Intended Use

Cryojet Plus is indicated to be used to: Minimize pain and thermal injury during laser and dermatological treatments and for temporary topical anesthetic relief for injections.

Device Description

Cryojet Plus is composed of main body (including power supply system, cooling system), power cable, supporting arm rod (a group of supporting rod, two support clips), wind tube, adapters, casters and other accessories. Control system consists of LCD screen display and control board, etc; Cooling system includes compressor, evaporator, radiator, fan, water tank; Power supply system is composed of power cable, filter and ON/OFF switch power.

AI/ML Overview

The provided text describes a medical device called "Cryojet Plus" and its substantial equivalence to a predicate device "Cryo 7". However, it does not contain a study comparing the device's performance against specific acceptance criteria in a comparative effectiveness study involving human readers or standalone AI performance.

Therefore, many of the requested sections cannot be filled from the provided document.

Here's a breakdown of what information is available and what is not:

1. A table of acceptance criteria and the reported device performance

The document provides a comparison of technical specifications between the proposed device (Cryojet Plus) and the predicate device (Cryo 7). While these are performance characteristics, they are not explicitly presented as "acceptance criteria" met by a study designed to prove them. Instead, they are presented as part of a substantial equivalence comparison.

SpecificationPredicate Device (Cryo 7) PerformanceProposed Device (Cryojet Plus) PerformanceDiscussion of Similarities/Differences (Acceptance/Conformance)
Product Code and RegulationGEX – Powered Laser Surgical InstrumentGEX – Powered Laser Surgical InstrumentIdentical
Device DescriptionCompressor, evaporator, fan to cool skinCompressor, evaporator, fan to cool skinIdentical
Intended UseMinimize pain & thermal injury; temporary topical anesthetic reliefMinimize pain & thermal injury; temporary topical anesthetic reliefIdentical
Outlet Temperature-30°C ± 3°C-25°C ± 3°CSimilar, not significantly different
Outlet Air Speed210 ± 40 to 600 ± 100 liters/minute200 ± 50 to 620 ± 60 liters/minuteSimilar, not significantly different
Cooling Method, Operation PrincipleCooling agent-based condenser, evaporation systemCooling agent-based condenser, evaporation systemIdentical
Cooling MaterialGas (Air)Gas (Air)Identical
Mains Voltage100V - 120V / 50-60Hz100V - 120V / 50-60HzIdentical
Main Fuse16A circuit breakerWithout circuit breakerNo impact
Power ConsumptionMax. 10AMax. 8ANo impact (Lower consumption is generally seen as positive/neutral)
IP ClassificationIP20IPX0No impact
DesignInternal metal chassis, plastic shellInternal metal chassis, industrial designed ABS plastic shell, adapter ABS+PCSimilar, no impact on safety/effectiveness
MaterialsR452a, closed loop cooling systemR404a, closed loop cooling systemNo impact (Both environmentally approved)
Cooling Spot SizeVaries, 10cm² at 5cm distanceVaries, 10cm² at 5cm distance (with 3 adapter sizes)Difference in adapter quantity does not raise different safety/effectiveness questions
Fan Speed Settings1-91-10Similar, not significantly different
InterfaceUSB 2.0, RS232TTL, PWMDifference doesn't affect safety/effectiveness
Therapy Time1:00 - 100:00 min - adjustable1:00 - 100:00 min - adjustableIdentical
Time Delay to OperateImmediate from pressing START buttonImmediate from pressing START buttonIdentical
Compatibility with EnvironmentR452a Ozone destruction level 0R404 Ozone destruction level 0Both environmentally approved, no impact
Display5.0 inch touch screen10.1 inch touch screenDifferent screen sizes have no impact on safety/effectiveness
Net Weight60kg65kgSimilar, not significantly different
DimensionsH1060 mm x W500 mm x D560 mmH810 mm x W400 mm x D845mmDifferent dimensions have no impact on safety/effectiveness
Working Environment Temperature10 - 35°C10 - 35°CIdentical
Working Environment Humidity20-80% RH (no condensation)20-80% RH (no condensation)Identical
Working Environment Pressure900-1030 hPa900-1030 hPaIdentical
Storage & Transport Temperature-10 - 50°C-10 - 50°CIdentical
Storage & Transport Humidity10-90% RH (no condensation)10-90% RH (no condensation)Identical
Storage & Transport Pressure700-1060 hPa700-1060 hPaIdentical
Environmental SpecificationsFor indoor use onlyFor indoor use onlyIdentical

