K Number
K233331

Validate with FDA (Live)

Device Name
MediOMx
Manufacturer
Date Cleared
2024-06-17

(262 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MediOMx is a software-based viewer intended to be used with off-the-shelf hardware for the display of DICOM data to aid in diagnosis for healthcare professionals. It performs operations relating to the transfer, storage, and display of image data. MediOMx allows users to perform image manipulations, including window/level, pan, zoom, and rotation. MediOMx provides 2D display, Multi-planar Reformatting of medical image data. MediOMx is not intended for primary mammography interpretation.

Mobile usage is for reference and referral only.

Device Description

MediOMx is a software-based medical image viewer used with web browsers for the 2D visualization of DICOM and non-DICOM medical images. MediOMx is intended for storage, display, manipulation, and processing of radiological data, including images, reports and other clinical information. It has the following primary features and functions:

  • Zero-footprint HTML5 medical image upload, transfer and display of medical images between facilities
  • Easy access to images for all participants in the healthcare process, including radiologists, technologists, physicians, nurses and other patient care practitioners
  • Serve as information and data management system for for DICOM and non-DICOM medical images
  • Tools for image manipulation (e.g.pan, zoom, rotate, window level)
  • Metadata information and orientation labels display
  • Advanced image manipulation functions like view synchronization across series, 3D visualization like MIP and MPR
  • Encrypted transmission of medical images through secured networks
  • Encrypted storage of medical images
  • HIPAA-compliant data management, including centralized storage of user activity via audit trails.
  • Management of users, roles, and permissions
  • It supports optional integration with FDA-cleared 3rd party AI models. It sends the input study to the 3rd party AI model, receives the AI output and displays outputs of the 3rd party AI model "as-is" in MediOMx for visualization by the Radiologist to assist in diagnosing the study. Outputs are displayed in accordance with the 3rd party provider's regulatory clearance. The original image is always accessible

MediOMx consists of a cloud-based application that displays and processes digital medical images, and associated medical information to aid in the day-to-day operations and workflow of clinicians and healthcare practitioners. The web browser based medical image viewer serves as the frontend module which users interact with in viewing data. The backend module handles the connection and processing of data from a variety of sources within the health system, in view of preparing visualizations to be rendered by the viewer. MediOMx can connect and access the medical images across different sources in a health system: an existing PACS or VNA, cloud storage or local server-based storage. Users can also upload images securely into MediOMx which can be shared and enables collaboration with other users. The data connection and imaging data processing is handled by the MediOMx backend module which supports the standardized transmission protocol as defined in the DICOM standard. Users interact with MediOMx through a standard web browser, thus providing access to full quality images from anywhere and supporting a greater efficiency for care. MediOMx utilizes authorization and authentication mechanisms that enforces authorized users to access the imaging data. The system extends beyond the hospital and its internal network. With a proper network, MediOMx can be accessed by clinical users outside of the hospital network. This way referring physicians can easily call up the imaging data of their patients or external expert accessing the imaging data for additional opinion. MediOMx provides end-users with the ability for industry standard features such as Window/ Level, Image Flip and Rotate. Images are initially displayed in the 2D view mode, but with the ability to toggle into advanced viewing mode of MPR for relevant exam types. It supports processing and displaying Multiplanar Reconstruction (MPR). MediOMx provides an image rendering mechanism that preloads lower resolution images during image scrolling to improve interactivity and performance for users operang in lower network bandwidth while the full quality image is loaded in the background. The use of a secure data transmission protocol and data encryption ensure high data security for data management via the internet. MediOMx tracks user activity via audit trails and stores the audit data on the centralized server.

AI/ML Overview

This document does not contain the detailed information necessary to answer all parts of your request. It's a 510(k) summary for a medical image viewer called MediOMx, which focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study report for specific acceptance criteria and performance metrics.

Here's what can be extracted based on the provided text, and where information is missing:


1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state acceptance criteria or provide a table of reported device performance in the way a clinical study report would. It generally states that "Safety and performance of the PMX MediOMx device has been evaluated and verified in accordance with software specifications and applicable performance standards through software verification, validation, standalone and clinical performance testing." However, specific numerical performance metrics (e.g., accuracy, sensitivity, specificity, or response times for image display) and their corresponding acceptance criteria are not presented.

