is a culture-based media that is intended to be used in standard laboratory procedures for virus culture and diagnostic assays that utilize stable recoverable infectious viral particles.

VII. Substantial Equivalence Comparison

The following table demonstrates the substantial equivalence comparison of Hardy Diagnostics' Viral Transport Medium to Copan Universal Transport Medium (UTM-RT) System:

New DevicePredicate Device
Hardy Diagnostics' ViralTransport MediumCopan Universal TransportMedium (UTM-RT) System
K233534K042970
Device Similarities
Device Product Codeand ClassificationJSM, Class 1JSM, Class 1
Intended Use /Indications for UseHardy Diagnostics' Viral TransportMedium (VTM) is intended for thecollection and transport of clinicalspecimens for the preservation ofviral agents including, Influenza A,Influenza B, Adenovirus, andEchovirus from the collection site tothe testing laboratory HardyCopan Universal TransportMedium (UTM-RT) System isintended for the collection andtransport of clinical specimenscontaining viruses, Chlamydiae,Mycoplasma or Ureaplasmafrom the collection site to thetesting laboratory UTM RT can
New DeviceHardy Diagnostics' ViralTransport MediumK233534Predicate DeviceCopan Universal TransportMedium (UTM-RT) SystemK042970
Device Similarities
Diagnostics' VTM is a culture-based media that is intended to be used in standard laboratory procedures for virus culture and diagnostic assays that utilize stable recoverable infectious viral particles.be processed using standard clinical laboratory operating procedures for viral, chlamydial, mycoplasma and ureaplasma culture.
Tube MaterialPlastic Screw-Cap TubeSame
pH$7.3 \pm 0.2$ at 25°CSame
Shelf-life365 daysSame
Storage Temperature2-25°CSame
Single Use DeviceYesSame
Device Differences
Media Formulation• Hank's Balanced Salts Solution (HBSS)• Fetal Bovine Serum• Sucrose• Amphotericin B• Gentamicin Sulfate• Phenol Red• Hank's Balanced Salts solution (HBSS)• Bovine Serum Albumin (BSA)• Vancomycin• Amphotericin B• Colistin• L-Glutamic Acid• L-Cysteine• HEPES Buffer• Phenol Red• Gelatin• Sucrose
Supported Strains• Adenovirus• Echovirus• Influenza A• Influenza B• Adenovirus• Cytomegalovirus• Echovirus Type 30• Herpes Simplex Virus Type 1• Herpes Simplex Virus Type 2• Influenza A• Parainfluenza 3• Respiratory Syncytial Virus• Varicella Zoster Virus• Chlamydia pneumoniae• Chlamydia trachomatis• Mycoplasma hominis• Mycoplasma pneumoniae• Ureaplasma urealyticum

{5}------------------------------------------------

{6}------------------------------------------------

VIII. Shelf-life Stability

The shelf life of Hardy Diagnostics' Viral Transport Medium was determined to be 12 months from the date of manufacture when stored at 2-25°C. Three lots of VTM were assessed qualitatively at each time point for functionality and physical characteristics using real time ageing studies. In the real time stability study, media lots were held at 2-8°C and 20-25°C for 12 months. At each timepoint of testing, performance (virus viability), pH, appearance, and fill volume of the media were examined.

A microbial contamination check was performed to verify the stability of the incorporated antibiotics. 100uL of VTM was inoculated to blood agar and SabDex plates, and incubated at 35°C for a minimum of 48 hours to ensure that no contamination was present. No growth was observed on any of the plates tested.

Overall, all results met the study acceptance criteria and support a shelf-life stability of Hardy Diagnostics' VTM for up to 12 months or 365 days.

Viral Recovery Performance IX.

The performance of Hardy Diagnostics' Viral Transport Medium was evaluated for virus viability across various conditions, including two different incubation temperatures and multiple days. The viral recovery study was conducted by spiking a known virus concentration into pooled negative clinical matrix. The contrived viral samples were then inoculated to a minimum of three lots of Viral Transport Medium and held at 2-8°C and 20-25°C for 0, 24, and 48 hours. Aliquots of each lot were pulled at 0, 24, and 48 hours, serially diluted, and inoculated in triplicate into the susceptible host cell line. Host cells were plated at a suitable density in culture medium in microwell plates prior to testing. Virus induced cytopathic effect (CPE) was observed and the viral titer was determined by calculating the fifty-percent tissue culture infective dose (TCIDsomL) using the Reed-Muench method. Results were considered acceptable if the average viral recovery for each time point and storage condition demonstrate any percent changes within ±90% (i.e., 1 log change) from baseline (T=0).

The results are presented in Tables 1 and 2.

Test StrainViral recovery(TCID50/mL)at T=0 hrs.Percent changes in viral recoveryfrom the baseline (T= 0 hr.)(-ve indicates reduction)
24 hrs.48 hrs.
Influenza A2.12x103-92.24*-59.76
Influenza B7.23x102-23.86-3.45
Echovirus1.40x105-20.55-25.23
Adenovirus1.92x10318.11-7.71
  • Considered acceptable because subsequent timepoints, i.e., 48 h time point showed < 90% increase.

{7}------------------------------------------------

Test StrainViral recovery(TCID50/mL)at T=0 hrs.Percent Changes in viral recoveryfrom the baseline (T=0 hr.)(-ve indicates reduction)
24 hrs.48 hrs.
Influenza A$1.32x10^3$-5.98-58.59
Influenza B$8.05x10^2$-36.32-44.27
Echovirus$1.27x10^5$21.91-3.98
Adenovirus$1.47x10^3$-20.1923.11

Table 2: Viral recovery performance of Viral Transport Medium at 2-8°C

X. Conclusion

The shelf-life stability results support the storage of Hardy Diagnostics' Viral Transport Medium at 2-25°C for up to 12 months. Viral recovery study results support the recovery of Influenza A, Influenza B, Adenovirus, and Echovirus when stored at 2-8°C and 20-25°C for up to 48 hours from sample collection. Based on the device's technological characteristics, intended use, and performance, the Hardy Diagnostics' VTM is substantially equivalent to the predicate device (K042970).

§ 866.2390 Transport culture medium.

(a)
Identification. A transport culture medium is a device that consists of a semisolid, usually non-nutrient, medium that maintains the viability of suspected pathogens contained in patient specimens while in transit from the specimen collection area to the laboratory. The device aids in the diagnosis of disease caused by pathogenic microorganisms and also provides epidemiological information on these diseases.(b)
Classification. Class I (general controls).