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable. This device is not an AI/ML diagnostic or image analysis device, and no clinical testing was performed or reported. The submission focuses on substantial equivalence based on technical characteristics and non-clinical testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. No clinical test set with ground truth established by experts was used.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. No clinical test set was used.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. No MRMC comparative effectiveness study was done. The device is a non-AI therapeutic cooling device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI/ML algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

Not applicable. No clinical ground truth was established or used for the purpose of this 510(k) submission. The "ground truth" for the device's technical performance would be engineering measurements as per the standards listed.

8. The sample size for the training set

Not applicable. This is not an AI/ML algorithm and therefore does not have a training set.

9. How the ground truth for the training set was established

Not applicable. This is not an AI/ML algorithm and therefore does not have a training set.

Study/Evidence that Proves the Device Meets Acceptance Criteria:

The document states that the Cryojet Plus has been considered to comply with several voluntary standards through investigation and test. These standards cover aspects like electrical safety, electromagnetic compatibility, usability, software lifecycle processes, risk management, biocompatibility assessment (lack of direct patient contact negates further testing), and quality management systems.

Specifically mentioned are:

  • IEC 60601-1:2005;A1:2012;A2:2020 (Medical electrical equipment - General requirements for basic safety and essential performance)
  • IEC 60601-1-2:2014 (Medical electrical equipment – Electromagnetic disturbances - Requirements and tests)
  • IEC 60601-1-6:2013 (Medical electrical equipment – Usability)
  • IEC 62366-1:2015 (Medical devices – Application of usability engineering to medical devices)
  • IEC 62304:2015 (Medical devices software – software life cycle processes) - Software verification and validation tests were performed successfully and met acceptance criteria.
  • ISO 14971:2019-12 (Medical devices – Application of risk management to medical devices)
  • ISO 15223-1:2016 (Medical devices - Symbols to be used with medical device labels, labeling, and information to be supplied - Part 1: General requirements)
  • ISO 13485:2016 (Medical devices – Quality management systems--Requirements for regulatory purposes)
  • ISO 10993-1:2018 (Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process) - Biocompatibility was assessed, and no additional testing was deemed necessary due to no direct patient contact.

Non-clinical testing for Cryojet Plus included:

  • Electrical safety testing according to IEC 60601-1.
  • Electromagnetic compatibility testing according to IEC 60601-1-2.
  • Software verification and validation.

The conclusion is that the Cryojet Plus is substantially equivalent to the predicate device (Cryo 7) in intended use, device characteristics, and application parameters, based on these technical comparisons and non-clinical testing. No clinical testing was included in this 510(k).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. The words "U.S. FOOD & DRUG" are on the top line, and the word "ADMINISTRATION" is on the bottom line.

February 6, 2024

Beijing Sano Laser S&T Development Co., Ltd % Zhao Huifang Correspondent Microkn Business Consulting (Shanghai) Co., Ltd. Room 1319, Block A, No 3699, Gonghexin Road Jingan District Shanghai, 200040 China

Re: K233416

Trade/Device Name: Cryojet Plus (SHE-ACP001-1) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: December 26, 2023 Received: December 26, 2023

Dear Zhao Huifang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to. 21 CFR 820.30. Design controls: 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).