The "Substantial Equivalence" section (page 5-7) compares features and functions of the proposed device (MediOMx) with the predicate device (MD.ai Viewer (K223425)). This comparison implicitly serves as a form of "performance" assessment in the context of a 510(k) submission, confirming that MediOMx largely matches the predicate's capabilities.

Feature/FunctionProposed Device (MediOMx) PerformancePredicate Device (MD.ai Viewer) PerformanceSubstantially Equivalent
User Install RequirementsThin Client - no install, runs within browserThin Client - no install, runs within browserYes
CommunicationDICOM, non-DICOMDICOM, non-DICOMYes
ModalitiesCT, MR, X-ray, USCR, CT, DX, IVOCT, MR, MG, NM, OCT, OT, PT, RF, SC, US, XAYes
Window Level, Rotate/Pan/Zoom, Reset, Presets, InvertYesYesYes
Multi-study viewing, Image Export, Image Sharing compliantYesYesYes
Metadata Display/HideYesYesYes
Orientation LabelsYesYesYes
Keyboard ShortcutsNoYesYes
Measurements, AnnotationsNoYesYes
Full Screen ModeYesYesYes
Multi-monitor, LayoutsNoYesYes
Linking SeriesNoYesYes
Image Scrolling, Linked Scrolling, Reference LinesYesYesYes
GSPS, KINNoNoYes
Multiplanar Reformat (MPR)YesYesYes
Maximum Intensity Projection (MIP)YesYesYes
Oblique, Volume Rendering, Opacity Presets, Scalpel tool, bone removalNoNoYes
Sharpen, blur, emboss, edge filtersNoYesYes
Histogram Equalization filterNoYesYes

2. Sample size used for the test set and the data provenance
The document mentions "software verification, validation, standalone and clinical performance testing" but does not specify the sample size of any test sets (e.g., number of images, number of cases) or the data provenance (e.g., country of origin, retrospective/prospective nature).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided in the document.

4. Adjudication method for the test set
This information is not provided in the document.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This device is described as a "software-based viewer" that "supports optional integration with FDA-cleared 3rd party AI models." It "receives the AI output and displays outputs of the 3rd party AI model 'as-is' in MediOMx for visualization by the Radiologist to assist in diagnosing the study." It does not appear to be an AI-powered diagnostic algorithm itself, nor does the document describe an MRMC study to measure the effect size of human readers improving with AI assistance from this specific device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The document states "standalone and clinical performance testing" was performed, but it does not elaborate on what "standalone testing" specifically entailed in terms of algorithmic performance metrics or without human-in-the-loop performance. Given its role as a viewer, it's more likely standalone testing would relate to its functionality and display accuracy rather than a diagnostic algorithm's performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
This information is not provided in the document.

8. The sample size for the training set
This device is an image viewer, not an AI algorithm that would typically have a "training set" in the machine learning sense. While software is developed and tested, the concept of a training set as used for AI models is not applicable here.

9. How the ground truth for the training set was established
As above, this is not applicable for an image viewer.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

PMX Inc. % Rory Carrillo Regulatory Consultant Cosm 45 Bartlett St. San Francisco, California 94110

June 17, 202

Re: K233331

Trade/Device Name: MediOMx Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: September 29, 2023 Received: May 13, 2024

Dear Rory Carrillo:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb

Jessica Lamb Assistant Director Imaging Software Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K233331

Device Name MediOMx

Indications for Use (Describe)

MediOMx is a software-based viewer intended to be used with off-the-shelf hardware for the display of DICOM data to aid in diagnosis for healthcare professionals. It performs operations relating to the transfer, storage of image data. MediOMx allows users to perform image manipulations, including window/level, pan, zotation. MediOMx provides 2D display, Multi-planar Reformatting of medical image data. MediOMx is not intended for primary mammography interpretation.

Mobile usage is for reference and referral only.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "PMIX" written with a collection of colorful lines. The lines are mostly blue and green, with some red and yellow. The lines are arranged to form the letters of the word, and they appear to be flowing or moving. The background is white.

510(k) Summary

1. General Information

510(k) SponsorPMX Inc,
Address50 N Brockway St, Suite 3-3Palatine, IL 60067
Correspondence PersonRory A. CarrilloRegulatory ConsultantCosm
Contact InformationEmail: rory@cosmhq.comPhone: 415-580-0916
Date PreparedSeptember 29, 2023

Proposed Device 2.

Proprietary NameMediOMx
Classification NameSystem, Image Processing, Radiological
Regulation Number892.2050
Regulation NameMedical image management and processing system
Product CodeLLZ
Regulatory ClassII

3. Predicate Device

Proprietary NameMD.ai Viewer
Premarket NotificationK223425
Classification NameSystem, Image Processing, Radiological
Regulation Number892.2050
Regulation NameMedical image management and processing system
Product CodeLLZ
Regulatory ClassII

{4}------------------------------------------------

Device Description 4.

MediOMx is a software-based medical image viewer used with web browsers for the 2D visualization of DICOM and non-DICOM medical images. MediOMx is intended for storage, display, manipulation, and processing of radiological data, including images, reports and other clinical information. It has the following primary features and functions:

  • . Zero-footprint HTML5 medical image upload, transfer and display of medical images between facilities
  • Easy access to images for all participants in the healthcare process, including . radiologists, technologists, physicians, nurses and other patient care practitioners
  • Serve as information and data management system for for DICOM and non-DICOM ● medical images
  • Tools for image manipulation (e.g.pan, zoom, rotate, window level)
  • Metadata information and orientation labels display ●
  • . Advanced image manipulation functions like view synchronization across series, 3D visualization like MIP and MPR
  • Encrypted transmission of medical images through secured networks ●
  • Encrypted storage of medical images ●
  • HIPAA-compliant data management, including centralized storage of user activity via ● audit trails.
  • Management of users, roles, and permissions ●
  • It supports optional integration with FDA-cleared 3rd party AI models. It sends the input study to the 3rd party AI model, receives the AI output and displays outputs of the 3rd party AI model "as-is" in MediOMx for visualization by the Radiologist to assist in diagnosing the study. Outputs are displayed in accordance with the 3rd party provider's regulatory clearance. The original image is always accessible

MediOMx consists of a cloud-based application that displays and processes digital medical images, and associated medical information to aid in the day-to-day operations and workflow of clinicians and healthcare practitioners. The web browser based medical image viewer serves as the frontend module which users interact with in viewing data. The backend module handles the connection and processing of data from a variety of sources within the health system, in view of preparing visualizations to be rendered by the viewer. MediOMx can connect and access the medical images across different sources in a health system: an existing PACS or VNA, cloud storage or local server-based storage. Users can also upload images securely into MediOMx which can be shared and enables collaboration with other users. The data connection and imaging data processing is handled by the MediOMx backend module which supports the standardized transmission protocol as defined in the DICOM standard. Users interact with

{5}------------------------------------------------

MediOMx through a standard web browser, thus providing access to full quality images from anywhere and supporting a greater efficiency for care. MediOMx utilizes authorization and authentication mechanisms that enforces authorized users to access the imaging data. The system extends beyond the hospital and its internal network. With a proper network, MediOMx can be accessed by clinical users outside of the hospital network. This way referring physicians can easily call up the imaging data of their patients or external expert accessing the imaging data for additional opinion. MediOMx provides end-users with the ability for industry standard features such as Window/ Level, Image Flip and Rotate. Images are initially displayed in the 2D view mode, but with the ability to toggle into advanced viewing mode of MPR for relevant exam types. It supports processing and displaying Multiplanar Reconstruction (MPR). MediOMx provides an image rendering mechanism that preloads lower resolution images during image scrolling to improve interactivity and performance for users operang in lower network bandwidth while the full quality image is loaded in the background. The use of a secure data transmission protocol and data encryption ensure high data security for data management via the internet. MediOMx tracks user activity via audit trails and stores the audit data on the centralized server.

Indications For Use 5.

MediOMx is a software-based viewer intended to be used with off-the-shelf hardware for the display of DICOM data to aid in diagnosis for healthcare professionals. It performs operations relating to the transfer, storage, and display of image data. MediOMx allows users to perform image manipulations, including window/level, pan, zoom, and rotation. MediOMx provides 2D display, Multi-planar Reformatting of medical image data. MediOMx is not intended for primary mammography interpretation

Mobile usage is for reference and referral only.