{2}------------------------------------------------

Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Tanisha L. Tanisha L.Hithe -S 2024.02.06 Hithe -S 17:09:18 -05'00'

Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical & Infection Control Devices, Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K233416

Device Name Cryojet Plus (SHE-ACP001-1)

Indications for Use (Describe)

Cryojet Plus is indicated to be used to: Minimize pain and thermal injury during laser and dermatological treatments and for temporary topical anesthetic relief for injections.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

I. Contact Details

Applicant Name:Beijing Sano Laser S&T Development Co., Ltd.
Room 201, 2nd Floor, No.7 Building, Maohuagongchang, NO.1CAIDA Third Street, Nancai Town, Shunyi District, Beijing, China
TEL: 086-10-57150601
Contact:Hongbo Zhang, CEO
360349720@qq.com

Date Prepared: Oct 06, 2023

II. Device Name

Device Name : Cryojet Plus (Model: SHE-ACP001-1)

Generic Name: Powered Laser Surgical Instrument

Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology (21CFR 878.4810)

Regulatory Class: II

Product Code: GEX

III. Predicate Device

510(k) NumberProduct CodeTrade NameApplicant
K220020GEXCryo 7Zimmer MedizinSysteme GmbH

This predicate has not been subject to a design-related recall.

No reference devices were used in this submission.

IV. Device Description

Cryojet Plus is intended to minimize pain and thermal injury during laser and dermatological treatments and for the temporary topical anesthetic relief for injections. And it shall be used by professionally trained medical operators.

{5}------------------------------------------------

Cryojet Plus is composed of main body (including power supply system, cooling system), power cable, supporting arm rod (a group of supporting rod, two support clips), wind tube, adapters, casters and other accessories. Control system consists of LCD screen display and control board, etc; Cooling system includes compressor, evaporator, radiator, fan, water tank; Power supply system is composed of power cable, filter and ON/OFF switch power.

How It Works: Gaseous refrigerant is compressed into high-temperature and high-pressure gas by high-power compressor and sent to condenser for cooling. Refrigerant can produce cold air by condenser and evaporation system for skin cooling through conduction, evaporation and forced convection. Evaporator absorbs heat in air and vaporizes the heat, then heat is gaseous, then the gas is back to compressor and continue to be compressed. Thus cooling is done through such circulation. Defrosting system melts the frost inside evaporation system into liquid water and discharge the water through drain valve.

Software is used for system initialization, displaying user interface and cooling system control. Parameter setting (therapy time and wind level) can be done by touching LCD screen, thus START/STOP and defrosting of device can be controlled. The software can be used in conjunction with hardware system during treatment to minimize thermal damage to tissues and reduce pain.

V. Intended Use/Indications for use

Cryojet Plus is indicated to be used to: Minimize pain and thermal injury during laser and dermatological treatments and for temporary topical anesthetic relief for injections.

Main components of device are listed below:

ComponentsFunction Description
Air SwitchControl of Device Power ON/OFF

Table 2 Main Components of Proposed Device

{6}------------------------------------------------

Touch ScreenDisplay User Interface, Operation Control System, TreatmentParameter Setting
Wind TubeDeliver cold air from blower to adapter.
AdapterDifferent sizes of adapters are applicable to different body parts to betreated.

VI. Substantial Equivalence Comparison / Comparison of technological characteristics with the predicate device

The predicate device for this submission is Cryo 7(K220020) from Zimmer MedizinSysteme GmbH that has been approved under K220020.

The submitted proposed device (Cryojet Plus) and predicate device (Cryo 7) are both classified as II. They are both for skin cooling by blowing cold wind to patient's local skin parts (the device is not intended to be in contact with patient). The indications of proposed device (Cryojet Plus) and predicate device (Cryo 7) are the same.

The submitted proposed device (Cryojet Plus) and predicate device (Cryo 7) are based on the same scientific theories:

  1. The working theories principles are identical. They both apply compressor cooling. Refrigerant is compressed into high-temperature and high-pressure gas (by high-power compressor) and sent to condenser for cooling and liquefaction. Refrigerant goes through expansion valve for throttle in evaporator and it absorbs heat in air and generate cold air surrounding it. Meanwhile, the fan blows natural air into evaporator for coldness and heat exchange, then the cold air is conducted to skin by wind tube. Forced convection makes skin cooling. Refrigerant absorbs heat in evaporator and becomes evaporated, then back to compressor to continue compression. Thus cooling is done through such circulation.