Substantial Equivalence 6.

Proposed DeviceMediOMxPredicate Device:MD.ai Viewer (K223425)SubstantiallyEquivalent
Indicationsfor UseMediOMx is a software-basedviewer intended to be used withoff-the-shelf hardware for thedisplay of DICOM data to aid indiagnosis for healthcareprofessionals. It performsoperations relating to the transfer,storage, and display of imagedata. MediOMx allows users toMD.ai Viewer is a software-basedviewer intended to be used withoff-the-shelf hardware for thedisplay of DICOM andnon-DICOM medical images andother healthcare data to aid indiagnosis for healthcareprofessionals. It performsoperations relating to the transfer,storage, display, andYes

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "PHNX" written with colorful lines. The lines are mostly blue, green, and red. The lines are arranged to form the shape of the letters. The letters are all uppercase.

Proposed DevicePredicate Device:Substantially
MediOMxMD.ai Viewer (K223425)Equivalent
perform image manipulations,including window/level, pan,zoom, and rotation. MediOMxprovides 2D display, Multi-planarReformatting of medical imagedata. MediOMx is not intendedfor primary mammographyinterpretationMobile usage is for reference andreferral only.measurement of image data.MD.ai Viewer allows users toperform image manipulations,including window/level, rotation,measurement and markup. MD.aiViewer provides 2D display,Multi-Planar Reformatting and3D visualization of medicalimage data. Mobile usage is forreference and referral only. MD.aiViewer is not intended forprimary mammographyinterpretation.
Feature/FunctionProposed DeviceMediOMxPredicate Device:MD.ai Viewer (K223425)SubstantiallyEquivalent
User InstallRequirementsThin Client - no install, runswithin browserThin Client - no install, runswithin browserYes
CommunicationDICOM, non-DICOMDICOM, non-DICOMYes
ModalitiesCT, MR, X-ray, USCR, CT, DX, IVOCT, MR,MG, NM, OCT, OT, PT, RF,SC, US, XAYes
Window Level,Rotate/Pan/Zoom,Reset, Presets,InvertYesYesYes
Multi-studyviewing, ImageExport, ImageSharing compliantYesYesYes
MetadataDisplay/HideYesYesYes
Orientation LabelsYesYesYes
Keyboard ShortcutsNoYesYes
Feature/FunctionProposed DeviceMediOMxPredicate Device:MD.ai Viewer (K223425)SubstantiallyEquivalent
Measurements,AnnotationsNoYesYes
Full Screen ModeYesYesYes
Multi-monitor,LayoutsNoYesYes
Linking SeriesNoYesYes
Image Scrolling,Linked Scrolling,Reference LinesYesYesYes
GSPS, KINNoNoYes
Multiplanarreformat (MPR)YesYesYes
Maximum IntensityProjection (MIP)YesYesYes
Oblique, VolumeRendering, OpacityPresets, Scalpeltool, bone removalNoNoYes
Sharpen, blur,emboss, edgefiltersNoYesYes
HistogramEqualization filterNoYesYes

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "PHRX" written with a series of colorful lines. The lines are mostly blue and green, with some red and yellow. The lines are arranged to form the letters of the word, and they appear to be flowing or moving. The background is white.

The subject device and predicate device have substantially equivalent indications for use and technological characteristics. The minor differences do not raise questions of safety or effectiveness as the underlying technology is similar and risks associated with these differences are mitigated with similar general and special controls.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the letters "PMX" formed by a dense network of colorful lines. The lines vary in color from blue to green to red, suggesting a gradient or a range of values. The letters are well-defined, with the lines closely packed together to create the shapes.

Performance Data 7.

Safety and performance of the PMX MediOMx device has been evaluated and verified in accordance with software specifications and applicable performance standards through software verification,validation,standalone and clinical performance testing. Additionally, the software validation activities were performed in accordance with IEC 62304:2006/41:2016 Medical device software - Software life cycle processes, in addition to the FDA Guidance documents, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions".

8. Conclusion

Based on the information submitted in this premarket notification, and based on the indications for use, technological characteristics and performance testing, the MediOMx raises no new questions of safety and effectiveness and is substantially equivalent to the predicate device in terms of safety, effectiveness, and performance.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).