{7}------------------------------------------------

Defrosting system melts the frost inside evaporation system into liquid water and discharge the water through drain valve.

    1. While the wind current flow level range, lowest temperature of output wind, cooling spot coverage area at a distance of 5cm, therapy time and defrost for both devices are identical or similar to reach the indicated effect, they do differ in appearance, size (dimension), weight and the product code of control system. The whole device is supported and moved through casters.
  • Product Components: The two devices are both substantially composed of compressor, 3. evaporator, condenser, DC brushless motor, fan, temperature sensor and wind tubes.
    1. Their energy sources are also identical.

The comparison of technical characteristics between proposed device (Cryojet Plus) and predicate device (Cryo 7) are as follows:

    1. For Cryojet Plus produced by SANO Laser, there are 3 sizes of adapters (listed below) for therapy on different body parts; for Cryo 7 produced by Zimmer, there is no adapter for choice, wind is blown out from end of wind tube directly. (Vary in cooling spot coverage area at a distance of 5cm).
ItemAdapter SpecificationsSizeIntended for use
1Small Sized RoundAdapterImage: Small Sized Round AdapterFor small areatreatment and skincooling meanwhile(e.g. eyebrow,armpits, forehead.)
2Medium SizedSquare AdapterImage: Medium Sized Square AdapterFor medium areatreatment and skincooling meanwhile(especially used forhair removal on arms,armpits and legs, inconjunction with hair

{8}------------------------------------------------

removal machine.)
3Large Sized SquareAdapterImage: AdapterFor large areatreatment and skincooling meanwhile(especially used forhair removal onthighs, abdomen, inconjunction with hair
    1. Different Display Screen Size: Proposed device (Cryojet Plus) screen size is 10.1 inch and predicate device (Cryo 7) screen size is 5 inch;
  • IP Protection Level Difference: Proposed device (Cryojet Plus) is IPX0 and predicate 3. device (Cryo 7) is IP20;
  • Weight Difference: Proposed device (Cryojet Plus) is 65kg and predicate device (Cryo 7) 4. is 60kg.

However, these differences do not affect the substantial equivalence; they are still based on the same technical characteristics to achieve the same purpose. For technical characteristics and operating principle associated with therapy, proposed device (Cryojet Plus) and predicate device (Cryo 7) are identical similarities and differences between the proposed device and the predicate device are described below in the comparison table. The differences do not raise any new issues of safety or effectiveness.

{9}------------------------------------------------

SpecificationPROPOSED DEVICEBeijing Sano Laser S&T DevelopmentCo., Ltd.Cryojet Plus (This Submission)PREDICATE DEVICEZimmer MedizinSysteme GmbHCryo 7 (K220020)Discussion ofSimilarities/Differences
Product Code andRegulationGeneral & Plastic Surgery 21 CFR878.4810 GEX – Powered Laser SurgicalInstrumentGeneral & Plastic Surgery 21 CFR 878.4810GEX - Powered Laser Surgical Instrument.Identical
DeviceDescriptionThe Cryojet Plus Air Device consists of acompressor, evaporator and fan to cool thepatient's skin locally without contact toskin.The Cryo 7 Cold Air Device consists of acompressor, evaporator and fan to cool thepatient's skin locally without contact to skin.Identical
Intended UseThe Cryojet Plus is intended to minimizepain and thermal injury during laser anddermatological treatments and fortemporary topical anesthetic relief forinjections.The Cryo 7 is intended to minimize pain andthermal injury during laser and dermatologicaltreatments and for temporary topical anestheticrelief for injections.Identical
PerformanceCharacteristicsMeasurement atdevice outlet-25°C ± 3°C, At an air speed of 200 ± 50to 620 ± 60 liters/minute.-30°C ± 3°C, At an air speed of 210 ± 40 to600 ± 100 liters/minute.Similar, notsignificantlydifferent
Cooling Method,OperationPrincipleCooling agent-based condenser andevaporation system. Cold air is producedwhich is then delivered to skin for skincooling by conduction, evaporation, andforced convection.Cooling agent-based condenser and evaporationsystem. Cold air is produced which is thendelivered to skin for skin cooling by conduction,evaporation, and forced convection.Identical
SpecificationPROPOSED DEVICEBeijing Sano Laser S&T DevelopmentCo., Ltd.Cryojet Plus (This Submission)PREDICATE DEVICEZimmer MedizinSysteme GmbHCryo 7 (K220020)Discussion ofSimilarities/Differences
Cooling MaterialGas (Air)Gas (Air)Identical
Mains Voltage100V - 120V / 50-60Hz100V - 120V / 50-60HzIdentical
Main FuseWithout circuit breaker in main switch16A circuit breaker in main switchNo impact
PowerConsumptionMax. 8AMax. 10ANo impact
IP ClassificationIPX0IP20No impact
DesignInternal metal chassis and framework byindustrial designed ABS plastic shell,adapter material is ABS+PCInternal metal chassis covered by industrialdesigned plastic shellSimilar, noimpact ondevice safetyandeffectivenessevaluation
MaterialsEnvironmentally approved R404a, closedloop cooling systemEnvironmentally approved R452a, closed loopcooling systemNo impact
SpecificationPROPOSED DEVICEBeijing Sano Laser S&T DevelopmentCo., Ltd.Cryojet Plus (This Submission)PREDICATE DEVICEZimmer MedizinSysteme GmbHCryo 7 (K220020)Discussion ofSimilarities/Differences
Cooling Spot SizeVaries in cooling spot size (coverage area)with a distance of 5cm from wind outletpart to skin - 10cm $^{2}$Varies in cooling spot size (coverage area) witha distance of 5cm from wind outlet part to skin -10cm $^{2}$Thedifference ofadapterquantity doesnot raisedifferentquestions ofsafety andeffectiveness:SANOCryojet Plusmachine has 3adapter sizesfor choiceaccording todifferent bodyparts to betreated;Zimmermachine hasonly oneadapter size.
SpecificationPROPOSED DEVICEBeijing Sano Laser S&T DevelopmentCo., Ltd.Cryojet Plus (This Submission)PREDICATE DEVICEZimmer MedizinSysteme GmbHCryo 7 (K220020)Discussion ofSimilarities/Differences
Fan SpeedSettings1-101-9Similar, notsignificantlydifferent
InterfaceTTL, PWMUSB 2.0, RS232Difference ininterfacesdoesn't affectdevice safetyandeffectivenessevaluation
Therapy time,Treatment time1:00 - 100:00 min - adjustable for users1:00 - 100:00 min - adjustable for usersIdentical
Time Delay toOperateImmediate from pressing STARTbutton when system turns to standby modeImmediate from pressing STARTbutton when system turns to standby modeIdentical
Compatibility withthe environmentR404 Ozone destruction level 0R452a Ozone destruction level 0Bothenvironmentally approved,no impact
SpecificationPROPOSED DEVICEBeijing Sano Laser S&T DevelopmentCo., Ltd.Cryojet Plus (This Submission)PREDICATE DEVICEZimmer MedizinSysteme GmbHCryo 7 (K220020)Discussion ofSimilarities/Differences
Display5.0 inch touch screen display10.1 inch touch screen displayDifferentscreen sizeshave noimpact on thedevice safetyandeffectivenessevaluation
Net Weightwithoutaccessories65kg60kgSimilar, notsignificantlydifferent
DimensionsH810 mm x W400 mm x D845mmH1060 mm x W500 mm x D560 mmDifferentdimensionshave noimpact on thedevice safetyandeffectivenessevaluation
SpecificationPROPOSED DEVICEBeijing Sano Laser S&T DevelopmentCo., Ltd.Cryojet Plus (This Submission)PREDICATE DEVICEZimmer MedizinSysteme GmbHCryo 7 (K220020)Discussion ofSimilarities/Differences
WorkingEnvironmentTemperature andHumidityTemperature: 10 - 35°CTemperature: 10 - 35°CIdentical
Humidity: 20-80% relative humidity (nocondensation)Humidity: 20-80% relative humidity (nocondensation)
Atmosphere Pressure: 900-1030 hPaAtmosphere Pressure: 900-1030 hPa
Storage andTransportTemperature: -10 - 50°CTemperature: -10 - 50°CIdentical
Humidity: 10-90% relative humidity (nocondensation)Humidity: 10-90% relative humidity (nocondensation)
Atmosphere Pressure: 700-1060 hPaAtmosphere Pressure: 700-1060 hPa
EnvironmentalSpecificationsFor indoor use onlyFor indoor use onlyIdentical

Table Summary of Similarities and Differences

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

Cryojet Plus 510(k) Notification K233416 Performance Data

By investigation and test, Cryojet Plus has been considered to comply with the following voluntary standards:

StandardsStandardsOrganizationStandards Title
IEC 60601-1:2005;A1:2012;A2:2020IECMedical electrical equipment-Part1:Generalrequirements for basic safety and essentialperformance
IEC 60601-1-2:2014IECMedical electrical equipment – Part 1-2: Generalrequirements for basic safety and essentialperformance – Collateral standard: Electromagneticdisturbances - Requirements and tests
IEC 60601-1-6:2013IECMedical electrical equipment – Part 1-6: Generalrequirements for basic safety and essentialperformance - Collateral standard: Usability
IEC 62366-1:2015IECMedical devices – Part 1: Application of usabilityengineering to medical devices
IEC 62304:2015IECMedical devices software –software life cycleprocesses
ISO 14971:2019-12ISOMedical devices – Application of risk management tomedical devices
ISO 15223-1:2016ISOMedical devices - Symbols to be used with medicaldevice labels, labeling, and information to besupplied - Part 1: General requirements
ISO 13485:2016ISOMedical devices – Quality management systems--
Requirements for regulatory purposes
ISO 10993-1:2018ISOBiological evaluation of medical devices – Part 1:Evaluation and testing within a risk managementprocess

{16}------------------------------------------------

Software Verification and Validation Testing

After evaluation on specified use conditions, it could be proved that the technical design (including software flow and device application) of Cryojet Plus is suitable to fulfill the specified intended use. In both formative and summative testing performed, no application problems were found that could lead to any risk for the user, patient or the third parties.

The verification of software requirements was performed according to IEC 62304. All the tests were performed successfully and met their acceptance criteria.

Biocompatibility Testing

The device has been assessed for the requirement of biocompatibility testing as per ISO 10993-1. Since the device is not intended to be in contact with the patient, and the user wears medical gloves when touching the device, there is no additional biocompatibility requirement for device.

Electrical Safety and Electromagnetic Compatibility (EMC)

The device has undergone electrical and mechanical safety performance testing and electromagnetic compatibility testing as a result of the changes referenced. The system complies with IEC 60601-1:2005 + A1:2012+ A2:2020, and IEC 60601-1-2:2014+ A1:2020 .

Non-clinical Testing

Non-clinical testing for Cryojet Plus included electrical safety testing according to IEC 60601-1, electromagnetic compatibility testing according to IEC 60601-1-2. The device has been assessed for the requirement of biocompatibility testing as per ISO 10993-1. Since the device is not intended to be in contact with the patient, and the user wears medical gloves when touching the device, there is no additional biocompatibility requirement for device. In addition, software verification and validation was completed.

Clinical Testing

No clinical testing is included in this 510(k).

{17}------------------------------------------------

Summary / Conclusion

The Cryojet Plus are substantially equivalent to the Cryo 7 in intended use, device characteristics and application parameters.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.